reproducibility of corneal topographic data derived from anterior segment spectral-domain oct

11
Rosachiara Leaci, Alberto Neri, Patrizia Scaroni, Claudio Macaluso Ophthalmology, University of Parma, Parma, Italy The authors have no financial interest in the subject matter of this poster

Upload: monifa

Post on 06-Jan-2016

31 views

Category:

Documents


3 download

DESCRIPTION

Rosachiara Leaci, Alberto Neri, Patrizia Scaroni, Claudio Macaluso Ophthalmology, University of Parma, Parma, Italy The authors have no financial interest in the subject matter of this poster. Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Rosachiara Leaci, Alberto Neri, Patrizia Scaroni, Claudio Macaluso

Ophthalmology, University of Parma, Parma, Italy

The authors have no financial interest in the subject matter of this poster

Page 2: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Purpose:

To evaluate the reproducibility of corneal curvature measurements entirely derived from OCT scan data, and to compare them to measurements obtained by means of a conventional Placido disc topographer.

Page 3: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Methods: 101 KERATOCONUS PATIENTS and 40 NORMAL

SUBJECTS were examined in their right eye with two instruments:

1) a spectral domain OCT, dedicated to anterior segment examination (Casia SS-1000, Tomey Co., Japan)

1) a conventional Placido-disc topographer (Keratron Scout, Optikon, Italy)

Examinations were carried out in the same session twice with each device, in order to assess reproducibility.

Statistical analysis was performed with the SPSS 18.0 software.

Page 4: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Casia SS-1000 (Tomey Co., Japan) is an OCT dedicated to the anterior segment that is equipped with a fully automated topography routine, deriving curvature data by fitting OCT data scans along 16 meridians

Immagine riconoscimento profili topo casia

Page 5: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

The software automatically identifies outer and inner corneal boundaries on each of the 16 B-scan. The operator is asked to check that the position of the boundaries is correct before to visualize topographic data.

Page 6: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

The OCT shows maps of anterior, posterior corneal curvature and pachimetry.

Page 7: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Results: validity of OCT keratometry

Pathologic eyes

Normal eyes

ICC = 0.985

ICC = 0.959

A

C

B

D

A, C: scatter plots of the average corneal curvature values obtained by Casia AS-OCT (x-axis) and Keratron topographer (y-axis). B, D: Bland-Altman plots of Casia-Keratron comparison, representing the differences of corresponding measurements (y-axis) in relation to their means (x-axis). Corneal curvature is expressed in Dioptries.

Curvature measurements obtained with the two instruments showed excellent agreement both in the normal and in the patients group, with Intra-Class Correlation Coefficients (ICC) of 0.959 and 0.985, respectively.

Page 8: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Results: reproducibilityIn the normal group reproducibility (ICC) of curvature measurements was 0.988 for the OCT and 0.976 for the conventional topographer, while in the patients group ICC for OCT and topographer was 0.997 and 0.989, respectively. Scatter plots of test (x-axis) vs. re-test (y-axis) average corneal curvature values are shown below.

CASIA AS-OCT KERATRON TOPOGRAPHER

Pathologic eyes

Normaleyes

ICC=0,997 ICC=0,989

ICC=0,988 ICC=0,976

Page 9: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Further analysis of reproducibility with Bland-Altman plots revealed a higher variability with the conventional topographer as compared to the OCT in the patients group (p<0.01), but not in the normal group (p=0.085). ICC and Coefficient of Repeatability (CR) are reported in the figures below. [CR = 1,96 x SD of the differences between repeated measures].

Results: reproducibility

CASIA AS-OCT KERATRON TOPOGRAPHER

Normaleyes

ICC=0,997 ICC=0,989

ICC=0,988 ICC=0,976

Pathologic eyes

CR=0,62D CR=1,07D

CR=0,35D CR=0,46D

Page 10: Reproducibility of corneal topographic data derived from anterior segment spectral-domain OCT

Conclusions:Conclusions:

Corneal curvature measurements entirely Corneal curvature measurements entirely derived from OCT scan data showed derived from OCT scan data showed excellent reproducibility and were found excellent reproducibility and were found comparable to measurements obtained comparable to measurements obtained by means of a conventional Placido disc by means of a conventional Placido disc topographer.topographer.