RAPID ONLINE ANALYSIS OF ACETYLENE FOR HYDROGENATION REACTOR CONTROL OPTIMIZATION Linh D. Le, J.D. Tate, and Mary Beth Seasholtz The Dow Chemical Company 2301 N. Brazosport Blvd. B-1463 Freeport, TX 77541 Manish Gupta and Doug Baer Los Gatos Research 67 East Evelyn Ave., Suit 3 Mountain View, CA 94041 Trevor Knittel, Alan Cowie, and Jie Zhu Analytical Specialties Inc. 910 Gemini Houston, TX 77058 KEYWORDS Off-Axis Integrate Cavity Output Spectroscopy, Process Optimization, Hydrogenation Reactors, Rapid On-line Analyzer, Acetylene ABSTRACT Rapid analysis of acetylene (C 2 H 2 ) in crack gas has been a significant technology gap in the olefin's industry for many years. Currently, the accepted best practice in the industry is on-line Gas Chromatography (GC); however, it suffers from a slow response time. A much faster, near real-time, analysis would allow tighter acetylene hydrogenation process control (patent pending). Since 1999, The Dow Chemical Company has been aggressively pursuing solutions to this unmet need. To catalyze a commercial solution, Dow formed a strategic partnership with several external partners (Analytical Specialties, Dept. of Energy and Los Gatos Research). These collaborations have led to the development and commercialization of a near real-time analyzer capable of measuring C 2 H 2 in cracked gas at the part-per-billion level. This technology is based on Off-Axis Integrated Cavity Output Spectroscopy (ICOS). The presentation will outline the ICOS technology, application feasibility study, and analyzer performance and field validation undertaken by Dow.

Upload: vokien

Post on 30-Dec-2016

224 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: rapid online analysis of acetylene for hydrogenation reactor control

RAPID ONLINE ANALYSIS OF ACETYLENE FOR HYDROGENATION REACTOR CONTROL

OPTIMIZATION

Linh D. Le, J.D. Tate, and Mary Beth Seasholtz The Dow Chemical Company

2301 N. Brazosport Blvd. B-1463 Freeport, TX 77541

Manish Gupta and Doug Baer Los Gatos Research

67 East Evelyn Ave., Suit 3 Mountain View, CA 94041

Trevor Knittel, Alan Cowie, and Jie Zhu Analytical Specialties Inc.

910 Gemini Houston, TX 77058

KEYWORDS

Off-Axis Integrate Cavity Output Spectroscopy, Process Optimization, Hydrogenation Reactors, Rapid On-line Analyzer, Acetylene

ABSTRACT Rapid analysis of acetylene (C2H2) in crack gas has been a significant technology gap in the olefin's industry for many years. Currently, the accepted best practice in the industry is on-line Gas Chromatography (GC); however, it suffers from a slow response time. A much faster, near real-time, analysis would allow tighter acetylene hydrogenation process control (patent pending). Since 1999, The Dow Chemical Company has been aggressively pursuing solutions to this unmet need. To catalyze a commercial solution, Dow formed a strategic partnership with several external partners (Analytical Specialties, Dept. of Energy and Los Gatos Research). These collaborations have led to the development and commercialization of a near real-time analyzer capable of measuring C2H2 in cracked gas at the part-per-billion level. This technology is based on Off-Axis Integrated Cavity Output Spectroscopy (ICOS). The presentation will outline the ICOS technology, application feasibility study, and analyzer performance and field validation undertaken by Dow.

Page 2: rapid online analysis of acetylene for hydrogenation reactor control

INTRODUCTION

In 2002, Dow’s Process Analytical group commenced a collaborative effort with an industry supplier, Analytical Specialties, Inc. (ASI), and a small research company, Los Gatos Research (LGR) to develop and commercialize an Off-Axis Integrated Cavity Output Spectroscopy (ICOS) analyzer for rapid on-line acetylene (C2H2) measurements. The specific technical objectives for the Off-Axis ICOS platform are pathlength enhancement for sensitivity improvement and to provide near or real-time data for improving advanced process control (APC) and product quality control (PQC).

The speed, reliability and accuracy of C2H2 analysis for hydrogenation reactors are critical to process control optimization, safety and environmental drivers, and product quality specifications. However, the fastest process Gas Chromatography (GC) methods-the industry accepted best practice at present-- have a 90-second to 5 minutes response-time, much less than optimum which are 15 seconds or less.

This laser-based technology development effort has been funded through the Department of Energy [DOE] under a cost-share grant and their Small Business Innovation Research (SBIR) program grant. The work proceeded from laboratory testing and “proof-of-concept” to prototype construction and field validation, and commercial implementation. Developmental concepts and aspects such as the application economic impact, technology fundamentals and evaluation, application spectroscopy and chemometric feasibility, and field on-line validation and implementation are addressed and reported.

TECHNOLOGY DESCRIPTION

Off-Axis ICOS is an optical absorption spectroscopic technique which is based on the principle that when a molecule interacts with a light beam at a specific wavelength it will cause the molecule to vibrate/rotate, and thereby absorb the energy (the light source). The wavelengths at which these absorptions occur are unique and are a function of the analyte and its molecular bond types. Utilizing the measured absorption light intensity, the concentration of the probed species can be determined from Beer’s Law.

The basic operating concept of the Off-Axis ICOS is based on injecting the laser light into the optical cavity in an off-axis configuration with respect to the on-axis cavity alignment formed by the two highly reflective dielectric mirrors 1, 2, as illustrated in Figure 1.

Unlike traditional absorption measurements, the optical pathlength in the Off-Axis ICOS method is effectively infinite as the entering laser light bounces back and forth retracing the same path on each cycle. The effective pathlength of the cavity is given by the mirror reflectivity. Therefore, using highly reflective astigmatic mirrors with R > 99.99% the laser beam can travel 10,000 passes, lengthening a 20cm long cavity to an optical pathlength of 2,200 meters. This allows Off-Axis ICOS to have very high sensitivity and low detection limit 3.

Page 3: rapid online analysis of acetylene for hydrogenation reactor control

FIGURE 1- SCHEMATIC OF OFF-AXIS ICOS SYSTEM

In addition, by implementing LGR patented off-axis alignment configuration Off-Axis ICOS is intrinsically robust. The cavity effective pathlength does not depend on the exact laser beam alignment, only on the losses in the cavity, and thus it’s not necessary to meticulously align the input beam or the optical cavity allowing for a high degree of mechanical vibration tolerance.

The absorption intensity signal is acquired as the injection laser light wavelength is scanned over the entire spectral feature of absorption peak. Quantitative measurement of the probed species; concentration, can be derived from the transmitted intensity signal, the analyte absorption line-strength, sample temperature and total pressure, and the cavity’s effective optical pathlength. The detector, electronics and laser software control module process the detected amplitude signal and analyze the acquired spectra to derive the targeted gas concentration in real-time.

PROOF-OF-CONCEPT TECHNOLOGY EVALUATION Off-Axis ICOS is employed to perform quantitative trace gas detection by means of a high-finesse (low loss) optical cavity. The cavity is formed by two high reflectivity (HR) (typical reflectivity is > 99.9%), dielectric-coated fused silica mirrors. The technology detection sensitivity would be impaired if the mirrors are fouled (reduction of mirror reflectivity) when exposed to process sample streams. Thus, it was pertinent to determine if this condition could potentially be an unacceptable factor of reliable performance.

Figure 2 shows the testing apparatus to determine potential mirror fouling of the Off-Axis ICOS sample cavity under process conditions. The mirrors were installed on a typical extractive GC sample flow system for a 1-week, 1-month, and 6-month long term studies to verify if there was any affect on the window’s reflectivity and/or its coating’s integrity from the exposure to hydrocarbon process gas line contents. To determine the mirror’s reflectivity before and after exposure to the process, LGR calculated the mirror loss from the measured the cavity’s ring-down time parameter.

LensT P

Detector

ElectronicsLaser Controller

Fiber-CoupledDiode Laser

Page 4: rapid online analysis of acetylene for hydrogenation reactor control

FIGURE 2- INSTALLED TEST CAVITY IN HYDROCARBON PROCESS

APPLICATION FEASIBILITY STUDY Besides the target acetylene analyte, there are several other gases present in the ethylene cracked process steam. Figure 3 is a plot of several spectra corresponding to the major constituents present in the stream.

FIGURE 3-SPECTRAL GLOBAL VIEW OF MAJOR CONSTITUENTS IN ETHYLENE CRACKED PROCESS STEAM

Page 5: rapid online analysis of acetylene for hydrogenation reactor control

Conventional laser sensors are incapable of performing quantitative analysis of process gas samples with spectroscopic interference backgrounds. Utilizing chemometrics as an analytical prediction tool enables the same laser-based technology to analyze and monitor analyte concentrations in the presence of spectral interferences. Chemometrics model spectral data illustrated in Figure 4 allows the Off-Axis ICOS analyzer to accomplish the following, which conventional spectrometry cannot:

Ability to monitor near real-time component(s) in process streams with complex compositions by incorporating the model algorithm into the laser spectrometer.

Ability to predict multi-components.

Ability to provide concentration estimates in the presence of spectral interferences and non-linear effects including pressure and temperature effects.

FIGURE 4-CHEMOMETRIC MODEL PARENT SPECTRA

PROCESS OFF-AXIS ICOS PROTYPE The prototype analyzer, shown in Figure 5, consisted of an electronic control unit, sample optical cavity unit, and sample handling system (SHS) equipped with a vacuum-pump. The data acquisition and operation of the analyzer was achieved through the use of the software module as shown in Figure 6.

Model Parent Spectra

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

0.45

0.50

1 101 201 301 401 501 601 701 801 901 1001Data Point

AU

Page 6: rapid online analysis of acetylene for hydrogenation reactor control

FIGURE 5-OFF-AXIS ICOS PROTOTYPE

FIGURE 6- ICOS PROTOTYPE SOFTWARE CONTROL MODULE

6.446

2572.8

29.0

100.1

802.20.6692

6.446

2572.8

29.0

100.1

802.20.6692

Page 7: rapid online analysis of acetylene for hydrogenation reactor control

RESULTS CHEMOMETRIC MODEL VALIDATION Seven gas mixtures with various concentration ranges for acetylene, methylacetylene, ethylene, methane, and ethane were obtained to perform model validation based on process operating conditions.

Figure 7 illustrates data of the model performance validation where Mixture 1 was in use. The upper graph shows acetylene and methylacetylene precision with auto calibration time-stamp for each day, and the lower graph shows each component prediction residual results. The first graph segment of 1 to 200 data point represents the model uncertainty in the comparison between measured Mixture 1 spectra and the model prediction spectra for acetylene component, and the segment data point of 201 to 561 is for methylacetylene component.

MIX 1 Model Validation (Sept 5-6, 13-14, 04)

0.300.350.400.450.500.550.600.650.700.750.800.850.900.951.001.051.101.151.201.251.30

16:50

:00

19:17

:25

21:44

:51

0:12:2

3

2:39:5

8

5:07:3

1

7:35:0

2

10:02

:36

12:30

:10

18:57

:01

21:24

:25

23:51

:51

2:19:0

9

4:46:3

1

7:14:0

0

9:41:2

4

12:08

:49

14:36

:14

Time

ppm

900

920

940

960

980

1000

1020

1040

ppm

C2H2 MA

Stdev = 0.0159, %Error = 2.56

Stdev = 3.91, %Error = 0.97 Stdev = 2.95, %Error = 0.39

Stdev = 0.0154, %Error = 0.0

MIX 1 Prediction Residual

-0.010-0.0050.0000.0050.010

1 41 81 121 161 201 241 281 321 361 401 441 481 521 561Data Point

AU

FIGURE 7- MODEL PERFORMANCE PRECISION DATA AND PREDICTION RESIDUAL FOR MIXTURE 1 SAMPLE

Page 8: rapid online analysis of acetylene for hydrogenation reactor control

Summarized in Table I is the performance specification (precision and accuracy) of the ICOS. For comparison with the GC, the model validation data were grouped by concentration range. The GC specifications for acetylene measurement were from the Dow optimum GC method. As shown in Table I, the accuracy and precision of the ICOS prototype (with 6 second analyzer response time) for C2H2 application are as good or improved over GC specifications. Due to GC slow response-time for the existing methylacetylene analysis, this parameter has not been previously used to aid in the acetylene hydrogenation process optimization control. Thus, the methylacetylene measurement required specifications are not available for comparison.

TABLE I- MODEL VALIDATION PERFORMANCE SPECIFICATION COMPARISON WITH GC

ICOS * Expected GC Spec. ICOS * Expected GC

Spec.

0.50 0.49 0.03 0.06 0.01 0.080.96 0.88 0.07 0.06 0.08 0.080.99 1.02 0.04 0.06 0.03 0.081.48 1.51 0.07 0.14 0.03 0.202.86 2.58 0.06 0.14 0.28 0.208.00 7.74 0.14 0.14 0.26 0.20

Accuracy/Bias (± ppm)

Acetylene (Sep 5-20, 04)Precision @ 95% RSD

(± ppm)BOC Reported

(ppm)

ICOS Ave. (ppm)

1017 1020.9 5.8 0.41021 1021.1 13.9 0.03022 3082.3 63.0 2.05349 5144.0 130.7 3.8

Methylacetylene (Sep 5-20, 04)BOC Reported

(ppm)ICOS Ave.

(ppm)Precision @ 95% RSD

(± ppm)Accuracy

(%)

ON-LINE FIELD TRIAL PERFORMANCE

The field test installation compared the ICOS with the GC in parallel, as illustrate in Figure 8, at Dow, Oyster Creek Hydrocarbon plant.

Page 9: rapid online analysis of acetylene for hydrogenation reactor control

FIGURE 8- ICOS AND GG ARE IN PARALLEL FIELD COMPARISON Plant operation protocol generally prohibits making changes to process conditions until a sufficient number of data points (minimum 2 to 3) indicate a consistent or trending alarm condition. Note that the analyzer input to the plant’s process control system is once per minute. Every GC measurement (5 minute cycle) in Figure 9 repeats the same data point 5 times.

FIGURE 9- REACTOR ACETYLENE SPIKE

Because of the batch process nature of a gas chromatograph (GC), an excursion may be well underway before the GC can detect it. The continuous measurement made by the ICOS analyzer provides near real-time results and updates measurements quickly enough to make a reliable determination.

A

B

Cracked Gas

C

ICOS GC

Analyzer House

[<1ppm C2H2]

C2H2 On-Line Process Sample

0.00

0.50

1.00

1.50

2.00

2.50

3.00

21:15 21:25 21:35 21:45 21:55 22:05 22:15 22:25 22:35 22:45Time

C2H

2 (pp

m)

GC ICOS

Action Action TakenTaken

11stst Data Pt. Data Pt.

22ndnd Pt Pt No Action No Action

Page 10: rapid online analysis of acetylene for hydrogenation reactor control

Figure 10 illustrates a process methylacetylene spike simultaneously measured by the ICOS analyzer and a GC (data rate of 1 data point per ~20min). In this example, the GC reported a measurement exceeding alarm condition almost ten minutes after the excursion took place and with only two measurements exceeding alarm value. The measurement was not considered reliable. The ICOS analyzer detected the excursion as it was occurring and made approximately 200 measurements during the twenty minute period of the excursion. Therefore, the measurement quality of the ICOS gives plant operations confidence that the data reflects actual process conditions to take action to control and optimize reactors.

The continuous measurement and speed of the ICOS analyzer also allows more accurate process measurements. In the example shown in Figure 10, the ICOS analyzer measured and reported changes in process conditions much more quickly and accurately than the GC. While this process spike does not represent a plant upset or excursion, it did indicate that process conditions fluctuated. No matter how accurate the GC method was, it did not accurately reflect actual process conditions, primarily because of its intrinsic slow response time.

MA On-Line Process Sample

200

250

300

350

400

450

500

550

600

12:00 12:05 12:10 12:15 12:20 12:25 12:30 12:35 12:40 12:45 12:50 12:55

Time

MA

(ppm

)

GC17 ICOS

FIGURE 10- REACTOR METHYLACETYLENE SPIKE

In addition to speed and accuracy, the ICOS analyzer’s signal to noise ratio enables it to be sensitive enough to reliably report small changes in measured values. The sensitivity of the GC measurement is such that small changes in the process are difficult or impossible to differentiate a process spike from baseline noise. Some process spikes, like those in Figure 11, are not large enough to warrant process control actions, but can be useful for correlating other process events for troubleshooting.

Page 11: rapid online analysis of acetylene for hydrogenation reactor control

0.0

0.5

1.0

1.5

2.0

10:00 10:15 10:30 10:45 11:00 11:15 11:30 11:45 12:00Time

C2H

2 (p

pm)

GC18 ICOS

FIGURE 11- MEASUREMENT SENSITIVITY

While the ICOS analyzer prediction model was optimized for a measurement range of 0-10 ppm, the measurement range extends to 1,000 ppm. During the plant restart, the analyzer recorded measured values far above the reporting range of the GC as shown in Figure 12, thus providing additional data about the plant during the startup process.

The ICOS analyzer recorded measurement results, diagnostic and event data with every measurement. Additionally, the analyzer was set to record spectra at a rate of about one per hour during normal run conditions and at a higher rate during warning or fault alarm conditions.

On-Line Process Sample

0

200

400

600

800

1000

22:00 22:15 22:30 22:45 23:00 23:15Time

C2H

2 (pp

m)

ICOS GC18

Process Event?

Page 12: rapid online analysis of acetylene for hydrogenation reactor control

FIGURE 12- EXTENDED RANGE MEASUREMENT

During the plant startup, at approximately 11:04 pm, a downward spike occurred in the acetylene and methylacetylene measurements as magnified in Figure 13. It is not uncommon to doubt the validity of a GC measurement during the pressures of a plant startup because of the reliability of GC’s under these conditions. However, because an alarm condition existed, the ICOS analyzer recorded additional spectral data that was used to confirm the reliability of the measurement.

Spectral data was extracted from files stored by the analyzer, tabulated and charted at different intervals representing different measurement values. An example of a spectra data table and chart are shown in Figure 13.

0

200

400

600

800

1000

1200

21:50 21:54 21:58 22:02 22:06 22:10 22:14 22:18 22:22 22:26 22:30Time

C2H

2 (pp

m)

0

500

1000

1500

2000

2500

3000

MA

(ppm

)

C2H2 (ppm) MA (ppm)

FIGURE13- EXTENDED RANGE MEASUREMENT (EXPANDED)

Page 13: rapid online analysis of acetylene for hydrogenation reactor control

9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/2004 9/6/200413:46:08 14:05:55 14:25:41 14:45:26 15:05:13 15:24:58 15:44:44 16:04:31 16:24:18 16:44:04 17:03:51

0.93 ppm 0.93 ppm 0.95 ppm 0.94 ppm 0.94 ppm 0.95 ppm 0.96 ppm 0.96 ppm 0.95 ppm 0.94 ppm 0.93 ppm 3072.4 ppm 3070.3 ppm 3055.9 ppm 3060.6 ppm 3057.5 ppm 3062.1 ppm 3056.7 ppm 3052.9 ppm 3063.6 ppm 3067.4 ppm 3077.3 ppm84.3 F 84.2 F 84.3 F 84.3 F 84.4 F 84.4 F 84.4 F 84.2 F 84.3 F 84.3 F 84.3 F 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi 1.93 psi

0.701186907 0.701377573 0.701256854 0.701113665 0.701508007 0.701294025 0.701277129 0.700967603 0.701373772 0.701310498 0.7012220.48255602 0.482709601 0.482889792 0.48279391 0.483022168 0.482732663 0.482691354 0.482648946 0.482695155 0.482655451 0.482687

0.486125289 0.486168457 0.486201826 0.48621399 0.48641623 0.486072913 0.48592364 0.485908434 0.48602324 0.486106535 0.4862530.489994115 0.490178614 0.49039361 0.490204887 0.490549472 0.490197622 0.490243155 0.49026909 0.49028928 0.490132405 0.490110.494389998 0.494728838 0.494446514 0.494515278 0.494709324 0.494315742 0.494255678 0.494150757 0.494324359 0.494427422 0.4945710.477513123 0.477652933 0.477575045 0.477597938 0.477624042 0.477393671 0.477350081 0.477228264 0.477256057 0.47740837 0.4774810.482463686 0.482539885 0.482788925 0.482609495 0.48298956 0.482708925 0.482655958 0.482698619 0.48275429 0.482580012 0.4825980.496715754 0.496887497 0.49710866 0.496863506 0.497062028 0.496788405 0.496737127 0.496803442 0.496757317 0.496743378 0.4967120.497031025 0.49691622 0.497038966 0.496985238 0.497231406 0.497031448 0.49702452 0.496888089 0.496901183 0.496939789 0.4971760.49089186 0.491143097 0.490879949 0.490987911 0.491076528 0.490561805 0.49046567 0.490610042 0.490661151 0.490842187 0.4908930.48307125 0.483196615 0.483161557 0.483217228 0.483468211 0.483060944 0.482862168 0.482809116 0.483042443 0.483204134 0.483322

0.481711497 0.481557578 0.481780261 0.481696797 0.481883493 0.481707611 0.481597029 0.481587821 0.481731687 0.481900558 0.4819480.485014829 0.485014745 0.48520634 0.485153119 0.485392191 0.485146614 0.485212169 0.485065685 0.485177449 0.485033837 0.4849070.494737033 0.494783918 0.495176148 0.495067594 0.495286898 0.494903792 0.494935471 0.495212642 0.494855639 0.494692006 0.4944270.504260291 0.504428993 0.504944054 0.504619068 0.5051234 0.504777801 0.50490097 0.504830347 0.504779153 0.504516343 0.5044570.511258848 0.511338764 0.511578427 0.511464974 0.511692472 0.511477138 0.511435322 0.511521996 0.511372808 0.511297961 0.5113110.515077495 0.515128604 0.515382797 0.515210463 0.515430105 0.515542967 0.515419798 0.515491098 0.515325183 0.515275933 0.5152460.518393921 0.518389106 0.518432105 0.518537871 0.518603426 0.518205451 0.518109485 0.518379138 0.518019347 0.518524608 0.5181260.520032451 0.520313593 0.520294079 0.520315536 0.520408884 0.519966812 0.519916125 0.520084489 0.519982609 0.519989283 0.5199660.522018353 0.521811805 0.521978142 0.521821858 0.522251934 0.52172378 0.521423715 0.521883105 0.521622829 0.52184982 0.5217230.522634872 0.5225705 0.522740638 0.522713352 0.522967038 0.522457046 0.522220593 0.522433139 0.522370626 0.522382115 0.5225660.523318804 0.523446449 0.523394073 0.523441634 0.523793568 0.523553821 0.523390525 0.523097556 0.523393397 0.52311158 0.52370.523355467 0.523346935 0.523685352 0.523501191 0.523882186 0.523587358 0.52365494 0.523435129 0.523629935 0.52365494 0.5238010.524080539 0.524002397 0.524188333 0.524046072 0.52442656 0.524002482 0.523979504 0.523842988 0.524058068 0.524197372 0.524218

-0.05

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

1 101 201 301 401 501 601 701 801 901 1001

Mix 2 11ppm CH2 Standard

FIGURE 14- EXAMPLE OF SPECTRA DATA TABLE AND CHART

The acetylene spectra shown in Figure 15 was extracted at intervals representing 1) the high concentration before the downward spike, 2) the concentration at the lowest point of the downward spike and 3) normal concentration as measured before the plant upset occurred.

0

5

10

15

20

25

30

35

40

0 100 200 300 400 500 600 700 800 900 1000Data Points

AU

22:02:22 22:04:19 16:24:41

FIGURE 15- ACETYLENE SPECTRA AT VARIED CONCENTRATIONS

Page 14: rapid online analysis of acetylene for hydrogenation reactor control

The peak height and area indicate that the measurement depicted in Figure 15 accurately represents process conditions and is not an instrument issue. Because the spectra can be used to validate the analyzer measurement, spectra data records become a powerful troubleshooting and diagnostic tool.

CONCLUSION

Under a variety of process conditions, the ICOS analyzer proved to be much more accurate, precise and durable than current gas chromatography methods. The speed and reliability of analysis provided insight into process run conditions that were previously unavailable and unexpected. Powerful diagnostic tools designed into the ICOS capabilities allowed for validation of reported results.

The continuous measurement method and speed of analysis of the ICOS analyzer proved to be more valuable than gas chromatography because, no matter how accurate the GC method was, it did not correctly reflect real-time process conditions, primarily because of the GC’s intrinsically slow response time. The speed, reliability and accuracy of acetylene (C2H2) analysis for hydrogenation reactors are critical to process control optimization, safety and environmental drivers, and product quality specifications.

ACKNOWLEDGMENTS

The authors would like to thank Craig Snook, Phillip Proctor, Jeff Owens, Mark Tomasek, Andy Weber, Ronny Van De Lavoir, Jeff Weigle, Kevin Welch, Paul O’Drisscoll, and Courtland Sears for their support and contributions to the success of this development effort.

REFERENCES 1. 1 A. O’Keefe, “Integrated Cavity Output Analysis of Ultra-Weak Absorption,” Chem. Phys.

Lett., 293, 331 (1998). 2. J.B. Paul, J.J. Scherer, A. O'Keefe, L. Lapson, J.G. Anderson, C. Gmachl, F. Capasso, and

A.Y. Cho, “Infrared Cavity Ring-down and Integrated Cavity Output Spectroscopy for Trace Species Monitoring,” SPIE, 4577, 1-11, (2001).

3. D.S. Baer, J.B. Paul, M. Gupta, and A. O’Keef, “Sensitive Absorption Measurements in The

Near-Infrared Region Using Off-Axis Integrated Cavity Output Spectroscopy,” Appl. Phys. B, Special Issue, (2002).