1 Aberdeen Test Center June 22, 2004 Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen (ATC) 21 st Transducer Workshop Aberdeen Test Center 2 Two Decades of History Calibration differences between government and / or contractor installations have been under scrutiny for years. In a 1983 report (Ref. #1) the author concluded there was a significant difference between calibration results with the probable cause being calibration methods. In a 1993 report (Ref. #2) the author concluded that there was still a significant difference in the calibration results obtained from 5 different facilities and listed the probable causes. +/-1.25% Bad Transducer +/-0.4% Transducer +/-0.7% Calibration Method +/-0.2% Calibration Electronics +/-0.2% Pressure Generation Estimated Effect Probable Cause

Upload: others

Post on 25-May-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

1

Aberdeen Test Center

June 22, 2004

Study of Calibration Techniques

A Demonstration of a Transducers Response to High Pressure Dynamic Loads

Presented By:Ken McMullen (ATC)

21st Transducer Workshop

Aberdeen Test Center 2

Two Decades of History

• Calibration differences between government and / or contractor installations have been under scrutiny for years.• In a 1983 report (Ref. #1) the author concluded there was a significant

difference between calibration results with the probable cause being calibration methods.

• In a 1993 report (Ref. #2) the author concluded that there was still a significant difference in the calibration results obtained from 5 different facilities and listed the probable causes.

+/-1.25%Bad Transducer

+/-0.4%Transducer

+/-0.7%Calibration Method

+/-0.2%Calibration Electronics

+/-0.2%Pressure Generation

Estimated Effect Probable Cause

Page 2: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

2

Aberdeen Test Center 3

Two Decades of History (cont.)

• In 2002 a calibration discrepancy was identified between Tourmaline, Quartz Transducers and Copper Crushers.

• So based on the 2002 discrepancies and the information from the 1983 and 1992 reports this 2003 investigation set out to further understand the possible causes of calibration error that still exist two decades later.

• Testing was conducted at:• YTC’s Transducer Manufacturing and Calibration facility • ATC’s ISO 17025 accredited Transducer Calibration Laboratory

(TCL)

Aberdeen Test Center 4

Static Pressure Primary Standard

ATC’s Pseudo-Dynamic Negative Calibration Method or “Pressure Release Method”.

Page 3: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

3

Aberdeen Test Center 5

Static Pressure Secondary Standard

YTC’s Pseudo-Dynamic Positive Calibration Method or “Positive Pressure Step Method”

Aberdeen Test Center 6

Pressure Generation Results

+0.14%100,024100,000100,04099,580100,000

+0.10%80,02880,00080,03079,68080,000

+0.07%59,98060,00060,02059,74060,000

+0.04%39,98040,00040,00039,80040,000

-0.09%20,00020,00019,99019,84020,000

ATC / YTCMangannon Cell

DH Primary

Std.

Mangannon Cell

Strain CellDH Primary

Std.

DifferenceATCYTC

Primary / Secondary Calibration Comparison for ATC and YPG

Page 4: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

4

Aberdeen Test Center 7

Calibration Electronics

• Minimum Transducer Calibration Electronics• Signal Conditioning (Charge Amplifier)

• Calibration Capacitor (Weakest Link)

• Readout Instrumentation (Simple – Sophisticated)

• YTC uses the Kistler 504 Series Charge Amplifier, National Instruments A/D

• ATC uses the Kistler 5011Series Charge Amplifier, Data Com A/D

Aberdeen Test Center 8

Calibration Capacitor Results

-0.19%25078 pC25125.12 pC3

-0.26%25061 pC25126.12pC1

Diff.MeasuredCalculated Channel No.

Pre-Instrumentation Check (Using Indicated Cap Values)

-0.12%9,988pF9,976pF1025C / 3

-0.26%10,014pF9,987pF3502 / 1

Diff.Measured Value

RecordedValue

Serial No. / Channel No.

Calibration Capacitor Check

0.02%25123.0pC25118.11pC3

0.08%25140.0pC25119.11pC1

Diff.MeasuredCalculatedChannel No.

Post-Instrumentation Check (Using Corrected Cap Values)

Induced Error

Calibration Correction

Cap. Error

Page 5: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

5

Aberdeen Test Center 9

YPG Calibration Methods

Delta Meas. Loc. = 12.3mS

1024cts71.86mS

1200-1210cts84.0-84.7mS900-910cts

63.0-63.7mS

800cts56.0mS

1400cts98.0mS

Measurement Locations

Screen Display Range

Pressure Step Wave

YPG DAQ Specifications:12 Bit A/DPre-trigger 1kPost-trigger 1kSample Rate 14285Hz, 70us/sampleTrigger point 1024cts, 71.86ms

Aberdeen Test Center 10

ATC Calibration Method

Trigger Meas. Loc.

1024cts655.56mS

0 - 500cts0 -320.0mS

1548 - 2047cts0.991 – 1.31S

Measurement Location (avg.)

Screen Display Range

Pressure Step Wave

ATC DAQ Specifications:12 Bit A/DPre-trigger 1kPost-trigger 1kSample Rate 1,562Hz, 640us/sampleTrigger point 1024cts, 327.68ms

Page 6: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

6

Aberdeen Test Center 11

Calibration Method Results

• A +/-0.7% error was attributed to the use of different calibration methods based on the 1992 report.

• The data in this report shows that the calibration method has little or no effect on the results when the transducer produces an ideal flat response.

• No transducer produces a perfect flat response, so the calibration method, the data acquisition approach and the data reduction are critical.

Aberdeen Test Center 12

Ideal vs. Practical Transducer Response

80x103

60

40

20

0

Pres

sure

(kPs

i)

-2000 -1500 -1000 -500 0 500 1000 1500 2000Time (mSec.)

Transducer Response

Ideal Transducer Response “A”Practical Response Transducer “B”Measurement Far from TransitionMeasurement Near Transition

Page 7: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

7

Aberdeen Test Center 13

Transducer Types

• Tourmaline Crystals• YTC / Kistler E30MA (0-120kPsi)• YTC RP120 (0-30kPsi)

• Cusp / Overshoot Anomaly• Time Constant / Drift Issues

• Quartz Crystals• Kistler 6213B (0-120kPsi)• Kistler 6205AM01 (0-100kPsi)

• Torque / Mount Issues

Aberdeen Test Center 14

Tourmaline CrystalsE30MA Overshoot or “Cusp” Phenomenon

80.0x103

79.5

79.0

78.5

78.0

Pres

sure

(Psi

)

0.50.40.30.20.10.0-0.1

Time (s)

Calibration Test 1 Location: YPG Date: 07-10-2003

E30MA SN:611

Overshoot

Positive Pressure Step Method

Page 8: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

8

Aberdeen Test Center 15

Tourmaline CrystalsE30MA Overshoot or “Cusp” Phenomenon

Pressure Release Method

-80.0x103

-79.5

-79.0

-78.5

-78.0Pr

essu

re (P

si)

6005004003002001000-100

Time (mSec.)

Calibration Test Location: ATC Date: 04-25-2003

Overshoot

E30MA SN:611

Aberdeen Test Center 16

Tourmaline CrystalsRP120 Overshoot or “Cusp” Phenomenon

30.0x103

29.5

29.0

28.5

28.0

27.5

27.0

26.5

26.0

Pres

sure

(Psi

)

1086420

Time (s)

RP120 Dynamic Pressure Increase Test Location: YPG Date: 10-1-2003

RP120 Sn:1006

Peak = 849.6 Psi or 2.9% Above Steady State

Time Constant Drift

Steady State

Positive Pressure Step Method

Page 9: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

9

Aberdeen Test Center 17

Tourmaline CrystalsRP120 Overshoot or “Cusp” Phenomenon

Pressure Release Method

2000

1800

1600

1400

1200

1000

800

600

400

200

0

-200

-400

-600

-800

-1000

Pres

sure

(Psi

)

108642

Time (s)

RP120 Dynamic Release Test Location: YPG Date: 09-30-2003

Peak = -623Psi or2.0% Below Baseline

Time Constant Drift

RP120 Sn: 24

Baseline

Aberdeen Test Center 18

Tourmaline Crystals

• Must be Calibrated Dynamically• To Account for:

• Overshoot / Cusp• Short Time Constant• Drift

• The calibration method is critical• The calibration method should mimic the

measurement to be made.

• Calibration measurement locations are critical

Page 10: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

10

Aberdeen Test Center 19

Quartz CrystalsMount & Torque Sensitivity

• Problem Resolution:• Mounting Surface Preparation

• Proper Thread Depth • Insure Sealing Surface is:

• Perfectly Flat• Perfectly Clean

• Firing Adapters / Calibrate Assembly• Triple Torque

• Torque Transducer / Remove Torque (3 Times)

• Thermal Protective Shield• Causes Transducer to Indicate a Lower Pressure

Aberdeen Test Center 20

Kistler 6213B vs. E30MA Wave Shape

82.0x103

81.5

81.0

80.5

80.0

79.5

79.0

78.5

78.0

77.5

77.0

Pres

sure

(Psi

)

0.50.40.30.20.10.0-0.1

Time (s)

Calibration Test Test #: 2 Location: ATC Date: 07-10-2003

6213B Undershoot

E30MA Overshoot

E30MA SN:611 6213B SN:626578

Page 11: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

11

Aberdeen Test Center 21

Quartz CrystalsMount & Torque Sensitivity

Firing

Adapter

Assembly

Aberdeen Test Center 22

Kistler 6213BThermal Protection Sensitivity

-0.5479,34279,780NegativeYes

-0.4479,65280,008PositiveYes

-0.4078,94279,305PositiveYes

+0.0280,02380,000NegativeNo

+0.1879,44779,100PositiveNo

+0.0279,12279,104PositiveNo

Percent Difference

(%)

Measured Pressure

(PSI)

“True Pressure”

(PSI)

Dynamic Pressure Step

Thermal Protection

Kistler 6213B SN:626578

Page 12: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

12

Aberdeen Test Center 23

Other Issues

• Transducer Durability• Since 1990 ATC has had 225 E30MA’s

• To Date 30 Remain in Inventory• On average 15 E30MA transducers are removed from

inventory each year.

• Since 1990 ATC has had 13 Kistler 6213B’s• To Date 12 Remain in Inventory.

Aberdeen Test Center 24

Other Issues (cont.)

P051-602.28%1.35%0.06%0.06%1.00%4-6213B’s (From Shelf)

P011-200.23%0.17%0.03%0.04%0.05%Re-Calibrate (In Place)

P001-100.86%1.10%0.05%0.05%0.28%4-6213B’s (From Shelf)

PH41-601.81%1.07%0.06%0.07%0.65%4-E30MA’s (From Shelf)

PH11-200.48%0.58%0.12%0.04%0.05%Re-Calibrate (In Place)

PH01-101.93%1.68%0.12%0.06%0.63%4-E30MA’s (From Shelf)

RoundNo.

Extreme Spread

WorstDeviation

Round to Round Var.

Within Gage Var.

Gage to Gage Var.

Configuration

Sensitivity Level Shift Test (10 repetitions at each configuration)

Page 13: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

13

Aberdeen Test Center 25

Other Issues (cont.)

P051-602.28%1.35%0.06%0.06%1.00%4-6213B’s (From Shelf)

P041-502.06%1.64%0.06%0.05%0.80%Changed Ports (Retorqued)

P011-200.23%0.17%0.03%0.04%0.05%Re-Calibrate (In Place)

P001-100.86%1.10%0.05%0.05%0.28%4-6213B’s (From Shelf)

PH41-601.81%1.07%0.06%0.07%0.65%4-E30MA’s (From Shelf)

PH31-400.64%0.35%0.11%0.06%0.07%Changed Ports (Retorqued)

PH11-200.48%0.58%0.12%0.04%0.05%Re-Calibrate (In Place)

PH01-101.93%1.68%0.12%0.06%0.63%4-E30MA’s (From Shelf)

RoundNo.

Extreme Spread

WorstDeviation

Round to Round Var.

Within Gage Var.

Gage to Gage Var.

Configuration

Torque / Mount Sensitivity Test (10 repetitions at each configuration)

Aberdeen Test Center 26

Calibration Discrepancies

• In 1989 ATC forced the Copper Crusher Tarage Tables to agree with the Tourmaline transducers.

• In 1990 ATC’s analysis of the Bourges Copper Crusher data (based on quartz transducers) indicated that the quartz sensors used by the French were reading 1.0% lower than the tourmaline sensors used by the USA.

• In 2002 the 1989 Copper Crusher Tarage table was again evaluated in the USA. The data indicated that the Copper Crushers no longer agree with the tourmaline transducers but now agree with the quartz transducers (approximately 0.8% lower than tourmaline).

Page 14: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

14

Aberdeen Test Center 27

Calibration Test Setup

• Testing was limited to ATC & YTC• The same transducers were used at each facility.

• 2 each YTC E30MA’s• 2 each Kistler 6213B’s

• The same electronics equipment (signal conditioning, data acquisition) was used at each facility.

• Each facility used it’s own calibrator and calibration method.

Aberdeen Test Center 28

Calibration Results

81.0x103

80.8

80.6

80.4

80.2

80.0

79.8

79.6

79.4

79.2

79.0

78.8

78.6

78.4

78.2

78.0

Pres

sure

(Psi

)

5.65.45.25.04.8

Time (s)

ATC Cal. Results vs YTC Cal. Results Date: 07-10-2003

E30MA SN:577 YTC Cal. E30MA SN:577 ATC Cal.

Peak = 79320.23Psi

Peak = 80271.42Psi

"True Pressure" = 79104.7PsiCalibration Difference = 1.1%

Same Tourmaline Transducer Calibrated at both ATC & YTC

Page 15: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

15

Aberdeen Test Center 29

Calibration Results (cont.)

Same Quartz Transducer Calibrated at both ATC & YTC

81.0x103

80.8

80.6

80.4

80.2

80.0

79.8

79.6

79.4

79.2

79.0

78.8

78.6

78.4

78.2

78.0

Pres

sure

(Psi

)

5.65.45.25.04.8

Time (s)

ATC Cal. Results vs YTC Cal. Results Date: 07-10-2003

6213B SN:626578 YTC Cal. 6213B SN:626578 ATC Cal.

Peak = 79561.69Psi

Peak = 79452.60Psi

"True Pressure" = 79104.7PsiCalibration Difference = 0.13%

Aberdeen Test Center 30

Results Summary

• Quantification of bias and errors uncovered:

0.30% ATC Calibration Error +0.40% E30MA Overshoot Error0.70% Total Error Identified During Copper Crusher Tests

- 0.26% YTC Instrumentation Error*0.44% Total Remaining Error

* Gage-to-gage variability of the E30MA is on the order of approximately 0.6%.

Page 16: Study of Calibration Techniques · 2010-11-18 · Study of Calibration Techniques A Demonstration of a Transducers Response to High Pressure Dynamic Loads Presented By: Ken McMullen

16

Aberdeen Test Center 31

Conclusion

1. Understand the one-size-fits-all for calibration does not always work.

2. Know how your transducer responds to your test environment.3. Know how your transducer responds to your calibration method.4. Understand the weak links in the calibration process.

a. Calibration Capacitors (charge mode transducers).b. Measurement Locations (A/D digitizers)c. Secondary Sources (check often)

5. When using new transducers for the first time test them thoroughly.6. Don’t solely rely on statistics for your assurance.

Aberdeen Test Center 32

Acknowledgments / References

ATCBill Burget

David PorterScott Walton

YTCWelton Phillips

David Smith

No. 1 - Weddle, Arlen L., Methodology Investigation Final Report Pressure Transducer Calibration Comparison; Report No. APG-MT-5884, U. S. Army Aberdeen Proving Ground, MD, September, 1983

No. 2 - McMullen, Kenneth R., Final Report of Calibration Data on Transducers, Report No. CSTA-7388, U. S. Army Combat Systems Test Activity, Aberdeen Proving Ground, MD, December, 1992