of handwheel controls used in a thesis - tdl
TRANSCRIPT
HUMAN FACTORS IN THE DESIGN AND OPERATION
OF HANDWHEEL CONTROLS USED IN
A DYNAMIC MANUAL TASK
by
LARRY BERNARD JORDAN, B. E. S.
A THESIS
IN
INDUSTRIAL ENGINEERING
Submitted to tha Graduate Faculty of Texas Technological College in Partial Fulfillment of tha Requirements for
The Dagraa of
MASTER OF SCIENCE
IN
INDUSTRIAL ENGINEERING
Approved
May, 1969
hOo. S-\ rs /. TABLE OF CONTENTS
LIST OF TABLES iv
LIST OF ILLUSTRATIONS v
I. INTRODUCTION . . . • c 1
Human Factors Research Areas 3
The Force Platform 9
Purpose and Scope . • . 13
Review of Previous Research l4
II. EXPERIMENTAL PROCEDURE • . 21
Apparatus 21
Performance of the Experiment 32
III. DESIGN OP THE EXPERIMENT 39
Selection of Variables 39
Discussion of Variables Selected . « . . . 46
Design of the Experiment 52
IV. ANALYSIS OF RESULTS. . . . . . . . . . . . . . 56
Analysis of Force Traces 56
Analysis of Variance .57
Analysis of Graphic Data . .61
V. SUMMARY, CONCLUSIONS, AND RECOMMENDATIONS. . . 88
Summary 89
Recommendations for Further Research . .,. 92
LIST OP REFERENCES 9^
ii
ill
TABLE OF CONTENTS Continued
APPENDIX 97
A. Force Platform Calibration Record . . . . . 9^
B. Graph of Subject Main Effect „ 99
LIST OP TABLES
Table Page
1. Optimum Diameter of Handwheels
Various Torques and Positions l6
2. Subject Data 22
3^ Applied Equilibrium Force For Various Torques and Handwheel Sizes ^7
4• EMS for Pour-Factor Experiment in a
Randomized Block Design 5^
5. Summary of Experimental Variables 55
6. Analysis of Variance 60
iv
LIST OP ILLUSTRATIONS
Figure Page
1^ Whitney's Method 12
2, Beckman Dynograph Recorder 23
3. Force Platform 25
H, Force Platform Schematic 26
5^ Torque Device Schematic . . . . 30
6. Torque Device 31
7. Calibration Table 33
8. Torque Device and Force Platform 35
9. Experiment in Progress . 36
10, Handwheels Used 43
11, Hand Positions 45
12, Typical Dynograph Recording 58
13• Relationship Between Resultant Bodily Reaction Force and Handwheel Size 62
l4• Relationship Between Resultant Bodily Reaction Force and Resistant Torque 66
15• Relationship Between Resultant Bodily Reaction Force and Relative Height 68
l6^ Angle of Elbow Flexion for the Average Subject 70
17. Relationship Between Resultant Bodily Reaction Force and Hand Position 72
18. Relationship Between Resultant Bodily Reaction Force and Resistant Torque for Various Handwheel Sizes 76
V
CHAPTER I
INTRODUCTION
Before World War II the human operator was not
given a great deal of attention by designers of his tasks
and equipment. This is understandable when one considers
that equipment during this period was not nearly as elabor
ate as that developed during and after World War II and
the deficiencies caused by this neglect of the human oper
ator were not nearly as great or as noticeable. Designers
became acutely aware of the need for close coordination of
men and machines when the complexity of some types of mod
ern military and civilian equipment began to outstrip the
abilities of the men trained to operate them. The effi
ciency of performance that was theoretically achievable
was never realized due to the limitations of the overlook
ed "weak link", the human operator.
The realization of the importance of the "human
factor" brought a new era in equipment design. It also
spawned a new field of scientific endeavor known as human
engineering (human factors engineering, ergonomics, bio
technology) . This new area "is not a single scientific
discipline but a synthesis which integrates the biological
sclences--psychology, anthropology, physiology, and
madicine--with engineering (5)." Human enginaaring means
"fitting the machine to the man, and keeping him function
ing with efficiency, with safety, and without discomfort
in any environment (l6)."
Before the advent of this interdisciplinary science,
the design of equipment was primarily the responsibility of
the engineer who was concerned mainly with engineering con
siderations. "As often as not this did not matter vary
much because the man who operated the equipment was tha
most efficient part of the man-machine unit (9)." This
neglect of the human factor by design engineers becomes
apparent when one considers tha following definition:
Engineering design is ".,. the process of applying various
techniques and scientific principles for the purpose of
defining a device, a process, or a system in sufficient
detail to permit its physical realization (27)."
Reliable performance within an existing environ
ment was considered a primary basis for acceptance of a
design• Also of concern to the design engineer was socio
economic conditions.
There are long and short-time economic cycles that are related directly and indirectly to natural, social, business, and political factors• Particular designs are formed within limited economic boundaries and associated conditions. In addition to economic influences, there are influences on designs established by various groups of people• (7)
Finally, much serious attention was directed to the
3
selection of materials based mainly upon three primary
considerations:
1. Mechanical functions of materials• 2, Non-mechanical functions of materials
such as appearance, safety, etc^ 3* Economics of material selection•
These, then, were the major factors with which design
engineers were primarily concerned in the past^ Modern
design engineers have recognized the importance of the
human factor and have added it to the list of major fac
tors influencing design.
Fortunately, through research, a considerable
volume of valuable human factors Information has been
generated which the designer may apply to the design of
tools, equipment, and systems involving man as a user or
component. Let us now briefly examine some of the more
Important areas that have been given attention by human
factors researchers•
Human Factors Research Areas
In the early stages of human engineering it was
desired to know the various dimensions of the human which
could be used In the development of improved working con
ditions. Research was done in the area of anthropometry
in order to determine statistical data for the basis of
man-machine relationships. The data thus obtained were
gathered together, and normal curves were developed
representing the sizes of entire individuals as well as
various components of the individuals such as arm lengths,
foot sizes, eye-level positions, leg lengths, etc., and
certain physical abilities of individuals. The human
measurements and the statistical data obtained not only
provide for a fairly good man-machine relationship, but
they also give a certain amount of insight into such
things as physical endurance, motivation, strength of
individual members of the human body, etc. Helpful in
formation in the area of health and safety has also been
provided. (7)
The availability of this information stimulated
research in the areas of workplace design and equipment
layout. Workplaces were designed to accommodate the
varying bodily dimensions of the greatest possible per
centage of the operator population. Researchers attempt
ing to determine the optimum arrangement of controls and
displays found that there was a scarcity of data on the
sensory-motor capabilities of the human being. This need
sparked research into the range, strength, and speed of
human movements, studies of h"uman fatigue and endurance,
and studies of the sensory mechanisms such as vision,
hearing, etc^
It was found that, in most cases, environmental
and psychological as well as physiological or biological
factors were Involved in human performance• Consequently,
much attention was given the areas of motivation, learn
ing, emotional state, etc., as well as the effects of
temperature and himiidlty, altitude, noise, vibration,
illumination, radiation, and other environmental factors
on the human organism. All of this research has essen
tially tha same goal, to maximize man's efficiency in the
performance of a task^
Many researchers felt that this goal could be
achieved by simply minimizing the amount of physical
effort or the energy expenditure of the human operator.
However, before physical effort can be minimized, one
must be able to measure It, Just what is "little physical
effort," and what is "considerable physical effort"? It
seems indeed understandable that so many attempts have
been made to find some measureable dimensions for human
energy expenditure that may permit a better rating and
understanding of this component of industrial work^ (19)
Almost every technique of "technical measurement"
may be classified in one of four categories;
1. Straight output measure
2. Fatigue measurements
3. The mechanical-physical system
4. The physiological system
Straight output measure is the system known as piece work. While it never actually claimed to be a measure of effort or energy expenditure, high piece work performance too often became synonymous with high effort. However, the mere
count of the number of pieces produced by the worker during the day does not tell us very much about energy expenditure. It has no application at all to the results of effort which due to their nature cannot be expressed by time consumption. (19)
Fatigue measurements and the Introduction of
fatigue allowances were the first step to understanding
the complex human contributions to some counted output.
Many industrial engineers thought this was the answer to
their problem^ Such enthusiasm was somewhat hasty since
mainly jobs with a fairly high degree of physical exertion
had been examined and these jobs are rare in modern In
dustry. These early fatigue measurements were made by
counting the number of pieces produced during consecutive
hours of the working day and drawing fatigue cur'ves. (19)
It soon became apparent that under conditions re
quiring less than near-to-maximum physical effort, the form
of such curves was determined more by incentive or motiva
tion than by fatigue. This approach actually tended to
hide the real amount of individual energy expenditure.
Due to its many subjective aspects, fatigue cannot be con
sidered a good measure of expended energy. (19)
The mechanical-physical system makes use of the
basic laws of mechanics. Numerous attempts have been made
to fit the principles to human v/ork However, mechanical
work formulas must be modified before they can be applied
to human tasks. The problem is one of finding some
7
additional factor that makes sense to the "human motor".
As yet, we do not know what this factor might be. (19)
We now come to the category of technical measure-
ment known as the physiological system. Physiological
measurements represent a more direct attack on the
phenomena of energy expenditure in h-uman work.
The CO2 method represented one of the earliest
approaches to physiological measurements. An attempt was
made to obtain reliable knowledge about human energy
expenditures by measuring the amount of oxygen converted
into carbon dioxide. Extremely valuable information
concerning the energy efficiency of certain prototype
jobs was provided by this method, (19)
As more sophisticated measuring devices became
available, additional methods of determining the "physio
logical cost" to the human being in the performance of a
task came into use. Some of the more common and widely
used of these physiological indices of performance are?
1. Metabolic rate 2. Heart rate 3. Blood pressure 4. Blood flow rate 5. Pulmonary ventilation 6. Oxygen consumption 7. Carbon dioxide production 8. Rate of perspiration 9s Skin temperature 10. Electromyography
All of these techniques have certa.ln dis"f""'nrt dis
advantages in that most require the attacl iment of apparatus
8
to the subject, some require precise laboratory analysis,
and others are just inherently difficult to measure.
Some techniques reflect all of these disadvantages.
The acceptance of heart rate as a measure of
physiological cost was based on the assumption that a
faster heart rate is associated with more effort or energy
exerted. Recently, however, it has been found that there
is not a linear relationship between the energy exerted
and heart rate. In one study subjects pedaled a bicycle
ergometer at 30 rpm for several hours, and it was found
that the heart beat rate rises for approximately 30 min
utes after the start of work even though a constant amount
of energy is being exerted. After 30 minutes, the heart
rate was again relatively constant. (21)
The oxygen consumption method, which has been the
most widely accepted approach, requires the attachment of
apparatus to the subject which in itself may well distort
the interaction between the subject and the work situation.
(18)
In the carbon dioxide method the worker is required
to wear a mask which has a physical as well as a psychologi
cal effect on his performance. "Another problem is that the
oxygen 'debt' Incurred has a time lag which makes it diffi
cult to assign a quantitative value to energy used (ex
pended) on a specific part of the task (21)•"
Many of these measurement problems can be overcome
by the use of a fairly recent device for measuring physio
logical cost--the force platform. Since this device has
been selected for use in this particular experiment, let
us examine it in greater detail.
The Force Platform
This device was developed by Ls Lauru in 1953^
improved by Greene and Morris (1959), and further modi
fied by Barany (1961) with the aid of a grant from the
National Science Foundations It is capable of detecting
minute forces in three independent perpendicular planes.
The subject stands on the platform and there is no other
contact between the subject and apparatus.
Movements of the platform are detected by a
pressure-sensing device, such as a strain gauge or
plezioelectrlc crystal, for each of the frontal, vertical,
and lateral planes. Independence of the three axes is
assured through the use of an equilateral triangular
support of the vertical forces and by single point trans
mittal for the lateral and frontal forces (21). Greene
and Morris (1959) demonstrated the fact that resolution
of forces into the three planes of motion is possible
with this type construction (12)s Hudson (1962) examined
the dynamic characteristics of the frequency responses
and found them to be inconsequential (l8)^ He makes the
10
following statements concerning his investigation:
The general conclusion as of now is that the evaluation did not reveal any characteristic which constitutes a serious obstacle to use of (this) apparatus as a primary experimental instrument. The platform.•.was subjected to an extensive series of tests and the force response curve for all axes is linear within very narrow limits. There is no evidence of interaction between axes. The sensitivity exceeds all experimental requirements. In fact the platform appears sensitive enough to detect tha human heart beat of the subject. (l8)
The output of the force platform can be recorded
by either an ink-supplied, pen type stylus on regular re
cording paper or by a resistance type stylus on special
heat sensitive paper. The force trace originates from an
established zero mark and deviates up or down. The dis
tance deviated is directly proportional to the force ex
erted in the specific plane. According to Konz and Day
(21), two analysis techniques can be used:
1, The total area recorded (i.e. energy) can be calculated either by using a planimeter or by feeding the signal into an analog integrating circuit.
2. The maximum height of the trace can be used as an index of the forces exerted by the operator when performing the task. Barany (1963) demonstrated that using the maximum height is sufficient for some tasks. In his experiment he found a high correlation between the area under the force trace and the maximum height of the force trace. Therefore, if the energy exerted is required, the area analysis should be used; if only the maximum force exerted is required, measuring the maximum height of the line is sufficients
11
A third method of analysis was described by Whit
ney (28) in 1958 in an experiment using the force plat
form to determine the strength of the lifting action in
man^ He wanted to determine the average deflection of
the central two-thirds of each record from Its respective
baseline. The central portion was considered more repre
sentative of the subject's performance under the specified
conditions. He measured the area under the central 2 cm
of each record with a planimeter and then divided this
area (in cm ) by 2 cm. This average deflection in cm was
then converted into its equivalent force by means of the
calibration record of the platform obtained at the con
clusion of each series of trials. Figure 1 illustrates
the use of this methods The shaded area A represents the
central portion previously mentioned• The value of c
is obtained from the calibration record of the platform •
The use of the force platform as a measure of
physiological cost was justified when an investigation by
Greene (1957) of the relationship between two widely ac
cepted measures of physiological cost and the force trace
records of the platform indicated that the force traces
compare favorably with the two other measures when the
trace is properly interpreted• In fact, he suggested ttiat
the force platform method is superior to both the oxygen
consumption method and pulse rate method as a measure of
the physiological cost of certain types of dynamic work, (13)
A = 10 cm'
>2 _ Average deflection = 10 cm^ 5 cm
c = 4 lb per cm 2 cm
Average force = 5c = 5(4) — 20 lb
12
Flg^ 1•--Whitney's method
13
Purpose and Scope
The purpose of this experiment is to determine the
affect of certain operational and design parameters on
tha bodily reaction forces developed by an operator of a
handwheel used in a strenuous dynamic task and to deter
mine the optimum values of these parameters, thereby pro
viding information of great use in the design and/or oper
ation of handwheel controls• The specific parameters to
be evaluated are:
1• The diameter of the handwheel 2. The resistant torque that must be
overcome 3^ The mode of operation (i^e^, place
ment of the hands on the wheel)
4, The height of the operation
The results of this study should also enable conclusions
to be drawn concerning optimiim combination of these para
meters for similar tasks involving the application of high
torque.
"By far the greatest amount of work on handwheels
has been to examine the influence of various factors on
their effectiveness as tracking controls." (4)
The measure selected to evaluate handwheels used in
tracking tasks has typically been one such as accuracy or
maximum turning rate, etc. Little attention has been given
to the evaluation of handwheels used for tasks other than
tracking, such as the application of torque in order to
open a valve or move a weight through a distance, and the
14
measure of physiological cost to the operator has been
largely neglected• Also, low values of resistant or
braking torque have been used in most of the previous
experiments•
In this experiment, handwheels will be evaluated
for a task other than tracking in which, for the most
part, high values of resistant torque must be overcome•
Tha output of the force platform will be used as a measure
of the physiological cost to the operator.
Before proceeding further, let us examine some of
the previous research in the areas of handwheel design
and operation•
Review of Previous Research
One of the most extensive investigations of hand-
wheels was made by Davis (6) in a two-part study performed
in 19^9 and 1951. In the first part of the study he intro
duced the following variables: size of the handwheel,
frictional torque applied to the handwheel shaft, and
location of the handwheel in relation to the operator both
as to height and angle of the axis of the handwheel shaft.
The task the subjects were required to perform was the
setting of an indicator to a specific position by turning
the handwheel one complete revolution.
The handwheel diameters evaluated were 3 in^, 6 in.,
8 in^, 10 in., and l6 in^ The values of torque used were
15
0, 20, 40, and 90 inch pounds. The desired torques were
achieved by applying friction loads to the handwheel shaft
by adjusting C-clamps around two oak vise jaws surrounding
the shafts Performance was evaluated at heights of 24,
36, 39, ^0, 42, 48, and 58 inches above the floor, and
with the axis of rotation of the handwheel at 0 degrees
(perpendicular to the sub ject), -f 45 degrees, -45 degrees,
and 90 degrees (parallel to the subject).
Davis found that:
For each control location there is a breaking point or a divergence point at which the relationship between size of device used and performance changes. This divergence point is usually at the 40 Inch-pound torque. Above this point, the larger the control device, the better the performance achieved; below this point, the smaller the size of control device, the better tha results• Above the divergence point, the larger the frictional torque, the larger the difference grows in favor of the larger devices. (6)
Table 1 summarizes the more important results of
Davis' investigation with regard to optimum sizes of hand-
wheels. It must be remembered that these findings are based
on one revolution of the handwheel and that the operator
turned them with one hand by means of a handle on the wheel
rather than by using the rim.
In part two of Davis' study (1951), be Investigated
the same variables of height, angle, torque, size, and type
of device as in the prior experiment, but this time he
required that the operators make the setting in one-half
revolution. He found that:
16
TABLE 1
OPTIMUM DIAMETER OP HANDWHEELS, IN INCHES, FOR VARIOUS TORQUES AND POSITIONS (FROM DAVIS)
Height from
floor, inches
36
36
40
42
Position degrees*
0 (front)
0 (side)
- 45
-f 45
Torque, inch-pound • •
0
3,6,8,
3,6
3,6
3.6
20
10,16
10
10,16
6,10
40
10,16
10
6,10,16
10
'. !• . ' • t
s
90
16
10
10,16
10,16
^Degrees from horizontal of shaft of wheel
17
For small control devices such as 1^ and 3 inch radius wheels...best performance is obtained at minimum torque for all locations. Any increase in torque above zero inch-pound results in a marked reduction in performance. Performance of the intermediate size wheel, 5 in. radius, is only slightly influenced by torque. Optimal performance for all locations comas at 20 to 40 inch-pound-torque. However, performance at these torques is not greatly different from that at the highest or lowest torques. Regardless of the,,.size of the control device or the torque at which it is operated, better performance is obtained at the 40 inch height, minus 45 degrees location than at 36 inch horizontal axis which can be considered as typical. (6)
Halson (19^9) investigated the performance of oper
ators using handwheels in a tracking task and concluded
that: 1, A large-sized wheel Is consistently super
ior to a small-sized one up to 100 rpm when the advantage passes to the latter.
2. Accuracy of tracking increases with speed of turning.
3^ Past turning speeds are superior for prolonged as well as short periods of tracking.
4. The advantage of heavy over light hand-wheels at all tracking speeds is clearly evident. (15)
In this experiment the operators were seated and
the handwheels were either 28,5 or 30 Inches from the
floor and l4 inches to the right of the target. Runs
lasted either 3 or 4 minutes.
The Poxboro Company conducted a number of experi
ments to determine the effects of orientation, inertia,
friction, and diameter of the wheel on tracking accuracy
(l4). The display device consisted of two black pointers
on a white scale, surrounded by a gray matte surface. The
18
bottom pointer served as a stationary zero point. The top
pointer was controlled by a metal wheel, 6 Inches in dia
meter, with a 3 inch wooden handle at a right angle to the
plane of the wheel. Tha wheel center was 28i Inches from
tha floor and 24 inches from the subject.
The following conclusions were drawn from the results
of these experiments:
1. There is no permanent advantage in using either horizontal or vertical position of handwheels for tracking performance.
2. Little effect on accuracy was traceable to the difference in position of the hand-wheel (vertical or oblique),
3. Inertia reduced tracking error materially and had a marked smoothing effect on performance .
4. The difference between the 4J and 9 inch diameter sizes did not make for large differences in accuracy or smootv ness of tracking, except that with friction, the larger wheel was better.
5. Friction is definitely undesirable. As the affective frictional torque at the handwheel increases, performance gets rapidly worse, especially at low speeds. (14)
In an experiment performed by Katchmar (1957), sub
jects cranked handwheels of different radii and loads for
varying periods of time. Seventy-five subjects and three
wheel radii (4, 5, and 7 in.) were used. The loads ranged
from zero to 90 inch-pounds and the cranking time varied
from one to ten minutes. Results Indicated that the loca
tion and orientation of handwheels make little difference
in cranking speed and accuracy as long as they can be oper
ated comfortably. (5)
19
Morgan (25) says:
1. The diameter of the handwheel rim should not exceed 3A "^ 2 inches.
2. For most effective use, handwheel displacement should not exceed plus or minus 6o degrees from the normal (null) position because larger arcs require the hands to shift position on the control.
Corlett (l4) conducted a preliminary experiment in
1961 to examine tha effects of certain factors that he con
sidered relevant to the problem, and to determine which
should be examined in greater detail. The factors consid
ered relevant were:
1. Handwheel diameter
2. Dial diameter
3. Line thickness
4. Spindle friction
5. Working height
Results of the survey experiment Indicated tnat the most promising fields for future experimentation would be tha variation of (a) dial diameter and line thickness, and (b) handwheel diameter and friction on the spindle. The value at which the other listed factors should be held constant were shown to have little effect on the results, (l4)
Lehman (23) reports on a study of a considerably dif
ferent nature. One of the purposes of this study was to
determine tha most suitable location of the steering wheel
of a tractor from an investigation of the energy cost of
steering and the force and speed achieved. The techniques
employed were measurement of oxygen consumption and heart
rata of the operator and of forces required in turning the
20
steering wheel. Lehman summarizes the findings of the
study as follows:
The largest force can be exerted on an almost horizontal steering wheel. On the other hand, the steering wheel may be turned with the greatest velocity if it is nearly vertical. With this vertical position of the wheel, however, the energy consimiption is very high, and is smallest with the steering wheel inclined so that its axis (i.e., the steering column) makes an angle of 50 - 6o degrees with the horizontal. In this range only 70 per cent of the maximum force is achieved but this position of the steering column must be considered the most favorable one, (23)
These findings must be viewed in the light of the
fact that only one (unspecified) wheel size was used and
the effects of varying the heights and frictional torque
were not investigated.
By now it is apparent that additional research con
cerning handwheel controls is warranted and desirable.
Research involving non-tracking applications is especially
needed. It is this apparent need that stimulated research
culminating in the experiment treated in the succeeding
chapters.
CHAPTER II
EXPERIMENTAL PROCEDURE
Subjects
The subjects in the experiment were volunteers from
the male population of the university. Five volunteers
whose schedules suited the experiment were selected. The
subjects selected were of similar build and thus consti
tuted a relatively homogeneous group with respect to
anthropometric measurements. All subjects were checked
to Insure that they had no injuries to hands or arms.
Their personal data relevant to this research appears in
Table 2. The average age as shown by Table 2 was 26.2
years. The standard deviation was less than one year
(0.955) so age can be disregarded as a factor in this
experiment. All subjects were right-handed.
Apparatus
The following equipment was used in the experiment:
1. A Beckman Dynograph Recorder
2. Force platform
3. Torque device
4. Metronome
The Beckman Dynograph Recorder (see Figure 2) is
a highly sensitive oscillograph capable of simultaneously
recording signals in different modes from many sources,
21
22
CVI
<
<
< p tH O
CO
4J -^^ ^ P . O bOC P -H .H M 0'-—• W P
PH 0 4J^-v T S P • M bO C ps -H -H O 0 * - ^ P P CO
o 4.> 0
M ^ - s ^ M . O O PI P P5 tH M P I ^ - -W W
P. ^ 0 o ^ - ^ TS P • M M Pi ps W - H o - P o CO 4^
4:> P ^ - ^ bOP •H M 0 :s
4 J y - s ,
P • bOPi •H •H 0 ^ ^ P
0 bO <
4^ o 0 •
•r-3 O p : ^ : ps CO
O •
OJ ^
o • i n i n
l A ,
.=j-r-{
o »
cn M
CO !>-M
i n .
t ^ VD
I -CM
M
O •
LA ^
O .
!S-i n
i n ,
i n M
o • OJ M
o 1--H
o , M t -
i n CVJ
CVJ
i n .
cn ^
o 0
VD i n
o .
i n M
i n ,
CM M
O 00 M
o •
a VD
i n CM
(Y^
o ,
i n .:^
i n .
00 i n
o VO i H
i n •
cn M
00 i ^ H
o o
M l ^
o CY
.=t
o .
Cvl ^
o ,
^ i n
i n ,
^ M
'" '
o • CM M
CM t>-M
O .
CO VD
^ CM
i n
23
Pig . 2.--Beckman Dynograph Racordar
24
It was used to amplify and record the three outputs of tha
force platform.
The force platform shown in Figure 3 was used to
detect the bodily reaction forces produced in performing
the experimental task. This force platform was construct
ed at Texas Technological College using a design based
on one developed principally by Barany and Whetsel of
Purdue University in conjunction with a National Science
Foundation grant. Their platform is a redesigned version
of the original model constructed by Greene (12), The
new design maintains the geometric properties of the
equilateral triangular placement of cantilever beams and
tha utilization of Linear Variable Differential Trans
formers (LVDT's) in the sensing units but possesses the
added features of being compact, portable, and relatively
inexpensive to reproduce. The device weighs less than
100 pounds, with overall dimensions of 25" x 22" x 5".
Mechanically, the force platform consists of a 3/4
inch thick hexagonal aluminum top-plate and truss section
suspended vertically by six horizontal cantilever beams
and restricted horizontally by six vertical cantilever
beams. In turn, the twelve beams plus three sensing units
are supported by an aluminum base plate. The schematic
diagram In Figure 4 illustrates the arrangement of the
beams and sensing units.
Forces VI, V2, and V3 in Figure 4 represent the
25
Pig. 3.—Force platform
26
^ i g . 4 . _ orc
FRONTAL
' Platform schemati.
27
weight of the suspended portion of the platform plus any
other downward vertical force that might be exerted on tha
instrument. These three forces are exerted at three points
which form the vertices of an equilateral triangle. Forces
V4, V5, and V6 represent the upward vertical forces that
might be exerted on the platform during transportation or
large displacamants. The three upward forces form the
vertices of a second equilateral triangle superimposed on
the triangle formed by the downward forces. The common
intersection of the perpendicular bisectors of the sides
of the triangle formed by the upward forces is on the same
vertical axis as the intersections of the perpendicular
bisectors of the sides of the triangle formed by the down
ward forces. This arrangement not only restricts the plat
form in both vertical directions, but, a constant force
exerted in either direction anywhere on the platform will
result in the same amount of deflection at the center point
of the hexagonal top-plate•
Forces Fl, P2, P3, and F4 represent the frontal
forces and forces LI and L2 represent the lateral forces
exerted on the platform. Horizontal movement of the sus
pended portion of the platform is restricted by the two
sets of vertical beams shown in Figure 4,
The restrictive forces in the vertical and horizontal
directions are exerted at point contacts consisting of steal
balls pushing against flat tool-steel surfaces attached to
28
the cantilever beams. Therefore, only forces perpendicular
to the restricting beams will cause beam deflection.
Deflection in the three orthogonal directions is
detected by shielded Linear Variable Differential Trans
formers (LVDT's). These devices consist of three adjacent
colls wound on the same insulated spool. A magnetic core
is initially located in the center of the middle coll.
Linear deflection of the core along the axis of the colls
causes increased coupling of the center coil with one of
tha end coils while linear deflection in the opposite
direction causes Increased coupling of the other end coil.
This voltage differential in the end coils is used as an
input signal to the Dynograph Recorder.
The three sensing units SI, S2, and S3 are located
as shown in Figure 4, The vertical deflection sensing
unit SI is placed at the common center of the two equi
lateral triangles. An adjustable set screw through the
center brace of the truss section is used as the sensing
face. The frontal sensing unit S2 is placed along the
frontal axis passing through the center of the platform,
A vertical plate perpendicular to the frontal axis is used
as the sensing face. Only deflections parallel to the
frontal axis will displace the core. The lateral sensing
unit S3 is placed along the transversal axis which passes
through the center of the platform• A sensing face simi
lar to the frontal one is used so that only deflections
29
parallel to the lateral axis will displace tha core in the
lateral sensing unit.
A locally manufactured device was used for producing
the desired values of resistant torque (20)• Figure 5 shows
a sketch of the device• It is essentially a double block
brake. The blocks are made of wood with leather bearing
surfaces; the drum is made of metal and is fixed to a
central shaft. When a weight is suspended from the cable
the two wooden blocks are forced against the drum causing
a constant braking torque to be applied to the central
shaft. The device was clamped to a surface which could be
raised or lowered to the desired height. A picture of the
device is shown in Figure 6.
A metronome was used to pace the experimental task.
Pacing the task Insured that, for any given trial, all
subjects turned the handwheel at relatively the same angu
lar velocity.
Task
The task that the subjects were required to perform
was a simple, dynamic manual task such as that performed in
opening a valve or loosening a vise. In this type of task
high initial resistant torque must be overcome during a par
tial revolution of the handwheel. The subject, while stand
ing on the force platform, grasped the handwheel by the rim
with both hands and turned it counterclockwise through an
30
Pulley- Upper -1 Arm
Drum Shaft
Lower Arm
Upper Block
Lower Block
Pig* 5.—Torque device schematic
31
P ig . 6.--Torque device
32
arc of 90 degrees. This value was arbitrarily chosen but
is appropriate for the task and does not exceed the maxi
mum displacement of 120 degrees recommended by McCormick
( 2 2 ) .
Performance of the Experiment
The entire experiment was performed in an air con
ditioned research room where interruptions could be held
to a minimum. The experimenter and recorder were placed
at 90 degrees to the subject to minimize distractions.
Before beginning the experiment, the recorder was
turned on and the torque device was calibrated by means
of a torque wrench which was turned in time to the metro
nome. The metronome was set at the same setting as that
at which the experimental task was paced during the actual
experiment. This insured a relatively constant resistant
torque for this particular speed of rotation, which was
used throughout the experiment.
After the recorder had warmed up sufficiently, the
force platform was then calibrated. Calibration of the
platform for vertical forces was accomplished by placing
standard weights on the platform and noting changes in pen
deflection as a function of Increased load, A calibration
table utilizing pulleys was necessary in checking the meas
urement of horizontal forces. This device is shown in
Figure 7. Calibration records for the force platform
33
Fig. 7.--Calibration table
34
appear in Appendix A^
After calibrating the platform, the subject was
brought into the experimental room and given a thorough
briefing on the nature and conduct of the experiment.
After this had been done, the subject was directed to
stand at the center of the platform with his heels six
inches apart and feet aligned with lines on the platform.
This standard standing position insured that the verti
cal force due to the subject's weight passed through the
center of rotation of the platform. The subject stood in
a relaxed position on the platform with his arms hanging
naturally at his sides.
The specified diameter handwheel for the trial was
mounted on the shaft of the torque device which was affixed
to an adjustable stand resting on the floor. The stand was
positioned so that a fixed distance of 15-3/^ Inches as
measured from the acromion process at the shoulder of the
subject to a vertical plane passing through the midpoint
of the longitudinal axis of the handwheel hub was maintain
ed. The layout of the equipment and the experimental work
place are shown in Figures 8 and 9.
The torque and height were then adjusted to those
values required for the particular trials The recorder was
turned on for a few seconds to record the normal weight
reading for the subject and the recording pens were adjust
ed to a null position representing a zero activity levels
35
Flg^ 8.--Torque device and force platform
36
Pig. 9.—Experiment In progress
37
With the recorder off, the subject was told to
grasp the handwheel by the rim with both hands, placing
his hands in the positions for that trials The metronome
was turned on and the subject was given practice in rotat
ing the wheel through 90 degrees, regrasping in the same
position, rotating again through 90 degrees, etc, in time
with the metronome. When the experimenter was satisfied
that the subject was sufficiently trained and the hand
placements were correct for the trial, the recorder was
turned on and the pens re-zeroed, if necessary. The sub
ject was then told to grasp the wheel, turn it counter
clockwise through the prescribed arc, regrasp, and turn
again In time with the metronome until the wheel had been
turned five times. The metronome was set at 40 beats par
minute throughout the experim.ent. Upon completion of the
required number of turns, the subject returned to the
"relaxed position" where his null reading was again re
corded. The recorder was then turned off and preparations
made for the next trial.
The null or zero line established bench marks both
before and after the task which reduced the problem of any
electronic drift of the equipment as well as other mis
cellaneous errors. The subject standing with his arms at
his sides was used as a baseline for this experiment as the
output may then be Interpreted as the total force required
for the complete task (above that required to stand in
38
place).
Each subject performed ninety-six trials repre
senting the ninety-six possible combinations of height,
torque, etc. After every five trials, the subject was
given a sufficient amount of time to rest in order to
minimize the chance of fatigue affecting the results.
The time-order of performance for each trial was deter
mined by a table of random numbers.
At the end of each series of trials the subject
was asked which hand position or positions he preferred
at each of the heights. Also at this time the calibra
tion of both the force platform and torque device was
checked again. If the readings at the beginning and end
had been significantly different, the series would have
been rerun.
CHAPTER III
DESIGN OF THE EXPERIMENT
Selection of Variables
The variables considered relevant in an experiment
of this type fall Into two categories:
1^ Variables that are a function of the differences between subjects
2. Variables that are a function of the control design and/or workplace layout
The first category includes:
a. Subject height b. Subject weight c. Shoulder breadth and height d. Arm length a. Lateral location of the control with
respect to the subject f. Fore-and-aft location of tha control g. Maximum strength of the operator h. Susceptibility to fatigue 1, Mode of operation (hand positions) j• Spaed of rotation
The second category Includes such factors as:
a. Diameter of the handwheel control b. The resistant torque that must ba over
come 0, Vertical location d^ Control orientation a• Direction of movement f, Amount of movement
The affects of the first four variables in the
first category were controlled so that their influence was
negligible by selecting a relatively homogeneous sample of
subjects and using subjects as blocks in a randomized
block design.
39
40
Lateral location was fixed. The control was lo
cated directly in front of the operator's midline. This
is the recommanded location for positioning movements in
volving both hands as given by Briggs (5).
For rotary movements there is little difference in
the forces exertable at most fore-and-aft locations,
though the farthest points should be avoided (5). This
being the case, a fixed distance of 15-3/^ inches as
measured from the acromion process at the shoulder to a
vertical plane passing through the center of the longi
tudinal axis of the handwheel hub was used since this
would accommodate very close to 100 per cent of the popu
lation (24,8 inches will accommodate all but one per cent)^
The maximum strength of the operator was not a
factor in this experiment. Each subject was tested be
fore experimentation using the most disadvantageous com
bination of resistant torque and handwheel size to insure
that the force required to turn the handwheel did not ex
ceed the maximum strength of any subject. Resistant torque
did not exceed 120 inch-pounds and handwheel diameter was
no smaller than 7 inches. For this combination the com
bined force that must be exerted by both hands is less
than 35 pounds.
Subjects were given frequent rest periods of ade
quate length in order to eliminate fatigue as a source of
variation. Although some subjects are more susceptible
41
to fatigue than others, by closely observing the force
traces the experimenter can recognize its presence since
the recording st:jrlus will begin to make sudden rapid de
viations from the zero activity line while the subject is
in a relaxed position on tha platform. The subject will
be unable to maintain the stylus on the zero line due to
loss of muscular control, especially over those muscles
controlling posture and balance. This phenomenom has
been observed many times by this experimenter In prior
work with the force platform. Had it been observed in
this experiment, the subject would have been given an
unscheduled rest period.
Speed of rotation of the handwheel was maintained
relatively constant for all subjects for any given trial.
This was accomplished by having the subjects perform the
task in time with a metronome. Subjects were given ade
quate training in timing their movements before performing
each trial.
The last three factors in the second category ware
also controlled. One fixed control orientation was used
in this experiment. The handwheel was oriented so that
the axis of rotation was perpendicular to the frontal
longitudinal plane of the subject's body^ This was the
only orientation feasible due to the limitations of the
torque device.
Direction of movement was fixed. The handwheel was
42
rotated in a counterclockwise direction which is the most
realistic for the task chosen.
The handwheel was rotated through an arc of 90 de
grees. This value does not exceed the maximum recommended
value of 120 degrees and, again, is realistic enough for
the particular t^sk that was performed.
The four remaining independent variables are the
ones that were chosen for consideration in this experiment.
They are:
1, Handwheel diameter
2, Vertical location
3, Resistant torque
4, Mode of operation (hand position)
Handwheel diameters of 7, 10, and l4 inches were
selected (see Figure 10). These are sizes frequently
encountered in industry and the size Increments are such
that if there is a breaking point or divergence point on
the torque scale at which the relationship between size
of device and performance changes, the probability of de
tecting this point is relatively high.
Two relative heights were used. One was six inches
above the elbow height of the subject and the other six
inches below• The lower height would correspond to a con
trol height of 36 Inches for an average subject. Hand-
wheels used in industrial applications are frequently found
at this height. The second height was different enough from
43
Pig^ 10.--Handwheels used
44
the first that it enabled the effect of vertical location
to be more readily determined. Having these particular
heights enabled a determination to be made as to whether
location above or below the elbow is more advantageous.
Resistant torques of 20, 40, 80, and 120 inch-
pounds were applied to the handwheel shaft. Here again,
the increments were large enough that they would have re
flected any significant effect produced by varying the
torque. One of the reasons why high torque values were
selected is that they are representative of the task being
evaluated in this experiment. Also, these high torque
values required the application of forces of large enough
magnitude such that they did not demand that the sensitiv
ity of the recorder be as high nor as critical as in tasks
where smaller forces are applied. This, in turn, promoted
greater reliability since a small error in measurement
would not have appreciably affected the resultant force;
whereas, in the case of small applied forces, a small er
ror would have precipitated a completely erroneous inter
pretation of the force trace•
The subject grasped the rim of the handwheel in four
different ways^ Consideration of hand placements as a legi
timate variable was justified by a trial run which indicated
that this factor does Indeed affect the force platform out
put. Tha various hand placements can best be explained in
conjunction with Figure 11, They are:
45
0 Mode 1:0= 60°
R ® Mode II: e = 60°
R
Mode III: 0 = 120° Mode IV: 9 = 180°
Fig^ 11•--Hand positions
46
(a) Mode I—Left hand at ten o»clock position and right hand at twelve o'clock position.
(b) Mode II--Left hand at twelve o'clock position and'right hand at two o'clock position,
(c) Mode III--Left hand at ten o'clock position and right hand at two o'clock position.
(d) Mode IV--Left hand at nine o'clock position and right hand at three o'clock position.
Each of the ninety-six possible combinations of
the four variables selected was run in a random order for
each subject. In every case the dependent variable was
the force platform output.
Discussion of Variables Selected ^ ' I I I ' I I Ill » * B l l l ^ l H f • • I I P . • • • ^ 1 W m I I I . ' W ^ . 1 1
It was anticipated that the effect of increasing
the handwheel diameter would be predictable up to a pointy
Increasing the diameter increases the moment arm of the
force that must be applied to overcome the resistant tor
que. For the same value of resistant torque. Increasing
the diameter decreases the amount of applied force requir
ed. In general, the larger the frictional torque, the
larger the handwheel size would have to be in order to
lessen the physiological cost to the operator.
Table 3 clearly illustrates these facts. Of course,
there is a limit to how large the handwheel can be before
it exceeds the capabilities of the operator and/or becomes
TABLE 3
47
APPLIED EQUILIBRIUM FORCE IN POUNDS FOR VARIOUS TORQUES
AND HANDWHEEL SIZES .-- , .' 1 -
Torque in
inch-pounds
20
40
80
120
7
5.71
11,43
22.86
34.29
Handwha ( Diamate (inches
10
4.00
8,00
16.00
24.00
el r )
14
2.86
5.71
11.43
17.14
(Forces Just greater than those listed will cause movement of the handwheel)
48
more costly to the operator. If the handwheel diameter
becomes too large, a greater number of muscle groups and
of different type than at the smaller diameters are brought
into play. The sizable moment arms created by very large
handwheels require the use of the large muscles of the
back and trunk. This tends to rotate the entire body and
to greatly displace its normal center of gravity, creat
ing a condition of Imbalance. Great reactive forces in
consistent with the task evaluated here would then have to
be applied by certain muscles of the body in order to
maintain balance. To avoid this condition and the possi
bility of introducing additional variables, handwheel
sizes large than l4 inches were not considered. This would
Insure that major muscle activity would ba confined to rela
tively the same muscle groups, primarily those of the arms
and shoulders.
Increasing the diameter and, in effect, the moment
arm of the applied force, should Increase the bodily reac
tion forces to some extent for those handwheels that are
not inordinately large. In particular, an Increase in the
transverse forces should be apparent. The question of
primary Interest here is, is there a divergence point above
which the advantage lies with the larger devices and below
which the smaller devices are more advantageous? Davis (6)
found such a divergence point which he says is usually at
the 40 inch-pound torque. However, the task which he
49
evaluated was considerably different from the task eval
uated here. There is no guarantee that such a point exists
or, if it exists, can be readily determined for a task in
which accuracy is not a factor and is not the measure of
performance being evaluated.
Increasing the resistant torque will, for the same
diameter wheel, require that the applied force be Increased,
In general, the lowest value of frictional torque will pro
vide the least physiological cost to the operator. What
is Important here is not the resistant torque main effect,
per se, but the interaction of resistant torque with the
other main effects. Of particular interest is the inter
action with handwheel size. This Interaction should pin
point the divergence point previously discussed if such a
point exists. Although the resistant torque was not of
primary interest, it was expected that it would be highly
significant.
A number of studies have been performed in an at
tempt to determine optimimi heights for certain tasks.
However, the majority of these studies were concerned with
horizontal work surfaces. Ellis (8) performed an experi
ment involving a manipulative task by persons working in a
standing posture and found a level around 3 inches below
the elbow to be optimum. This was, on the average, about
42 inches above the floor. From this study and from the
experience reported by Barnes (3), there is a substantial
50
basis for arranging work surface height for standing work
ers at a level somewhat below elbow-level height (about 2
to 4 inches), at least for light assembly work or similar
manipulative tasks. Where movements involving consider
able exertion are required, somewhat lower levels probably
would be in order (24). These findings were not expected
to be necessarily applicable in this experiment involving
a vertical work surface. It should also be noted that the
effect of height is closely related to the arm length and
shoulder height of the subject. For any particular sub
ject, fixing the fore-and-aft distance and the control
height will completely determine the elbow position (height)
and angle of flexion. The optimum height was expected to
be the one for which the angle of flexion of the elbow was
closer to 90 degrees. The value of 90 degrees was sub
stantiated in an experiment by Provins and Salter as being
the optimum angle of elbow flexion for a task requiring a
relatively high degree of exertion (24).
The effect of the mode of operation (hand position)
is not easily predicted. Static blomechanical analysis
becomes very difficult in short order and is unrealistic
at best. Dynamic analysis, which is more realistic, pre
sents the problems of measuring such quantities as angular
displacement, time, length, weight, and weight distribu
tion. A prohibitively large number of kinematic quantities
and physical constants must be determined by measurement.
51
Also, very sophisticated techniques of measurement are
required. However, some hypotheses concerning mode of
operation were made based on a little logic, experience,
and intuition. This experimenter felt that the more
balanced hand placements would be less costly from a
physiological standpoint. These positions enable forces
to be applied more smoothly and with greater control.
Positions where both hands are on the same side of the
null position would cause an unbalanced force to be ap
plied which would tend to cause a shift in body weight.
A corresponding reactive force would be applied by cer
tain muscles of the body in an effort to maintain balance,
As the angle between the hands in these unbalanced posi
tions becomes smaller, the bodily reaction forces that
must be applied become larger•
The preceding comments may be summarized in the
following Initial hypotheses that were made prior to ex
perimentation:
1. Larger handwheel sizes are more advantageous at higher torque values requiring large applied force.
2. Lower torque values are less costly,
3. Vertical locations for which the elbow angle is close to 90 degrees are better for handwheel controls•
4. More balanced hand positions are less costly.
The first two hypotheses were dictated by common
52
sense and are readily apparent. The latter two may not
be so apparent and were based on preliminary analysis and
experience.
Design of the Experiment
A four by four by three by two factorial experi
ment In a randomized block design was used. Each subject
was treated as a block with a complete factorial experi
ment randomized within each block.
The mathematical model of the experiment is:
Where:
%1klm ~ Resultant bodily reaction force
u = A common effect in all observations
SjL = Subjects (blocks)
D^ = Handwheel diameters
Hj^ = Heights
T-j_ = Resistant torque
Pj^ = Hand positions
eijklm = Random error in the experiment
1 = 1, 2, 3, 4, 5 1 = 1 , 2, 3, 4
j = 1, 2, 3 m - 1, 2, 3, 4
k = 1, 2
53
The expected mean squares (EMS) and degrees of
freedom (df) are given in Table 4, These EMS values in
dicate that all main effects and interactions can be
tested against the 380 df error term. All of the inde
pendent variables considered In the experiment are sum
marized in Table 5,
TABLE 4
EMS FOR FOUR-FACTOR EXPERIMENT IN A RANDOMIZED BLOCK DESIGN
54
1
Source
h ^i
" k
T l
Pm
^'^Jk
""hi
^P jm
HTkl
I^^km
TPim
DHTjkl
jkm
^•^^klm
^ ^ ^ ^ j k l m
® i j k l m
T o t a l
d f
4
2
1
3
3
2
6
6
3
3
9
6
6
18
9
18
380
479
5 R 1
1
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
1
3 P J
3
0
3
3
3
0
0
0
3
3
3
0
0
0
3
0
1
2 F k
2
2
0
2
2
0
2
2
0
0
2
0
0
2
0
0
1
4 F 1
4
4
4
0
4
4
0
4
0
4
0
0
4
0
0
0
1
4 F m
4
4
4
4
4
4
4
0
4
0
0
4
0
0
0
0
1
2 (Te
<rl <r'i
2 ^ e
<TI
^l 2
^ e
<rl <rl <rl <rl <
<
<rl 2
CTe
<^l •G^
8
EMS
- 2 4 96(^3
+ i6o(r^
+ 24oo | 2
+ 12aTrp
4 120(Tp
4 8oac,H . ^2
4 40(Tj).p
4 60Jj.jrp
^ ^2 4 60^Hp
4 300"^p
2 4 20(Tj)j^ip
^ 20<JDHP
2 4 100-j^ipp
^2 4 1 5 0 H T P
2 + 5^DHTP
55
TABLE 5
SUMMARY OF EXPERIMENTAL VARIABLES
Factor
Subject
Handwheel Diameter
Resistant Torque
Height
Hand Position
Level
1 2 3 4 5
rrJt
10" 14"
20 in. 40 in, 80 in. 120 in.
6" bel
-lb -lb -lb -lb
ow e 6" above e
60" left 60"right 120" 180*
Ibow Ibow
Code
1 2
5
1 2 3
1 2 3 4
1 2
1 2 3 4
Type
Random
Fixed
Fixed
Fixed
Fixed
"-.- ••!'
Identification
S
D
T
H
P
CHAPTER IV
ANALYSIS OP RESULTS
Analysis of Force Traces
As mentioned previously, there are three primary
techniques popularly used in analyzing recorded force
platform output. Two of the techniques require measure
ment of the area under the force-time curve which has
stimulated some debate over just what this area repre
sents and the validity of using it as an index of work,
energy expended, or fatigue. One may use the third method,
which does not require measurement of this area, if the
task involved is not complex and is of short duration.
In this method the maximum height of the trace is used
as an index of the forces exerted by the operator in per
forming the task.
Since the previously mentioned requirements were
met by the task considered here, and since it was felt
that force would be a simple yet meaningful measure, a
variation of the third method was selected for use in
this experiment. The resultant of the average maximum
forces over five cycles (partial revolutions) produced
for the frontal, lateral, and vertical axes was used as
a single measure of operator performance. This resultant
average maximum force for a particular trial was computed
56
57
using values obtained from the force trace for that trial
produced by the Dynograph Recorder. A typical recording
is shown in Figure 12.
Determination of the resultant force for each
trial Involved five steps: (a) noting the maximum distance
either above or below the zero or static weight line for
each cycle for the three axes (see Figure 12), (b) adding
the five values to obtain the total maximum deflection
over five cycles for each axis, (c) converting these totals
to total maximum force for each axis by multiplying by a
constant obtained from the force platform calibration re
cord (see Appendix A), (d) dividing the resulting totals
by five to obtain the average maximum force for each axis,
and (e) computing the resultant force by taking the square
root of the sum of squares of the average maximum force of
each axis.
The data thus obtained were punched on cards and fed
into a digital computer which performed an analysis of
variance.
Analysis of Variance
The analysis of variance was performed using com
puter program BMD02V--Analysis of Variance for Factorial
Design—prepared by the Health Sciences Computing Facil
ity, UCLA, and modified for the Texas Technological College
computer center. The resulting analysis of variance was
58
^^-^T^^tt^^ •Q
0 0
P - Frontal axis L - Lateral or transverse axis V " Vertical a cis
Pig, 12.--Typical Dynograph recording
59
then adjusted to reflect the desired design for this
experiment--a factorial experiment in a randomized block
design. The analysis of variance for this design appears
in Table 6, Only the main effects and interactions that
were significant at the one and five per cent levels were
considered for analysis and discussion. Note that all
main effects and all first order interactions with the
exception of the height by hand position interaction were
found to be significant at these levels• Although the
variation between subjects was found to be significant
(as might be expected), this effect will not be discussed
in great detail as it is not of major concern here and adds
little of significant value to the discussion. A graph of
this main effect is presented in Appendix B, It shows that
subject number four had a considerably higher mean result
ant bodily reaction force than the other subjects. Subject
4 was the eldest of the group and had consistently larger
anthropometric measurements (see Table 2, Chapter II),
The subject interactions showed relatively little differ
ence in subject reactions to changes in design parameters
and Indicated the same basic trends for each subject. Con
sequently, subject interactions will not be discussed here.
The means of the remaining main effects and interactions,
however, were plotted to facilitate subsequent analysis
and will be discussed in detail.
TABLE 6
ANALYSIS OF VARIANCE
60
Source of
Variation
Subjects Handwheel diameter Torque
Height Hand position Handwheel dla, X torque Handwheel dla, X height
Handwheel dia, X hand position Torque x height
Torque x hand position
Height X hand position
Handwheel dia, x torque x height
Handwheel dla, x torque x hand pos.
Handwheel dia, x height X hand pos. Torque x height x hand position Handwheel dia. x torque x height x hand position
Residual (error)
Total
Degrees of
Freedom
4 2 3 1 3
6
2
6
3
9
3
6
18
6
9
18
380
479
^ — . - • . —
Sum of Squares
3806.09 7793.85 21432.42
1006.98 21878.01
1672.14
295.96
449.75
329.90
4592,92
134.00
171.70
602.81
163.90
208,15
225023
8241.44
73005.25
Mean Square
951.52 3896.92 7144,14
1006.98 7292,67
278.69
147.98
74,96
109.97
510.32
44.67
28.62
33.48
27.32
23.13
12.51
21.69
F
43.87* 179.66* 329.37* 46.43* 336.22*
12.85*
6.82*
3.46*
5.07*
25.53*
2.06
1.32
1.54
1.26
1.07
0.58
*Slgnificant at .01 level
61
Analysis of Graphic Data
In the remainder of this chapter we shall analyze
and discuss the significant effects graphically presented,
Handwheel Diameter Effect (Figure 13)
As is apparent from the graph, increasing the size
of the handwheel control decreased the amount of bodily
force produced in performing the task. The smallest hand-
wheel required the most force and the largest handwheel
required the least. This result was expected and is not
particularly surprising. Had this result not been obtain
ed, however, the validity of the experiment would have
been suspect.
A Duncan Multiple Range Test as described in Hicks
(17) was performed on the three means. The following sym
bols will be used in performing the test:
MSg = Error mean square (from Table 6)
k = number of means
p — '^9 r> > ' • • i k
N = number of observations in the mean
n = degrees of freedom of error term (Table 6)
s = standard error of a mean = YMS^/N
R = significant range from Appendix Table
E of Hicks (17)
LSR = least significant range = Rs
62
5 0 "
40 -
ra -CJ C ps o PL,
0 o u o p
30 "
20 •-
10 -•
0 1 1 h-7 10 14
Handwheel Diameter (in^)
Fig. 13.—Relationship between the resultant bodily reaction force and handwheel size
63
The procedure used in performing the test is shown
below in some detail in this instance, but hereafter only
the final results will be presented. For handwheel diam
eters of 7, 10, and l4 inches the means were 30,00, 23.50,
and 20^51 respectively.
s = MS E 21.69 = 0^368 \ N 1 l6o
P = _2 3_
R = 3.64 3.80 (at one per cent level)
LSR = 1.34 1,40
Arranging the means in ascending order:
3 2 1 20.51 23.50 30.60
Range 1 - 3 - 9 . 4 9
Range 1 - 2 rr 6.50
Range 2 - 3 = 2.99
Comparing observed ranges between means with the
least significant ranges (LSR), we find:
1 versus 3 - 9.49 > 1.40 1 versus 2 = 6.50 > 1.34 2 versus 3 = 2^99 > 1.34
Hence, all means are significantly different•
They may be depicted as follows:
Handwheel diameter 7" 10" l4"
Mean 30.00 23.50 20^51
Hereafter, any means not underscored by the same line are
significantly different and any means underscored by the
64
same line are not significantly different.
It can be concluded from the previous discussion
that a significantly different amount of force is pro
duced in the operation of each size handwheel and that
the 14 inch diameter handwheel is the optimum size of
those sizes tested since the bodily reaction force was
lowest for this size. The low bodily reaction force for
this diameter reflects the low applied force required.
Increasing the diameter increases the moment arm of the
applied force, thereby decreasing the applied force since
this moment arm and applied force constitute a torque in
opposition to any resistant or braking torque.
Note, however, that there is not as much advantage
gained in going from the 10 inch to the l4 inch diameter
as there is in going from the 7 inch to the 10 inch diam
eter. This reduction in rate of gain is due to that dis
advantageous aspect of increasing handwneel size previously
mentioned. As the handwheel diameter is Increased from 10
to l4 inches, the center of gravity of the body is displac
ed enough to create a slightly unbalanced condition. This
slight imbalance brought about by the increased moment arm
causes the body to apply counteracting reactive force in
order to maintain balance. This lessens, somewhat, the
advantage in going to larger size handwheels. As handwheel
diameters continue to increase in size, these reactive
forces Increase also and the advantage gained by increasing
65
the handwheel size becomes less and less.
Resistant Torque Effect (Figure l4)
The relationship between resistant torque and the
bodily reaction force produced in overcoming this resist
ant torque is clear-cut. The greater the torque that
must be overcome, the greater the bodily reaction force
produced. Note the almost linear relationship depicted
in Figure 14. The results of the Duncan Multiple Range
Test are given below:
N :. 120
s - 0o424
P = 2 3 4
R ^ 3o64 3.80 3.90
LSR = 1.54 1.61 1.65
Torque (inch-pounds) 20 40 80 120
Mean l6,62 20.22 27.08 34,76 s s — i ^ — M H B i V S M ^ K i V ^ O — P ^ H M W O H W a * ^ . ! ' ^ O W ^ M B B ^ B B S J H K V ^ O a ^ V s ^ n E I H ^ a a i i * ^ ^
All means are significantly different at the one
per cent level. Thus, the optimum value of torque for
those tested is the smallest value (i,e,, 20 inch-pounds).
This is a reasonable result since the bodily reaction
force is a function of the applied force required to over
come any particular value of resistant torque. As the
resistant torque increases in value, the applied force at
the handwheel rim must Increase correspondingly if the
resistant torque is to be overcome. Increases in the amount
66
5 0 -
4 0 -
ra Td
o OH
3 0 -
0 O PH O P
20..
10"-
0 -H 1 1 1 1 1— 20 40 60 80 100 120
Torque ( i n , - l b . )
Fig. l4.--Relationship between the resultant bodily reaction force and resistant torque
67
of applied force are reflected as increases in the com
puted force platform output or, in other words, the re
sultant bodily reaction force since an applied force
effects the transmittal of a related reactive force along
the main body members to the force platform. The final
reactive force sensed by the force platform will not
necessarily be of the same magnitude as the original ap
plied force required to overcome a particular value of
resistant torque but there probably should be a linear
relationship between the two as implied by the graph in
Figure l4.
Relative Height Effect (Figure 15)
The graph in Figure I5 indicates that the bodily
force produced decreases as the relative height is in
creased from six inches below the elbow to six inches
above the elbow. To determine if the force is signifi
cantly different for these two relative heights, a Duncan
Multiple Range Test was performed. The results are given
below:
N = 240
s = 0.301
P = 2
R =3.64
LSR = 1,096
Height 6" below elbow 6" above elbow
Mean 26.84 22.50
68
5 0 - -
4o
ra Td PI ps o P.4
3 0 ••
0 o Pl o
20 ••
1 0 •"
1 6"
below elbow
— I 6"
above elbow
Relative Height
Fig. 15,--Relationship between the resultant bodily reaction force and relative height
69
Thus, produced force at the two relative heights
are significantly different at the one per cent level.
This result in conjunction with the result of the analysis
of the graph In Figure 15 indicates that a relative height
of six inches above the elbow is the desired vertical hand-
wheel location for the task evaluated in this experiment.
However, this experimenter suspected that the basic con
tributing factor here was not height, in itself, but the
angle of elbow flexion generated at each of the heights.
To check this hypothesis, the average anthropometric
measurements of the subject group (see Table 2, Chapter
II) was used to determine the average angle of elbow flex
ion at each of the relative heights. From Figure l6 we
see that an angle of 75 degrees was generated for a rela
tive height of six inches above the elbow. McCormick (24)
gives the results of a study by Provins and Salter which
showed that the optimum elbow angle is about 90 degrees
for a task requiring application of relatively large
amounts of force. The elbow angle at a relative height of
six Inches above the elbow is much nearer to this value
than that at the lower height.
Another important point is that at the lower height
a rotation action localized primarily at the elbow is gen
erated, whereas at the greater height a rotation action of
the shoulder is generated. McCormick (24) also mentions
another study in which Provins was Involved which concluded
T Shoulder height
Elbow height
56,1'
43.5"
Angle A — 75 degrees
Angle B =: 132 degrees
70
Pig^ l6,--Angle of elbow flexion for the average subject
71
that "a rotation action of the shoulder has approximately
li times the force of a rotation action of the elbow, and
it has nearly 3 times as much staying power (the ability
to maintain a force)•" Thus, at the greater height more
force can be supplied as the demand for more applied
force increases before the aid of additional muscle groups
is required. Bringing additional muscle groups into play
is a more costly and less effective means of providing the
required force than where the force requirements can be
met by one muscle group without undue strain.
Finally, the fact that there is such a great differ
ence in magnitude between the two angles probably accounts
for the significantly different amounts of reactive force
produced at the two heights.
Hand Position Effect (Figure 17)
As mentioned earlier in Chapter III, there were
four different sets of hand positions considered in this
experiment (see Figure ll). Two were rather unbalanced in
that there was an arc of only 60 degrees separating the
right and left hands. For one set, the right hand was
placed at top dead center of the handwheel and the left
hand to the left; for the other, the left hand was placed
at top dead center and the right hand to the right. These
sets are denoted as 60 degrees left (60°L) and 60 degrees
right (60°R) respectively In Figure 17• The other two
sets of hand positions are more balanced, with 120 and 180
72
50-
ra Td
o PL,
40-
30
ra 0 o Pl o &H
20"
10
0
60^L 60°R 120 180°
Hand Position
Fig. 17.--Relationship between the resultant bodily reaction force and hand position
73
degree arcs separating the right and left hands. The graph
in Figure 17 shows that as the hand positions are changed
from unbalanced to more balanced ones, the resultant
bodily reaction force becomes less. This seems to indi
cate that the more balanced hand positions are less cost
ly from a physiological standpoint. To determine if the
mean forces for these hand positions are significantly
different from each other, a Duncan Multiple Range Test
was run.
N = 120
s = 0.424
P = 2 3 4
R = 3.64 3^80 3.90
LSR = 1,54 1.6l 1.65
Hand Position 60°L 6OOR 120° l80°
Mean 34.21 28.00 21,67 14.82
As shown above, the mean forces produced using each
of the hand positions are all significantly different from
one another at the one per cent level of significance.
Thus, the optimum set of hand positions of those considered
in this experiment seems to be the one where the hands are
180 degrees apart with the left hand placed midway along
an arc running counterclockwise from top to bottom dead
center of the handwheel rim and the right hand placed
directly opposite. The 60 degrees left position appears
to be the most costly and least desirable of the four.
74
An explanation for the preceding results can be
found upon examining the initial hand positions and what
happens as the right and left hands move through the 90
degree arc to their final positions. When the hands move
counterclockwise from an initial 60 degrees left position,
body weight is shifted almost entirely to the left side.
The right and left hands begin and end their movements
on the left side of a vertical line dividing the hand-
wheel into equal right and left portions. Consequently,
the forces applied by the right and left hands are applied
entirely to the left of the longitudinal axis of the sub
ject's body. An extremely unbalanced downward force is
produced which contributes substantially to the resultant
bodily reaction force.
In moving from the 60 degree right position to the
final position the right hand traverses a 6o degree arc
on the right side of the handwheel and a 30 degree arc on
the left, while the left hand traverses its entire 90
degree arc on the left side of the wheel. Thus, the right
hand applies a primarily upward force while traversing 6o
degrees of its arc. However, a primarily downward force
is applied over an arc of 120 degrees. Here again an
unbalanced force is applied, though not as unbalanced as
in the 60 degrees left position where the total l80 degree
arc traversed by both hands is located on one side of the
handwheel.
75
The 120 degrees position is a relatively balanced
position so there is a fairly small shift in body weight
as the right and left hands move through the first 60
degrees of their arcs. However, as the right hand moves
counterclockwise through the final 30 degrees of its arc,
it passes top dead center of the handwheel and joins the
left hand on the left side of the handwheel. Thus, an
unbalanced force is applied during a total arc of 60 de
grees (30 for each hand) as compared to 180 and 120 de
grees for the 60 degrees left and 60 degrees right posi
tions, respectively.
The 180 degrees position is a perfectly balanced
one. The right and left hands are diametrically opposite.
The right hand traverses its 90 degree arc entirely on the
right side and the left hand traverses its 90 degree arc
entirely on the left side. Consequently, there is no large
unbalanced force generated which might substantially in
crease the normal reactive force since the right and left
hands remain exactly opposite one another throughout their
arcs, and the forces they apply are in opposite directions
and of approximately the same magnitude,
Handwheel Diameter and Resistant Torque Interaction (Figure I8)
Figure 18 indicates that there is not much difference
in the bodily reaction force for the three handwheel sizes
at the lowest torque value. However, as the resistant
76
ra -d P: P3 O pL,
0 O u o p
50
40 -
30
O 7" diameter A 10" diameter D 14" diameter
20 ..
10 .-
0 + '+-20 40 80
Torque (in.-lb)
120
Pig, l8,--Relationship between resultant bodily reaction force and resistant torque for various handwheel sizes
77
torque increases, the difference in the forces becomes
more pronounced. The effect of increasing torque seems
to have the greatest impact on the smallest handwheel
diameter. Note that the largest handwheel actually had
a higher value of reactive force than the 10 inch diameter
wheel at the lowest torque, but that the largest handwheel
became more advantageous as the torque increased. The
advantage shifts to to the l4 inch handwheel somewhere
between 20 and 40 inch-pounds of torque. This is the
divergence point that Davis (6) found in his study. He
says that this point usually lies around the 40 inch-
pound torque. Although this study is of a considerably
different nature than Davis', it seems to offer addition
al evidence that such divergence points exist and should
be considered in the design of handwheel controls.
These divergence points are due to the two primary
effects of increasing the handwheel diameter. The major
effect noted when the diameter becomes larger is a reduc
tion in the required applied force and, consequently, the
reactive force. The second effect acts to increase the
reactive force. It is usually relatively minor but be
comes more important as handwheel diameters get larger
and larger^ This effect was discussed in Chapter III
where it was noted that increasing the diameter and, in
effect, the moment arm of the applied force should increase
the bodily reaction forces to some extent for even those
78
handwheels that are not inordinately large. These two
antagonistic effects, then, interact to determine the
bodily reaction force for a particular size handwheel.
At low torque values and where larger handwheels are used,
the normally minor adverse effect of increasing handwheel
diameter may become the dominant factor in determining
the reactive force• As resistant torque increases, how
ever, this adverse effect becomes less and less important.
Figure 18 shows that the most advantageous handwheel diam
eter at 120 inch-pounds of torque is a diameter of l4
inches•
Handwheel Diameter and Relative
Height Interaction (Figure 19)
The graph in Figure 19 indicates that Increasing the
relative height does not have as pronounced an effect on
the larger handwheels as on the 7 inch diameter handwheels
This is reasonable when one considers that a rotation action
of the elbow is produced in turning the small handwheel at
the lower height, which is less effective in the application
of required force. As the relative height increases, the
locus of rotation moves upward toward the shoulder. For
the larger handwheels, a rotation action of the elbow and
of the shoulder is generated at the lower relative height,
although the elbow action is the primary action. As the
relative height becomes greater, the relative difference
In rotation action is much greater for the 7 incn diameter
79
ra Td Pl ps o pu,
0 O u o p
50 -
40 -
O 7" diameter A 10" diameter D i4" diameter
30 -•
20 "
10 ••
0 •+• •f
6" below elbow
6" above elbow
Relative Height
Pig, 19•--Relationship between resultant bodily reaction force and relative height for various handwheel sizes
80
wheel.
Figure 19 also shows that differences in reactive
force among the three handwheel sizes are not as great
at a relative height of six inches above the elbow. This
is probably due to the fact that the primary rotation
action is now centered about the shoulder for all three
handwheels, and the difference in elbow angles is com
paratively small.
The most desirable combination appears to be a
handwheel diameter of 14 Inches at a relative height of
six Inches above the elbow.
Handwheel Diameter and Hand
Position Interaction "" (Figure 20)
The deleterious effect of the small handwheel is
quite apparent in Figure 20, The bodily reaction force is
substantially greater than for handwheel diameters of 10
and l4 inches. There is a marked difference in the reac
tive forces of all handwheels for the 6o degrees left
position. This is not surprising as this is the most un
balanced and least desirable hand position of the four and
the large unbalanced forces generated for this hand posi
tion greatly accentuate the normal differences in reactive
force that are a function of handwheel diameter. At the
other three hand positions there is not a great deal of
difference in the bodily reaction forces for the 10 and 14
inch wheels, although the l4 inch handwheel appears to
81
ra Td c ps o
(XH
.\ 0 O PH
o p
50 •
40-
30 "
20 .
10 .
0 1—
0 7" diameter A 10" diameter D 14" diameter
1 1 —1
6 0 O L 6 0 O R 1200 i8oo
Hand Position
Fig. 20*--Relationship between resultant bodily reaction force and hand position for various handwheel sizes
82
consistently show somewhat lower bodily reaction forces.
For these larger diameters and relatively more balanced
positions, the handwheel effect has the greater impact
and acts to attenuate the hand position effect to some
degree,
At hand positions of 60 degrees right, 120 degrees,
and 180 degrees the final displacement of the right hand
to the left of the vertical diameter of the 10 inch hand-
wheel is 2,5, 2.5, and 0 inches, respectively. For the
l4 inch diameter, these displacements of the hand that
determines to a large extent the relative imbalance of
the applied force is very slight for these hand positions.
However, for the 60 degrees left position the right hand
traverses its full 90 degree arc on the left side of the
vertical handwheel diameter. In other words, the final
displacement of the right hand is equal to the radius of
the handwheel for this hand position. The difference In
displacement for the two handwheels now becomes twice what
it was at the 60 degrees right position. The greater dis
placement for the larger handwheel creates a larger moment
arm which probably decreases the magnitude of the unbalanc
ed force applied by the right hand enough to more than off
set a slight increase in imbalance also generated by this
larger moment arm.
The general trend of the curves in Figure 20 indi
cates the desirability of using balanced hand positions.
83
The most effective combination appears to be a hand posi
tion of 180 degrees and a handwheel diameter of either
10 or 14 inches.
Resistant Torque and Relative
Height Interaction" (Figure 21)
The trend of the curves in Figure 21 reflects the
general torque effect of increasing reactive forces as re
sistant torque increases. Note, however, that there does
not seem to be much difference in the forces for the two
relative heights at 20 and 40 inch-pounds of resistant tor
que, but that the difference becomes more pronounced as
torque continues to increase. This is not an unreasonable
result since one would expect the advantages gained by using
the "better" height to become increasingly more apparent
as greater applied forces are exerted to overcome compara
tively large amounts of resistant torque. Note also that
not only is there little difference in reactive forces be
tween the two heights at the two lower torque values, but
that there is little difference between forces generated
at the 20 and 40 inch-pound torque values for each height.
But then, the increase in torque is only 20 inch-pounds as
compared to increments of 40 inch-pounds as torque is further
increased.
The analysis of Figure 21 seems to indicate that in
creasing the relative height does not have a particularly
great impact on the bodily reaction forces produced at low
85
values of resistant torque, but that the impact is consid
erably greater at higher torque values. At a resistant
torque value of 120 inch-pounds, for example, the advan
tage gained by Increasing the relative height is quite
substantial.
Resistant Torque and Hand
Position Interaction (Figure 22)
Figure 22 shows that lower forces resulted from
the more balanced hand positions. As torque increases
above 40 inch-pounds, the differences in reactive force
among the various hand positions become appreciably more
distinct. The slopes of the top three curves indicate that
the resultant forces for these hand positions are markedly
affected as resistant torque is increased beyond 40 inch-
pounds • At the 20 inch-pound torque the difference be
tween the two balanced and two unbalanced hand positions
is quite apparent, however, there is not much difference
between the 60 degrees right and 60 degrees left positions
nor between the 120 and l80 degrees positions at this torque value.
Note the relatively flat slope of the bottom curve
in Figure 22, This would seem to indicate that resistant
torque had considerably less effect when the most balanced
hand position was used. Such a hand position seems to
counteract somewhat the adverse effect of increased re
sistant torque. The most logical reason for this is
86
50
40 -
ra Td P: P5
o pL. 0 O PH
o p
30 •-
20 "
10 ..
0
O A D d
60 degrees left 60 degrees right 120 dei
20 40 80 120
Torque (in,-lb)
Fig. 22,--Relationship between resultant bodily reaction force and resistant torque for various hand positions
87
probably the fact that the force platform is so much more
sensitive to unbalanced positions than balanced ones.
Finally, recall that after each subject had complet
ed the complete series of trials, he was asked which combi
nation (s) of variables he preferred. The majority of sub
jects selected the l4 inch handwheel with 20 inch-pounds
of resistant torque, a height of six inches above the elbow,
and the l80 degree hand position• This closely agrees with
the results of the analysis presented in the previous sec
tions of this chapter, and seems to bear out the hypothe
sis that the human being instinctively seeks to perform,
a task requiring considerable effort in such a manner as
to minimize the amount of energy expended or force exerted.
In other words, the human being, when given a choice, will
usually select optimum or near optimum values of variables
affecting his performance of a task requiring major expen
diture of energyo Konz and Day (21) found this to be true
in their study and Ellis (8) reported similar results in
an experiment evaluating work-surface height. At the con
clusion of the experimental session, Ellis requested each
subject to adjust the work surface to the height at which
he (the subject) believed it would be most comfortable to
work^ An average preferred work-surface height of 4l,3
inches was obtained from this procedure, which was in close
agreement with the optimal value experimentally determined
to be 42,0 Inches (8).
CHAPTER V
SUMMARY, CONCLUSIONS, AND RECOMMENDATIONS
The material in this chapter is presented in three
sections. The initial section gives a brief summary of
the entire study. The next section presents the conclu
sions drawn from this research along with some personal
observations of the author. The final section points out
some areas worthy of further research.
Summary
This experiment was performed to determine the ef
fects of handwheel size, resistant torque, control height,
and hand position on a simple, dynamic task involving the
operation of handwheel controls.
Three different sizes of handwheel controls and
four values of resistant torque were selected for consid
eration. Also evaluated were two relative control heights
and four different hand positions.
A homogeneous sample of five subjects was selected.
Each of the subjects was required to rotate a handwheel
counterclockwise against resistant torque while standing
on a force platform, and were required to perform this
task for all combinations of variables.
The experimental design chosen was a factorial
88
89
experiment in a randomized block design, where subjects
were treated as blocks. Using this design, an analysis
of variance was performed on the data obtained from the
recorded force platform outputs in order to determine if
the variables had a significant effect on task performance.
A graphical analysis was then prepared to ascertain the
exact nature of these ef.fects.
Conclusions
Recall that several hypotheses were made prior to
this investigation (see Chapter III), Recorded evidence
with subsequent engineering and statistical calculations
sustained them all. These hypotheses were made based upon
logic and proven techniques of engineering analysis. Al
though it is doubtful whether some of the more subtle re
lationships between design parameters could have been
predicted in this fashion, it does indicate that a thorough
engineering analysis can be extremely valuable in predict
ing the more fundamental effects of these parameters on
human performance. This is especially true where experi
mental results may prove too restrictive, or where experi
mentation might be considered too costly from the stand-
poing of time, labor, and money.
In many cases, however, the engineer has become so
infatuated with experimentation that he has forgotten the
basic tools of his profession. Care was taken to utilize
90
engineering principles and analysis techniques wherever
applicable and appropriate in conjunction with this ex
periment as an aid in promoting sound, effective research.
Nonetheless, upon examining the results of this experi
ment, it soon became apparent that the effects of some of
the variables could have been predicted and/or justified
using the basic principles of motion economy. Especially
applicable were those principles advocating symmetrical,
balanced motions of the body members. Although these prin
ciples come from the very core of a specific branch of
industrial engineering, they apparently are generally ap
plicable to many other related areas. This seems to pro
vide additional evidence that research involving experimen
tation which also incorporates effective techniques of en
gineering analysis and the application of proven engineer
ing principles usually produces results that are more gen
erally applicable, more conclusive, and more easily sub
stantiated than results produced by experimentation alone.
The conclusions drawn from this research were based
mainly upon evidence shown graphically and statistically
to be of practical significance. These conclusions are
additionally supported by basic scientific and engineering
principles as well as results of previous research. They
are :
1^ Resultant bodily reaction force can be used as
an effective index of physiological cost to the
91
operator in the performance of a simple, dynamic
task of the nature of the one evaluated in this
experiments The results of this study also of
fer additional evidence that the force platform,
when properly used, can be a valuable tool in
the design and evaluation of hand controls.
2. Larger handwheels are more advantageous to the
operator at resistant torque values above 40
inch-pounds. Handwheel diameter should be in
creased as resistant torque increases above
this value in order to decrease the applied
force requirements. Care should be taken,
however, to insure that diameters do not be
come so large that they become unwieldy, and
more costly from a physiological standpoint,
3. As the resistant torque that must be overcome
decreases, the applied force requirements and
the physiological cost to the operator become
less until a torque value of 40 inch-pounds is
reached. As torque values decrease below this
value, the resultant bodily reaction force may
or may not decrease depending upon the handwheel
size. When designing a handwheel control to be
used in a task such as that evaluated here, the
combined effects of the proposed handwheel size
and resistant torque should be taken into
92
consideration. The two parameters should be
evaluated as a combination and not separately.
4. Less physiological cost is generated when hand-
wheel controls are operated such that an elbow
angle of near 90 degrees is maintained and,
where this angle cannot be maintained, angles
less than this value are generally preferable
to angles exceeding 90 degrees.
5. Balanced hand positions are more advantageous
than positions where both hands are on the same
side of the handwheels This is especially true
at higher values of resistant torque where a
hand position of l8o degrees proved to be un
questionably the best of those evaluated•
This experiment was, of necessity, somewhat restric
tive in nature• Extreme caution should always be exercised
in attempting to generalize the results of such an experi
ment. Nevertheless, it is felt that the findings of this
study are generally applicable to tasks similar to the one
considered in this experiment, and provide valuable Infor
mation that can be extremely useful in the design of hand-
wheel controls for such tasks.
Recommendations for Further Research
The experimental evidence and conclusions obtained
from this study indicate areas in which further research
93
is likely to produce more specific and useful results.
The following items deserve the attention of further re
search:
1. Smaller increments of resistant torque at the
lower values should be evaluated experimentally
to pinpoint more accurately the divergence
point above which the advantage definitely
lies with the larger handwheels.
2. An effort should be made to experimentally
determine the effect of very large handwheels.
3. In addition to the preceding direct variations
of this experiment, a number of other possible
applications of the force platform are indicated.
Use of the force platform could probably indicate
the onset of physical fatigure by indicating lack
of coordination and decreased use of an ability
in worker performance. It may also enable the
detection of differences in work methods more
precisely and easily than other physiological
measuring methods. Finally, the force platform
may be found useful as a means of determining
industrial efficiency and classifying jobs by
physiological cost.
LIST OF REFERENCES
1. Adams, S. I965. Determination of human mechanical energy and work output from independent measures of force and motion. Unpublished PH.D. dissertation, Arizona State University,
2. Barany, J. 1963. The nature of individual differences in bodily forces exerted during a motor task as measured by a force platform. Journal of Industrial Engineerings l4: 332-41.
3. Barnes, R, - 1963. Motion and time Study. New York: John Wiley.
4. Corlett, E. 196I. The accuracy of setting of machine tools by means of handwheels and dials. Ergonomics. 4(l): 53-62.
5. Damon, A., Stoudt, H,, and McFarland, R. 1966. The human body in equipment design. Cambridge, Mass.: Harvard University Press.
6. Davis, L. 1949. Human factors in design of manual machine controls. Mechanical Engineering. 71:811-16.
7. Edel, Jr., D., ed. I967. Introduction to creative design. Englewood Cliffs, N. J.: Prentice-Hall, Inc.
8. Ellis, D. 1951. Speed of manipulative performance as a function of work-surface height. Journal of Applied Psychology, 35^ 289-96.
9. Floyd, W,, and Welford, A., ed. 1954• Human factors in equipment design. London: H, R, Lewis and Company,
10. Greene, J,, and Morris, W. 1958, The force platform: an industrial engineering tool^ Journal of Industrial Engineering. 9° 128-132.
11. Greene, J., Morris, W., and Wiebers, J. 1959. Nine methods for measuring physiological cost of work. Journal of Industrial Engineerings 10: i8oi^5ir:; ^ _ — -
94
95
12. Greene, J., and Morris, W. 1959. The design of a force platform for work measurement. Journal of Industrial Engineering. 10: 312-17, — ^ --^
13. Greene, J. 1957. Design and initial evaluation of a force platform for measuring human work. Unpublished PH. D. dissertation. State University of Iowa, (cited in Hudson, 1962).
14. Handbook of htmian engineering data. 1952• Medford, Mass.: Tufts University.
15. Helson, H. 1949. Design of equipment and optimal human performance. Am. Journal of Psychology. 62: 473-97. " """^
16. Hertzberg, H. I96O. Some contributions of applied physical anthropology to human engineering. USAF, WADD, Technical Report 6O-I9.
17. Hicks, C. 1964, Fundamental concepts in the design of experiments. New York: Holt, Rinehart, and Winston.
18. Hudson, W. 1962. The study of the dynamic characteristics of a reaction free' apparatus for measur' ing physiological cost of dynamic manual work^ Journal of Industrial Engineering, 13: 15-I8.
19. Ingenohl, I, 1959. Measuring physical effort. Journal of Industrial Engineering, 10: 99-114,
20. Karnasiewicz, E. 1968. An electromyographic study of a rotary task. Unpublished thesis, Texas Technological College,
21. Konz, S,, and Day, R. 1966. Design of controls using force as a criterion, H'oman Factors, 8: 121-27.
22. Lauru, L., and Brouha, L. 1957- Physiological study of motions. Advanced Management. 22: 17-24,
23. Lehmann, G, 1958. Physiological basis of tractor design. Ergonomics. 1: 197-206.
24• McCormick, E. 1964. Human factors engineering. New York: McGraw-Hill.
96
25. Morgan, C, Chapanis, A,, Cook, J., and Lund, M, 1963. Human engineering guide to equipment design. New York: McGraw-Hill,
26. Slnaiko, H., ed, 1961. Selected papers on human factors in the design and use of control systems. New York: Dover Publications, Inc.
27. Taylor, E, 1959- The interim report on engineering design, Cambridge, Mass,: Massachusetts Institute of Technology.
28. Whitney, R, 1958. The strength of the lifting action in man. Ergonomics, 1(2): 101-28
29. Woodson, W, 1954. Human engineering guide for equipment designers. University of " California Press.
APPENDIX
A. Force Platform Calibration Record
B. Graph of Subject Main Effect
97
98
APPENDIX A: FORCE PLATFORM CALIBRATION RECORD
Applied Force
(lb)
0
5
10
15
20
25
30
Pen Deflection in Millimeters
Frontal Axis
0.0
4,0
8.0
12.0
l6.0
20.0
24.0
Lateral Axis
0.0
4.0
8.0
12.0
16.0
20,0
24.0 •
Vertical Axis
0.0
2.0
4.0
6.0
8.0
10.0
12.0
ra Td
o pL.
0 o u o
99
APPENDIX B: GRAPH OP SUBJECT MAIN EFFECT
50
40-
30--
20--
10 .-
O O
o o
0 3
Subject Number