Real-time 3D MR Respiratory Motion Acquisition for Simultaneous Human PET-MR Imaging Liheng Guo 1 , Mark Ahlman 3 , Tao Feng 2 , Michael Guttman 4 , Elliot McVeigh 1 , David Bluemke 3 , Benjamin Tsui 2 , Daniel Herzka 1 1 Biomedical Engineering, Johns Hopkins University, Baltimore, MD. 2 Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, MD. 3 Radiology and Imaging Sciences, National Institutes of Health, Bethesda, MD. 4 Cardiology, School of Medicine, Johns Hopkins University, Baltimore, MD. References 1. Koesters, et al. Motion-compensated EM PET Reconstruction for Simultaneous PET-MR Exams. ISMRM 2014 0789. 2. Kolbitsch, et al. A 3D MR-acquisition scheme for non-rigid bulk motion correction in simultaneous PET-MR. ISMRM 2014 0788. 3. Huang, et al. A Novel Golden-Angle Radial FLASH Motion-Estimation Sequence for Simult. Thoracic PET-MR. ISMRM 2013 2684. 4. Kolbitsch, et al. T1 and T2 Weighted MR Acquisition for Bulk Motion Correction for Simultaneous PET-MR. ISMRM 2013 3745. 5. Grimm, et al. MR-PET Respiration Compensation Using Self-Gated Motion Modeling. ISMRM 2013 0829. 6. Fürst, et al. Respiratory motion compensation in PET/MR with a self-gating MR sequence. SNMMI 2013. 7. Würslin, et al. Respiratory Motion Correction in Oncologic PET Using T1-Weighted MR Imaging. J Nucl Med 54:3 (2013). Objective This work aims to develop a fast MR acquisition technique that can capture 3D respiratory motion in real time during a simultaneous PET- MR scan, providing an alternative to retrospective motion acquisition. The resulting motion vector field is to be used in the PET image reconstruction to correct for the respiratory motion. Motivation Positron emission tomography (PET) is the primary imaging modality for cancer imaging due to its superior sensitivity to detect tumors. PET images is severely degraded by respiratory motion (Fig. 1). Respiratory motion tracking in 3D using MRI during PET imaging is now possible with newly available simultaneous human PET-MR scanners, and enables correction of respiratory motion effects. Current motion tracking methods [1-8] uses retrospective binning guided by the diaphragm navigator (Fig. 2,3), which assumes the repeatability of motion and is vulnerable to irregular motion (Fig. 4). This work implements a fast real-time 3D acquisition on the MR that can measure 3D motion at arbitrary time, without the use of a navigator or assuming the repeatability of motion. The associated MR image reconstruction exploits both the spatial and temporal sparsity of respiratory motion. Figure 1. Example of respiratory motion. Left: normal torso PET image reconstructed using attenuation correction map acquired at a proper respiratory state. Right: respiratory motion caused misalignment of attenuation map, resulting in large portions of image missing (arrows). (Adapted from: Goerres, et al. Respiration-induced Attenuation Artifact at PET/CT: Technical Considerations. Radiology. 2003; 226:906–910.) MRI: continuous image/motion data collection PET: normal imaging Figure 2. The diaphragm navigator: during a simultaneous PET-MR scan, conventional techniques use the MR to execute the diaphragm navigator to track one point on the lung-liver interface, providing a 1D displacement (red) that drives the retrospective binning of continuously acquired MR data. Figure 3. Retrospective binning of MR data: the 1D displacement of the diaphragm navigator (red) is used to divide continuously acquired MR data into several bins, which are then reconstructed separately to independent 3D images. Motion vector fields are derived from these 3D images and are used to correct PET image reconstruction. Lung Liver Time Figure 4. Irregular breathing: Retrospective binning using the 1D diaphragm navigator assumes repeatability of respiratory motion and is vulnerable to irregularity shown here. 1. For the MR pulse sequence, 3D radial projection imaging was chosen for its tolerance to undersampling. An in-house fast 3D radial sequence was developed for this work. 2. Projection angles were arranged through time using 2-parameter golden angle [9], which can distribute any number of projections in the 3D angular space in a uniform, pseudo-random, and nonrepeating fashion (Fig. 5). As a result: a) one can run an open-ended scan without planning for the number of projections and be assured that all projections will be evenly distributed, and b) one can use a sliding-window reconstruction with arbitrary window width and be assured that the projections within each window are always evenly distributed too (Fig. 6). Human volunteer experiments: gradient-echo 3D projection scans were performed on Siemens Trio (3T) and Avanto (1.5T) scanners over a FOV of 400×400×400mm at ~2ms per projection (128 points/projection). A sliding window width of 1000 projections and increments of 500 projections were used, resulting in temporal resolution of 2 sec and update rate of 1 frame/sec. 3. To obtain 3D image for each sliding-window frame, the projections in each sliding window were reconstructed using L1 minimization in a sparse domain (3D wavelet: Daubechies-4 level 4) in order to exploit the spatial sparsity of anatomical images. A large-scale L1-minimization algorithm was used [10]. 4. To obtain the 3D motion vector field that describe the 3D deformation across all temporal frames, a 4D image registration technique [11,12] was used to analyze all frames at once and to enforce smoothness of motion across all frames. This exploits the temporal sparsity of respiratory motion. 4) 4D Image Registration 3) Compressed-Sending Image Reconstruction 2) Two-Param Golden Angle, Sliding Window Frames 1) 3D Radial Projections, Continuous Acquisition Methods k z k x k y Continuous 3D Projection Acquisition Figure 5. 1000, 4000, 8000 projections in the 3D k-space, arranged by a 2- parameter golden angle scheme, which is capable of maintaining uniformity and pseudo-randomness of projection distribution given arbitrary number of projections, without repeating any projection twice. First Proj. Last Proj. Figure 6. Projections in 3D k-space inside sliding windows of several widths and positions during an open-ended scan, as distributed by the 2-parameter golden angle, which can maintain distribution uniformity and pseudo-randomness at arbitrary window widths and positions. Projections 1 ~ 200 789 ~ 1234 2500 ~ 3500 1000 Projections 4000 Projections 8000 Projections Results The proposed 3D acquisition and reconstruction technique can generate real-time images that clearly delineate the diaphragm and the cardiac boundary, as seen in Fig. 7. As seen in Fig. 8, these 3D images provide sufficient details to generate motion vector field for motion correction in PET image reconstruction. Due to the use of golden angle, one has the option of running open-ended scans and free choice of temporal resolution (trades off smoothly against image quality) of reconstruction. Figure 7. Several frames of the coronal view of a real-time 3D scan showing respiratory motion, acquired at 128×128×128 pixels over 400×400×400 mm FOV, reconstructed to 2-sec temporal resolution and update rate of 1 frame/sec. Slice 60 of 128-slice volume is shown in (a~d), and volume projections are shown in (e~h). a b c d e f g h Figure 8. 4D registration results of the real-time 3D scan shown in Fig. 7. Several temporal frames shown here (a~d) are aligned to a mean-deformation frame (e), which minimizes total deformation across all frames. The motion vector field generated here will be used in PET reconstruction to correct for motion effects. a b c d e Conclusion The proposed technique enables the acquisition of 3D respiratory motion in real time, without the assumption of motion repeatability, offering an alternative to the diaphragm navigator and retrospective binning, which takes several minutes of scanning. Given its flexibility in arbitrary scan duration, the real- time motion scan can be run continuously during a PET-MR exam to monitor motion at all time, or can be run at select times during the exam and leave room for other diagnostic MR scans. 8. Petibon, et al. Cardiac motion compensation and resolution modeling in simul. PET-MR: a cardiac lesion detection study. Phys Med Biol 58 (2013) 2085–2102. 9. Chan, et al. Temporal Stability of Adaptive 3D Radial MRI Using Multidimensional Golden Means. Magnetic Resonance in Medicine 61:354–363 (2009). 10. J. Yang and Y. Zhang. Alternating direction algorithms for L1-problems in compressive sensing, SIAM Journal on Scientific Computing, 33, 1-2, 250-278, 2011. (Available at: YALL1: Your ALgorithms for L1 Optimization. http://yall1.blogs.rice.edu/ ) 11. S. Klein, M. Staring, K. Murphy, M.A. Viergever, J.P.W. Pluim, "elastix: a toolbox for intensity based medical image registration," IEEE Transactions on Medical Imaging, vol. 29, no. 1, pp. 196-205, January 2010. http://elastix.isi.uu.nl/. 12. C.T. Metz, S. Klein, M. Schaap, T. van Walsum and W.J. Niessen. Nonrigid registration of dynamic medical imaging data using nD+t B-splines and a groupwise optimization approach, Medical Image Analysis, 15 (2011) 238–249.

Upload: liheng-guo

Post on 17-Feb-2017

64 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: NIH TGM Guo 38 Real-Time 3D

Real-time 3D MR Respiratory Motion Acquisition for Simultaneous Human PET-MR Imaging Liheng Guo1, Mark Ahlman3, Tao Feng2, Michael Guttman4, Elliot McVeigh1, David Bluemke3, Benjamin Tsui2, Daniel Herzka1

1Biomedical Engineering, Johns Hopkins University, Baltimore, MD. 2Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, MD. 3Radiology and Imaging Sciences, National Institutes of Health, Bethesda, MD. 4Cardiology, School of Medicine, Johns Hopkins University, Baltimore, MD.

References 1.  Koesters, et al. Motion-compensated EM PET Reconstruction for Simultaneous PET-MR Exams. ISMRM 2014 0789. 2.  Kolbitsch, et al. A 3D MR-acquisition scheme for non-rigid bulk motion correction in simultaneous PET-MR. ISMRM 2014 0788. 3.  Huang, et al. A Novel Golden-Angle Radial FLASH Motion-Estimation Sequence for Simult. Thoracic PET-MR. ISMRM 2013 2684. 4.  Kolbitsch, et al. T1 and T2 Weighted MR Acquisition for Bulk Motion Correction for Simultaneous PET-MR. ISMRM 2013 3745. 5.  Grimm, et al. MR-PET Respiration Compensation Using Self-Gated Motion Modeling. ISMRM 2013 0829. 6.  Fürst, et al. Respiratory motion compensation in PET/MR with a self-gating MR sequence. SNMMI 2013. 7.  Würslin, et al. Respiratory Motion Correction in Oncologic PET Using T1-Weighted MR Imaging. J Nucl Med 54:3 (2013).

Objective This work aims to develop a fast MR acquisition technique that can capture 3D respiratory motion in real time during a simultaneous PET-MR scan, providing an alternative to retrospective motion acquisition. The resulting motion vector field is to be used in the PET image reconstruction to correct for the respiratory motion.

Motivation •  Positron emission tomography (PET) is the primary imaging modality

for cancer imaging due to its superior sensitivity to detect tumors. •  PET images is severely degraded by respiratory motion (Fig. 1). •  Respiratory motion tracking in 3D using MRI during PET imaging is now

possible with newly available simultaneous human PET-MR scanners, and enables correction of respiratory motion effects. •  Current motion tracking methods [1-8] uses retrospective binning

guided by the diaphragm navigator (Fig. 2,3), which assumes the repeatability of motion and is vulnerable to irregular motion (Fig. 4). •  This work implements a fast real-time 3D acquisition on the MR that

can measure 3D motion at arbitrary time, without the use of a navigator or assuming the repeatability of motion. The associated MR image reconstruction exploits both the spatial and temporal sparsity of respiratory motion.

Figure 1. Example of respiratory motion. Left: normal torso PET image reconstructed using attenuation correction map acquired at a proper respiratory state. Right: respiratory motion caused misalignment of attenuation map, resulting in large portions of image missing (arrows). (Adapted from: Goerres, et al. Respiration-induced Attenuation Artifact at PET/CT: Technical Considerations. Radiology. 2003; 226:906–910.)

w

MRI: continuous image/motion data collection

PET: normal imaging

Figure 2. The diaphragm navigator: during a simultaneous PET-MR scan, conventional techniques use the MR to execute the diaphragm navigator to track one point on the lung-liver interface, providing a 1D displacement (red) that drives the retrospective binning of continuously acquired MR data.

� � �

Figure 3. Retrospective binning of MR data: the 1D displacement of the diaphragm navigator (red) is used to divide continuously acquired MR data into several bins, which are then reconstructed separately to independent 3D images. Motion vector fields are derived from these 3D images and are used to correct PET image reconstruction.

Lung

Liver Time

Figure 4. Irregular breathing: Retrospective binning using the 1D diaphragm navigator assumes repeatability of respiratory motion and is vulnerable to irregularity shown here.

1.  For the MR pulse sequence, 3D radial projection imaging was chosen for its tolerance to undersampling. An in-house fast 3D radial sequence was developed for this work.

2.  Projection angles were arranged through time using 2-parameter golden angle [9], which can distribute any number of projections in the 3D angular space in a uniform, pseudo-random, and nonrepeating fashion (Fig. 5). As a result: a) one can run an open-ended scan without planning for the number of projections and be assured that all projections will be evenly distributed, and b) one can use a sliding-window reconstruction with arbitrary window width and be assured that the projections within each window are always evenly distributed too (Fig. 6). Human volunteer experiments: gradient-echo 3D projection scans were performed on Siemens Trio (3T) and Avanto (1.5T) scanners over a FOV of 400×400×400mm at ~2ms per projection (128 points/projection). A sliding window width of 1000 projections and increments of 500 projections were used, resulting in temporal resolution of 2 sec and update rate of 1 frame/sec.

3.  To obtain 3D image for each sliding-window frame, the projections in each sliding window were reconstructed using L1 minimization in a sparse domain (3D wavelet: Daubechies-4 level 4) in order to exploit the spatial sparsity of anatomical images. A large-scale L1-minimization algorithm was used [10].

4.  To obtain the 3D motion vector field that describe the 3D deformation across all temporal frames, a 4D image registration technique [11,12] was used to analyze all frames at once and to enforce smoothness of motion across all frames. This exploits the temporal sparsity of respiratory motion.

4) 4D Image Registration

3) Compressed-Sending

Image Reconstruction

2) Two-Param Golden Angle,

Sliding Window Frames

1) 3D Radial Projections,

Continuous Acquisition

Methods

kz  

kx  

ky  

Continuous 3D Projection Acquisition

Figure 5. 1000, 4000, 8000 projections in the 3D k-space, arranged by a 2-parameter golden angle scheme, which is capable of maintaining uniformity and pseudo-randomness of projection distribution given arbitrary number of projections, without repeating any projection twice.

First Proj.

Last Proj.

Figure 6. Projections in 3D k-space inside sliding windows of several widths and positions during an open-ended scan, as distributed by the 2-parameter golden angle, which can maintain distribution uniformity and pseudo-randomness at arbitrary window widths and positions.

Projections 1 ~ 200 789 ~ 1234 2500 ~ 3500

1000 Projections 4000 Projections 8000 Projections

Results The proposed 3D acquisition and reconstruction technique can generate real-time images that clearly delineate the diaphragm and the cardiac boundary, as seen in Fig. 7. As seen in Fig. 8, these 3D images provide sufficient details to generate motion vector field for motion correction in PET image reconstruction. Due to the use of golden angle, one has the option of running open-ended scans and free choice of temporal resolution (trades off smoothly against image quality) of reconstruction.

Figure 7. Several frames of the coronal view of a real-time 3D scan showing respiratory motion, acquired at 128×128×128 pixels over 400×400×400 mm FOV, reconstructed to 2-sec temporal resolution and update rate of 1 frame/sec. Slice 60 of 128-slice volume is shown in (a~d), and volume projections are shown in (e~h).

a b c d e f g h

Figure 8. 4D registration results of the real-time 3D scan shown in Fig. 7. Several temporal frames shown here (a~d) are aligned to a mean-deformation frame (e), which minimizes total deformation across all frames. The motion vector field generated here will be used in PET reconstruction to correct for motion effects.

a b c d e

Conclusion The proposed technique enables the acquisition of 3D respiratory motion in real time, without the assumption of motion repeatability, offering an alternative to the diaphragm navigator and retrospective binning, which takes several minutes of scanning. Given its flexibility in arbitrary scan duration, the real-time motion scan can be run continuously during a PET-MR exam to monitor motion at all time, or can be run at select times during the exam and leave room for other diagnostic MR scans.

8.  Petibon, et al. Cardiac motion compensation and resolution modeling in simul. PET-MR: a cardiac lesion detection study. Phys Med Biol 58 (2013) 2085–2102. 9.  Chan, et al. Temporal Stability of Adaptive 3D Radial MRI Using Multidimensional Golden Means. Magnetic Resonance in Medicine 61:354–363 (2009). 10.  J. Yang and Y. Zhang. Alternating direction algorithms for L1-problems in compressive sensing, SIAM Journal on Scientific Computing, 33, 1-2, 250-278, 2011.

(Available at: YALL1: Your ALgorithms for L1 Optimization. http://yall1.blogs.rice.edu/) 11.  S. Klein, M. Staring, K. Murphy, M.A. Viergever, J.P.W. Pluim, "elastix: a toolbox for intensity based medical image registration," IEEE Transactions on Medical

Imaging, vol. 29, no. 1, pp. 196-205, January 2010. http://elastix.isi.uu.nl/. 12.  C.T. Metz, S. Klein, M. Schaap, T. van Walsum and W.J. Niessen. Nonrigid registration of dynamic medical imaging data using nD+t B-splines and a groupwise

optimization approach, Medical Image Analysis, 15 (2011) 238–249.