Page 1: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Oxygen Saturation Monitors

& Pulse Oximetry

© D. J. McMahon 141105 rev cewood 2017-11-142

Page 2: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

3

Page 3: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Key PointsOxygen Saturation Monitors & Pulse Oximetry :

• Role of hemoglobin in respiration• Basic principle of how SpO2 is determined by absorption of red and infra-red• Know the wavelengths of red and infra-red used in a SpO2 sensor• Oxygen saturation SpO2 vs oxygen content in the blood

i.e. the ‘oxygen dissociation curve’• Tone modulation in pulse oximeters - why?• Know the forms Hemoglobin can take• What does Perfusion Index (PI) measure• Sources of inaccuracy in SpO2 measurement• Detection of cerebral O2 saturation• Know the names Nellcor & Masimo

www.youtube.com/watch?v=WXOBJEXxNEoor http://safeYouTube.net/w/owkc

4

Page 4: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

‘Ventilation’ refers to the physics of gasmovement into the lungs. (mechanical)

‘Respiration’ refers to the exchange ofoxygen and carbon dioxide in cellsby means of hemoglobin in the blood.

(chemical)

5

Page 5: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Hemoglobin -

Each hemoglobin unit has 4 heme groups, each of whichcan carry one oxygen molecule.

6

Page 6: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Heme -

7

Page 7: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Oxygen transport in the blood -

8

Page 8: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Saturation defined:

Assume 100 hemoglobin molecules together carry a maximum of 400 (100 x 4) oxygen molecules.

If these 100 hemoglobin molecules were carrying380 oxygen molecules they would be carrying(380 / 400) x 100 = 95% of the maximum number of oxygen molecules that could carry.

They would be “95% saturated”.

Oxygen saturation is referred to as “SpO2”.

(Peripheral capillary Oxygen Saturation) 9

Page 9: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

PaO2 and PvO2 are the partial pressures of O2 in thearterial or venous blood. Same applies to CO2

SaO2 and SpO2 (saturation) both describe the percentage of hemoglobin that is carrying oxygen, but:

SaO2 is determined by laboratory analysis (a co-oximeter).SpO2 is determined by a pulse-oximeter.

Pulse oximeters are widely used, but SaO2 is still consideredthe ‘gold standard’ for assessing oxygenation.

10

Page 10: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

The Oxygen Dissociation Curve:

11

Page 11: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Effects of blood CO2, temperature, or pH on the curve:

12

Page 12: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Basis for SpO2 monitoring:De-oxygenated hemoglobin (Hb) partially absorbs light at 660nm. Oxygenated hemoglobin partially absorbs light at 940nm. (visible light is 400-720 nm)

If we measure the absorption of both wavelengths at any given instant and determine their ratio, we have the percentage of total hemoglobin that is carrying oxygen. i.e. the “Oxygen saturation”.

In practice, two LED’s each send a 660nm or 940nm signal through any perfused tissue to a sensor, which receives the partially absorbed signals.

Those signals are sent to the monitor where they are processed and display the % saturation, pulse rate, and (optionally) a plethysmographic (volume) waveform.

13

Page 13: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

(absorption)

lower absorption;high signal level

higher absorption;low signal level

OxyHemoglobin

de-oxygenatedHemoglobin

14

Page 14: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

a way to remember this...

15

Page 15: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

660 nM 940 nM

IRUV16

Page 16: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

What the monitor sees:

17

Page 17: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

660 nM red940 nM IR

18

Page 18: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

19

Page 19: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Each manufacturer has a specific R-curve for their monitors, determined by the components and algorithm they use.

(When testing monitors, we must know the manufacturer’s R-curve.)

R-curve for an SpO2 monitor:

(note the non-linear line)

20

Page 20: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

21

Page 21: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Tone Modulation in Pulse Oximeters:Almost all pulse oximeters emit a brief audio tone with each pulse, whose pitch is harmonized with the saturation value.

The higher the saturation, the higher the pitch.

listen... http://www.hark.com/clips/wshsvssnrl-pulse-oximeter 22

Page 22: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

23

Page 23: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

24

Page 24: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

25

Page 25: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

26

Page 26: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

27

Page 27: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Forms of Hemoglobin:Hemoglobin Hb: Hemoglobin without attached O2)

Oxyhemoglobin HbO2: Hemoglobin with attached O2)

Carboxyhemoglobin COHb: Hemoglobin with attached CO, Carbon Monoxide) gives inaccurate SpO2

Methemoglobin MetHb: (pronounced MET-hemoglobin) cannot bind to O2 and gives inaccurate SpO2

Sources of Inaccuracy in Pulse Oximetry:Motion of the perfused tissue (Motion Artifact)

Poor tissue perfusion due to cold temp or vascular disease

Carboxyhemoglobin: typically high in smokers. (May be mis-read as HbO2.)

Methemoglobin: from toxins. (May be mis-read as either HbO2 or Hb.)

Diagnostic dyes in bloodstream: methylene blue, cardio green, & others

Nail polish

Skin pigmentation

Ambient light that is rich in infrared (e.g. surgical lights) 28

Page 28: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Inaccuracy due to Carboxy-Hb and Met-Hb:

660 nM (red) 940 nM (infrared) 29

Page 29: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Perfusion Index (PI):

• A measurement of signal strength, which indicates the amount of pulsatile blood flow at the specific site on the patient’s body where a sensor is placed.

• References the pulse signal to the offset.• Especially useful in newborns.

30

Page 30: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Pleth Variability Index (PVI):

Measures the changes in PI (Perfusion Index) during each respiratory cycle.

Useful to check balance between the intrathoracic (chest) pressure and intravascular fluid volume.

e.g. during asthma attack, or to adjust the patient’s blood volume. 31

Page 31: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Oxygen Saturation (SpO2)

Pulse Rate (PR)

Perfusion Index (PI)

Hemoglobin (SpHb ®)

Oxygen Content (SpOC ®)

Carboxyhemoglobin (SpCO ®)

Methemoglobin (SpMet®)

Pleth variability index (PVI®)

note: all ® Masimo

Advanced hematological monitoring:

32

Page 32: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

33

Page 33: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Proprietary algorithm for Masimo’s Signal Extraction Technology (“SET”)

34

Page 34: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

“SatSecond” as analarm parameter:

The change in saturation and the time span factored together as a determinate for alarms.

Two examples where no alarm is activated:

35

Page 35: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

“SatSecond” as analarm parameter:

The change in saturation and the time span factored together as a determinate for alarms.

Two examples where an alarm is activated:

36

Page 36: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

A small, digital memory chip resides within every Nellcor OxiMaxsensor, transmitting sensor-specific data to the monitor to enhance pulse oximetry performance and provide responsive,clinician-friendly features.

Nellcor’s OxiMax™ technology -

37

Page 37: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Earlier calibration of sensor(1980’s & 1990’s)

Nellcor’s ‘OxiMax’ calibration

38

Page 38: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Nellcor’s “Max-Fast” sensor:

Reflects signal from bone at the forehead.

Greater sensitivity,

Eliminates motion artifact 39

Page 39: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

40

Page 40: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

41

Page 41: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

42

Page 42: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

43

Page 43: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Disposable sensors -

44

Page 44: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

45

Page 45: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

46

Page 46: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Reusable sensors -

47

Page 47: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Nonin’s “Onyx” miniature pulse oximetry monitor:

First self-contained miniature SpO2 monitor48

Page 48: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Onyx knock-offs -

49

Page 49: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

50

Page 50: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Regional O2 Saturation (rSO2):Detection of O2 saturation in the brain

Detects red and infrared signals by reflection in brain tissue

51

Page 51: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

rSO2: Bilateral cerebral oxygenation during surgery:

52

Page 52: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Regional Oxygen saturation (rSO2)Cerebral Oxygen Monitoring:

Somanetics InVos System

CASMed ForeSight system53

Page 53: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

http://www.somanetics.com/our-technology

Somanetics ‘INVOS’ rSO2 monitor:

54

Page 54: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Most major manufacturers include either Nellcor or Masimo SpO2 technology in their monitors as ‘OEM boards’:

55

Page 55: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Manufacturers of stand-alone monitors:- Nellcor

- Masimo- Nonin

- Invivo Research

Manufacturers of multi-parameter monitors who licenseSpO2 subassemblies from Nellcor or Masimo:

- Philips- Spacelabs

- Datex-Ohmeda- G.E. Medical

Major Players in Pulse Oximetry:

56

Page 56: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Pulse Ox testers for biomed shops -

There is no such thing as an SpO2 ‘simulator’ !

A simulator mimics a parameter with a directly equivalent voltage, impedance, pressure, etc.

There are no true SpO2 simulators because O2 saturation cannot be directly duplicated in a test device. SpO2 test devices only mimic the SpO2 by means of an optical or opto-mechanical component.

ref: www.hospimedica.com/whitepapers/52c1c251319ad_SP02.pdf57

Page 57: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Netech’s “FingerSim” -

Pronk’s “OxSim OX-1” -

Pulse Ox testers for biomed shops -

Fluke Spot Light -

58

Page 58: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

Pulse Ox testers for biomed shops -

Fluke Index 2 -

Fluke ProSim 8 withSpO2 tester

Clinical Dynamics“SmartSat 59

Page 59: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

SpO2 monitors should be tested at a variety of saturation values, normal and abnormal, including variations in the pulse rate, and pulse volume (or amplitude), if possible.

Example of tester options:

60

Page 60: Oxygen Saturation Monitors & Pulse Oximetryfacweb.northseattle.edu/cwood/EET286-Fall-2017...Oxygen Saturation Monitors & Pulse Oximetry : • Role of hemoglobin in respiration •

61