> r I" U. S. Departme nt of Commer ce National Burea u of Standards Research Paper RP1875 Volume 40, Ap ril 1948 Part of the Jo u rnal of Resea r ch of the National Burea u of Standa rds Magnetic Measurement of the Thickness of Composite Copper and Nickel Coatings on Steel By Abner Brenner and Eugenia Kellogg A m ag n etic method has been developed for measuring the t hickn ess of comp os ite coat- ings of cop per an d nickel, el ect rode positpd on steel. Tb e m et hod consists in m eas uring the attractive force between tbe specimen and two permane nt m ag n ets of different st rengths. Th e total t hi ckness of the deposit and the relative th ickness of co pper and nickel l aye rs can be d ete rmined from a set of calibra t ion curv e. . In the plating indu stry composite coatings of copper and nickel are ext ensively used, although they do not generally yield as good a protective value as an equal thiekness of pur e nickel. These composite coatings on steel consist of two lay ers- the first, co pper, which is eleetrodeposited directly on the steel, and the second, nickel, deposited on the co pper. Th e eo pp er is less expensive t han nickel, and it may be readily buffed as an under- coat for bright nickel. The present trend is to- ward the use of heavier eomposite coatings to secure increased prote ct ive value. Th e au tomo- t.i ve industry commonly uses copper-nickel de- posits up to 0.002 in. thick on steel parts. It is t.herefore d esirable to obtain a nondestru ctive met hod of measuring the total thickness and the relative thicknesses of the copper and ni ck el layers such as described in this pap er . An instrument known as the Magne-gage has been used to measur e th e thiclmess of nonmag- netic and of nickel coatings on steeU Thi s in- st rument measur es the attractive force b etween a small permanent magnet and the plat ed sample. This attractive force is dependent upon the t hi ckn ess of the coating and the magnetic proper- t ie s of the eo ating and base mat erial. When a coating is interposed between the magn et and the magnet ic base m aterial, the attractiv e for ce is de- crea se d, provided that the co ating is less magnetic than the base mat erial, the d ecrease being greater the thicker the coating. I Abner Brenner, J. Research NBS 20 ,357 (1938) RP1081. Th i ckness of Composite Coatings The Magne-gage is a spring balance, on one ann of which the magnet is suspended. A helical spring is attached in such a mann er that as the knob is turn ed, the spring is wound up. Th e for ce required to d etach the magn et is a measur e of the attra ction b et ween the magnet and the specimen. The attractive force of the magn et for the sample is read from a dial attached to the turning knob . Th e thiclmess of the coating is determined from this reading with the aid of a calibration curve mad e with coatings of known valu es of thickness. To det ermin e the thickness of composit e coatings, two magnets of different strengths are used. In order to compensate for the differen ce in the attra ctive force of the two magnets, it would be necessary to use two differ- en t gages unless different springs were used or two magn ets were hung at diIfer ent distances from the pivot on the balance arm. For the measurement s her e report ed, the str onger magn et was hung- near er to the pivot. The modified gage is shown in figure 1. Nickel is somewhat magnetic, although not as much so as steel, hen ce the calibration curve fo r the nickel coatings differs from that for copper coatings as copper is practically nonmagn eti c. As can be seen from figure 2, the calibration curve for pur e nickel coatings of given values of thick- ness falls considerably to the left of the curve for pure copper, and is steeper. Th e calibration curv e for any composite coating mu st fall within tbe boundaries of these two cur ves as, for ex- 295

Upload: truongtuyen

Post on 09-Jun-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Magnetic measurement of the thickness of composite copper ... · FIGURE 1. Magne-gagefor measuring thickness of composite coatings. ample, does the curve for equal thickness of cop

>

r

I"

U. S. Department of Commerce National Bureau of Standards

Research Paper RP1875 Volume 40, April 1948

Part of the Journal of Research of the National Bureau of Standards

Magnetic Measurement of the Thickness of Composite Copper and Nickel Coatings on Steel

By Abner Brenner and Eugenia Kellogg

A m agnetic method h as been developed for measurin g t he thickness of compos ite coat­ings of copper and nickel, electrodepositpd on steel. Tbe method co ns ists in measurin g the attractive force betwee n t be specime n and t wo perma nent magnets of differen t strengths. The total thick ness of t he deposit a nd the relat ive th ickness of copper and nickel layers can be determined from a set of calibrat ion curve . .

In the plating industry composite coatings of copper and nickel are extensively used, although they do not generally yield as good a protective value as an equal thiekness of pure nickel. These composite coatings on steel consist of two layers­the first, copper , which is eleetrodeposited directly on the steel, and the second, ni ckel, deposited on the copper . The eopper is less expensive than nickel, and it may be readily buffed as an under­coat for bright nickel. The present trend is to­ward t he use of heavier eomposite coatings to secure increased protective value. The au tomo­t.ive industry commonly uses copper-nickel de­posits up to 0.002 in. thick on steel parts. It is t.herefore desirable to obtain a nondestructive method of measuring the total thickness and the relative thicknesses of the copper and ni ckel layers such as described in this paper .

An instrument known as the Magne-gage has been used to measure the thiclmess of nonmag­netic and of ni ckel coatings on steeU This in­strument measures the attractive force between a small permanent magnet and the plated sample. This attractive force is dependent upon the t hickness of the coating and the magnetic proper­ties of the eoa ting and base material. When a coating is interposed between the magnet and the magnetic base material, the attractive force is de­creased, provided that the coating is less magnetic than the base material, the decrease being greater the thicker the coating.

I Abner Brenner, J . Research NBS 20,357 (1938) RP1081.

Thickness of Composite Coatings

The Magne-gage is a spring balance, on one ann of which the magnet is suspended. A helical spring is attached in such a mann er that as the knob is turned, the spring is wound up. The force required to detach the magnet is a measure of the attraction between the magnet and the specimen. The attractive force of the magnet for the sample is read from a dial attached to the turning knob . The thiclmess of the coating is determined from this reading wi th the aid of a calibration curve made with coatings of known valu es of thickness. To determine the thickness of composite coatings, two magnets of different strengths are used. In order to compensate for the difference in the attractive force of the two magnets, it would be necessary to use two differ­en t gages unless different springs were used or two magnets were hung at diIferent distances from the pivot on the balance arm. For the measurements here reported, the stronger magnet was hung- nearer to the pivot. The modified gage is shown in figure 1.

Nickel is somewhat magnetic, although not as mu ch so as steel, hence the calibration curve for the nickel coatings differs from that for copper coatings as copper is practically nonmagnetic. As can be seen from figure 2, the calibration curve for pure nickel coatings of given values of thick­ness falls considerably to the left of the curve for pure copper, and is steeper. The calibration curve for any composite coating must fall within t be boundaries of these two curves as, for ex-

295

Page 2: Magnetic measurement of the thickness of composite copper ... · FIGURE 1. Magne-gagefor measuring thickness of composite coatings. ample, does the curve for equal thickness of cop

FIGURE 1. Magne-gagefor measuring thickness of composite coatings.

ample, does the curve for equal thickness of cop­per and nickel. In the absence of knowledge of the relative thickness of copper and nickel, the total thickness cannot be determined from a single reading as a certain reading can represent a number of thicknesses . For instance, in figure 2, a dial reading of 70 would represent a total thickness of 0.00049 in. of pure copper; 0.00085 in. of a half copper and half nickel coating; or 0.00168 in. of a pure nickel coating.

It is possible, by making measurements on a

30

o x 25 I U Z

- 20 If) If)

w z i'i 15 I t-

~ ,0 t­o t-

/' /'

,/ ,/

/'

/' /'

/

/ /

/

/ /

/ /

cl I

70 80 90 100 110 120 130 14 0 150 DI AL READING

FIGURE 2. Calibration curves for coatings of various per­centages of copper and nickel on sleel .

N, loo-perccn t-nickcl coating; C-N, 50-percen t-copper- 50-percent-nickel coating; C, lOG-percent-copper coating.

296

coating with two magnets of different strengths, to determine not only the total thiclmess but also the relative thickness of the two layers. This method is based on the fact that the calibration curves of magnets of different strengths have quite different curvatures. This is illustrated in figure 3, where the stronger magnet has about tlll'ee times the magnetic strength of the weaker magnet. Curves for composite coatings of differ­ent compositions, such as IOO-percent copper, half copper-half nickel, and 100-percent nickel were prepared, as shown in figures 4 and 5. Suppose that on a given sample a dial reading of 78 was obtained with the weaker magnet, and a reading of 80 was obtained with the stronger

30

25

'" X

I 20

" " vi ~ 15 z '" " i I- 10

.J .. I­o I-

WEAK MAGNET

DIAL READING

FIGURE 3. Calibration curves for two magnets of different slrengths with copper coatings on sleel.

30

x

If)

~ 15 z " u I 10 t-

-' <l b 5 t- ..--­..-----

----,///'/'

/ /

/ /

/ /

I I

/ /

c/ I

0 ~0~~~~~~~~~~~8~0~9~0~10~0~11~0-'~2~0~1~30~14~0~15~0~ DIA L READING

FIGURE 4. Calibration curves of coatings containing various percentages of copper and nickel on steel f or the strong magnet .

N, lOG-percent-nick cl coating; C- N, 50-pcrcen t-copper- 50-percent-nickel coating; C, IOO-percent-co pper coating.

Journal of Research

Page 3: Magnetic measurement of the thickness of composite copper ... · FIGURE 1. Magne-gagefor measuring thickness of composite coatings. ample, does the curve for equal thickness of cop

30

25

<t

0 20

x :x: <.)

~

vi <f) 15 w z '>< <.)

~ l-

-' « l- ID 0 l-

5

o

N

/ /

C-N

/ /

/

/ /

I I

CI

I , I J

I I I I /

DIAL READING

FIGURE 5. Calibration curves oj coatings containing various percentages oj coppel' and nickel on steel J 0 1' the weak magnet.

N, l()()-percen t·nickel coating; C- N , 50-perccnt-co pper-50-perccnL-nickcl coating; C, ]()()-percenL·coppcr coating.

magnet. Assuming that the sample has percent­ages of copper and nickel represented by one of the calibration curves, there are six possibilities as to the thickness as follows:

If 100 % coppeL ____ _____ _ .. ___ _ If 50 % copper- 50 % nickeL ___ _ _ If 100 % nickeL _____ .. ___ _____ _ _

Thickness, inch X 10- 4

Magnet 1 Magnp.t 2

7. 2 13.4

> 30. 0

9. 0 13.4 26.2

Thickness of Composite Coatings

'fhe thickness must be the same when measured by each magnet and in this ca e i 0.001 34 in. 'fhe percentage of copper or nickel i defined by the point where the thickness readings for the two magnets coincide. 'fherefore in the above case half the total thickness must be copper . If the relative percentage of copper does not correspond to a specified curve, it becomes nec­essary to interpolate between the curves.

'fo avoid cumbersome, lengthy tabulations, the following graphical method has been developed. 'fhemethodinvol ves the superposition of the graphs for one magnet on those for the other so that the pertinent gage readings coincide. 'fhe points of intersection of the curves are noted. In figure 6, the reading of 78 with the weaker magnet is super­posed on ' the reading of 80 with the stronger magn et, at the line AB. Although the curves intersect at a number of points, those of like composition cross at only thne points, X , Y, and Z. 'fhrough those points line XZ is drawn. 'fhe intersection Q of the line AB a nd the line XZ defines the total thickness, in this case 0.001 34 in., measured on tlle left scale. As the thickness falls on tbe calibration curve for half copper and half nickel, it is not necessary to interpolate to determine the percentage of copper or nickel. If the interpolation is necessary, it may be done in the followin g manner. Suppose that a reading

30

" ~ 25 x I '-' z -.20 If) If)

w z '" 15 '-' I >-.J 10 « >-r?

120 130 14 0 150

DI A L READI NG

FIGURE 6. Superposition of calibration curves of the weak (fig. 5) and strong (fig. 4) magnets.

Intersection occurs on t he half·copper-half-nickel curve. N, l()()-pe rccnt­nickel coati ng; C-N, 5O-percent-coppcr- 50-percent-nickcl coati ng; C, ]00-percent·copper coating.

297

Page 4: Magnetic measurement of the thickness of composite copper ... · FIGURE 1. Magne-gagefor measuring thickness of composite coatings. ample, does the curve for equal thickness of cop

of 68 is obtained with the weak magnet and a reading of 75 with the strong, and the curves are superposed as above. In figure 7, the two dial readings are represented by line AB. A line XZ is drawn through the points X, Y, and Z where curves of like percentages of copper intersect. The position of point Q, where AB cuts XZ, defines the thiclmess (0.00172 in. ), which is read off from the ordinate, and also the percentage of the total thickness that is copper. Point Q lies somewhere between points X and Y, i. e., between the calibration curves for pure nickel and half copper and half nickel , hence the percentage of copper must fall between these two. As Q lies along line XZ about 40 percent of the distance from the half copper curve to the nickel curve, the copper content is (40jl00 )X 50 = 20-percent copper. Although the variation in composition is not linear along line XZ, we may assume it to be linear, without appreciably increasing the error in the determination of the percentage of copper of the coating.

The gage as designed will measure with a satis­factory accuracy composite coatings with total thicknesses from 0.0005 to 0.0030 in. Below and above this range the superimposed curves for the two magnets run virtually parallel. Within this range the error of the gage was found to be abou t 10 percent for the total thiclmess and 15 percent for the copper percentage determination, as shown in table 1. M easurements were made with the composite magne-gage on especially pre­pared specimens and on commercially plated coatings to determine their thickness; and the thiclmesses were checked by microscopic and chemical analysis. The chemical analyses were made by dissolving the coatings and analyzing the solutions by standard methods. As the gage was calibrated with samples plated with standard Watts nickel, a study was made of the effects of bright nickel. It was found that the bright nickel deposits caused no error beyond the experi­mental errors of the gage itself. The accuracy of the gage may be seen in figure 8 and table 1, which include bright nickel coatings as well as Watts nickel.

The sample must meet certain requirements to yield satisfactory measurements. An area at least I-in. square is required, especially for the stronger magnet. If this area is unobtainable, the sample

298

30

(;>25 x I <.J

~20 (f) (f)

w 215

'" <.J

I f-

lO --'

'" f-o f- 5

0. 0 0 173 " -- //

/

Y I / /

1 \ / ,,/'

~z __ " 1/ \

-- /(8 ,,"" ,I"""" , o .50 100 1:>0

/ /

/

/

c/ I

/ /

10 20 30 40 50 60 70 80 90 10 0 110 120 130 14 0 150 DIAL READING

FIGU R E 7. Superposition oJ calibration curves of the weak and strong magnets.

Intersection interpolated between curves. N, IOO·percent· nickel coat ing; C-N, 5O·percent·copper-50·percen t·nickel coating; C, l00·percent·copper coati ng.

T ABLE 1. Accuracy oJ composite gage

'r he percentage deviation is based on the total thickness obtained by chern· ical analysis . In most cases microscopic thickness mea.surements were mad e on areas adjacent to t hose used for the chemical analyses. Where the sample was found to be uniform, the microscopic determinations were concordan t with the chemical resnlts.

Spccimen 'rotal thickness T hickness of copper, percentage of total

Devia· Differ· T ype Com pos· tion C hem· Com· eoce

1'0. Plating of C hemical ite gage ical posite in per-by· nick· analysis resu lts anal· gage cen t·

e l b ysis results age of copper

--._-------------------Per·

Inch Inch cent 1 Corn ..... B 40X IO- ' 31X IO-' -23 63 100 37 2 ~ .. do ..... ~ B 61 52 - 15 44 50 6 3 ~ .. do ...... B 63 62 -2 21 30 9 4 ... do ...... B 66 109 +65 84 76 -8 5 N BS .. .. . B 71 73 3 0 10 + 10

6 ... do ...... B 76 72 -6 35 45 10 7 Com ..... B 77 72 - 7 38 54 16 8 NBS ..... B 78 76 -3 35 40 5 9 Com ..... B 87 86 -1 86 76 -10

10 ~ .. do ...... B 86 87 +1 38 30 -8

11 ~ .. do .... .. B 93 82 - 12 24 30 +6 12 ... do ...... W 97 93 -4 57 60 3 13 ~ .. do ...... W 102 86 - 16 56 70 14 14 N BS ..... W 105 100 -4 76 75 -1 15 ~ .. do ...... B 113 132 +17 24 25 +1

16 Com ..... B 130 128 - 2 20 8 - 12 17 ... do ...... W 136 134 -1 59 55 -4 18 N B S ..... W 148 152 +3 75 80 + 5 19 ~ .. do ...... W 149 166 11 75 60 - 15 20 Com ..... W 150 145 -3 59 55 -4

21 N BS ..... W 191 185 -3 25 25 0 22 ... do ...... W 199 205 +3 75 80 + 5 23 ~ .. do ~ ..... W 201 235 17 25 33 6 24 ~ .. do ... ~ .. W 247 222 -10 25 40 15

• Com refers to commercially plated samples; N BS to those plated a t t he National Bureau of Standards.

b B relers to bright nickel ; W to Watts n ickel.

Journal of Research

Page 5: Magnetic measurement of the thickness of composite copper ... · FIGURE 1. Magne-gagefor measuring thickness of composite coatings. ample, does the curve for equal thickness of cop

30 ..,. b x :r 25 0

~

<II

<II 20 >-

oJ <l Z <l

>- 15 CD

<II <II

'" Z

" 10 0

:r t-

oJ <l t-O t-

10 15 20 25 30

TOTAL THICK NESS AS GA GE D, INCH X 10- 4

FIGURE 8. Accuracy of the compos'ite coating M agne-gage. T he dotted lices reprcscnt a IO·pcrccnt dc,' ia tion of thickness obtained by

gaging from tbe t hick ness obtai ned by analysis. Watts nickel, 0 , chemical analysis; ~, metallurgical analysis; bright nickel, X, chemical analysis .

,

Thickness of Coznposite Coatings

should be placed on a larger steel backing. The specimen should have a radius of curvatm e of not less than half an inch. An error of one dial divi­sion will cause an appreciable error in thicknes . For example, with the weaker magnet in the range from 0.001 in. to 0.0020 in. , one dial division will cause an error of 0.00014 in. , or about 8 percen t, if the sample is pure nickel; and with the stronger magnet an error of 0.00006 in., or about 5 percen t. Care therefore must be taken in making the read­ings, and in providing an adequate area and a sufficient radius of CUTvat ure.

For the initial calibration and subseq uent check­ing of this gage, standard-thickness samples will be provided. For each composition of coating, each set will consist of four specimens with dif­ferent thicknesses of coating on steel. It is prob­able that all req uircments can be met by three sets of specimens, consis ting respectively of (a) pure nickel, (b) half copper and half nickel, and (c) pure copper .

W ASHING'fON, October 20, 1947.