General Specifications Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 Model FLXA21 2-Wire Analyzer GS 12A01A02-01E n General The model FLXA ® 21 2-Wire Analyzer, one model of FLEXA ® series, offers single or dual sensor measurement. The modular-designed analyzer offers 4 kinds of measurements – pH/ORP (oxidation-reduction potential), contacting conductivity (SC), inductive conductivity (ISC) or dissolved oxygen (DO) – with the respective sensor module. For dual sensor measurement, the combination of two same type sensor inputs – pH/ORP and pH/ORP (analog sensor only), SC and SC, and DO and DO – are available with two sensor modules. Dual sensor measurement offers additional functionalities; calculated data function and redundant system. Variety of calculated data from two measuring parameters is selectable for each measurement. On the redundant system built on two measuring parameters of two sensor inputs, main output parameter is automatically switched over to the second sensor output in case of the main sensor’s failure condition. Addition to conventional analog pH/ORP sensors, the analyzer FLXA21 can be connected to Yokogawa’s digital sensor, FU20F pH/ORP SENCOM ® Sensor. In the FLXA21 Human Machine Interface (HMI), 2-wire type analyzer FLXA21 offers easy touch screen operation and simple menu structure in 12 languages. Menus of display, execution and setting are displayed in a selected language. The analyzer FLXA21 automatically recognizes the installed sensor module and prepares the necessary menus for right configuration, even for dual sensor measurement. For immediate measurement, the FLXA21 offers quick setup functionality. The quick setup screen appears when the analyzer is powered. Only a few setups – date/time, language, basic sensor configurations and output – will start the measurement. The FLXA21 offers the best accuracy in measurement with temperature compensation functionality and calibration functionality. Sensor diagnostics and sensor wellness indication make measurement reliable. Logbook of events and diagnostic data is a useful information source for maintenance. For the wide range of industrial environment, the FLXA21 is designed with the enclosure of plastic, stainless steel or stainless steel with corrosion- resistant coating. And, for hazardous location, the FLXA21 has approvals of ATEX, IECEx, FM, CSA and NEPSI. n Features • 4 kinds of measurements; pH/ORP, SC, ISC and DO • Dual sensor measurement on 2-wire type analyzer; pH/ORP and pH/ORP, SC and SC, and DO and DO • Calculated data from dual sensor measurement • Redundant system on dual sensor measurement • Connection of digital FU20F pH/ORP SENCOM Sensor • Easy touch screen operation on 2-wire type analyzer • Simple HMI menu structure in 12 languages • Quick setup menu for immediate measurement • Indication of sensor wellness • Enclosure – plastic, stainless steel or stainless steel with corrosion-resistant coating • Hazardous location approvals – ATEX, IECEx, FM, CSA and NEPSI GS 12A01A02-01E ©Copyright Mar. 2010 11th Edition Feb. 02, 2016

Upload: others

Post on 18-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

GeneralSpecifications

Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 JapanTel.: 81-422-52-5617 Fax.: 81-422-52-6792

Model FLXA212-Wire Analyzer

GS 12A01A02-01E

n GeneralThe model FLXA®21 2-Wire Analyzer, one model of FLEXA® series, offers single or dual sensor measurement. The modular-designed analyzer offers 4 kinds of measurements – pH/ORP (oxidation-reduction potential), contacting conductivity (SC), inductive conductivity (ISC) or dissolved oxygen (DO) – with the respective sensor module.For dual sensor measurement, the combination of two same type sensor inputs – pH/ORP and pH/ORP (analog sensor only), SC and SC, and DO and DO – are available with two sensor modules. Dual sensor measurement offers additional functionalities; calculated data function and redundant system.Variety of calculated data from two measuring parameters is selectable for each measurement. On the redundant system built on two measuring parameters of two sensor inputs, main output parameter is automatically switched over to the second sensor output in case of the main sensor’s failure condition.Addition to conventional analog pH/ORP sensors, the analyzer FLXA21 can be connected to Yokogawa’s digital sensor, FU20F pH/ORP SENCOM® Sensor.

In the FLXA21 Human Machine Interface (HMI), 2-wire type analyzer FLXA21 offers easy touch screen operation and simple menu structure in 12 languages. Menus of display, execution and setting are displayed in a selected language.The analyzer FLXA21 automatically recognizes the installed sensor module and prepares the necessary menus for right configuration, even for dual sensor measurement.For immediate measurement, the FLXA21 offers quick setup functionality. The quick setup screen appears when the analyzer is powered. Only a few setups – date/time, language, basic sensor configurations and output – will start the measurement.

The FLXA21 offers the best accuracy in measurement with temperature compensation functionality and calibration functionality. Sensor diagnostics and sensor wellness indication make measurement reliable. Logbook of events and diagnostic data is a useful information source for maintenance.

For the wide range of industrial environment, the FLXA21 is designed with the enclosure of plastic, stainless steel or stainless steel with corrosion-resistant coating. And, for hazardous location, the FLXA21 has approvals of ATEX, IECEx, FM, CSA and NEPSI.

n Features•4kindsofmeasurements;pH/ORP,SC,ISCandDO•Dualsensormeasurementon2-wiretypeanalyzer;pH/ORPandpH/ORP,SCandSC,andDOandDO

•Calculateddatafromdualsensormeasurement•Redundantsystemondualsensormeasurement•ConnectionofdigitalFU20FpH/ORPSENCOMSensor

•Easytouchscreenoperationon2-wiretypeanalyzer•SimpleHMImenustructurein12languages•Quicksetupmenuforimmediatemeasurement•Indicationofsensorwellness•Enclosure–plastic,stainlesssteelorstainlesssteelwithcorrosion-resistantcoating

•Hazardouslocationapprovals–ATEX,IECEx,FM,CSAandNEPSI

GS12A01A02-01E©CopyrightMar.2010

11thEditionFeb.02,2016

Page 2: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

2

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

nGeneralSpecifications1. Basic MeasurementObject/SensorType•pH/Oxidation-reductionPotential(pH/ORP)(analog

sensor)•Conductivity(SC)•InductiveConductivity(ISC)•DissolvedOxygen(DO)•pH/Oxidation-reductionPotential(pH/ORP)(digital

sensor)Note: Theavailablemeasurementobjectdependsona

sensormoduleinstalledontheanalyzer. AnalyzerStructureModulestructure

CompositionofAnalyzerOne(1)HousingassemblyOne(1)ortwo(2)Sensormodules

CombinationofSensorModulewhentwomodulesareinstalledCombinationsoftwosamesensormodulesareavailable; pH/ORPandpH/ORP(analogsensor) SCandSC DOandDO

2. Measurement2-1. pH/Oxidation-reductionPotential(pH/ORP)

withanalogsensors InputSpecificationDualhighimpedanceinput(≥1012Ω)

InputRangepH: -2to16pH(withoption/K:0to14pH)ORP: -1500to1500mVrH: 0to100rHTemperature:

Pt1000: -30to140ºCPt100: -30to140ºC6.8k: -30to140ºCPTC10k: -30to140ºCNTC8k55: -10to120ºC3kBalco: -30to140ºCPTC500: -30to140ºC

OutputRangepH: min.span1pH max.span20pHORP: min.span100mV max.span3000mVrH: min.span2rH max.span100rHTemperature:min.span25ºC max.span170ºC

Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)pH

Linearity: ±0.01pHRepeatability:±0.01pHAccuracy:±0.01pH

ORPLinearity: ±1mVRepeatability:±1mVAccuracy:±1mV

TemperaturewithPt1000,6.8k,PTC10k,NTC8k55,3kBalco,

PTC500Linearity: ±0.3ºCRepeatability:±0.1ºCAccuracy:±0.3ºC

withPt100Linearity: ±0.4ºCRepeatability:±0.1ºCAccuracy:±0.4ºC

2-2. Conductivity(SC) InputSpecificationTwoorfourelectrodesmeasurementwithsquarewaveexcitation,usingmax60m(200ft)cable(WU40/WF10)andcellconstantsfrom0.005to50.0cm-1

InputRangeConductivity:

min.: 0µS/cmmax.: 200mSx(Cellconstant) (overrange2000mS/cm)

Resistivity:min.: 0.005kΩ/(Cellconstant)max.: 1000MΩxcm

Temperature:Pt1000: -20to250ºCPt100: -20to200ºCNi100: -20to200ºCNTC8k55: -10to120ºCPb36(JISNTC6k): -20to120ºC

OutputRangeConductivity: min.0.01µS/cm max.2000mS/cm(max90%zero

suppression)Resistivity: min.0.001kΩxcm max.1000MΩxcm(max90%zero

suppression)Temperature: min.span25ºC max.span270ºC

Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)Conductivity

2µSxKcm-1to200mSxKcm-1

Accuracy:±0.5%F.S.1µSxKcm-1to2µSxKcm-1

Accuracy:±1%F.S.Resistivity

0.005kΩ/Kcm-1to0.5MΩ/Kcm-1

Accuracy:±0.5%F.S.0.5MΩ/Kcm-1to1MΩ/Kcm-1

Accuracy:±1%F.S.Temperature

withPt1000,Pb36,Ni100Accuracy:±0.3ºC

withPt100,NTC8k55Accuracy:±0.4ºC

TemperaturecompensationNaCltable: ±1%Matrix: ±3%

Stepresponse: 90%(<2decades)in7secondsNote: "F.S."meansmaximumsettingvalueofanalyzeroutput. "K"meanscellconstant. YOKOGAWAprovidesconductivitysensorsofwhich

cellconstantsare0.1to10cm-1.

Feb.02,2016-00

Page 3: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

3

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

2-3. InductiveConductivity(ISC) InputSpecificationCompatiblewiththeYokogawainductiveconductivityISC40serieswithintegratedtemperaturesensor:NTC30korPt1000.

InputRangeConductivity: 0to2000mS/cmat25ºCreference

temperature.Temperature: -20to140ºCCablelength: max.60meterstotallengthoffixedsensor

cable+WF10(J)extensioncable. Influenceofcablecanbeadjusted

bydoinganAIRCALwiththecableconnectedtoadrycell.

OutputRangeConductivity:

min.span:100µS/cmmax.span: 2000mS/cm(max90%zero

suppression)Temperature: min.span25ºC max.span160ºC

Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)(Outputspanis0-100µS/cmormore)Conductivity:

Linearity: ±(0.4%F.S.+0.3µS/cm)Repeatability:±(0.4%F.S.+0.3µS/cm)

Temperature:±0.3ºCStepresponse: 90%(<2decades)in8secondsNote: "F.S."meansmaximumsettingvalueofanalyzer

output.2-4. DissolvedOxygen(DO) InputSpecificationTheFLXA21acceptsoutputfrommembranecoveredDissolvedOxygensensors.ThesesensorscanbeGalvanictype,wherethesensorgeneratesitsowndrivingvoltageorPolarographictype,wherethesensorusesexternaldrivingvoltagefromtheconverter.Theinputrangeis0to50µAforGalvanicsensorsand0to1microAforPolarographicsensors.Fortemperaturecompensation,theFLXA21acceptsPt1000(DO30sensor)andNTC22kelements(OXYFERMandOXYGOLDsensors).

InputRangeDO30sensor: DissolvedOxygen:0to50mg/l(ppm) Temperature: -20to150ºC Note:ProcesstemperatureforDO30is0to40ºC

Hamiltonsensors:Oxyferm:

Measurementrange:10ppbto40ppmTemperaturerange: 0to130ºC

OxygoldG:Measurementrange:2ppbto40ppmTemperaturerange: 0to130ºC

OxygoldB:Measurementrange:8ppbto40ppmTemperaturerange: 0to100ºC

OutputRangeDOconcentration:

mg/l(ppm):min.: 1mg/l(ppm)max.: 50mg/l(ppm)

ppb:min.: 1ppbmax.: 9999ppb

%saturation:min.: 10%max.: 600%

Temperature:min.span25ºCmax.span170ºC

Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)Performanceinppmmode:

Linearity: ±0.05ppmor±0.8%F.S.,whicheverisgreater

Repeatability:±0.05ppmor±0.8%F.S.,whicheverisgreater

Accuracy:±0.05ppmor±0.8%F.S.,whicheverisgreater

Performanceinppbmode:Linearity: ±1ppbor±0.8%F.S.,whicheveris

greaterRepeatability:±1ppbor±0.8%F.S.,whicheveris

greaterAccuracy:±1ppbor±0.8%F.S.,whicheveris

greaterTemperature

Linearity: ±0.3ºCRepeatability:±0.1ºCAccuracy:±0.3ºC

Note: "F.S."meansmaximumsettingvalueofanalyzeroutput.

2-5. pH/Oxidation-reductionPotential(pH/ORP)withdigitalsensor,FU20FpH/ORPSENCOMSensor

InputSpecificationBi-directionaldigitalcommunication(RS-485)

betweenFU20FandFLXA21 InputRange(dependingonFU20F)pH: 0to14pHORP: -1500to1500mVrH: 0to100rHTemperature:-10to105ºC

OutputRangepH: min.span1pH max.span20pHORP: min.span100mV max.span3000mVrH: min.span2rH max.span100rHTemperature:min.span25ºC max.span170ºC

Feb.02,2016-00

Page 4: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

4

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

3. Electrical OutputSignalGeneral: Oneoutputof4-20mADC Note:Tolerance:±0.02mA

Bi-directionalHARTdigitalcommunication,superimposedonmA(4-20mA)signal

Outputfunction: LinearorNon-linear(21-steptable)Burnoutfunction:(NAMUR43exceptISC)

WithoutHART/PH201G: Down:3.6mA (signal:3.8to20.5mAforpH/ORP,SC

andDO) (signal:3.9to20.5mAforISC) Up:22mAWithHART/PH201G: Down:3.6mAforpH/ORP,SCandDO Down:3.9mAforISC (signal:3.8to20.5mAforpH/ORP,SC

andDO) (signal:3.9to20.5mAforISC) Up:22mA

PowerSupplyNominal24VDClooppoweredsystem

One(1)Sensormodule(1input): 16to40VDC(forpH/ORP(analog

sensor),SCandDO) 17to40VDC(forISC) 21to40VDC(forpH/ORPSENCOM

sensor)Two(2)Sensormodules(2inputs): 22.8to40VDC(forpH/ORP(analogsensor),

SCandDO)Note: WhentheFLXA21isusedinthemulti-dropmodeof

HARTcommunication,theoutputsignalischangedfrom12.5mADCto4mADCjustafterthepoweristurnedon.Enoughpowersupplyfortheinstrumentsistobeprovided.

MaximumLoadResistancepH/ORP(analogsensor),SCandDO: RefertotheFigure1.ISCandpH/ORPSENCOMsensor: RefertotheFigure2.

Messagelanguage: 12(English,Chinese,Czech,French,

German,Italian,Japanese,Korean,Polish,Portuguese,RussianandSpanish)

Oneanalyzerhasall12languages.Note: Descriptionforaselectionoflanguageand

languagenamesarewritteninEnglish.Note: OnlyEnglishalphabetandnumericareavailable

foratagnumber,anadditionaldescriptionforeachvalueonthedisplayscreenandpasswords.

Note: Onlyformessagelanguageonthescreen,12languagesareprovided.

4. Mechanicalandothers HousingCase: •Plastic(Polycarbonate) •Stainlesssteelwithoutpainting •Stainlesssteelwithepoxycoating •Stainlesssteelwithurethanecoating •Stainlesssteel+highanti-corrosioncoatingCasecolorandfinish:

Color: Silvergray(equivalenttoMunsell3.2PB7.4/1.2)

(forplasticcase,stainlesssteelcaseswithcoating)

Finish: Electropolishing(forstainlesssteelcasewithoutpainting)

Window: Polycarbonate(flexible)Windowframeforstainlesssteelcases: Polycarbonate,color:silvergray

(equivalenttoMunsell3.2PB7.4/1.2)Protection:IP66(exceptCanada),Type4X(except

Canada),Type3S/4X(Canada) PlateMainnameplate:insidecasecoverRegulationplate: onthecaseoutside

CableandTerminalCablesize:

Outerdiameter:6to12mm(suitableforM20cablegland)3.4to7mm(groundingcableforplasticcase)

Terminalscrewsize:M4 torqueofscrewup:1.2N•mWireterminal: Pinterminal,ringterminalandspade

terminalcanbeusedforanalyzer’spowersupplyterminalsandsensorterminals.

Forthegroundingterminalonthestainlesssteelcase,ringterminalshouldbeused.

Pinterminal:pindiameter:max.1.9mm Ringandspadeterminal:width:max.7.8mm

Feb.02,2016-00

250

600

1000

0 24.718

V - 11.50.022

R =

1295

1617 22.8 4040Voltage (V)

2-sensor measurement

Load

resi

stan

ce (Ω

)

Digital Communication Range (HART)

Figure1 SupplyVoltageandLoadResistance forpH/ORP(analogsensor),SCandDO

250304

600516

1000

0 24.718

V - 11.50.022

R =

1295

17 4022.86 40

Except SENCOM

Voltage (V)

Load

resi

stan

ce (Ω

)

Digital Communication Range (HART)

18.2 21

Figure2 SupplyVoltageandLoadResistance forISCandpH/ORPSENCOMsensor

DisplayLCDwithatouchscreen: Black/White:213x160pixelsContrastadjustmentavailableonthetouchscreen

Page 5: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

5

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

CableEntryPlasticcase:

1-Sensormeasurement: 3holes, M20cableglandx3pcs, Sleevex1pc(forgroundingcableline)2-Sensormeasurement: 4holes, M20cableglandx4pcs, Sleevex1pc(forgroundingcableline)

Stainlesssteelcase: 3holes, M20cableglandx3pcs Closeupplugx1pcNote: Cableglandandplugaredeliveredwithananalyzer,

butnotassembledintotheanalyzer. MountingMountinghardware(option): •Universalmountingkit(Note) •Pipeandwallmountinghardware •PanelmountinghardwareNote: Thiskitcontainsthepipeandwallmounting

hardwareandthepanelmountinghardware.Hood(option): •Stainlesssteel •Stainlesssteelwithurethanecoating •Stainlesssteelwithepoxycoating

StainlessSteelTagPlateWhentheadditionalcode“/SCT”withatagnumberisspecified,thetagplateonwhichthetagnumberisinscribedisdeliveredwiththeanalyzer.Tagplateishangingtype.

ConduitAdapterUsingoptionaladapter •G1/2(quantity:4) •1/2NPT (quantity:4) •M20x1.5 (quantity:4)Theseconduitadaptersaredeliveredwithananalyzer,butnotassembledintotheanalyzer.

SizeofHousingCasePlastic: 144(W)x144(H)x151(D)mm(without

cablegland)Stainlesssteelcase: 165(W)x165(H)x160(D)mm(without

cablegland) WeightApprox.1kg(Plastichousing)Approx.2kg(Stainlesssteelhousing)

AmbientOperatingTemperature-20to+55ºC

StorageTemperature-30to+70ºC

Humidity10to90%RHat40ºC(Non-condensing)

DocumentFollowingdocumentsaredeliveredwithananalyzer;

Papercopy:Start-upManual writteninEnglishSafetyPrecautions writteninEnglish

CD-ROM:Start-upManual writteninEnglishUser'sManual writteninEnglish

SafetyRegulationsManual forEuropeanregion writtenin25languagesGeneralSpecifications writteninEnglishTechnicalInformation forHARTCommunication writteninEnglishUserSettingTable of5kindsofmeasurement/sensortype writteninEnglish

RegulatoryComplianceSafety: UL61010-1 UL61010-2-030 CAN/CSAC22.2No.61010-1 CAN/CSA-C22.2No.61010-2-030 EN61010-1 EN61010-2-030EMC: EN61326-1ClassA,Table2(Forusein

industriallocations) EN61326-2-3 AS/NZSCISPR11 RCM:EN55011ClassA,Group1 KoreaElectromagneticConformity

StandardClassA한국 전자파적합성 기준

Installationaltitude:2000morlessCategorybasedonIEC61010:I(Note1)PollutiondegreebasedonIEC61010:2(Note2)Note1:Installationcategory,calledover-voltagecategory,

specifiesimpulsewithstandvoltage. Equipmentwith"CategoryI"(ex.twowire

transmitter)isusedforconnectiontocircuitsinwhichmeasuresaretakentolimittransientover-voltagestoanappropriatelylowlevel.

Note2:Pollutiondegreeindicatesthedegreeofexistenceofsolid,liquid,gasorotherinclusionswhichmayreducedielectricstrength.Degree2isthenormalindoorenvironment.

Intrinsicsafetyandnonincendive(suffixcodeType:-CB,-CD,-DD):

ATEXIntrinsicsafetyapprovalApplicablestandard

ExplosiveAtmospheresEN60079-0:2012/A11:2013Equipment-

GeneralrequirementsEN60079-11:2012Equipmentprotectionby

Intrinsicsafety“i”Typeofprotection

II1GExiaIICT4Ga Group:II Category:1G T4:forambienttemperature:–20to55ºC Atmospherepressure:80kPa

(0.8bar)to110kPa(1.1bar)IECExIntrinsicsafetyapproval

ApplicablestandardIEC60079-0:2011 Part0:Equipment-

GeneralrequirementsIEC60079-11:2011 Part11:Equipment

protectionbyIntrinsicsafety“i”Typeofprotection

ExiaIICT4Ga T4:forambienttemperature:–20to55ºC Atmospherepressure:80kPa

(0.8bar)to110kPa(1.1bar)

Feb.02,2016-00

Page 6: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

6

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

FMIntrinsicsafetyandnonincendiveapprovalApplicablestandard

Class-3600:2011ApprovalStandardforElectricEquipmentforuseinHazardous(Classified)LocationsGeneralRequirement

Class-3610:2010ApprovalStandardforIntrinsicallySafeApparatusandAssociatedApparatusforUseinClassI,II,andIII,Division1,Hazardous(Classified)Locations

Class-3611:2004NonincendiveElectricalEquipmentforUseinClassIandII,Division2andClassIII,Divisions1and2,Hazardous(Classified)Locations

Class-3810:2005ElectricalEquipmentforMeasurement,ControlandLaboratoryUse

ANSI/NEMA250:2014 EnclosuresforElectricalEquipment(1000VoltsMaximum)

ANSI/ISA60079-02013 Part0:GeneralRequirements

ANSI/ISA60079-112014Part11:Equipmentprotectionbyintrinsicsafety“i”

TypeofprotectionClassI,Division1,GroupsA,B,CandD

(IntrinsicallySafe)ClassI,Division2,GroupsA,B,CandD

(Nonincendive)ClassI,Zone0,GroupIIC,inHazardous

(Classified)Locations(IntrinsicallySafeAExiaIIC)

ClassI,Zone2,GroupIIC,inHazardous(Classified)Locations(Nonincendive)

Forallprotectiontype, T4:forambienttemperature:-20to55°CAtmospherepressure:80kPa(0.8bar)to

110kPa(1.1bar)

CSAIntrinsicsafetyandnonincendiveapprovalApplicablestandard

CAN/CSAC22.2No.94-M1991 SpecialPurposeEnclosures

CAN/CSAC22.2No.157-2 IntrinsicallySafeandNon-

incendiveEquipmentforUseinHazardousLocations

C22.2No213-M1987 Non-incendiveElectricalEquipmentforUseinClassI,Division2HazardousLocations

CAN/CSA-E60079-0-07 Electricalapparatusforexplosivegasatmospheres-Part0:Generalrequirements

CAN/CSA-E60079-11-02Electricalapparatusforexplosivegasatmospheres-Part11:Intrinsicsafety“i”

Typeofprotection(C22.2)ClassI,Division1,GroupsA,B,CandD

(IntrinsicallySafe)ClassI,Division2,GroupsA,B,CandD

(Nonincendive)Forallprotectiontype, T4:forambienttemperature:-20to55°CAtmospherepressure:80kPa(0.8bar)to

110kPa(1.1bar)AmbientHumidity:0to100%(Nocondensation)

Typeofprotection(E60079)ExiaIICT4 T4:forambienttemperature:-20to55°CAtmospherepressure:80kPa(0.8bar)to

110kPa(1.1bar)AmbientHumidity:0to100%(Nocondensation)

NEPSIIntrinsicsafetyapprovalApplicableStandard

GB3836.1-2010 Explosiveatmospheres-Part1:Equipment-Generalrequirements

GB3836.4-2010 Explosiveatmospheres-Part4:Equipmentprotectionbyintrinsicsafety“i”

GB3836.20-2010Explosiveatmospheres-Part20:Equipmentwithequipmentprotectionlevel(EPL)Ga

TypeofprotectionExiaIICT4GaT4:forambienttemperature:-20°Cto55°CAtmospherepressure:80kPa(0.8bar)to

110kPa(1.1bar)

Feb.02,2016-00

Page 7: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

7

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

ElectricalParameters(Exia)Eachhousingassembly(basemodule)andeachsensormodulearerespectivelycertificated.Inputparametersofsensormodulemeetoutputparametersofhousingassembly.

Housingassembly

Inputparameters

Supplyandoutputcircuit(terminals+and-): Ui,Vmax =30V Ii,Imax =100mA Pi,Pmax =0.75W Ci =13nF Li =0mH (Linearsource)

Outputparameters

Measuringmoduleinputcircuit(CN2orCN3onBackboard) UoVt,Voc =13.65V Io,It,Isc =50mA Po =0.372W Co,Ca =80nF Lo,La =7.7mH

pH/ORPmodule,SCmodule,andDOmodule

Inputparameters

Ui,Vmax =13.92V Ii,Imax =50mA Pi,Pmax =0.374W Ci =40nF Li =2.9mH

Outputparameters

Sensorinputcircuit(pH:terminals11through19,SC:terminals11through16,DO:terminals11through18) UoVt,Voc =11.76V Io,It,Isc =116.5mA Po =0.3424W Co,Ca =100nF Lo,La =1.7mH

ISCmodule

Inputparameters

Ui,Vmax =13.92V Ii,Imax =50mA Pi,Pmax =0.374W Ci =40nF Li =7.7mH

Outputparameters

Sensorinputcircuit(terminals11through17) UoVt,Voc =11.76V Io,It,Isc =60.6mA Po =0.178W Co,Ca =100nF Lo,La =8mH

SENCOMSensormodule

Inputparameters

Ui,Vmax =13.92V Ii,Imax =50mA Pi,Pmax =0.374W Ci =40nF Li =7.2mH

Outputparameters

Sensorinputcircuit(terminals82,83,84,86and87) UoVt,Voc =5.36V Io,It,Isc =106.16mA Po =0.1423W Co,Ca =31µF Lo,La =0.45mH

ATEXandIECEx ControlDrawing(for4-20mAtype)

Measuring module 1

Measuring module 2

Sensor 1

Sensor 2

Associated Apparatus

Housing AssemblyFLXA21/FLXA202 Analyzer

Supply

Supply

Non-hazardous AreaHazardous Area

Supply +, Supply – Ui: 30 V Ii: 100 mA Pi: 0.75 W Ci: 13 nF Li: 0 mH

-

+

-

+

MeasuringModule1,2TypeofMeasuringModule

pH,SC,DO ISC SENCOMUo 11.76V 11.76V 5.36VIo 116.5mA 60.6mA 106.16mAPo 0.3424W 0.178W 0.1423WCo 100nF 100nF 31μFLo 1.7mH 8mH 0.45mH

SpecificConditionsofUse-WhentheenclosureofFLXA21/202Analyzerismadeofaluminumalloy,andwhenitisusedinanexplosiveatmosphererequiringequipmentofCategory1GorEPLGa,theAnalyzermustbeinstalledinsuchawaythat,evenintheeventofrareincidents,anignitionsourceduetoimpactfrictionsparksisexcluded.

-Precautionsshallbetakentominimizetheriskfromelectrostaticdischargeofnon-metallicpartsoftheenclosureorpaintedpartsofthemetallicenclosure.

Notes:1. Theassociatedapparatusmustbealinearsource.2.MeasuringModule2isnotnecessarilyinstalled.

AsforISCmoduleandSENCOMmodule,onlyonemoduleispermittedtobeinstalledatatime.

3.Sensor1andSensor2maybesimpleapparatusorintrinsicallysafeapparatus.

4.WARNING–POTENTIALELECTROSTATICCHARGINGHAZARD–SEEUSER’SMANUAL

FM ControlDrawing(for4-20mAtype)

Class I, Division 1, Groups A, B, C, D, orClass I, Zone 0, Group IICorClass I, Division 2, Groups A, B, C, D, orClass I, Zone 2, Group IIC,

Temperature Class: T4

Measuring Module 1

Measuring Module 2

Sensor 1

Sensor 2

Voc (or Uo) ≤ UiIsc (or Io) ≤ Ii (only for IS)Po ≤ Pi (only for IS)Ca (or Co) ≥ Ci + CcableLa (or Lo) ≥ Li + Lcable

Associated Apparatus withEntity Parameters orNIFW parameters

Housing AssemblyFLEXA Series Analyzer

Supply

Supply

Unclassified LocationHazardous (Classified) Location

Supply +, Supply – Ui: 30 V Ii: 100 mA (only for IS) Pi: 0.75 W (only for IS) Ci: 13 nF Li: 0 mH

-

+

-

+

Feb.02,2016-00

Page 8: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

8

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

MeasuringModule1,2TypeofMeasuringModule

pH,SC,DO ISC SENCOMUo 11.76V 11.76V 5.36VIo 116.5mA 60.6mA 106.16mAPo 0.3424W 0.178W 0.1423WCo 100nF 100nF 31μFLo 1.7mH 8mH 0.45mH

SpecificConditionsofUse-Precautionsshallbetakentominimizetheriskfromelectrostaticdischargeofnon-metallicpartsandpaintedpartsoftheenclosure.Whentheequipmentusedinhazardouslocations,avoidanyactionwhichgenerateelectrostaticdischargesuchasrubbingwithadrycloth.

-InthecasewheretheenclosureoftheanalyzerismadeofAluminum,ifitismountedinZONE0,itmustbeinstalledsuch,thatevenintheeventofrareincidents,ignitionsourcesduetoimpactandfrictionsparksareexcluded.

Notes:1. Thisdrawingreplacestheformercontroldrawing

IKE039-A12.2.Norevisiontothisdrawingwithoutpriorapprovalof

FM.3. InstallationmustbeinaccordancewiththeNational

ElectricCode(NFPA70),ANSI/ISARP12.06.01andrelevantlocalcodes.

4. TheassociatedapparatusmustbealinearsourcewhichisFM-approved.

5.Controlequipmentconnectedtotheassociatedapparatusmustnotuseorgeneratemorethan250Va.c.r.m.sord.c.

6. Thecontroldrawingoftheassociatedapparatusmustbefollowedwheninstallingtheequipment.

7.MeasuringModule2maynotbeinstalled.AsforISCmoduleandSENCOMmodule,onlyonemoduleispermittedtobeinstalledatatime.

8. ForDivision1/Zone0installation,Sensor1andSensor2maybesimpleapparatus,orintrinsicallysafeapparatuswiththefollowingentityparameters.ForDivision2/Zone2installation,theymaybeequipmentsuitableforDivision2/Zone2,orsimpleapparatus,ornonincendivefieldwiringapparatuswiththefollowingnonincendivefieldwiringparameters.

Ui(orVmax)≥Uo Ii(orImax)≥Io Pi≥Po Ci≤Co–Ccable Li≤Lo–Lcable9. ForDivision2/Zone2installation,general-purpose

powersupplymaybeusedifawiringmethodsuitableforDivision2,butotherthannonincendivefieldwiring,istaken.

10.WARNING–POTENTIALELECTROSTATICCHARGINGHAZARD-WHENTHEEQUIPMENTUSEDINHAZARDOUSLOCATIONS,AVOIDANYACTIONWHICHGENERATEELECTROSTATICDISCHARGESUCHASRUBBINGWITHADRYCLOTH.

11.WARNING–INTHECASEWHERETHEENCLOSUREOFTHEANALYZERISMADEOFALUMINUM,IFITISMOUNTEDINZONE0,ITMUSTBEINSTALLEDSUCH,THATEVENINTHEEVENTOFRAREINCIDENTS,IGNITIONSOURCESDUETOIMPACTANDFRICTIONSPARKSAREEXCLUDED

12.WARNING–SUBSTITUTIONOFCOMPONENTSMAYIMPAIRINTRINSICSAFETYANDSUITABITLITYFORDIVISION2/ZONE2.

CSA ControlDrawing(for4-20mAtype)

Measuring module 1

Measuring module 2

(Refer to Note 7)

Sensor 1

Sensor 2

Housing AssemblySupply -

Supply +

→ Non Hazardous LocationHazardous Location ←

(*)

Refer to NoteSafety Barrier

-

+

Intrinsically SafeGroup IIC, Zone 0Class I, Division 1

Non-incendiveClass I, Division 2,Groups A, B, C, D

2 wire analyzer

Not use safety barrier but CSA certified equipment use in Non-incendive

Note: The measuring module on this drawing means the sensor module on this General Specifications.

Electricalparameters(Intrinsicallysafe)HousingAssembly

Supplyandoutputcircuit(terminals+and-) Ui(Vmax)=30V,Ii(Imax)=100mA,

Pi(Pmax)=0.75W,Ci=13nF,Li=0mHMeasuringmoduleinputcircuit(CN2orCN3on

Backboard) Uo(Vt,Voc)=13.65V,Io(It,Isc)=50mA,

Po=0.372W,Co(Ca)=80nF,Lo(La)=7.7mH

pHmodule,SCmoduleandDOmodule Ui(Vmax)=13.92V,Ii(Imax)=50mA,

Pi(Pmax)=0.374W,Ci=40nF,Li=2.9mHSensorinputcircuit(terminals11through19) Uo(Vt,Voc)=11.76V,Io(It,Isc)=116.5mA,

Po=0.3424W,Co(Ca)=100nF,Lo(La)=1.7mH

ISCmodule Ui(Vmax)=13.92V,Ii(Imax)=50mA,

Pi(Pmax)=0.374W,Ci=40nF,Li=7.7mHSensorinputcircuit(terminals11through17) Uo(Vt,Voc)=11.76V,Io(It,Isc)=60.6mA,

Po=0.178W,Co(Ca)=100nF,Lo(La)=8mH

Installationrequirementsbetweenhousingassemblyandsafetybarrier

Uo≤UiIo≤IiPo≤PiCo≥Ci+CcableLo≥Li+LcableVoc≤VmaxIsc≤ImaxCa≥Ci+CcableLa≥Li+LcableUo,Io,Po,Co,Lo,Voc,Isc,CaandLaareparametersofbarrier.

ElectricalParameters(Nonincendive)HousingAssembly

Supplyandoutputcircuit(terminals+and-) Ui(Vmax)=30V,Ci=13nF,Li=0mHMeasuringmoduleinputcircuit(CN2orCN3on

Backboard) Uo(Vt,Voc)=13.65V,Io(It,Isc)=50mA,

Co(Ca)=80nF,Lo(La)=7.7mHpHmodule,SCmoduleandDOmodule Ui(Vmax)=13.92V,Ci=40nF,Li=2.9mH

Feb.02,2016-00

Page 9: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

9

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Feb.02,2016-00

Sensorinputcircuit(terminals11through19) Uo(Vt,Voc)=11.76V,Io(It,Isc)=116.5mA,

Co(Ca)=4uF,Lo(La)=4.5mHISCmodule Ui(Vmax)=13.92V,Ci=40nF,Li=7.7mHSensorinputcircuit(terminals11through17) Uo(Vt,Voc)=11.76V,Io(It,Isc)=60.6mA,

Co(Ca)=4uF,Lo(La)=19mHNoteforIntrinsicallySafeInstallation:

1: Inanysafetybarrierused,theoutputcurrentmustbelimitedbyaresistor“R”suchthatIo=Uo/RorIsc=Voc/R.

2: ThesafetybarriermustbeCSAcertified.3: Inputvoltageofthesafetybarriermustbelessthan

250Vrms/Vdc4: Whenusingnonisolationbarrierconnect(*)toIS

earthingsystem.5: pHandSCSensor(s)areofapassivetypeto

beregardedas‘simpleapparatus’sameas06ATEX0218X,06ATEX0219,IECExKEM06.0052X,FM3028779,06ATEX0220X,06ATEX0221,IECExKEM06.0053Xortheoneindividuallycertifiedwithrelevantparameters.

6: ISCSensor(s)areISC40Sof00ATEX1067Xortheoneindividuallycertifiedwithrelevantparameters.

7: DOSensor(s)areofapassivetypetoberegardedas‘simpleapparatus’ortheoneindividuallycertifiedwithrelevantparameters.

8: Measuringmodule2maynotmounted.AsforISCmodule,onlyonecanbemounted.

9: InstallationshouldbeinaccordancewithCanadianElectricalCodePartIandLocalElectricalCode.

10:DonotalterdrawingwithoutauthorizationfromCSA.NoteforNonincendiveInstallation:

1: Theparameterforsensorinputcircuitmustbetakenintoaccountwheninstalled.

2: InstallationshouldbeinaccordancewithCanadianElectricalCodePartIandLocalElectricalCode.

3: DonotalterdrawingwithoutauthorizationfromCSA.NEPSI ControlDrawing(for4-20mAtype)

Measuring module 1

Note: The measuring module on this drawing means the sensor module on this General Specifications.

Measuring module 2

Sensor 1

Sensor 2

(*)

(Note 1)Safety BarrierHousing Assembly

Supply -

Supply +

→ Non Hazardous LocationHazardous Location ←

-

+

Electricaldataareasfollows; MaximumVoltage(Ui)=30V MaximumCurrent(Ii)=100mA MaximumPower(Pi)=0.75W InternalCapacitance(Ci)=13nF InternalInductance(Li)=0mHNote1:Theoutputcurrentmustbelimitedbyaresistor“R”

suchthatImaxout=Uz/R(linearsource).Note2:SafetybarriercertifiedbyNEPSIshouldbeused.Note3:Whenusingnonisolationbarrier,connect(*)toIS

earthingsystem.Note4:Measuringmodule2isinstalledwhenrequired.

Whenmeasuringinductiveconductivity,onlyonemodulecanbeinstalled.

5. DigitalCommunication KindofDigitalCommunication•HARTorPH201GdedicateddistributorNote: Onlyonekindofdigitalcommunicationisavailable

foroneanalyzer. OutputValueParameter(HART)Fourvalueparameters(measuredvalues)areavailableforonedigitalcommunication.•For1-sensormeasurement,theseparametersare

measuredvalues.•For2-sensormeasurement,refertothenextitem.

DigitalCommunicationof2-SensorMeasurement(HART)Evenwhentwosensormodulesareinstalled,onlyonedigitalcommunicationisavailablefor2-sensormeasurement.Fourvalueparameterscanbeselectedfromthefollowings; Measuredvaluesoftwosensors Calculateddataof2-sensormeasurement Redundantsystemoutput

SpecificContactOutputwithdedicateddistributor,modelPH201G(StyleB)Thedistributor,modelPH201G,isdesignedtoconnectwiththe2-WireAnalyzer.Thisdistributorsuppliesdrivepowertotheanalyzerandreceivessimultaneously4-20mADCsignalfromtheanalyzer.Thissignalisconvertedto1-5VDCsignalinthedistributor.Thisdistributoralsoreceivesdigitalsignalssuperimposedonthe4-20mADCsignal,andprovidescontactoutputsInput/Outputsignal:

Numberofavailabledrive/signalpoint: 1Outputsignal:1-5VDC(2points)(Note)Loadresistance: 2kΩorless(1-5VDCoutput)Isolationsystem: Loopisolationtype

Note: Twooutputsignalsforoneanalyzer'sanalogoutputareprovided.Two1-5VDCoutputsignalsaresame.

Contactoutput:Contactrating: 250VAC,maximum100VA 220VDC,maximum50VAHoldcontactoutput: NCcontact,normallyenergized Contactcloseswhenpowerisoffor

duringHoldsituation.Failcontactoutput: NCcontact,normallyenergized Contactcloseswhenpowerisoffor

duringFail/Warningconditions.Washcontactoutput: NOcontact Contactclosesduringwashcycles.

RegulatoryComplianceKoreaElectromagneticConformityStandardClassA한국 전자파적합성 기준

Page 10: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

10

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

6. Model&SuffixCodes

Model Suffixcode Optioncode Description

FLXA21 ······················································································· ···················· 2-Wire Analyzer

Powersupply -D ···················· Always-D

Housing -P-S-U-E-W

·····················································································

PlasticStainlesssteelStainlesssteel+urethanecoatingStainlesssteel+epoxycoatingStainlesssteel+highanti-corrosioncoating(Note20)

Display -D ···················· Anti-glareLCD

Type -AA-AB-AD-AG-AQ-AR-CB-CD-EA-ES-EG-EQ-ER-DD

··············································································································································································································································

Generalpurpose(Note19)GeneralpurposeforCE,RCMGeneralpurposeforCSAGeneralpurposeforKCGeneralpurposeforEACwithPA(Note13)GeneralpurposeforEAC(Note14)ISforATEX,IECEx(Note11)ISforFM,(CSA(pending))(Note17)ISforCSA,NEPSI(Note5)ATEX,IECExforSENCOMsensor(Note9)(Note19)ISforKOSHA(Note12)ISforEACExwithPA(Note15)ISforEACEx(Note16)NIforFM,(CSA(pending))(Note18)

1stinput -P1-C1-C5-D1-S1

····································································································

pH/ORP(Note7)Conductivity(SC)Inductiveconductivity(ISC)Dissolvedoxygen(DO)pH/ORP(SENCOMsensor)(Note8)

2ndinput(Note1) -NN-P1-C1-D1

················································································

WithoutinputpH/ORP(Note7)Conductivity(SC)Dissolvedoxygen(DO)

Output(Note10) -A ···················· 4-20mA+HART

— -N ···················· Always-N

Languageset(Note2) -LA ···················· Englishand11languages

Country(Note3) -N-J

········································

GlobalexceptJapanJapan

— -NN ···················· Always-NN

Option Mountinghardware

Hood

TagplateConduitadapter

Measurementlaw

/UM/U/PM/H6/H7/H8/SCT/CB4/CD4/CF4/K

Universalmountingkit(Note4)PipeandwallmountinghardwarePanelmountinghardwareHood,stainlesssteelHood,stainlesssteel+urethanecoatingHood,stainlesssteel+epoxycoatingStainlesssteeltagplateConduitadapter(G1/2x4pcs)Conduitadapter(1/2NPTx4pcs)Conduitadapter(M20x1.5x4pcs)WithMeasurementLawcertificate(Note6)

Notes:1: Whena2ndinputisselected,onlythesamekindofthe1stinputisavailable. Forexample,whena1stinputis“-P1”,the2ndinputmustbethesame“-P1”. ThecombinationofISCandISCisnotavailable.And,thecombinationofSENCOMsensorandSENCOMsensorisnot

available,either.2: Theselanguagesaremessagelanguagesontheanalyzer’sdisplay. OneanalyzerhasEnglishand11languages. Alllanguagesareasfollows;English,Chinese,Czech,French,German,Italian,Japanese,Korean,Polish,Portuguese,

RussianandSpanish.3: WhenananalyzerisusedinJapan,itmustmeettheJapaneseMeasurementLaw. OnlySIunitsmustbeusedontheanalyzeranditsdocumentsinJapan.4: Theuniversalmountingkitcontainsthepipeandwallmountinghardware(/U)andthepanelmountinghardware(/PM).5: Thetype“-EA”isintrinsicsafetyofCSAandNEPSI,andnon-incendiveofCSA.TemperatureclassesareT4.Thetype“-EA”

cannotbeusedwithSENCOMsensor.

Feb.02,2016-00

Page 11: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

11

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

6: TheanalyzerwithJapaneseMeasurementLawcertificateisavailableonlyforthefollowingmodel; FLXA21-D-[Housingcode]-D-AA-P1-NN-A-N-LA-J-NN/[optioncodeexcept/K]/K OnlyonepHmeasurementwithananalogsensoriscertified.Theoutputsignalof4-20mAiscertified.HARTcommunication

isnotcertified.7: ThisinputistobecomefromananalogpH/ORPsensor.8: Whentheanalyzerisconnectedwiththedigitalsensor,FU20FpH/ORPSENCOMSensor,onlythefollowingmodelisavailable; 2ndinput:Withoutinput(-NN) Option: except“/K”9: Thetype“-ES”isintrinsicsafetytypeofATEX,IECExforSENCOMsensor.TemperatureclassisT4.10: TheFLXA21hasotheroutputtypesof“FOUNDATIONFieldbus”communication(suffixcode:-F)and“PROFIBUSPA”

communication(suffixcode:-P).RefertoGS12A01A02-71EandGS12A01A02-72E.11: Thetype“-CB”intrinsicsafetytypeofATEXandIECExcanbeusedwithSENCOMsensor.TemperatureclassisT4.12: Thetype“-EG”intrinsicsafetytypeofKOSHAforKorea.TemperatureclassisT4.13: Thetype“-AQ”isGeneralpurposetypeofEACwithPattanApprovalforRussia.14: Thetype“-AR”isGeneralpurposetypeofEACforKazakhstanandBelarus.15: Thetype“-EQ”intrinsicsafetytypeofEACwithPattanApprovalforRussia.TemperatureclassisT4.16: Thetype“-ER”intrinsicsafetytypeofEACforKazakhstanandBelarus.TemperatureclassisT4.17: Thetype“-CD”intrinsicsafetytypeforFM.TemperatureclassisT4.18: Thetype“-DD”nonincendivetypeforFM.TemperatureclassisT4.19: Forthetype“-AA”and“-ES”,themainnameplateisinsideofthehousing.Thissuffixcodewillbeterminated.

Therefore,incaseof“-AA”pleaseselect“-AB”,“-AD”or“-AG”forgeneralpurpose. Incaseof“-ES”pleaseselect“-CB”forATEX,IECExforSENCOMsensor.

20: Thehousingwithstainlesssteel+highanti-corrosioncoatingisnotavailableforthetype“-EA”.

nDimensionsandMountingPlasticHousing

44.9137

120.2

151

22.5 80

4-M6 depth 5(For sensor 2 cable)

For sensor 1 cable

For power supply

For grounding cable

137

80144

24.7

26.5

26.5

141144

FB4_01.ai

Unit: mm

ConduitAdapter(Optioncode:/CB4,/CD4,/CF4)

Adapter

49(1.93")

G1/2 screw (/CB4), 1/2 NPT screw (/CD4)M20x1.5 screw (/CF4)

Approx.55(2.2")

Packing

Unit: mm(inch)Nut

F0204.ai

Feb.02,2016-00

Page 12: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

12

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

(Note)Theuniversalmountingkit(/UM)containsthepipeandwallmountinghardware(/U)andthepanelmountinghardware(/PM).

Panelmountinghardware(Optioncode:/PM,/UM)

2-M5 length 35

185

178

100

195

135

Pan

el th

ickn

ess

1 to

12

Spacing panel cutout

4-M6 *

138+10

138

+1 0

FB4_03.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Wallmountinghardware(Optioncode:/U,/UM)

144 164

13

10070

144

224

200

50

15

For wall mounting3-ø10 holes

4-M6 *

FB4_02.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Note: Thewallonwhichtheanalyzerismountedshouldbestrongenoughtobeartheweightofmorethan8kg.

Pipemountinghardware(Optioncode:/U,/UM)

151

100

200

5017

4

144(205)

Pipe mounting (Horizontal) Pipe mounting (Vertical)

Pipe 50A(ø60.5)

M8 U-bolt

4-M6 *

FB4_04.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Feb.02,2016-00

Page 13: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

13

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

Stainlesssteelhood(Optioncode:/H6,/H7,/H8)184 220

(70)(100)

9144

144

7257

(50)

199

FB4_05.ai

Unit: mm

StainlessSteelHousing

80165

13722.244

150160

32.5

4-M6 depth 5(For sensor 2 cable)For sensor 1 cable

For power supply

Plate(Only with coating)

80137

165

25.3

26.5

FB4_06.ai

Unit: mm

ConduitAdapter(Optioncode:/CB4,/CD4,/CF4)

Adapter

49(1.93")

G1/2 screw (/CB4), 1/2 NPT screw (/CD4)M20x1.5 screw (/CF4)

Approx.55(2.2")

Packing

Unit: mm(inch)Nut

F0204.ai

Feb.02,2016-00

Page 14: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

14

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

(Note)Theuniversalmountingkit(/UM)containsthepipeandwallmountinghardware(/U)andthepanelmountinghardware(/PM).

Panelmountinghardware(Optioncode:/PM,/UM)

2-M5 length 35

185

178

100

195

134

Pan

el th

ickn

ess

1 to

12

Spacing panel cutout

4-M6 *Plate(Only with coating)

138+10

138

+1 0FB4_08.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Wallmountinghardware(Optioncode:/U,/UM)

165 17313

70100

4-M6 *

165

200

5015

234.

5

For wall mounting3-ø10 holes

FB4_07.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Note: Thewallonwhichtheanalyzerismountedshouldbestrongenoughtobeartheweightofmorethan8kg.

Feb.02,2016-00

Page 15: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

15

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

Pipemountinghardware(Optioncode:/U,/UM)

160

100

200

5018

4.5

165(214)

Pipe mounting (Horizontal) Pipe mounting (Vertical)

Pipe 50A(ø60.5)

M8 U-bolt

4-M6 *

FB4_09.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Stainlesssteelhood(Optioncode:/H6,/H7,/H8)184

220

(70)(100)

9165

165

7257

(50)

199

FB4_10.ai

Unit: mm

Feb.02,2016-00

Page 16: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

16

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

nWiringDiagrams

(Plastic housing)(Stainless steel housing)

*2 *2

*4

*5

*6

*6

n n n n n n n n n

n n n n n n n n n

+–

*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.*2: Connect the analyzer to ground. (Class D ground: 100 ohm or less) The way of connecting the grounding cable varies depending on the plastic housing and stainless steel

housing. In the case of the plastic housing, connect the grounding cable to the terminal of the power module inside,

and in the case of the stainless steel housing, connect the grounding cable to the terminal of the housing. Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing. Although, on the stainless steel housing, the ground terminal symbol is (protective ground), the ground is

really functional ground. *3: This line is connected to a distributor or 24V DC power supply.*4: Terminal numbers for each sensor module are shown below.*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive

conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.*6: The terminal box may be necessary depending on the sensor cable length and the distance between the

analyzer and the sensor. The SENCOM sensor is to be connected directly to the analyzer without a terminal box.*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.

*1

WTB10 or BA10Terminal box FLXA21 2-Wire Analyzer

Sen

sor

WTB10 or BA10Terminal box

Sen

sor

PH Module

SC Module

ISC Module

DO Module

SENCOM Module

161519171318141211

PH

NC1817141315161211

DO

15161417131211

ISC

NC161514131211

SC

8786NC848382NC

SENCOM

Power supply

Output 1 (1-5V DC)

20 to 130V DCor80 to 138V AC, 47 to 63Hz

Power supply20 to 130V DCor80 to 138V AC, 47 to 63Hz

HOLD WASHFAIL

Output 2 (1-5V DC)

Output 1 (1-5V DC)

Output 2 (1-5V DC)

Case of DistributorPH201G (Style B)

Case of DistributorSDBT

(*3)

*3

*7

*7

(*3)

A+

– CMN CMNB

CD

FH

L N

A+

–+

–B

1+

–2

FH

L N

ba

dc

fe

Feb.02,2016-00

Page 17: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

17

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

nInquirySpecificationsSheetforFLXA212-WireAnalyzer

Makeinquiriesbyplacingcheckmarks()inthepertinentboxesandfillingintheblanks.

1. GeneralInformationCompanyname ContactPerson; Department; Plantname; Measurementlocation; Purposeofuse; Indication,Recording,Alarm,Control

2. MeasurementConditions(1)Processtemperature; to Normally [°C](2)Processpressure; to Normally [kPa](3)Flowrate; to Normally [l/min](4)Flowspeed; to Normally [m/s](5)Slurryorcontaminants;No,Yes(6)Nameofprocessfluid; (7)Componentsofprocessfluid; (8)Others;

3. InstallationSite(1)Ambienttemperature; to [°C](2)Location;Outdoors,Indoors(3)Others;

4. Requirements1stInput; pH/ORP(analogsensor)Conductivity(SC)Inductiveconductivity(ISC) Dissolvedoxygen(DO)pH/ORP(digitalsensor,FU20F)2ndInput; With(sameas1stInput)Without

4.1 pH/ORP(analogsensor) 1st Input(1)Measuringrange;pH0to14ORP to mV (2)Transmissionoutput;4to20mADCpHORPTemperature(3)Systemconfigurationselection;Electrode,Holder,pHConverter,Cleaningsystem,Terminalbox, Accessories(4)Electrodecablelength; 3m,5m,7m,10m,15m,20m, m(5)Electrodeoperatingpressure; 10kPaorless,Morethan10kPa(6)Typeofholder; Guidepipe,Submersion,Flow-through,Suspension,Angledfloatingball, Verticalfloatingball(7)Cleaningmethod;Nocleaning,Ultrasoniccleaning,Jetcleaning,Brushcleaning(8)Sampletemperature;-5to105°C,-5to100°C,-5to80°C(9)Others;

2ndInput(1)Measuringrange;pH0to14ORP to mV (2)Transmissionoutput;4to20mADCpHORPTemperature(3)Systemconfigurationselection;Electrode,Holder,pHConverter,Cleaningsystem,Terminalbox, Accessories(4)Electrodecablelength; 3m,5m,7m,10m,15m,20m, m(5)Electrodeoperatingpressure; 10kPaorless,Morethan10kPa(6)Typeofholder; Guidepipe,Submersion,Flow-through,Suspension,Angledfloatingball, Verticalfloatingball(7)Cleaningmethod;Nocleaning,Ultrasoniccleaning,Jetcleaning,Brushcleaning(8)Sampletemperature;-5to105°C,-5to100°C,-5to80°C(9)Others;

Feb.02,2016-00

Page 18: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

18

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

4.2 Conductivity

1st Input(1)Measuringrange; (2)Transmissionoutput;4to20mADC(3)Detector/sensor; SC4AJ Twoelectrodesystem(0.02cm-1)Twoelectrodesystem(0.1cm-1) SC8SG Twoelectrodesystem(0.01cm-1)Twoelectrodesystem(10cm-1), Fourelectrodesystem(10cm-1) SC210G Twoelectrodesystem(0.05cm-1)Twoelectrodesystem(5cm-1)(4)Detector/sensormountingmethod; SC4AJ Adaptermounting,Weldingsocket,Weldingclamp SC8SG Screw-in,Flow-through SC210G Screw-in,Flange,Flow-through,Screw-inwithgatevalve(5)Electrodecablelength; SC4AJ3m,5m,10m,20m SC8SG 5.5m,10m,20m SC210G 3m,5m,10m,15m,20m(6)Others;

2ndInput(1)Measuringrange; (2)Transmissionoutput;4to20mADC(3)Detector/sensor; SC4AJ Twoelectrodesystem(0.02cm-1)Twoelectrodesystem(0.1cm-1) SC8SG Twoelectrodesystem(0.01cm-1)Twoelectrodesystem(10cm-1), Fourelectrodesystem(10cm-1) SC210G Twoelectrodesystem(0.05cm-1)Twoelectrodesystem(5cm-1)(4)Detector/sensormountingmethod; SC4AJ Adaptermounting,Weldingsocket,Weldingclamp SC8SG Screw-in,Flow-through SC210G Screw-in,Flange,Flow-through,Screw-inwithgatevalve(5)Electrodecablelength; SC4AJ3m,5m,10m,20m SC8SG 5.5m,10m,20m SC210G 3m,5m,10m,15m,20m(6)Others;

4.3 Inductiveconductivity(1)Measuringrange; (2)Transmissionoutput;4to20mADC(3)Systemconfigurationselection;ISC40GJSensor,Holder,Converter,BA20Terminalbox, WF10JExtensioncable(4)Sensormountingmethod; ISC40FDJImmersionholder,ISC40FFJFlow-throughholder, ISC40FSJDirectinsertionadapter(5)ISC40GJSensorcablelength; 5m,10m,15m,20m(6)WF10JExtensioncablelength; 5m,10m,20m,30m,40m(7)Others;

Feb.02,2016-00

Page 19: Model FLXA21 2-Wire Analyzerweb-material3.yokogawa.com/FLXA21-GS-11.pdf · 02/02/2016  · • pH/Oxidation-reduction Potential (pH/ORP) (digital sensor) Note: Theavailablemeasurementobject

19

All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E

4.4 Dissolvedoxygen

1st Input(1)Measuringrange; 0to50mg/L (2)Transmissionoutput;4to20mADC(3)Systemconfigurationselection;Electrode,Holder,Converter,Cleaningsystem, Terminalbox,Maintenancepartsset,Calibrationset(4)Electrodecablelength; 3m,5m,10m,15m,20m(5)Typeofholder; Guidepipe,Submersion,Flow-through,Suspension, Angledfloatingball,Verticalfloatingball(6)Cleaningmethod; Nocleaning,Jetcleaning(7)Others;

2ndInput(1)Measuringrange; 0to50mg/L (2)Transmissionoutput;4to20mADC(3)Systemconfigurationselection;Electrode,Holder,Converter,Cleaningsystem, Terminalbox,Maintenancepartsset,Calibrationset(4)Electrodecablelength; 3m,5m,10m,15m,20m(5)Typeofholder; Guidepipe,Submersion,Flow-through,Suspension, Angledfloatingball,Verticalfloatingball(6)Cleaningmethod; Nocleaning,Jetcleaning(7)Others;

4.5 pH/ORP(digitalsensor,FU20F)(1)Measuringrange;pH0to14ORP to mV (2)Transmissionoutput;4to20mADCpHORPTemperature(3)Systemconfigurationselection;Electrode,Holder,pHConverter,Cleaningsystem,Accessories(4)Electrodecablelength; 3m,5m,10m,20m, m(5)Electrodeoperatingpressure; 10kPaorless,Morethan10kPa(6)Typeofholder; Guidepipe,Submersion,Flow-through,Suspension,Angledfloatingball, Verticalfloatingball(7)Cleaningmethod;Nocleaning,Jetcleaning(8)Sampletemperature;-5to105°C,-5to100°C,-5to80°C(9)Others;

Subjecttochangewithoutnotice.Feb.02,2016-00