MAGNETICALLY DRIVEN MICRO-MOVABLE ELECTRODE FOR CELL COUPLING Y. Yamanishi 1* , T. Kawahara 2 , T. Iyanagi 3 , M. Hagiwara 2 , Fumihito Arai 2 1 JST, PRESTO, JAPAN 2 Nagoya University, JAPAN 3 Tohoku University, JAPAN ABSTRACT We developed a novel method of cell coupling on a chip by using magnetically driven microtool (MMT). It is made of nickel and is used as an electrode as well as a cell pusher. The novelty of this device is the electrode itself whose tip part is supported by a spring and is positioned by external magnetic field. We propose a new cell coupling chip which provides larger coupling force of cells by using a physical motion of MMT. The operation of the microelectrode showed that the oocyte is deformed and the donor cell is successfully coupled physically with the oocyte. The contact area be- tween the cells was increased, and which is good for consequent fusion process. The FEM analysis on the electric field density also showed that a high concentration between the membrane of oocyte and donor cell. The proposed method is suited for cell fusion on a chip, and will contribute to high throughput cell manipulation. KEYWORDS: Magnetically Driven Microtool, Cell Manipulation, Cell Coupling, Microelectrode INTRODUCTION Eliminating the manual manipulations problems of contamination, poor success, and low replicability, together with the skill-dependent human element, requires that cell manipulation be automated through the use of disposable biochips in cell manipulations such as embryo cloning and microfertilization. Complex cell manipulation such as microfertiliza- tion is currently done by skilled operators using micromanipulators. The automation of on-chip micromanipulation has been reported to ensure efficient, accurate, high cell-manipulation throughput using magnetic actuators [1][2], microfluid force [3], and dielectrophoresis[4]. We have been developed a chip for automation of embryo cloning (cloning chip) based on a new protocol of nuclear transfer. The conventional design of the coupling of oocyte and donor cell was performed only by a dielectrophoretic force which was generated by opposite electrode mounted on a chip [5]. However, due to the limited force applied for coupling cells, and the size of the oocyte is 10 times as large as the donor cells, the success rate of the continuous fusion process was limited. From the empirical investigation, it is recognized in manual operation using glass capillary that the coupling of cells with physical force increased the success rate of cell fusion. Here we propose a novel device to apply physical force between the cells using movable electrode on a chip. EXPERIMENTAL SETTING Figure 1 shows the concept view of the cell coupling chip. Movable electrode has a spring structure to position its tip part toward the opposite fixed electrode. Figure 2 shows the process flow to fabricate the chip. Nickel electroplating was applied after pattering of Cr/Au, then the MMT with spring structure can be obtained. This MMT was made of Nickel and it is also has a function as electrode. The novelty of the chip is the electrode itself can be moved by external magnet- ic field using a spring structure to use as a cell pusher. Figure 1: Design of cell coupling chip using movable electrode. (oocyte) Microtool Fixed electrode Fixed Spring Structure Movable Electrode Inlet (egg cell) Outlet Inlet (donor) 978-0-9798064-3-8/μTAS 2010/$20©2010 CBMS 905 14th International Conference on Miniaturized Systems for Chemistry and Life Sciences 3 - 7 October 2010, Groningen, The Netherlands

Upload: others

Post on 02-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MAGNETICALLY DRIVEN MICRO-MOVABLE ELECTRODE FOR …O2 Plasma Cover glass Micro-channel Patterned SU-8 Ni-MMT (a) Overall view (b) Magnified view 2 m 300 m FIXED k=8.3 N/m 2 mN Movable

MAGNETICALLY DRIVEN MICRO-MOVABLE ELECTRODE FOR CELL COUPLING

Y. Yamanishi1*, T. Kawahara2, T. Iyanagi3, M. Hagiwara2, Fumihito Arai2 1JST, PRESTO, JAPAN

2Nagoya University, JAPAN 3Tohoku University, JAPAN

ABSTRACT

We developed a novel method of cell coupling on a chip by using magnetically driven microtool (MMT). It is made of nickel and is used as an electrode as well as a cell pusher. The novelty of this device is the electrode itself whose tip part is supported by a spring and is positioned by external magnetic field. We propose a new cell coupling chip which provides larger coupling force of cells by using a physical motion of MMT. The operation of the microelectrode showed that the oocyte is deformed and the donor cell is successfully coupled physically with the oocyte. The contact area be-tween the cells was increased, and which is good for consequent fusion process. The FEM analysis on the electric field density also showed that a high concentration between the membrane of oocyte and donor cell. The proposed method is suited for cell fusion on a chip, and will contribute to high throughput cell manipulation. KEYWORDS: Magnetically Driven Microtool, Cell Manipulation, Cell Coupling, Microelectrode

INTRODUCTION

Eliminating the manual manipulations problems of contamination, poor success, and low replicability, together with the skill-dependent human element, requires that cell manipulation be automated through the use of disposable biochips in cell manipulations such as embryo cloning and microfertilization. Complex cell manipulation such as microfertiliza-tion is currently done by skilled operators using micromanipulators. The automation of on-chip micromanipulation has been reported to ensure efficient, accurate, high cell-manipulation throughput using magnetic actuators [1][2], microfluid force [3], and dielectrophoresis[4].

We have been developed a chip for automation of embryo cloning (cloning chip) based on a new protocol of nuclear transfer. The conventional design of the coupling of oocyte and donor cell was performed only by a dielectrophoretic force which was generated by opposite electrode mounted on a chip [5]. However, due to the limited force applied for coupling cells, and the size of the oocyte is 10 times as large as the donor cells, the success rate of the continuous fusion process was limited. From the empirical investigation, it is recognized in manual operation using glass capillary that the coupling of cells with physical force increased the success rate of cell fusion. Here we propose a novel device to apply physical force between the cells using movable electrode on a chip.

EXPERIMENTAL SETTING

Figure 1 shows the concept view of the cell coupling chip. Movable electrode has a spring structure to position its tip part toward the opposite fixed electrode. Figure 2 shows the process flow to fabricate the chip. Nickel electroplating was applied after pattering of Cr/Au, then the MMT with spring structure can be obtained. This MMT was made of Nickel and it is also has a function as electrode. The novelty of the chip is the electrode itself can be moved by external magnet-ic field using a spring structure to use as a cell pusher.

Figure 1: Design of cell coupling chip using movable electrode.

(oocyte)

Microtool

Fixed electrode

Fixed

Spring Structure

Movable Electrode

Inlet (egg cell) Outlet

Inlet (donor)

978-0-9798064-3-8/µTAS 2010/$20©2010 CBMS 905 14th International Conference onMiniaturized Systems for Chemistry and Life Sciences

3 - 7 October 2010, Groningen, The Netherlands

Page 2: MAGNETICALLY DRIVEN MICRO-MOVABLE ELECTRODE FOR …O2 Plasma Cover glass Micro-channel Patterned SU-8 Ni-MMT (a) Overall view (b) Magnified view 2 m 300 m FIXED k=8.3 N/m 2 mN Movable

・Ni Movable Electrode Fabrication 1.Sputter deposition 3.Electroplating

2.Patterning 4.Etching

・Micro-Channel Fabrication1. Patterning 3. Peeling MMT from the mold

2. Molding 4. Bonding and Wiring

Figure 2: Fabrication processes Figure 3: Fabricated cell coupling chip using electrode

To apply current to the electrode, a wire was connected to the electrode through a hole mounted in a chip and sealed

by silver-based adhesive paste (Figure 3). The wire is completely fixed and the tip part of movable electrode can be moved by the magnetic force and precise positioning is possible without any disturbance from the wire.

The microchannel was fabricated by the conventional photolithography technique as shown in Figure 2. Before bounding the PDMS (polydimethylsiloxane) chip with cover glass by using O2 plasma, the fabricated Nickel electrodes were assembled on a chip. The fixed electrode was assembled following a guide of PDMS columns attached with the PDMS microchannel. The movable electrode was also assembled in microchannel where the tip of the electrode aligned to a level of microchannel as shown in Figure 3.

The FEM analysis of the displacement of the movable electrode showed that the displacement of the tip was more than 240 m and hence the tip can reach to the fixed electrode (Figure 4) at a given applied magnetic force (2 mN). Fig-ures 5 shows the operation of the movable electrode actuated by a permanent magnet below the chip. It was confirmed that the movable electrode was moved to the PDMS wall which has a suction hole to trap an oocyte. Therefore this MMT can be used for coupling oocyte and donor cell. The transportation channel of oocyte and donor cell used the con-ventional design of the chip for coupling by a dielectrophoretic force [5]. This method did not provide enough force for cell coupling, because the size of the oocyte is 10 times as large as the donor cells. By using the movable electrode, we could obtain the additional physical force to increase the success rate of cell fusion.

Patterned SU-8 Ni

Si

Cr/Au

PDMS VV

Cover glassO2 Plasma

Micro-channel Patterned SU-8 Ni-MMT

(a) Overall view

(b) Magnified view

2 m

300 m

Spring Constant k=8.3 N/m FIXED

2 mN

Movable Electrode

Micro- channel

400 m

Magnet Movable Electrode

100 m

Vacuum Hole for Trapping Oocyte

Movable Electrode

Movable Electrode

(b) Position after moving

(a) Default position

Micro- channel

Magnified View

Magnified View

(a) FEM analysis of Mises Stress

Figure 4: FEM analysis of Mises Stress on spring structure of movable electrode. (a) FEM analysis, (b) spring structure of movable electrode.

Figure 5: Operation of movable electrode.

(a) default position (b) position after moving

(b)

Ni

Movable Electrode

Egg cell inlet port

Donor cell inlet port

Egg cell inlet port

Donor cell inlet port Fixed Electrode

PDMS Column

Suction channel

Assist flow

906

Page 3: MAGNETICALLY DRIVEN MICRO-MOVABLE ELECTRODE FOR …O2 Plasma Cover glass Micro-channel Patterned SU-8 Ni-MMT (a) Overall view (b) Magnified view 2 m 300 m FIXED k=8.3 N/m 2 mN Movable

Figure 6 shows the result of the cell coupling experiment using swine oocyte and donor cell. The oocyte is deformed

and the donor cell is successfully coupled physically with the oocyte. It is important to note that the contact area between the cells was increased ,and which is good for consequent fusion process. The FEM analysis on the electric field density also showed that a region of a high concentration was distributed equally between the membrane of oocyte and donor cell (Figure 7). For further work, the design of the electrode and magnitude of the applied voltage can be optimized for the consequent sufficient cell fusion process. It is also important to compare the success rate with the manual operation of cell coupling and fusion with glass capillaries.

CONCLUSION

We succeeded in a novel method of cell coupling on a chip by using MMT. This MMT was made of Nickel and it is also has a function as electrode as well as a cell pusher. The novelty of the chip is the electrode itself can be moved by an external magnetic field using a spring structure. This magnetically micro-movable electrode was fabricated by an electroplating techniques and operated successfully. FEM analysis confirmed its operation on a chip by an external mag-netic force, and also it confirmed that a high concentration of electric field between the membrane of oocyte and donor cell was distributed equally due to increase of the contact area between the deformed oocyte and donor cell. This chip contributes to the high throughput cell manipulation as well as electrical measurement. ACKNOWLEDGEMENTS

This work was supported in part by the Research and Development Program for New Bio-industry Initiatives and JST-SENTAN. REFERENCES [1] Y. Yamanishi, S. Sakuma, K. Onda, F. Arai, “Powerful Actuation of Magnetized Microtools by Focused Magnet-ic

Field for Particle Sorting in a Chip”, Biomed Microdevices, vol. 12, p. 745-752, (2010). [2] M. Hagiwara, T. Kawahara, Y. Yamanishi, F. Arai, “Driving Method of Microtool by Horizontally-arranged Per-

manent Magnets for Single Cell Manipulation”, Applied Physics Letters, vol. 97, pp. 013701-1 -013701-3, (2010). [3] H. C. Zeringue, D. J. Beebe and M. B. Wheeler, “Removal of Cumulus from Mammalian Zygotes Using Microfu-

lidic Techniques”, Biomedical Microdevices, Volume 3, p.219–224, (2001). [4] M. Gel, S. Suzuki, Y. Kimura, O. Kurosawa, B. Techaumnat, H. Oana, and M. Washizu,”Microorifice-Based High-

Yield Cell Fusion on Microfluidic Chip: Electrofusion of Selected Pairs and Fusant Viability”, IEEE Transcation on Nanobioscience, Vol.8, No.4, p.294-299, (2009).

[5] S. Sakuma, Y. Yamanishi, F. Arai et al., “All-in-one Unified Microfluidic Chip for Automation of Embryonic Cell Manipulation,”, The 13th International Conference on Miniaturized Systems for Chemistry and Life Sciences -TAS, p.1883-1885, (2009).

CONTACT *Y. Yamanishi, tel: +81-52-789-5656; [email protected]

Figure 6. Tucking donor cell into oocyte (a) before tucking donor cell, (b) coupling, (c)magnified view of tucking oo-cyte, (d) coupling of oocyte and donor cell (e) after coupl-ing of donor cell

Figure 7. FEM analysis of concentration of electric field (a)entire view (b) point contact of oocyte and donor cell (b) larger area contact due to deforma-

100 m

Oocyte

Donor cell

Movable Electrode

Movable Electrode Movable Electrode Movable Electrode

Oocyte Pushed Oocyte

Oocyte

Donor Cell

Coupled Donor Cell Donor Cell

Donor Cell

Oocyte Oocyte

Movable Electrode

Oocyte

(a)

(b) (c)

(a) (b)

(c) (d) (e)

907