qa for linac kv imaging systems -4442-40814...kv p accuracy & stability kvp fluctuations 100 105...

69
QA for Linac KV Imaging QA for Linac KV Imaging Systems Systems J-P Bissonnette, D Moseley, T Purdie, E White, D Jaffray

Upload: others

Post on 16-Aug-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

QA for Linac KV Imaging QA for Linac KV Imaging SystemsSystems

J-P Bissonnette, D Moseley,T Purdie, E White, D Jaffray

Page 2: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Learning ObjectivesLearning Objectives

• Understand the reasoning behind QA– Define quality metrics– Analyse metrics for quality improvement

• Apply to a device: cone-beam CT

• Apply to an existing RT process

Page 3: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

What is Quality?What is Quality?

• Several dimensions of quality– Technical specs– Staff/client education & credentials– Consistency of performance– Efficiency of service delivery– Continuity and timeliness– Safety– Human factors engineering

Applicable to devices and processes both

Page 4: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Quality MetricsQuality Metrics

• Analyse a process or device in each of its dimensions

• For each dimension:– What is it that can go wrong?– How does this risk impact quality?

• Define an output that can be controlled, measured, or quantified

Page 5: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Quality MetricsQuality Metrics

• Quantifiable measures of success– Recommended maintenance performed– Error rate less than 0.5%– 1 mm objects can be seen

• Can be statistically analysed

• Can be tracked and monitored

Page 6: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Risk AnalysisRisk Analysis

• Faulty mechanical design causing harm to patient?– FDA, EC, Health Canada– Preventative maintenance & daily visual checks

• System unreliable?– Buggy software, unstable components– R&D identifies these issues & report to users

Page 7: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Risk AnalysisRisk Analysis

• Fault resulting in poor image quality?– Portal imaging as a back-up system

• Excessive dose to the patient?– Validation of imaging techniques– Monitor tube output periodically

• Reports incorrect geometrical info?– Subtle, not immediately obvious– May impact success or failure of RT

Page 8: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ConeCone--beam CT: QA of a Devicebeam CT: QA of a Device

• Safety

• Geometric

• System stability

• Image quality

• System infrastructure

• Dose

Page 9: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Safety TestsSafety Tests

• Collision detectors on flat panels– Motion inhibits when activated

• X-ray tube deployment inhibit

• X-ray area monitors

• Grounding tests• Visual inspection for electrical hazards

Page 10: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ConeCone--beam CT: QA of a Devicebeam CT: QA of a Device

• Safety

• Geometric

• System stability

• Image quality

• System infrastructure

• Dose

Page 11: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

GeometryGeometry

• Pixel size calibration

• MV and kV source alignment

• Flexmaps– Refresh periodically– Repeatability

• kV source positioning• Flat panel positioning• Each imaging position

Page 12: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Pixel Size at IsocentrePixel Size at Isocentre

0.52 mm/pixel

Page 13: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

MV GeometryMV Geometry

• Imaging and treatment beams coincide

Page 14: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

MV/kV CoincidenceMV/kV Coincidence

• Treatment is orthogonal to imaging

Page 15: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

kV/MV Calibration ConceptkV/MV Calibration Concept

kV

BB (Reconstruction Iso-centre)

MV Mechanical isocentre

x

y

z

MV Radiation isocentre

Calibrated isocentre

Page 16: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

MV/kV Calibration Procedure MV/kV Calibration Procedure

u

v

Page 17: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Results for Six UnitsResults for Six Units

-5-4.5

-4-3.5

-3-2.5

-2-1.5

-1-0.5

00.5

11.5

22.5

33.5

44.5

5

-180 -135 -90 -45 0 45 90 135 180

Gantry angle (degrees)

Abs

olut

e U

dis

plac

emen

t (m

m)

Unit 7Unit 8Unit 9Unit 10Unit 14Unit 17

Page 18: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

-5-4.5

-4-3.5

-3-2.5

-2-1.5

-1-0.5

00.5

11.5

22.5

33.5

44.5

5

-180 -135 -90 -45 0 45 90 135 180

Gantry angle (degrees)

Abso

lute

V d

ispl

acem

ent (

mm

)

Unit 7Unit 8Unit 9Unit 10Unit 14Unit 17

Page 19: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

LongLong--term Stability:term Stability: FlexmapFlexmap

12 calibrations over 28 months

u

v

-1.5

-1.25

-1

-0.75

-0.5

-0.25

0

0.25

0.5

0.75

1

1.25

-180 -135 -90 -45 0 45 90 135 180

Gantry angle (degrees)

Resi

dual

dis

plac

emen

t (m

m)

95% confidence interval = 0.25 mm

Page 20: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Targeting StudyTargeting Study• 21 sessions:

– Over a 3 month period – Three different operators– Single calibration map

• Geometric calibration maps monitored weekly in parallel study.

Page 21: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

MV/kV DifferenceMV/kV Difference

Page 22: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Daily Geometry QADaily Geometry QA

• Align phantom with lasers

• Acquire portal images (AP & Lat) & assess central axis

• Acquire CBCT• Difference between

predicted couch displacements (MV & kV) should be < 2 mm

Page 23: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Daily Geometry QADaily Geometry QA

• Align phantom with lasers

• Acquire portal images (AP & Lat) & assess central axis

• Acquire CBCT• Difference between

predicted couch displacements (MV & kV) should be < 2 mm

Page 24: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70
Page 25: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70
Page 26: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70
Page 27: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

-2.00

-1.50

-1.00

-0.50

0.00

0.50

1.00

1.50

2.00

01-Jan-06 20-Feb-06 11-Apr-06 31-May-06 20-Jul-06

Dev

iatio

n fr

om is

ocen

tre,

X d

irect

ion

(cm

)

ES07

ES08

NS09

NS10

WS17

Page 28: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Effect of Incorrect CalibrationEffect of Incorrect Calibration

Page 29: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ConeCone--beam CT: QA of a Devicebeam CT: QA of a Device

• Safety

• Geometric

• System stability

• Image quality

• System infrastructure

• Dose

Page 30: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

System StabilitySystem Stability

• Dark image calibration– Acquired prior to each scan

• Gain stability & defect maps– Refresh floods every month

Page 31: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Dark image performanceDark image performanceDark fields

0

1

2

3

4

5

6

7

8

14-J

an-0

4

23-A

pr-0

4

1-Au

g-04

9-N

ov-0

4

17-F

eb-0

5

28-M

ay-0

5

5-Se

p-05

14-D

ec-0

5

24-M

ar-0

6

2-Ju

l-06

10-O

ct-0

6

Date

Stan

dard

dev

iatio

n in

pix

el v

alue

CenteredFull offsetMid offset

Page 32: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Flood Fields: image SNRFlood Fields: image SNR

0

20

40

60

80

100

120

14-J

an-0

4

23-A

pr-0

4

1-Au

g-04

9-N

ov-0

4

17-F

eb-0

5

28-M

ay-0

5

5-Se

p-05

14-D

ec-0

5

24-M

ar-0

6

2-Ju

l-06

10-O

ct-0

6

Date

Sta

ndar

d de

viat

ion

in p

ixel

val

ue

X ray tube changes

Page 33: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

X ray Generator StabilityX ray Generator Stability

• Reproducibility and Accuracy– kVp, mA, ms

Page 34: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

XX--ray Generator: ms Accuracyray Generator: ms Accuracy

0

10

20

30

40

50

60

70

80

90

100

0 10 20 30 40 50 60 70 80 90

Programmed time (ms)

Mea

sure

d tim

e (m

s)

Second experimentLinear fitFirst experiment

Linear fit: Y = 1.002 x + 0.789

Page 35: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

mAs linearity: fixed currentmAs linearity: fixed currentmAs linearity

0

50

100

150

200

250

0 5 10 15 20 25 30

programmed mAs

Mea

sure

d ex

posu

re (m

R)

40 mA80 mA32 mA320 mA160 mA

Page 36: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

kVkVpp accuracy & stabilityaccuracy & stability

kVp fluctuations

100

105

110

115

120

125

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70 73 76

shot

Mea

sure

d kV

p

fixed mA

40 mA 80 mA 160 mA 320 mA 32 mA

117.6 ± 4.6 kVp

Page 37: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ConeCone--beam CT: QA of a Devicebeam CT: QA of a Device

• Safety

• Geometric

• System stability

• Image quality

• System infrastructure

• Dose

Page 38: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Image QualityImage Quality

CatPhan 500 phantom

Page 39: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ScaleScale

• Geometric calibration to tie isocentre to centre of volumetric reconstruction

• Scale to relate all pixels to isocentre

5 cm

Page 40: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ScaleScale

• Geometric calibration to tie isocentre to centre of volumetric reconstruction

• Scale to relate all pixels to isocentre

Page 41: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

UniformityUniformity

• Standard CT tests– Cupping, capping

• Baseline non-uniformity index:

minmax

minmax

CTCTCTCT

+−

Page 42: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Linearity of CT NumbersLinearity of CT Numbers

Page 43: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Linearity of CT numbersLinearity of CT numbers

0

200

400

600

800

1000

1200

1400

1600

1800

2000

0 200 400 600 800 1000 1200 1400 1600 1800 2000

Theoretical Housfield unit

Mea

sure

d Ho

unsf

ield

uni

t Unit 7Unit 8Unit 9Unit 10Unit 12Unit 16Unit 16 with annulusUnit 17

Page 44: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Add ScatterAdd Scatter

IEC standard 61675-1

Page 45: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Linearity of CT numbersLinearity of CT numbers

0

200

400

600

800

1000

1200

1400

1600

1800

2000

0 200 400 600 800 1000 1200 1400 1600 1800 2000

Theoretical Housfield unit

Mea

sure

d Ho

unsf

ield

uni

t Unit 7Unit 8Unit 9Unit 10Unit 12Unit 16Unit 16 with annulusUnit 17

Page 46: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Linearity of CT NumbersLinearity of CT Numbers

• Fairly linear (χ2 > 0.99) for all systems

• Beam hardening

• Scatter conditions

• Non-standard metric; use only as a baseline

Page 47: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Spatial ResolutionSpatial Resolution

GE CT scanner2.5 mm slices 0.5 mm pixels

120 kVp, 100 mA

Page 48: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Low Contrast ResolutionLow Contrast Resolution

1 % contrastvisible

0.5 % contrastvisible

1 % contrastvisible

GE CT scanner2.5 mm slices 1 mm pixels

120 kVp, 100 mA

Page 49: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ConeCone--beam CT: QA of a Devicebeam CT: QA of a Device

• Safety

• Geometric

• System stability

• Image quality

• System Infrastructure

• Dose

Page 50: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

System InfrastructureSystem Infrastructure

• Data safety

• Storage space available

• Review of QA data

Page 51: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Daily CBCT QA ProgramDaily CBCT QA ProgramDimension Procedure Tolerance

Detector stability Dark image calibrationGeometry Localising lasers < 1 mm

MV/kV/laser alignment < 1 mmAccuracy of shifts < 1 mm

Safety Interlocks: interrupts or prevents irradiation

Functional

Warning lights FunctionalWarm-up Generator operation Functional

Detector operation FunctionalDetector signal Within expected rangeCollimator operational Functional

Clinical process issues Database integrityStorage space availability

Page 52: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Monthly CBCT QA ProgramMonthly CBCT QA ProgramDimension Procedure Tolerance

Imaging system Gain stability Replace or refreshperformance Defect maps Replace or refreshImage quality Scale and distances ± 0.5 mm

CT number linearity & stability

Baseline

Image uniformity BaselineHigh contrast spatial resolution

Baseline

Artefacts AbsenceGeometric Geometric calibration Replace / refresh

Accuracy of couch shifts

< 1 mm

Clinical process issues Review of daily test results

Page 53: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Annual CBCT QA Program (service)Annual CBCT QA Program (service)Dimension Procedure Tolerance

X-ray generator kVp accuracy Baselinestability mAs linearity Baseline

Radiation quality (HVL) BaselineAccuracy of mA & mAs Baseline

Geometry Couch scales 1 mmCouch motion accuracy (manual or remote)

1 mm

Detector tilt BaselineDetector skew BaselineDetector scale Baseline

Page 54: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Annual CBCT QA Program Annual CBCT QA Program (upgrades)(upgrades)

Dimension Procedure ToleranceData transfer Link to treatment

planningFunctional and

accurateLong term and short term storage

Functional

Dosimetry Axial and skin doses BaselineClinical process issues Database integrity and

maintenance Baseline

Documentation of imaging procedure

Up-to-date

Review of daily and monthly test results

Completeness

Page 55: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Radiation Therapy ProcessRadiation Therapy Process

Imaging Planning

TreatmentdeliveryVerify

Page 56: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ImageImage--Guided RT: ProcessGuided RT: Process

Verify

Adjust

Imag

e tra

nsfer

Image

Target

Adjust

Verify

BeginTx

InitialSetup

Within tolerance

With

in tol

eran

ceRe-try

Shift

Page 57: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Verify: Quality MetricsVerify: Quality Metrics

• Patient ID

• Prescription

• Location

• Patient set-up

Page 58: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Illustrative Example: SBRT LungIllustrative Example: SBRT Lung

• Small, inoperable lung lesions

• Radio ablative doses (60 Gy in 3 fx)

• High level of confidence– Targeting– Planning– Treatment

Page 59: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Method: Lung SRT ProcessMethod: Lung SRT Process

Patient consentand education

Immobilisation

Steretactic framefor abdominalcompression

Helical and 4DCTimaging: derive

GTV

Planning

Image guidedpositioning at the

treatment unitCBCT

± 3 mm

Treat

Page 60: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Image Guidance: CBCTImage Guidance: CBCT

Planning CTPlanning CT

CBCT Fx #1CBCT Fx #1

CBCT Fx #2CBCT Fx #2

CBCT Fx #3CBCT Fx #3

GTVGTVPTVPTV

Page 61: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

CBCT Image Guidance AccuracyCBCT Image Guidance AccuracyDisplacements (mm)

Medio-lateral Antero-posterior Cranio-caudal

4.7 ± 5.3 7.3 ± 9.4 5.6 ± 4.8

2.1 ± 2.32.7 ± 2.22.0 ± 1.8

Cranio-caudalAntero-posteriorMedio-lateral

Residual Error (mm)

10 patients

Page 62: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ObservationsObservations

•IGRT clearly reduces both the systematic and random errors.

•Residual errors reflect our current tolerance (± 3 mm).

Page 63: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Radiation TherapyRadiation Therapy

• Who?– Record & verify / electronic file

• What?– Planning CT, diagnostic work-up

• When?– Record & verify / electronic file

Page 64: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Radiation TherapyRadiation Therapy

• How?– Treatment plan

• Where?– Daily CBCT

IGRT is a QA tool for externalbeam radiation therapy

Page 65: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Summary and ConclusionsSummary and Conclusions

• Introduced framework for a QA program

– Goal: reduce errors and uncertainties• Devices • Processes

Page 66: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Summary and ConclusionsSummary and Conclusions

• Quality has several dimensions

– Dimensions can be formulated as metrics• Quantifiable• Traceable

– Use quality metrics• Assess performance• Improve quality with time

Page 67: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

Summary and ConclusionsSummary and Conclusions

• Have applied QA framework to a device– Cone-beam CT

• Have applied QA framework to a process– SBRT lung

Page 68: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

ConclusionsConclusions

• Demonstrated that IGRT is routinely used to reduce systematic and random errors & uncertainties

• In this sense, IGRT is a QA tool

Page 69: QA for Linac KV Imaging Systems -4442-40814...kV p accuracy & stability kVp fluctuations 100 105 110 115 120 125 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 67 70

AcknowledgementsAcknowledgements

• Doug Moseley • Tom Purdie• Elizabeth White• David Jaffray