Thermo Scientific Orion VERSA STAR™ RDO/DO/Temperature Module Reference Guide

Upload: others

Post on 22-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

Thermo Scientific Orion VERSA STAR™ RDO/DO/Temperature Module

Reference Guide

Page 2: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.

AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.

A+, All in One, Aplus, AUTO-BAR, AUTO-CAL, Auto-ID, AUTO-READ, AUTO-STIR, Auto-Test, AutoTration, CISA, digital LogR, DuraProbe, EZ Startup, ISEasy, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, Optimum Results, Orion Dual Star, Orion Star, SAOB, SMART AVERAGING, SMART STABILITY, Star LogR, Star Navigator 21, Stat Face, Triode are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.

Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific, Inc. and its subsidiaries.

© 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries.

The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.

This publication supersedes all previous publications on this subject.

Page 3: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

Thermo Scientific Orion VERSA STAR™ RDO/DO/Temperature Module

Table of Contents

Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Chapter 2 Module Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Module Connections and Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Module Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Information From Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Chapter 3 Setup Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Setup Menus for Primary Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Read Type Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6RDO Probe Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

Setup Menu for Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Chapter 4 RDO/DO Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Calibration Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Chapter 5 Temperature Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Chapter 6 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Chapter 7 Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Chapter 8 Data Storage and Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Data Storage Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Measurement Read Type . . . . . . . . . . . . . . . . . . . . . . . . . . .17Datalog, Computer and Printer Export Settings . . . . . . . . . .17

Datalog Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Calibration Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Chapter 9 Customer Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Warranty and Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19WEEE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20RDO/DO/Temperature Module Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Appendix Advanced Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Dissolved Oxygen Barometric Pressure Compensation . . . . . . . . . . . . . . . . . . 27Dissolved Oxygen Salinity Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Page 4: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

This page intentionally left blank.

Page 5: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

1Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 1 Introduction

Thank you for your purchase of the Thermo Scientific Orion VERSA STAR™ RDO/DO/Temperature Module.

This module is for use with the Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter.

Some key features of this module include:

• WorkswithandautomaticallyrecognizesOrionpolarographicDOandRDOoptical/

luminescence-base probes

• PolarographicDOprobemeasurementrangeof0to90mg/Land0to600%saturation

• RDOprobemeasurementrangeof0to50mg/Land0to500%saturation

• Readingscompensatefortemperature,salinityandbarometricpressure

• Automaticandmanualdataloggingoptions

• Quickcalibrationusingwater-saturatedair,air-saturatedwater,Winklertitrationorzeropoint

Pleasereadthisreferenceguidethoroughly.Anyuseoutsideoftheseinstructionsmayinvalidateyourwarrantyand

causepermanentdamagetothemeter.

Page 6: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

This page intentionally left blank.

Page 7: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

3Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 2 Module Overview

Toconnectthemodulestothemeterandforspecificinformationregardingthemeteroutputs,pleaseseethe

Orion VERSA STAR Reference Guide. The Reference Guide is on the CD included with the meter and also available

at www.thermoscientific.com/water.

Module Connections and Inputs

1. SelecteithertheRDOorPolarographicDOprobebasedonsamplerangeandtestingrequirements.

2. Placeselectedprobeinthemeter’selectrodearm.IfusingtheRDOprobe,attachtheincludedadapter.

3. Attachthe9pinmini-DINconnectionintotheRDO/DOmodule.

4. Attachthestirrertothesideofthemeterandintothemeter'selectrodearm,ifneededforyour

test method.

5. Setupareawithdeionizedrinsewater,sampleandothersupplies.

6. Connectthemeterfordatacollection(Hyperterminal,LIMS,StarPrinter,etc),ifdesired.

7. PowertheOrionVERSASTARmeter,stirrer,anddatacollectiondevice(s).

RDO/DO Module

9pinmini-DINconnectorfor RDOorpolarographicDOprobe

Page 8: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Module Maintenance

Forroutinemaintenance,dustandwipethemodulewithadampcloth.Ifnecessary,warmwaterormildwater-based

detergentcanbeused.Maintenancecanbeperformedonadaily,weeklyormonthlybasis,asrequiredbyoperating

environment.Immediatelyremoveanyspilledsubstanceusingthepropercleaningprocedureforthatspilltype.

Display Information from Module

ExampleRDOMeasurementScreen–SingleChannel

Date and time

Primaryreading

Barometric pressure

Sample Temperature

Calibration information

RDO cap life information

(changestored colorat30days)

Theinformationonthedisplaycanbecustomized.Enterthesetupmenutomakechanges.1.PressSetup.2.UsethearrowkeystohighlightInst.Settings, press (f3/select) key.3.UsethearrowkeytohighlightDisplayView. Press(f3/select) key.4. Follow the onscreen prompts to choose what you would like to see on the measurement screen.

Page 9: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

5Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 3 Setup Menus

Setup Menus for Primary Measurements

1. SelecttheRDO/DOchannelbypressingthechannelkey,thenpressf3 (Setup).

The meter detects the RDO or DO probe type automatically and selects mode menu.

2. Selecttheoptionsforyourtesting.

3. Pressthef1 (Back) button to step back to the main screen.

RDO/DO Setup Menu (Default values are in bold in the table below)ProbeSerialNumber InputSampleID Off,Manual,AutoIncrement

ResolutionDependent on measurement unit: 0 .1 (default),1for%sat 0.01(default),0.1formg/L

Stability Smart,Fast,Medium,SlowAveraging Off,Automatic SmartRead Type AutoRead,Timed,ContinuousMeasurement Unit mg/L,%PressureCompensation Auto,ManualSalinity Correction Auto*,ManualAlarm** Limit,CalDue,SetPoint*ForautomaticsalinitycorrectionofRDO/DOreading, ensureoxygenprobeandconductivitycellareinthesamesample.**Limitandsetpointalarmdefaultsetto‘off’. Cal due alarm set to 12 hours.

Page 10: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

6 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Read Type Information

There are three main read types available as shown in the setup menus.

• Auto-Read – Themeterwilldisplaythemeasurementasitstabilizesandlockandholdthe

measurementwhenitisstable.(TheARiconwilllockonthescreen.)Oncethemeasurementislocked,

themeterwillautomaticallyexportthemeasurementtothedatalog,ifthedatalogisenabledin

thesetupmenu,andtoaprinterorcomputer,ifaprinterorcomputerisconnectedtothemeterand

enabledinthesetupmenu.Totakeanewmeasurement,pressmeasure (esc)/1.

• Timed – Themeterwillreadanddisplayameasurementatthesettimeintervalthatisprogrammed

by the operator. There are two options:

1. Bychecking/selectingthe"Interval"box,atthesettimeinterval,themeterwillautomatically

exportthemeasurementtothedatalog,ifthedatalogisenabledinthesetupmenu,andtoa

printerorcomputer,ifaprinterorcomputerisconnectedtothemeterandenabledinthesetup

menu.Ifselected,enterthetimeintervalvalueinahours:minutes:secondsformatusingthe

numeric keypad and press f3 (select) key.

2. Ifthe"Interval"boxisunchecked/notselected,asingletimedmeasurementwillbetaken.

Themeterwillapplythesamesavingandexportingdataconditionsaswiththesettimeinterval

torecordandexportdata,butlockafteronetimedintervalhasexpired.Pressmeasure (esc)/1

toresetthecountdownclockandtakeanewreading.

• Continuous – The meter will continuously measure and update the display. This read type is useful

whenperforminganexperimentthatrequirescontinuousmeasurementstobetaken.Pressthe

log/print/0keytoexportthemeasurementtothedatalog,ifthedatalogisenabledinthesetup

menu,andtoaprinterorcomputer,ifaprinterorcomputerisconnectedtothemeterandenabledin

the setup.

RDO Probe Information

TheRDOprobeworkswithauniqueopticalcap.Informationexclusivetothiscap,suchas:

• Serialnumber

• Remainingcaplife

• Dateofmanufacture

canbefoundinthesetupmenuwhentheRDOprobeisconnectedandrecognizedasthemeasurementmode.

1. Verify RDO probe is connected.

2. Press setup/3.

3. UsethearrowstohighlightthechannelwithRDOifneeded.Pressf3 (Select) key.

4. UsetherightarrowtohighlightRDOElectrodeInfotoreviewinformation.

Page 11: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

7Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Setup Menu for Temperature

AllofthetemperaturemeasurementsettingsarefoundinTemperatureSetupmenuforeachmeasurement.

*Thisnumberwillchangetomatchthesetupmenuchannel.

Temperatureprobescanbesetupinthemeasurementmodeofthesampleyouaremeasuring.Ch-1ATCallowsthe

temperatureprobetobeusedinmorethanonechannelfortemperature-compensatedmeasurements,suchaswhen

apHelectrodeanda2-cellconductivitycellareinthesamemeasurementsample.TemperatureisacriticalpHand

conductivitymeasurementasthemeasurementvaluesrequireatemperaturecorrectiontodeterminethereference

value,often25ºCintheAmericaand20ºCinEurope.

Temperature Unit ºCorºFselectionTemperature Input and Type Ch-1ATC,Ch-2ATC,Ch-3ATC,Ch-4ATC,MTC

Temperature Cal Ch-1* ATC only or All ATC allows all ATCs to be calibrated at once

Page 12: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

8 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

This page intentionally left blank.

Page 13: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

9Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 4 RDO/DO Calibration

PolarographicDOprobesonly–Theseprobesneedtobepolarizedpriortouse.Itiscontinuouslypolarizedwhen

connectedtothemeter.Iftheprobeisneworhasnotbeenconnectedtothemeter:connecttheprobetothemodule

connectedtothemeter,connectthemetertoapowersourceandwait30minutesforpolarization.

Calibration Methods

• Water-Saturated Air (Air) – Thesimplestandmostaccuratemethod,thismethodusesthe

calibration sleeve included with the DO probe. Due to the inherent differences between water

saturatedairandair-saturatedwater,102.3%saturationwillbedisplayedwhenthecalibration

readingisstableusingapolarographicDOprobe.Forbestaccuracy,thecalibrationtemperature

shouldmatchthemeasuringtemperature.Moistenthespongeorabsorbentclothinthecalibration

sleevewithdistilledwaterandinserttheprobeintothesleevewithouttouchingthe

water-saturated material.

• Air-Saturated Water (Water) – Usewaterthatis100%saturatedwithair.Bubbleairintoawater

sampleforanextendedperiodoftime,preferablyovernight.

• Winkler Titration (Manual) – This method uses a water sample with a known concentration

ofdissolvedoxygenandcanbeusedtocalibratetheDOprobetothevalueachievedbyaWinkler

titration.AmanualcalibrationinvolvesperformingaWinklertitrationandusingthatsampleasa

calibrationstandard.Theoxygenlevelresultingfromthetitrationisentered.Thiscorrelatesthe

meterinputtotheWinklertitration.Duetothepossibilityoftitrationerrors,thismethodisinherently

less accurate.

• Set Zero – Azeropointcalibrationrequiresoxygen-freesolution.Thismethodneedstoberepeated

eachtimetheprobeisreplacedormodified(capchangewithRDOprobes;membranechangeor

freshelectrolytewithpolarographicDOprobes).Thiscalibrationisgenerallynotrequiredunless

measurementswillbetakenbelow10%saturationor1mg/L.Anairorwatercalibrationneedstobe

performedpriortoazeropointcalibration.

Calibration Procedure

1. Presscal/. to enter calibration mode.

2. Follow the screen prompts for calibration.

3. The calibration can be saved or printed when complete.

4. Informationforthelastcalibrationcanbedisplayedonthemainmeasurementscreen.Refertothis

referenceguide,Chapter2ModuleOverview,DisplayInformationsectionformoreinformation.

Page 14: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

10 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

This page intentionally left blank.

Page 15: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

11Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 5 Temperature Calibration

TheATCtemperaturedisplayhasarelativeaccuracyof±0.1°C.TemperaturesensorsbuiltintoRDO/polarographic

DOprobeshavevaryingtemperatureaccuracies,usually±0.5°Cto±2°C.Use this function only if necessary. Since the

temperatureoffsetcalculatedduringthecalibrationisappliedtoallfuturetemperaturemeasurements,recalibrateifa

differentprobeisused.Fortemperaturecalibration,theprobeneedstobeconnectedtothemeterandthecalibration

solutionshouldhaveaknown,stabletemperature.ItisrecommendedthattwoNIST-traceablethermometersbeused

to measure and verify solution temperature.

1. Pressf3 (setup) key.

2. Usethearrowkeystohighlightthechannelthatthetemperatureprobeisconnectedto.

Pressf3 (select).

3. UsethearrowkeystohighlightTemperatureandpressf3 (select).

4. UsethearrowkeystohighlighttheappropriateTemperatureCaloptionandpressf3 (select).

5. Followtheonscreenprompts.

Page 16: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

12 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

This page intentionally left blank.

Page 17: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

13Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 6 Measurement

Aftersuccessfulcalibration,thesamplescanbetested.

1. RinsetheRDOorDOprobewithdeionizedwater,blotdrywithlint-freetissue.

2. PlaceRDO/DOprobe,conductivitycellifrequired,andoptionalstirrerinfirstsample.

3. IfthemeterisinAutoReadmode(meterdefault)pressmeasure (esc) (1)key.Ifthemeterisin

continuousortimedreadmode,themeterwillimmediatelystarttakingreadings.

4. IfinAutoReadmodeanddatalogisenabled(default),thereadingwillautomaticallybestoredwhen

the“AR”appears.Ifincontinuousreadmodeanddatalogisenabled,presslog/print (0) key to store

intothemeter’smemory.

Note:Toenableordisabledatalogging,presssetupandselectInst.Settings.

Page 18: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

14 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

This page intentionally left blank.

Page 19: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

15Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 7 Methods

The meter stores up to ten methods per channel with the last calibration data stored for each method.

Use the channel setup to access and modify methods.

1. Pressf3 (setup) key.

2. Usethearrowkeystohighlighttheappropriatechannel.Pressf3 (select).

3. UsethearrowkeystohighlightMethodandpressf3 (select).

4. Themethodscanbeloaded,copied,savedandeditedbyfollowingtheonscreenprompts.

Page 20: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

16 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

This page intentionally left blank.

Page 21: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

17Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 8 Data Storage and Review

Data Storage Settings

Measurement Read Type

The measurement read type determines when the meter sends measurements to outputs as selected in the setup menu

and when properly connected to the meter. These are set individually for each channel. Measurement Read Type is set

perchannelinthesetupmenu.Pleasereferchapter3SetupMenusinthisreferenceguideformoreinformation.

Datalog, Computer and Printer Export Settings

OrionVERSASTARMetershavea2000-datalogcapability.Tomakeyourselection:

1. Frommeasurementmode,presssetup/3.

2. PressthearrowkeystohighlightInst.Settings.Presssetup/3.

3. Usethearrowkeys,f2 (page) key and f3 (select) key to:

a. set the communication protocol

b.turnonprinting

c. select the print format

d.turnondatalogging

Formoreinformationonthesesettings,refertotheOrionVERSASTARReferenceGuide,Chapter3InstrumentSettings

toturnonDataLogtoenabledatastorage.

Datalog Review

This option allows for stored data to be:

• reviewed

• printedindividually,inagrouporall

• deleted

• analyzedwithgeneralstatisticalinformation

Thedatalogwillreflectascreencaptureofalltheinformationatthetimeofdatalogging.

Toreviewdataloginformation

• Presslog view (9) key

• Pressf3 (select) key

• Followpromptstoreviewinformation

Page 22: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

18 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Example Datalog View

Calibration Review

Thisfeatureallowsforreviewandprintingofeachofthelast10calibrationsforeachparameterofthemeter.

The information will match the information recorded for that calibration.

To review calibration information

1. Presslog view (9) key.

2. Pressright arrow (6)keytoselectCalLoginformationanddown arrow (8)tohighlightthe

desired parameter.

3. Press f3 (select) key and follow the prompts.

Example Calibration Log Display

Page 23: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

19Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Chapter 9 Customer Services

Troubleshooting Guide

Assistance

Aftertroubleshootingallcomponentsofyourmeasurementsystem,contactTechnicalSupport.WithintheUnitedStatescall1.800.225.1480andoutsidetheUnitedStatescall978.232.6000orfax978.232.6031.InEurope,theMiddleEastandAfrica,contactyourlocalauthorizeddealer.Forthemostcurrentcontactinformation,orthelatestapplicationandtechnicalresourcesforThermoScientificOrionproducts,visitwww.thermoscientific.com/water.

Warranty and Registration

Toregisteryournewmeterandforthemostcurrentwarrantyinformation,visitwww.thermoscientific.com/water.

WEEE Compliance

ThisproductisrequiredtocomplywiththeEuropeanUnion’sWasteElectrical&ElectronicEquipment(WEEE)Directive2002/96/EC.Itismarkedwiththesymbolabove. ThermoFisherScientifichascontractedwithoneormorerecycling/disposalcompaniesineachEUMemberStateandthisproductshouldbedisposedoforrecycledthroughthem.Furtherinformationoncompliancewiththesedirectives,therecyclersinyourcountry,andinformationonThermoScientificOrionproductsthatmayassistthedetectionofsubstancessubjecttotheRoHSDirectiveareavailableatwww.thermoscientific.com.

Error Recommended Actions

Measurement is flashing9999andoverrangeorunderrange.

Measurementisoutsidetheallowablemeasurementrange.Checkelectrodeconnectionandsettingsinthesetupmenu.Cleantheelectrodeaccordingtotheelectrodeuserguideandrecalibratetheelectrodewithnewbuffersorstandards. Iftheerrorcontinues,performthemeterselftestinthediagnosticsmenu.

The measurement freezesandwill notchange.

ThemeterisinAutoReadmode.(ARiconappearssolidonthedisplay.)Press measure (esc)/1keytotakeanewreadingorgotosetuptochangethereadtype.

Page 24: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

20 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Declaration of Conformity

Manufacturer: Thermo Fisher Scientific Inc .

Address: Ayer Rajah Crescent Blk55#04-16/24 Singapore139949 Singapore

Hereby declares that the following products:Orion VERSA STAR Advanced Electrochemistry Benchtop Meters with Modules are rated 100to240VAC,50/60Hz,0.5A.

Equipment Class:

Measurement,controlandlaboratoryOrion Star A-series meters are EMC Class A

Conforms with the following directives and standards:

EN61326-1:2006 Electromagnetic Compatibility (EMC Directive) Electricalequipmentformeasurement, controlandlaboratoryuse-EMCrequirements

EN61010-1:2001 Safety StandardsUL61010-1:2004 Safetyrequirementsforelectricalequipmentformeasurement,CAN/CSA C22 .2 No . 61010-1-04 controlandlaboratoryuse-generalrequirements

CheowKwangChan QA/RegulatoryManager

PlaceandDateofIssue: 12December,2011 Singapore

Page 25: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

21Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

RDO/DO/Temperature Module Specifications

NOTE:Wereservetherighttomakeimprovements.Specificationssubjecttochangewithoutnotice.

RDO/DO Module (VSTAR-RD) SpecificationsDissolved Oxygen

Concentration-PolarographicRange 0to90mg/LResolution 0.01,0.1Relative Accuracy ±0.2

%Saturation-PolarographicRange 0to600Resolution 0.1,1Relative Accuracy ±2%

Concentration - RDORange 0to50mg/LResolution 0.01,0.1Relative Accuracy

±0.1mg/Lupto8mg/L; ±0.2mg/Lfrom8to20mg/L; 10%ofreading,20to50mg/L

%Saturation-RDORange 0to500Resolution 0.1,1

Relative Accuracy ±2%upto200%saturation; ±10%ofreadingabove200%saturation

Automatic Barometric PressureCorrection

450.0to850.0mmHg

Salinity Factor Correction Automatic when used with conductivity module and probeormanual;0to45ppt

Calibration Features

Watersaturatedair,airsaturatedwater, manual(Winkler)andzeropoint

ProbeType PolarographicorRDOTemperature Range 0to50°C,32to122ºF

Resolution 0.1Relative Accuracy ±0.1Offset Calibration 1 point

Input 9-pinmini-DIN

Page 26: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

22 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Ordering Information

CML # DescriptionVSTAR00 Orion VERSA STAR Benchtop Meter

-Electrodearmwithredesignedholder - Universal power adapter

VSTAR10 Orion VERSA STAR pH Benchtop Meter Set -OrionVERSASTARmeterwithpH/temperaturemodule -Electrodearmwithredesignedholder - Universal power adapter

VSTAR12 Orion VERSA STAR pH Benchtop Meter Kit -OrionVERSASTARmeterwithpH/temperaturemodule -8302BNUMDOrionROSSTriodepH/ATCprobe -096019OrionStarstirrerprobe -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -Electrodearmwithredesignedholder - Universal power adapter

VSTAR20 Orion VERSA STAR Conductivity Benchtop Meter Set - Orion VERSA STAR meter - Orion VERSA STAR conductivity/temperature module -Electrodearmwithredesignedholder - Universal power adapter

VSTAR22 Orion VERSA STAR Conductivity Benchtop Meter Kit - Orion VERSA STAR meter - Orion VERSA STAR conductivity/temperature module -013005MDOrionDuraProbeconductivitycell,K=0.475 -011007conductivitystandard,1413µS,5x60mL -Electrodearmwithredesignedholder - Universal power adapter

VSTAR30 Orion VERSA STAR RDO/Dissolved Oxygen Benchtop Meter Set - Orion VERSA STAR meter -OrionVERSASTARdissolvedoxygen/temperaturemodule -Electrodearmwithredesignedholder - Universal power adapter

VSTAR32 Orion VERSA STAR RDO/Dissolved Oxygen Benchtop Meter Kit - Orion VERSA STAR meter -OrionVERSASTARdissolvedoxygen/temperaturemodule -083005MDOrionpolarographicDOprobe -080017calibrationsleeveforDOprobe -080513DOprobemaintenancekit -970802BODfunnel/stirrer -080360BODadapter -Electrodearmwithredesignedholder - Universal power adapter

Page 27: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

23Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

CML # DescriptionVSTAR40A Orion VERSA STAR pH/ISE Benchtop Meter Set

- Orion VERSA STAR meter -OrionVERSASTARpH/ISE/temperaturemodule -Electrodearmwithredesignedholder - Universal power adapter

VSTAR40A2 Orion VERSA STAR pH/ISE Benchtop Meter Kit - Orion VERSA STAR meter -OrionVERSASTARpH/ISE/temperaturemodule -8102BNUWPOrionROSSUltrapHelectrode -096019OrionStarstirrerprobe -927007MDATCprobe,stainlesssteel -ROSSsolutionkit(475mLeachofpH4,7and10buffers;storagesolution; cleaningsolution;andpHelectrodestoragebottle) -Electrodearmwithredesignedholder - Universal power adapter

VSTAR40B Orion VERSA STAR Benchtop Meter with Two pH/ISE Modules - Orion VERSA STAR meter -TwoOrionVERSASTARpH/ISE/temperaturemodules -Electrodearmwithredesignedholder - Universal power adapter

VSTAR40B2 Orion VERSA STAR Benchtop Meter with Two pH/ISE Modules Kit - Orion VERSA STAR meter -TwoOrionVERSASTARpH/ISE/temperaturemodules -8102BNUWPOrionROSSUltrapHelectrode -096019OrionStarstirrerprobe -927007MDATCprobe,stainlesssteel -ROSSsolutionkit(475mLeachofpH4,7and10buffers;storagesolution; cleaningsolution;andpHelectrodestoragebottle) -Twoelectrodearmswithredesignedholders - Universal power adapter

VSTAR50 Orion VERSA STAR pH/Conductivity Benchtop Meter Set - Orion VERSA STAR meter -OrionVERSASTARpH/temperaturemodule - Orion VERSA STAR conductivity/temperature module -Electrodearmwithredesignedholder - Universal power adapter

VSTAR52 Orion VERSA STAR pH/Conductivity Benchtop Meter Kit -OrionVERSASTARmeterwithpH/temperaturemodule - Orion VERSA STAR conductivity/temperature module -8157BNUMDOrionROSSUltraTriodepH/ATCprobe -013005MDOrionDuraProbeconductivitycell,K=0.475 -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -011007conductivitystandard,1413µS,5x60mL -Electrodearmwithredesignedholder - Universal power adapter

Page 28: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

24 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

CML # DescriptionVSTAR80 Orion VERSA STAR pH with LogR Benchtop Meter Set

- Orion VERSA STAR meter -OrionVERSASTARpH/temperaturewithLogRtechnologymodule -Electrodearmwithredesignedholder - Universal power adapter

VSTAR82 Orion VERSA STAR pH Benchtop Meter Kit - Orion VERSA STAR meter -OrionVERSASTARpH/temperaturewithLogRtechnologymodule -8172BNXOrionROSSSure-FlowpHelectrode -927007MDATCprobe,stainlesssteel -ROSSsolutionkit(475mLeachofpH4,7and10buffers;storagesolution; cleaningsolution;andpHelectrodestoragebottle) -810007ROSSfillsolution,5x60mL -Electrodearmwithredesignedholder - Universal power adapter

VSTAR90 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen Benchtop Multiparameter Meter Set - Orion VERSA STAR meter -OrionVERSASTARpH/ISE/temperaturemodule - Orion VERSA STAR conductivity/temperature module -OrionVERSASTARdissolvedoxygen/temperaturemodule -Electrodearmwithredesignedholder - Universal power adapter

VSTAR91 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen Benchtop Multiparameter Meter Set - Orion VERSA STAR meter -TwoOrionVERSASTARpH/ISE/temperaturemodules - Orion VERSA STAR conductivity/temperature module -OrionVERSASTARdissolvedoxygen/temperaturemodule -Two096019OrionStarstirrerprobes -Twoelectrodearmswithredesignedholders - Universal power adapter

Page 29: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

25Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

CML # DescriptionVSTAR92 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen

Benchtop Multiparameter Meter Set - Orion VERSA STAR meter -OrionVERSASTARpH/ISE/temperaturemodule - Orion VERSA STAR conductivity/temperature module -OrionVERSASTARdissolvedoxygen/temperaturemodule -8157BNUMDOrionROSSUltraTriodepH/ATCprobe -013005MDOrionDuraProbeconductivitycell,K=0.475 -083005MDOrionpolarographicDOprobe -080017calibrationsleeveforDOprobe -080513DOprobemaintenancekit -970802BODfunnel/stirrer -080360BODadapter -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -011007conductivitystandard,1413µS,5x60mL -Electrodearmwithredesignedholder - Universal power adapter

VSTAR93 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen Ultimate Benchtop Multiparameter Meter Set - Orion VERSA STAR meter -OrionVERSASTARpH/ISE/temperaturemodule - Orion VERSA STAR conductivity/temperature module -OrionVERSASTARdissolvedoxygen/temperaturemodule -8157BNUMDOrionROSSUltraTriodepH/ATCprobe -013005MDOrionDuraProbeconductivitycell,K=0.475 -083005MDOrionpolarographicDOprobe -Two096019OrionStarstirrerprobes -080017calibrationsleeveforDOprobe -080513DOprobemaintenancekit -970802BODfunnel/stirrer -080360BODadapter -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -011007conductivitystandard,1413µS,5x60mL -Twoelectrodearmswithredesignedholders - Universal power adapter

Page 30: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

26 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

AccessoriesCML # DescriptionVSTAR-PH OrionVERSASTARpH/temperaturemoduleVSTAR-ISE OrionVERSASTARpH/ISE/temperaturemoduleVSTAR-CND Orion VERSA STAR conductivity/temperature moduleVSTAR-RD Orion VERSA STAR RDO/DO/temperature moduleVSTAR-LR OrionVERSASTARpH/temperaturewithLogRtechnologymoduleSTARA-BEA Benchtop electrode arm for Orion Star A-series metersSTARA-HB FreestandingbaseforusewithOrionStarA-seriesbenchtopelectrodearmSTARA-PWR Replacement universal power adapter for Orion VERSA STAR meters096019 Orion Star stirrer probe8102BNUWP OrionROSSUltrapHelectrode,refillable,glassbody8172BNWP OrionROSSSure-FlowpHelectrode,glassbody8156BNUWP OrionROSSUltrarefillablepHelectrode,epoxybody8157BNUMD OrionROSSTriodepH/ATCelectrode,refillable,epoxybody8302BNUMD OrionROSSTriodepH/ATCelectrode,refillable,glassbody8172BNX OrionROSSSure-FlowpHelectrode,glassbody8611BNWP OrionROSSsodiumcombinationISE927007MD OrionATCprobe,stainlesssteelbody927005MD OrionATCprobe,epoxybody013005MD OrionDuraProbeconductivitycell,4-cell,K=0.475013016MD Orionconductivitycell,flowthrough,K=0.1083005MD Oriondissolvedoxygenprobe,polarographic086030MD Orionauto-stirDO/BODprobe,polarographic087010MD OrionRDOopticalDOprobe,3mcable810199 OrionROSSall-in-onepHbufferkit910104 OrionpH4.01buffer,475mL910107 OrionpH7.00buffer,475mL910110 OrionpH4.01buffer,475mL011008 Orionconductivitystandard,100µS/cm,5x60mLbottles011007 Orionconductivitystandard,1413µS/cm,5x60mLbottles011006 Orionconductivitystandard,12.9mS/cm,5x60mLbottles1010001 Orion conductivity calibration resistor kit9512HPBNWP Orionhigh-performanceammoniaISE9512BNWP OrionstandardammoniaISE951007 Ammoniastandard,1000ppm,475mL951210 Low-levelammoniaISA951211 AmmoniaISA,475mL951209 FillsolutionforOrionhigh-performanceammoniaISE,60mL951202 FillsolutionforstandardammoniaISE,60mL951213 AmmoniaISEstoragesolution9609BNWP OrionfluoridecombinationISE040906 Fluoridestandard,1ppmwithTISABII,475mL040907 Fluoridestandard,2ppmwithTISABII,475mL040908 Fluoridestandard,10ppmwithTISABII,475mL940909 TISABII,1gallon8611BNWP OrionROSSsodiumcombinationISE

Page 31: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

27Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Appendix Advanced Features

Dissolved Oxygen Barometric Pressure Compensation

The Orion VERSA STAR Meter with RDO/DO/Temperature module systems have an internal barometer that is used forpressurecompensateddissolvedoxygenreadings.Themetercanalsousemanualbarometricpressurecompensationifdissolvedoxygenismeasuredwithasubmergedprobeorinapressurizedvessel.ThepressuremustbeenteredasmmHg.

1 mm Hg = 0.03937 inch Hg = 1.3332 hPa (mBar) = 0.01934 PSI.

Dissolved Oxygen Salinity Correction

Automaticsalinitycorrectionfordissolvedoxygenreadingsreportedasmg/LisavailableontheOrionVERSASTARMeterwithRDO/DO/TemperaturemodulesystemsifbothdissolvedoxygenandOrionVERSASTARConductivity/Temperature modules are used simultaneously. The meter uses the conductivity value read by the conductivity cell to calculatethesalinitycorrectionfactorandappliesthefactortodissolvedoxygenreadingsreportedasmg/L.

Manualsalinitycorrectioncanbeusedfordissolvedoxygenreadingsreportedasmg/LonallOrionVERSASTARMeter withRDO/DO/Temperaturemodulesystems.Themanualsalinitycorrectionfactormustbeenteredasppt(parts perthousand).

Conductivity at 20 °C (mS/cm)

Salinity Correction Value (ppt)

Conductivity at 20 °C (mS/cm)

Salinity Correction Value (ppt)

Conductivity at 20 °C (mS/cm)

Salinity Correction Value (ppt)

5 3 20 13 35 256 4 21 14 36 257 4 22 15 37 268 5 23 15 38 279 6 24 16 39 2810 6 25 17 40 2911 7 26 18 42 3012 8 27 18 44 3213 8 28 19 46 3314 9 29 20 48 3515 10 30 21 50 3716 10 31 22 52 3817 11 32 22 54 4018 12 33 23 56 4219 13 34 24

ThistablewascalculatedfromtheInternationalOceanographicTables,Vol.1,NationalInstituteofOceanographyofGreatBritain,Womley,Godaming,Surrey,EnglandandUnesco,Paris1971.

Page 32: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

28 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module

Notes

Page 33: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

This page intentionally left blank.

Page 34: Thermo Scientific Orion VERSA STAR™ RDO/DO ......2011/12/12  · 4 Thermo Scientific Orion VERSA STAR RDO/DO/Temperature Module Module Maintenance For routine maintenance, dust and

9 0 0 1 : 2 0 0 8

Water Analysis Instruments

North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 [email protected]

NetherlandsTel: (31) 033-2463887 [email protected]

India Tel: (91) [email protected]

Japan Tel: (81) [email protected]

ChinaTel: (86) 21-68654588 [email protected]

Singapore Tel: (65) [email protected]

Australia

Tel: (613) 9757-4300in Australia (1300) [email protected]/water

© 2011 Thermo Fisher Scientific Inc. All rights reserved.

68X592003 Rev. A 12-11