Introduction Bacteria Acinetobacter calcoaceticus-baumannii complex Klebsiella oxytoca Serratia marcescens Enterobacter aerogenes Klebsiella pneumoniae group Staphylococcus aureus Enterobacter cloacae complex Moraxella catarrhalis Streptococcus agalactiae Escherichia coli Proteus spp. Streptococcus pneumoniae Haemophilus influenzae Pseudomonas aeruginosa Streptococcus pyogenes Antimicrobial Resistance Genes CTX-M IMP KPC NDM OXA-48-like VIM mecA/C and MREJ Atypical Bacteria Chlamydia pneumoniae Mycoplasma pneumoniae Legionella pneumophila Viruses Adenovirus Human Rhinovirus/Enterovirus Parainfluenza Virus Coronavirus Influenza A Respiratory Syncytial Virus Human Metapneumovirus Influenza B Routine evaluation of lower respiratory specimens for diagnosis of pneumonia and other important lower respiratory tract infections involves various testing paradigms to identify the microorganisms that may be responsible for the infection. Though molecular methods may be used for viral detection, culture remains the gold standard for diagnosis of bacterial infection from lower respiratory specimens. Quantitative and/or semi- quantitative culture can help distinguish clinically relevant bacteria from colonizing bacteria or normal oral flora based on the abundance of single or multiple bacteria cultured from the specimen 1 . However, culture can be a time-consuming (~24-48 hours) and insensitive procedure, particularly for fastidious bacteria or when evaluating specimens collected after empiric antibiotic therapy has been administered. The FilmArray® Pneumonia Panel is a molecular, multiplex device that can rapidly identify viruses, bacteria, and antimicrobial resistance genes in sputum-like and bronchoalveolar lavage (BAL)-like specimens obtained from individuals with signs of a lower respiratory tract infection (Table 1). Model for Bin Result Precision When tested repeatedly, different measured values within the range of a bin will exhibit variable precision of the reported bin results. Bin precision may be as low as 50% for values at a bin limit and precision will increase (up to 90% or higher) as the distance of the measured value from a bin limit increases. If the bin size is at least two standard deviations (2σ) of the measured value, then the precision of bin results will follow the model illustrated in Figure 4: o >90% at a bin center (Scenario 1) o ~60 – 90% between a bin limit and bin center (Scenario 2) o ~50% at bin limits (Scenario 3) Precision of the FilmArray® Pneumonia Panel Considerations for Interpreting Relative Abundance of Bacterial Nucleic Acids in Lower Respiratory Specimens M Buccambuso, L O’Connor, M Brooks, J Manwaring, T Edwards, M Hockin, J Arce, R Lems, J Larsen, A Fratto, D Abbott, J Southwick, A Judd, and E Amiott BioFire Diagnostics, LLC, Salt Lake City, UT, USA Summary and Conclusion Precision of Bin Results Figure 4. Model for Precision of FilmArray Pneumonia Panel Bin Results Top: The probability of the same bin results for each replicate tested varies based on proximity of the measured value to a bin limit. Bottom: Expected distribution of bin results at different mean measured values. Table 1. Analytes Detected by the BioFire FilmArray Pneumonia Panel Figure 5. Summary of the Precision of Bin Results for Bacteria Detected by the FilmArray Pneumonia Panel Reported percent (of 90 replicates) is graphed as a bar in each bin (x-axis) at each dilution tested (y-axis). Figure 2. Approaches to Diagnosis of Bacterial Lower Respiratory Tract Infections: Semi-quantitative Culture and FilmArray Pneumonia Panel The FilmArray Pneumonia Panel is a molecular, multiplex device developed to identify bacterial and viral pathogens of lower respiratory tract infections and to rapidly and reliably report the approximate abundance of bacterial nucleic acids (±1-log copies/mL) in polymicrobial BAL and sputum-like specimens. Reporting results in bins with discrete limits dictates that at some concentrations, bin precision will be low (~50%) and at other concentrations, bin precision will be high (>90%). Variability in the precision of results should be considered when establishing criteria for device performance evaluations, verifications, and quality control procedures. Contact Maggie Buccambuso Information BioFire Diagnostics, LLC [email protected] ~24-48 hours ~1 hour #P0570 Note: The FilmArray Pneumonia Panel and the FilmArray Pneumonia Panel plus, which includes the detection of Middle East Respiratory Syndrome Coronavirus (MERS-CoV), have not been evaluated by the US FDA or other regulatory agencies. The objective of the bin result is to provide simple and accurate information about the relative abundance of bacteria in lower respiratory specimens in a manner that is similar in concept to current quantitative and semi-quantitative culture-based practices, yet faster and more sensitive (Figure 2). Materials and Methods To evaluate the precision of bin results, testing was performed with contrived BAL-like samples consisting of multiple bacteria titrated in 1-log dilutions (Figure 3; ~10 2 - 10 7 copies/mL of bacterial genomic DNA, measured by digital PCR). Samples were tested repeatedly over multiple days by different operators using different FilmArray systems (FilmArray, FilmArray 2.0 and FilmArray Torch) and instruments, as well as three different Investigational Use Only (IUO) reagent lots. FilmArray Pneumonia Panel Figure 3. Variables Included in the Evaluation of Bin Precision for Bacteria Detected by the FilmArray Pneumonia Panel Reagent Lot Operator Instrument Day lung In the Precision of Bin Results section, the distribution of measured values (data points) and % bin precision (pie charts) for 90 replicates of three bacteria (Staphylococcus aureus, Pseudomonas aeruginosa, and Haemophilus influenzae) at six different dilutions is shown. Results for twelve additional bacteria are shown in Figure 5 and demonstrate, that the FilmArray Pneumonia Panel bin results have the expected precision of ~50% to >90% for all bins over the reportable range of the device. Figure 1. Schematic of FilmArray Pneumonia Panel Bins and Bin Limits Not Detected 10^4 copies/mL 10^5 copies/mL 10^6 copies/mL ≥10^7 copies/mL 10 3.5 10 4.5 10 5.5 10 6.5 ≥10^7 copies/mL 10^6 copies/mL 10^5 copies/mL 10^4 copies/mL Not Detected For bacteria, the panel provides an estimate of nucleic acid abundance for 15 different groups or species. The estimate is based on real-time PCR relative to an internal standard material of known quantity and is reported in bins over a relevant range from 10^4 to ≥10^7 copies/mL. Each bin represents a 1-log range of values with a discrete lower and upper limit. (Figure 1). Semi-Quantitative Culture +++ ++++ ++++ When evaluating and validating the performance of the FilmArray Pneumonia Panel, it is important to recognize that the precision (repeatability and/or reproducibility) of the bin results is a function of the bin structure. Precision will vary depending on the quantity of bacterial nucleic acid measured and its relation to the upper and lower limits of the bins (see model in Figure 4). An evaluation of the predicted and observed precision of bin results is described in this poster. Not Detected 10^4 10^5 10^6 ≥10^7 Reference: 1 Garcia, Lynne S. Clinical Microbiology Procedures Handbook, 3 rd Edition. Washington, D.C. American Society of Microbiology, 2010.

Upload: others

Post on 18-Mar-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Precision of the FilmArray® Pneumonia Panel · 2020. 3. 13. · Klebsiella oxytoca Serratia marcescens Enterobacter aerogenes Klebsiella pneumoniae group Staphylococcus aureus Enterobacter

Introduction

Bacteria Acinetobacter

calcoaceticus-baumanniicomplex

Klebsiella oxytoca Serratia marcescens

Enterobacter aerogenes Klebsiella pneumoniae group Staphylococcus aureus

Enterobacter cloacae complex Moraxella catarrhalis Streptococcus agalactiae

Escherichia coli Proteus spp. Streptococcus pneumoniae

Haemophilus influenzae Pseudomonas aeruginosa Streptococcus pyogenes

Antimicrobial Resistance Genes

CTX-M IMP KPC NDM OXA-48-like VIM

mecA/C and MREJAtypical Bacteria

Chlamydia pneumoniae Mycoplasma pneumoniae Legionella pneumophila

Viruses

Adenovirus Human Rhinovirus/Enterovirus Parainfluenza Virus

Coronavirus Influenza A Respiratory Syncytial Virus

Human Metapneumovirus Influenza B

Routine evaluation of lower respiratory specimens for diagnosis ofpneumonia and other important lower respiratory tract infections involvesvarious testing paradigms to identify the microorganisms that may beresponsible for the infection. Though molecular methods may be used forviral detection, culture remains the gold standard for diagnosis of bacterialinfection from lower respiratory specimens. Quantitative and/or semi-quantitative culture can help distinguish clinically relevant bacteria fromcolonizing bacteria or normal oral flora based on the abundance of singleor multiple bacteria cultured from the specimen1. However, culture can bea time-consuming (~24-48 hours) and insensitive procedure, particularlyfor fastidious bacteria or when evaluating specimens collected afterempiric antibiotic therapy has been administered.

The FilmArray® Pneumonia Panel is a molecular, multiplex device thatcan rapidly identify viruses, bacteria, and antimicrobial resistance genesin sputum-like and bronchoalveolar lavage (BAL)-like specimens obtainedfrom individuals with signs of a lower respiratory tract infection (Table 1).

Model for Bin Result PrecisionWhen tested repeatedly, different measured values within the range of abin will exhibit variable precision of the reported bin results. Bin precisionmay be as low as 50% for values at a bin limit and precision will increase(up to 90% or higher) as the distance of the measured value from a binlimit increases. If the bin size is at least two standard deviations (2σ) of themeasured value, then the precision of bin results will follow the modelillustrated in Figure 4:

o >90% at a bin center (Scenario 1) o ~60 – 90% between a bin limit and bin center (Scenario 2)o ~50% at bin limits (Scenario 3)

Precision of the FilmArray® Pneumonia PanelConsiderations for Interpreting Relative Abundance of Bacterial Nucleic Acids in Lower Respiratory Specimens

M Buccambuso, L O’Connor, M Brooks, J Manwaring, T Edwards, M Hockin, J Arce, R Lems, J Larsen, A Fratto, D Abbott, J Southwick, A Judd, and E AmiottBioFire Diagnostics, LLC, Salt Lake City, UT, USA

Summary and Conclusion Precision of Bin Results

Figure 4. Model for Precision of FilmArray Pneumonia Panel Bin ResultsTop: The probability of the same bin results for each replicate tested varies based on proximityof the measured value to a bin limit.Bottom: Expected distribution of bin results at different mean measured values.

Table 1. Analytes Detected by the BioFire FilmArray Pneumonia Panel

Figure 5. Summary of the Precision of Bin Results for Bacteria Detected bythe FilmArray Pneumonia Panel Reported percent (of 90 replicates) is graphed as a barin each bin (x-axis) at each dilution tested (y-axis).

Figure 2. Approaches to Diagnosis of Bacterial Lower Respiratory Tract Infections: Semi-quantitative Culture and FilmArray Pneumonia Panel

The FilmArray Pneumonia Panel is a molecular, multiplex devicedeveloped to identify bacterial and viral pathogens of lower respiratorytract infections and to rapidly and reliably report the approximateabundance of bacterial nucleic acids (±1-log copies/mL) in polymicrobialBAL and sputum-like specimens. Reporting results in bins with discretelimits dictates that at some concentrations, bin precision will be low(~50%) and at other concentrations, bin precision will be high (>90%).Variability in the precision of results should be considered whenestablishing criteria for device performance evaluations, verifications, andquality control procedures.Contact Maggie Buccambuso

Information BioFire Diagnostics, [email protected]

~24-48 hours

~1 hour

#P0570

Note: The FilmArray Pneumonia Panel and the FilmArray Pneumonia Panel plus, whichincludes the detection of Middle East Respiratory Syndrome Coronavirus (MERS-CoV), havenot been evaluated by the US FDA or other regulatory agencies.

The objective of the bin result is to provide simple and accurateinformation about the relative abundance of bacteria in lower respiratoryspecimens in a manner that is similar in concept to current quantitativeand semi-quantitative culture-based practices, yet faster and moresensitive (Figure 2).

Materials and MethodsTo evaluate the precision of bin results, testing was performed withcontrived BAL-like samples consisting of multiple bacteria titrated in 1-logdilutions (Figure 3; ~102 - 107 copies/mL of bacterial genomic DNA,measured by digital PCR). Samples were tested repeatedly over multipledays by different operators using different FilmArray systems (FilmArray,FilmArray 2.0 and FilmArray Torch) and instruments, as well as threedifferent Investigational Use Only (IUO) reagent lots.

FilmArray Pneumonia Panel

Figure 3. Variables Included in the Evaluation of Bin Precision for Bacteria Detected by the FilmArray Pneumonia Panel

Reagent Lot Operator InstrumentDay

lung

In the Precision of Bin Results section, the distribution of measuredvalues (data points) and % bin precision (pie charts) for 90 replicates ofthree bacteria (Staphylococcus aureus, Pseudomonas aeruginosa, andHaemophilus influenzae) at six different dilutions is shown. Results fortwelve additional bacteria are shown in Figure 5 and demonstrate, thatthe FilmArray Pneumonia Panel bin results have the expected precisionof ~50% to >90% for all bins over the reportable range of the device.

Figure 1. Schematic of FilmArray Pneumonia Panel Bins and Bin Limits

Not Detected

10^4copies/mL

10^5copies/mL

10^6copies/mL

≥10^7copies/mL

103.5 104.5 105.5 106.5

≥10^7copies/mL

10^6copies/mL

10^5copies/mL

10^4copies/mL

Not Detected

For bacteria, the panel provides an estimate of nucleic acid abundancefor 15 different groups or species. The estimate is based on real-timePCR relative to an internal standard material of known quantity and isreported in bins over a relevant range from 10^4 to ≥10^7 copies/mL.Each bin represents a 1-log range of values with a discrete lower andupper limit. (Figure 1).

Semi-Quantitative Culture+++ ++++ ++++

When evaluating and validating the performance of the FilmArrayPneumonia Panel, it is important to recognize that the precision(repeatability and/or reproducibility) of the bin results is a function of thebin structure. Precision will vary depending on the quantity of bacterialnucleic acid measured and its relation to the upper and lower limits of thebins (see model in Figure 4). An evaluation of the predicted and observedprecision of bin results is described in this poster.

Not Detected

10^4 10^5 10^6 ≥10^7

Reference: 1 Garcia, Lynne S. Clinical Microbiology Procedures Handbook, 3rd Edition. Washington, D.C.American Society of Microbiology, 2010.