rtm assignment 2

12
RTM ASSIGNMENT 2: THE WATER EQUIVALENCE OF LIF TLDS

Upload: ouma-otieno-moses

Post on 14-Apr-2017

135 views

Category:

Health & Medicine


0 download

TRANSCRIPT

RTM ASSIGNMENT 2: THE WATER EQUIVALENCE OF LIF TLDS

i

TABLE OF CONTENTS

TABLE OF CONTENTS ................................................................................................................................................... i

ABSTRACT ................................................................................................................................................................... ii

INTRODUCTION .......................................................................................................................................................... 1

METHODOLOGY ......................................................................................................................................................... 2

Clinical measurements ........................................................................................................................................... 2

Equipment .............................................................................................................................................................. 2

Entrance dose calibration ...................................................................................................................................... 2

Midline dose determination .................................................................................................................................. 3

RESULTS AND DISCUSSION ......................................................................................................................................... 4

CONCLUSION .............................................................................................................................................................. 8

References ................................................................................................................................................................. 9

ii

ABSTRACT

The dosimetry was carried out for radiotherapy patients, and measurements were performed using

LiF and thermoluminescent dosimeters (TLDs). Evaluations were done for water-equivalent

(effective) thicknesses and target dose with transmission data. Considerations were made for the

accuracy of the parameter for the ratio of measured to expected value for each quantity. The

entrance dose was estimated as 1.01 ± 0.07. The mean ratio of effective to contour depth was 1.00

± 0.13, showing a wide distribution reflecting the influence of contour inaccuracies. The mean

ratio of the measured contour dose prescription was 1.00 ± 0.07. The difference in depths that is

patient and effective depth is a reflection of target dose discrepancies. Graphical simulations were

done using Monte-Carlo Simulations and presented.

1

INTRODUCTION

Probability relationship approximates the control of local tumor and normal tissue injury (Leunens,

et al., 1990; Leunens, et al., 1990). Thus procedural therapy for weak point determination is key

to treatment. The most and effective method for this probabilistic determination weak point is in

vivo dosimetry. The commonly used method is thermoluminescent dosimeter (TLD) (Tung, et al.,

2004). This application si popular since it permits the use of small dependencies on dose rate,

temperature, energy and high spatial sensitivity, and resolution (Moscovitch, 1999). TLDs are

popular also since they are unattached to the bias voltage and electrometer though they are passive

and their answers are not prompt.

In vivo dosimetry are in various levels: intracavity dose; entrance dose and exit dose. The coherent

objective of dosimetry is the determination of value of doses. Therefore, the study is first-quality

assurance program for determination of target dose based on water-equivalent (effective) diameter.

This was done for comparison with the actual contour diameter for investigation of the effect of

the corresponding variation of the delivered dose to the target.

Photons on the other hand interacts with atoms in many ways (Corvan, et al., 2014): the multi-

MeV regime, where photoionization occurs, electron-positron pair production and compton

scattering. There are also a number of ways of detection of MeV gamma rays which in most cases

exploits quantum electromagnetic cascade initiated by photon solidification and compton

scattering.

2

METHODOLOGY

Clinical measurements

The study was forced on the Source Axis Distance (SAD) since the study is focused on the midline.

Measurements were taken for dosage at the entrance and exit doses. TLDs of 50 𝑚𝑚 build up was

used to cover material for placement of the surface perpendicular to the beam central axis at each

measuring point. Shadow effect was avoided by either the entrance detector or the exit detector by

shifting the beam slightly out of his axis. Accuracy of the treatment procedure is the ratio of the

measured to the expected values.

Equipment

Since TLD is water equivalent, the analysis of TLDs were done with the Harshaw-3500 manual

reader. With an absorbed dose of 1 Gy at the center of a 10 × 10 cm2 field size and 95 cm source-

skin distance (SSD), the residual signal was noted to be < 1%. Thus batch uniformity = 3.5%

which is equivalent to 1 SD.

Entrance dose calibration

TLDs were positioned on phantom surface for measurement of entrance dose (Den) measurement

and calibration were done for the determination of dose inside the phantom at a depth of the

maximum dose (𝑑𝑚).

Sensitivity loss factor was estimated by:

𝑆𝑓 =

(�̅�𝑟𝑒𝑓

𝐼𝐶 𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒)1

(�̅�𝑟𝑒𝑓

𝐼𝐶 𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒)𝑓

3

Midline dose determination

Determination of Midline dose was based on measurements taken during transmission. From the

theory, 𝑇𝑒𝑥 =𝐷𝑒𝑥

𝐷𝑒𝑛, while 𝑇𝑚𝑖𝑑 =

𝐷𝑚𝑖𝑑

𝐷𝑒𝑛. However, 𝑇𝑒𝑥 and 𝑇𝑚𝑖𝑑 could be calculated from

𝑇𝑒𝑥(𝐴′, 𝑑𝑧−𝑒𝑛) =𝑇𝑀𝑅(𝐴′, 𝑑𝑧−𝑑𝑒𝑛

) (𝑆𝑆𝐷 + 𝑑𝑚

𝑆𝑆𝐷 + 𝑍 − 𝑑𝑚)

2

𝐵𝑆𝐹(𝐴′)

𝐵𝑆𝐹(𝐴0)

𝑇𝑚𝑖𝑑 (𝐴, 𝑑𝑍2

) =

𝑇𝑀𝑅 (𝐴′, 𝑑𝑍2

) (𝑆𝑆𝐷 + 𝑑𝑚

𝑆𝑆𝐷 +𝑍

𝑑𝑚

)

2

𝐵𝑆𝐹(𝐴)

𝐵𝑆𝐹(𝐴0)

where 𝐴0, 𝐴, and 𝐴′ are the field sizes at the entrance, midline ((𝑆𝑆𝐷+

𝑍

2

𝑆𝑆𝐷+𝑑𝑚) 𝐴0) and exit levels

((𝑆𝑆𝐷 + 𝑧 − 𝑑𝑚

𝑆𝑆𝐷 + 𝑑𝑚) 𝐴0)

and 𝑍 is the water-equivalent (effective) depth, 𝑑𝑚 maximum dose depth.

4

RESULTS AND DISCUSSION

Correction factors

Correction factors have to be determined when irradiation geometry differs from the reference

geometry used in the calibration procedure. These variations in TLDs’ response were examined in

different SSDs and field sizes. Correction factor (CF) is defined as

𝐶𝐹 =

�̅�𝑟𝑒𝑓

𝐼𝐶 (𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒)𝑟𝑒𝑓𝑒𝑟𝑒𝑛𝑐𝑒 𝑔𝑒𝑜𝑔𝑚𝑒𝑡𝑟𝑦

�̅�𝑟𝑒𝑓

𝐼𝐶 (𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒)𝑔𝑒𝑜𝑔𝑚𝑒𝑡𝑟𝑦 𝑜𝑓 𝑖𝑛𝑡𝑒𝑟𝑒𝑠𝑡

The attention of the experiment was focused on the LiF, a material of the lowest value of Z. Based

on the database, the Compton scattering was noted as the overriding interaction between

mechanisms for a multi-MeV gamma-ray beam up to energies of 20 MeV as shown in the figure

below.

Figure 1: Energy loss for a multi-MeV gamma ray beam as it interacts with LiF.

5

The lines in the square corresponds to nuclear filed while the blue is for diamond for Compton

scattering. Compton scattering is dominant up to around 1.5 MeV, after which pair production

dominates.

To estimate the effect of LiF on the spectrum of the electrons, simulation was done using the

Monte-Carlo code FLUKA and the simulation present in the figure below.

Figure 2: Spectra of the electrons generated.

However, the Gaussian gave appropriate estimation of the relationship between the peak values of

the conversion factor and is show in the Figure 3 below. Furthermore, the peak of the conversion

factors in the electron spectra plotted as a function of the initial photon energy. The peaks rise in

magnitude up to 5 MeV where the peaks remain endless up to 10 MeV after which they start to

decrease again. This shown in Figure 4.

6

Figure 3: The peak of the conversion factors in the electron spectra plotted as a function of the

initial photon energy

7

Figure 4: Fluka simulations.

8

CONCLUSION

Lif and TLD was used to carry a compressive series evaluation of in vivo dosimetry measurements.

Determination of water-equivalent depth and midline dose showed a close association between the

midline dose imprecision and discrepancy of water-equivalent and contour depths.

Furthermore, the modification of compact gamma-ray spectrometer for in vivo dosimetry for

treatment presented showed that for energy windows around 20 MeV, using Monte-Carlo

simulations, the performance of the system can be checked. The results of this experiments were

in line with other conducted by Aad, et al., (2014).

9

References

Aad, G. a. A. T. a. A. B. a. A. J. a. K. S. A. a. A. A. et al., 2014. Measurement of the inclusive

isolated prompt photons cross section in p p collisions at s= 7 TeV with the ATLAS

detector using 4.6 fb- 1. hysical Review D, 89(5), p. 052004.

Corvan, D. J., Sarri, G. & Zepf, M., 2014. Design of a compact spectrometer for high-flux MeV

gamma-ray beams. Review of Scientific Instruments, 85(6), p. 065119.

Leunens, G., Van Dam, J., Dutreix, A. & Van der Schueren, E., 1990. Quality assurance in

radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedure.

Radiotherapy and Oncology, 17(2), pp. 141-151.

Leunens, G., Van Dam, J., Dutreix, A. & Van der Schueren, E., 1990. Quality assurance in

radiotherapy by in vivo dosimetry. 2. Determination of the target absorbed dose.

Radiotherapy and Oncology, 19(1), pp. 73-87.

Moscovitch, M., 1999. Personnel dosimetry using LiF: Mg, Cu, P. Radiation protection

dosimetry, 85(1-4), pp. 49-56.

Tung, C. et al., 2004. In vivo dosimetry for external photon treatments of head and neck cancers

by diodes and TLDS. Radiation protection dosimetry, 111(1), pp. 45-50.