pacemakers 2013 12 -7 (a)

Upload: rofi-irman

Post on 13-Apr-2018

221 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    1/55

    Muhammad Diah

    Instalasi Kateterisasi Jantung

    Cardiology Sub Div, Depart Internal Medicene

    RSUD Zainoel Abidin, Banda Aceh

    Cardiology Sub Div, Depart Internal Medicene

    RSHM Palembang

    1

    ECG - Pacemakers

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    2/55

    2013-12-7. M.Diah2

    *First described in 1952

    *Introduced into clinical practice in 1960

    *First endocardial defibrillators in 1980

    *1991 in USA 1 million people had permanent

    pacemakers

    *Now

    - Approximately 3 million with pacemakers

    - Approximately 1 million with ICD device

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    3/55

    2013-12-7. M.Diah3

    *Indications

    *Basics, Pacemaker Components and Code

    *ECG in Pacemaker

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    4/55

    2013-12-7. M.Diah4

    *Provides electrical stimuli to cause cardiac

    contraction when intrinsic cardiac activity is

    inappropriately slow or absent*Sense intrinsic cardiac electric potentials

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    5/55

    * Stimulate cardiac depolarization

    * Sense intrinsic cardiac function

    * Respond to increased metabolic demand by

    providing rate responsive pacing

    * Provide diagnostic information stored by the

    pacemaker

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    6/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    7/55

    2013-12-7. M.Diah7

    Clinical Indication

    for Pacer1. Symptomatic bradycardia

    2. Symptomatic heart block

    - 2nddegree heart block

    - 3rdor complete heart block

    - Bifasicular or transfasicular bundlebranch blocks.

    3. Prophylaxis

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    8/55

    2013-12-7. M.Diah 8

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    9/55

    *

    *Pulse Generator

    *Electronic Circuitry

    *Lead System

    2013-12-7. M.Diah9

    Lead

    Pace

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    10/55

    2013-12-7. M.Diah10

    *Subcutaneous or submuscular

    *Lithium battery*4-10 years lifespan

    *long life and gradual decrease in power

    sudden pulse generator failure is anunlikely cause of pacemaker malfunction

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    11/55

    2013-12-7. M.Diah11

    PPM

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    12/55

    2013-12-7. M.Diah12

    *Sensing circuit*Timing circuit

    *Output circuit

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    13/55

    2013-12-7. M.Diah13

    Bipolar

    *Lead has both negative,

    (Cathode) distal andpositive, (Anode) proximalelectrodes

    *Separated by 1 cm

    *Larger diameter: moreprone to fracture

    *Compatible with ICD

    Unipolar

    * Negative (Cathode)

    electrode in contactwith heart

    * Positive (Anode)electrode: metal casingof pulse generator

    * Prone to oversensing

    * Not compatible withICD

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    14/55

    *current travels only a

    short distance

    between electrodes

    *small pacing spike:

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    15/55

    *current travels a

    longer distancebetween electrodes

    *larger pacing spike:

    >20mm

    2013-12-7. M.Diah15

    -

    Anode

    Cathode

    +

    -

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    16/55

    2013-12-7. M.Diah16

    I

    Chamber

    Paced

    II

    Chamber

    Sensed

    III

    Response

    to Sensing

    IV

    Programmable

    Functions/RateModulation

    V

    Antitachy

    Function(s)

    V: Ventricle V: Ventricle T: Triggered

    P: Simple

    programmable P: Pace

    A: Atrium A: Atrium I: InhibitedM: Multi-

    programmableS: Shock

    D: Dual (A+V) D: Dual (A+V) D: Dual (T+I) C: Communicating D: Dual (P+S)

    O: None O: None O: None R: Rate modulating O: None

    S: Single(A or V)

    S: Single(A or V)

    O: None

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    17/55

    *1stletter chamber paced

    *2

    nd

    letter

    chamber sensed*3rdletter Response to chamber sensed

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    18/55

    2013-12-7. M.Diah18

    *AAT

    *Paces atria

    *Senses atria

    *Triggers generator to fire if atria sensed

    *VVI

    Ventricular Pacing : Ventricular sensing; intrinsicQRS Inhibits pacer discharge

    *VVIR

    As above + has biosensor to provide Rate-responsiveness

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    19/55

    2013-12-7. M.Diah19

    *DDDPaces + Senses both atrium + ventricle,intrinsic cardiac activity inhibits pacer d/c, noactivity: trigger d/c

    *DDDR

    As above but adds rate responsiveness toallow for exercise

    *VVI*Paces ventricle

    *Senses ventricle

    *Inhibited by a sensed ventricular event

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    20/55

    2013-12-7. M.Diah20

    When the need for oxygenated blood increases,

    the pacemaker ensures that the heart rate

    increases to provide additional cardiac outputAdjusting Heart Rate to Activity

    Normal Heart Rate

    Rate Responsive Pacing

    Fixed-Rate Pacing

    Daily Activities

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    21/55

    * Wallet card: 5

    letter code

    * CXR: code visible

    * Single lead in

    ventricle: VVI* Separate leads

    DDD or DVI

    2013-12-7. M.Diah21

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    22/55

    *VVI - lead lies in

    right ventricle*Independent of atrial

    activity

    *Use in AV conductiondisease

    2013-12-7. M.Diah22

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    23/55

    2013-12-7. M.Diah23

    VVI / 60

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    24/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    25/55

    *Typically in pts withnonfibrillating atria and

    intact AV conduction*Native P, paced P, native

    QRS, paced QRS

    *ECG may be interpreted

    as malfunction whennone is present

    *May have fusion beats

    2013-12-7. M.Diah25

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    26/55

    Rate = 60 bpm / 1000 ms

    A-A = 1000 ms

    APVP

    APVP

    V-AAV V-AAV

    2013-12-7. M.Diah26

    Atrial Pace, Ventricular Pace (AP/VP)

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    27/55

    Atrial Spike Ventricular Spike*

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    28/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    29/55

    Rate = 60 ppm / 1000 ms

    A-A = 1000 ms

    APVS

    APVS

    V-AAV V-AAV

    2013-12-7. M.Diah29

    Atrial Pace, Ventricular Sense (AP/VS)

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    30/55

    ASVP

    ASVP

    Rate (sinus driven) = 70 bpm / 857 ms

    A-A = 857 ms

    2013-12-7. M.Diah30

    Atrial Sense, Ventricular Pace (AS/ VP)

    V-AAV AV V-A

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    31/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    32/55

    Rate (sinus driven) = 70 bpm / 857 ms

    Spontaneous conduction at 150 ms

    A-A = 857 ms

    ASVS

    ASVS

    V-AAVAV

    V-A

    2013-12-7. M.Diah32

    Atrial Sense, Ventricular Sense (AS/VS)

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    33/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    34/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    35/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    36/55

    *EKG abnormalities due to

    *Failure to output

    *Failure to capture

    *Sensing abnormalities

    *Operative failures

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    37/55

    *Definition

    *No pacing spike present despite indication to

    pace

    *Etiology*Battery failure, lead fracture, break in lead

    insulation, oversensing, poor lead connection,

    cross-talk

    *Atrial output is sensed by ventricular lead

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    38/55

    *Definition

    *Pacing spike is not followed by either an atrial or

    ventricular complex

    *Etiology*Lead fracture or dislodgement, break in lead

    insulation, elevated pacing threshold, MI at lead

    tip, drugs, metabolic abnormalities, cardiac

    perforation, poor lead connection

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    39/55

    *Oversensing

    *Senses noncardiac electrical activity and is inhibited fromcorrectly pacing

    *Etiology

    *Muscular activity (diaphragm or pecs), EMI, cell phoneheld within 10cm of pulse generator

    *Undersensing

    *Incorrectly misses intrinsic depolarization and paces

    *Etiology

    *Poor lead positioning, lead dislodgement, magnetapplication, low battery states, MI

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    40/55

    Due to pacemaker placement

    *Pneumothorax

    *Pericarditis

    *Perforated atrium or ventricle

    *Dislodgement of leads

    *Infection or erosion of pacemaker pocket*Infective endocarditis (rare)

    *Venous thrombosis

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    41/55

    Pacemaker syndrome

    *Patient feels worse after pacemaker placement

    *Presents with progressive worsening of CHF symptoms

    *Due to loss of atrioventricular synchrony, pathway now

    reversed and ventricular origin of beat

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    42/55

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    43/55

    *Can interfere with function of pacemaker or

    ICD*Device misinterprets the EMI causing

    *Rate alteration

    *Sensing abnormalities

    *Asynchronous pacing

    *Noise reversion

    *Reprogramming

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    44/55

    Examples

    *Metal detectors

    *Cell phones

    *High voltage power lines

    *Some home appliances (microwave)

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    45/55

    *Intensity of electromagnetic field decreases inversely

    with the square of the distance from the source*Newer pacemakers and ICDs are being built with

    increased internal shielding

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    46/55

    *CC: Chills, rigors

    *HPI:

    *65 yom c/o fevers, chills, rigors x 1 day.Positive n/v and anorexia. Pt states hehad recent pacemaker insertion 4 days

    ago for an arrhythmia.

    *PMH:

    *HTN

    *Arrythmia*Hypercholesterolemia

    *PSHx:

    *As stated above

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    47/55

    *

    *Physical exam

    *Temp 101.2, HR 110, BP 90/55

    *EKG

    *Diagnosis?

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    48/55

    *

    *Pocket Infection

    *Pacemaker insertion is a surgical procedure

    *1% risk for bacteremia*2% risk for pocket infection

    *Usually occurs within 7 days of pacemakerinsertion

    *May have tenderness and redness overpacemaker site

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    49/55

    *

    *CC: SOB

    *HPI:

    *65 yom states he had onset of shortnessof breath and left sided pleuritic chestpain. Pt states he awoke with pain anddifficulty breathing. Had pacemaker

    placed yesterday.*PMHx:

    *HTN, Diabetes, Hypercholesterolemia,Arrythmia, CAD

    *PSHx:*Pacemaker, left knee surgery, b/lcataract

    *

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    50/55

    *

    *Physical Exam*BP 146/85, HR 80s, RR 30s, O2 Sat 88%

    *Lungs

    *Decreased breath sounds on left

    *EKG

    *Diagnosis?

    *

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    51/55

    *

    *Pneumothorax

    *Occurs during cannulation of central veins

    *Incidence

    *Cardiologist dependent

    *Treatment

    *Small or asymptomatic observation

    *Large or symptomatic Chest tube

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    52/55

    *

    *CC: Cardiac arrest

    *HPI: 59 yom found on couch. Wife states they

    were watching TV when patient let out a moanand then became unconscious. She states, hehas a bad heart and had somethingput in afew years ago.

    *PMHx: unknown

    *Meds: bottles in bathroom

    *

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    53/55

    *

    *Physical Exam*Airway patent, no visible chest rise, no

    pulses

    *Generally: cool, clammy, diaphoretic

    *EKG:

    *Diagnosis?

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    54/55

    *

    *Cardiac Arrest with ICD (V-fib)

    *2% annual incidence with ICD

    *Etiology

    * ICD delivered predetermined shocks for

    identified event and patient failed to respond

    * ICD failed to recognize event and failed to shock

    appropriately

    *Failure to sense, lead fracture, EMI, inadvertent ICDdeactivation

  • 7/27/2019 Pacemakers 2013 12 -7 (a)

    55/55

    *

    *Cardiac Arrest with ICD

    *Treat using ACLS protocols

    *Secure airway

    *CPR

    *Defibrillate/shock as warranted

    *Keep sternal pad 10 cm away from pulse generator

    *Meds