Quality Control in Refractive Surgery

Stefan Pieger*, M.Sc.

Wendelstein, Germany

* Ingenieurbüro Pieger GmbH

Nidek Germany Office

Page 2: Quality Control in Refractive Surgery

Introduction

Personal experience as application specialist for refractive excimer lasers since 1987. (Meditec, Schwind, Nidek)

Progress in PRK&LASIK was usually based on outcomes analysis.

Excimer Laser Surgery and refractive surgery in general well suited for a systematic approach on quality control.

Page 3: Quality Control in Refractive Surgery

Why Quality Control?

Increase confidence level about refractive procedures offered in your center.

Verify current nomogram settings and make adjustments if necessary.

Reduce enhancement rate. Use for marketing and advertisement. Discover trends and technical problems in

order to react more rapidly. Fulfill requirements of ophthalmic societies.

Page 4: Quality Control in Refractive Surgery

How to collect your data?

Patients files

Excel Spread Sheet

Database Software (Access; Filemaker; etc.)

Outcomes Analysis Software (Datagraph; ASSORT; Refr. Consultant; etc.)

Page 5: Quality Control in Refractive Surgery

How to analyze refractive data?

Standard Refractive Outcomes Stability / Safety / Predictability / Efficacy

Additional Outcome Parameters Astigmatism Outcomes: Surgically Induced change in

Cylinder (SIA); Double Angle Scatter Plot Defocus Equivalent / Contrast Sensitivity in mesopic

conditions / pre OP BSCVA vs. post OP UCVA Wavefront Based Outcomes

Defocus + Cyl (‘aberrometer refraction’) Higher Order RMS / Spherical Aberration (Z12) / Coma /

Trefoil

Page 6: Quality Control in Refractive Surgery

Standard Refractive Outcomes: Safety

1% 0% 2% 0%

22%

13%

54%57%

19%25%

2% 3%0% 0%

0%

10%

20%

30%

40%

50%

60%

70%

80%

90%

100%

lost > 2 lost 2 lost 1 unchanged gained 1 gained 2 gained > 2

2. SAFETY: Change in BSCVA - Percent

1 (770)

3 (653)

month (eyes)

Follow up time interval

Number of eyes per Follow up visit.

2 or more lines lost 3 % at 1 m post OP

Page 7: Quality Control in Refractive Surgery

Standard Refractive Outcomes: Efficacy

4%6%

18%

26%

46%40%

14%14%10%

7%3%4% 5%3%

0%

10%

20%

30%

40%

50%

60%

70%

80%

90%

100%

20/12 orbetter

20/15 20/20 20/25 20/30 20/40 20/50 orworse

4. EFFICACY: UCVA - Percent

1 m (691)

3 m (562)

month (eyes)

Follow up time interval

Number of eyes per Follow up visit.

72% 20/20 or better 3 month post Op

Page 8: Quality Control in Refractive Surgery

Standard Refractive Outcomes: Stability (SEQ)

1. STABILITY: Achieved Change in Refr. over Time

-0,22-0,18-0,20-0,15

-4,70

1189 744 662 42 25-8,00

-7,00

-6,00

-5,00

-4,00

-3,00

-2,00

-1,00

0,00

1,00

pre OP 1 m 3 m 6 m 12 m± 1 StDev

Follow up time interval

Mean value of SEQ

Number of eyes per Follow up visit.

Page 9: Quality Control in Refractive Surgery

Standard Refractive Outcomes: Predictability (SEQ)

PREDICTABILITY: Attempted vs Achieved (Scatter) 744 eyes

y = -0.00x2 + 0.94x + 0.21

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Attempted delta SR equiv. [D]Da

tag

rap

h

Achieved [D]

overcorrected

undercorrected

Number of eyes at follow up

±1 D ‘happiness’ Zone

Trend line

y = -0.00x2 + 0.94x + 0.21Regression Formula: ‘Achieved = 0.94*Attempted’(~6% undercorrection)

Page 10: Quality Control in Refractive Surgery

Astigmatism Outcomes:SIA (based on Vector Analysis)

Attempted Cyl vs SIA (Scatter) 3 month postOP 490 eyes

y = -0.00x2 + 0.90x + 0.04

0

1

2

3

4

5

6

7

0 1 2 3 4 5 6 7

Attempted Cyl [D]

Da

tag

rap

h

Achieved [D]

overcorrected

undercorrected

y = 0.90 * x(~10% undercorrection)

-2/-1@180°

0/-0.5@90°(-0.5/+0.5@180°)

0.5 D CylOvercorrection!

0/-0.5@180°

0.5 D CylUndercorrection!

Page 11: Quality Control in Refractive Surgery

Astigmatism Outcomes:Double Angle Scatter Plot

Double Angle Scatter Plot 3 m postOP 491 eyes

-5

-4

-3

-2

-1

-0

-1

-2

-3

-4

-5

-5 -4 -3 -2 -1 -0 -1 -2 -3 -4 -5

Mean Cyl 0,07 @ 14,4°

Datagraph

45°

90°

135°

Double Angle Scatter Plot 918 eyes

-5

-4

-3

-2

-1

-0

-1

-2

-3

-4

-5

-5 -4 -3 -2 -1 -0 -1 -2 -3 -4 -5

Mean Cyl 0,46 @ 179,6°

Datagraph

45°

90°

135°

PreOP Cyl & Axis PostOP Cyl & Axis

Page 12: Quality Control in Refractive Surgery

Additional Outcomes:pre OP BSCVA vs. post OP UCVA

7%

1%

39%34%

79%

96%91%

98% 95%99%

98%100%

99%100%

1%0%0%

10%

20%

30%

40%

50%

60%

70%

80%

90%

100%

'20/10' 20/12 orbetter

20/15 orbetter

20/20 orbetter

20/25 orbetter

20/30 orbetter

20/40 orbetter

20/50 orbetter

20/60 orworse

preOP BSCVA vs. postOP UCVA - Percent

3 m (397)

preSCVA (703)

month (eyes)

Page 13: Quality Control in Refractive Surgery

Additional Outcomes:Defocus Equivalent

60%64%

86%88%

99%99% 100%100% 100%100% 100%100%

0%

10%

20%

30%

40%

50%

60%

70%

80%

90%

100%

<=0,5D <=1D <=2D <=3D <=4D <=5D

DEFOCUS EQUIVALENT - Percent

1 m (615)

3 m (554)

month (eyes)

SEQ =

SPH + ½ CYL

DEQ =

|SEQ|+|½ CYL|

Page 14: Quality Control in Refractive Surgery

72%76%

92% 92%

99% 99% 100%100%

0%

10%

20%

30%

40%

50%

60%

70%

80%

90%

100%

+- 0,5 +- 1,0 +- 2,0 +- 3,0

Refractive outcome - % within | Attempted | 1 m (744) 3 m (662)

Defocus Equivalent vs. Refractive Outcome

60%64%

86% 88%

99% 99% 100%100%

0%

10%

20%

30%

40%

50%

60%

70%

80%

90%

100%

<=0,5D <=1D <=2D <=3D

DEFOCUS EQUIVALENT - Percent

1 m (615)

3 m (554)

-5.0/+10.0@90°SEQ PlanoDEQ +5

DEQ ~ ‚Blur Circle‘

Page 15: Quality Control in Refractive Surgery

Additional Outcomes:Mesopic Contrast Sensitivity

Contrast Sensitivity Mesopic

1,63

1,88

1,57

1,21

1,68

1,92

1,61

1,22

0,40

0,60

0,80

1,00

1,20

1,40

1,60

1,80

2,00

2,20

A (3cpd) B (6cpd) C (12cpd) D (18cpd)

pre op (112)

1 m (72)

month (eyes)

Spatial Frequency [cycles/degree]

Normal Population Range (Vector Vision CSV 1000)

VA in LogMar Scale

Page 16: Quality Control in Refractive Surgery

Wavefront based Outcomes – 2ndOrder / Coma / Trefoil

„Work in progress“ ! Using Aberrometer Refraction rather

than Manifest Refraction? Presenting horizontal and vertical

Coma individually? Vector calculation to present magnitude and axis in [D]! (0.5 D Coma @ 230°)

Trefoil: Axis? Present only magnitude?

ff

f f

Page 17: Quality Control in Refractive Surgery

Wavefront based Outcomes – Higher Order RMS [µm ±StDev]

RMS higher order over time

0.540.540.510.53

0.31

52 26 29 11 30.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

0.80

0.90

pre OP 1 m 3 m 6 m 9 m

Pupil Diameter: 6.0 mm

Page 18: Quality Control in Refractive Surgery

Wavefront based Outcomes – Spherical Aberration [µm] or [D]

Spherical Aberration (Z12) over time [µm]

-0,24-0,25

-0,23

-0,03

18 13 11 6-0,40

-0,35

-0,30

-0,25

-0,20

-0,15

-0,10

-0,05

0,00

0,05

0,10

pre OP 1 m 3 m 9 m

Spherical Aberration (Z12) DEQ over time [D]

1,521,421,361,31

0,41

18 13 11 7 60,00

0,50

1,00

1,50

2,00

2,50

3,00

pre OP 1 m 3 m 6 m 9 m

Dat

agra

ph

Pupil Diameter: 6.0 mm

Page 19: Quality Control in Refractive Surgery

Making Outcome-based Nomogram Adjustments

Comparison of Laser Settings vs. Achieved change in refraction (and not Attempted vs. Achieved).

Reduce random errors as far as possible as nomograms can only compensate systematic errors!

Must be specific for major laser parameters like OZ, TZ, ablation profile type as well as for refraction types.

Page 20: Quality Control in Refractive Surgery

Nomograms for Individual Patient Groups

Data must be filtered on certain parameters: Refraction Type

(Myp/MyoAsti/Hyp/HypAsti…) Surgery Type (PRK; LASIK; LASEK;

Custom…) Optical Zone Diameter Others (age, laser software version;

humidity…)

Page 21: Quality Control in Refractive Surgery

Nomogram Improvements – Laser Settings vs. Achieved

Laser Settings vs Achieved 1166 eyes

y = -0,01x2 + 1,22x - 0,08

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Laser setting [D]

Da

tag

rap

h

Achieved [D]

overcorrected

undercorrected

y = -0,01x2 + 1,22x - 0,08

reduce attempted SEQ by 22%!

Page 22: Quality Control in Refractive Surgery

Laser Settings vs Achieved 1166 eyes

y = -0,01x2 + 1,22x - 0,08

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Laser setting [D]

Da

tag

rap

h

Achieved [D]

overcorrected

undercorrected

Identify and exclude outliers

Page 23: Quality Control in Refractive Surgery

Laser Setting CYL vs. Surgical Induced change in Astigmatism

Laser Setting (cyl) vs SIA 1480 eyes

y = 0.03x2 + 0.60x + 0.26

0

1

2

3

4

5

6

7

0 1 2 3 4 5 6 7

LaserSetCyl [D]

Da

tag

rap

h

Achieved [D]

overcorrected

undercorrected

1. High Scatter! (further analysis necessary!)

2. 40 % systematic undercorrection

Page 24: Quality Control in Refractive Surgery

Nomograms: General Comments

1. Reduce Scatter by Standardized Surgery and OR Environment

2. Exclude Outliers from Data Analysis

3. Exclude Enhancements

4. Choose appropriate follow up interval (≥ 3 m)

5. Create Formula („-10%“) / Lookup Table or use Nomogram Software

Page 25: Quality Control in Refractive Surgery

Summary

Improving the results of refractive surgery procedures must be based on an individual quality control system.

Nomograms can compensate for systematic errors, but not for random errors.

Modern outcomes analysis software allows constant monitoring of your results.

Conventional Outcomes will be extended by HO-RMS, Spherical Aberration and Coma.