preparation · preparation ① prepare ph standard solutions ② connect the electrode to the meter...

2
Remove the protective cap. Open the internal solution filler port. Preparation Prepare pH standard solutions Connect the electrode to the meter Press the power key of the meter pH9 pH9 pH 4 pH 4 pH 7 pH 7 LIGHT ENT ˞Illustration is image of D-72 Remove the protective cap from the electrode, and open the internal solution filler port Temperature connector Press CODE:GZ0000336672A January, 2017 © 2017 HORIBA Advanced Techno Co., Ltd. Quick Manual Portable pH Meter D-71/D-72 MEAS CAL DATA LIGHT SET MODE ENT http://www.horiba-water.com Electrode connector

Upload: others

Post on 15-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Preparation · Preparation ① Prepare pH standard solutions ② Connect the electrode to the meter ③ Press the power key of the meter pH4 pH9 p H 7 CAL MEAS DATA LIGHT MODE SET

Remove the protective cap.Open the internal solution �ller port.

Preparation① Prepare pH standard solutions

② Connect the electrode to the meter

③ Press the power key of the meter

pH9pH9pH4pH4

pH7

pH7

MEASCAL DATA

LIGHT

SETMODE

ENT

※Illustration is image of D-72

④ Remove the protective cap from the electrode, and open the internal solution �ller port

Temperature connector

Press

CODE:GZ0000336672AJanuary, 2017 © 2017 HORIBA Advanced Techno Co., Ltd.

Quick Manual Portable pH Meter D-71/D-72

MEASCAL DATA

LIGHT

SETMODE

ENT

http://www.horiba-water.com

Electrode connector

Page 2: Preparation · Preparation ① Prepare pH standard solutions ② Connect the electrode to the meter ③ Press the power key of the meter pH4 pH9 p H 7 CAL MEAS DATA LIGHT MODE SET

Press

MEASCAL DATA

LIGHT

SETMODE

ENT④ Start the third point calibration

Immerse the electrode in the pH 7 standard solution until the liquid junction is fully submerged.

Immerse the electrode in the pH 4 standard solution until the liquid junction is fully submerged.

To measurement or second point calibration

To measurement or third point calibration

② Start the �rst point calibration

MEASCAL DATA

LIGHT

SETMODE

ENT

MEASCAL DATA

LIGHT

SETMODE

ENT

MEASCAL DATA

LIGHT

SETMODE

ENT

MEASCAL DATA

SETMODE

MEASCAL DATA

LIGHT

SETMODE

ENT

② Start the measurement

③ Start the second point calibrationMEASCAL DATA

LIGHT

SETMODE

ENT

MEASCAL DATA

LIGHT

SETMODE

ENT

① Enter the calibration mode

Immerse 3 cmor more.

Immerse 3 cmor more.

Immerse 3 cmor more.

① Enter the measurement mode

When the measurement sample is known to be acidic, perform two-point calibration with pH 7 and pH 4. When it is known to be alkaline, perform two-point calibration with pH 7 and pH 9. When the sample pH is not known, perform three-point calibration with pH 4, pH 7 and pH 9. Calibration

To measurement

Immerse the electrode in the pH 9 standard solution until the liquid junction is fully submerged. ※The calibration method is the same as first and second point calibration.

Measurement

Con�rm that the instantaneous value is stable, then press .

Con�rm that the instantaneous value is stable, then press .

When changes from the blinking state tothe lit state and appears, calibration is complete.

When changesfrom the blinking stateto the lit state and appears, calibration iscomplete.

When changesfrom the blinking stateto the lit state and appears, calibration iscomplete.

When performing the second point calibration,press .

When performing the third point calibration,press .

Press

Press

When changes from the blinking stateto the lit state, measurement is complete.

When the batteries run out or the meter will not be used for a long period, remove the batteries.

NOTE