towards enhanced amine detection for plasma samples with ... · towards enhanced amine detection...
TRANSCRIPT
Universiteit van Amsterdam
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
Master Research Thesis by
W.F. Duvivier
Master of Science
Analytical Sciences
2010 – 2011
Master Research Thesis by
W.F. Duvivier
Daily Supervisor
Ing. S. Shi
Supervisors
Dr. W.Th. Kok
Dr. R.J. Vreeken
Second Reviewer
Prof. Dr. P.J. Schoenmakers
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier
Table of contents
List of abbreviations 1
Abstract 2
Samenvatting 3
Preface 4
Chapter 1: Introduction 5
1.1 Metabolomics 5
1.2 Metabolites 6
1.3 Analytical approaches in metabolomics 7
1.4 Compound specific analysis: amino acids and amines 9
1.5 Aim of this project 11
Chapter 2: Synthesis of a new derivatization reagent 13
2.1 Introduction 13
2.2 Materials and methods 13
2.3 Results and discussion 15
2.4 Recommendations 18
Chapter 3: Analytical optimisation 20
3.1 Introduction 20
3.2 Materials and methods 20
3.3 Results and discussion 26
3.4 Recommendations and further research 35
Chapter 4: General conclusions 36
Acknowledgements 38
References 39
Attachments 41
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 1
List of abbreviations
ACN Acetonitrile
AMQ 6-aminoquinoline
ANOVA Analysis of Variance
AQC 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate
CSF Cerebrospinal Fluid
DCE Dichloroethane
DCM Dichloromethane
DMF Dimethylformamide
DSC di(N-succinimidyl)carbonate
EDTA Ethylenediaminetetraacetic acid
ESI Electrospray Ionization
FMOC 9-fluorenylmethylchloroformate
FT-MS Fourier Transform-Mass Spectrometry
GC Gas Chromatography
HILIC Hydrophilic Interaction Chromatography
HMDB Human Metabolome Database
HPLC High Performance Liquid Chromatography
IR Infrared
LC Liquid Chromatography
LOD Limit of Detection
LOQ Limit of Quantification
m/z Mass-to-charge ratio
MeOH Methanol
MRM Multiple Reaction Monitoring
MS Mass Spectrometry
NMR Nuclear Magnetic Resonance
OPA Ortho-phthalaldehyde
PITC Phenylisothiocyanate
RP Reversed Phase
RSD Relative Standard Deviation
SRM Selected Reaction Monitoring
SSA Sulfosalicylic Acid
t-BuOK Potassium-tert-butoxide
UPLC Ultra Performance Liquid Chromatography
UV Ultraviolet
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 2
Abstract
To acquire better understanding of diseases and their biomarkers, profiling methods are a
powerful tool to study a certain metabolic compound class. Recently, an amine profiling
method using AccQ·Tag (AQC) derivatization and UPLC-MS/MS is developed and validated
for CSF samples [1].
In the search for a new derivatization reagent with enhanced performance in negative ion-
mode MS-detection, 5-nitro-6-isocyanatoquinoline has been synthesized. This isocyanate
will, theoretically, result in AQC-like derivatized amines with an additional nitro functional
group. The applicability of this reagent has to be tested in future studies.
In addition, the established profiling method [1] has been optimised for application to
various kinds of plasma samples. Therefore, an adjustment has been made to the sample
preparation: the amount of methanol added to the sample for protein precipitation has been
raised to 150 µL. A full validation for mouse plasma has been performed, and two
comparisons have been made to gain more insight in the extendibility of this validation to
other types of plasma samples.
The method is successfully validated for almost 60 compounds in mouse plasma with a LOD
range of 7 ng/mL – 1.2 µg/mL. One important outcome of the validation is an observed day-
to-day batch effect.
The results of the comparisons show that there are no significant differences in the
performance of the method when analysing plasma samples of different species (human,
mouse or rat) or taken with different anti-coagulants (citrate, EDTA or heparin).
Overall can be concluded that the applicability of the amine profiling method is extended
and several leads to enhancing the method by using a new derivatization reagent are
proposed. Especially raising the yield and purity of the synthesized isocyanate requires high
priority for future projects. From an analytical point of view, the anomalies which came up
during the comparisons need more in-depth investigations. Next to this, validation of the
method for rat and human plasma samples has to be completed.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 3
Samenvatting
Profiling methoden zijn een krachtig middel om meer begrip te krijgen van ziekten en hun
signaalstoffen. Met deze methoden is het mogelijk om een bepaalde groep metabolieten te
bestuderen. In de afgelopen jaren is een UPLC-MS/MS methode voor de profiling van aminen
opgezet en gevalideerd voor CSF monsters, hierbij wordt gebruik gemaakt van de AccQ·Tag
(AQC) derivatizering [1].
In de zoektocht naar een nieuw derivatizeringsreagens met verbeterde prestaties in negatieve
ion-mode bij MS-detectie, is 5-nitro-6-isocyanatoquinoline gesynthetiseerd. Dit isocyanaat
zal, theoretisch, resulteren in gederivatiseerde AQC-achtige aminen met een extra nitro-
functionele groep. De toepasbaarheid van dit reagens moet worden getest in toekomstige
studies.
Daarnaast is de gevestigde profiling methode [1] geoptimaliseerd voor toepassing op
verschillende soorten plasmamonsters. Hiervoor is een aanpassing gemaakt aan de
monstervoorbereiding: de toegevoegde hoeveelheid methanol voor denaturatie van het eiwit
in de monsters is verhoogd naar 150 µL. Een volledige validatie voor muisplasma is
uitgevoerd, en twee vergelijkingen zijn gemaakt om meer inzicht te verkrijgen in de
toepasbaarheid van deze validatie op andere plasmasoorten.
De methode is succesvol gevalideerd voor bijna 60 amine verbindingen in muisplasma met
een detectielimiet van 7 ng/ml tot 1.2 µg/ml. Een belangrijke uitkomst van de validatie is het
waargenomen dag-tot-dag batch effect.
Uit de resultaten van de vergelijkingen blijkt dat er geen significante verschillen zijn wat
betreft de werking van de methode bij de analyse van plasmamonsters afkomstig van
verschillende diersoorten (mens, muis of rat) of die zijn genomen met verschillende
antistollingsmiddelen (citraat, EDTA of heparine).
In het algemeen kan geconcludeerd worden dat de toepasbaarheid van de amine profiling
methode is vergroot en een aantal aanknopingspunten voor de synthese van een nieuw
derivatizeringsreagens zijn beschreven. Vooral het verhogen van de opbrengst en zuiverheid
van het gesynthetiseerde isocyanaat vereist hoge prioriteit voor toekomstige projecten.
Vanuit een analytisch oogpunt vereisen de afwijkingen die naar voren kwamen uit de
vergelijkingen meer onderzoek. Verder is volledige validatie van de methode voor rat en
humaan plasma gewenst.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 4
Preface
The main goal of this master research project is to enhance amine profiling for metabolomic
studies. To acquire better understanding of diseases and their biomarkers, profiling methods
are a powerful tool to study a certain metabolic compound class. In the last few years, an
amine profiling method is compiled and validated for CSF samples [1]. This method is the
starting point of this thesis. Evaluation of the derivatization and analytical optimisation of the
method should provide us with more insight in possibilities to enhance the sensitivity and
applicability to other biological samples types.
Chapter 1 of this thesis is an introduction into metabolomics, analytical approaches and
amino acid analysis. Also, the objectives of this project are described in this chapter. The
reported study exists of two parts: an organic synthesis and an analytical part. The aim of the
first part, described in chapter 2, is to develop a new derivatization reagent, resulting in
derivatized amines with better characteristics for detection by mass spectrometry in negative
ion-mode. Chapter 3 describes the applicability of the amine profiling method to plasma
samples of different species. General conclusions are stated in the last chapter, chapter 4.
All the research described in the thesis is carried out at Leiden University in the Analytical
BioSciences group under supervision of ing. S. Shi and dr. R.J. Vreeken.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 5
Chapter 1: Introduction
1.1 Metabolomics
Metabolomics, sometimes also referred to as metabonomics, studies small-molecule
metabolite profiles and provides us with information about the metabolic state of cellular
systems in different conditions (e.g. healthy and diseased) [2]. Metabolite profiles reflect the
combined effects of many influences, such as drugs, environment and nutrition, and result
into a better understanding of cellular biochemistry. Metabolomics is one of the newest
“omics” and is applicable for medical, diagnostic, food and industrial microbiology among
others [3].
Figure 1.1: Metabolomics as part of functional genomics [4].
In the search for a better view on the role of different biochemical processes and related
diseases, metabolomics is one of the key techniques. By studying the chemical profiles of
these processes, metabolic pathways and their workings can be unscrambled, hopefully
leading to better understanding of diseases and their causes. Especially the search for
biomarkers, which could contribute to the discovery of a certain disease in an early state, has
been a hot topic in metabolomics over the last few years.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 6
Figure 1.2: Alanine, aspartate and glutamate metabolism in humans [5].
1.2 Metabolites
The metabolites which are studied in metabolomics belong to a wide range of different
compounds classes (figure 1.3). This study focuses on biogenic primary and secondary
amines, among which amino acids.
← More non-polar More polar →
Carotenoids
Steroids
Phenolics
Alcohols
Alkaloids
Organic acids
Organic amines
Sugars
Nucleotides
Flavenoids
Catecholamines
Polar organics
Nucleosides
Amino acids
Metals
Ionic compounds
Figure 1.3: Rough overview of metabolomic compound classes [6].
Amines are organic compounds which contain a basic nitrogen atom with a lone electron
pair. They can be distinguished in primary, secondary and tertiary amines based on the
number of substituents (respectively one, two and three) on the nitrogen atom. Amino acids
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 7
are vital as building blocks of proteins and have many functions in metabolism. Each amino
acid contains an amino group, a carboxylic functionality and a so-called “R” side chain, which
differs for each amino acid (e.g. –CH3 for Alanine). Humans must include 9 of the 20 standard
amino acids in their diet; these amino acids can not be metabolized or synthesized from
other compounds.
Amines play an important role as signal compounds and parts of metabolomic pathways
which are influenced by diseases such as Parkinson and other central nervous system
diseases. They may have multiple roles in mechanisms that among others can affect body
temperature, psychomotor functioning and pain. In the case of Parkinson, disorders of amine
metabolisms comprise a wide spectrum of symptoms, with motor dysfunction being the most
prominent clinical feature [7].
This project focuses on the analysis of amines in different matrices, mostly blood plasma and
cerebrospinal fluid (CSF). Blood plasma is the whole blood minus the blood cells and contains
dissolved proteins, glucose, clotting factors, mineral ions, hormones and carbon dioxide. On
the other hand, CSF is a clear body fluid that is taken from around the spinal cord. CSF
contains less plasma proteins and is produced at a rate of 500 ml/day. The concentration of
most biogenic amines is lower in CSF than in blood plasma [8].
1.3 Analytical approaches in metabolomics
In most cases, sample preparation depends much on the targeted group of metabolic
compounds. The extraction type and procedure, for example, are matched with the chemical
properties of the investigated compound group [4]. Due to this focusing, metabolomics
might seem to be a little less challenging from an analytical point of view. Nevertheless, the
high chemical complexity within a specific group of compounds, the wide dynamic range and
the large biological variance make metabolomics an analytical challenging exercise.
Most analytical methods in metabolomics are based on separation techniques like GC (gas
chromatography) and HPLC (high performance liquid chromatography) or UPLC (ultra
performance liquid chromatography) coupled to different mass spectrometry (MS)
techniques or nuclear magnetic resonance (NMR) spectroscopy. Generally, mass
spectrometers are more selective and more sensitive compared to other types of detectors.
However, as stated by Zhang et al. [9], NMR is in some cases very suitable for metabolomic
research: “NMR might be less sensitive, but the data from NMR experiments is often more
easily quantitated and highly reproducible. In particular, the same nuclei detected (i.e. all 1H)
in an NMR experiment have the same sensitivity, independent of the properties of metabolite
molecules. Therefore, the absolute quantities of different metabolites can be measured with a
single internal or external standard. In addition, NMR requires minimal or no sample
preparation or separation, and is non-destructive.” This research project focuses though on
the use of mass spectrometry based methods.
Before compounds can be detected using MS, they have to be ionized and, preferably,
separated. Ionization techniques may vary, especially for GC-MS and LC-MS couplings. Also
the separation mechanisms in LC are various: hydrophilic and small, charged molecules are
best suited for capillary electrophoresis [10], reversed-phase chromatography can be used for
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 8
hydrophobic compounds [11] and hydrophilic and neutral compounds are well separated by
hydrophilic interaction chromatography (HILIC) [3][12].
Metabolic MS-analysis can be roughly divided into two approaches: targeted and non-
targeted. The aim in non-targeted methods is to cover the metabolome as broadly as
possible while at the same time maintaining the ability to at least differentially quantify the
metabolites. This way, it is possible to detect changes in metabolomic profiles, but also allows
detection of previously unknown or poorly characterized metabolites [13].
The aim in targeted methods is to quantitatively analyse a specific, biologically relevant,
metabolite class. More and more common are methods which cover over 100 metabolites of
a specific category. In targeted strategies predefined metabolite-specific signals are often
used to precisely and accurate determine relative abundances and concentrations of a limited
number of pre-known and expected endogenous metabolites [14].
A common used MS/MS (or MS2) detection mode for targeted metabolomics is selective
reaction monitoring (SRM). As shown in figure 1.4a, ions with a selected m/z value are
isolated into the collision cell. After collision induced dissociation, only the fragments with a
certain, forehands chosen, m/z value are lead into the detector. No time is wasted on
acquiring non-relevant data, which leads to maximum sensitivity [15].
Figure 1.4: Selective reaction monitoring (a) and precursor-ion scan (b) in MS/MS [15].
When using chromatographic separation, SRMs can be time-programmed to avoid
measuring irrelevant m/z values during analyte elution and thereby maximizing the
sensitivity. This technique is known as multiple reaction monitoring (MRM) [14][15].
While studying a metabolite class for which all the different metabolites fragment into a
specific fragment, it is possible to use a precursor-ion scan to obtain an overview of the most
abundant compounds. In a precursor-ion scan, MS2 can be set to a m/z value characteristic
to the studied metabolite class, while MS1-detection is scanned. This way, only ions which
fragment into the predefined fragment will be detected (figure 1.4b). [14] An example of the
use of a precursor-ion scan is described by Rochfort et al. [16]. This study shows a wide range
of aliphatic, alkenyl and indole glucosinolates which all fragment into a fragment of m/z 259
(figure 1.5).
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 9
Figure 1.5: Breakdown of glucosinolates to produce the characteristic m/z 259 ion [16].
1.4 Compound specific analysis: amino acids and amines
For years, the reference method for quantitative amino acid analysis was ion-exchange
chromatography followed by post-column ninhydrin derivatization [17][18]. Due to the need
for analysis of lower concentrations, methods based on reversed-phase (RP) liquid
chromatography have been developed over the last decades. The combination of pre-column
derivatization and RP-LC results not only in faster analysis (<60 min), but also in a rise of
sensitivity.
Ortho-phthalaldehyde (OPA) [19], phenylisothiocyanate (PITC) [20] and 9-
fluorenylmethylchloroformate (FMOC) [21] are among the most frequently used
derivatization reagents for amino acid analysis, but have various disadvantages. OPA, PITC
and FMOC respectively fail to react with secondary amino acids [22], bring about the need to
remove excess reagent by drying [20] and yield multiple derivatives [21].
Cohen et al. [23] demonstrated the use of 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate
(AQC) as an amine derivatization reagent. This reagent reacts with amino acids in a rapid one
step procedure (figure 1.6) to form stable urea derivatives which can be analyzed by RP-LC
with fluorescence detection. The derivatization shows excellent results concerning
derivatization efficiency and tolerance of common buffer salts and detergents. Supported by
the good chromatographic characteristics of AQC-derivates, this derivatization has been used
on a wide range of products [24][25][26].
Figure 1.6: One step procedure for amino acid derivatization using AQC [23].
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 10
Boogers et al. [27] described a changeover from the Pico·Tag HPLC method to the
AccQ·Tagultra UPLC method for amino acid analysis in protein hydrolysates. The Pico·Tag and
AccQ·Tag methods are commercially available derivatization kits, which, respectively, use PITC
and AQC for the derivatization of primary and secondary amino acids. Boogers et al. found
that the total chromatographic run time of the AccQ·Tagultra UPLC method was only 40% of
the time required for the Pico·Tag HPLC method, while the quantitative results for both
methods compared fairly well.
Another (commercially developed) method for amino acid analysis is EZ:faast (Phenomenex).
It involves solid phase extraction (SPE) and liquid/liquid extraction for sample clean-up, and
subsequent reaction of the amino acids with ethylchloroformate to produce N-
ethoxycarbonyl ethylesters. These derivatives can be separated and detected using GC(-MS)
or LC-MS [28].
Recently, hydrophilic interaction chromatography (HILIC) is an innovative separation
technique and can also be used for amino acid analysis. HILIC is developed for the separation
of hydrophilic and neutral compounds among which amino acids and amines [29]. However,
HILIC is still an upcoming technique and lacks, for the moment, good reproducibility.
The last few years, mass spectrometry has been more and more used as a detection method
for AQC-derivatized amino acids [30][31]. This development is especially promising for
selective measurement of compounds at low concentrations and in complex matrices. In
theory, it is unnecessary to, baseline, separate the compounds prior to MS detection,
although this has some bottlenecks concerning co-eluting compounds (which may cause ion
suppression) or compounds with the same m/z value. Good chromatography can, almost
completely, solve these difficulties and enhance the sensitivity of the method [31].
The use of precursor-ion scans and full scan analysis of a concentrated CSF sample on a FT-
MS, in combination with information in the HMDB (Human Metabolome Database), made it
possible to expand the AQC-method from amino acid analysis to the detection and
quantification of about 60 amine containing compounds. This expanded method is validated
for CSF samples [1].
Despite of the advantages mentioned above, the AQC derivatization method is not perfect.
One of its drawbacks, is good performance in negative ion-mode. Therefore, a compound
requires negative charge stabilisation after ionization.
Measuring in negative ion-mode can enhance the sensitivity of a method. Almost all neutral
substances are able to yield positive ions, while acidic groups or electronegative elements are
needed to produce negative ions. [32] Due to the lower chemical background signal, a
compound with good performance in negative ion-mode will have a higher signal-to-noise
ratio compared with positive ion-mode. Theoretically, the decreased noise will result in a
lower detection limit for compounds with good performance in negative ion-mode [33].
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 11
1.5 Aim of this project
The aim of this project is to develop an enhanced amine profiling method, which will be used
for the detection and identification of amines in biological samples such as CSF and blood
plasma. To achieve this, a two-sided approach has been chosen: a new, improved,
derivatization reagent will be synthesized and in addition, the existing amine profiling
method [1] for CSF samples will be evaluated and validated for application to various plasma
types.
1.5.1 Organic synthesis of a new derivatization reagent
Until now, the amine profiling method has been used with positive ion-mode mass
spectrometry. After AQC derivatization and electrospray ionisation, positive ions (mostly
[M+H]+ ions) are more stable and formed more efficiently then negative ions (mostly [M-H]-).
In this project, the derivatization reagent will be modified to enhance the use of negative ion
mode. In theory, this should improve the sensitivity of the method. Fewer molecules are
ionized in the negative mode, so considerably less chemical background is observed,
providing an increased linear range and lower limits of detection [33].
AQC is easily synthesized by the addition of 6-aminoquinoline (AMQ, in dry acetonitrile), to
di(N-succinimidyl)carbonate (DSC, in dry acetonitrile) which is heated to reflux (figure 1.7).
After about 30 minutes of continued refluxing, the reaction mixture can be concentrated by
rotary evaporation and cooled for 24 hours, resulting in AQC-crystals [23].
Figure 1.7: Synthesis of AQC [23].
AQC as derivatization reagent has many advantages; the derivative has good
chromatographic characteristics and the reaction is a fast, one-step procedure. Furthermore,
it reacts instantaneous with primary and secondary amines, which makes it very suitable as
derivatization reagent for the method used in this project.
Other established methods for amino acid and amine analysis, such as ninhydrine and dansyl
derivatization [34][35] or EZ:faast [36], could be considered as well. However, because of the
previous described benefits, AQC is used as starting position for the synthesis of a new
derivatization reagent suitable for LC-MS in negative ion mode.
1.5.2 Analytical optimisation of the amine profiling method
The analytical part of this project focuses on implementation of the extended method to be
used with plasma samples and afterwards, validation of the method for this sample type.
Plasma samples are more easily obtained and available in bigger amounts, which make them
more suitable for large screening projects than CSF samples.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 12
Validation of the method is important to ensure that the method is robust related to the limit
of detection and quantification (LOD and LOQ), linearity and repeatability. Therefore, a
validation protocol developed within the Analytical BioSciences group of Leiden University is
used, which evaluates the performance of the used method. Since plasma samples can be
taken with support of different anti-coagulants, a possible effect of these agents on the
performance of the method has to be investigated as well.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 13
Chapter 2: Synthesis of a new derivatization reagent
2.1 Introduction
As stated in chapter 1, the AQC-derivatization method suits very well for amino acid analysis,
both with UV- and MS-detection. AQC-derivatives are best detected using positive ion-mode;
the lack of negative charge stabilisation in the derivatives causes low sensitivity in negative
ion-mode. The amine profiling method used nowadays is not sensitive enough for the
quantification of compounds of low endogenous levels. Compared to positive ion-mode, the
signal-to-noise ratio in negative ion-mode of a compound with good characteristics for
negative ionization is higher due to the lower background signal.
Derivatives with a good performance in negative ion-mode, in combination with the benefit
of negative ion-mode mentioned above, could make the method more sensitive. As stated in
the introduction, the AQC-reagent is chosen as starting point as a result of its advantages
and experience gained before.
Better performance in negative ion mode requires negative charge stabilisation in the
derivative. This can be achieved by incorporating a nitro-group in the reagent and
subsequently in the derivative. The best possibility to realize this, is the modification of
6-aminoquinoline (AMQ) before reacting with DSC (figure 2.1). The synthesis of an
aminoquinoline with an ortho-placed nitro-group (6-amino-5-nitroquinoline) was found in
literature and therefore chosen as starting point [37].
6-amino-5-nitroquinoline
N
NH2
NOO
N
O
O
O
N
O
O
O
O
+
DSC
N
NH
NOO
O
ON
O
O
+ N
O
O
OH
New derivatizing reagent
NHS Figure 2.1: Theoretical reaction scheme of 6-amino-5-nitroquinoline with DSC (based on [23]).
2.2 Materials and methods
All reactions were monitored using thin layer chromatography (TLC) and various colouring
reagents (e.g. ninhydrin and potassium permanganate).
2.2.1 Synthesis of N-tetramethylethiocarbamoylsulfenamide
4 mL ammonia solution (28-30 wt%) was injected in a 50 mL round-bottom flask which was
standing in ice mixed with salt (NaCl; -15°C) to cool the reaction . Next sodium hypochlorite
was diluted with water (1:1) to achieve a concentration of 7% sodium hypochlorite (1N). 16.5
mL of this solution was added dropwise to the ammonia, slowly enough (± 30 minutes) to
maintain the temperature below 0°C.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 14
Ammonium N,N-tetramethylenedithiocarbamate solution (2.5 g ; 17.1 mmol in 15 mL H2O)
was added dropwise to the flask in about 15 minutes. The mixture was stirred vigorously for
30 minutes after which about 1 gram of pure sulfenamide was filtered off [38].
NS
SH
N+
H
HH
H
NS
S NH2
N-tetramethylethiocarbamoylsulfenamideAmmonia
N,N-tetramethylenedithiocarbamate
+NH4OH
NaOCl
Figure 2.2: Synthesis of N-tetramethylethiocarbamoylsulfenamide [38].
The purity of the sulfenamide was checked by 1H and 13C NMR (see attachment 1 and 2).
2.2.2 Synthesis of 6-amino-5-nitroquinoline
1.12 g (1.00 mmol) potassium-tert-butoxide (t-BuOK) was dissolved in 6 mL
dimethylformamide (DMF) under argon. A solution of 0.66 g (3.80 mmol) 5-nitroquinoline
and 0.74 g (4.56 mmol) sulfenamide (N-tetramethylethiocarbamoyl-sulfenamide, synthesis
described in paragraph 2.2.1) in 2 mL DMF was added dropwise to the t-BuOK solution over a
period 5 minutes. After 50 minutes of continued stirring, the mixture was poured into 100 mL
cold water. Subsequent to extraction with three times 60 mL dichloroethane (DCE), the DCE-
fraction was washed with 200 mL water and dried with Na2SO4 [37].
+N
NO O
N
NO O
NH2
6-amino-5-nitroquinoline
NS
S NH2
5-nitroquinoline
t-BuOK
carbamoylsulfenamideN-tetramethylethio-
Figure 2.3: Synthesis of 6-amino-5-nitroquinoline [37].
After filtration, the filtrate is chromatographically purified and characterized using 1H NMR
and TLC-MS (see attachment 3 and 4). 0.21 g (1.11 mmol) 6-amino-5-nitroquinoline was
yielded (29% of crude product).
2.2.3 Synthesis of an AQC-based derivatization reagent
0.15 g (0.60 mmol) of DSC (di(N-succinimidyl) carbonate) was dissolved in 10 mL dry
acetonitrile and heated to reflux. 6-amino-5-nitroquinoline (0.1 mg ; 0.50 mmol), dissolved in
8 mL of dry acetonitrile, was added dropwise to the refluxing carbonate solution in 30
minutes. After an additional 30 minutes of reflux the reaction mixture was concentrated by
rotary evaporation to about half its volume. The solution was cooled for 24 hours and the
resulting crystals were filtered.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 15
2.2.4 Synthesis proposal for 5-nitro-6-isocyanatoquinoline
0.19 g (1.00 mmol) 6-amino-5-nitroquinoline was dissolved in 10 mL dry dichloromethane
(DCM) at 0°C under argon. 0.6 mL (1.14 mmol) phosgene (20% in toluene) was added
dropwise. The solution was stirred for 15 minutes, and NEt3 (0.4 mL; 2.9 mmol) was added.
After an additional 15 minutes of stirring, the solution was concentrated using rotary
evaporation [39].
N
NO O
NH2
+O
Cl Cl N
NO O
N
O + ClH
ClH
N
NO O
N
O+ Cl
-
6-amino-5-nitroquinoline Phosgene
5-nitro-6-isocyantoquinoline
5-nitro-6-isocyantoquinoline
Et3H+-HNEt3
Figure 2.4: Reaction scheme of 6-amino-5-nitroquinoline with phosgene [39].
2.2.5 Modified synthesis proposal for 5-nitro-6-isocyanatoquinoline
0.050 g (0.26 mmol) 6-amino-5-nitroquinoline (dissolved in 10 mL DCM, under argon) was
added dropwise to 0.7 mL (1.33 mmol) phosgene (20% in toluene, under argon at 0°C). A
yellow, turbid solution is formed, which becomes clear and more orange after the addition of
three equivalents of NEt3 (0.15 mL ; 1.09 mmol). Also, a white gas is formed. After rotary
evaporation, the residue was analyzed again.
2.3 Results and discussion
The main goal of the first part of this project was to develop a new derivatization reagent for
amines which has good properties for negative ion mode. As noted in the introduction of this
chapter, the chosen route to achieve this goal was to modify the AQC reagent by adding a
nitro-group to the part of the reagent that will be attached to the target compound.
The first step involved the synthesis of N-tetramethylethiocarbamoylsulfenamide, which was
necessary to enable synthesis of the desired quinoline: 6-amino-5-nitroquinoline. Both the
sulfenamide as the amino-nitroquinoline were obtained with only minor contaminations as
shown by the NMR spectra in the attachments.
2.3.1 Synthesis of new, AQC-based, reagent
The reaction of 6-amino-5-nitroquinoline with DSC, as described above, did not result in the
desired product; a new AQC-based reagent. The formed crystals did not contain the product
aimed for, but consisted of pure 6-amino-5-nitroquinoline and DSC (the starting products).
Several new experiments with adjustments to the protocol were performed (e.g. longer
refluxing, etc.), but without any improvements concerning the resulting compounds.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 16
There are a couple of possibilities, or combinations of these possibilities, why this synthesis
protocol did not result in the compounds aimed for. Theoretically, the amine-group of the
synthesized amino-nitroquinoline is less nucleophile as the amine-group of the original
quinoline, 6-aminoquinoline, which is not very nucleophile to begin with. The nitro-group
pulls electrons away from the amine-group, causing lower reactivity.
Another possibility is stabilization due to hydrogen bonding. By tautomeric structures, a
hydrogen bond could be formed between the nitro- and the amine-group. Besides that,
there could be a steric effect from the nitro-group, which is positioned next to the amine-
group.
It is not likely that a different modification of the quinoline would have given better results.
When the nitro-group is situated on the other side of the amine-group, but still ortho-placed,
it would have the same disadvantages as the used quinoline.
Meta-placed nitro-groups (figure 2.5a) are not easy, or not even possible, to synthesis. A
para-placed group would theoretical give a more reactive compound, but changes the
reaction mechanism by dislocating the amine-group (figure 2.5b).
N
NOO
NH2N
NH2
NO O
N
NOO
NH2
A B
Figure 2.5: Possible amino-nitroquinolines with a; meta-placed groups and b; para-placed groups
2.3.2 Second synthesis proposal
After the disappointing results of the first synthesis procedure, several different options were
considered. Instead of continuing with trying to synthesis an AQC-like reagent, another
pathway was chosen, in which the amine-group of the amino-nitroquinoline was replaced by
an isocyanate (figure 2.4). This isocyanate can theoretically react with amines, as shown in
figure 2.6, resulting in the same derivatized amine as previously aimed for.
N
NO O
N
O+ NH
R1
R2
N
NO O
NH
O
N
R1
R2
5-nitro-6-isocyanatoquinolineAmine or amino acid
Derivatized amine Figure 2.6: Theoretical reaction scheme of an isocyanate with an amine or amino acid
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 17
Direct-infusion mass spectrometry analysis of the reaction mixture showed a spectrum with
high abundant m/z-values of 102, 190, 203, 239 and 405 (figure 2.7). Interpretation of these
m/z-values implies that 6-amino-5-nitroquinoline was still present in the reaction mixture
(m/z 190) and dimerization of the amino-nitroquinoline with the isocyanate takes place (m/z
405). This implies that the isocyanate is formed, but reacts with the surplus of amino-
nitroquinoline. M/z 102, 203 and 239 are most likely by-products of the synthesis.
Figure 2.7: Mass spectrum after direct infusion of reaction mixture with a possible dimmer structure
A new synthesis protocol was composed to achieve better results (paragraph 2.2.5). In this
protocol, some changes were made to avoid dimerization: 6-amino-5-nitroquinoline was
dissolved in a larger volume and this time added dropwise to the phosgene to prevent the
formed isocyanate to react with a surplus of amino-nitroquinoline.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 18
Figure 2.8: Mass spectrum after direct infusion of reaction mixture
The resulting reaction mixture was again analyzed by direct-infusion MS. The results of this
analysis (figure 2.8) show that the desired product is actually formed, indicated by an m/z-
value of 216. Also m/z-values 190, 239 and 405 are still relatively high abundant. Next to that,
a lot of other, lower, m/z-values are present, which most likely are by-products.
Different quick, tests were done to examine a possible derivatization reaction between the
synthesized isocyanate and several amino acids. The results did not show new products to be
formed, but the yield of the synthesis was too small to achieve significant and thorough
testing.
2.4 Recommendations
A few recommendations regarding the synthesis are already mentioned in the discussion
section. The taken synthesis pathway was chosen based on the complexity of the synthesis
steps and the availability of the required chemicals. There are more possibilities to synthesis
the same compounds, which could be investigated in future studies to study the yield and
purity. The use of a catalyst could be considered to achieve this goal.
Adding another functional group to the reagent should be considered as well. Theoretically,
good results could be expected when adding a carboxylic acid or Fluor group.
Also, looking into other possible derivatization reagents could pay off. Next to AQC, EZ:faast
is another frequently used method for amino acid analysis which includes sample clean up by
SPE and liquid/liquid extraction [36].
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 19
2.4.1 Synthesis of isocyanate
With more time, experience with and knowledge of organic chemistry, it should be possible
to optimize the synthesis of the isocyanate. After modification of the reaction conditions, the
difficulty of the dimerization of the isocyanate with 6-amino-5-nitroquinoline was partly
solved. Another, but harder, option is to cool the reaction with liquid nitrogen or an acetone
bath. This is a more time and work consuming option, but could be investigated to gain yield
and purity. With a higher yield and purity, more and better testing could be done regarding
its derivatization properties.
In the synthesis protocol used in this project, phosgene is used for the formation of an
isocyanate. The toxicity of phosgene could be a drawback for the production of larger
amounts of the reagent. There are other synthesis pathways that lead to the sought
isocyanate, one of them is described below:
The first step in this pathway is the conversion of the amino-group of 6-amino-5-
nitroquinoline into an urethane-group. This can be achieved by the addition of triethylamine
and methylchloroformate, after which the reaction mixture is extracted, washed, dried and
filtered as described by Wilson et al. [40]. The urethane-group synthesized compound can be
converted into an isocyanate by the addition of chlorocatecholborane and refluxing [41].
N
NO O
NH2
N
NOO
NH
O
OCH3
N
NO O
N
O
6-amino-5-nitroquinoline 5-nitro-6-urethanequinoline
5-nitro-6-isocyanatoquinoline Figure 2.9: Alternative synthesis pathway for 5-nitro-6-isocyanatoquinoline [40][41].
2.4.2 Reagent testing
Due to a shortage of time, it was not possible to test the product carefully. The tests that
have been done did not show any results, but more testing need to be done to get to a more
considered conclusion.
To gain more reagent, more 5-nitro-6-isocyananatoquinoline could be synthesized (possibly
in several batches) and chromatographically purified. This way more and more accurate
testing is possible.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 20
Chapter 3: Analytical optimisation
3.1 Introduction
This part of this research project focuses on the further development and optimisation of the
amine profiling method described by Shi et al. [1]. With this UPLC-MS/MS method based on
Waters’ AccQ·Tag method, around 60 amine containing compounds can be detected and
quantified using the benefits of UPLC and multiple reaction monitoring (MRM). UPLC results
in a higher chromatographic resolution and enables sufficient separation of the compounds
within a short run time. Besides that, MRM-mode provides us with a adequate sensitivity.
This expanded method is validated for CSF samples; validation for plasma samples of
different species is an important objective to achieve. This project was focused at human,
mouse and rat plasma samples, based on the need out of the field. Many studies are
performed on test animals before application on humans. Mice and rats are most frequently
used for these kind of studies, causing the need for an amine profiling method to be
applicable for these matrices.
Subsequently, the effect of different anti-coagulants on the method will be evaluated. Plasma
samples can be taken with support of different anti-coagulants, which may affect the
performance of the method. To ensure a wide application range, this method should be
applicable to plasma samples, irrespective of the used anti-coagulant.
3.2 Materials and methods
3.2.1 Chemicals
The acetonitrile (ACN), methanol (MeOH) and water (H2O) used for these experiments are all
Ultra LC-MS grade from Biosolve (Valkenswaard, The Netherlands). AccQ·Tag Ultra eluent A
and B concentrate and the AccQ·Tag Ultra derivatization kit are from Waters (Etten-Leur, The
Netherlands).
3.2.2 Standard mixes for validation and protein precipitation test
For the validation of the method and the protein precipitation test, a set of standard mixtures
is used. The composition of these mixes can be found in attachment 5.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 21
3.2.3 Internal standards
The following labelled compounds are used as internal standards in the experiments
described below.
Labelled “Cell free” amino acid mix (U-13C, 98%; U-15N, 98%) from Cambridge Isotope
Laboratories, Inc. (Andover, USA):
Table 3.1: Content of labelled amino acid mix
Amino Acid Molar % Weight %
Aspartic Acid 5.8 6.0
Glutamic Acid 8.1 9.3
Asparagine 5.7 5.9
Serine 3.0 2.5
Glutamine 3.6 4.1
Histidine 0.6 0.7
Glycine 7.9 4.6
Threonine 5.5 5.1
Alanine 10.7 7.5
Arginine 2.8 3.9
Tyrosine 2.5 3.6
Alanine 6.8 6.2
Methionine 1.5 1.8
Tryptophan 2.5 4.0
Phenylalanine 5.5 7.1
Isoleucine 5.2 5.4
Leucine 9.1 9.3
Lysine 1.8 2.1
Proline 2.1 1.9
Cysteine 9.6 9.1
Deuterated compounds:
- β-alanine-2,2,3,3-D4
- (±)-Epinephrine-D3
- Histamine-α,α,β,β-D4-2HCl
- L-2-Aminobutyric-D6 Acid
- L-NT-methyl-D3-l-hisditine
- L-3-(4-Hydroxy-3-methoxy-D3-phenyl)alanine
- 2-(4-Hydroxy-3-methoxyphenyl)ethyl-1,1,2,2-D4-amine HCl
- (±)-Norepinephrine-2,5,6,α,β,β-D6 HCl, L-Ornithine-3,3,4,4,5,5-D6 HCl
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 22
3.2.4 Standard solutions
The above mentioned (internal) standards were dissolved in Ultra LC-MS-grade water in the
concentrations noted in table 3.2. From these stock solutions, the other concentration levels
are prepared by dilution factors of 2 between each concentration level.
Table 3.2: (Internal) Standard stock solutions
Mix Stock Solution
1 1 mg/mL
2 0.2 mg/mL
3 0.01 mg/mL
4 0.4 mg/mL
Labelled amino acids 0.1 mg/mL
Deuterated compounds 1 mg/mL*
*Already dissolved (1 mg/mL): 20 µL of each compound, 4820 µL water added (5 mL in total)
3.2.5 Plasma samples
For these experiments, pooled human (female), rat and mouse plasma samples were used,
taken from healthy populations (with heparin as anti-coagulant).
The samples used for the anti-coagulant comparison were taken from one person (female) at
one moment with different anti-coagulants: EDTA, heparin and citrate.
3.2.6 Sample pre-treatment
For biological fluids a simple protein precipitation was applied as sample pre-treatment. 10
µL of each (internal) standard dilution was added to 5 µL biological sample followed by the
addition of 150 µL* of MeOH for protein precipitation. The mixture was vortexed for 10s and
centrifuged at 10.000 rpm (9408 g) for 10 minutes at 10 °C. The supernatant was pipetted
into a sample vail and dried under N2.
The residue was dissolved in 80 µL borate buffer (Waters AccQ·Tag Ultra derivatization kit, pH
8.8), after 10s vortexing, 20 µL of AccQ·Tag reagent (dissolved in 500 µL AccQ·Tag Ultra
Reagent Diluent) was added and the mixture was vortexed immediately. This step was carried
out one by one.
The sample was heated 10 minutes at 55 °C. The heating converts a minor side product of
tyrosine to a major mono-derivatized compound.
1 µL of the derivatized mixture was injected into the UPLC-MS/MS system.
* Amount of MeOH added varied for some experiments
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 23
3.2.7 LC method
An Acquity UPLC system with autosampler (Waters) in combination with a Xevo TQ Tandem
Quadrupole mass spectrometer (Waters) was used for the experiments described below. The
column used was a AccQ·Tag Ultra 2.1x100mm (1.7 µm particle size) column (Waters), used
with mobile phase A (AccQ·Tag eluent A concentrate diluted 10 times with water) and mobile
phase B (AccQ·Tag eluent B). The LC-gradient used is shown in table 3.3.
Table 3.3: LC-gradient used for separation of the targeted compounds
Time Flow (mL/min) %A %B Curve
1 Initial 0.700 99.9 0.1 Initial
2 0.54 0.700 99.9 0.1 6
3 5.74 0.700 90.9 9.1 7
4 7.74 0.700 78.8 21.2 6
5 8.04 0.700 40.4 59.6 6
6 8.05 0.700 10.0 90.0 6
7 8.64 0.700 10.0 90.0 6
8 8.73 0.700 99.9 0.1 6
9 9.50 0.700 99.9 0.1 6
3.2.8 Mass spectrometry
MS/MS detection of the targeted amines was achieved by using selective reaction monitoring
(SRM) for each of the compounds. The SRM parameters were time-programmed to avoid
measuring irrelevant m/z values during analyte elution and thereby maximizing the
sensitivity. This technique is also known as multiple reaction monitoring (MRM). The used
SRM transitions can be found in attachment 6.
The MS/MS source and analyzer parameters used for the analysis of the derivatized amines
and internal standard are shown in attachment 7.
3.2.9 Calculations
The used calculations are described in attachment 8.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 24
3.2.10 Experimental design
3.2.10.1 Method validation
Validation of the method is performed using a 3-day procedure which is shown in table 3.4.
To determine the recovery, plasma-containing samples are spiked after sample preparation.
The same goes for ion suppression samples, which are academic samples with no sample
pre-treatment. To calculate the recovery, ion suppression and matrix effect, the labelled
standards are used as calibrants and 4 labelled compounds (asparagine, glutamine, lysine
and valine) are used as internal standards. This way, the endogenous level of the compounds
is not interfering with the results of the measurement.
Academic samples did not contain plasma. The plasma added to FreezThaw+1 and +2
samples is thawed and frozen, respectively, one and two times more compared to the plasma
used for the other samples. Every sample is prepared in triplo and injected three times.
Table 3.4: Experimental design of method validation (total number of samples = 96)
Cal Academic Recovery Ion Suppr. Cal Recovery Cal Recovery FreezThaw+1 FreezThaw+2 C0 3 3 C1 3 3 C2 3 3 C3 3 3 3 3 3 3 3 3 C4 3 3 C5 3 3 3 3 3 3 3 3 3 3 C6 3 3 C7 3 3 3 3 3 3 3 3 C8 3 3 27 27 9 9 9 3 9 3 9 9
All samples consisted of 10 µL of each standard mix dilution (4 in total), 10 µL of both labelled internal
standard dilutions and 5 µL plasma. Except for the academic samples, those did not contain plasma.
Every sample is injected three times.
3.2.10.2 Protein precipitation test
For testing the protein precipitation, samples with three standard concentration levels and
three different amounts of methanol are prepared in duplo (table 3.5). Each sample is
injected three times.
Table 3.5: Experimental design of a protein precipitation test (total number of samples = 36)
100 µL MeOH 150 µL MeOH 500 µL MeOH
C3 2 2 2
C5 2 2 2
C7 2 2 2
All samples consisted of 10 µL of each standard mix dilution (4 in total), 10 µL of both labelled internal
standard dilutions and 5 µL human plasma. Every sample is injected three times.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 25
3.2.10.3 Comparison between mouse, rat and human plasma
The comparison between plasma from different species is carried out by using the labelled
standards as calibrants and 4 labelled compounds (asparagine, glutamine, lysine and valine)
as internal standards. This way, the endogenous level of the compounds, which may vary
between the species, is not interfering with the results of the measurement. The used
samples are described in table 3.6, samples are spiked with calibrants both before (“Before”-
samples) and after (“After”-samples) protein precipitation. Each sample is prepared in triplo
and injected three times.
Table 3.6: Experimental design of comparison between mouse, rat and human plasma
(total number of samples = 72)
Conc. - Spiked Mouse Rat Human Academic
C3 - Before 3 3 3 3
- After 3 3 3 3
C5 - Before 3 3 3 3
- After 3 3 3 3
C7 - Before 3 3 3 3
- After 3 3 3 3
All samples consisted of 10 µL of both labelled internal standard dilutions and 5 µL plasma of the
relevant species. The academic samples did not contain plasma. Every sample is injected three times.
3.2.10.4 Comparison between different anti-coagulants
The comparison between EDTA, heparin and citrate as anti-coagulant is carried out the same
way as the comparison between plasma from different species, the experimental design is
described in table 3.7. Each sample is prepared in triplo and injected three times.
Table 3.7: Experimental design of comparison between EDTA, heparin and citrate
(total number of samples = 72)
Conc. - Spiked EDTA Heparin Citrate Academic
C3 - Before 3 3 3 3
- After 3 3 3 3
C5 - Before 3 3 3 3
- After 3 3 3 3
C7 - Before 3 3 3 3
- After 3 3 3 3
All samples consisted of 10 µL of both labelled internal standard dilutions and 5 µL human plasma
contain the relevant anti-coagulant. The academic samples did not contain plasma. Every sample is
injected three times.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 26
3.3 Results and discussion
3.3.1 General chromatography
The samples are analysed using 99 different MRM windows which, in total, measure 57
compounds and 28 labelled internal standards using 99 transitions (figure 3.1). The
robustness of the used chromatography is demonstrated below.
Figure 3.1: MRM windows and number of transitions used for the analysis of the compounds
Figure 3.2 shows the chromatograms of two different injections; one of the first injections of
day one (run 13) compared to one of the last of day three (run 97). No changes in retention
times are observed, which indicates the robustness of this method.
Figure 3.2: Chromatography: day 1 run 13 (above) vs. day 3 run 97 (below)
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 27
The separation between isoleucine and leucine is a good indication for the suitability of the
chromatography for this purpose, because these isomeric amino acids are hard to separate.
Below, the chromatographic separation of isoleucine and leucine is shown for two samples
measured on different days (figure 3.3).
Figure 3.3: Separation of isoleucine and leucine (R=1)
during day 1 run 37 on the left and day 3 run 97 on the right
The resolution is 1 for all samples, which implies an acceptable separation of these isomers.
There is some change in retention time, but the difference is negligible (~0.02 minute) and
possibly caused by renewing the mobile phases in between days. This is another example of
the consistent chromatography of this method, even during large sample runs covering
multiple days.
3.3.2 Validation for human plasma samples
The starting point for this part of this project was validation of the extended amine profiling
method for human plasma samples. The 3-day procedure was carried out as described above
in section 3.2.10.1. The results, however, were not satisfying: the linearity (R2) was poor and
the relative standard deviation (RSD) of the (internal) standards was very high (>45%).
After evaluation of the (raw) data, a presumption was that the protein precipitation was not
sufficient or consistent enough. This can be caused by the high protein concentration in
plasma samples compared to CSF samples. Another approach was chosen to adjust for this
difficulty. The first step of this renewed approach was testing the protein precipitation to
make sure this was complete and consistent by using different volumes of methanol.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 28
Leftover protein can interfere in the ionisation process and on the other hand, there could
also be an amount of compound lost in the precipitate. Both phenomena can cause a higher
RSD, which is screened for in this experiment.
3.3.3 Protein precipitation test
The relative standard deviations of around 20 compounds (both amino acids and amines) are
averaged and plotted with their deviation into figure 3.4.
Clearly, a decrease in the RSD is observed when using 150 µL MeOH for protein precipitation
instead of 100 µL. Also a lower deviation of the RSD is found. When using 500 µL, the RSD is
higher than for 150 µL, but lower compared to 100 µL. From this trend, it can be assumed
that the precipitation is not complete when using 100 µL MeOH, possibly causing
interference during sample preparation and analysis.
The high RSD values for 100 and 500 µL imply that the protein precipitation was not
consistent for these volumes. For 100 µL of added MeOH, this phenomenon is probably
caused by the interference of leftover protein.
By adding 500 µL of MeOH, another cause of the elevated RSD is more likely. A high volume
of MeOH, such as 500 µL, is not easy to handle in this sample pre-treatment protocol. The
sample vails were filled up to such an extend that, during the drying under N2, sample can
spatter out of the vail more easily than when handling lower volumes. This might explain the
higher RSD.
Protein precipitation test
0%
2%
4%
6%
8%
10%
12%
14%
16%
100 150 500
µL MeOH
Ave
rag
e R
SD
of
~20
com
po
un
ds
Figure 3.4: Relative standard deviation for samples deproteinated with different volumes of methanol
Based on these results, the unsatisfactory results of the human plasma validation can be
explained and consequently, 150 µL was used in the follow-up experiments discussed below.
After evaluation of the experiments thus far, this adjusted MeOH-volume was applied on
validation of the method for mouse plasma samples (which was already in preparation). Due
to a shortage of time, no new validation for human plasma samples was done. Instead, a
comparison between mouse, rat and human plasma based on recovery, matrix effect and ion
suppression was composed. The same experiment was applied to human plasma samples
taken with different anti-coagulants. The outcome of these comparisons provides us with a
indication of the applicability of the method on other plasma types.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 29
3.3.4 Validation for mouse plasma
The same 3-day validation procedure as discussed in paragraph 3.3.2 was carried out for
mouse plasma samples. The resulting data is processed using a validation protocol used for
all the validations of the Analytical BioSciences group of Leiden University. This protocol
evaluates the performance of the method for compounds mentioned in paragraph 3.2.2,
regarding linearity, limit of detection and quantification (LOD/Q), recovery, matrix effect, ion
suppression and batch and freeze thaw effect.
A full overview of the results can be found in attachment 9, a brief overall impression and a
more detailed look into the results of one compound (citrulline) is given in this part of this
thesis.
The most important conclusions of the validation for mouse plasma samples are:
- Good linearity is achieved (average R2 of 0.974; range 0.875-0.999)
- General limit of quantification between 7 ng/mL and 1.2 µg/mL, which is comparable
to the results achieved during validation for CSF samples
- Average ion suppression, matrix effect and recovery of, respectively, 96.6 % (66.8 –
111.0), 92.3 % (58.9 – 108.4) and 86.9 % (range 57.2 – 96.0)
- A batch effect is observed for all compounds
- No freeze thaw effect is observed
- Robust chromatography is achieved (see attachment 9 for plotted retention times of
all injections; n>350)
Figure 3.5 shows the calibration curve for citrulline in sample matrix, with A, B and C as code
for the triplicate sample preparation and I, II and III indicating the injection. The R2 of the
curve (0.998), is a sign of adequate linearity for this compound.
Figure 3.5: Citrulline calibration curve in sample matrix, 8 concentration levels
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 30
For all compounds, a batch effect is observed (figure 3.6A for citrulline). The RSD within a day
is 3.54% (average for three concentration levels; n=9), the RSD between different days 6.34%
(average for three concentration levels; n=9). ANOVA shows that this difference is significant
and thus a batch effect is determined.
Figure 3.6: Batch (A) and freeze thaw (B) effect for three concentration levels (n=9)
No substantial difference is found between plasma samples with different amounts of freeze
and thaw cycles, respectively 0, 1 and 2 (figure 3.6B). The RSD of samples within a cycle (3.0%,
average for three concentration levels; n=9) is very similar to the RSD between samples with
different freeze and thaw cycles (2.9%, average for three concentration levels; n=9). ANOVA
shows that no significant difference is observed; freeze thaw cycles do not influence the
analysis.
Overall, it can be concluded that validation of the method for mouse plasma samples was
completed for all 57 compounds and difficulties described in paragraph 3.3.2 are solved by
the adjusted sample preparation. One important remark has to be taken into account; a day-
to-day batch effect occurs.
Due to a lack of time, a new validation for human plasma samples could not be completed.
To get more insight about the applicability of this validation to other types of plasma
samples, two comparisons are made: a first between mouse, rat and human plasma, another
between human plasma samples taken with different anti-coagulants. For these comparisons,
only labelled standards are used to rule out the variation in endogenous levels between the
different plasma samples. The recovery, ion suppression and matrix effect calculated from the
results of these analysis will be discussed below.
3.3.5 Comparison between mouse, rat and human plasma
For experimental observations, there appears to be a difference between the amount of
proteins in the different plasma samples. Human plasma seems to contain the highest
amount of proteins, based on the amount of precipitate after protein precipitation and
centrifugation.
The first thing to look at in this comparison is the recovery. Figure 3.7 shows the average
recoveries for the three different types of plasma. The average RSD of the experimental
observations is 7.5% (range 2.4 – 15.5%), histidine and norepinephrine have higher RSD-
values; respectively 22.7 and 19.2%.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 31
It should be noted that the found recoveries are in general around 80%. This is caused by a
flaw in the experimental design; the internal standards are, for all samples, added before
sample preparation, causing a higher values for the ‘after’-samples and thus artificial lower
recoveries (see attachment 10 for a calculated explanation).
Recovery
0%
20%
40%
60%
80%
100%
120%
His
tidin
e
As p
a ra g
ine
1-m
e th
ylh i
s tid
ine
Se r
i ne
Glu
t am
ine
His
t am
ine
Arg
inin
e
Gly
cin
e
As p
a rt ic
Ac i
d
Gl u
tam
ic A
cid
Be
ta-a
lan
ine
Thr
eo
nine
Ala
nin
e
Nor
epin
eph r
ine
Pr o
li ne
L -a l
ph a
-am
i no b
u ty r
ica
c id
Orn
i thin
e
Lys
ine
Tyr
os i
ne
Me t
hio
n in e
Va
line
3-m
eth
ox y
tyro
sine
3- m
eth
ox y
t yr a
min
e
Isol
eu c
ine
Leu
cine
Ph
e nyl
a la
nin e
Try
pto
p ha
n
Mouse
Rat
Human
Figure 3.7: (Average) Recovery of samples with mouse, rat and human plasma (n=9)
A few irregularities can be found in the data presented in figure 3.7. Norepinephrine shows
the most different results for the recovery. The low recovery and elevated RSD imply that this
compound is instable during protein precipitation. More research is needed to confirm this
phenomenon.
Ornithine and lysine also have lower recoveries compared to the other compounds. Both
compounds contain two amine-groups and are double derivatized by the AccQ·Tag reagent.
The double derivatization might not be complete, or double charged ions could be formed
resulting in a lower recovery of these compounds. The elevated relative standard deviation
for these compounds is a plausible result of this phenomenon.
When focusing on the ion suppression, very similar results are found for all species (figure
3.8). Only three compounds have irregular results: norepinephrine and thryptophan, and
histidine to a lower extend. The first is already discussed above, thryptophan shows
unaccountable numbers for academic samples which cause abnormally high ion suppression.
No direct explanation for this occurring is found within the experiments of this research
project.
Histidine has a 2-3 times higher RSD, which is caused by the irregular shapes peaks that are
hard to integrate. Those irregular peaks are most likely caused by the short retention time of
this compound (tR = 1.68 ; tm = 1.12). This could possibly be prevented by using a higher
concentration level for this compound or adjustment of the gradient to achieve a longer
elution time.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 32
Ion suppression
0%
20%
40%
60%
80%
100%
120%
140%
160%
180%H
istid
ine
Asp
a rag
ine
1-m
e th
ylh
i sti d
i ne
Se r
ine
Glu
tam
i ne
Hi s
t am
ine
Arg
inin
e
Gly
cin
e
As p
a rt ic
Ac i
d
Glu
tam
i c A
c id
Bet
a -a l
ani
ne
Th
reon
ine
Ala
nin
e
No r
e pin
ep h
r ine
Pro
li ne
L-a
l ph
a-am
ino
buty
ricac
id
Orn
ithin
e
L ys i
n e
Tyr
osin
e
Met
h io n
i ne
Val
ine
3-m
eth
o xy t
y ro s
ine
3 -m
e th
oxyt
yram
ine
Iso l
eu c
ine
Leu c
ine
Ph e
n yl a
lan i
ne
Try
pto
p ha n
Mouse
Rat
Human
Figure 3.8: (Average) Ion suppression of samples with mouse, rat and human plasma (n=9)
Results similar to the recovery and ion suppression can be found for the matrix effect
(attachment 11). The matrix effect figures show somewhat more fluctuations due to the fact
that the samples used for these calculations are spiked before sample pre-treatment and thus
put through more steps causing higher variation.
The last step in this comparison is a t-test between the different sample groups concerning
recovery, ion suppression and matrix effect. Concluding from these calculations, there are no
structural significant differences for this method when used with mouse, rat or human plasma
samples regarding these parameters (attachment 11). This observation is consistent with
figures 3.7 and 3.8, which both do not show big differences between application on the
different plasma types.
Exception on this conclusion is methionine, which has a relatively low recovery in human
plasma. This is strengthened by the results from the t-test: the t-value from the t-test
between mouse and rat plasma is distinctively below the critical value of 2.78 (two-sided, 95%
confidence)[42], the t-values between mouse/rat and human plasma are far above this critical
value (table 3.8). This implies a difference in the performance of the method concerning
methionine in human plasma samples.
Table 3.8: T-values for the recovery
t-value
Mouse <-> Rat 1.54
Mouse <-> Human 10.22
Rat <-> Human 6.94
3.3.6 Comparison between different anti-coagulants
The recovery of the samples with different anti-coagulants is very consistent through almost
all compounds. The average RSD of the experimental observations is 12.2% (3.3 – 25.2%),
histidine has a much higher RSD; 42.1% (a possible explanation for this trend is explained
above in paragraph 3.3.5).
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 33
Norepinephrine, again, is a major abnormality, something already explained above
(paragraph 3.3.5). The same goes for histidine, which also has some differences compared to
the other compounds. 1-methylhistidine and histamine have somewhat higher recoveries,
without a direct biological explanation. They do have a elevated RSD compared to the other
compounds (2 times the average) which causes less accurate results.
Recovery
0%
20%
40%
60%
80%
100%
120%
His
tidin
e
Asp
a ra g
i ne
1-m
eth y
lhis
t idin
e
Se r
ine
Gl u
t am
ine
Hi s
t am
ine
Ar g
i ni n
e
Gly
cin e
Asp
a rti c
Ac i
d
Glu
tam
ic A
cid
Be t
a-al
anin
e
Th r
e on i
n e
Ala
n in e
Nor
epin
e ph r
ine
Pro
line
L -a l
pha -
a min
obut
yric
a cid
Or n
i thi n
e
L ysi
n e
Ty r
o si n
e
Met
hio n
ine
Val
i ne
3 -m
e th o
xyty
r os i
n e
3 -m
e th o
xyty
ram
ine
Iso l
e uc i
n e
L euc
ine
Phe
nyla
lani
ne
Tr y
p to p
h an
Citrate
EDTA
Heparin
Figure 3.9: (Average) Recovery of samples with different anti-coagulants (n=9)
One other thing that should be mentioned is the low value for methionine concerning
heparin plasma. This was also noticed in the comparison between mouse, rat and human
plasma, but concluding from the results of this experiment, has to be a phenomenon limited
to analysis of heparin plasma samples.
The ion suppression is very similar for all compounds, expect the previously discussed
exceptions norepinephrine, tryptophan and, at a lower extend, histidine. There are only small
differences in the performance of the method between the sample groups (figure 3.10). The
same thing goes for the matrix effect (attachment 12), although it has some more
fluctuations because the samples used for these calculations are spiked before sample pre-
treatment.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 34
Ion Suppression
0%
20%
40%
60%
80%
100%
120%
140%
160%
180%H
ist id
ine
Asp
ar a
gin
e
1-m
eth
ylhi
s tid
ine
Se r
ine
Glu
tam
ine
His
t am
ine
Arg
inin
e
Gly
cin e
Asp
art
i c A
c id
Glu
t am
ic A
cid
Bet
a-a l
ani
ne
Thr
eoni
ne
Ala
nine
No
repi
ne p
h rin
e
Pro
l ine
L-al
pha-
amin
obu
tyr ic
acid
Orn
ithin
e
L ys i
n e
Ty r
o si n
e
Met
hio n
ine
Val
i ne
3 -m
eth
oxyt
yro s
ine
3-m
eth
oxyt
yram
ine
Isol
eu c
ine
Leu c
ine
Phe
nyla
lani
ne
Try
ptop
han
Citrate
EDTA
Heparin
Figure 3.10: (Average) Ion suppression of samples with different anti-coagulants (n=9)
The results of the t-tests from this dataset show, in most cases, no significant difference of
the method performance when used with human plasma taken with different anti-coagulants
(attachment 12). One anomaly has been observed; for a few compounds (beta-alanine, L-
alpha-aminobutyric acid, isoleucine and leucine), a significant difference in matrix effect has
been observed between citrate on one hand and heparin and EDTA on the other. The same
effect is not observed in the ion suppression results. This implies that citrate has an effect on
these compounds only during sample preparation and not during MS-analysis.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 35
3.4 Recommendations and further research
Below are suggestions for further research, based on the outcomes of this project;
- The use of another protein precipitant should be studied. Acetonitrile and SSA (sulfosalicylic
acid) are much described in the literature and especially acetonitrile might be a good
alternative, while SSA is not favourable when using MS-detection.
- Possibly, the ESI-source is disturbed on the retention times of the compounds with irregular
results described in the results section of this chapter. By performing a full scan in MS-mode,
these disturbances in the ESI-source, for instance caused by left-over protein or co-eluting
compounds, might be detected.
- A MS-scan for multiple charged compounds could be done. Ornithine and lysine are
suspected of gaining multiple charges during electrospray ionization because of their double
derivatization. This presumption can be confirmed by scanning for m/z-values for multiple
charged compounds.
- Norepinephrine should be investigated more intensively to determine what happens during
protein precipitation. A more detailed literature study and/or focused experiments should be
considered.
- The low values for thryptophan in academic samples (which do not contain plasma) are
another anomaly found during this research project: this does not seem to be an anomaly
(the same trend is observed during different experiments), more research on this subject is
needed.
- For some compounds (beta-alanine, L-alpha-aminobutyric acid, isoleucine and leucine) a
significant difference between citrate on one hand and heparin and EDTA on the other is
observed concerning the matrix effect. This trend is not observed in the ion suppression
results, implying an effect of citrate on these compounds during sample preparation. More
experiments to confirm or reject this conclusion are needed.
- Alike, the effect of heparin on the recovery of methionine is another focus point. Lower
recoveries are found for heparin human plasma samples compared to EDTA or citrate as anti-
coagulant. To fully understand this phenomenon, literature should be consulted or new
experiments could be done.
In addition, several general research focus points can be noted. First of all validation of the
method for human and rat plasma samples should be completed using the adjusted MeOH
volume for protein precipitation, including validation for different anti-coagulants. Expanding
the method to monitor even more compounds is the next step in enhancing amine profiling
for metabolomic studies.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 36
Chapter 4: General conclusions
The main aim of this thesis was to develop an enhanced amine profiling method suitable for
biological samples. The first approach to reach this goal was the development of a new
derivatization reagent to be used with, the more sensitive, negative ion-mode. Several
synthesis protocols were applied and eventually 5-nitro-6-isocyanatoquinoline was
synthesized.
N
NO O
N
O+ NH
R1
R2
N
NO O
NH
O
N
R1
R2
5-nitro-6-isocyanatoquinolineAmine or amino acid
Derivatized amine Figure 4.1: Amine derivatization with 5-nitro-6-isocyanatoquinoline
This isocyanate should theoretically form AQC-like derivatized amines with an additional
NO2-functional group (figure 4.1) to gain negative charge stabilisation for better performance
in negative ion-mode. The reagent has to be synthesized and purified in higher quantities
before significant testing can be done. Due to a lack of time, this was not possible during this
research project.
Furthermore, the already established amine profiling method for CSF samples [1] has been
optimised for application to various kinds of plasma samples. Therefore, an adjustment has
been made to the sample preparation: the amount of methanol added to the sample for
protein precipitation has been raised to 150 µL. Afterwards, the method is successfully
validated for 57 compounds (attachment 9) in mouse plasma with a LOD range of 7 ng/mL –
1.2 µg/mL. One important outcome of the validation has to be taken into account; a day-to-
day batch effect is observed for all compounds.
To gain more insight in the extendibility of the validation to other types of plasma samples,
two comparisons have been made. The results of these comparisons show that there are no
significant differences in the performance of the method when analysing plasma samples
from different species or samples taken with different anti-coagulants. These findings are
graphically presented in figure 4.2. Two exceptions on the stated conclusions are observed;
the performance regarding recovery for methionine in heparin human plasma, and regarding
matrix effect for four compounds in citrate human plasma differ compared to the other
samples.
Figure 4.2: Graphical representation of the applicability of the method
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 37
Overall can be concluded that the applicability of the method is extended. In addition, several
leads to enhancing the method by using a new derivatization reagent are proposed.
Especially raising the yield and purity of the synthesized isocyanate requires high priority.
Supplementary, the anomalies which came up during the analytical part of this thesis need
more research, next to validation of the method for rat and human plasma samples.
Ideally, this project is continued in a collaboration between a student with high affinity for
organic synthesis and a student with an analytical chemistry background. The synergy in such
a duo-project might lead to new steps towards enhanced amine profiling.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 38
Acknowledgements
In the first place, I want to thank Shanna Shi for her role as my daily supervisor. Thank you for
all the time you invested in my project and all the questions you answered. Your contribution
to my thesis even involved giving your blood for my experiments!
Secondly, I would like to thank Rob Vreeken for keeping track of my project during my time
in Leiden and correcting my thesis. Next to that, thank you for your help with my questions
about future jobs and applying at DSM.
Richard van den Berg was very important during the organic synthesis part of my project;
thank you for all your suggestions, instructions and showing me there is more than just
analytical chemistry.
I also want to thank everyone in the Analytical BioSciences and Bio-organic Synthesis groups
of Leiden University for their direct and indirect influence on this thesis. Adrie Dane is
gratefully acknowledged for his calculations on the experimental results. Finally, I have to
thank Wim Kok for being my study coordinator and his supervision out of the University of
Amsterdam.
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 39
References
[1] Shi, S., manuscript in preparation
[2] Nicholson, J.K., Xenobiotica, 1999, 29 (11), 1181-1189
[3] Krastanov, A., Biotechnology & Biotechnological Equipment, 2010, 24 (1), 1537-1543
[4] Fukusaki, E., Journal of Bioscience and Bioengineering, 2005, 100 (4), 347-354
[5] KEGG PATHWAY Database, 21-10-2010, http://www.genome.jp/kegg/pathway.html
[6] Introduction to metabolomics, Dave Barrett
http://www.docstoc.com/docs/2699139/Introduction-to-Metabolomics
[7] Pons, R., Journal of Inherited Metabolic Disease, 2009, 32 (3), 321-332
[8] Human Metabolome Database, 21-10-2010, http://www.hmdb.ca/
[9] Zhang, S., Analyst, 2010, 135 (7), 1490-1498
[10] Soga, T., Journal of Proteome Research, 2003, 2 (5), 488-494
[11] Tolstikov, V.V., Analytical Chemistry, 2003, 75 (23), 6737-6740
[12] Tolstikov, V.V., Analytical Biochemistry, 2002, 301 (2), 298-307
[13] Orešič, M., Nutrition, Metabolism and Cardiovascular diseases, 2009, 19 (11), 816-824
[14] Griffiths, W.J., Angewandte Chemie-International Edition, 2010, 49 (32), 5426-5445
[15] Griffiths, W.J., Chemical Society Reviews, 2009, 38 (7), 1882-1896
[16] Rochfort, S.J., Phytochemistry, 2008, 69 (8), 1671-1679
[17] Moore, S., Journal of Biological Chemistry, 1948, 176 (1), 367-388
[18] Moore, S., Analytical Chemistry, 1958, 30 (7), 1185-1190
[19] Turnell, D.C., Clinical Chemistry, 1982, 28 (3), 527-531
[20] Bidlingmeyer, B.A., Journal of the Association of Official Analytical Chemists, 1987, 70 (2),
241-247
[21] Einarsson, S., Journal of Chromatography A, 1983, 282 (Dec), 609-618
[22] Bidlingmeyer, B.A., Journal of Chromatography B, 1984, 336 (1), 93-104
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 40
[23] Cohen, S.A., Analytical Biochemistry, 1993, 211 (2), 279-287
[24] Bosch, L., Journal of Chromatography B, 2006, 831 (1-2), 176-183
[25] Callejon, R.M., European Food Research and Technology, 2008, 227 (1), 93-102
[26] Kabelova, I., Journal of Food Composition and Analysis, 2008, 21 (8), 736-741
[27] Boogers, I., Journal of Chromatography A, 2008, 1189 (1-2), 406-409
[28] Phenomenex Application Note TN-110: EZ:faast
[29] Langrock, T., Amino Acids, 2006, 30 (3), 291-297
[30] Mayer, H.K., Journal of Chromatography A, 2010, 1217 (19), 3251-3257
[31] Hou, S.M., Talanta, 2009, 80 (2), 440-447
[32] Mass Spectrometry; Principles and Applications (Third Edition); Edmond de Hoffman and
Vincent Stroobant
[33] Bigwarfe Jr., P.M., Rapid Communications in Mass Spectrometry, 2002, 16 (24), 2266-
2272
[34] Spackman, D.H., Analytical Chemistry, 1958, 30 (7), 1190-1206
[35] Loukou, Z., Journal of Chromatography A, 2003, 996 (1-2), 103-113
[36] U.S. Patent 6, 770, 246, 2004
[37] Mąkosza, M., Journal of Organic Chemistry, 1998, 63 (15), 4878-4888
[38] Smith, G.E.P. Jr., Journal of Organic Chemistry, 1949, 14 (6), 935-945
[39] Soulier, J.L., Journal of Medicinal Chemistry, 1997, 40 (11), 1755-1761
[40] Wilson, A.A., Organic & Biomolecular Chemistry, 2010, 8 (2), 428-432
[41] Valli, V.L.K., Journal of Organic Chemistry, 1995, 60 (1), 257-258
[42] Miller, J.N. & Miller J.C., Statistics and Chemometrics for Analytical Chemistry (Fifth
edition) 2005
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 41
Attachments
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 42
Attachment 1
1H NMR spectrum of N-tetramethylethiocarbamoylsulfenamide
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 43
Attachment 2
13C NMR spectrum of N-tetramethylethiocarbamoylsulfenamide
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 44
Attachment 3
1H NMR spectrum of 6-amino-5-nitroquinoline
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 45
Attachment 4
TLC-MS scan of 6-amino-5-nitroquinoline ([M+H]+ = 190.1)
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 46
Attachment 5
Composition of standard mixes for validation and protein precipitation test
Mix 1 Mix 3
Weight % Weight %
Aspartic acid 1,0 Beta-Alanine 19,8
Glutamic acid 1,5 Norepinephrine 13,7
Asparagine 1,8 S-adenosylhomocysteine 11,1
Serine 3,1 dopamine 12,5
Glutamine 24,9 Epinephrine 29,8
Histidine 2,6 serotonin 7,9
Glycine 4,3 3-methoxytyramine 5,2
Threonine 4,0 Total 100
Alanine 6,7
Arginine 4,9 Mix 4
Tyrosine 3,1 Weight %
Valine 7,4 DL-5-hydroxy-lysine 4,1
Methionine 1,2 gamma-L-glutamyl-L-alanine 4,5
Tryptophan 2,5 L-4-hydroxy-proline 3,9
Phenylalanine 2,7 Sarcosine 4,2
Isoleucine 2,2 Hydroxylamine 4,6
Leucine 4,5 L-2-aminoadipic acid 4,0
Lysine 5,0 5-hydroxy-L-tryptophan 4,3
Proline 4,0 L-carnosine 4,6
Cystine 4,6 methyldopa 4,3
Taurine 2,2 L-homoserine 4,1
Citrulline 2,0 S-(5)-adenosyl-L-homocysteine 3,3
L-alpha-aminobutyric acid 1,1 phosphoserine 4,0
Ornithine 1,9 Nε,Nε,Nε-trimethyllysine 4,0
Homocysteine 1,1 homocystine 4,3
Total 100 DL-3-aminoisobutyric acid 3,8
5-aminolevulinic 4,2
Mix 2 Glycylglycine 4,3
Weight % ethanolamine 4,3
O-Phosphoethanolamine 15,1 o-acetyl-L-serine 4,5
3-Methyl-L-Histidine 12,6 s-methyl-L-cysteine 4,3
Taurine 26,4 L-methionine sulfoxide 4,0
1-Methyl-L-Histidine 1,5 L-pipecolic acid 4,1
Histamine 0,7 gamma-aminobutyric acid 4,3
Citrulline 16,5 L-alpha-aminobutyric acid 3,9
Ornithine 16,2 Total 100
Putrescine 2,0
L-Kynurenine 1,1
Homo-L-arginine 7,8
Total 100
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 47
Attachment 6
Used SRM transitions for MS/MS detection (tm = 1.12 minutes)
Compound Mass Transition tR
L-Histidine 326 171 1,68
L-4-hydroxy-proline 302 171 1,69
O-Phosphoethanolamine 312 171 1,86
L-Asparagine 303 171 1,94
L-Asparagine_C13N15 309 171 1,94
3-Methylhistidine 340 171 2,08
Taurine 296 171 2,22
1-Methylhistidine 340 171 2,36
1-Methylhistidine_d3 343 171 2,36
Glycylglycine 303 171 2,52
L-Serine 276 171 2,54
L-Glutamine 317 171 2,74
L-Glutamine_C13N15 324 171 2,74
N6,N6,N6-Trimethyl-L-lysine 359 171 2,69
L-Arginine 345 171 2,81
L-Arginine_C13N15 355 171 2,81
Histamine 282 171 2,82
Histamine_d4 286 171 2,82
Glycine 246 171 2,91
Glycine_C13_N15 249 171 2,91
L-carnosine 397 171 3,01
L-homoserine 290 171 3,25
L-Aspartic acid 304 171 3,26
L-Aspartic acid_C13N15 309 171 3,26
ethanolamine 232 171 3,29
L-methionine sulfoxide 336 171 3,4
L-Glutamic acid 318 171 3,86
L-Glutamic acid_C13N15 324 171 3,86
Sarcosine 260 171 3,87
Citrulline 346 171 3,88
Homo-L-arginine 359 171 4,05
Beta-Alanine 260 171 4,09
Beta-Alanine_d4 264 171 4,09
L-Threonine 290 171 4,33
L-Threonine_C13N15 295 171 4,33
gamma-L-glutamyl-L-alanine 389 171 4,54
5-aminolevulinic 302 171 4,68
L-Alanine 260 171 4,77
L-Alanine_C13N15 264 171 4,77
gamma-aminobutyric acid 274 171 4,95
L-2-aminoadipic acid 332 171 5,14
o-acetyl-L-serine 318 171 5,14
Norepinephrine 340 171 5,22
Norepinephrine_d6 346 171 5,22
DL-3-aminoisobutyric acid 274 171 5,41
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 48
L-Proline 286 171 5,45
L-Proline_C13N15 292 171 5,45
Epinephrine 354 171 5,84
Epinephrine_d3 357 171 5,84
S-(5)-adenosyl-L-homocysteine 555 171 5,85
L-Alpha-aminobutyric acid 274 171 6,04
L-Alpha-aminobutyric acid_d6 280 171 6,04
Ornithine 473 171 6,14
Ornithine_d6 479 171 6,14
5-hydroxy-L-tryptophan 391 171 6,44
L-Lysine 487 171 6,54
L-Lysine_C13N15 495 171 6,54
Dopamine 324 171 6,56
L-Tyrosine 352 171 6,61
L-Tyrosine_C13N15 362 171 6,61
methyldopa 382 171 6,66
Putrescine 429 171 6,68
L-Methionine 320 171 6,76
L-Methionine_C13N15 326 171 6,76
L-pipecolic acid 300 171 6,85
Serotonine 347 171 6,85
L-Valine 288 171 6,9
L-Valine_C13N15 294 171 6,9
3-Methoxytyrosine 382 171 6,92
3-Methoxytyrosine_d3 385 171 6,92
3-Methoxytyramine 338 171 7,33
3-Methoxytyramine_d4 342 171 7,28
homocystine 609 171 7,49
s-methyl-L-cysteine 306 171 7,5
L-Kynurenine 379 171 7,59
L-Isoleucine 302 171 7,65
L-Isoleucine_C13N15 309 171 7,65
L-Leucine 302 171 7,73
L-Leucine_C13N15 309 171 7,72
L-Phenylalanine 336 171 7,83
L-Phenylalanine_C13N15 346 171 7,82
L-Tryptophan 375 171 7,92
L-Tryptophan_C13N15 388 171 7,91
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 49
Attachment 7
MS/MS source and analyzer parameters used for analysis of the derivatized amines and internal
standards
Source parameters Analyzer parameters
Polarity ES+ Low Mass 1 Resolution 3.0
Capillary (kV) 3.20 High Mass 1 Resolution 3.0
Cone (V) 52.00 Ion Energy 1 0.5
Extractor (V) 3.00 MS Mode Collision Energy 4.00
Source Temperature (°C) 140 MSMS Mode Collision Energy 20.00
Desolvation Temperature (°C) 450 MS Mode Entrance 0.50
Cone Gas Flow (L/Hr) 50 MS Mode Exit 0.50
Desolvation Gas Flow (L/Hr) 1000 Low Mass 2 Resolution 3.0
Collision Gas Flow (mL/min) 0.10 High Mass 2 Resolution 3.0
Ion energy 2 1.0
Multiplier 509.16
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 50
Attachment 8
• Calculation of recovery, matrix effect and ion suppression
Sample classes:
Recovery: ∗1100%
2 Matrix effect: ∗1
100%3
Ion suppression: ∗2100%
4
• t-test
−=
+
1 2
1 2
( )
1 1
x xt
sn n
where − + −
=+ −
2 22 1 1 2 2
1 2
( 1) ( 1)
( 2)
n s n ss
n n
• ANOVA
21 1
( )( )
n m
iji j
total total
XTotal of all observations
CMN N
= == =∑ ∑
2
1 1
n m
total iji j
SS X CM= =
= −∑ ∑
2
1
ni
i i
TSST CM
n== −∑
totalSSE SS SST= −
1
SSTMST
k=
−
SSEMSE
N k=
−
MSTF
MSE=
4
No sample prep.
3
Spiked before
sample prep.
Academic
2
Spiked after
sample prep.
1
Spiked before
sample prep.
Matrix
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 51
Attachment 9
Overview of the validation results for mouse plasma samples
Validation characteristics for each compound
Component Slope Intercept R2
LOQ
[ug/ml]
Precision
RSD
Batch
Effect
L-histidine 0,332 0,134 0,988 0,235 < 6% Yes
L-4-hydroxyproline 2,696 0,441 0,997 0,005 < 10% Yes
o-phosphoethanolamine 0,955 0,035 0,993 0,008 < 10% Yes
L-asparagine 3,140 1,451 0,998 0,040 < 10% Yes
3-methylhistidine 1,773 0,212 0,992 0,019 < 10% Yes
taurine 1,826 5,364 0,978 0,691 < 6% Yes
L-serine 6,644 5,552 0,998 0,168 < 6% Yes
N-methylhistidine 2,586 0,178 0,977 0,008 < 10% Yes
Histamine 5,373 -0,006 0,987 0,003 < 10% Yes
L-arginine 4,525 3,998 0,992 0,227 < 6% Yes
L-glutamine 9,508 36,064 0,996 0,641 < 6% Yes
glycine 4,113 6,050 0,994 0,373 < 6% Yes
glycylglycine 25,450 0,052 0,991 0,002 < 10% Yes
L-carnosine 1,271 0,005 0,982 0,045 < 10% Yes
L-homoserine 22,616 0,074 0,979 0,010 < 10% Yes
N6,N6,N6-trimethyllysine 1,798 0,019 0,954 0,043 < 10% Yes
L-aspartic acid 2,892 0,094 0,998 0,021 < 10% Yes
L-glutamic acid 1,809 0,458 0,999 0,038 < 10% Yes
sarcosine 4,317 0,039 0,995 0,008 < 10% Yes
citrulline 0,977 0,491 0,998 0,197 < 6% Yes
Beta-alanine 3,880 0,056 0,984 0,003 < 10% Yes
ethanolamine 22,133 1,767 0,958 0,076 < 10% Yes
L-methioninesulfoxide 3,584 0,117 0,996 0,011 < 10% Yes
gamma-aminobutyricacid 5,043 0,005 0,995 0,001 < 10% Yes
L-threonine 3,567 4,697 0,998 0,202 < 6% Yes
L-alanine 1,803 6,977 0,947 4,743 < 6% Yes
5-aminolevulinic 0,157 0,004 0,949 0,071 < 10% Yes
DL-3-aminoisobutyricacid 2,411 0,001 0,997 0,000 < 10% Yes
homo-L-arginine 0,026 0,003 0,965 0,225 < 6% Yes
gamma-L-glutamyl-L-alanine 0,175 0,004 0,984 0,083 < 10% Yes
epinephrine 0,920 0,001 0,905 0,010 < 10% Yes
L-2-aminoadipicacid 3,895 0,548 0,996 0,070 < 10% Yes
S-(5)-adenosyl-L-
homocysteine 0,173 0,000 0,987 0,001 < 10% Yes
o-acetyl-L-serine 0,752 -0,001 0,962 0,016 < 10% Yes
Norepinephrine 9,174 0,001 0,965 0,001 < 10% Yes
putrescine 0,089 0,000 0,904 0,019 < 10% Yes
3-methoxytyrosine 3,230 0,006 0,988 0,002 < 10% Yes
L-alpha-aminobutyricacid 2,462 0,090 0,997 0,011 < 10% Yes
3-methoxytyramine 3,738 0,000 0,993 0,000 < 10% Yes
ornithine 3,325 1,652 0,991 0,116 < 6% Yes
dopamine 8,396 0,034 0,979 0,003 < 10% Yes
methionine 7,706 3,998 0,996 0,065 < 10% Yes
L-lysine 7,751 16,706 0,992 0,475 < 6% Yes
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 52
L-tyrosine 3,490 3,611 0,995 0,267 < 6% Yes
L-valine 1,994 4,192 0,915 2,340 < 6% Yes
L-proline 8,830 8,448 0,997 0,246 < 6% Yes
L-kynurenine 0,404 0,005 0,988 0,023 < 10% Yes
L-tryptophan 4,840 3,660 0,998 0,149 < 6% Yes
L-isoleucine 2,391 2,002 0,998 0,136 < 6% Yes
L-leucine 1,227 1,661 0,934 1,601 < 6% Yes
L-phenylalanine 7,940 4,579 0,996 0,068 < 10% Yes
cystine 1,413 0,000 0,911 2,999 < 6% Yes
5-hydroxy-L-tryptophan 4,559 -0,016 0,972 0,024 < 10% Yes
methyldopa 0,929 -0,003 0,982 0,011 < 10% Yes
homocystine 0,014 0,000 0,904 0,008 < 10% Yes
s-methyl-L-cysteine 0,248 0,000 0,890 0,006 < 10% Yes
L-pipecolicacid 0,076 0,003 0,875 0,014 < 10% Yes
serotonine 2,322 0,029 0,952 0,006 < 10% Yes
Plotted retention times for all injections (n>350)
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 53
Attachment 10
Example calculations of the low recovery
“Before” samples: internal standards added before, calibrants added before
“After” samples: internal standards added before, calibrants added after
If a loss of compound during sample preparation is assumed, the following effect on the
relative area will be observed:
Area Internal standards Area Calibrants Relative area
Before Decreased Decreased No change (e.g. 100)
After Decreased No change Increased (e.g. 120)
Relative area = Area Calibrants / Area internal standards
Example of resulting recovery:
Recovery = ∗100%Before
After
= ∗ =100100% 83%
120
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 54
Attachment 11
Additional figures and tables for the comparison of mouse, rat and human plasma
Matrix effect
0%
20%
40%
60%
80%
100%
120%
140%
His
tidin
e
Asp
arag
ine
1-m
eth
y lh
ist id
i ne
Ser
ine
Glu
tam
i ne
His
t am
ine
Ar g
inin
e
Gl y
c in
e
As p
a rt ic
Ac i
d
Glu
tam
ic A
cid
Be t
a-al
an i
ne
Th
reo n
ine
Ala
nin
e
Nor
epin
ep h
rin e
Pro
line
L-a
lph
a-am
ino
b uty
r icac
id
Orn
ithin
e
Lysi
ne
Ty r
o si n
e
Me t
h io n
i ne
Va l
ine
3 -m
e th
oxyt
yros
ine
3 -m
e th
oxyt
yra m
ine
Iso l
eu c
ine
L eu c
i ne
Phe
nyla
lani
ne
Try
p to
p ha n
Mouse
Rat
Human
Matrix effect for the comparison of mouse, rat and human plasma (n=9)
t-test results compared to critical value 2.78 (two-sided, 95% confidence)[42]:
green=no significant difference, red=significant difference
Recovery Ion suppression Matrix effect
Histidine Mouse<->Human 0,94 1,39 0,95
Rat<->Human 0,44 0,82 3,50
Mouse<->Rat 1,39 0,25 1,29
Asparagine etc. 2,41 1,10 0,07
3,00 0,63 0,51
0,48 0,68 0,51
1-methylhistidine 0,54 1,01 0,31
0,23 0,18 0,10
0,28 0,77 0,38
Serine 0,67 4,11 1,43
0,09 1,67 1,19
0,66 0,78 0,70
Glutamine 0,12 0,48 0,94
0,55 0,94 0,77
0,24 1,21 1,30
Histamine 2,31 1,07 7,00
1,08 1,29 4,41
1,07 0,57 0,89
Arginine 1,72 0,43 1,35
0,61 0,01 0,64
2,95 0,54 4,82
Glycine 0,18 0,77 0,86
0,25 0,12 0,05
0,73 0,68 1,19
Aspartic Acid 1,35 0,10 0,54
0,04 0,12 0,34
1,10 0,00 0,70
Glutamic Acid 3,48 0,02 2,23
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 55
0,43 0,06 0,04
1,89 0,05 1,07
Beta-alanine 2,86 0,24 4,44
0,97 0,02 1,19
1,17 0,24 2,97
Threonine 1,80 0,73 3,82
0,49 0,29 1,52
0,66 0,36 1,62
Alanine 1,01 0,79 2,85
0,38 0,45 0,41
1,73 0,39 1,68
Norepinephrine 2,77 0,21 1,36
1,56 0,34 0,73
0,35 0,45 0,25
Proline 0,69 0,72 0,96
0,15 0,22 0,06
0,31 0,90 0,70
L-alpha-aminobutyricacid 0,85 0,42 6,49
0,15 0,14 0,30
0,75 0,28 2,55
Ornithine 1,14 0,00 0,84
0,62 0,05 0,93
0,07 0,05 0,34
Lysine 1,93 0,74 1,61
1,18 0,02 1,43
1,17 0,79 0,09
Tyrosine 2,31 0,27 2,56
0,32 0,04 0,18
2,79 0,33 1,58
Methionine 10,22 0,97 1,99
6,94 0,65 0,94
1,53 2,16 2,38
Valine 4,38 0,79 2,19
0,56 0,26 0,32
2,33 0,52 1,01
3-methoxytyrosine 2,10 1,23 8,98
0,12 0,45 0,64
1,26 0,82 1,97
3-methoxytyramine 2,23 0,75 2,94
0,82 0,73 1,57
1,64 0,02 3,70
Isoleucine 2,93 0,19 1,27
0,43 0,53 0,24
1,81 0,33 0,91
Leucine 1,31 0,35 1,02
0,12 0,14 0,08
1,72 0,56 0,68
Phenylalanine 1,37 0,33 1,31
0,10 0,21 0,17
0,91 0,19 0,74
Tryptophan 2,74 0,02 1,27
0,34 0,06 0,08
3,27 0,04 1,02
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 56
Attachment 12
Additional figures and tables for the comparison of EDTA, heparin and citrate
Matrix Effect
0%
20%
40%
60%
80%
100%
120%
140%H
isti d
ine
As p
ara
gin
e
1 -m
eth
y lhi
sti d
ine
Ser
ine
Glu
t am
ine
Hi s
tam
ine
Ar g
inin
e
Gly
cine
Asp
ar t
ic A
c id
Glu
t am
ic A
cid
Be t
a -a l
an i
n e
Th r
eoni
ne
Ala
nin e
No
r ep i
nep
hri n
e
Pro
l ine
L-a l
p ha -
amin
o bu
tyr ic
a cid
Or n
it hin
e
L ys i
n e
Tyr
o si n
e
Met
h io n
ine
Va l
ine
3 -m
eth
o xy t
y ros
ine
3 -m
et h
o xy t
yram
ine
Isol
eu c
ine
Leuc
ine
Ph e
n yl a
lani
ne
Try
ptop
han
Citrate
EDTA
Heparin
Matrix effect for the comparison of EDTA, heparin and citrate as anti-coagulant (n=9)
t-test results compared to critical value 2.78 (two-sided, 95% confidence)[42]:
green=no significant difference, red=significant difference
Recovery Ion suppression Matrix effect
Histidine Citrate<->Heparin 0,74 0,14 0,70
Citrate<->EDTA 1,11 0,15 0,85
EDTA<->Heparin 1,16 0,29 0,96
Asparagine etc. 0,02 0,46 0,64
etc. 0,30 0,74 0,90
etc. 0,22 0,35 0,25
1-methylhistidine 1,26 0,42 1,15
0,83 0,80 0,85
0,10 0,73 0,01
Serine 0,21 0,15 0,26
0,70 0,78 0,80
0,34 0,08 0,76
Glutamine 0,26 0,05 0,34
0,29 0,10 0,42
0,04 0,06 0,03
Histamine 0,10 0,13 0,36
0,60 0,03 1,08
0,47 0,17 1,77
Arginine 0,40 0,37 0,09
1,44 0,01 1,39
0,96 0,33 0,76
Glycine 0,22 0,33 0,20
0,65 0,39 0,15
0,67 0,66 0,11
Aspartic Acid 4,60 1,15 1,01
0,58 0,92 0,13
2,49 1,75 0,72
Glutamic Acid 1,23 0,42 1,20
0,56 1,70 0,36
Towards enhanced amine detection for plasma samples
with UPLC-MS/MS for metabolomic studies
W.F. Duvivier 57
2,20 1,05 1,86
Beta-alanine 2,86 0,24 4,44
1,17 0,24 2,97
0,97 0,02 1,19
Threonine 0,67 0,74 0,11
0,19 0,14 0,26
0,53 0,79 0,38
Alanine 1,39 0,11 1,57
0,36 1,85 0,77
1,29 0,48 0,69
Norepinephrine 3,11 0,02 0,91
1,19 0,44 0,89
5,66 0,49 3,36
Proline 0,74 1,32 2,67
0,99 6,25 2,66
0,64 0,18 0,92
L-alpha-aminobutyricacid 2,01 0,22 2,87
1,14 1,50 3,18
1,08 0,32 0,69
Ornithine 0,28 0,50 0,23
0,10 0,52 0,32
0,41 0,92 0,54
Lysine 0,06 0,52 0,65
0,83 1,18 0,20
0,77 1,57 0,82
Tyrosine 0,38 0,87 0,15
0,19 2,25 1,18
0,89 1,52 2,26
Methionine 1,03 0,71 0,95
2,34 1,53 1,94
1,55 3,13 1,61
Valine 1,03 0,28 1,66
1,14 1,37 2,14
0,29 0,47 1,01
3-methoxytyrosine 0,07 0,41 0,13
1,07 0,58 2,51
1,47 0,64 2,95
3-methoxytyramine 0,17 0,44 0,26
0,47 0,46 4,10
1,44 0,05 1,73
Isoleucine 2,40 0,33 4,98
1,70 2,47 5,32
0,02 1,35 2,03
Leucine 1,46 0,56 3,56
1,20 2,97 4,34
0,07 1,46 2,15
Phenylalanine 0,07 1,08 0,65
0,77 1,41 2,88
1,16 0,76 2,52
Tryptophan 0,07 0,67 0,91
0,93 1,37 3,31
1,14 0,75 2,82