new sensitivity and selectivity gains with the … notes/new_sensitivity...ruo-mkt-11-1287-a new...
TRANSCRIPT
RUO-MKT-11-1287-A
New Sensitivity and Selectivity Gains with the QTRAP® 6500 System and SelexION™ Technology for Steroid Analysis For Research Use Only. Not for use in diagnostic procedures.
Michael Jarvis
SCIEX, Canada
2 © 2016 AB Sciex RUO-MKT-11-1287 A
Introduction
• The two major challenges facing steroid researchers (in fact, all
researchers!) using LC/MS/MS are:
‒ Sensitivity
‒ Selectivity
3 © 2016 AB Sciex RUO-MKT-11-1287 A
Challenges for LC-MS/MS Analysis of Steroids
Sensitivity:
• Low ionization
efficiencies
Selectivity:
• Common precursor ion
masses
• Similar fragmentation
patterns
• Similar chromatographic
retention properties
• Lots of endogenous
interferences http://en.wikipedia.org/wiki/Steroidogenesis#Steroidogenesis
4 © 2016 AB Sciex RUO-MKT-11-1287 A
XIC of +MRM (6 pairs): Exp 1, 287.2/97.0 Da ID: ASD from Sample 15 (Androgens mix 5) of Quant-RestekC18-150msec.wiff (Turbo Spr... Max. 8.7e4 cps.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5Time, min
0.00
5000.00
1.00e4
1.50e4
2.00e4
2.50e4
3.00e4
3.50e4
4.00e4
4.50e4
5.00e4
5.50e4
6.00e4
6.50e4
7.00e4
7.50e4
8.00e4
8.50e4
9.00e4
9.50e4
1.00e5
1.05e5
1.10e5
1.15e5
1.20e5
1.25e5
1.29e5
Inte
ns
ity
, c
ps
3.17
ASD
T DHT
AND
DHEA 5a-ASD
75% B 80% B
98% B
55% B
(~400pg/mL)
(~500pg/mL)
(~30ng/mL) (~10ng/mL)
(~4ng/mL)
(~10ng/mL)
Analysis of 6 androgens in 60:40 methanol:water
Similar retention times
Common fragment ions complicate the analysis
5 © 2016 AB Sciex RUO-MKT-11-1287 A
Analysis of androstenedione in urine sample
Enzymatic hydrolysis performed, to remove glucuronides
Numerous interferences observed
Androstenedione
For Research Use Only. Not for Use in Diagnostic Procedures.
6 © 2016 AB Sciex RUO-MKT-11-1287 A
Agenda
• Pushing the limits of sensitivity with the QTRAP® 6500
LC/MS/MS system
• Enhanced selectivity using SelexION™ ion mobility
technology and LC/MS/MS
SCIEX Triple Quad™ and QTRAP® 6500 LC/MS/MS System
8 © 2016 AB Sciex RUO-MKT-11-1287 A
SCIEX Triple Quad™ and QTRAP® 6500 System
• IonDrive™ technology provides a
new level of performance – Up to 10x greater sensitivity
– Up to 6 orders dynamic range
– Improved robustness
• Advanced dual RF electronics
allows 2 modes of operation: – High Sensitivity (5-1250 Da)
– Extended mass range (5-2000 Da)
• Compatible with SelexION™ ion
mobility technology
• 2x faster scanning MRM3
• Same compact footprint as 4500
and 5500 system (32” x 32” x 24”)
9 © 2016 AB Sciex RUO-MKT-11-1287 A
IonDrive™ Technology Create more ions. Capture more ions. Detect more ions.
IonDrive™
Turbo V
Source
IonDrive™
High Energy
Detector
IonDrive™
QJet Ion
Guide
10 © 2016 AB Sciex RUO-MKT-11-1287 A
6x signal gain
>4x S/N gain
QTRAP® 5500
system
Actual MRM Sensitivity Gains +ESI mode
QTRAP® 6500
system
Verapamil
m/z 455
Erythromycin
m/z 734
5x signal gain
5x S/N gain
Ginsenoside Rb2
m/z 1101
20x signal gain
>4x S/N gain
Thyroid Research:
Analysis of Free T3 and Free T4
12 © 2016 AB Sciex RUO-MKT-11-1287 A
Thyroid hormones, T3 and T4
• Thyroid hormones circulate in the blood as mostly protein bound
hormone (about 99.9% bound)
• The major binding protein is Thyroxine Binding Globulin (TBG)
and transthyretin. Only the unbound form of the hormones are
biologically active.
• Free T3 (2-4 pg/mL) and free T4 (8-15 pg/mL) are present in
very low concentrations in serum.
O
O
I
I
I
OH
NH2
OH
O
O
I
I
I
I
OH
NH2
OH
3,3,5-triiodothyronine (T3) Thyroxine (T4)
13 © 2016 AB Sciex RUO-MKT-11-1287 A
Sample Preparation
• Free (unbound) fraction of T3 and T4 was isolated from the
protein-bound fraction by ultrafiltration, using a filter having
10kDa molecular-weight cut-off.
• 500uL of serum was subjected to ultrafiltration at 30C, for 60
minutes.
• 25uL of internal standard was added to 100uL of the ultrafiltrate
• 50uL was directly injected onto LC/MS/MS system
14 © 2016 AB Sciex RUO-MKT-11-1287 A
MS Method
• Triple Quad™ 6500 system
• Positive ESI mode
Name Q1 Q3
T3 (1) 651.7 605.8
T3 (2) 651.7 479.0
T4 (1) 777.7 731.6
T4 (2) 777.7 604.8
LC Method
Phenomenex Kinetex C18, 50x4.6mm, 2.6um @ 0.6 mL/min
A = Water + 0.1% acetic acid
B = Methanol + 0.1% acetic acid
5.0 8.0
8.1 10.0 0
Time %B
0 20
5 95
8 95
8.1 20
10 20
15 © 2016 AB Sciex RUO-MKT-11-1287 A
LOQ < 0.5 pg/mL
T3
Blank injection
XIC of +MRM (6 pairs): 651.700/605.800 Da ID: T3 1 from Sample 3 (blank) of Feb22-2014_final.wiff (Turbo Spray IonDrive), S... Max. 2921.9 cps.
5.00 5.05 5.10 5.15 5.20 5.25 5.30 5.35 5.40 5.45 5.50 5.55 5.60 5.65 5.70 5.75 5.80 5.851112 1123 1134 1145 1156 1167 1178 1189 1201 1212 1223 1234 1245 1256 1267 1278 1289 1301
Time, min
0
50
100
150
200
250
300
350
400
450
500
550
600
619
In
te
ns
ity
, c
ps
5.61 5.83
5.885.47
5.79
5.65
5.715.76
5.22
4.975.53
5.38
5.55
XIC of +MRM (6 pairs): 651.700/605.800 Da ID: T3 1 from Sample 6 (0.5pg) of Feb22-2014_final.wiff (Turbo Spray IonDrive), S... Max. 2781.3 cps.
5.05 5.10 5.15 5.20 5.25 5.30 5.35 5.40 5.45 5.50 5.55 5.60 5.65 5.701123 1134 1145 1156 1167 1178 1189 1201 1212 1223 1234 1245 1256 1267
Time, min
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2781
In
te
ns
ity
, c
ps
5.37
5.06 5.68
5.16 5.61 5.635.565.53 5.665.22
XIC of +MRM (6 pairs): 777.700/731.600 Da ID: T4 1 from Sample 6 (0.5pg) of Feb22-2014_final.wiff (Turbo Spray IonDrive), S... Max. 2784.4 cps.
5.40 5.45 5.50 5.55 5.60 5.65 5.70 5.75 5.80 5.85 5.90 5.95 6.00 6.05 6.10 6.15 6.20 6.251201 1212 1223 1234 1245 1256 1267 1278 1289 1301 1312 1323 1334 1345 1356 1367 1378 1389
Time, min
0
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
In
te
ns
ity
, c
ps
5.74
5.93
6.25
5.97 6.216.02 6.11 6.155.69 6.065.45 6.095.61 5.64 5.855.83
5.38 5.515.43
XIC of +MRM (6 pairs): 777.700/731.600 Da ID: T4 1 from Sample 3 (blank) of Feb22-2014_final.wiff (Turbo Spray IonDrive) Max. 7076.0 cps.
5.40 5.45 5.50 5.55 5.60 5.65 5.70 5.75 5.80 5.85 5.90 5.95 6.00 6.05 6.10 6.15 6.20 6.251201 1212 1223 1234 1245 1256 1267 1278 1289 1301 1312 1323 1334 1345 1356 1367 1378 1389
Time, min
0
50
100
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
875
In
te
ns
ity
, c
ps
5.72
6.23
5.845.47 6.21
5.77
5.68
6.125.87 6.10
6.155.63
5.66 6.19
5.91
5.805.59
5.99 6.06
T4
Blank injection
T3
0.5 pg/mL (S/N = 48)
T4
0.5 pg/mL (S/N = 60)
16 © 2016 AB Sciex RUO-MKT-11-1287 A
Measurements in serum
T3, sample 2
1.4 pg/mL
T4, sample 5
7.8 pg/mL
T3, sample 4
3.4 pg/mL
T4, sample 4
17.3 pg/mL
For Research Use Only. Not for Use in Diagnostic Procedures.
Steroid Research:
Analysis of Aldosterone
18 © 2016 AB Sciex RUO-MKT-11-1287 A
5.0
7.5
7.6 10.0 0 1.0
6.0
6.5
MS Method
• Triple Quad™ 6500 system
• Negative ESI mode
Name Q1 Q3
Aldosterone (1) 359.2 189.0
Aldosterone (2) 359.2 331.1
Aldosterone-d7 366.2 338.2
LC Method
Phenomenex Gemini-NX C18, 150x3.0mm, 5um @ 0.5 mL/min
A = Water + 2mM ammonium acetate
B = Methanol + 2mM ammonium acetate Time %B
0 20
1 20
5 70
6 70
6.5 90
7.5 90
7.6 20
10 20
19 © 2016 AB Sciex RUO-MKT-11-1287 A
LOQ = 1 pg/mL
14 pg/mL (39.6 pmol/L) aldosterone in human
serum sample
LOQ = 1 pg/mL (2.8 pmol/L) aldosterone in human serum
14 pg/mL Aldosterone
(unknown sample) 1 pg/mL Aldosterone
(serum calibrator)
For Research Use Only. Not for Use in Diagnostic Procedures.
20 © 2016 AB Sciex RUO-MKT-11-1287 A
Linear Dynamic Range: 1pg/mL – 1mg/mL Aldosterone
1 mg/mL
6 orders LDR!
1 pg/mL 1 pg/mL
5 pg/mL
10 pg/mL
100 pg/mL
SelexION™ Ion Mobility Technology
22 © 2016 AB Sciex RUO-MKT-11-1287 A
Revolutionizing the way we do LC-MS/MS
• Benefits of ion mobility separations ‒ Separation of isobaric interferences
‒ Removal of chemical background noise
‒ Enables simplified sample preparation
‒ Enables faster LC run-times
‒ Improves LLOQ
‒ Improves data quality
• Features of the SelexION™ device ‒ Installation without breaking vacuum
‒ Ability to operate in “transparent” mode
‒ User-defined resolution setting
‒ Rapid MRM scanning
‒ Enhanced separations using chemical
modifiers
SCIEX Triple Quad™ 6500 system,
equipped with SelexION™ technology
23 © 2016 AB Sciex RUO-MKT-11-1287 A
SelexION™ Ion Mobility Device Components:
New hardware for
SelexION™ device
Existing ion path
24 © 2016 AB Sciex RUO-MKT-11-1287 A
SelexION™ Ion Mobility Device Components:
Robust, easy-to-install, hardware components:
No tools required
No cables
No need to break vacuum
Installation in about 2 minutes
1. Orifice Plate 2. Ion Mobility Cell 3. Curtain Plate
25 © 2016 AB Sciex RUO-MKT-11-1287 A
a0
a0
a=0
t1
t2
)0(
)0()()(
K
KEKE
a
Tuning the SelexION™ Ion Mobility Device
Ions migrate towards one of the planar electrodes
– If an ion’s high-field mobility K(E) is larger than its low-field mobility K(0), a>0
– If an ion’s low-field mobility K(0) is larger than it high-field mobility K(E), a<0
Any ion can be steered back onto the center-line, by application of a compound-specific DC compensation voltage (CV).
Compensation
Voltage (COV)
Separation
Voltage (SV)Compensation
Voltage (COV)
Separation
Voltage (SV)
ions in
transport gas to mass
spectrometer
26 © 2016 AB Sciex RUO-MKT-11-1287 A
Tuning the SelexION™ Ion Mobility Device
In infusion mode, the compensation voltage (CV) parameter is ramped in order to determine the optimized value.
Note that this is analogous to tuning any other compound-specific parameter, e.g. the collision energy (CE)
-14 -12 -10 -8 -6 -4 -2 0 2 4 6 8 10 12 14 16 180%
50%
100%
Rel. I
nt.
(%
)
Compensation Voltage (CV)
-14 -12 -10 -8 -6 -4 -2 0 2 4 6 8 10 12 14 16 180%
50%
100%
Rel. I
nt.
(%
)
Compensation Voltage (CV)
The ion residence time, and
hence the resolution of the ion
mobility device, can be controlled.
CV peak widths range from
approximately 0.5 – 2.5 V FWHM.
31 © 2016 AB Sciex RUO-MKT-11-1287 A
Resolving isobaric prostaglandins with SelexION™ technology
• Optimization of the SelexION™ device for isomers is accomplished by ramping the COV parameter.
• At COV=10, only PGF2a is transmitted. At COV=7, only 8-iso-PGF2a is transmitted.
COV=6
COV=7
COV=8
COV=9
COV=10
COV=11
COV=9
PGF2a
8-iso
COV=6
COV=7
COV=8
COV=9
COV=10
COV=11
COV=9
PGF2a
8-iso
COV=6
COV=7
COV=8
COV=9
COV=10
COV=11
COV=9
PGF2a
8-iso
COV=6
COV=7
COV=8
COV=9
COV=10
COV=11
COV=9
PGF2a
8-iso
O
OH
OH
OH
OH
CH3
PGF2a 8-iso PGF2a
O
OH
OH
OH
OH
CH3
Chromatographic run-time (minutes)
32 © 2016 AB Sciex RUO-MKT-11-1287 A
Resolving isobaric prostaglandins with SelexION™ technology
SelexION™ device
“off”
SelexION™ device “on”
Tuned for PGF2a
SelexION™ device “on”
Tuned for 8-iso-PGF2a
(a)
(b)
(c)
PGF2a
8-iso-PGF2a
PGF2a
8-iso-PGF2a
O
OH
OH
OH
OH
CH3
PGF2a
O
OH
OH
OH
OH
CH3
PGF2a
8-iso PGF2a
O
OH
OH
OH
OH
CH3
8-iso PGF2a
O
OH
OH
OH
OH
CH3
Q1 Q3 CE
353.2 309.2 -28
Q1 Q3 CE CV
353.2 309.2 -28 10.0
Q1 Q3 CE CV
353.2 309.2 -28 7.0
33 © 2016 AB Sciex RUO-MKT-11-1287 A
Protein Precipitation (PPT) – LC-MS/MS
289/109 289/97 Testosterone
peak in
female serum
Testosterone
peak in
female serum
Protein Precipitation (PPT) – LC-MS/MS with SelexION™ Technology
289/109 289/97
For Research Use Only. Not for Use in Diagnostic Procedures.
Analysis of the Neuro-active Steroid Allopregnanolone
35 © 2016 AB Sciex RUO-MKT-11-1287 A
Metabolism of Neurosteroids
Progesterone 5a-Dihydroprogesterone Allopregnanolone
(3a,5a-Tetrahydroprogesterone)
Pregnanolone
(3a,5b-Tetrahydroprogesterone)
Pregnenolone
3b-HSD
5a-reductase
3a-HSD
3a-HSD
5b-Dihydroprogesterone
• The most familiar steroid hormones, commonly measured in blood for research
purposes, are the mineralocorticoids, glucocorticoids and the androgens and
estrogens.
• One alternative pathway that has been of great interest to researchers is the
production of compounds by 5α-reductase , followed by 3α-hydroxysteroid
oxido reductase.
36 © 2016 AB Sciex RUO-MKT-11-1287 A
Currently, the assays in use are an RIA (but supplies of the
antibody are very low) and GC/MS (which is only performed by a
few academic labs).
Availability of the assay is a real backlog in trying to perform
research studies.
• Sensitivity ‒ Potential for increased sensitivity by LC/MS/MS
• Sample prep ‒ 5-day process with GC/MS
‒ Very labor intensive for GC/MS
‒ Many opportunities for errors
‒ Limited number of samples/run
‒ Long sample run time for GC/MS
Allopregnanolone: Why LC-MS/MS?
37 © 2016 AB Sciex RUO-MKT-11-1287 A
Current Method Early days for LC-MS/MS…
10 pg/mL
(31.5 pM)
Sample Name: "THP, 0.5 ng/mL" Sample ID: "" File: "THP LLE calibration curve with IS 2012-10-09.wiff"Peak Name: "THP 1" Mass(es): "301.200/135.100 Da"Comment: "" Annotation: ""
Sample Index: 12
Sample Type: Standard
Concentration: 0.500 ng/mL
Calculated Conc: 0.457 ng/mL
Acq. Date: 2012/10/09
Acq. Time: 5:44:30 PM
Modified: Yes
Proc. Algorithm: Specify Parameters - MQ III
Noise Percentage: 100
Base. Sub. Window: 1.00 min
Peak-Split. Factor: 2
Report Largest Peak: Yes
Min. Peak Height: 0.00 cps
Min. Peak Width: 0.00 sec
Smoothing Width: 3 points
RT Window: 10.0 sec
Expected RT: 7.46 min
Use Relative RT: No
Int. Type: Manual
Retention Time: 7.46 min
Area: 5.73e+005 counts
Height: 2.00e+005 cps
Start Time: 7.43 min
End Time: 7.52 min
7.00 7.05 7.10 7.15 7.20 7.25 7.30 7.35 7.40 7.45 7.50 7.55 7.60 7.65 7.70 7.75 7.80 7.85 7.90 7.951910 1923 1937 1950 1964 1978 1991 2005 2019 2032 2046 2059 2073 2087 2100 2114 2128 2141 2155 2169
Time, min
0.0
2.0e4
4.0e4
6.0e4
8.0e4
1.0e5
1.2e5
1.4e5
1.6e5
1.8e5
2.0e5
2.2e5
2.4e5
2.6e5
2.8e5
3.0e5
3.2e5
3.4e5
3.6e5
3.8e5
4.0e5
4.2e5
Inte
ns
ity
, c
ps
7.56
7.12 7.46
7.35
7.32
7.25
7.41
7.05
7.79
7.67
7.957.89
500 pg/mL
(157.5 pM)
Keys to success
1. Derivatization with the SCIEX Amplifex™ Keto reagent
– Add reagent at room temp; 30 minutes; no need for fume hood
2. Removal of interferences with SelexION™ ion mobility technology
38 © 2016 AB Sciex RUO-MKT-11-1287 A
Plasma samples Collected from pregnant, postpartum and normal females
Liquid-liquid extraction Extract 100uL of plasma, using 1:1 (v:v) ethyl acetate:hexane mixture
Derivatization Add 50uL of SCIEX Amplifex™ Keto reagent to the dried, extracted sample
Allow to react for 1 hour.
Sample preparation
Allopregnanolone Derivatized Allopregnanolone
Steroids are non-polar, and are difficult to ionize
Derivatization enhanced the sensitivity, by improving the ionization efficiency
For Research Use Only. Not for Use in Diagnostic Procedures.
39 © 2016 AB Sciex RUO-MKT-11-1287 A
MS/MS and Ion Mobility
Ion
mobility MS/MS LC
a0
a0
a=0
t1
t2
Compensation
Voltage (COV)
Separation
Voltage (SV)Compensation
Voltage (COV)
Separation
Voltage (SV)
ions in
transport gasto mass
spectrometer
Q1 Q2 Q3
Mass selection
of parent ion
Fragmentation Mass selection
of fragment ion
To
Detector
• MS/MS separates compounds
based on differences in the
masses of the ions (both
precursor ions and fragment
ions).
Ion mobility separates compounds based on differences in the size/shape of the ionized compounds.
40 © 2016 AB Sciex RUO-MKT-11-1287 A
Assessing the effect of the SelexION™ ion mobility device
Postpartum sample #P5
(without ion mobility)
Postpartum sample #P5
(with ion mobility)
LC MS/MS
Ion
mobility MS/MS LC
allopregnanolone
allopregnanolone
For Research Use Only. Not for Use in Diagnostic Procedures.
41 © 2016 AB Sciex RUO-MKT-11-1287 A
10 pg/mL
allopregnanolone
(31.5 pM)
50 pg/mL
allopregnanolone
(157.2 pM)
200 pg/mL
allopregnanolone
(628.9 pM)
Sensitivity of the LC-ion mobility-MS/MS method
ALLO derivatized LC MRM calibration curve 2012-10-24 reinjection with ACN MP.rdb (THP 1): "Linear" Regression ("1 / (x * x)" weighting): y = 0.37 x + 0.00146 (r = 0.9994)
0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55Analyte Conc. / IS Conc.
0.00
0.01
0.02
0.03
0.04
0.05
0.06
0.07
0.08
0.09
0.10
0.11
0.12
0.13
0.14
0.15
0.16
0.17
0.18
0.19
An
aly
te A
rea
/ I
S A
rea
50 pg/mL 10 pg/mL
500 pg/mL
200 pg/mL
Using this very sensitive method, it was possible to measure <10pg/mL
Only 10uL were injected
The instrument response was linear over the calibration range covering from 10 - 25,000 pg/mL
For Research Use Only. Not for Use in Diagnostic Procedures.
42 © 2016 AB Sciex RUO-MKT-11-1287 A
Pregnanolone
(3a,5b-THP)
Allopregnanolone
(3a,5a-THP)
Epipregnanolone
(3b,5b-THP)
Epiallopregnanolone
(3b,5a-THP)
Allopregnanolone
Pregnanolone
Epi-allopregnanolone
+ Epi-pregnanolone
Representative chromatogram: “Pregnant” sample
For Research Use Only. Not for Use in Diagnostic Procedures.
43 © 2016 AB Sciex RUO-MKT-11-1287 A For Research Use Only. Not for Use in Diagnostic Procedures.
44 © 2016 AB Sciex RUO-MKT-11-1287 A
• Keys to success, for LC-MS/MS method: ‒ Ion mobility separation using the SelexION™ device
‒ Derivatization using the SCIEX Amplifex™ Keto reagent
‒ Highly sensitive SCIEX QTRAP® 6500 LC/MS/MS system
• Great potential for research into: ‒ Detecting individual differences in metabolism of reproductive and
neurosteroid hormones
‒ Predicting individual differences in responses to SSRIs and other
medications.
• The QTRAP® 6500 system, combined with SelexION™ ion
mobility technology, provides researchers with the sensitivity
and selectivity required to measure low-level endogenous
steroids in complex biological matrices.
Conclusions
46 © 2016 AB Sciex RUO-MKT-11-1287 A
Trademarks/Licensing
AB Sciex is doing business as SCIEX.
For Research Use Only. Not for use in diagnostic procedures.
© 2016 AB Sciex. The trademarks mentioned herein are
the property of AB Sciex Pte. Ltd. or their respective owners.
AB SCIEX™ is being used under license.