UNIVERSITY OF MIAMI COLLEGE OF ENGINEERING Transforming Lives Through Teaching, Research, & Service Force-Feedback System For Laryngoscopy Quratulain Amin, Eli J. Candelaria, Edugie Osunde, Wame Chibamo, Karen Johnsson Rodriguez Weizhao Zhao Department of Biomedical Engineering Senior Design BS Project May - 2020 Abstract During the intubation process, forces greater than 39N applied onto the laryngoscope can cause damage (Carassiti et al., 2012). It may affect: the teeth, vocal cords of patients, perforate the hypopharynx, or cause subcutaneous cervical and facial emphysema (Berry et al., 2013). Currently, there are limited ways to protect the patient's mouth from excessive force. Also, available solutions interfere with the procedure, and don’t offer full protection. Therefore, a force transducer that displays the force being generated is beneficial. The objective of this device is to bypass the need for protection by alerting the user of excessive force. When the tongue’s soft tissue contacts the sensor on the laryngoscope, the device elicits sensory feedback. The physician can then see how much force he is exerting and adjust accordingly. This system would consist of a 3-D printed adaptor fitted onto a traditional Macintosh laryngoscope blade. Force sensing resistors (FSRs) would be mounted on the adaptor and connected to a feedback circuit that would provide a sensory signal for the users. Introduction Laryngoscopes are devices used for intubation, and to observe the laryngeal area. When forces greater than 39N are applied by technicians, damage may occur in the patients’ throat, tongue and teeth. Current devices have no way of measuring the amount of force being generated. Therefore, our project focuses on developing a feedback system to prevent damaging due to laryngoscopy Results References Conclusion Research has indicated that 39 N is the threshold for damage. Any exceeding force to an adult without underlying health conditions can lead to adverse effects, such as nerve, teeth and tissue damage. Current laryngoscope models lack feedback systems, which informs healthcare professionals of potential damage they may be inflict. Our device would fill a gap currently seen in the healthcare system globally. Berry, J. M., & Harvey, S. (2013). Chapter 17 - Laryngoscopic Orotracheal and Nasotracheal Intubation. In C. A. Hagberg (Ed.), Benumof and Hagberg's Airway Management (Third Edition) (pp. 346-358.e341). Philadelphia: W.B. Saunders. Carassiti, M., Zanzonico, R., Cecchini, S., Silvestri, S., Cataldo, R., & Agrò, F. E. (2012). Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal and difficult airways: a manikin study. British Journal of Anaesthesia, 108(1), 146-151. doi:10.1093/bja/aer304 Acknowledgments We would like to thank University of Miami’s Department of Biomedical Engineering and our mentor Dr. Zhao, for he advised us throughout this enterprise. Finally, we would like to thank Dr. Manns and Dr. Bohorquez for their dedication throughout this process. Calibration Fig1: Force Feedback System for laryngoscopy. For this project FSR was used which decrease the resistance as the force applied on it increased (A). FSR mounted onto the 3D adaptor which further connected with the LED. If the force applied by the physician passed 39N threshold, the transducer will provide a sensory feedback in the form of LED and buzzer(B). Adaptor Design Fig2:ElectroForce Gauge was used to calibrate the 0-5Kg sensor. The force increased incrementally and the change in resistance was recorded. ElectroForce Gauge Fig3: The thin adaptor can easily clip on and off from the laryngoscope. It is single-use, and disposable. Final Prototype Methods | Design | Analysis Fig4: 0-5Kg sensors were mounted onto the adaptor(1). The FSRs were connected to an amplifying circuit(2). The amplified FSR signal was fed to the Arduino, which was connected to an LED and a buzzer(3). The force applied by the user caused resistance changes, which were transformed into sensory feedback (4). FSR Force(N) Conductance(mS) Fig5: It can be concluded that the sensor is accurate, and that force has a linear relationship with conductance. Force is proportional to conductance, and inversely proportional to resistance. Based on the measurements obtained, this sensor works in our desired range (0-50N) . § The chosen sensor for our product can measure the force between 0-50N. § The 3D designed adaptor is thin and does not affect the intended use of the laryngoscope. § The feedback transducer informs the user if damaging forces are being applied in the form of a sensory output Adaptor Circuit to Amplify the FSR signal Sensory Output Amplified signal 1 2 3 4 A B Fig6: The adaptor does not interfere with the functions of the laryngoscope, fulfilling the design specifications.

Upload: others

Post on 18-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Senior DesignBS Project May -2020 Force-Feedback System ......Methods | Design | Analysis Fig4: 0-5Kg sensors were mounted onto the adaptor(1). The FSRs were connected to an amplifying

UNIVERSITY OF MIAMICOLLEGE OF ENGINEERING

Transforming Lives Through Teaching, Research, & Service

Force-Feedback System For LaryngoscopyQuratulain Amin, Eli J. Candelaria, Edugie Osunde, Wame Chibamo, Karen Johnsson Rodriguez

Weizhao Zhao

Department of Biomedical Engineering

Senior Design BS ProjectMay - 2020

Abstract

During the intubation process, forces greater than 39N appliedonto the laryngoscope can cause damage (Carassiti et al., 2012). Itmay affect: the teeth, vocal cords of patients, perforate thehypopharynx, or cause subcutaneous cervical and facialemphysema (Berry et al., 2013). Currently, there are limited ways toprotect the patient's mouth from excessive force. Also, availablesolutions interfere with the procedure, and don’t offer fullprotection. Therefore, a force transducer that displays the forcebeing generated is beneficial. The objective of this device is tobypass the need for protection by alerting the user of excessiveforce. When the tongue’s soft tissue contacts the sensor on thelaryngoscope, the device elicits sensory feedback. The physiciancan then see how much force he is exerting and adjustaccordingly. This system would consist of a 3-D printed adaptorfitted onto a traditional Macintosh laryngoscope blade. Forcesensing resistors (FSRs) would be mounted on the adaptor andconnected to a feedback circuit that would provide a sensory signalfor the users.

IntroductionLaryngoscopes are devices used for intubation, and to observe thelaryngeal area. When forces greater than 39N are applied bytechnicians, damage may occur in the patients’ throat, tongue andteeth. Current devices have no way of measuring the amount offorce being generated. Therefore, our project focuses ondeveloping a feedback system to prevent damaging due tolaryngoscopy

Results

References

Conclusion

Research has indicated that 39 N is the threshold for damage. Any exceeding force to an adult without underlying health conditions can lead to adverse effects, such as nerve, teeth and tissue damage. Current laryngoscope models lack feedback systems, which informs healthcare professionals of potential damage they may be inflict. Our device would fill a gap currently seen in the healthcare system globally.

Berry, J. M., & Harvey, S. (2013). Chapter 17 - Laryngoscopic Orotracheal and Nasotracheal Intubation. In C. A. Hagberg (Ed.), Benumof and Hagberg's Airway Management (Third Edition) (pp. 346-358.e341). Philadelphia: W.B. Saunders.

Carassiti, M., Zanzonico, R., Cecchini, S., Silvestri, S., Cataldo, R., & Agrò, F. E. (2012). Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal and difficult airways: a manikin study. British Journal of Anaesthesia, 108(1), 146-151. doi:10.1093/bja/aer304

AcknowledgmentsWe would like to thank University of Miami’s Department ofBiomedical Engineering and our mentor Dr. Zhao, for he advisedus throughout this enterprise. Finally, we would like to thank Dr.Manns and Dr. Bohorquez for their dedication throughout thisprocess.

Calibration

Fig1: Force Feedback System for laryngoscopy. For this project FSR was used which decrease the resistance as the force applied on it increased (A). FSR mounted onto the 3D adaptor which further connected with the LED. If the force applied by the physician passed 39N threshold, the transducer will provide a sensory feedback in the form of LED and buzzer(B).

Adaptor Design

Fig2:ElectroForce Gauge was used to calibrate the 0-5Kg sensor. The force increased incrementally and the change in resistance was recorded.

ElectroForce Gauge

Fig3: The thin adaptor can easily clip on and off from the laryngoscope. It is single-use, and disposable.

Final Prototype

Methods | Design | Analysis

Fig4: 0-5Kg sensors were mounted onto the adaptor(1). The FSRs were connected to an amplifying circuit(2). The amplified FSR signal was fed to the Arduino, which was connected to an LED and a buzzer(3). The force applied by the user caused resistancechanges, which were transformed into sensory feedback (4).

FSR

Force(N)

Cond

uctance(mS)

Fig5: It can be concluded that the sensor is accurate, and that force has a linear relationship with conductance. Force is proportional to conductance, and inversely proportional to resistance. Based on the measurements obtained, this sensor works in our desired range (0-50N) .

§ The chosen sensor for our product can measure the force between 0-50N.

§ The 3D designed adaptor is thin and does not affect the intended use of the laryngoscope.

§ The feedback transducer informs the user if damaging forces are being applied in the form of a sensory output

Adaptor

Circuit to Amplify the FSR signal

Sensory Output

Amplified signal

1 2 3 4

A B

Fig6: The adaptor does not interfere with the functions of the laryngoscope, fulfilling the design specifications.