Page 1: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Methods for sampling emissions ofDioxins, Mercury and Heavy Metals

AQE workshop 13 March 2013

Page 2: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• Outline of the session:

• Sampling using the Standard Reference Method

• Continuous sampling

AQE workshop 13 March 2013

Page 3: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Dioxins

AQE workshop 13 March 2013

Furans

Page 4: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• PCDDs 7 of the 75 PCDDs are specifically toxic• PCDFs 10 of 135 congeners have “dioxin-like”

properties• PCBs 12 have “dioxin-like” properties• The reference congener is the most toxic dioxin

2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD) whichper definition has a Toxic Equivalence Factor of one.

• The group does NOT include 1,4 dioxin C4H4O2

What are dioxins and dioxin-likecompounds?

AQE workshop 13 March 2013

Page 5: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• EN1948 Standard reference method for emissionsmonitoring

Sampling according to EN1948-1:2006Determination of the Mass Concentrationof PCDDs/PCDFs and dioxin-like PCBs.Sampling of PCDDs/PCDFs

AQE workshop 13 March 2013

Page 6: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• Modified method for PCBs

Sampling according to EN1948-4:2007Determination of the Mass Concentrationof PCDDs/PCDFs and dioxin-like PCBs.Sampling and analysis of dioxin-like PCBs

AQE workshop 13 March 2013

Page 7: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• May be produced in emissions from– Waste incinerators– Cement plants– Steel mills– Forest fires, burning cars

• Dioxins are persistent organic pollutants (POPs)– A suspect human carcinogens (US-EPA)– Long lifetime in the environment– Are difficult to break down

Why monitor dioxin emissions?

AQE workshop 13 March 2013

Page 8: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

EN1948-1 Sample train diagram

AQE workshop 13 March 2013

Page 9: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

EN1948-1 sample train using Isostack G4

AQE workshop 13 March 2013

Page 10: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

EN14385 sample train with Isostack G4

AQE workshop 13 March 2013

Page 11: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Isostack G4 automatic isokinetic samplerwith heated probe and box

AQE workshop 13 March 2013

Page 12: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• Duct survey - suggests nozzle size based upon gasvelocity and temperature profile

• Calculates probe insertion depths based upon stackdimensions

• During sampling - maintains isokinetic velocity at thenozzle

• Alters the operator when the probe must be moved• Logs sampling data

Features of an automatic isokineticsampler

AQE workshop 13 March 2013

Page 13: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• Time/date• Stack i.d• Stack Diameter• Gas composition• Water content• Nozzle diameter• # of ports and points• Insertion depths

Stored data – about the stack

AQE workshop 13 March 2013

Page 14: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Stored data – at each point

• Stack temperature –max, min, average

• Probe temperature• Heated box

temperature• Condenser temperature• Ice box temperature• DGM temperature

• Stack gas velocity– max,min, average

• Stack pressure• Sample line pressure• Sample flow rate• Isokinetic deviation• Stack flow rate (actual)• Stack flow rate (ref)

AQE workshop 13 March 2013

Page 15: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Stored data – in summary

• Sample volume at STP• Sample volume at DGM

conditions• Stack differential

pressure• Ambient pressure• DGM encoder impulses• DGM pressure

• Dry and wet actual ductflow rate

• Dry and wet standardflow rate

AQE workshop 13 March 2013

Page 16: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Report transferred as XML file that can bedisplayed in a spreadsheetsample of data displayed here

AVERAGE VALUESTotal Points [#] 13

Velocity v'a [m/sec] 21.089[9.766;22.983]

Stack temperature ta [°C] 57.651[42.755;70.292]

Stack Static Pressure Pastatic [kPa] 102.07[102.027;102.123]

Isokinetic Deviation DI [%] 0.064[-1.000;0.866]

Probe temperature tprobe [°C] 28.122[-72.414;111.067]

Box temperature tbox [°C] 24.057[7.290;40.922]

Aux1 temperature taux1 [°C] 147.142[135.334;156.379]

Aux2 temperature taux2 [°C] 201.133[181.497;217.734]

Stack Differential Pitot Pressure dPpitot [Pa] 314.017[134.667;385.312]

Ambient Pressure Pamb [kPa] 102.142[102.142;102.142]

DUCT FLOW RATE

Dry actual QVa [m3/h] 115705[53579;126089]

Moist actual Q'Va [m3/h] 116874[113257;119747]

Moist standard [Tnorm Pnorm] Q'Vn [m3/h] 97285[93952;102860]

Dry standard [Tnorm Pnorm] QVn [m3/h] 96312[93013;101831]

AQE workshop 13 March 2013

Page 17: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• Uncertainty of data from sampling and analysis hashigh uncertainty. Control of sensors in samplingsystems is crucial

• Each instrument has UKAS traceable annual checksof DGM, pressure and temperature sensors

• In-house checks of sensors periodically• G4 will store these corrections and recalibrations.• Calibration report is exported in XML format

Quality control of sampling systems

AQE workshop 13 March 2013

Page 18: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• DECS systems manufactured by Tecora• Dozens supplied in Europe in the last 5 years• Several plants with multiple units installed• France, Italy, Finland, USA• Incinerators, Steel Plants, Cement Plants

Proposed EN1948-5A standard for long-term sampling ofPCDDs/PCDFs and dioxin-like PCBs.

AQE workshop 13 March 2013

Page 19: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Dioxin Emissions Continuous SamplerControl Unit Sampling Unit

AQE workshop 13 March 2013

Page 20: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Parallel Sampling – two CU and two SU

AQE workshop 13 March 2013

Page 21: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Display to program settings

AQE workshop 13 March 2013

Page 22: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Set sampling duration or volume

AQE workshop 13 March 2013

Page 23: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Display during operation

AQE workshop 13 March 2013

Page 24: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

Sidestream sampling for Hg and heavymetals

AQE workshop 13 March 2013

Page 25: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals

TECORA Isostack G4 and DECS – Stack Emissions Sampling

• Automatic sampler used by stack team for QAL2checks

• Comprehensive report of all sampling parameters

• Long term sampling of dioxins• Same high standards of monitoring and reporting of

sampling parameters

Summary of Tecora capabilities

AQE workshop 13 March 2013

Page 26: Methods for sampling emissions of Dioxins, Mercury and Heavy Metals