IntroductionThank you for selecting the pocket pH tester. This manual provides a step-by-step guide to help you operate the meter, please carefully read the followinginstructions before use.
Inserting the Batteries
insert batteries into the battery compartment.— Remove the battery compartment lid.
— Insert the batteries into the battery compartment, note polarity.
— Replace the battery compartment lid and turn clockwise until it is tight.
KeypadThe tester includes 3 multifunction keys, names and symbols describe functionkey controls.
Take out the three G13A button cell batteries from packaging. Follow the stepsbelow to
— Power the tester ON/OFF.— Freezes the measured values on the display, press the key again to resume measuring.— In calibration mode, exits calibration and returns to measurement mode.
— Press and hold the key to enter calibration mode.— Press the key again to select the pH buffer option.
— Confirms the calibration or option.
KEY FUNCTION
MEAS/HOLD
CAL
ENTER
49G961/49G962/49G964
01
49G961/49G962/49G964 Pocket pH Tester
03
— Remove the electrode protective cap from tester.
— If the electrode bulb dries out, soak the electrode in storage solution or tap water for at least 15 minutes. Do not use ,
—
—
Prior to Use
moduledistilled or deionized water
it will shorten the life of sensor.
Preparation of pH Standard Buffer SolutionsThe tester is packaged with three pH buffer packets required for calibration.
Open the pH7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until reagent is completely dissolved.
Preparation of pH4.01 and 10.01 standard buffer solutions are the same as above. Prepared standard buffer solutions should be stored in hermetically sealed glass containers.
Electrode Bulb
49G961/49G962/49G964 49G961/49G962/49G964
04
— Press MEAS key to turn on the tester, the meter shows measured values.— — If
—
—
Press and hold the CAL key the display shows “CAL 7-4”. If necessary, press CAL key again, the tester will show “CAL 10-7”.
Choose the pH buffer option you want to use, press ENTER key to confirm. The tester shows first calibration point, pH7.00.
4. Dip the electrode into the pH7.00 standard buffer solution, the end of the electrode must be completely immersed into the calibration solution. Stir the tester gently.
Power On/Off
Press and hold the MEAS key for 5 seconds, the tester will turn off.you do not press any key for 8 minutes, the tester will automatically turn
off to conserve energy.
CalibrationThe tester must be calibrated prior to first use or new electrode replaced.
To ensure accuracy, regular calibration is recommended.Do not reuse calibration solution after calibration, contaminants in solution
will affect the calibration and eventually the accuracy of the measurement.
1. Rinse the pH electrode with distilled water, press MEAS key to turn on the tester.2. for 3 seconds,
3.
HoldMeas
05
49G961/49G962/49G964
5. Press ENTER key to confirm. W
Dip the electrode into the pH4.01 or 10.01 standard buffer solution. Stir the tester gently.
7. Press ENTER key to confirm. W
ait for the measured value to stabilize, the reading will flash 3 times indicating that first calibration point is completed.
The tester will automatically show “CAL 4” or “CAL 10” (Depending on the option selected in the step 2).
6. Rinse the electrode with distilled water.
ait for the measured value to stabilize, the reading will flash 3 times, “CAL” indicator will disappear from display, the tester returns to measurement mode. Calibration is completed.
EXIT THE CALIBRATION: During the calibration, If you want to exit calibration mode. Press MEAS key, the tester will immediately return to measurement mode.
or
49G961/49G962/49G964
06
pH MeasurementRinse the electrode thoroughly with distilled water,
ait for the to stabilize, record
Electrode ReplacementWhen the tester fails to calibrate or gives fluctuating readings for calibration standards, you need to replace the electrode module.1. Twist the electrode collar counter clockwise, pull the old electrode module away from the tester.
2. Align the slot on the new electrode module, gently push the module into the tester.
3. Twist the electrode collar clockwise until it is tight.
1. press MEAS key to turn on the tester.2. Dip the electrode into the sample solution, stir the tester gently.3. W reading the measured value on the display.
Hold Function1. During the measurement, press HOLD key,
, “HOLD” indicator appears on the display.2. Press the key again to release the measured reading, you can continue to take measurement.
Electrode Care and Maintenance— For best results, please always keep the sensor’s glass bulb wet.— Ensure that the sensor is thoroughly washed with distilled water after each
use.— If you do not use the tester for long periods, please store the sensor with
electrode storage solution. DO NOT use de-ionized or distilled water.
Installation is completed.
the tester immediately freezes current displayed value
Slot
0807
49G961/49G962/49G964
Troubleshooting
Tester Included
×
Appendix: How to prepare electrode storage solution1. Dissolve 223.6 grams of potassium chloride reagent (KCL) in 1 liter distilled water.2. Pour the 50mL of KCL solution into beaker.3. Pour the 50mL of pH4.01 standard buffer solution into the same beaker.4. Stir the solution until the solution has thoroughly mixed, preparation is done.
Specifications
— pH Buffer Pouches (pH4.01/7.00/10.01)— Batteries (G13A 3)— OIPM
Model
pH Range
Accuracy
Calibration Points
pH Buffer Options
Operating Temperature
49G961/49G962/49G964
0.0~14.0pH
±0.1pH
2 points
USA (pH4.01, 7.00, 10.01)o o0~60 C, 32~140 F
Power Requirement
Dimensions
Weight
3 1.5V “ Batteries× G13A”
185(L) 40(Dia.)mm×
100 grams
LCD DISPLAY CAUSE
— Electrode dried out. Soak the electrode in tap water for 10 minutes. — Measured value is out of range. Check the electrode whether clogged, dirty or broken.
module
— Incorrect buffer used or contaminated buffer solution. Use fresh pH buffer solutions for calibration.— Electrode is broken. Replace the electrode .module
— Keypad is not working properly. Replace the batteries.
09
49G963/49G966/49G967/49G965/49G968/49G969Pocket pH Tester
10
49G963/49G966/49G967/49G965/49G968/49G969
IntroductionThank you for selecting the pocket pH tester. This manual provides a step-by-step guide to help you operate the meter, please carefully read the followinginstructions before use.
KeypadThe tester includes 3 multifunction keys, names and symbols describe functionkey controls.
The tester is equipped with a clear LCD display used to show measured values,mode indicators and unit indicators.
Display
Me
as
Ho
ld
— Power the tester ON/OFF.— Freezes the measured values on the display, press the key again to resume measuring.— In calibration mode, exits calibration and returns to measurement mode.
— Press the key to enter calibration mode.— Press and hold the key to enter the system menu.— In setup mode, press the key to select default options.
— Confirms the calibration or displayed option.
KEY FUNCTION
MEAS/HOLD
CAL
ENTER
HOLD
DESCRIPTION
Indicates tester is in the measurement mode
Indicates tester is in the calibration mode
Indicates tester is in SETUP mode
Indicates the measured value has been frozen
Indicates temperature compensation is enabled
MEAS
CAL
SETUP
ATC
INDICATOR
11
Inserting the BatteriesBefore using the tester, insert batteries into the battery compartment, followthe procedure outlined below.—
— Insert two “AAA” batteries into the battery compartment, note polarity.
—
— Installation is completed.
— Remove the electrode protective cap from tester.— If the electrode bulb dries out, soak the electrode in storage solution or tap water for at least 15 minutes. Do not use ,
Twist the electrode collar counter clockwise, pull the electrode module away from the tester.
Align the slot on the electrode module, push the module into the tester.
Twist the electrode collar clockwise until it is tight.
Prior to Use
moduledistilled or deionized water
it will shorten the life of sensor.
+
Electrode Bulb
49G963/49G966/49G967/49G965/49G968/49G969
12
49G963/49G966/49G967/49G965/49G968/49G969
—
—
Power On/Off— Press MEAS key to turn on the tester, the display shows measured values.
— —
Preparation of pH Standard Buffer SolutionsThe tester is packaged with three pH buffer packets required for calibration.
Open the pH7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until reagent is completely dissolved.
Preparation of pH4.01 and 10.01 standard buffer solutions are the same as above. Prepared standard buffer solutions should be stored in hermetically sealed glass containers.
Press and hold the MEAS key for 5 seconds, the tester will turn off.If you do not press any key for 8 minutes, the tester will automatically turn
off to conserve energy.
To disable the auto-off function, please read the SYSTEM SETUP section.
HoldMeas
HoldMeas
49G963/966/967 49G965/968/969
49G963/49G966/49G967/49G965/49G968/49G969
14
1. Press and hold the CAL key for 3 seconds to enter setup menu, the tester shows default pH buffer option.
Press CAL key again to select the available standards, press ENTER key to confirm.
The tester goes to calibration point selection mode, the display shows “2/ CAL” (2
3. Press CAL key to select the number of calibration points, press ENTER key to confirm.
o The tester goes to temperature unit selection mode, the display shows “ C/ UNIT”
o o Press CAL key to select the temperature unit ( C or F), press ENTER key to confirm.
o The tester goes to temperature calibration mode, the display shows “ C/CAL”.
2.
points calibration).
.
4.
13
49G963/49G966/49G967/49G965/49G968/49G969
The tester contains an integrated system menu that allows you to customizeeach displayed option to meet personal preference.
System Menu
PARAMETER OPTIONS
pH Buffer Sets USA Standard (pH4.01, 7.00, 10.01, Default)
NIST Standard (pH4.01, 6.86, 9.18)
Single point calibrationCalibration Point
2 points calibration (Default)
3 points calibration
Degrees Fahrenheit
Temperature Unit Degrees Celsius (Default)
Do not use this function (Default)
Reset Restores the parameters back to default settings
Disables the auto-off function
Auto-Off Function Enables the auto-off function (Default)
Disables the auto-hold function (Default)
Auto-Hold Function Enables the auto-hold function
Temperature Calibration Enters the calibration mode
15
49G963/49G966/49G967/49G965/49G968/49G969
If you do not want to perform temperature calibration, press ENTER key, the tester will go to next option.
5. Press CAL key, the tester enters the temperature calibration mode. The display shows current temperature reading.
6. Press CAL key again to set temperature value, press ENTER key to confirm. The tester shows “YES/HOLD” the auto-hold function is enabled.
— If the auto-hold function is enabled, the tester will automatically sense a stable end-point reading and will freeze it. — If you disable this function, the tester allows you to freeze the reading by pressing the HOLD key.
7. Press CAL key to enable or disable the auto-hold function.
8. Press ENTER key to confirm. The tester shows “YES/OFF”
auto-off function is enabled.
indicating that
indicating that the
49G963/49G966/49G967/49G965/49G968/49G969
16
When the auto-off function is enabled,
9. Press CAL key to enable or disable the auto-off function.
10. Press ENTER key to confirm. The tester shows “NO/RST”
the tester should not be reset.
Reset function will restore the tester back to factory default parameters,calibration values and selected system parameters will be lost or reset.
11. Press CAL key to enable or disable the reset function.
12. Press ENTER key to confirm. The tester returns to measurement mode.Setting is completed.
if you do not press any keys for 8 minutes, the tester will automatically turn off to conserve energy.
indicating that
WARNING: all
Enter
17
49G963/49G966/49G967/49G965/49G968/49G969
allows up to 3 points calibration.
tester can
— — —
Press CAL key the tester shows “pH7.00/CAL1”(or “pH6.86/CAL1”).
4. Dip the pH electrode into the pH7.00 (or pH6.86) standard buffer solution, the end of the electrode must be completely immersed into the calibration solution. Stir the tester gently.
5. Press ENTER key to confirm. W
pH CalibrationDuring pH mode, the tester We recommendthat you perform at least 2 points calibration for high accuracy measurement.The
Single point calibration should only be carried out with pH 7.00 or pH 6.86, otherwise calibration will not be accepted.
automatically recognise and calibrate to following standardbuffer values.
USA Standard: pH4.01, 7.00, 10.01NIST Standard: pH4.01, 6.86, 9.18
The tester must be calibrated prior to first use or new electrode replaced. Toensure accuracy, regular calibration is recommended. Do not reuse calibrationsolution after calibration, contaminants in solution will affect the calibration andeventually the accuracy of the measurement.
Single Point Calibration1. Make sure that you have selected 1 point calibration in the system menu.2. Rinse the pH electrode with distilled water, press MEAS key to turn on the tester.3. ,
ait for the measured value to stabilize, thedisplay shows “END”. Single point calibration is done.
49G963/49G966/49G967/49G965/49G968/49G969
18
2 Points CalibrationMake sure that you have selected 2 points calibration in the system menu.Repeat steps 2 to 5 above. When the first calibration point is completed, the
display will show “CAL2”. The tester prompts you to continue with second point calibration.
Rinse the electrode with distilled water,
ait for the measured value to stabilize, the display shows electrode slope and “END”. Second point calibration is done.
3 Points CalibrationMake sure that you have selected 3 points calibration in the system menu.Repeat steps 2 to 5 above. When the first calibration point is completed, the
display will show “pH4.01/CAL2”. The tester prompts you to continue with second point calibration.
Rinse the electrode with distilled water,
ait for the measured value to stabilize, the display shows electrode slope and “pH10.01/CAL3” (or “pH9.18/CAL3”).
— —
— dip the electrode into the pH4.01 or 10.01 (pH4.01 or 9.18) standard buffer solution. — Press ENTER key to confirm. W
— —
— dip the electrode into the pH4.01 standard buffer solution. — Press ENTER key to confirm. W
Enter
Enter
19
49G963/49G966/49G967/49G965/49G968/49G969
Rinse the electrode with distilled water,
ait for the measured value to stabilize, the display shows electrode slope and “END”. The tester returns to pH measurement mode. Calibration is completed.
EXIT THE CALIBRATION: During the calibration, if you want to exit calibration mode, press MEAS key. The tester will immediately return to measurement mode.
Temperature CalibrationDuring the measurement, if the temperature reading displayed differs fromthat of an accurate thermometer, you need to calibrate the tester.
6. Press MEAS key, the tester returns to measurement mode. Calibration is completed.
— dip the electrode into the pH10.01 (or pH9.18) standard buffer solution. — Press ENTER key to confirm. W
1. Press and hold the CAL key to enter system menu.o o2. Press ENTER key until display shows the “ C/CAL” or “ F/CAL”.
3. Press CAL key to enter the temperature calibration mode.o4. Press CAL key again to set temperature value (Displayed value ±0.5 C).
5. Press ENTER key to confirm, the display shows next option.
Enter
49G963/49G966/49G967/49G965/49G968/49G969
20
pH MeasurementRinse the electrode thoroughly with distilled water,
ait for the to stabilize, record
Electrode ReplacementWhen the tester fails to calibrate or gives fluctuating readings for calibration standards, you need to replace the electrode module.1. Twist the electrode collar counter clockwise, pull the old electrode module away from the tester.
1. press MEAS key to turn on the tester.2. Dip the electrode into the sample solution, stir the tester gently.3. W reading the measured value on the display.
Auto-Hold FunctionWhen the auto-hold function is enabled, the tester will automatically sense astable end-point measured value and freeze it, “HOLD” indicator will appear onthe display. Press HOLD key, the tester resumes measuring.
Manual Hold Function1. During the measurement, press HOLD key,
, “HOLD” indicator will appear on the display.2. Press key again to release the measured reading, you can continue to take measurement.
Electrode Care and Maintenance— For best results, please always keep the sensor’s glass bulb wet.— Ensure that the sensor is thoroughly washed with distilled water after each
use.— If you do not use the tester for long periods, please store the sensor with
electrode storage solution. DO NOT use de-ionized or distilled water.
the tester immediately freezes currently displayed value
49G963/49G966/49G967/49G965/49G968/49G969
22
Specifications
Calibration Points Up to 3 points
pH Buffer Options USA (pH4.01, 7.00, 10.01)
NIST (pH4.01, 6.86, 9.18)
Temperature Compensation o o0~60 C, 32~140 F, Automatic
Auto-Off After 8 minutes from last key press
Operating Temperature o o0~60 C, 32~140 F
LCD Display Custom LCD
Connector 6-pin
Power Requirement 2 1.5V “× AAA” Batteries
Weight 100 grams
Dimensions 185(L) 40(Dia.)mm×
Model 49G963/966/967 49G965/968/969
Temperature Range — o0~60 C
Resolution — o0.1 C
Accuracy — o±1 C
pH Range -1.00~15.00pH -1.00~15.00pH
Resolution 0.01pH 0.01pH
Accuracy ±0.01pH ±0.01pH
21
49G963/49G966/49G967/49G965/49G968/49G969
Installation is completed.
— pH Buffer Pouches (pH4.01/7.00/10.01)— Batteries (AAA 2)— OIPM
2. Align the slot on the new electrode module, gently push the module into the tester.
3. Twist the electrode collar clockwise until it is tight.
Troubleshooting
Appendix: How to prepare electrode storage solution1. Dissolve 223.6 grams of potassium chloride reagent (KCL) in 1 liter distilled water.2. Pour the 50mL of KCL solution into beaker.3. Pour the 50mL of pH4.01 standard buffer solution into the same beaker.4. Stir the solution until the solution has thoroughly mixed, preparation is done.
Tester Included
×
Slot
— Electrode dried out. Soak the electrode in tap water for 10 minutes. — Measured value is out of range. Check the electrode whether clogged, dirty or broken.
module
— Incorrect buffer used or contaminated buffer solution. Use fresh pH buffer solutions for calibration.— Electrode is broken. Replace the electrode .module
LCD DISPLAY CAUSE
— Keypad is not working properly. Replace the batteries.