PRODUCT DATASHEET: atmosFIR CEM Call us on: +44(0)1270 872 000 www.protea.ltd.uk Protea Limited. 10 Prosperity Court, Middlewich, Cheshire, CW10 0GD UK. in measurement solutions technology leadership Protea PD-701 : November 2016 atmosFIR CEM is a complete emissions monitoring system incorporating the atmosFIR 19” rack FTIR analyser. Providing a completely hot and wet emissions measurement, atmosFIR CEM is ideal for monitoring of multiple gases from applications such as incineration and power. A standard gas suite of NO, NO 2 , N 2 O, SO 2 , NH 3 , HCl, CH 4 , O 2 , CO 2 and H 2 O and more is provided, but with the powerful FTIR analyser, further gases can be added easily in software. At the heart of system is the high-resolution, robust and proven FTIR spectrometer offering high signal throughput, low-noise and long lifetime of components. atmosFIR has been developed to incorporate the latest improvements and advantages in technology, The atmosFIR FTIR analyser contains an in-built sampling system and is designed for ppm- level emissions monitoring. The analyser can be used as a portable or bench-top unit, as well as part of the installed CEM system. For emissions applications, atmosFIR CEM contains management of the heated sampling system from within the system – heated lines, probes – as well as comprehensive span gas options, including 6 span valves that can be used to calibration the analyser directly or “to probe”. Protea’s PAS-Pro operating software not only collects and analyses the infrared spectra with the latest in analytical algorithms (with no limit on the number of gases that can be measured) but also manages the complete automation of the system. Modbus Serial, Modbus TCP/IP and OPC Server data exchange is provided as standard, and the monitoring results are reported both Raw and Corrected in real-time. atmosFIR CEM is a complete multigas analyser system using FTIR technology. [ Waste Incineration Emissions monitoring [ Combustion emissions from power generation [ Core FTIR module available for integration [ Meets the requirements of EN 15267-3, QAL1 and EN14181. Certification due Q2 2017 FTIR Specifications Double-pivot interferometer with increased robustness. Permanently aligned optics, giving repeatable measurements and high light throughput. The scanning mechanism has a lifetime guarantee. Resolution 1cm -1 ,2cm -1 , 4cm -1 , 8cm -1 variable resolutions. 1cm -1 for standard emissions monitoring Optics: Zinc Selenide beamsplitter (non-hygroscopic) Spectral Range: 485 - 8500cm -1 Reference laser: Solid state laser (no scheduled maintenance required). Long lifespan (10 years) compared with HeNe laser Source: Mid-IR source, with electronic stabilization for long lifespan Detector: DTGS with signal sampling at 24-bit ADC Sample Cell Materials: Ni-coated Al cell. Proprietary alloy mirror substrate with multi-layer coating. Volume: 300ml Pathlength: 4.2m standard pathlength Temperature: 180°C FTIR Sampling system: Heated pre-cell filter for extra protection against dust Zirconia oxygen sensor for parallel O2 measurement Automated Nitrogen Purge Valve atmosFIR CEM

Upload: others

Post on 08-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: technology leadership in measurement solutio ns atmosFIR CEM cem... · 2017-02-01 · PRODUCT DATASHEET: atmosFIR CEM Call us on: 44(0)1270 872 000 Protea Limited. 10 Prosperity Court,

PRODUCT DATASHEET: atmosFIR CEM

Call us on: +44(0)1270 872 000www.protea.ltd.ukProtea Limited. 10 Prosperity Court, Middlewich, Cheshire, CW10 0GD UK.

in measurement solutionstechnology leadershipProtea

PD-701 : November 2016

atmosFIR CEM is a complete emissions monitoring system incorporating the atmosFIR 19” rack FTIR analyser. Providing a completely hot and wet emissions measurement, atmosFIR CEM is ideal for monitoring of multiple gases from applications such as incineration and power. A standard gas suite of NO, NO2, N2O, SO2, NH3, HCl, CH4, O2, CO2 and H2O and more is provided, but with the powerful FTIR analyser, further gases can be added easily in software.

At the heart of system is the high-resolution, robust and proven FTIR spectrometer offering high signal throughput, low-noise and long lifetime of components. atmosFIR has been developed to incorporate the latest improvements and advantages in technology, The atmosFIR FTIR analyser contains an in-built sampling system and is designed for ppm-level emissions monitoring. The analyser can be used as a portable or bench-top unit, as well as part of the installed CEM system.

For emissions applications, atmosFIR CEM contains management of the heated sampling system from within the system – heated lines, probes – as well as comprehensive span gas options, including 6 span valves that can be used to calibration the analyser directly or “to probe”.

Protea’s PAS-Pro operating software not only collects and analyses the infrared spectra with the latest in analytical algorithms (with no limit on the number of gases that can be measured) but also manages the complete automation of the system. Modbus Serial, Modbus TCP/IP and OPC Server data exchange is provided as standard, and the monitoring results are reported both Raw and Corrected in real-time.

atmosFIR CEM is a complete multigas analyser system using FTIR technology.

[ Waste Incineration Emissions monitoring

[ Combustion emissions from power generation

[ Core FTIR module available for integration

[ Meets the requirements of EN 15267-3, QAL1 and EN14181. Certification due Q2 2017

FTIR Specifications

Double-pivot interferometer with increased robustness. Permanently aligned optics, giving repeatable measurements and high light throughput. The scanning mechanism has a lifetime guarantee.

Resolution 1cm-1,2cm-1, 4cm-1, 8cm-1 variable resolutions. 1cm-1 for standard emissions monitoring

Optics: Zinc Selenide beamsplitter (non-hygroscopic)

Spectral Range: 485 - 8500cm-1

Reference laser: Solid state laser (no scheduled maintenance required). Long lifespan (10 years) compared with HeNe laser

Source: Mid-IR source, with electronic stabilization for long lifespan

Detector: DTGS with signal sampling at 24-bit ADC

Sample Cell Materials: Ni-coated Al cell. Proprietary alloy mirror substrate with multi-layer coating.Volume: 300mlPathlength: 4.2m standard pathlengthTemperature: 180°C

FTIR Sampling system: Heated pre-cell filter for extra protection against dustZirconia oxygen sensor for parallel O2 measurementAutomated Nitrogen Purge Valve

atmosFIR CEM

Page 2: technology leadership in measurement solutio ns atmosFIR CEM cem... · 2017-02-01 · PRODUCT DATASHEET: atmosFIR CEM Call us on: 44(0)1270 872 000 Protea Limited. 10 Prosperity Court,

PRODUCT DATASHEET: atmosFIR CEM

Call us on: +44(0)1270 872 000www.protea.ltd.ukProtea Limited. 10 Prosperity Court, Middlewich, Cheshire, CW10 0GD UK.

in measurement solutionstechnology leadershipProtea

PD-701 : November 2016

Measurements for atmosFIR CEM

atmosFIR CEM can be provided with core atmosFIR FTIR gas analyser set-up for any number of different process or emissions applications. Contact Protea for specific gas analysis requirements.The standard incineration emissions application includes the following gases and ranges

Typical Response Time 120secs at 1cm-1 resolution. (T90)

Linearity <2% range Repeatability (σ) <1% range Temperature Drift <2% range per 10K

Standard Combustion Emissions Model

Component Ranges / mg/m3 Lower detection Limit (LDL) / mg/m3

Component Ranges / mg/m3 Lower detection Limit (LDL) / mg/m3

CO 0-75; 0-1000 0.6 CH4 (Methane) 0-50; 0-1000 0.1

NO 0-200; 0-600 1.0 C2H6 (Ethane) 0-50; 0-1000 0.1

NO2 0-200; 0-600 0.6 C3H8 (Propane) 0-50; 0-1000 0.8

N2O 0-50; 0-400 0.4 C2H4 (Ethene) 0-50; 0-1000 0.4

SO2 0-75; 0-1000 0.6 HCHO (Formaldehyde) 0-20; 0-100 0.2

NH3 0-15; 0-50 0.1 TOC (Indication only) 0-50; 0-1000 -

HCl 0-15; 0-100 0.2 H2O 0-40% 0.02%

HF 0-15; 0-50 0.2 CO2 0-20% 0.005%

High Resolution FTIR Analysis – no limit to number of gases that can be measured via multigas analysis of full IR spectrum.

Page 3: technology leadership in measurement solutio ns atmosFIR CEM cem... · 2017-02-01 · PRODUCT DATASHEET: atmosFIR CEM Call us on: 44(0)1270 872 000 Protea Limited. 10 Prosperity Court,

PRODUCT DATASHEET: atmosFIR CEM

Call us on: +44(0)1270 872 000www.protea.ltd.ukProtea Limited. 10 Prosperity Court, Middlewich, Cheshire, CW10 0GD UK.

in measurement solutionstechnology leadershipProtea

PD-701 : November 2016

Installation and Utilities

Cabinet Approx. 1800(h) x 600(w) x 800(d) plus 100mm plinth for installation

Weight 250kg

Supply and Power Consumption

1 x 13A for cabinet. Probe powered locally at stack2.5kW (including 10m heated line and sample probe)

Ambient Temperature 5°C < T < 30°C without air conditioner on cabinet

Gas Utilities Instrument Air @ 4 bar g for eductor. Dry, oil and particle free.99.999% N2 @ 0.5 bar g for instrument zero gas. Automatic zero every 24 hours.Span gases @ 1 bar g for system checking

atmosFIR CEM Sampling Specifications

Sampling Rate 2.5lpm, controlled via with venturi orifice

Sample Control Heated eductor post-FTIR

Sample Line 4/6mm PTFE core heated to 180°C. Sample line temperature alarm

Sample Probe 2um PTFE filter heated to 180°C. Sample probe temperature alarm

Span Gases 6 span gas valves, assignable to various gases and concentration levels

Span Gas Delivery “Direct” to analyser or “Span to Probe” selectable

Page 4: technology leadership in measurement solutio ns atmosFIR CEM cem... · 2017-02-01 · PRODUCT DATASHEET: atmosFIR CEM Call us on: 44(0)1270 872 000 Protea Limited. 10 Prosperity Court,

PRODUCT DATASHEET: atmosFIR CEM

Call us on: +44(0)1270 872 000www.protea.ltd.ukProtea Limited. 10 Prosperity Court, Middlewich, Cheshire, CW10 0GD UK.

in measurement solutionstechnology leadershipProtea

PD-701 : November 2016

Supplier:

[ Complete Emissions Supplied by Protea[ FTIR and Sampling Control integrated in software[ Expandable system – [ Multiple Sampling Points [ No limit on number of gases