DESCRIPTION
Técnicas Químicas del Microlab 200TRANSCRIPT
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALBUMIN CATLOG BT 10.002.00 METODOLOGY COLORIMETRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME ALBUMIN TYPE FINAL POINT FILTER 620 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 . CONTROL - 1 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALPHA-AMYLASE CATLOG BT 11.001.00 METODOLOGY KINETIC COLORIMETRICDETERMINATIONS 60
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME AMY TYPE KINETIC FILTER 405 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 6950 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.800 ABS HIGH 2.300 . CONTROL - 1 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALKALINE PHOSPHATASE
CATLOG BT 11.005.00 METODOLOGY KINETIC COLORIMETRICDETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME ALP TYPE KINETIC FILTER 405 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 2757 SAMPLE BLANK NO RETARD 60 READING TIME 60 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT ALT/TGP CATLOG BT 11.008.00 METODOLOGY KINETIC UV DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME ALT/TGP TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 1746 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.800 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT AST/TGO CATLOG BT 11.007.00 METODOLOGY KINETIC UV DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME AST/TGO TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 1746 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.800 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT BILIRUBIN DIRECT CATLOG BT 10.003.00 METODOLOGY COLORIMETRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME BDI TYPE FINAL POINT FILTER 546 TEMPERATURE 25 UNIT mg/dL REAG BLANK NO ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.220
............- ABS HIGH 0.330 FACTOR 1.0 SAMPLE BLANK SI RETARD 5 READING TIME 2 REACTION VOL 900 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT BILIRUBIN TOTAL CATLOG BT 10.003.00 METODOLOGY COLORIMETRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME BTO TYPE FINAL POINT FILTER 546 TEMPERATURE 25 UNIT mg/dL REAG BLANK NO ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.220
............- ABS HIGH 0.330 FACTOR 1.0 SAMPLE BLANK SI RETARD 5 READING TIME 2 REACTION VOL 900 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CLCIUM ASX CATLOG BT 12.002.00 METODOLOGY ARSENAZO III DETERMINATIONS 200
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CALCIUM TYPE FINAL POINT FILTER 620 TEMPERATURE 25 UNIT mg/dL REAG BLANK Si ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.220
............- ABS HIGH 0.330 FACTOR 1.0 SAMPLE BLANK NO RETARD 5 READING TIME 2 REACTION VOL 900 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CLCIUM CATLOG BT 12.001.00 METODOLOGY CPC DETERMINATIONS 200
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix the Reagent A + Reagent B in a proportion 1:1 and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CALCIUM TYPE FINAL POINT FILTER 578 TEMPERATURE 25 UNIT mg/dL REAG BLANK Si ABS - LOW STANDARD
ABS - HIGH 1 CALIB - NAME @ ............- NR 1 ............- CONC 0.000 ............- REP 2.300 ............- ABS LOW 1.00 ............- ABS HIGH No FACTOR 10 SAMPLE BLANK 2 RETARD 500 READING TIME 1 REACTION VOL * REPLICATES * REF. LOW 0.000 REF. HIGH 2.300 . * CONTROL - 1 * LOT * TARGET VALUE * LOW LIMIT HIGH LIMIT
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CK NAC CATLOG BT 11.002.00 METODOLOGY KINETIC UV DETERMINATIONS 40
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CK NAC TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 4127 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CK MB CATLOG BT 11.003.00 METODOLOGY KINETIC UV DETERMINATIONS 40
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CK MB TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 8254 SAMPLE BLANK NO RETARD 5 READING TIME 120 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHLORIDE CATLOG BT 12.003.00 METODOLOGY COLORIMTRIC DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CHLORIDE TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mEq/L REAG BLANK Si ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1.00 SAMPLE BLANK NO RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHOLESTEROL CATLOG BT 10.004.00 METODOLOGY ENZIMTIC
COLORIMTRIC DETERMINATIONS 1000
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CHOL TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CHOLESTEROL HDL CATLOG BT 10.005.00 METODOLOGY PRECIPITATION +
ENZIMATIC COLORIMTRIC
DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME HDL CHOL TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT CREATININE CATLOG BT 10.007.00 METODOLOGY COLORIMTRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME CRE TYPE TWO POINTS FILTER 505 TEMPERATURE 37 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 12 READING TIME 60 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT GAMMA GT CATLOG BT 11.006.00 METODOLOGY KINTIC COLORIMTRICDETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME GGT TYPE KINETIC FILTER 405 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 2757 SAMPLE BLANK NO RETARD 60 READING TIME 60 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT GLUCOSE CATLOG BT 10.008.00 METODOLOGY ENZIMTIC
COLORIMTRIC DETERMINATIONS 1000
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME GLU TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT IRON CRX CATLOG BT 12.004.00 METODOLOGY COLORIMETRIC DETERMINATIONS 200
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME IRON TYPE FINAL POINT FILTER 620 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT LACTATE DEHYDROGENASE
CATLOG BT 11.004.00 METODOLOGY KINTICO UV DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME LDH TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO
CALIB NAME NO
FACTOR 8095 SAMPLE BLANK NO RETARD 60 READING TIME 60 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT MAGNSIUM CATLOG BT 12.007.00 METODOLOGY MAGON DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME MAG TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT PHSPHORUS UV CATLOG BT 12.006.00 METODOLOGY Molibidate UV DETERMINATIONS 100
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME PUV TYPE FINAL POINT FILTER 340 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT PROTEN TOTAL CATLOG BT 10.009.00 METODOLOGY COLORIMTRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME PRO TYPE FINAL POINT FILTER 546 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT PROTEN URINRY CATLOG BT 10.016.00 METODOLOGY COLORIMTRIC DETERMINATIONS 200
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME PRO-U TYPE FINAL POINT FILTER 600 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT TRIGLICRIDES CATLOG BT 10.010.00 METODOLOGY ENZIMTIC
COLORIMTRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME TGR TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URA UV CATLOG BT 10.012.00 METODOLOGY KINTICO UV DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME UUV TYPE TWO POINTS FILTER 340 TEMPERATURE 37 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 20 READING TIME 60 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.600 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URIC ACID - MONOREAGENT
CATLOG BT 10.001.00 METODOLOGY ENZYMATIC
COLORIMETRIC DETERMINATIONS 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME UAC TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUCT URIC ACID - TWOREAGENT
CATLOG BT 10.014.00 METODOLOGY ENZYMATIC
COLORIMETRIC DETERMINATIONS 250
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME UAC TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *
-
Equipment Applications
MICROLAB 200
TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]
PRODUTO UREA CATLOGO BT 10.013.00 METODOLOGIA ENZYMATIC
COLORIMETRIC N DETERMINAES 500
- The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:
REAGENT PREPARATION Please follow the Instructions of the kit. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00
To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS None
# Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00
TEST NAME URE TYPE FINAL POINT FILTER 578 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD
............- NR 1
............- CONC @
............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *