Indication for use: As a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions. Before use, please read the Instructions For Use which accompany the product for indications, contraindications, warnings and precautions. Caution: Federal law restricts this device to sale by or on the order of a physician. Step by step description - Use of Gangi-SoftGuard ® 29376 Rev 01 2016-08

Upload: buihanh

Post on 25-Feb-2018

218 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Indication for use: As a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions. Before use, please read the Instructions For Use which accompany the product for indications, contraindications, warnings and precautions. Caution: Federal law restricts this device to sale by or on the order of a physician.

Step by step description - Use of Gangi-SoftGuard®

29376 Rev 01 2016-08

Page 2: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Introduction Gangi-SoftGuard® is a coaxial biopsy needle provided with a trocar stylet and a spring loaded blunt stylet. This step by step guide describes the function of the spring loaded blunt stylet and shows an example procedure where Gangi-SoftGuard® is used.

29376 Rev 01 2016-08

Page 3: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Introduction – Package contents Each Gangi-SoftGuard® package contains:

! "! #! $! %! &! '! (! )! *! "!!

""!

"#!

"$!

"%!

"&!

"'!

"(!

")!

"*!

#!!

#"!

##!

#$!

#%!

#&!

#'!

Needle length

"(" mm

"#" mm

(" mm

#(!

2931

0 Re

v. 0

1 20

15-0

1

mm

Needle

Trocar stylet

Spring loaded blunt stylet

Reference guide

29376 Rev 01 2016-08

Page 4: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Introduction – The blunt stylet The blunt stylet is spring loaded. When resistance is met at the blunt tip, the tip retracts into the needle. This exposes the sharp needle, and cutting through tissue can be performed. Simultaneously with the retracting tip, the pin at the needle hub springs back, giving visual and tactile feedback of the position of the blunt tip.

29376 Rev 01 2016-08

Page 5: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Introduction – Exchanging stylets The device is provided with two stylets. The stylets are easily exchanged and locked into position. Stylets can be exchanged before and during procedure.

UNLOCK REMOVE INSERT LOCK

29376 Rev 01 2016-08

Page 6: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 1 Determine which stylet is preferred. The needle is delivered with the trocar stylet inserted. If the blunt stylet is preferred, exchange stylets. In this procedure, the trocar stylet is used for skin entry.

TROCAR BLUNT

29376 Rev 01 2016-08

Page 7: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 2 Insert device superficially. Monitor position using imaging technique. Note: If trocar stylet is not used, make a small skin incision using a scalpel to make penetration of skin easier.

29376 Rev 01 2016-08

Page 8: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Note The indicator on the needle hub reveals the bevel position at the needle tip.

29376 Rev 01 2016-08

Page 9: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 3 Exchange to blunt stylet.

29376 Rev 01 2016-08

Page 10: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 4 Advance device. Monitor position using imaging technique.

29376 Rev 01 2016-08

Page 11: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 5 If resistance is met at the blunt tip, the tip will retract into the needle and the sharp needle will be exposed. Simultaneously, the pin at the stylet hub will spring back indicating the position of the blunt tip. If continuing to advance the device, the sharp needle will cut through the tissue causing the resistance.

29376 Rev 01 2016-08

Page 12: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Note If exposure of the sharp needle is not wanted, the retraction of the blunt tip can be prevented by pressing the pin in the hub with your thumb.

29376 Rev 01 2016-08

Page 13: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 6 Advance device proximal to lesion. Monitor position using imaging technique.

29376 Rev 01 2016-08

Page 14: Step by step description - Use of Gangi-SoftGuardapriomed.com/wp-content/uploads/29376-gangi-softguard-step-by-ste… · Introduction Gangi-SoftGuard® is a coaxial biopsy needle

Step 7 Remove the stylet. Perform biopsy as per normal routine.

29376 Rev 01 2016-08