Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University Students: Cristin Faria & Diego A. Espinosa Faculty Advisor: Dr. Sudhir Shrestha Industry Advisor: Ben Valvodinos Client: North Bay Vision Center Website: http://diabeticretinopathyssu.weebly.com Email: [email protected] [email protected]

Upload: others

Post on 28-May-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

PortableRetinaEyeScanningDevice

1

EngineeringScienceDepartmentSonomaStateUniversityStudents:CristinFaria&DiegoA.EspinosaFacultyAdvisor:Dr.SudhirShresthaIndustryAdvisor:BenValvodinosClient:NorthBayVisionCenter

Website:http://diabeticretinopathyssu.weebly.comEmail:[email protected]@sonoma.edu

Page 2: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

1. ProblemandSolution2. GeneralSystemOverview3. SystemDescriptionandTechnicalComponents4. MarketingandEngineeringRequirements5. TestingandResults6. MaterialsandCosts7. Timeline8. Challenges9. LessonsLearned10. FutureWork11. Conclusion

2

Overview

Page 3: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

3

Problem

• Millionsofpeopleintheworldarediagnosedwithdiabeteseveryyear.

• Leadingtodiabeticretinopathy,adiseasefoundintheretinaoftheindividual.

• Oftenleadingtoblindnessifleftuntreated.

• Thisdiseaseisnoteasilydetected,andnormallynotreversiblewhenfoundinthepatient.

Page 4: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

4

PreviousWorksandourSolution

PeekVision:● Createdcamerasandsoftwareapplicationsonsmartphonestoscanpatients’retinas,butdoes

notutilizemachinelearningsorunsslowly.

Epipole:● Handheldretinalfunduscamera,mustbeconnectedtoWindowsorAndroidandtheInternet.

OurSolution:● AstandaloneproductthatisabletogiveinstantresultsbyutilizingMachineLearningand

ImageProcessing.● NoInternetconnectionnecessary.

Ourdeviceisadiagnostictool,andresultsgivenfromdeviceshouldbetakentoalicensedmedicalprofessional

Page 5: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

5

FinalProduct

● Imageofourfinalproduct.

● Howcomputer,raspberrypi,camera,LEDandbatteryareallconnected.

Page 6: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

Explainalltheparts– diagram

6

SystemOverview

● Systemoverviewthatlooksoverentiredeviceandshowshoweachpieceisconnected.

● Basicoverviewofhowoursystemworksandcanbeimplemented.

Page 7: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

7

HardwareDesign

● Visualrepresentationofallofthehardwarecomponentsweutilized.

● RaspberryPiisthecentralunit.

● Demonstrateshowallcomponentsinteractwitheachother.

Page 8: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

8

SoftwareDesign

● Thisflowchartgivesageneraldescriptionofwhatourprogramdoes.● Initializes,takesapicture,determineswhetherornottheretinaishealthyand

displaystheresultviaLED.● ThisprogramutilizesthemachinelearningtoolboxinMATLAB,RaspberryPi3

ModelB,andLEDtogiveaccurateresults.

Page 9: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

Requirements

MarketingRequirements● Thedevicewillbereliableindeterminingthe

healthstatusofapictureofapatient'sretina.● Thedevicewillnotbefullyautomated,butwill

besimpletouse.● Thedevicewillcapturepicturesofaretina

fundusimageswithhighresolution.● Thedevicewillbemarketedtowardtrained

physiciansandhealthpractitionersinvolvedintreatingpatientsinruralregionswithlimitedaccesstoophthalmologytestingtechnologythroughouttheworld.

9

EngineeringRequirements● Theaccuracyofthescanningmustbeover90%

inorderforthedevicetobereliableindeterminingthestatusofthepicture.

● Thedevicewillconsistoffourcomponents:RaspberryPi3ModelB,ArduCAM(camera),BatteryPack,andLEDboxlight.

● TheprojectrequiresinterfacingMATLABwiththedevicebut,minimalcommandsareusedtoexecutethedevice.

● Functionalityofthedevicewillbeaprecursortotakingascanofapatient’sretina,apracticedonebylicensedindividuals.

Foracompletelistofourrequirementsvisitourwebsite:http://diabeticretinopathyssu.weebly.com

Page 10: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

10

MachineLearningandtheConfusionMatrix

WhatisMachineLearning?● Itisatypeofcomputerprogrammingwhichusesdatatoperformatask.● Themoredata,orimagesyouaddtoaprogram,thebetteritperforms.

WhatistheConfusionMatrix?● Itisatablethatisusedtodescribetheperformanceofaclassifier(methodusedtoclassify

data)onasetoftestdatawherethetruevaluesareknown.● Inourprojectthemostimportantpartofthematrixlookedatthefalsenegativerateand

truepositiveratetodeterminetheoverallaccuracyofouralgorithm.

Page 11: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

11

MachineLearningandourProject

● Ourprojectutilizesmachinelearningandimageprocessinginordertorunthealgorithm,anddifferentiatebetweenimages.

● Inordertocheckbetweenhealthyretinaimagesanddiabeticretinaimagesouralgorithmlooksatbloodvessels,darkspots,damagetotheretinashapetounderstandthedifferencesbetweenthetwocategories.

Page 12: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

12

ClassifierTest

● Testusedtodeterminewhichtypeofclassifierwouldbebestsuitedtotestouralgorithmmostaccurately.

● Tested21differentclassifiersandfoundthat8hadsamebestaccuracyof60%.

● Thistestwascompletedbeforeweimprovedtheaverageaccuracyofouralgorithm.

● LookedatconfusionmatricesandconcludedthatLinearSupportVectorMachine(SVM)wouldgiveusthebestresult.

Page 13: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

13

AccuracyImprovementTest

● InitiallythebestaccuracyfortheLinearSupportVectorModelwas60%whichwas30%lowerthanourminimumgoalaccuracyof90%.

● ToincreaseouraccuracyfirstweeliminatedourGlaucomagroupfromourcode,thisincreasedouraverageaccuracyto83%.

● IncreasedthenumberofimagesinDiabeticRetinopathyfrom15to45,thisactionincreasedouraverageaccuracyto91.7%.

● OurFalsediscoveryrateforDiabeticRetinopathyis2%andis22%forhealthyretinas.

Page 14: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

14

AlgorithmAccuracyTest

● Testedtheaccuracyofthealgorithm,bytesting60images,30withdiabeticretinopathyand30healthyretinalimages.

● Imagestestedwerenotincludedinourtestingalgorithm,weretakenbyourcamera,andhadnotbeenpreviouslyseenbyouralgorithm.

● Wewereabletoconclude5healthyimagesweregivenaninaccurateresult,andonlyasingleimagewithdiabeticretinopathywasgivenaninaccurateresult.

● Givenanaverageaccuracyof90%,diabeticretinopathyaccuracyof97%,andhealthyaccuracyof83%.

● Diabeticretinopathyhadafalsediscoveryrateof3%andhealthyhadafalsediscoveryrateof17%.

Page 15: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

15

AutomationTest

● Automatedtakingpictures,savingthem,andtestingthem.

● Itautomaticallytakeimageswithanexternalcamera,savesthemtoaspecificlocation,andteststhemwithouralgorithm.

● Thisprogramworkswithasingleclickofamouse.

Page 16: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

16

HealthIndicatorTest

● TestedtoconfirmthattheLEDwascompatiblewithMATLABandwouldgiveusthecorrectresult.

● IftheLEDturnsonitmeanstheimagehasdiabeticretinopathy,ifnottheimageishealthy.

● TestedImageresultisDiabeticRetinopathy.

Page 17: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

17

MaterialsandCosts

● SpecialThankstoSOURCEwhogaveus$400tocompleteourproject.

● SpecialthankstoSharahmMarivaniforprovidingcomponents.

Page 18: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

18

Timeline

● TestingandDebuggingourprojecttooklongerduetocameraandRaspberryPiissues.

● Communicationledtosomedrawbacksandtimeconstraints.

● GettingMATLABtocommunicatecorrectlywithRaspberryPicauseddelays.

● Evenwiththeseissueswewerestillabletomakeacompletedproject.

Page 19: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

• WehadtochangetheideaofourprojectfromNVIDIAtoRaspberryPiduetolackofNVIDIAsupport.

• Utilizingmachinelearningandimageprocessing.

• GettingMATLABonRaspberryPi.

• Testingandupdatingouralgorithmtogetittoover90%accuracy.

19

Challenges

Page 20: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

● Progressisnotastraightpath,especiallywhenyouhaveaprojectthatismainlysoftwarebased.

● Thingsgowrong,andbythingswemeaneverything.

● Teamwork,anddedicationarevitalwhentakingonanewproject.

20

LessonsLearned

Page 21: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

21

FutureWork

● The next phase of our project will be to attach a retinal fundus camera to our device in order to take an image of a patient’s retina.

● Introducing, the product to doctors and health practitioners in regions that to do not have access to this technology, and make it so patients in these different regions have access to better medical care and are able to get treated faster.

Page 22: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

22

SpecialThanks

● We would like to thank our Professor Dr. Farid Farahamand, Faculty Advisor Dr. Sudhir Shrestha, Industry Advisor Ben Valvodinos, and Client North Bay Vision Center for giving us feedback, and helping us with the completion of our project.

● We would also like to thank our family and friends, for their encouragement and support throughout our senior design project, those actions and kind words did not go unnoticed.

Page 23: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

• https://www.aoa.org/patients-and-public/eye-and-vision-problems/glossary-of-eye-and-vision-conditions/diabetic-retinopathy

• http://www.diabetes.co.uk/news/2014/may/portable-eye-scanner-to-revolutionise-detection-of-diabetic-retinopathy-96133928.html

• http://www.who.int/mediacentre/news/releases/2003/pr86/en/• https://www.mathworks.com/discovery/machine-learning.html• https://www.sas.com/en_us/insights/analytics/machine-learning.html#• https://www.engineersgarage.com/articles/image-processing-tutorial-

applications• https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4463765/• http://onlinelibrary.wiley.com/doi/10.1046/j.1464-5491.2000.00333.x/full

23

References

Page 24: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

Questions/Comments

24

Questions

Page 25: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

25

ClassifierTestTable1

● Took 21 classifiers and looked at the average accuracy using the confusion matrix.

● Found that 8 different classifiers had same initial accuracy of 60%.

● These accuracies were our initial accuracies before adding anything to our algorithm.

Page 26: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

26

ClassifierTestTable2

● Looked at the True Positive Rate and False Negative Rate of top 8 classifiers.

● Was able to conclude from the test that Linear SVM gave us the best overall accuracy rating in each category.

● False Negative Rate was lower in all three categories and gave us a more well-rounded accuracy.

Page 27: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

27

AlgorithmImprovementTestMatrix1

● Single confusion matrix for Linear Support Vector Model.

● The top row specifies Diabetic Retinopathy, where 41 out of 45 images were correctly categorized, with an accuracy rating of 91%.

● Bottom row specifies the healthy category where 14 out of 15 were categorized correctly with an accuracy rating of 93%.

● Which gave an average accuracy of 92%.

Page 28: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

28

AlgorithmImprovementTestMatrix2

● Confusion matrix for Linear Support Vector Model, that shows true positive and false negative rates.

● For Diabetic retinopathy the true positive rate was 91% and false negative rate was 9%.

● For Healthy the true positive rate was 93% and false negative rate was 7%.

Page 29: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

29

AlgorithmAccuracyTestGraph

● BluebarisDiabeticRetinopathyandorangeisHealthy.

● Tested30imageswithDiabeticRetinopathy,andwasgiven29correctresults.

● Resultedina3%Falsenegativerate.

● Tested30HealthyImagesandwasgiven25correctresults.

● Resultedina17%Falsenegativerate.

Page 30: Portable Retina Eye Scanning Device · Portable Retina Eye Scanning Device 1 Engineering Science Department Sonoma State University ... Visual representation of all of the ... This

30

AutomationTest

● Codeusedinordertoautomateourprogram.

● WritteninMATLABandutilizesourmachinelearningalgorithm.