oscillometric noninvasive blood pressure measurements: the influence of external artifacts

14
OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS V. Jazbinsek, J. Luznik, S. Mieke* and Z. Trontelj Institute of Mathematics, Physics and Mechanics, University of Ljubljana, Ljubljana, Slovenia *Physikalisch-Technische Bundesanstalt, Berlin, Germany

Upload: duard

Post on 10-Jan-2016

37 views

Category:

Documents


0 download

DESCRIPTION

OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS. V. Jazbinsek, J. Luznik, S. Mieke * and Z. Trontelj Institute of Mathematics, Physics and Mechanics, University of Ljubljana, Ljubljana, Slovenia - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS

V. Jazbinsek, J. Luznik, S. Mieke* and Z. Trontelj

Institute of Mathematics, Physics and Mechanics, University of Ljubljana, Ljubljana, Slovenia

*Physikalisch-Technische Bundesanstalt, Berlin, Germany

Page 2: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Introduction

Non-invasive blood pressure (NIBP) measuring devices are based on recording the arterial pressure pulsations in an inflated cuff wrapped around a limb during the cuff pressure deflation.

The recorded NIBP data contains the pressure pulses in the cuff, called oscillometric pulses, superimposed on the cuff pressure deflation.

Page 3: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Objective to separate the deflation and

external artifacts from the pressure pulses and to obtain the typical shapes of pressures changes in the cuff due to those artifacts

Page 4: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Measurements

EU-project “Simulator for NIBP”

Compressor for the cuff inflation and a sensor for pressure detection, built in a PC (LODE, NL)

25 healthy volunteers took part in this research

Page 5: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Measurement Measured data

Arterial pressure pulses, called oscillometric pulses, superimposed on the cuff pressure deflation.

Filtered pulsesDeflation can be removed from data by the digital high pass filtering. Subtraction of negative envelope (curve defined by blue dots) leads to

Oscillometric pulsesThe waveform describing the pulse amplitudes (red dots) as a function of the cuff pressure is a base for automated determination of the systolic and diastolic pressures.

Page 6: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

• Segmentation of data (red) into pulses (blue vert. lines)

• The deflation signal is calculated by the interpolation of data between subsequent segment borders (green)

• Subtraction of deflation from the measured data to obtain only pulses with positive deflections (oscillometric pulses - black)

Data presentation

Page 7: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Mesurements: tapping on the cuff with a plastic tube every 2 s

Segmentation of artefact was obtained from filtered data (band pass

2.5-250 HZ) by selecting 0.7 s time interval around the artefact peak.

Extraction of tapping artifact

Page 8: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Typical shapes of high frequency artifacts

Averaging of tapping artifacts

Page 9: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Projection into normalized heart beat view – reference waveform

Page 10: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Subtraction of reference waveform

a)3rd recording

b)Subtracted reference using waveform optimization (WO)

c) Subtracted reference using single beat optimization (SBO)

Page 11: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Extraction of fist closing artifactMesurements: fist closing every 5 s

Page 12: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Typical shapes of low frequency artifacts

Page 13: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Conclusion I

High frequency artifacts (tapping, vibrations) can be exctracted by applying digital fitering

Page 14: OSCILLOMETRIC  NONINVASIVE  BLOOD  PRESSURE MEASUREMENTS: THE  INFLUENCE  OF  EXTERNAL  ARTIFACTS

Conclusion II

Low frequency artifacts (moving) extraction require:

• Transformation into heartbeat view to obtain the reference signal.

• Reference can be applied as a template to extract a normal oscillometric waveform used for automated determination of systolic and diastolic pressures.

• Subtraction of normalized oscillometric waveform from contaminated data.

• Averaging of extracted artifacts to obtain typical shape.