SOURCE:-

The source depends upon the wavelength range that is required for the study.

Intense source are normally used in order to maximize the output signal and to increase the signal-to-noise ratio.

Tungsten filament lamp have been used in the spectral range from about 600 to 1000nm.

The most popular is high pressure(50 to 70 atm) xenon arc lamp in spectral range of 250 to 2500nm.

Hydrogen and dueterium arcs are useful from about 165 to 250nm.

Krypton lamp can be used from 125 to 165nm.

Open carbon arc lamp from 350 to 700nm can be used but the disadvantage is that carbon electrode in the source are consumed during emission and must be replaced periodically.

3

Page 4: Photoacoustic

Lasers are available for use throughout most of the

spectral range used for PAS.

the lasers can be operated in either the continous

or pulsed mode.

both lasers that emit light and lasers that emit

continuum are used.

E.g argon ion laser, nitrogen laser, Nd:YAG laser,

co2 laser, co laser etc.

The intensity of the source is usually modulated, at

a single wavelength, either by pulsing the source or

by using a constant- intensity source with a

chopper.4

Page 5: Photoacoustic

WAVELENGTH SELECTOR:-

In instruments that use a continuous source, such

as a xenon-arc lamp, the radiation is usually made

monochromatic with a diffraction grating

monochromator.

Sources that emit line radiation, such as some

lasers, can use a monochromator or filter to select

the proper line.

5

Page 6: Photoacoustic

CELLS:-

Cell design depends on the physical state of sample.

The cell must posses one transparent window through

which incident radiation can pass and acoustically

insulated to prevent interference from external signals.

The body is constructed from stainless steel or polished

aluminum, that has large thermal mass.

A substance that has a large thermal mass requires

absorption of a relatively large amount of energy before

its temperature is significantly raised.

In addition cells are also constructed from copper, silver,

and gold.

6

Page 7: Photoacoustic

Because of large thermal mass, scattered incident

radiation that strikes the cell body causes an

insignificant temperature rise and PAS signal

relative to those that occur when radiation strikes

an absorbing samples.

The cell windows are constructed from material that

is transparent.

Sapphire and quartz window are most commonly

used in UV region.

In the visible region glass and plastic windows have

been used.

In IR region – ZnSe is used.

7

Page 8: Photoacoustic

When analyte is a gas, the cell is most often

cylindrical.

1. Single-pass cells:

In those cells that have two windows on the end of

the cylinder, radiation makes a single pass through

the cell.

In some cells acoustic baffle are used to decrease

noise within cell.

2. Multiple-pass cells:

In this type both ends of the cylinder are coated

with a multilayer of a highly reflective dielectric

material.

A hole through the dielectric substance allows

radiation to enter the cell. 8

Page 9: Photoacoustic

The radiation is reflected through the sample

several times by dielectric material.

This type can be used for assays at lower

concentrations than single-pass cells.

3. Differential cell:

Often consist of two sequentially connected cells

within the same cylinder.

Each cell use separate detector. The incident

radiation sequentially passes through the cell.

The combined cell and detector is a spectrophone.

PAS in the IR region has been used as detector for

gas chromatography.

9

Page 10: Photoacoustic

10

Page 11: Photoacoustic

The distance between the sample and the window

should be greater than the length the length of thermal

diffusion in order to prevent loss of thermal energy from

the sample through the window.

The volume of the filler gas within the cell is kept as

small as possible in order to maximize the PAS signal.

The filler gas chosen to be non absorbing. Often air or

an inert gas is used. Used of an inert gas generally

results in greater analytical sensitivity.

If the sample is a thin layer of a thermally thin

substance, incident radiation can strike the sample

without first passing through the filler gas because heat

can rapidly travel to inner surface of the sample.

If the sample is not a thin layer of thermally thin

substance, radiation usually strikes the sample from

above after passing through a filler gas.11

Page 12: Photoacoustic

DETECTORS:-

Microphone and piezoelectric transducer(PZT) have been used as detector for PAS.

Of the several type of microphones, the condensermicrophone is most common used.

1. Condenser microphone:

Consist of circular, thin metallic diaphragm that is mounted by its edge near a rigid stationary metallic plate.

The two plate serve as the plate of capacitor in an electric circuit. The thin diaphragm is forced closer to the stationary plate whenever a pressure wave strikes it.

When pressure is no longer exerted on the diaphragm by the wave, the diaphragm is returned to its initial position by tension in the diaphragm.

12

Page 13: Photoacoustic

The result is diaphragm that vibrates at the frequency of the modulated radiation that strikes the sample.

The distance through which the diaphragm modulates is the function of amplitude of pressure wave that is emitted from the cell.

Microphone cannot be used in liquids.

Piezoelectric transducer are primarily used for detecting PAS signal in solutions.

2. Piezoelectric detector:

Is a solid material such as lead zirconate titanate which has a structure that is altered with application of pressure.

Structural alteration causes a rearrangement of charge within a substance and an alteration of the potential difference across the material.

A piezoelectric transducer changes a pressure wave into an electric signal which subsequently is amplified and monitered.

13