Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided

Lumbar Puncture

Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV

Department of Radiology and Radiological Science

Medical University of South Carolina, Charleston, SC

American Society of Neuroradiology MeetingChicago, April 2015

EP # 143

Page 2: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Myself, other authors, and my immediate family members do not have a financial relationship with a commercial organization that may have a direct or indirect interest in the content.

Disclosure of Commercial Interest

Page 3: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Background and Purpose• Fluoroscopic-guided lumbar puncture is an effective

alternative to bedside lumbar puncture in challenging patients.

• No published guidelines are available for an acceptable range of fluoroscopic time for this procedure.

• The purpose of this study is to retrospectively evaluate radiation dose and fluoroscopic time in fluoroscopic-guided lumbar puncture (FGLP).

Page 4: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Methods: Patient Population

• We retrospectively reviewed all FGLP performed at our institution during nine consecutive months (October 2, 2013 to June 30, 2014).

• The radiology information system at our hospital was searched by using the procedure code for FGLP to identify subjects.

• We excluded all myelograms and cases where fluoroscopic time and radiation doses were not recorded in the EMR.

Page 5: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

• All FGLP procedures were performed in the prone or left lateral decubitus position using a sterile technique.

• FGLPs were performed at L2-L3, L3-L4, L4-L5 or L5-S1 levels.

• 22-gauge spinal needles were most commonly used.

• Three neuroradiology fellows and a physician assistant performed the FGLP during the study period. Faculty directly or indirectly supervised trainees.

Methods: Procedure Technique

Page 6: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Methods: Data Collection• The procedure reports and stored images were reviewed

by using the Picture Archiving and Communication System (PACS).

• Fluoroscopic time, radiation dose, operator, clinical indication, puncture site, patient age and gender were collected from the radiology report.

Page 7: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

We performed group comparisons of patients stratified by:•Age•Gender•Puncture site•Operator•Study indication

Statistical analyses were conducted using Mann-Whitney tests. Differences were considered statistically significant when p < 0.05.

Methods: Data Analysis

Page 8: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

• 188 FGLPs were included in the study (median age [SD] = 47 [19], age range = 10 - 86; 68 males, 120 females).

• 18 FGLPs were excluded because fluoroscopic time and radiation dose were not available on PACS.

Results

Page 9: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Clinical indications for the FGLPs were as follows: •Altered mental status (17 cases)•Abnormal neurologic exam (13 cases)•Leptomeningeal tumor (33 cases)•Infection (30 cases)•Idiopathic intracranial hypertension (61 cases)•Demyelinating / inflammatory disease (22 cases)•Normal pressure hydrocephalus (6 cases)•Seizure disorder (2 cases)•Miscellaneous other indications (4 cases)

Results

Page 10: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Fluoroscopy time but not radiation dose were significantly different among operators (respectively p = 0.02 and p = 0.062). There was no significant difference in fluoroscopy time and radiation exposure between the three fellows and the physician assistant.

Results

Flu

oro

sco

pic

Tim

e (s

eco

nd

s)

Rad

iati

on

Do

se (

mG

y)

Page 11: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Fluoroscopy time but not radiation dose estimates were significantly lower in patients younger than 50 years (respectively p = 0.004 and p = 0.467).

ResultsR

adia

tio

n E

xpo

su

re (

mG

y)

Flu

oro

sco

pic

Tim

e (s

eco

nd

s)

Rad

iati

on

Do

se (

mG

y)

Page 12: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Top 3 indications for FGLP were tumor (N = 33), infection (N = 30), idiopathic intracranial hypotension (N = 61).Radiation dose but not fluoroscopic time were significantly different across indication categories (p = 0.001 and p = 0.425), with the greatest radiation dose in patients referred for idiopathic intracranial hypertension (IIH).

ResultsR

adia

tio

n E

xpo

su

re (

mG

y)

Flu

oro

sco

pic

Tim

e (s

eco

nd

s)

Rad

iati

on

Do

se (

mG

y)

Page 13: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

There were no significant differences in fluoroscopic time or radiation dose between puncture sites. There were no significant differences in fluoroscopic time or radiation dose between males and females.

Results

Flu

oro

sco

pic

Tim

e (s

eco

nd

s)

Rad

iati

on

Do

se (

mG

y)

Page 14: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Fluoroscopic Time and Radiation Dose Estimates by Clinical Indication

Indication Count RD* FT**

Altered mental status 17 14.9 (9.5) 33 (9.5)

Abnormal neuro exam 13 17.5 (12) 41.6 (13)

Infection 30 11.8 (20) 30.7 (24)

IIH 61 37 (44.6) 44 (38)

Demyelinating / Inflammatory 22 13.4 (25) 34 (36.5)

NPH 6 10.9 (8) 38.5 (29)

Seizure disorder 2 23 (19) 41.6 (31)

Leptomeningeal tumor 33 7 (5) 24 (11)

Miscellaneous 4 21.5 (33) 37.6 (36)

* Mean Radiation Dose in mGy (SD)** Mean Fluoroscopy Time in seconds (SD)

Page 15: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

• Patients with IIH present a greater challenge to successful bedside LP due to elevated BMI and difficulty in palpating the bony landmarks. Laxity of adipose tissues and need for longer needles to reach the thecal sac lengthen the procedure and potentially increase the amount of radiation dose.

• FGLP performed in patients with IIH had the highest mean fluoroscopy time (44 seconds) and mean radiation dose estimates (37 mGy).

Discussion

Page 16: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

Limitations• Retrospective study performed at a single institution

• Regionally specific patient population

• No control for patient anxiety, ability and willingness to cooperate, individual pain threshold, pre-existing degenerative changes or spinal stenosis, and postsurgical scar tissue or hardware, that could certainly impact the procedure time

• BMI (body mass index) data were not available

Page 17: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

• We evaluated the radiation dose and fluoroscopic time in fluoroscopic-guided lumbar punctures at our institution as part of a quality improvement project.

• Monitoring fluoroscopic time and radiation dose in fluoroscopic-guided lumbar puncture is essential to ensure the quality and safety of this procedure and to minimize radiation exposure for both patients and radiation health workers.

Conclusion

Page 18: Radiation Dose and Fluoroscopic Time of Fluoroscopic-guided Lumbar Puncture Zoghbi B, Sharma K, Hadi S, Al-Sabbagh M, Aquino B, Spampinato MV Department

References1. Boddu SR, Corey A, Peterson R et al. Fluoroscopic-Guided Lumbar Puncture: Fluoroscopic Time and Implications of Body Mass Index-A Baseline Study. AJNR Am J neuroradiol 35:1475-80 Aug 20142. Brook AD, Burns J, Dauer E, et al. Comparison of CT and fluoroscopic guidance for lumbar puncture in an obese population with prior failed unguided attempt. J Neurointerv Surg May 2014;6:324-283. Smuck M, Zheng P, Chong T, et al. Duration of fluoroscopic-guided spine interventions and radiation exposure is increased in over-weight patients. PM R 2013;5: 291-96, quiz 2964. Eskey CJ, Ogilvy CS. Fluoroscopy-guided lumbar puncture: decreased frequency of traumatic tap and implications for the assessment of CT-negative acute subarachnoid hemorrhage. AJNR Am J Neuroradiol 2001; 22:571-765. Edwards C, Leira EC, Gonzalez-Alegre P. Residency Training : A failed lumbar puncture is more about obesity than lack of ability. Neurology, 2015 Mar 10;84(10): e 69-72