THE NOVEL LIVANOVA 3D RECHORD SEMIRIGID RING FACILITATES MITRAL VALVE REPAIR WITH ARTEFICIAL CHORDS: A MATCHED PAIR ANALYSIS Gábor Szabó 1,2 ; Raffaele De Simone 1 ; Ursula Tochtermann 1 ; Sabreen Mkalaluh 1 ; Matthias Karck 1 1 Department of Cardiac Surgery, University Hospital Heidelberg, Heidelberg, Germany; 2 Heart and Vascular Center, Semmelweis University, Budapest, Hungary, Background Replacement of diseased chordae with Gore-Tex sutures (neochordae) in patients with degenerative mitral valve insufficiency has become a standard technique used by surgeons in mitral valve repair with good long-term results. Nevertheless, determining the correct length of the neochordae has remained problematic. Moreover, the nature of the suture material and the tendency of the knot to slide may result in in difficulty in establishing the correct length of the neochordae. The novel Livanova (Sorin) 3D Rechord ring with an additional chordal guiding system allows a quick „semi-automatic“ sizing of the neochordae and prevents the „sliding“ of knots. We compared patients who underwent mitral valve repair with neochordae using either the novel Livanova 3D Rechord ring or Carpentier- Edwads (CE) Physio II ring in a a retrospective matched-pair analysis. Methods & Results A matched pair analysis of retrospectively collected data of patients underwent mitral valve repair between January 2014 –December 2016 was undertaken. 20 Patients received mitral valve repair using neochordae and Livanova 3D Rechord ring were matched with 20 patients with mitral valve repair with neochordae and implantation of CE Physio II ring. Pre- intra- and early postoperative as well as echocardiographic data were analyzed. All values are given as mean±SD, p<0.05 was considered statistically significant. Patients did not differ regarding age , gender, NYHA status, comorbidities, ICU and in-hospital stay as well as in-hospital, and 30-day mortality. Conclusions The use of he novel Livanova 3D Rechord ring is safe and provides excellent functional results. The chordal guiding system markedly reduces the time oft the procrdure by facilitating sizing and knotting and reducing the number of necessary neochorde to achive valve competence. Although the long-term durability of mitral repair with this type of semirigid annuloplastic ring warrants further validation, our current clinical data are encouraging. . Basic principles and surgical technique Representative echocardiographic findings Peak CKMB values CPB time Crossclamp time 1) Radermecker MA. ATS 2010;89:344-45, 2) www.livanova.sorin.com REF 2. Financial Disclosures: none REF 1. Post-OP 0 20 40 60 80 100 120 140 160 0 50 100 150 200 250 0 20 40 60 80 100 120 0 0,5 1 1,5 2 2,5 3 3,5 4 4,5 5 Number of neochordae P=0.0001 P=0.004 P=0.002 P=0.08 Pre-OP

Upload: others

Post on 04-Oct-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: THE NOVEL LIVANOVA 3D RECHORD SEMIRIGID RING …eposterkiosk.com/mitral17/ePosters/239-Szabo.pdf · using neochordae and Livanova 3D Rechord ring were matched with 20 patients with

THE NOVEL LIVANOVA 3D RECHORD SEMIRIGID RING FACILITATES MITRAL VALVE REPAIR WITH ARTEFICIAL CHORDS: A MATCHED PAIR ANALYSIS

Gábor Szabó1,2; Raffaele De Simone1; Ursula Tochtermann1; Sabreen Mkalaluh1; Matthias Karck1

1Department of Cardiac Surgery, University Hospital Heidelberg, Heidelberg, Germany; 2Heart and Vascular Center, Semmelweis University, Budapest, Hungary,

BackgroundReplacement of diseased chordae with Gore-Tex sutures (neochordae) in patients with degenerative mitral valve insufficiency has become a standard technique used by surgeons in mitral valve repair with good long-term results. Nevertheless, determining the correct length of the neochordae has remained problematic. Moreover, the nature of the suture material and the tendency of the knot to slide may result in in difficulty in establishing the correct length of the neochordae. The novel Livanova (Sorin) 3D Rechord ring with an additional chordal guiding system allows a quick „semi-automatic“ sizing of the neochordae and prevents the „sliding“ of knots. We compared patients who underwent mitral valve repair with neochordae using either the novel Livanova 3D Rechord ring or Carpentier-

Edwads (CE) Physio II ring in a a retrospective matched-pair analysis.

Methods & ResultsA matched pair analysis of retrospectively collected data of patients underwent mitral valve repair between January 2014 –December 2016 was undertaken. 20 Patients received mitral valve repair

using neochordae and Livanova 3D Rechord ring were matched with 20 patients with mitral valve repair with neochordae and implantation of CE Physio II ring. Pre- intra- and early postoperative as well as echocardiographic data were analyzed. All values are given as mean±SD, p<0.05 was considered statistically significant. Patients did not differ regarding age , gender, NYHA status,

comorbidities, ICU and in-hospital stay as well as in-hospital, and 30-day mortality.

ConclusionsThe use of he novel Livanova 3D Rechord ring is safe and provides excellent functional results. The chordal guiding system markedly reduces the time oft the procrdure by facilitating sizing and

knotting and reducing the number of necessary neochorde to achive valve competence. Although the long-term durability of mitral repair with this type of semirigid annuloplastic ring warrants further validation, our current clinical data are encouraging. .

Basic principles and surgical technique Representative echocardiographic findings Peak CKMB values

CPB time

Crossclamp time

1) Radermecker MA. ATS 2010;89:344-45, 2) www.livanova.sorin.com

REF 2.

Financial Disclosures: none

REF 1.

Post-OP

0

20

40

60

80

100

120

140

160

1 2

0

50

100

150

200

250

1 2

0

20

40

60

80

100

120

1 2

0

0,5

1

1,5

2

2,5

3

3,5

4

4,5

5

1 2

Number of neochordaeP=0.0001

P=0.004

P=0.002

P=0.08

Pre-OP