silicon and germanium...

8
SILICON AND GERMANIUM PHOTODETECTORS OPTICS PHOTONICS MEASUREMENTS

Upload: others

Post on 22-Mar-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

SILICON AND GERMANIUM

PHOTODETECTORS

OPTICS PHOTONICS MEASUREMENTS

Page 2: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

UV - NIRS I L I C O N

The UV Enhanced Low Profile Silicon Photodetector (10-UVSI) is a large 10mm x 10mm area UV enhanced Si photodiode housed in a low profile detector housing.

The 10-UVSI provides NIST traceable power measurements across the 200nm to 1100nm spectral region. It is operated in photovoltaic (unbiased) mode to provide the lowest noise in DC and low frequency opera on since it avoids the electron shot noise associated with bias currents.

The housing provides #4-40 and M4 threads for post mounted applica ons, but also includes an SM1 thread for compa bility with many standard op cal components or mounts. Its low profile (11mm thick) is perfect for ght spaces.

All detectors are calibrated following protocols outlined by NIST publica on SP250-41-08 for spectral response transfer and are supplied with a NIST traceable calibra on report. The report details individual measured spectral responsivity and its uncertainty along all NIST traceability informa on. Detectors are BNC terminated, but an op onal Newport power meter compa ble DB15 connector is available.

To maintain accuracy and guarantee performance, ICC recommends annual photodiode detector recalibra on.

SILICON PHOTODETECTORS

NIST TRACEABLE

200 400 600 800 1000 1200

0.1

0.2

0.3

0.4

0.5

Spec

tral R

espo

nse

(A/W

)

Wavelength (nm)

1Uncalibrated spectral range. 2Typical values measured at 25°C ambient temperature

Parameter Min Typical Max

Model NumberDetector TypeMaterialPhotosensi ve Area (W x H, mm)Spectral Range (nm)�� 200 1100Window MaterialSpectral Response (A/W)�

200nm 0.12970nm 0.5

Calibra on Uncertainty200nm - 220nm (%)221nm - 350nm (%)351nm - 970nm (%)971nm - 1100nm (%)

Max. Measurable Power (mW)Max Power Density, Average (W/cm�)Uniformity (%)

25 @ 950nm*9*

± 2.0

6425

Sensor Characteris cs

10-UVSIHamamatsu S1337-BQ

UV Enhanced Silicon 10 x 10

Quartz

*Dependent on power meter

Page 3: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

Parameter Min Typical Max

Rise Time (μs)� 3Shunt Resistance (GΩ) 0.05 0.2Noise Equivalent Power (fW/Hz� � �) 18Opera ng Temperature (°C) 0 60Connector Type

Electrical Characteris cs

BNC

NOTES:1.) DIMENSIONS ARE IN INCHES UNLESS OTHERWISE NOTED.2.) MIN BNC CABLE LENGTH IS 48”

MECHANICAL DIMENSIONS:

Contact us for pricing informa on.

Page 4: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

The Large Area Ge Photodetector (05GE) features a 5mm diameter ac ve area germanium photodiode housed in a low profile detector housing and provides NIST traceable power measurements across the 800nm to 1800nm spectral region. It is operated in photovoltaic (unbiased) mode to provide the lowest noise in DC and low frequency opera on since it avoids the electron shot noise associated with bias currents.

The housing has both #4-40 and M4 threads for post mounted applica ons as well as an internal SM1 thread for compa bility with many standard op cal components or mounts. Its low profile (11mm thick) is perfect for ght spaces.

All detectors are calibrated following protocols outlined by NIST publica on SP250-41-08 for spectral response transfer and are supplied with a NIST traceable calibra on report. The report details individual measured spectral responsivity and its uncertainty along all NIST traceability informa on. Detectors are BNC terminated, but an op onal Newport power meter compa ble DB15 connector is available.

To maintain accuracy and guarantee performance, ICC recommends annual photodiode detector recalibra on.

GERMANIUM PHOTODETECTORS

NIST TRACEABLE

1Uncalibrated spectral range. 2Typical values measured at 25°C ambient temperature

Parameter Min Typical Max

Model NumberDetector TypeMaterialPhotosensi ve Area (W x H, mm)Spectral Range (nm)�� 800 1800Window MaterialSpectral Response (A/W)�

1300nm 0.651550nm 0.85

Calibra on Uncertainty800nm - 1700nm (%)1701nm - 1800nm (%)

Max. Measurable Power (mW)Max Power Density, Average (W/cm�)Uniformity (%)

15 @ 1550nm*5*

± 5.0

3.55

Sensor Characteris cs

05GEGPD Optoelectronics GM8

Germanium 5mm dia.

Quartz

850nm 0.26

800 1000 1200 1400 1600 18000.0

0.2

0.4

0.6

0.8

1.0

1.2

Spec

tral R

espo

nse

(A/W

)

Wavelength (nm)

NIR-SWIRG E RM A N I UM

*Dependent on power meter

Page 5: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

Parameter Min Typical Max

Rise Time (μs)� 3Shunt Resistance (kΩ) 2 4Noise Equivalent Power (pW/Hz� � �) 4Opera ng Temperature (°C) 0 60Connector Type

Electrical Characteris cs

BNC

NOTES:1.) DIMENSIONS ARE IN INCHES UNLESS OTHERWISE NOTED.2.) MIN BNC CABLE LENGTH IS 48”

MECHANICAL DIMENSIONS:

Contact us for pricing informa on.

Page 6: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

ORIEL 1.5” FLANGEADAPTERS

WE OFFER SPECIALLY DESIGNED FLANGE ADAPTERS to

connect our detectors to 1.5” flange connections from

Newport - Oriel Instruments.

PART NO:

External 1.5” Flange Adapter (O15E-SM1E)

Internal 1.5” Adapter (O15I-SM1E)

O15I-SM1E

O15E-SM1E

LENS, OPTICS, ANDFILTERS

USE EXTERNAL SM1 THREADED LENS TUBES to add

optical filters or lenses. Apertures, diffusures, or

integration spheres can also be added using SM1 adapters

- however, detector recalibration is required.

PART NO:

Thorlabs External SM1 Lens Tube, 0.3”, (SM1L03)

Thorlabs External SM1 Lens Tube, 0.5”, (SM1L05)

Thorlabs External SM1 Lens Tube, 1.0”, (SM1L10)

... and any external SM1 threaded components

SM1L10

SM1L05

SM1L03

EXTERNAL SM1FIBER OPTIC ADAPTERS

OUR SM1 THREADED DETECTOR HEAD can measure fiber optical

power when used with any fiber optic adapters from name brand

companies like Newport or Thorlabs.

PART NO:

Thorlabs ST/PC Fiber Adapter Plate with Ext. SM1 Thread, (SM1ST)

Thorlabs SMA Fiber Adapter Plate with Ext. SM1 Thread, (SM1SMA)

Thorlabs FC/PC Fiber Adapter Plate with Ext. SM1 Thread, (SM1FC)

Thorlabs APC Fiber Adapter Plate with Ext. SM1 Thread, (SM1APC)

SM1L10

SM1L05

SM1L03

10-UVSI

05GE

PHOTODETECTORS ACCESSORIES

NEW

O15I-SM1E

O15E-SM1E

PTICS, ANDLTERS

HREADED LENS TUBES to add

es. Apertures, diffusures, or

so be added using SM1 adapters

r recalibration is required.

ens Tube, 0.3”, (SM1L03)

ens Tube, 0.5”, (SM1L05)

SM1L10

SM1L05

SM1L03

M1 Thread, (SM1ST)

M1 Thread, (SM1SMA)

M1 Thread, (SM1FC)

1 Thread, (SM1APC)

SM1L10

SM1L05

SM1L03

10-UVSI

05GE

Page 7: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM
Page 8: SILICON AND GERMANIUM PHOTODETECTORSiccconsultingllc.com/.../08/Detector-Brochure-RevA-2018.pdfDetector Brochure.ai Author Julie Chen Created Date 8/21/2018 12:55:58 AM

ICC CONSULTING LLC67 FOREST, SUITE 287

MARLBOROUGH, MA 01752201.621.1720

www.iccconsultingllc.com

Rev A (2018)