sapphire™ - gilson eng manuals/tn/sapphire.pdf · ultrasonic point level system ... manual....

77
Part of Thermo Fisher Scientific Sapphire™ Ultrasonic Point Level System User Guide P/N 900-000068 Revision C

Upload: vonhi

Post on 17-Mar-2018

223 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Part of Thermo Fisher Scientific

Sapphire™Ultrasonic Point Level SystemUser GuideP/N 900-000068

Revision C

Page 2: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result
Page 3: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Sapphire™Ultrasonic Point Level SystemUser GuideP/N 900-000068

Revision C

Page 4: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result
Page 5: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

© 1992 Thermo Fisher Scientific Inc. All rights reserved.

“Hastelloy” is a registered trademark of Haynes International, Inc. “MONEL” is a registered trademark of Special Metals. “Carpenter 20” is a registered trademark of Carpenter Technology Corp. “Tri-Clover” is a trademark or registered trademark of Alfa Laval Inc.

All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries.

Thermo Fisher Scientific (Thermo Fisher) makes every effort to ensure the accuracy and completeness of this manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result of errors or omissions. Thermo Fisher reserves the right to make changes to the manual or improvements to the product at any time without notice.

The material in the manual is proprietary and cannot be reproduced in any form without expressed written consent from Thermo Fisher.

Page 6: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 7: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Thermo Fisher Scientific Sapphire User Guide v

Revision History

Revision Level Date Comments

8/92 08-92 Initial release

5/93 05-93 Added Canadian Standards Association (CSA) approval, Factory Mutual (FM) non-incendive approval, relay contact ratings, thermal isolator functionality, 4–20 mA setup procedures, and circuit board replacement section.

A00 05-94 Modified to reflect incorporation of gold-plated relay contacts, a lower operating temperature range, and new Hastelloy B and Carpenter 20 sensor materials. Added remote mounting, Tri-Clover, and normally open (NO) auxiliary relay capabilities.

A00 02-95 Modified to include the consolidated logic printed circuit board number 220-000448.

A00 09-95 Added PFA Teflon Center Shaft Gap Sensor.

A00 07-96 Modified to reflect addition of intrinsically safe approvals, field-selectable NO or NC auxiliary fault relay, manual loop test, noise self-test, variable system qualification, and 526/527 operating instructions.

A00 09-96 Modified to correct typographical errors.

A00 07-97 Modified to reflect address and company name changes.

01 04-00 Deleted PFA probes and added new temperature specifications.

02 02-01 Name change.

03 10-01 PROM change and eliminate noise test.

04 08-02 Revised per ECO 7803.

05 01-03 Revised per ECO 7880.

A 01-06 Revised per ECO 5099.

B 11-06 Revised per ECO 5412.

C 03-08 Revised per ECO 6251.

Page 8: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 9: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Contents Safety Summary .................................................................................................. ix

Warnings ...............................................................................................ix Cautions ................................................................................................ix Notes .....................................................................................................xi

Product Overview ............................................................................................. 1-1 Chapter 1 Introduction........................................................................................ 1-1 Using this Guide ................................................................................. 1-1

Chapter Overviews........................................................................... 1-1 Personnel ......................................................................................... 1-2

Components ....................................................................................... 1-2 Dual Compartment Housing ........................................................... 1-2 Power Supply ................................................................................... 1-3 Filter Board...................................................................................... 1-3 Logic PCB........................................................................................ 1-3 Sensors ............................................................................................. 1-3

Sensor Configurations ......................................................................... 1-4 Two-Wire Electronics ...................................................................... 1-5

Chapter 2 Theory of Operation.......................................................................................... 2-1 Ultrasonic Principles ........................................................................... 2-1 Time Gate Technology ....................................................................... 2-2 Continuous Self-Test .......................................................................... 2-2 Manual Loop Test............................................................................... 2-3 System Qualification Time.................................................................. 2-3 System Threshold Adjustment ............................................................ 2-3 Relay Output Models.......................................................................... 2-4

Process Relay Alarm Operation ........................................................ 2-4 Auxiliary Fault Relay Operation....................................................... 2-5

Two-Wire Models............................................................................... 2-6 Process Alarm Indication.................................................................. 2-6 Self-Test Fault Indication................................................................. 2-6

Optional Thermal Isolator .................................................................. 2-7

Special Handling Considerations.................................................................. 3-1 Chapter 3 ESD Procedures .................................................................................. 3-1 Unpacking, Inspection, & Storage ...................................................... 3-2 Hazardous Location Installations ........................................................ 3-2

Thermo Fisher Scientific Sapphire User Guide vii

Page 10: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Contents

viii Sapphire User Guide Thermo Fisher Scientific

Installation..........................................................................................................4-1 Chapter 4 Factory Default Settings ...................................................................... 4-1 Field Setup Procedures ........................................................................ 4-2 Threshold Adjustment Verification ..................................................... 4-4

Relay Output Models....................................................................... 4-4 Two-Wire Models............................................................................ 4-5

Self-Test Verification .......................................................................... 4-6 Relay Output Models....................................................................... 4-6 Two-Wire Models............................................................................ 4-7

Installation Guidelines ........................................................................ 4-7 Installation Procedures ........................................................................ 4-9

3/4” NPT Mount – Integral Systems ............................................... 4-9 Flange Mount Sensors – Integral Systems....................................... 4-10 Remote Mount Installations........................................................... 4-11

3/4” NPT Mount – Remote Systems .......................................... 4-12 Flange Mount Sensors – Remote Systems.................................... 4-14

Chapter 5 Maintenance ......................................................................................................5-1 Electronics .......................................................................................... 5-1 Sensors ................................................................................................ 5-1 Circuit Board Replacement ................................................................. 5-1 Procedure............................................................................................ 5-2 Field Adjustments ............................................................................... 5-6

System Sensitivity Adjustment ......................................................... 5-6 System Qualification Time Adjustment ........................................... 5-6

Final Inspections ................................................................................. 5-6

Chapter 6 Troubleshooting & Service .............................................................................6-1 General ............................................................................................... 6-1 Service................................................................................................. 6-2

Appendix A Ordering Information .......................................................................................A-1 Integral Mount System........................................................................A-1 Remote Mount System........................................................................A-3

Appendix B Specifications...................................................................................................B-1 Relay Operation ..................................................................................B-1 Two-Wire Operation ..........................................................................B-2

Appendix C Drawings............................................................................................................ C-1

Toxic & Hazardous Substances Tables* .....................................................D-1 Appendix D

Page 11: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Safety Summary

The following warnings, cautions, and notes appear throughout this guide.

Warnings Warnings alert users to operating procedures, practices, conditions, etc. that may result in injury or death if not carefully followed or observed.

Warning Sapphire systems may be used only in those hazardous locations identified on the product nameplates. ▲ (3–2)

Warning Installation must be carried out in accordance with local site requirements and regulations. ▲ (4–1)

Warning Electrical installation must be made in compliance with application area classification codes. Only qualified personnel should perform installation. ▲ (4–2)

Warning Hazardous voltages can be present on terminals. Voltages can cause injury or death. Proceed with extreme caution. ▲ (4–4)

Warning The system may be used only in those hazardous locations identified on the product nameplate. These installation and servicing instructions are for use by qualified personnel only. ▲ (4–7)

Warning When disconnecting power from the instrument, follow proper lockout/tagout procedures to avoid accidental exposure of personnel to high voltage circuits. ▲ (5–2)

Cautions Cautions alert users to operating procedures, practices, conditions, etc. that may result in damage to equipment if not carefully followed or observed.

Caution The manual loop test feature will not activate (reverse outputs) when the system is in a self-test fault or malfunction condition. ▲ (2–3)

Caution The thermal isolator is an integral part of the sensor assembly. Do not attempt to remove it from the sensor. Removing or loosening the thermal isolator can damage the sensor. ▲ (2–7)

Thermo Fisher Scientific Sapphire User Guide ix

Page 12: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Safety Summary Cautions

Caution These systems are electrostatic discharge (ESD) sensitive. Use proper ESD protective equipment and follow ESD handling procedures when servicing or handling the instrument. Failure to comply with ESD procedures can result in circuit damage. ▲ (3–1)

Caution These systems are ESD sensitive instruments. Use proper ESD protective equipment and follow ESD handling procedures when servicing or handling the instrument. Failure to comply with ESD procedures can result in circuit damage. ▲ (4–2)

Caution Sensors are fragile instruments. Handle with care. Do not support or carry systems by the sensors. Sensor breakage may cause fluids to enter electronics housing. The sensor, flange (if applicable), and electronics have been carefully assembled at the factory. Disassembly of the system is not required for installation and may result in damage to the sensor. ▲ (4–9)

Caution Do not support or carry flange mounted unit by the sensors. Handle these units by the flange to avoid damaging the sensor. ▲ (4–10)

Caution The sensor assembly is provided with a pre-determined length (per customer order) of sensor cables. Use care to avoid damaging the cable or phono plugs during installation. ▲ (4–12)

Caution Due to the nature of the piezoelectric crystals used in the probes of Sapphire systems, Thermo Fisher recommends discharging each signal cable by connecting the shield (case) to center conductor (pin) before connecting to the main logic board or remote mounting bracket signal cables. ▲ (4–13)

Caution The sensor assembly is provided with a pre-determined length (per customer order) of sensor cables. Use care to avoid damaging the cable or phono plugs during installation. ▲ (4–18)

Caution Due to the nature of the piezoelectric crystals used in the probes of Sapphire systems, Thermo Fisher recommends discharging each signal cable by connecting the shield (case) to center conductor (pin) before connecting to the main logic board or remote mounting bracket signal cables. ▲ (4–15)

Caution Do not clean with saturated steam greater than 50 psig. Repeated steam cleaning can potentially shorten sensor life. ▲ (5–1)

x Sapphire User Guide Thermo Fisher Scientific

Page 13: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Safety Summary Notes

Caution Sapphire systems are ESD sensitive instruments. Use proper ESD protective equipment and follow ESD procedures when handling the instrument. Failure to comply with ESD procedures can result in circuit damage. ▲ (5–2)

Caution Replacement boards are sent in sets. Do not reuse the old circuit board. Improper operation or damage can occur. ▲ (5–3)

Caution Components L1 and R37 of the logic PCB assembly are factory calibrated. Do not adjust these components. Improper operation or damage can occur. ▲ (5–4)

Caution Due to the nature of the piezoelectric crystals used in the probes of Sapphire systems, Thermo Fisher recommends discharging each signal cable by connecting the shield (case) to the center conductor (pin) before connecting to the main logic PCB or remote mounting bracket signal cables. ▲ (5–4)

Notes Notes are statements of company policy regarding operating procedures, practices, conditions, etc. Notes are also used to emphasize operating procedures, practices, conditions, etc.

Note The factory settings located on the underside of the main logic PCB are configured for sensor types and should not be changed. ▲ (1–3)

Note Due to the unique signal processing capability of the slot sensor, the threshold adjustment potentiometer (R9) is not required. This function is disabled on slot sensor units. ▲ (2–3)

Note System relay operation has been designed for fail-safe operation. A self-test failure or power loss will de-energize the process and auxiliary relays regardless of the actual state of the measurement gap. ▲ (2–4)

Note Inspection, adjustment, installation, and maintenance of this unit must be performed by experienced personnel. Refer to “Personnel” in Chapter 1. ▲ (3–2)

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲ (4–10)

Thermo Fisher Scientific Sapphire User Guide xi

Page 14: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Safety Summary Notes

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲ (4–10)

Note Sapphire system enclosures installed in Class I, Div. 1 Hazardous (Classified) Locations are factory sealed and require a poured conduit seal only if the conduit run leaves said location or if the conduit run enters another enclosure requiring a poured seal. ▲ (4–10)

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

(4–11)

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲ (4–11)

Note A system enclosure installed in Class I, Div. 1 Hazardous (Classified) Location is factory sealed and requires a poured conduit seal only if the conduit run leaves the installation location or if the conduit run enters another enclosure requiring a poured seal. ▲ (4–11)

Note Installation must be done in accordance with local electrical codes. ▲

(4–12)

Note If optional pipe or wall mounting brackets are used, mount the bracket to a solid structure or pipe prior to mounting the electronics to the bracket. ▲ (4–13)

Note A minimum of five full threads is required between all threaded joints. ▲ (4–13)

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

(4–13)

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲ (4–14)

xii Sapphire User Guide Thermo Fisher Scientific

Page 15: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Safety Summary Notes

Thermo Fisher Scientific Sapphire User Guide xiii

Note A system enclosure installed in Class I, Div. 1 Hazardous (Classified) Location) is factory sealed and requires a poured conduit seal only if the conduit run leaves the installation location or if the conduit run enters another enclosure requiring a poured seal. ▲ (4–14)

Note If optional pipe or wall mounting brackets are used, mount the bracket to a solid structure or pipe prior to mounting the electronics to the bracket. ▲ (4–15)

Note A minimum of five full threads is required between all threaded joints. ▲ (4–15)

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

(4–15)

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲ (4–16)

Note A system enclosure installed in Class I, Div. 1 Hazardous (Classified) Location) is factory sealed and requires a poured conduit seal only if the conduit run leaves the installation location or if the conduit run enters another enclosure requiring a poured seal. ▲ (4–16)

Note When installing a DC relay/power supply PCB (P/N 220-000450), secure the solder lug from the board against the board using the closest standoff. ▲ (5–4)

Note The interconnect cable is keyed to prevent incorrect orientation. ▲

(5–4)

Note Due to the unique signal processing capability of the slot sensor, the threshold adjustment potentiometer (R9) is not required. Slot sensor configured units have this function disabled. ▲ (5–6)

Note This troubleshooting section is a general list of possible conditions that could exist. This listing is not exhaustive and not meant to address all possible problems. Contact Thermo Fisher if a condition occurs that is not covered in this manual. ▲ (6–1)

Page 16: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 17: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Chapter 1 Product Overview

Introduction The Thermo Scientific Sapphire™ Ultrasonic Point Level System is the industry standard for reliable point level measurements. Its continuous self-test feature automatically checks system integrity including electronics, sensor, and crystal disbonding without needing additional crystals or sensor mounted electronics.

Dual compartment housing provides quick, easy access to field wiring connections while protecting the electronics from exposure to harsh environments, damage, or leaky connections.

Using this Guide This section provides an overview of each chapter and descriptions or qualifications of personnel called out in this guide.

Chapter Overviews ● Chapter 1: Provides a physical description of the system components.

● Chapter 2: Contains discussions regarding the theory of ultrasonic operation, Time Gate™ technology, manual loop test, system qualification time, and system threshold adjustment. Also describes the operation of the relay output and two-wire output models.

● Chapter 3: Discusses procedures for handling electrostatic discharge sensitive equipment, as well as procedures for unpacking, inspecting, and storing the system. Also included are physical mounting instructions for hazardous locations.

● Chapter 4: Contains information on factory configured default settings for field selected parameters. Specific information and installation guidelines are provided to allow the user to customize and verify operating parameters prior to final installation and placing the system into service.

● Chapter 5: Recommends preventative maintenance measures for system electronics and sensor components. Also contains series of instructions involving the installation of various circuit board replacement kits.

● Chapter 6: Provides a series of recommended troubleshooting measures to facilitate diagnosis and isolation of system faults.

● Appendix A: Provides system ordering information.

Thermo Fisher Scientific Sapphire User Guide 1-1

Page 18: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Product Overview Components

● Appendix B: Provides system specifications.

● Appendix C: Provides system drawings.

● Appendix D: Provides Toxic and Hazardous Substances tables.

Personnel ● Installation personnel: Should be an electrician or a person with a solid background in electronics instrumentation. The person should be familiar with the National Electrical Code (NEC), proper grounding and safety procedures, and local wiring regulations.

● Operating personnel: Should have knowledge of the process and should read and understand this guide before attempting any procedure pertaining to the operation of the system.

● Maintenance personnel: Should have a background in electricity and be able to recognize shock hazards. Should also be familiar with electronic process control instrumentation and have a good understanding of troubleshooting procedures.

Components Following are brief descriptions of the main system components. They are illustrated in Figure 1–1.

Dual Compartment Housing

All Sapphire systems are supplied with dual compartment housing in either integral or remote mounted configurations. This housing provides easy access to field connections without exposing the electronics to unnecessary handling, harsh environments, or physical damage. CSA approvals allow the Type 4 enclosure to be located in environmentally hazardous and corrosive locations.

The two housing compartments are called the circuit compartment and the connection compartment. The circuit compartment contains the power supply or filter board and the logic circuit board. The connection compartments houses two terminal strips for power and signal connections. A product label attached to the outside of the housing identifies the two compartments.

The compartments are mechanically isolated from each other using sealed electrical feed-throughs to connect the terminal strips with the circuit boards. Located at the top of the housing are two 1/2-inch NPT conduit connection ports for user supplied power and field wiring terminations.

1-2 Sapphire User Guide Thermo Fisher Scientific

Page 19: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Product Overview Components

Power Supply The power supply PCB is used with every relay unit and is located at the bottom of the circuit compartment. It is completely hidden from view by the logic PCB. The board is factory configured for the specific power specifications for each system. Field configuration for NO and NC contacts of the FAULT relay are available for customers. The NC contact is the factory default setting.

Filter Board The filter or EMC board replaces the power supply board in two-wire units.

Logic PCB The logic PCB is located at the top of the circuit compartment. This PCB contains field and factory selectable jumpers and an adjustment potentiometer. All field selectable settings are accessible on the exposed face of the logic PCB for easy access. Factory settings are located on the underside of the logic PCB, along with sensor and power supply connections. The factory settings on the underside of the logic PCB do not require field modification.

Note The factory settings located on the underside of the main logic PCB are configured for sensor types and should not be changed. ▲

Sensors ensors Sapphire systems are available in standard gap or optional slot sensor configurations. The gap sensor performs exceptionally well in most applications, including aerated and attenuating liquids which can cause problems for other ultrasonic technologies. The slot sensor provides reliable operation in even more severe or viscous applications (see Table 1–1). The gap sensor will handle over 98% of the applications.

Table 1–1. Sensor types

Sensor Style Gap Type Slot Type

Materials of construction 316SS, Hastelloy® B and C, MONEL®, Carpenter 20®

316SS, Hastelloy C

Minimum sensor “A” dimension 1” (25 mm) 4” (102 mm)

Maximum sensor “A” dimension 120” (3048 mm) 120” (3048 mm)

Remaining sensor length 1.44” (36.6 mm) 0.30” (7.6 mm)

Sensor operating temperature1 -40°F to +320°F (-40°C to +160°C) -310°F to +212°F (-190°C to +100°C)

-40°F to +320°F (-40°C to +160°C) -310°F to +212°F (-190°C to 100°C)

Maximum operating pressure2 -10 to 1500 psig (-69 to 10342 KPa) -10 to 1500 psig (-69 to 10342 KPa)

1The thermal isolator is required for 3/4" NPT units when process temperatures exceed 275°F (135°C). Flanged and remote mounted units do not require the thermal isolator. 2Pressure ratings are for 3/4” NPT units.

Thermo Fisher Scientific Sapphire User Guide 1-3

Page 20: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Product Overview Sensor Configurations

Figure 1–1. System components

Sensor Configurations

ensor Configurations

Gap sensors are available in a broad range of materials, which include 316SS, Hastelloy C, MONEL, Hastelloy B, and Carpenter 20. Slot sensors are available in 316SS and Hastelloy C only. The stainless steel versions of both sensor types can be delivered with a 150 grit (32 μinch) or better, Sanitary 3A finish and 316L stainless steel Tri-Clover® (16 AMP) fitting that assembles and disassembles quickly for fast cleaning. Sensor extensions, nipples, and flanges are available in like materials. All systems are available with 3/4-inch NPT mounting, various flange sizes, Tri-Clover fittings, or remotely mounted electronics. Systems may be installed vertically or horizontally.

Every system is imprinted with the sensor material type. For gap sensors, this information is located on the nipple flat or flange edge to orient the bridge location. For slot sensors, it is located on the sensor body side (Figure 1–2).

Sensor leads are routed into the circuit compartment of the housing and connected to the underside of the logic PCB.

1-4 Sapphire User Guide Thermo Fisher Scientific

Page 21: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Product Overview Sensor Configurations

Thermo Fisher Scientific Sapphire User Guide 1-5

Two-Wire Electronics

Sensors for use with two-wire electronics are approved explosion proof and intrinsically safe. They are identical in performance to earlier explosion proof only sensors but have additional components to meet the intrinsically safe requirements.

Figure 1–2. Sensor markings

Page 22: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 23: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Chapter 2 Theory of Operation

Sapphire systems are designed using state-of-the-art ultrasonic and electronic technologies. Advancements in acoustical design and materials, a patented Time Gate technology, and continuous self-test have produced the most reliable and technically advanced ultrasonic level sensor of its type.

Ultrasonic Principles

The system utilizes ultrasonic sound waves to detect the presence or absence of a liquid within a defined measurement gap. The basic principle that sound waves travel easily through a liquid media, while being severely attenuated when traveling though air, is incorporated into the design.

Most ultrasonic level switches utilize the continuous feedback technology developed and patented 30 years ago. A piezoelectric crystal positioned on one side of a measurement gap is continuously excited by an electrical signal. This signal causes an ultrasonic sound wave to travel through the measurement gap (see Figure 2–1).

If no liquid is present within the measurement gap, the ultrasonic sound wave is attenuated and the second crystal, called the receive crystal, detects only a small percentage of the original ultrasonic sound wave. This causes only a slight excitation of the receive crystal, resulting in a low amplitude output from the receive crystal.

If liquid is present within the measurement gap, a much larger percentage of the transmitted ultrasonic sound wave is detected by the receive crystal, resulting in a much higher amplitude output.

To measure aerated and attenuating liquids, the feedback gain must be increased to very high levels. The increased gain amplifies the signal received when the gap is filled with air or liquid.

Thermo Fisher Scientific Sapphire User Guide 2-1

Page 24: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Theory of Operation Time Gate Technology

Time Gate Technology

Thermo Fisher’s Time Gate technology pulses the transmit crystal, and the signal from the receive crystal is accepted in predictable, narrow time windows or time gates. The ultrasound received when the gap is dry travels along the bridge of the gap and reaches the receive crystal at a different time than the ultrasound that travels through the air or liquid in the gap. This method separates these signals and raises the sensitivity beyond the capabilities of the continuous feedback method.

Continuous Self-Test The Time Gate technology incorporates a continuous self-test feature to monitor system integrity. The unit uses the signal that travels along the bridge for its unique self-test. This signal is always present whether the gap is filled with air or liquid. It must always be present or the unit will indicate a fault. The crystals must be operational and properly bonded to the sensor to generate and receive the ultrasound. The continuous self-test is performed automatically and does not require the use of special self-test crystals or electronics in the sensor probe. The simpler self-test design and the unique ability to detect crystal disbonds add to the unit’s high level of reliability.

Figure 2–1. Ultrasonic point measurement

2-2 Sapphire User Guide Thermo Fisher Scientific

Page 25: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Theory of Operation Manual Loop Test

Manual Loop Test Loop Test Manual loop test is a feature that allows users to check or verify loop integrity, field wiring, and/or the operation of any downstream equipment activated by system outputs. This feature is momentarily activated when the supplied magnet is placed over the proper location on the housing cover marked “Loop Test.”

Manual loop test will change outputs to the opposite state when systems are not in a self-test fault or malfunction condition. Two-wire loop-powered units will change the current output to the opposite 8 or 16 mA state. Relay units will change to the opposite relay state.

Caution The manual loop test feature will not activate (reverse outputs) when the system is in a self-test fault or malfunction condition. ▲

System Qualification

Time

System Qualification

Time

This setting determines the amount of time that liquid must be continuously present within the measurement gap (in high level operation) or absent from the gap (in low level operation) before the process level alarm is indicated. At the point of switching from one condition to the next (from dry to wet), the system first "qualifies" that the new condition is continuously present and not a momentary occurrence. The new condition must be continuously present before the changing liquid level condition is indicated.

The implementation of a sensor qualification test assures that the actual liquid level has reached the measurement gap instead of a wave fluctuation due to agitation or turbulence. This setting can be field adjusted by the system qualification time potentiometer (R20) in the range of 0.3 to 12 seconds.

System Threshold

Adjustment

System Threshold

Adjustment

The threshold adjustment potentiometer (R9) determines the gap sensor's wet/dry switch point. The factory-configured default position for this setting is approximately mid-range and will operate properly for the majority of system applications. Counterclockwise rotation of the threshold potentiometer increases the threshold. Clockwise rotation decreases the threshold.

Note Due to the unique signal processing capability of the slot sensor, the threshold adjustment potentiometer (R9) is not required. This function is disabled on slot sensor units. ▲

Thermo Fisher Scientific Sapphire User Guide 2-3

Page 26: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Theory of Operation Relay Output Models

Relay Output Models

The relay output model provides one process relay for level alarm indication and a second auxiliary relay for self-test fault indication. Both relays use gold-plated contacts for reliable switching of lower current and voltage levels, eliminating the need for DCS users to install interposing relays. Process level alarm indication can be field selected for either high level or low level fail-safe operation by configuring the placement of the process level jumper (J4). Operation of the auxiliary fault relay can be field selected for either fault or mimic operation by the appropriate placement of the auxiliary fault jumper (J3). Table 2–1 defines relay alarm states for all possible combinations of process alarms, self-test faults, and power loss conditions.

Process Relay Alarm Operation

ocess Relay Alarm Operation

Relay output models indicate process level alarms through SPDT process relays. The SPDT process relay provides a normally closed (NC) and a normally open (NO) set of contacts. During normal operating conditions, the process relay is energized as long as the process level jumper (J4) is in the correct fail-safe position.

Relay output models operate in one of two field selectable fail-safe modes: high level or low level operation. The position of the field selectable process level jumper (J4) determines the fail-safe operation of the unit. When systems are field selected for high level operation, the process relay is energized when the measurement gap is dry and de-energized when the measurement gap is wet. When systems are set for low level operation, the process relay is energized when the measurement gap is wet and de-energized when the gap is dry.

Note System relay operation has been designed for fail-safe operation. A self-test failure or power loss will de-energize the process and auxiliary relays regardless of the actual state of the measurement gap. ▲

This is provided as a safety feature because reliable information regarding the state of the measurement gap cannot be assured during self-test failures. Therefore, interpretation of both the process and auxiliary relay status is necessary to verify actual process conditions.

2-4 Sapphire User Guide Thermo Fisher Scientific

Page 27: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Theory of Operation Relay Output Models

Auxiliary Fault Relay Operation

Relay output models indicate self-test fault conditions through a SPST auxiliary relay contact. The SPST auxiliary fault relay provides one field selectable NC or NO contact and is always energized during normal operating conditions.

Fault or mimic operation is also selectable through the auxiliary fault jumper (J3). When set for fault operation, the auxiliary relay de-energizes upon self-test failure or power loss. When set for mimic operation, a second set of contacts is available which responds in the same manner as the process alarm relay contacts, and the auxiliary fault relay de-energizes on process alarm, self-test failure, or power loss.

Table 2–1. Relay output combinations

Jumper & Sensor Condition No Power

MF MF Dry Dry Dry Dry Wet Wet Wet Wet

Level setting (J4): 1&2: Low level (8 wet/16 dry) 2&3: High level (8 dry/16 wet)

X

X

X

2–3

1–2

2–3

1–2

2–3

1–2

2–3

1–2

Malfunction jumper (J11): 1&2: 5 mA fault output 2&3: 19 mA fault output

X

X

X

X

X

X

X

X

X

X

X

Auxiliary relay (J3)*: 1&2: Fault operation 2&3: Mimic operation

X

X

X

1–2

1–2

1–2

1–2

1–2

1–2

1–2

1–2

Manual loop test (S3): 0 = No manual loop test (de-energized) 1 = Manual loop test activated (energized)

X

X

X

0

0

1

1

0

0

1

1

Process alarm relay output Auxiliary fault relay output*

0 0

0 0

0 0

1 1

0 1

0 0

1 0

0 1

1 1

1 0

0 0

Where: MF = Malfunction or self-test failure X = Jumper position (e.g. state) does not matter

WET = System sensor’s measurement gap/slot is physically Wet 0 = Relay is de-energized or manual loop test is not activated

DRY = System sensor’s measurement gap/slot is physically Dry 1 = Relay is energized or manual loop test is activated

1–2 = Jumper pins in position 1 & 2 2–3 = Jumper pins in position 2 & 3

* When auxiliary relay jumper (J3) is selected for mimic operation (pins 2 & 3), the auxiliary fault relay output will mimic the operation of the process alarm relay output.

Thermo Fisher Scientific Sapphire User Guide 2-5

Page 28: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Theory of Operation Two-Wire Models

Two-Wire Models

The two-wire, 4–20 mA DC model provides reliable level detection and lower installation costs for users with two-wire instrumentation. Table 2–2 defines the two-wire output for all possible process alarm, self-test, power loss, and main logic board jumper settings.

Process Alarm Indication

Two-wire models do not use relays to indicate process level alarm or self-test failures discussed above. Instead, these loop-powered models use a fixed mA output to indicate both process level alarms and self-test faults.

A wet measurement condition is always indicated by an output of 16 mA DC or 8 mA DC depending on the position of jumper J4. Likewise, a dry measurement condition is always indicated by an output of 8 mA DC or 16 mA DC depending on the position of jumper J4.

Self-Test Fault Indication

The position of the auxiliary fault jumper (J3) has no effect on the fixed mA output for these two-wire systems. By setting the malfunction jumper (J11), users can select whether 5 or 19 mA will indicate self-test failures.

In addition, self-test faults are only indicated when a self-test fault occurs. Process alarm indications are not mimicked or coupled to self-test failure indications.

Table 2–2. Two-wire output combinations

Jumper & Sensor Condition No Power

MF MF Dry Dry Dry Dry Wet Wet Wet Wet

Level setting (J4): 1&2: Low level (8 Wet/16 Dry) 2&3: High level (8 Wet/16 Dry)

X

X

X

2–3

1–2

2–3

1–2

2–3

1–2

2–3

1–2

Malfunction jumper (J11): 1&2: 5 mA fault output 2&3: 19 mA fault output

X

1–2

2–3

X

X

X

X

X

X

X

X

Auxiliary relay (J3): 1&2: Fault operation 2&3: Mimic operation

X

X

X

X

X

X

X

X

X

X

X

Manual loop test (S3): 0 = No manual loop test (de-energized) 1 = Manual loop test activated (energized)

X

X

X

0

0

1

1

0

0

1

1

Two-wire current output (mA) 0 mA 5 19 8 16 16 8 16 8 8 16

Where: MF = Malfunction or self-test failure X = Jumper position (e.g. state) does not matter

WET = System sensor’s measurement gap/slot is physically Wet 0 = Relay is de-energized or manual loop test is not activated

DRY = System sensor’s measurement gap/slot is physically Dry 1 = Relay is energized or manual loop test is activated

1–2 = Jumper pins in position 1 & 2 2–3 = Jumper pins in position 2 & 3

2-6 Sapphire User Guide Thermo Fisher Scientific

Page 29: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Theory of Operation Optional Thermal Isolator

Thermo Fisher Scientific Sapphire User Guide 2-7

Table 2–3. Relay terminal combinations

Unit Power Process Alarm

Self-Test Fault

Auxiliary Jumper

Process Relay Auxiliary Relay

NO NC If NC If NO

On No No Fault C O O C

On No No Mimic C O O C

On Yes No Fault O C O C

On Yes No Mimic O C C O

On No Yes Fault O C C O

On No Yes Mimic O C C O

On Yes Yes Fault O C C O

On Yes Yes Mimic O C C O

Off No No Fault O C C O

Off No No Mimic O C C O

Off Yes No Fault O C C O

Off Yes No Mimic O C C O

Off No Yes Fault O C C O

Off No Yes Mimic O C C O

Off Yes Yes Fault O C C O

Off Yes Yes Mimic O C C O

Optional Thermal Isolator

ptional Thermal Isolator

Systems with 3/4-inch metal NPT sensors can extend the operating temperature range of the sensor when ordered with the optional thermal isolator. The thermal isolator allows sensors with 3/4-inch NPT fittings to be installed in process applications above 275°F/135°C (up to 320°F/160°C). Flanged units have built-in thermal isolation in the flange nipple and remote mounted systems do not require the use of the thermal isolator. The thermal isolator is an integral part of the sensor assembly and can only be supplied with a new sensor. It is not available as an "add-on" option to an existing sensor.

Caution The thermal isolator is an integral part of the sensor assembly. Do not attempt to remove it from the sensor. Removing or loosening the thermal isolator can damage the sensor. ▲

Page 30: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 31: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Chapter 3 Special Handling Considerations

Caution These systems are electrostatic discharge (ESD) sensitive. Use proper ESD protective equipment and follow ESD handling procedures when servicing or handling the instrument. Failure to comply with ESD procedures can result in circuit damage. ▲

ESD Procedures SD Procedures These systems are ESD sensitive instruments. They contain electronic components that can be damaged from discharges of static electricity.

The following procedures should be observed when installing, setting up, servicing, troubleshooting, and repairing the system:

1. Use an antistatic bag. Electronic subassemblies are shipped in special antistatic bags. Keep the assembly in the bag as often as possible whenever the subassembly is not installed.

2. ESD sensitive subassemblies should be removed from their antistatic bags only:

a. At a designated static-free workstation or when the bag is grounded at the field site

b. After conductive area of container has been neutralized

c. After you have made firm contact with an antistatic mat and/or firmly gripped by a grounded individual

3. Personnel handling ESD sensitive devices should be neutralized to a static-free workstation by a grounding wrist strap that is connected to the station or to a good grounding point at the field site.

4. Do not allow clothing to make contact with ESD sensitive devices.

5. Avoid touching connectors and components.

Thermo Fisher Scientific Sapphire User Guide 3-1

Page 32: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Special Handling Considerations Unpacking, Inspection, & Storage

3-2 Sapphire User Guide Thermo Fisher Scientific

6. Avoid partial connection of ESD sensitive devices. These devices can be damaged by floating leads, especially the power supply connector.

7. Ground test equipment.

8. Avoid static charges during troubleshooting.

Unpacking, Inspection, & Storage

acking, Inspection, & Storage

Note Inspection, adjustment, installation, and maintenance of this unit must be performed by experienced personnel. Refer to “Personnel” in Chapter 1. ▲

Upon receipt, examine the instrument for damage that occurred in transit. If damage is found or there is evidence of rough handling, file a damage claim with the carrier and contact your representative.

Carefully inspect the packing material before discarding it to make certain that all equipment and instruction paperwork has been removed. Careful handling and installation will ensure satisfactory performance of the system.

Use the original packaging material and container for storage if necessary. The storage environment should be protected and free from extremes of temperatures and high humidity and fall within the environmental constraints listed in the performance specification section of this manual.

Hazardous Location Installations

Hazardous Location Installations

CSA units are available for installation in environmentally hazardous and corrosive locations. See the Appendix B for specific agency approval classifications.

Warning Sapphire systems may be used only in those hazardous locations identified on the product nameplates. ▲

Page 33: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Chapter 4 Installation

Warning Installation must be carried out in accordance with local site requirements and regulations. ▲

Reference the drawing appendix for system drawings.

Factory Default Settings

actory Default Settings

Each system is factory-configured and shipped with the following default settings. Refer to Figures 4–1 and 4–2 for locations of the field selectable jumpers and potentiometers.

Table 4–1. Default settings

Field Selectable Adjustment Factory Default Setting

Process level jumper (J4) High level operation (2 & 3)

Auxiliary fault jumper (J3) Fault mode operation (1 & 2)

System qualification time potentiometer (R20) Minimum time (0.3 seconds)

Threshold adjustment potentiometer (R9) Nominal (center position)

Malfunction jumper (J11) 19 mA (2 & 3)

NO or NC jumper (J6) NO (2 & 3)

Figure 4–1. Main logic PCB field selectable adjustments

Thermo Fisher Scientific Sapphire User Guide 4-1

Page 34: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Field Setup Procedures

Figure 4–2. Power supply board NO/NC field selection

Field Setup Procedures

ield Setup Procedures

Warning Electrical installation must be made in compliance with application area classification codes. Only qualified personnel should perform installation. ▲

All field selectable adjustments should be made prior to installing the system in process locations. Bench adjustments are recommended to avoid exposure of sensitive circuit board components to process environments.

Thermo Fisher recommends performing the following prior to final system installation in process locations.

Caution These systems are ESD sensitive instruments. Use proper ESD protective equipment and follow ESD handling procedures when servicing or handling the instrument. Failure to comply with ESD procedures can result in circuit damage. ▲

1. Remove the housing covers from the circuit and connection sides of the dual compartment housing.

2. Customize the main logic PCB’s field selectable adjustments for your specific application as described below:

a. Process level setting (J4): Locate jumper J4 on the main logic PCB. For high level operation (default), place the 2-pin jumper on pins 2 and 3. For low level operation, place the jumper on pins 1 and 2. Place J4 across pins 1 and 2 to have 8 mA for wet condition and 16 mA for dry condition. Place J4 across pins 2 and 3 to have 8 mA for dry conditions and 16 mA for wet condition.

4-2 Sapphire User Guide Thermo Fisher Scientific

Page 35: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Field Setup Procedures

b. Auxiliary relay (J3): Locate jumper J3. For fault operation (default), place the 2-pin jumper on pins 1 and 2. For mimic operation, place the jumper on pins 2 and 3.

c. Malfunction jumper (J11): Locate jumper J11. The position of J11 determines the current output during self-test or malfunction conditions. The malfunction jumper is applicable to only two-wire output systems. For 19 mA output during malfunction conditions (default), place J11 on pins 2 and 3. For 5 mA output during malfunction conditions, place J11 on pins 1 and 2.

d. System qualification time (R20): Locate potentiometer R20. Leave the potentiometer fully counterclockwise (0.3 seconds default) unless the process has excessive agitation and splashing.

e. NO/NC setting (J6): Remove the main logic PCB and locate jumper J6 on the power supply board. For NO operation (default), place the 2-pin jumper in pins 2 and 3. For NC operation, place jumper J6 in pins 1 and 2. Reinstall the logic board.

3. Connect incoming power to terminals L1 and L2 of the top terminal strip. Refer to Figure 4–3 for location and identification of terminal connections. Do not turn on power at this time.

Figure 4–3. Field wiring connections

Thermo Fisher Scientific Sapphire User Guide 4-3

Page 36: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Threshold Adjustment Verification

Threshold Adjustment Verification

Although the threshold adjustment potentiometer (R9) is factory calibrated for nominal conditions and should operate properly for the majority of applications, we recommend testing functionality using actual process liquids to verify the factory calibrated threshold adjustment.

Relay Output Models elay Output Models Follow this procedure to verify and adjust the threshold adjustment potentiometer (R9).

Warning Hazardous voltages can be present on terminals. Voltages can cause injury or death. Proceed with extreme caution. ▲

1. Connect the test leads of a digital voltmeter (DVM) to the process terminals 2C (process contact common) and 2NC (process contact NC). Set the DVM for continuity measurement.

2. Turn on power to instrument.

3. Check continuity across the process alarm relay contacts. For high level operation, the DVM should read open. For low level operation, the DVM should show continuity. Hold the loop test magnet next to housing at location marked “Loop Test.” The process alarm and fault relays should switch.

4. Fill a container with at least four inches of representative sample for typical process conditions.

5. Immerse the sensor in the liquid far enough to cover the entire gap or slot.

6. Within the field selected system qualification time (0.3 to 12.0 seconds), the process alarm relay should change state and de-energize. For high level operation, the DVM should show continuity. For low level operation, the DVM should read open.

7. Remove the sensor from liquid. Within the system qualification time, the process alarm relay should change state and energize.

4-4 Sapphire User Guide Thermo Fisher Scientific

Page 37: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Threshold Adjustment Verification

8. If the process alarm relay did not change states as described in steps 6 and 7, adjust the threshold adjustment potentiometer (R9) as follows:

a. If the process alarm relay did not change state within the qualification time when immersed in the liquid, decrease the threshold by rotating potentiometer R9 1/8-turn clockwise. Re-test the unit and, if necessary, adjust in 1/8-turn increments until the process alarm changes state within the qualification time.

b. If process alarm relay does not change state within the qualification time when removed from the liquid, increase the threshold by rotating potentiometer R9 1/8-turn counterclockwise. Re-test the unit, and if necessary, adjust in 1/8-turn increments until the pro-cess alarm relay changes state within the qualification time.

Two-Wire Models For two-wire, 4–20 mA units, the same threshold adjustment procedures can be followed as described in above. However, instead of monitoring relay contacts for wet/dry indication, the mA output should be measured for indication of 8 mA, ±1 mA (dry condition) or 16 mA, ±1 mA (wet condition) if jumper J4 is in 1-2 position.

1. Connect a current meter in series with the DC power supply as shown in Figure 4–4.

2. Turn on power to instrument.

3. Check the mA output on the meter for a dry sensor. The output should read 8 mA, ±1 mA. Hold the loop test magnet over the circuit housing cover at the location marked “Test.” The output should read 16 mA, ±1 mA.

4. Fill a container with at least four inches of a representative sample for typical process conditions.

5. Immerse the sensor in the liquid far enough to cover the entire gap or slot.

6. Within the system qualification time, the output should change to 16 mA, ±1 mA.

7. Remove the sensor from the liquid. Within the system qualification time, the output should switch back to 8 mA, ±1 mA.

Thermo Fisher Scientific Sapphire User Guide 4-5

Page 38: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Self-Test Verification

8. If the output did not change state as described in steps 6 and 7, adjust the threshold adjustment potentiometer (R9) as follows:

a. If the analog output did not change state within the qualification time when immersed in the liquid, decrease the threshold adjustment by rotating potentiometer R9 1/8-turn clockwise. Test the unit again and, if necessary, adjust in 1/8-turn increments until the analog output changes state within the qualification time.

b. If analog output does not change state within the qualification time when removed from the liquid, increase the threshold adjustment by rotating potentiometer R9 1/8-turn counterclockwise. Re-test the unit, and if necessary, adjust in 1/8-turn increments until the analog output changes state within the qualification time.

Figure 4–4. 4–20 mA test wiring diagram

Self-Test Verification With the sensor configured for your specific application, perform a final check of the auxiliary relay (relay models only) and the self-test function.

Relay Output Models 1. With the sensor outside of the liquid, check for continuity across the auxiliary relay terminals 1C and 1NC or 1NO (auxiliary relay common and normally closed or normally open contact depending on position of jumper J6 on the power supply board).

2. Refer to Table 2–1 to check the auxiliary relay contact operation for proper fault/mimic operation.

3. Submerge the sensor in the liquid. Refer to Table 2–1 for proper auxiliary relay contact operation. Verify with the DVM. Monitor the operation of the process relay. Verify that a self-test fault is not indicated.

4-6 Sapphire User Guide Thermo Fisher Scientific

Page 39: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Guidelines

Two-Wire Models ire Models 1. The two-wire, 4–20 mA unit does not use an auxiliary relay output to indicate a self-test failure. A properly functioning unit (i.e., no self-test failure) will only provide the appropriate mA output corresponding to the state of the measurement gap (either wet or dry) – either 8 mA or 16 mA, depending on the position of jumper J4. Refer to Table 2–2 for proper operation.

2. If a self-test failure is present, the output should read either 5 mA or 19 mA, depending on the position of jumper J11.

3. Verify that no self-test failure is present.

The system is now ready for field installation. Remove power from the unit, disconnect the wiring, and replace the housing covers.

Installation Guidelines

Installation Guidelines

Warning The system may be used only in those hazardous locations identified on the product nameplate. These installation and servicing instructions are for use by qualified personnel only. ▲

This system has been designed for long, trouble-free service and has been tested at the factory. However, a good installation is the key to good performance. Thermo Fisher recommends setting the configurable adjustments according to your application (see “Field Setup Procedures”) and testing prior to final installation. The following installation guidelines should be considered when selecting a mounting location.

● The unit should be kept out of the direct flow of liquid during filling operations to avoid erroneous indications. If necessary, a baffle may be used to divert the liquid.

● Avoid installing the sensor in areas of high turbulence or agitation such as pump discharge or suction. This can cause air pockets to be present in the measurement gap which may cause false process alarm indications.

● Make sure that a clear space exists inside the vessel or pipe for the sensor length. The measurement gap should be clear of the coupling and wall to avoid forming a pocket which can trap liquid or air. For aerated applications, mount the sensor horizontally, paying particular attention to the orientation of the sensor bridge (Figure 4–6).

● If reducing bushings are used directly in certain smaller pipe sizes, be sure the measurement gap is beyond the bushings.

Thermo Fisher Scientific Sapphire User Guide 4-7

Page 40: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Guidelines

● The unit may be mounted in any position desired. Horizontal mounting is recommended in highly viscous liquids or liquids containing sludge and debris.

● Proper sensor orientation is important for reliable operation of the instrument. In horizontal mounting applications, the measurement gap must be oriented so that liquid can easily flow through the gap or slot. Proper horizontal orientation of the sensor is indicated in Figure 4–5.

● For proper alignment of the gap or slot when installing the probe, a bridge alignment mark is stamped on the hex nipple or flange edge at the top of the sensor. This bridge alignment mark indicates the sensor bridge location for the gap sensor and the sensor body side for the slot sensor. Refer to Figure 4–6.

● Be certain that all wiring, conduit, and electrical fittings conform to local electrical codes for the location selected.

Figure 4–5. Horizontal installation

4-8 Sapphire User Guide Thermo Fisher Scientific

Page 41: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

Figure 4–6. Sensor bridge alignment markings

Installation Procedures

Caution Sensors are fragile instruments. Handle with care. Do not support or carry systems by the sensors. Sensor breakage may cause fluids to enter electronics housing. The sensor, flange (if applicable), and electronics have been carefully assembled at the factory. Disassembly of the system is not required for installation and may result in damage to the sensor. ▲

3/4” NPT Mount – Integral Systems 4” NPT Mount –

Integral Systems 1. Carefully remove the system from the shipping container. Inspect for

signs of damage that may have occurred during shipment.

2. Apply Armite Laboratories anti-seize compound LP-250 to the 3/4-inch NPT process threads.

3. Carefully insert the sensor into the 3/4-inch NPT connections. Lower the unit into position. Do not force it into position.

4. Slowly hand-tighten the sensor thread. Avoid cross-threading. Continue tightening the fitting with a wrench on hex nut until fitting is leak-tight. Do not twist electrical housing as damage to sensor leads can occur.

5. Make sure the measurement gap is positioned properly when complete.

Thermo Fisher Scientific Sapphire User Guide 4-9

Page 42: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

6. Remove the connection (field wiring) compartment housing cover. Feed process signal/alarm and power wires through the 1/2-inch NPT conduit inlet and into the connection compartment of the housing.

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

7. Secure the conduit to the 1/2-inch NPT electrical connections.

8. Refer to “Field Setup Procedures” earlier in this chapter for installation and connection of process signal/alarm and power wiring.

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲

Note Sapphire system enclosures installed in Class I, Div. 1 Hazardous (Classified) Locations are factory sealed and require a poured conduit seal only if the conduit run leaves said location or if the conduit run enters another enclosure requiring a poured seal. ▲

9. Replace the connection compartment housing cover, tighten until fully seated, and turn on power.

10. Installation is complete.

Flange Mount Sensors – Integral Systems

ange Mount Sensors – Integral Systems

1. Carefully remove the system from shipping container. Inspect for signs of damage that may have occurred during shipment.

Caution Do not support or carry flange mounted unit by the sensors. Handle these units by the flange to avoid damaging the sensor. ▲

2. Carefully insert the sensor into the flange opening. Lower the unit into position. Do not force it into position.

3. Bolt the flange with the appropriate hardware. Do not twist electrical housing as damage to sensor leads can occur.

4-10 Sapphire User Guide Thermo Fisher Scientific

Page 43: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

4. Make sure measurement gap or slot is positioned properly when complete.

5. Remove the connection (field wiring) compartment housing cover. Feed the process signal/alarm and power wires through the 1/2-inch NPT conduit inlet and into the connection compartment of the housing.

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

6. Secure the conduit to the 1/2-inch NPT electrical connections.

7. Refer to “Field Setup Procedures” earlier in this chapter for installation and connection of process signal/alarm and power wiring.

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲

Note A system enclosure installed in Class I, Div. 1 Hazardous (Classified) Location is factory sealed and requires a poured conduit seal only if the conduit run leaves the installation location or if the conduit run enters another enclosure requiring a poured seal. ▲

8. Replace the connection compartment housing cover, tighten until fully seated, and turn on power.

9. Installation is complete.

Remote Mount Installations mote Mount

Installations The remote mount Sapphire sensor design provides a non-integral sensor and electronics assembly. The non-integral feature allows installation of the electronics in an orientation where easy access to the circuit boards and field wiring is possible without concern to the constraints of the sensor locations.

Refer to drawing 952-000099 in the drawing appendix and the instructions that follow.

Thermo Fisher Scientific Sapphire User Guide 4-11

Page 44: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

Thermo Fisher provides the following items on a remote mount system:

● Electronics assembly and housing

● Adapter plate assembly (installed on electronics assembly by Thermo Fisher)

● Sensor assembly (flange or NPT mount)

● Wall or pipe mounting bracket (optional)

All interconnecting conduit, fittings, seals, and connection boxes required to connect the sensor assembly to the electronics assembly per applicable electrical codes are supplied and installed by the user.

Note Installation must be done in accordance with local electrical codes. ▲

1. Carefully remove system components from the shipping container, and inspect for damage that may have occurred during shipment.

3/4” NPT Mount – Remote Systems 4” NPT Mount –

Remote Systems

Caution The sensor assembly is provided with a pre-determined length (per customer order) of sensor cables. Use care to avoid damaging the cable or phono plugs during installation. ▲

2. Apply Armite Laboratories anti-seize compound LP-250 to the 3/4-inch NPT process threads of the sensor assembly.

3. Carefully insert the sensor into the 3/4-inch NPT connections. Lower the unit into position. Do not force it into position.

4. Slowly hand-tighten the sensor thread. Avoid cross-threading. Continue tightening the fitting with a wrench on hex nut until fitting is leak-tight.

5. Make sure the measurement gap is positioned properly when complete.

6. Ensure earth safety ground is properly connected to sensor.

7. Determine the location for electronics assembly. The location should be such that conduit routing between the sensor assembly and electronics assembly can be made without sharp bends which could kink the sensor leads.

4-12 Sapphire User Guide Thermo Fisher Scientific

Page 45: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

Note If optional pipe or wall mounting brackets are used, mount the bracket to a solid structure or pipe prior to mounting the electronics to the bracket. ▲

8. Install the electronics assembly to the mounting bracket (optional) or other solid structure using the supplied mounting hardware. Take care not to damage electronic leads or phono plugs during installation.

9. Install the certified/approved connection box to the adapter plate assembly by feeding the sensor leads into the box and tightening the box to 3/4-inch NPT threads.

10. Complete the conduit and seal connection between the sensor assembly and connection box while feeding sensor leads/cables through the conduit and into the connection box.

Note A minimum of five full threads is required between all threaded joints. ▲

11. Connect the phono plugs of the sensor assembly to the electronics assembly at the connection box (red-red, black-black). Replace the connection box cover.

Caution Due to the nature of the piezoelectric crystals used in the probes of Sapphire systems, Thermo Fisher recommends discharging each signal cable by connecting the shield (case) to center conductor (pin) before connecting to the main logic board or remote mounting bracket signal cables. ▲

12. Pour seals in accordance with manufacturer’s instructions.

13. Remove the connection (field wiring) compartment housing cover. Feed the process signal/alarm and power wires through the 1/2-inch NPT conduit inlet and into the connection compartment of the housing.

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

14. Secure the conduit to the 1/2-inch NPT electrical connections.

Thermo Fisher Scientific Sapphire User Guide 4-13

Page 46: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

15. Refer to “Field Setup Procedures” earlier in this chapter for installation and connection of process signal/alarm and power wiring.

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲

Note A system enclosure installed in Class I, Div. 1 Hazardous (Classified) Location) is factory sealed and requires a poured conduit seal only if the conduit run leaves the installation location or if the conduit run enters another enclosure requiring a poured seal. ▲

16. Replace the housing cover, tighten until fully seated, and turn on power.

17. Installation is complete.

1. Carefully remove system components from the shipping container, and inspect for signs of damage that may have occurred during shipment.

Flange Mount Sensors – Remote Systems

ange Mount Sensors – Remote Systems

Caution The sensor assembly is provided with a pre-determined length (per customer order) of sensor cables. Use care to avoid damaging the cable or phono plugs during installation. ▲

2. Install the flange gasket.

3. Carefully insert the sensor into the flange opening. Lower the unit into position. Do not force it into position.

4. Bolt the flange using the appropriate hardware.

5. Ensure the measurement gap is positioned properly when complete.

6. Ensure earth safety ground is properly connected to sensor.

7. Determine the location for electronics assembly. The location should be such that conduit routing between the sensor assembly and the electronics assembly can be made without sharp bends which could kink the sensor leads.

4-14 Sapphire User Guide Thermo Fisher Scientific

Page 47: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

Note If optional pipe or wall mounting brackets are used, mount the bracket to a solid structure or pipe prior to mounting the electronics to the bracket. ▲

8. Install the electronics assembly to mounting bracket (optional) or other solid structure using the supplied mounting hardware. Take care not to damage the electronic leads or phono plugs during installation.

9. Install the certified/approved connection box to the adapter plate assembly by feeding the sensor leads into the box and tightening the box to the 3/4-inch NPT threads.

10. Complete the conduit and seal connection between the sensor assembly and connection box while feeding the sensor lead through the conduit and into the connection box.

Note A minimum of five full threads is required between all threaded joints. ▲

11. Connect the phono plugs of the sensor assembly to the electronics assembly at the connection box (red-red, black-black). Replace the connection box cover.

Caution Due to the nature of the piezoelectric crystals used in the probes of Sapphire systems, Thermo Fisher recommends discharging each signal cable by connecting the shield (case) to center conductor (pin) before connecting to the main logic board or remote mounting bracket signal cables. ▲

12. Pour seals in accordance with manufacturer’s instructions.

13. Remove the connection (field wiring) compartment housing cover. Feed the process signal/alarm and power wires through the 1/2-inch NPT conduit inlet and into the connection compartment of the housing.

Note The connection compartment can be identified by the product label (“Conn”) located on the side of the dual compartment electrical housing. ▲

14. Secure the conduit to the 1/2-inch NPT electrical connections.

Thermo Fisher Scientific Sapphire User Guide 4-15

Page 48: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Installation Installation Procedures

4-16 Sapphire User Guide Thermo Fisher Scientific

15. Refer to “Field Setup Procedures” earlier in this chapter for installation and connection of process signal/alarm and power wiring.

Note Install process signal/alarm and power wiring in accordance with applicable electrical codes. ▲

Note A system enclosure installed in Class I, Div. 1 Hazardous (Classified) Location) is factory sealed and requires a poured conduit seal only if the conduit run leaves the installation location or if the conduit run enters another enclosure requiring a poured seal. ▲

16. Replace the housing cover, tighten until fully seated, and turn on power.

17. Installation is complete.

Page 49: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Chapter 5 Maintenance

Electronics Electronics Sapphire systems are precision electronic instruments which incorporate no moving parts in their design. Because of this, little maintenance is required. However, as with any electronic instrument, exposure of the circuit boards and connections to harsh plant environments can cause long-term erratic operation.

To prevent the exposure of electrical components, conduct periodic inspections of the electrical fittings (conduit couplings, housing, connector, etc.) for looseness or separation. Tighten as required. In addition, the compartment housing contains O-rings to seal the compartments from moisture. Inspect them periodically for indication of cracks or brittleness and replace if necessary.

Sensors Sensors Periodically inspect the sensor for buildup of product on the sensor body. Removal of product from the sensor to prevent improper operation should be done as required.

Caution Do not clean with saturated steam greater than 50 psig. Repeated steam cleaning can potentially shorten sensor life. ▲

Circuit Board Replacement Circuit Board Replacement

Table 5–1. Replacement electronics

Model Description P/N

J 12–55 Vdc input, 4–20 mA output (IS/EP/DIP/NI) (2 board set) 340-000251

F 15–35 Vdc input, relay output (NO or NC selectable auxiliary fault relay) (2 board set)

340-000248

G 115 Vac input, relay output (NO or NC selectable auxiliary fault relay) (2 board set)

340-000249

S 230 Vac input, relay output (NO or NC selectable auxiliary fault relay) (2 board set)

340-00250

15–35 Vdc input board (single board) 340-000482

230 Vac input board (single board) 340-000484

Thermo Fisher Scientific Sapphire User Guide 5-1

Page 50: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Maintenance Procedure

Model Description P/N

4/20 mA 12–55 Vdc input board (single board) 340-000485

Main logic board for all voltage ranges 340-000486

Table 5–2. Replacement kits

Kit # Description

340-000248 1 logic PCB assembly 1 relay/power supply PCB with 15–35 Vdc, field selectable NC/NO auxiliary relay 3 nylon screws, 6-32 x 3/8 1 installation instruction

340-000249 1 logic PCB assembly 1 relay/power supply PCB with 115 Vac, field selectable NC/NO auxiliary relay 3 nylon screws, 6-32 x 3/8 1 installation instruction

340-000250 1 logic PCB assembly 1 relay/power supply PCB with 230 Vac, field selectable NC/NO auxiliary relay 3 nylon screws, 6-32 x 3/8 1 installation instruction

340-000251 1 main logic PCB 1 EMC filter PCB with 12–55 Vdc, 4–20 mA IS output 3 nylon screws, 6-32 x 3/8 1 installation instruction

Procedure The following instructions apply only when using the kits listed in Table 5–2.

Warning When disconnecting power from the instrument, follow proper lockout/tagout procedures to avoid accidental exposure of personnel to high voltage circuits. ▲

Caution Sapphire systems are ESD sensitive instruments. Use proper ESD protective equipment and follow ESD procedures when handling the instrument. Failure to comply with ESD procedures can result in circuit damage. ▲

5-2 Sapphire User Guide Thermo Fisher Scientific

Page 51: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Maintenance Procedure

1. Turn off power to the system.

2. Remove the housing cover from the circuit side (side opposite the set screw).

3. Remove the three nylon screws that secure the logic PCB.

4. Lift the logic PCB until its back is accessible.

5. After disconnecting the circuit board interconnect cable, disconnect both sensor leads from the back of the logic board.

6. Note the part number and revision level of logic PCB as designated on the top side (front) of the board. It should read: 220-000453.

7. Locate and record the position (1–2 or 2–3) of the following jumpers :

J3: ___________________

J4: ___________________

J11: __________________

8. Using a trimpot adjuster, rotate the sensitivity potentiometer (R9) fully counterclockwise and record how much the potentiometer turned before it seated (e.g. 1/8, 1/4, 1/2 turn).

Potentiometer turns: _______

9. Using a trimpot adjuster, rotate the qualification time potentiometer (R20) fully counterclockwise and record how much the potentiometer turned before it seated (e.g. 1/8, 1/4, 1/2 turn).

Potentiometer turns: _______

10. Unscrew the three standoffs from the relay/power supply PCB or EMC filter PCB and remove this board. Do not lose these standoffs as you will use them with the new boards. Record the position of jumper J6 on the relay/power supply PCB.

J6: _______

Caution Replacement boards are sent in sets. Do not reuse the old circuit board. Improper operation or damage can occur. ▲

Thermo Fisher Scientific Sapphire User Guide 5-3

Page 52: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Maintenance Procedure

11. Remove the new logic PCB from the anti-static bag. Handle with care and follow ESD handling procedures.

12. Configure the jumpers on the front of the logic PCB exactly as recorded in step 7. If necessary, configure the jumpers per Figure 5–1.

Caution Components L1 and R37 of the logic PCB assembly are factory calibrated. Do not adjust these components. Improper operation or damage can occur. ▲

13. Locate jumper J6 on the new power supply PCB. For NO operation, place the 2-pin jumper on pins 1 and 2. For NC operation, place the jumper on pins 2 and 3.

14. Orient the new relay/power supply PCB in the housing so that the mounting holes are aligned. Push the board onto the terminal strip pins. Install the standoffs to secure the circuit board.

Note When installing a DC relay/power supply PCB (P/N 220-000450), secure the solder lug from the board against the board using the closest standoff. ▲

15. Connect the sensor leads to the back of the logic PCB. Orientation is not important.

Caution Due to the nature of the piezoelectric crystals used in the probes of Sapphire systems, Thermo Fisher recommends discharging each signal cable by connecting the shield (case) to the center conductor (pin) before connecting to the main logic PCB or remote mounting bracket signal cables. ▲

16. Connect the board interconnect cable from the relay/power supply or EMC filter PCB to the new logic PCB.

Note The interconnect cable is keyed to prevent incorrect orientation. ▲

17. Place the logic PCB in position and install the three new nylon screws, being careful not to damage sensitive board components.

18. Board replacement is now complete.

5-4 Sapphire User Guide Thermo Fisher Scientific

Page 53: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Maintenance Procedure

Figure 5–1. Circuit board replacement

Thermo Fisher Scientific Sapphire User Guide 5-5

Page 54: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Maintenance Field Adjustments

5-6 Sapphire User Guide Thermo Fisher Scientific

Field Adjustments If the system is not installed in the process or can be easily removed from the process, follow the instructions provided in Chapter 4.

System Sensitivity Adjustment

If the system cannot be removed from the process:

1. Rotate the sensitivity potentiometer (R9) on the logic PCB fully counterclockwise.

2. Rotate the potentiometer clockwise the number of turns recorded in step 8 of the previous section. In most cases this will provide proper operation.

3. Monitor system operation for any problems. Refer to the Chapter 6 for troubleshooting procedures if a problem occurs.

Note Due to the unique signal processing capability of the slot sensor, the threshold adjustment potentiometer (R9) is not required. Slot sensor configured units have this function disabled. ▲

System Qualification Time Adjustment

tem Qualification Time Adjustment

You can adjust system qualification time between 0.3 and 12 seconds using the system qualification time potentiometer (R20). If the system cannot be removed from the process:

1. Rotate the qualification time potentiometer (R20) on the logic PCB fully counterclockwise.

2. Rotate the potentiometer clockwise the amount of turns recorded in step 9 of “Procedure.”

Final Inspections Final Inspections 1. Inspect the O-ring for cracks or brittleness. Replace if necessary (P/N 664-000104).

2. Install housing cover. Turn on system power.

Page 55: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Chapter 6 Troubleshooting & Service

General Sapphire systems have been designed with state-of-the-art surface mount board technology. This technology, along with the continuous self-test circuitry, should provide long-term, reliable instrument operation.

The self-test circuit continuously monitors the operation of the electronics and sensor and provides a fault indication if an electrical problem occurs or if the crystals disbond. A problem which results in a self-test fault typically requires return of the unit to Thermo Fisher for evaluation and repair.

Should an operational problem occur and the self-test circuit does not indicate a fault condition, check the following list of possible conditions that could prevent proper indication of the level system although the unit is functioning normally.

Note This troubleshooting section is a general list of possible conditions that could exist. This listing is not exhaustive and not meant to address all possible problems. Contact Thermo Fisher if a condition occurs that is not covered in this manual. ▲

Table 6–1. General troubleshooting

Condition Possible Cause

Sensor does not indicate level alarm when liquid is present in the gap or slot.

- Process level jumper (J4) set to low level operation. Refer to “Field Setup Procedures” in Chapter 4. - Threshold adjustment set too high. Refer to “Threshold Adjustment Verification” in Chapter 4. - Process liquid is too highly aerated. Contact Thermo Fisher.

System indicates process level alarm with no liquid present in the gap or slot.

- Process level jumper (J4) set to high level operation. Refer to “Field Setup Procedures” in Chapter 4. - Process liquid has created buildup in the gap or slot measurement area. Clean gap or slot sensor and establish maintenance interval.

Thermo Fisher Scientific Sapphire User Guide 6-1

Page 56: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Troubleshooting & Service Service

6-2 Sapphire User Guide Thermo Fisher Scientific

Condition Possible Cause

Process level alarm fluctuates on and off.

- Process liquid level is fluctuating near sensor location due to agitation or turbulence. Increase sensor qualification time from 0.3 second to 12.0 seconds. Refer to “Field Setup Procedures” in Chapter 4.

Self-test fault continues after electronics or system replacement.

- High levels of environmental noise may be present. Contact Thermo Fisher.

Service If you have completed the troubleshooting steps in the previous section and the unit still is not performing satisfactorily, the local Thermo Fisher representative is your first contact for support and is well equipped to answer questions and provide application assistance. You can also contact Thermo Fisher directly.

In the United States: Thermo Fisher Scientific 1410 Gillingham Lane Sugar Land, TX 77478 Phone: 713-272-0404 Fax: 713-272-2272

In Canada: Thermo Fisher Scientific 14 Gormley Industrial Avenue Gormley, Ontario L0H 1G0 Phone: 905-888-8808 Fax: 905-888-8828

In Europe: Thermo Fisher Scientific Ion Path Road Three Winsford Cheshire CW7 3GA United Kingdom Phone: +44 1606549700

On the Web: www.thermo.com

Page 57: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Appendix A Ordering Information

Integral Mount System

Table A–1.

Code System

SASP Sapphire ultrasonic point level system

Code Approvals

C Canadian Standards Association (Canada and U.S.)

Code Electronics

J 12–55 Vdc input, 4–20 mA intrinsically safe output

F 15–35 Vdc input, relay output, NO or NC user-selectable auxiliary fault relay

G 115 Vac input, relay output, NO or NC user-selectable auxiliary fault relay

S 230 Vac input, relay output, NO or NC user-selectable auxiliary fault relay

Code Mounting Configuration

O Internal electronics and sensor

T Thermal isolator (required for 3/4” NPT units with process temperatures above 275°F/135°C)

Code Sensor Type

G Gap

S Slot (316SS, 316SS polished, Hastelloy C)

Code Sensor Material (same as fitting/flange material)1

S 316SS (gap and slot)

V 316SS polished (gap and slot)

H Hastelloy C (gap and slot)

U MONEL (gap only)

B Hastelloy B (gap only)

Q Carpenter 20 (gap only)

Code Measurement for “A” Dimension

T Inches

M Millimeters

Thermo Fisher Scientific Sapphire User Guide A-1

Page 58: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Ordering Information Integral Mount System

Code Sensor “A” Dimension3, 4

0##0

Sensors available in 0.5” increments only (316SS) 0010–1200 for gap sensors (0010 is 1.0”), minimum 1” (316SS, 316SS polished) 0040–1200 for slot sensors (1200 is 120.0”), minimum 4” (Hastelloy C)

0##0 Sensor available in 5 mm increments only (MONEL) 0025–3050 for gap sensors (0025 is 25 mm), minimum 25 mm (Hastelloy B) 0100–3050 for slot sensors (3050 is 3050 mm), minimum 100 mm (Carpenter 20)

Code Fitting/Flange Type1, 2

N 3/4” NPT (material same as sensor)

F Flange 150 lb ANSI Raised Face (metal)

G Flange 300 lb ANSI Raised Face (metal)

T Tri-Clover (16 AMP) sanitary fitting and 150 grit polished

Code Fitting/Flange Size2

00 3/4” NPT only

10 1” flange or Tri-Clover fitting

15 1.5” flange or Tri-Clover fitting

20 2” flange or Tri-Clover fitting

25 2.5” Tri-Clover only

30 3” flange or Tri-Clover fitting

40 4” flange or Tri-Clover fitting

60 6” flange only

Code Fitting/Flange Material (same as sensor)

S 316SS

V 316SS polished (Tri-Clover 3A fitting and 3/4" NPT only)

H Hastelloy C

U MONEL

B Hastelloy B

Q Carpenter 20

Code Accessories (optional)

T Stainless steel tag

1Fitting/flange materials must be the same as the sensor material for all but Tri-Clover fittings. Tri-Clover fittings are only available in 316L and must be used only with 316SS sensors.

2Order in 0.5” or 10 mm increments because manufacturing tolerances are ±0.2” (±5 mm) for metal sensors.

3When specifying sensor lengths, take into consideration that the total installed sensor length is equal to the “A” dimension plus the remaining sensor length. 4Metal flanges have a raised face.

A-2 Sapphire User Guide Thermo Fisher Scientific

Page 59: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Ordering Information Remote Mount System

Remote Mount System

Table A–2. Remote mount electronics

Code Remote Electronics

REM Original issue remote electronics

Code Approvals

C Canadian Standards Association (Canada and U.S.)

Code Electronics

J 12–55 Vdc input, 4–20 mA intrinsically safe output

F 15–35 Vdc input, relay output, NO or NC user-selectable auxiliary fault relay

G 115 Vac input, relay output, NO or NC user-selectable auxiliary fault relay

S 230 Vac input, relay output, NO or NC user-selectable auxiliary fault relay

Code Sensor Type

G Gap

S Slot

Code Mounting

O No mounting

P Pipe mounting bracket

W Wall mounting bracket

Table A–3. Remote mount sensor

Code System

RSPS Sapphire remote mount system

Code Remote Sensor

R Original issue remote sensor

Code Associated Sensor Wiring

ST Standard electronics (slot)

IS Intrinsically Safe electronics (gap and slot, default)

Code Cable Length

XX 1 to 5 ft (e.g. 01 = 1 ft)

XX 6 to 10 ft (e.g. 08 = 8 ft)

XX 11 to 15 ft (e.g. 13 = 13 ft)

XX 16 to 20 ft (e.g. 17 = 17 ft)

XX 21 to 25 ft (e.g. 24 = 24 ft)

Thermo Fisher Scientific Sapphire User Guide A-3

Page 60: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Ordering Information Remote Mount System

Code Sensor Type

G Gap (all materials, Intrinsically Safe only)

S Slot (316SS, 316SS polished, Hastelloy C)

Code Sensor Material (same as fitting/flange material)1

S 316SS (gap and slot)

V 316SS polished (gap and slot)

H Hastelloy C (gap and slot)

U MONEL (gap only)

B Hastelloy B (gap only)

Q Carpenter 20 (gap only)

Code Measurement for “A” Dimension

T Inches

M Millimeters

Code Sensor “A” Dimension3, 4

0##0

Sensors available in 0.5” increments only (316SS) 0010–1200 for gap sensors (0010 is 1.0”), minimum 1” (316SS, 316SS polished) 0040–1200 for slot sensors (1200 is 120.0”), minimum 4” (Hastelloy C)

0##0 Sensor available in 5 mm increments only (MONEL) 0025–3050 for gap sensors (0025 is 25 mm), minimum 25 mm (Hastelloy B) 0100–3050 for slot sensors (3048 is 3048 mm), minimum 100 mm (Carpenter 20)

Code Fitting/Flange Type4

N 3/4” NPT only (material same as sensor)

F Flange 150 lb ANSI Raised Face (metal)

G Flange 300 lb ANSI Raised Face (metal)

T Tri-Clover (16 AMP) sanitary fitting and 150 grit polished

Code Fitting/Flange Size

00 3/4” NPT only

10 1” flange or Tri-Clover fitting

15 1.5” flange or Tri-Clover fitting

20 2” flange or Tri-Clover fitting

25 2.5” Tri-Clover fitting only

30 3” flange or Tri-Clover fitting

40 4” flange or Tri-Clover fitting

60 6” flange only

A-4 Sapphire User Guide Thermo Fisher Scientific

Page 61: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Ordering Information Remote Mount System

Thermo Fisher Scientific Sapphire User Guide A-5

Code Fitting/Flange Material1, 4

S 316SS

V 316L SS polished (Tri-Clover 3A and 3/4" NPT only)

H Hastelloy C

U MONEL

B Hastelloy B

Q Carpenter 20

Code Accessories (optional)

T Stainless steel tag

NOTES: See NOTES following Table A–1.

Page 62: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 63: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Appendix B Specifications

Results may vary under different operating conditions.

Relay Operation Table B–1. Performance: Electronics, integral & remote relay operation

Supply voltage Electronics Model F: 15–35 Vdc Model G: 115 Vac, ±10%, 50/60 Hz Model S: 230 Vac, ±10%, 50/60 Hz

Power consumption 4 W maximum (2 W typical)

Outputs Two relay outputs - Process relay is field selectable for Hi/Lo operation - Auxiliary relay is field selectable for fault/mimic operation and NC/NO operation

Process relay SPDT relay with gold-plated contacts. De-energizes on process alarm, self-test, or power loss Maximum contact rating: 5 amps at 115/250 Vac or 5 amps at 30 Vdc Minimum contact rating: 100 mA at 5 Vdc or 15 mA at 24 Vdc

Auxiliary relay SPST (either NC or NO as ordered or selectable by customer) relay with gold-plated contacts. De-energizes on self-test and power loss. De-energizes on process alarm when mimic operation is field selected. Maximum contact rating: 5 amps at 115/250 Vac or 5 amps at 30 Vdc Minimum contact rating: 100 mA at 5 Vdc or 15 mA at 24 Vdc

Self-test feature Continuous. Checks operation of electronics and sensor including crystal bond to sensor body.

Manual loop test feature All output conditions momentarily change to the opposite state (except when in a fault condition) by placing the supplied magnet (P/N XB4353) over housing cover at the location marked “Loop Test.”

Sensor qualification time 0.3 to 12 seconds, field selectable

Repeatability ±0.1 inch

Thermo Fisher Scientific Sapphire User Guide B-1

Page 64: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Specifications Two-Wire Operation

Enclosure Dual compartment NEMA 4 (ANSI 250 type 4X) Enclosure 4 housing. Low-copper aluminum; polyurethane paint; Buna-N O-rings

Connections Two 1/2” NPT conduit connections. Wire range is 22 AWG to 12 AWG.

Operating temperature range

Electronics: -40°F to +158°F (-40°C to +70°C) Metal sensors:

-40°F to +320°F (-40°C to +160°C) -310°F to +212°F (-190°C to +100°C)

Humidity 10–95 % RH, non-condensing

Weight (with 3/4" NPT) Approximately 4 lb (1.8 kg)

Table B–2. CSA (C-US) approvals: Electronics, integral & remote relay operation

Explosion proof Class I, Div. 1, Groups B, C, & D

Dust-ignition proof Class II, Div. 1, Groups E, F, & G Class III, Hazardous Locations

Non-incendive Class I, Div. 2, Groups A, B, C, & D T4 Temperature Code

Indoor/Outdoor use Enclosure 4

3A Sanitary standards 3A Symbol Authorization

Two-Wire Operation

Table B–3. Performance: Integral & remote two-wire operation

Supply voltage Model J: 12–55 Vdc

Power consumption 1.1 W maximum

Current loop loads Rload,max = (Vsupply – 12) x 50, ohms

Analog output (±1 mA) 19 mA / 5 mA self-test fault (field selectable) 8 mA Dry / 16 mA Wet (field selectable) 16 mA Dry / 8 mA Wet (field selectable)

Self-test feature Continuous. Checks operation of electronics and sensor including crystal bond to sensor body.

Manual loop test feature All output conditions momentarily change to the opposite state (except when in a fault condition) by placing the supplied magnet (P/N XB4353) over housing cover at the location marked “Loop Test.”

Sensor qualification time 0.3 to 12 seconds, field selectable

Repeatability ±0.1”

B-2 Sapphire User Guide Thermo Fisher Scientific

Page 65: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Specifications Two-Wire Operation

Thermo Fisher Scientific Sapphire User Guide B-3

Enclosure Dual compartment NEMA 4X (ANSI type 4X) Enclosure 4 housing. Low-copper aluminum; polyurethane paint; Buna-N O-rings

Connections Two 1/2” NPT conduit connections. Wire range is 22 AWG to 12 AWG.

Operating temperature range

Electronics: -40°F to +158°F (-40°C to +70°C) Sensors:

-40°F to +320°F (-40°C to +160°C) -310°F to +212°F (-190°C to +100°C)

Humidity 10–95 % RH, non-condensing

Weight (with 3/4" NPT) Approximately 4 lb (1.8 kg)

Table B–4. CSA (C-US) approvals: Electronics, integral & remote two-wire relay operation

Explosion proof Class I, Div. 1, Groups B, C, & D

Dust-ignition proof Class II, Div. 1, Groups E, F, & G Class III, Hazardous Locations

Non-incendive Class I, Div. 2, Groups A, B, C, & D T4 Temperature Code Remote sensor and field wiring are approved non-incendive field wiring when installed in accordance with Class I, Div. 2 wiring diagram in this manual

Indoor/Outdoor use Enclosure 4

Intrinsically Safe (Entity) Class I, Div. 1, Groups A, B, C, & D T4 Temperature Code (135°C) when installed in accordance with I.S. wiring diagram in this manual Entity parameters: Vmax = 55 V, Imax = 118 mA, Ci = 0.024 μF, Li = 0 mH

Page 66: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 67: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Appendix C Drawings

Note Information presented in this chapter has been regenerated from original drawings. Every effort is made to maintain document accuracy. However, in order to enhance legibility, the documents may have been restructured and some information may have been intentionally excluded. Therefore, the drawings within this manual may not be exact duplicates of the original drawings. ▲

Note Drawings in this manual are included for reference only and may not be the current version. Contact the factory if you need a copy of the latest revision. ▲

Table C–1.

Drawing Rev. Description Page

- - Electronics dimensional drawing (1 sheet) C–2

- - Sensor dimensional drawing (1 sheet) C–3

951-100030 3 Remote mount dimensional drawing (1 sheet) C–4

959-000066 A IS/Exia Loop Diagram (1 sheet) C–5

959-000070 A Class I, Div. 2 Wiring Diagram (1 sheet) C–6

952-000099 A Installation drawing, remote mount system (1 sheet)

C–7

Thermo Fisher Scientific Sapphire User Guide C-1

Page 68: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Drawings

Figure C–1. Electronics dimensional drawing (sheet 1 of 1)

C-2 Sapphire User Guide Thermo Fisher Scientific

Page 69: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Drawings

Figure C–2. Sensor dimensional drawing (sheet 1 of 1)

Thermo Fisher Scientific Sapphire User Guide C-3

Page 70: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Drawings

Figure C–3. Drawing 951-100030 (rev. 3): Remote mount dimensional drawing (sheet 1 of 1)

C-4 Sapphire User Guide Thermo Fisher Scientific

Page 71: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Drawings

Figure C–4. Drawing 959-000066 (rev. A): IS/Exia Loop Diagram (sheet 1 of 1)

Thermo Fisher Scientific Sapphire User Guide C-5

Page 72: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Drawings

Figure C–5. Drawing 959-000070 (rev. A): Class I, Div. 2 Wiring Diagram (sheet 1 of 1)

C-6 Sapphire User Guide Thermo Fisher Scientific

Page 73: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Drawings

Thermo Fisher Scientific Sapphire User Guide C-7

Figure C–6. Drawing 952-000099 (rev. A): Installation drawing, remote mount system (sheet 1 of 1)

Page 74: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

This page intentionally left blank.

Page 75: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Appendix D Toxic & Hazardous Substances Tables*

*English and Chinese versions.

The Toxic and Hazardous Substances tables can be found on the following page.

Thermo Fisher Scientific Sapphire User Guide D-1

Page 76: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Toxic & Hazardous Substances Tables*

D-2 Sapphire User Guide Thermo Fisher Scientific

Page 77: Sapphire™ - Gilson Eng Manuals/tn/sapphire.pdf · Ultrasonic Point Level System ... manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result

Thermo Fisher Scientific81 Wyman StreetP.O. Box 9046Waltham, Massachusetts 02454-9046United States

www.thermofisher.com