on-line flare btu monitoring with process mass spec for ... · on-line flare btu monitoring with...

16
On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance Don Rodriguez, Robert Paddison Thermo Fisher Scientific

Upload: others

Post on 10-Aug-2020

7 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

On-Line Flare BTU Monitoring

with Process Mass Spec for RSR

complianceDon Rodriguez, Robert Paddison

Thermo Fisher Scientific

Page 2: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Regulatory Background

Refinery Sector Rule - The final rule is based on the risk and technology review of two

refinery emissions standards: • National Emission Standards for Hazardous Air Pollutants From Petroleum Refineries (40 CFR

Part 63 Subpart CC, "Refinery MACT 1“ – Max. Achievable Control Technology)

• National Emission Standards for Hazardous Air Pollutants for Petroleum Refineries: Catalytic

Cracking Units, Catalytic Reforming Units, and Sulfur Recovery Units (40 CFR Part 63 Subpart

UUU, "Refinery MACT 2").

Page 3: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

What is Process Mass Spectrometry?

• It is NOT GC/MS

• All molecules in the stream are ionized and fragmented

• Each compound produces a signature fragmentation pattern

(fingerprint)

• Separation and quantitation are solution to mathematical matrix

• Matrix must have as many equations as unknowns

• Process mass spec only quantifies those compounds that you tell it

to look for

Page 4: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Example mass spectra

Page 5: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Background

• In a flare, vented gases are combined with steam and/or air and

burned to produce water vapor, carbon dioxide, trace NOX & SOX

compounds

• Flaring occurs during start-up, shut-down and during unplanned

operational interruptions and upsets

• Increasing regulatory control on flares• US EPA 40 CFR Part 63 subpart CC – must measure flare vent gas to

determine Net Heating Value (BTU/SCF)

• May use GC (63.670 (j1)), Calorimeters (63.670 (j1)), or Process Mass

Spectrometers (EPA Generally Applicable AMP, Feb. 5th, 2018)

• Component analysis (GC & Mass Spec) assists with root-cause analysis

by identifying individual components that are present in the flare g

• Net Heating Value

X1*Ideal Net Heating Value1+X2*Ideal Net Heating Value2+X3*Ideal Net

Heating Value3...+Xn*Ideal Net Heating Valuen

• as

Page 6: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

RSR for Flare - SpecificsReductions in Flare Emissions

Under the new rule, provisions have been implemented to assure 98%

combustion efficiency, superseding the flare General Provision requirements.

Flares must now meet the following operating conditions:

• Continuously lit pilot flame at all times

• No visible emissions when flare vent gas is below the smokeless

capacity of the flare

• Net heating value for the combustion zone gas (NHVcz), ≥ 270

BTU/ft3 on a 15-minute average, calculated using flare vent gas

composition monitoring

• Compliance with one of two options for measuring flare tip velocity

Page 7: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Monitoring System Requirements – 40 CFR 63.671

Owner shall install, operate, calibrate and maintain the Continuous Process Monitoring System

(CPMS)) as . . .

1. … readout for compliance is readily accessible by operator (sic)

2. . . . must complete a minimum of one cycle (sampling, analyzing and data recording) for each

successive 15-minute cycle

3. Except for periods of malfunction, repairs, & QA/QC activities, owner shall operate & collect data

continuously when emissions are routed to flare

4. Operate, maintain & calibrate according to part (b)

5. Must comply with out of control procedures in (c)

6. Reduce data as specified in (d)

7. CPMS must be capable of measuring appropriate parameter over range expected at that location

Page 8: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Section (e) – addn. requirements for GC

For monitors used to determine compositional analysis:

1. Q/A requirements are in table 13 of 63.671

2. Cal Gases must meet one of 2 options:

a) Must include all of Hydrogen, Methane, Ethane, Ethylene, Propane, Propylene,

n-Butane, iso-Butane, Butenes, 1,3-Butadiene, n-Pentane

b) or may calibrate according to surrogate method

3. Owner may choose a compound list based on what is reasonably expected to be

present

4. Thermo Scientific recommends target compound list calibration for mass specs

Page 9: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Table 13Net heating Value by Gas Chromatograph.

• As specified in Performance Specification 9 of 40 CFR part 60, appendix B

• Follow the procedure in Performance Specification 9 of 40 CFR part 60,

appendix B, except that a single daily midlevel calibration check can be used

(rather than triplicate analysis), the multi-point calibration (linearity validation) can

be conducted quarterly (rather than monthly), and the sampling line temperature

must be maintained at a minimum temperature of 60 °C (rather than 120 °C).

Hydrogen Analyzer

• +/- 2 percent over the concentration measured or 0.1 volume percent, whichever is

greater.

• Specify calibration requirements in your site specific CPMS monitoring plan. Calibration

requirements should follow manufacturer’s recommendations at a minimum.

Page 10: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Performance Spec 9

Calibration precision, calibration error, and performance audit tests are conducted to

determine conformance of the CPMS with specifications. Daily calibration and

maintenance requirements are also specified

• 6.2 Sampling System – shall be heat traced with no cold spots.

• 7.1 Calibration Gases. Obtain three concentrations certified within 2%. Performance

test in Meth. 205 shall be repeated quarterly. Concentrations shall be: Low level. 40-

60% of measured concentration Mid level. 90-110% of measured concentration

High level. 140-160% of measured concentration

• 8.3 7-Day Calibration Error Test. At beginning of each 24 hr period, set instr setpts

by conducting a (multipoint) calibration for each compound. Throughout the 24 hr

per., analyze the stack gas. At end of 24 hrs., analyze each of 3 initial cal standards

in triplicate, determine avg instrument response. Each CE for all compounds must

be < 10%. Repeat procedure for a total of 7 consecutive days.

Page 11: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Performance Spec 9 (cont)

8.4 Performance Audit Test Periods. Conduct the performance audit once during the

initial 7-day CE test and quarterly thereafter. Analyze EPA audit gas three times, and

calculate the average instr response (if using a gas mixture, report the certified

concentration of each compound)

10.1 Initial (Multi-point) Calibration. After routine maintenance, or at least once per month

(subst. quarter), conduct a multipoint calibration of the analyzer for each target analyte.

Calibration shall meet requirements in section 13.3 (sample to be analyzed shall flow

continuously through the sample system).

10.2 Daily Calibration. Once every 24 hours, analyze the mid-level standard in triplicate.

Average analyte response cannot vary by more than 10% from the certified value.

Page 12: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

EPA Alternate Measurement Protocol 119

EPA approves substitution of continuous mass spectrometry for continuous gas

chromatography.

Must meet requirements in 40 CFR63.671 (e1 & e2), including Table 13.

May use alternative sampling line temp (60o C)

Must meet applicable Performance Spec 9

• Multi point calibration check at three concentrations

• Periodic mass spec calibrations as directed in Table 13

• May use n-Pentane response for unknowns and compounds with higher molecular

weight than Pentane

. . .”we will announce on EPA’s website that the alternative method is

broadly applicable to determination of NHV under this subpart”

Page 13: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Example 1: Flare Gas Application Summary

Name Formula Molecular Weight Conc %mol precision

(standard

deviation) %mol

Hydrogen H2 2 Bal ≤0.1

Methane CH4 16 10 ≤0.02

Water H2O 18 10 ≤0.05

Carbon Monoxide CO 28 10 ≤0.05

Nitrogen N2 28 1 ≤0.005

Ethane C2H6 30 0.1 ≤0.005

Methanol CH3OH 32 1 ≤0.005

Carbon Dioxide CO2 44 1 ≤0.005

Propane C3H8 44 1 ≤0.005

i- & n- Butane i- & n-C4H10 58 1 ≤0.005

i- & n- Pentane i- & n-C5H12 72 0.5 ≤0.005

NHV 0.3% rel

Page 14: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Why use process mass spec?

• Speed – measurement of all components in under 30 seconds

• Stability – some clients run their mass specs for >1 yr without

needing recalibration

• Excellent linearity - many components exhibit R2 of 1.0

• Flexible application – software based

• Versatility – can provide measurement of multiple sample points with

the same analyzer

Page 15: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Mass spec drawbacks

• Cost

• Low level of industry adoption

• Perceived as “complex”

• Persistent negative stereotypes from >30 years ago

• Lots of calibration cylinders required

• If you perform maintenance, must allow vacuum to be attained

before the analyzer can be returned to service

Page 16: On-Line Flare BTU Monitoring with Process Mass Spec for ... · On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance EPA Alternate Measurement Protocol 119 EPA approves

On-Line Flare BTU Monitoring with Process Mass Spec for RSR compliance

Thank you!!

Questions??