large surface biological/chemical decontamination …

89
LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION by JONATHAN NELSON BLACKWELL, B.S. A THESIS IN ELECTRICAL ENGINEERING Submitted to the Graduate Faculty of Texas Tech University in Partial Fulfillment of the Requirements for the Degree of MASTER OF SCIENCE IN ELECTRICAL ENGINEERING

Upload: others

Post on 18-Feb-2022

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

LARGE SURFACE BIOLOGICAL/CHEMICAL

DECONTAMINATION

by

JONATHAN NELSON BLACKWELL, B.S.

A THESIS

IN

ELECTRICAL ENGINEERING

Submitted to the Graduate Faculty

of Texas Tech University in Partial Fulfillment of the Requirements for

the Degree of

MASTER OF SCIENCE

IN

ELECTRICAL ENGINEERING

Page 2: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Copyright 2002, Jonathan Nelson Blackwell

Page 3: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

ACKNOWLEDGMENTS

The exploration into the large surface decontamination of biological and chemical

agents performed over the last 15 months could not have been completed without the help

of multiple people and departments at Texas Tech University (TTU). I offer special

thanks to Dr. O'Hair and Dr. Dickens, in the Electrical Engineering Department at TTU,

who gave me the opportunity to work on this project and have been constant sources of

advice, direction, and the main driving force throughout the project. I would also like to

recognize and thank Mike Hoffman for his significant contributions and work on the data

acquisition system. Mike was a great help when performing the tests and keeping me

company while we listened to the noise of 60,000 volts kill spores. Recognition to the

long standing efforts of Dr. Kristiansen and the rest of the faculty and staff working in the

Center for Pulsed Power and Power Electronics should also be made. (Dino Castro,

Daniel Garcia, and Lonnie Stephenson were especially helpful throughout the duration of

the project.) The facilities, equipment, and resources wdth which I was provided are a

result of countless years of work by these people.

Dr. Fralick and his lab assistant Kathryn Royce, who work in the Microbiology

Department at the Health Science Center at TTU, have been instrumental in providing the

biological agents, preparing the test specimens, transporting the specimens to the

Electrical Engineering lab, and evaluating the specimens after exposure. Dr. Dasgupta

and Dr. Karthikeyan Sathmgnan, a post-doctoral researcher, were responsible for

preparing, transporting, and evaluating the chemical agents used during the experiments.

Dr. Dasgupta and Dr. Sathmgnan work in the Chemistry department at TTU. The

judgment and professional experience of both Dr. Fralick and Dr. Dasgupta have been

invaluable resources throughout this project.

On a personal note, I would like to offer special thanks to my wife. Heather, who

has been a constant source of love and support. I would also like to thank my parents for

the love, support, and guidance they have offered throughout my life. They are

responsible, in a large part, for the person I have become.

11

Page 4: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

TABLE OF CONTENTS

ACKNOWLEDGEMENTS u

ABSTRACT v

LIST OF TABLES vi

LIST OF FIGURES vii

LIST OF ABBREVIATIONS ix

CHAPTER

L INTRODUCTION 1

1.1 Background 1

1.2 Project Objective 3

IL EXPERIMENTAL SETUP 7

2.1 Generation of Killing Mechanisms 7

2.1.1 Pulsed Corona Discharge 7

2.1.2 Heat 17

2.1.3 Hydroxide (OH) Radicals 21

2.2 Data Acquisition 22

2.2.1 Oscilloscope Settings 23

2.2.2 Agilent Data Acquisition/Switch Unit. 26

2.2.3 Solutions to Electromagnetic Interference (EMI) 28

2.2.4 Power Supplies... 30

III. CHEMICAL AND BIOLOGICAL AGENTS 32

3.1 Biological Agents 32

3.1.1 Background 32

3.1.2 Preparation 32

3.1.3 Coupon Analysis 33

3.2 Chemical Agents _ 34

rV. CHEMICAL AND E. COLI TEST RESULTS 35

4.1 Method for Calculating Results 35

4.2 Chemical Experimental Results 37

111

Page 5: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

4.2.1 Malathion Exposed to the Pulsed Corona Discharge... 37

4.2.2 Parathion Exposed to the Pulsed Corona Discharge.. 38

4.3 E. coli Experimental Results 41

4.3.1 E. coh Exposed to the Pulsed Corona Discharge 42

4.3.2 E. coh Exposed to Only Heat. 43

4.3.3 E. coU Exposed to Heat Then the PCD 45

4.3.4 E. coli Exposed to the PCD With Water Vapor 46

V. SPECIFIC APPLICATION TO THE ANTHRAX PROBLEM. 48

5.1 Motives and a Re-Evaluation 48

5.2 Bacillus Experimental Results 49

5.2.1 Bacillus Exposed to the Pulsed Corona Discharge 49

5.2.2 Bacillus Exposed to Only Heat 50

5.2.3 Bacillus Exposed to Heat Then the PCD 52

5.2.4 Bacillus Exposed to the PCD With Water Vapor. 52

5.2.5 Bacillus Exposed to the PCD With Water Vapor when Covered 54

VI. FUTURE RESEARCH 56

VIL CONCLUSIONS 58

REFERENCES 59

APPENDDC A. RESULT TABLES FOR ALL TESTS 60

B. SLEDE LAYOUTS FOR ALL TESTS 67

C. POWER CONDITIONING UNIT (PCU) 73

IV

Page 6: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

ABSTRACT

Contamination by microorganisms such as E, coli and Anthrax has been a source

of concern in the agricultural and food processing industries for many years. However,

the increasing presence and availability of the more robust microorganisms, such as

Anthrax, present a significant threat to the safety and quality of life to which we are

currently accustomed. As the knowledge and technology to manufacture and weaponize

lethal sfrains and different suspensions of biological and chemical agents become more

available, the knowledge and technology to detect and desfroy these biological and

chemical agents must also be developed. Methods currently used to decontaminate

biological agents include elecfron beams, exposure to very high heats in excess of 130°C,

exposure to Ultraviolet (UV) light, and the use of liquid solvents such as bleach, just to

name a few. Based on previous research, another very effective decontamination

technique is to expose the agent to an arc jet plume, which exhibits multiple killing

mechanisms but is energy inefficient. It is of great interest to explore additional

decontamination techniques that decontaminate more rapidly, are more energy efficient,

do not harm the contaminated surface, and do not leave behind a residue which must later

be cleaned.

The pulsed corona discharge (PCD), which generates many of the same killing

mechanisms as the arc jet, is thought to be a viable solution to the decontamination of

biological and chemical agents; especially when combined with environmental enhancers

that increase the amount of water vapor in the air or that increase the temperature of the

specimen. A combination of experiments has been performed with the PCD utiUzing the

environmental enhancements previously mentioned on both chemical and biological

specimens with the main goal pointing towards the decontamination of Anthrax in

various environments. Additional experiments have been performed to determine if

common barriers, such as paper or plastic envelopes, would inhibit the killing efficiency

of the pulsed corona discharge. The most immediate benefits of the research performed

with the pulsed corona discharge could be applied toward the destmction of Anthrax in

contaminated mail.

Page 7: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

LIST OF TABLES

A.l Malathion Exposed to the Pulsed Corona Discharge 61

A.2 Parathion Exposed to the Pulsed Corona Discharge 61

A.3 E. coh Exposed to the Pulsed Corona Discharge 62

A.4 E. coh Exposed to Only Heat 62

A.5 E. coh Exposed to Heat Then the Pulsed Corona Discharge 62

A.6 E. coli Exposed to the Pulsed Corona Discharge with Water Vapor 63

A.7 Bacillus Exposed to the Pulsed Corona Discharge 64

A.8 Bacillus Exposed to Heat Then the Pulsed Corona Discharge 64

A. 9 Bacillus Exposed to the Pulsed Corona Discharge with Water Vapor 64

A. 10 Bacillus Exposed to Heat Only 65

A.ll Bacillus Exposed to the PCD with Water Vapor when Covered 66

vi

Page 8: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

LIST OF FIGURES

2.1 First Wire Array Stmcture 10

2.2 Second Wire Array Stmcture 11

2.3 Layout for Heat Exposure Tests 19

2.4 Temperature Variation of Hot Plate at 80°C 21

2.5 Schematic of the Data Acquisition System 24

2.6 Location of Current and Voltage Monitors in A2 Unit of PCU 25

2.7 PINetworic 29

2.8 Opto-Isolator Circuit 29

4.1 Destmction Process of Malathion 38

4.2 Malathion Survivability When Exposed to the PCD 39

4.3 Percentage of DDTP Present When Exposed to the PCD 39

4.4 Parathion Survivability When Exposed to the PCD... 40

4.5 Paraoxon Present During Exposure to the PCD .40

4.6 Parathion Chromatograms .41

4.7 E. coh Survivability When Exposed to the PCD .43

4.8 E. coh Survivability When Exposed to 50°C .44

4.9 E. coh Survivability When Exposed to 75°C .45

4.10 E. coli Survivability When Exposed to Heat Then the PCD. .46

5.1 Bacillus Survivability When Exposed to the PCD 50

5.2 Bacillus Survivability When Exposed to 100°C 51

5.3 Bacillus Survivability When Exposed to 125°C .51

5.4 Bacillus Survivability When Exposed to Heat Then the PCD 52

5.5 Bacillus Survivability When Exposed to the PCD with Water Vapor 53

5.6 Bacillus Survivability When Placed Inside the Cushion Mailer 55

B. 1 Layout for Bacillus K, L, N, O Tests 68

B.2 Layout for Bacillus X, Y, Z, AA, BB, CC, DD, FF, KK, LL, NN,

0 0 and Parathion M Tests 68

B.3 Layout for Bacillus J, M, P, S, and E. coli Q Tests .69

B.4 Layout for Bacillus V and W Tests 69

vii

Page 9: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

LIST OF FIGURES

B.5 Layout for Bacillus T and U Tests 70

B.6 Layout for Bacillus EE and GG Tests 70

B.7 Layout for Bacillus HH Test 71

B.S Layout for Parathion K, Malathion H Tests 71

B.9 Layout for Malathion I Test 72

B. 10 Layout for Parathion H Test 72

C I Circuit Diagram of the Thyristor Charging Unit Al 77

C.2 Circuit Diagram of the High Voltage Nanosecond Pulse Shaper A2 .....77

C.3 Circuit Diagram of the Control Unit A3 78

VI11

Page 10: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

LIST OF ABBREVIATIONS

A ampere: the SI unit for electiical current

AC Alternating Current

AWG American Wire Gauge

BNC Bayonet Neill Concelman [Electronics]: Connector used with coaxial cable, named after inventor

DC Direct Current

GSa/s Giga-Samples per second

Hz Hertz: the SI unit for frequency

IR Infrared (as in IR temperature probe)

J Joule: a unit for Energy

kJ kilo-Joule (1000 Joules)

kV kilo-Volt or 1000 volts

kVA kilo-Volt Amp (1000 Vohs Amps)

MHz mega-Hertz (1,000,000 Hertz)

mL milliliter (0.001 liters)

mm millimeter (0.001 meters)

ms milhsecond (0.001 seconds)

ns nanosecond (1 x 10' seconds)

nmole nanomoles (1 x 10' moles)

Q Ohm: the SI unit for electrical resistance

PCD Pulsed Corona Discharge

PCU Power Conditioning Unit

SOS Semiconductor Opening Swdtch

fxs microsecond (1x10"^ seconds)

UV Ultraviolet (as in UV light)

V volt: the SI unit for voltage

IX

Page 11: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER I

BACKGROUND AND PROJECT OBJECTIVES

1.1 Background

On a daily basis, some portion of the nation's population comes in contact with

microorganisms and chemical compounds that are potentially lethal in the appropriate

concenfrations. Some of the more common hosts in industry to these agents include

feedlots, insecticides, and pesticides. Unfortunately, the probability of agents such as

these being used as weapons has also increased. As the presence of microorganisms such

as Anthrax and compoimds such as Parathion and Malathion become a more intmsive

presence and threat to the safety of the general public, technologies and decontamination

techniques must be developed which are increasingly energy efficient, decontaminate

more rapidly, do not harm the surface to be decontaminated, and leave no residue.

Throughout industry, many different decontaminafion systems are used which utilize

different sources of energy. Some of these destmction techniques include the use of

electron beams, killing with excessive amounts of heat, and liquid solvents such as

bleach. The efforts of this research group have been focused on determining the

capability of a pulsed corona discharge to desfroy biological and chemical agents. The

pulsed corona discharge is believed to generate killing mechanisms which will resuh in a

more energy efficient, rapid, non-destmctive solution to the problem of contamination by

biological and chemical agents.

The technology for this research was first investigated at Texas Tech University

(TTU) Center for Pulsed Power and Power Electronics in the late 1980s with NASA

sponsored research on arc jets, one of the satellite size electric space propulsion

technologies. The main parties involved with this research at TTU include Dr. E. O'Hair,

Dr. J. Dickens, and graduate students and lab technicians working imder their

supervision. The details concerning the background of the technology have been provided

by Dr. O'Hair. The primary focus of the research was directed toward reducing the

cathode erosion such that the operating time in the space environment could be extended

to at least 1000 hours. Both 10 and 30 kW size arc jets operating at 100 and 250 amps

1

Page 12: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

DC were investigated. The gas propellants used were nitrogen, hydrogen, or a

combination of the two at a flow rate of a few tenths of a gram per second. At these

current levels and small flow rates the nozzle exit gas temperature was very high, on the

order of about 10,000 °C. The plasma that was produced consisted of elecfrons, ions,

atoms and molecules. In addition to these elements, an elecfric field. Infrared (IR)

radiation, and Ulfraviolet (UV) radiation were also produced.

In 1997, the US Air Force, in an STTR solicitation, requested proposals on the

destiiicfion of biological and chemical warfare agents. The research group at Texas Tech

University (TTU) fonned a team v^th MONTEC, a Montana small business. The team

was awarded a one year contract to investigate the decontamination of biological and

chemical agents on large surface areas using a high temperature plasma.

The biological agent used during the experiments with the arc jet was a strain of

Bacillus spore. Bacillus spores were employed as the test specimen because they are

very good simulants for Anthrax, a well-known biological warfare agent. The spores

were plated on fiberglass slides. The exhaust environment produced by an arc jet is very

destmctive of aerosol deposited biological agents. Results showed that these agents

could be desfroyed, using air as the propellant, at a rate of 3 mph without causing any

damage to the surface. The proposed operational concept was to develop a 5 ton tmck

size mobile unit, using a row of arc jets positioned a couple of inches above the surface.

The decontamination unit would move fast enough to "clean-up" a road or mnway in a

reasonable time but not so fast that the residue agents would become airborne and blown

away.

The technical decontamination objectives of this research were met. However, to

operate a 3 meter long row of 10 to 15 kW arc jets would require a tmck-mounted

electric generator that could produce over 200 kW of power. The supersonic high

temperature gas plasma of the arc jet contains multiple killing mechanisms which, when

used individually at sufficient power levels and duration, have desfroyed surface

deposited chemical and biological agents. While the arc jet exhaust is a very effective

method for decontamination, it is not a very energy efficient solution. The two main

drawbacks with using the arc jet large surface decontamination system are: that it will

Page 13: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

require a large supporting electric generator and that it probably generates a more

desfructive environment than is required. These two drawbacks have lead to the first year

of research supported by SBCCOM and perfonned by the group at TTU.

A proposal was submitted to, and funded by US Army SBCCOM for the first year

of a three-year effort to research and develop the concept of a mobile field deployable

chemical and biological large surface decontamination system. In addition to effectively

decontaminating chemical and biological agents, the decontamination system should also

preserve the surface that is being decontaminated, leave no residue that requires clean-up,

decontaminate 15,000 square meters per hour, require no non-standard expendables, have

a long maintenance free "shelf-life," and have the ability to operate for longs hours with

minimum maintenance.

The technology exists in single sources such as: electron beams, UV lights, pulsed

corona discharges, etc., to reproduce portions of the arc jet environment. Electrically

driven sources like these listed will meet the general consfraints addressed in the previous

paragraph. From the published data there are no individual sources that meet the rapid

decontamination rate objective. So, the question that must be addressed is: What

combination of sources in series or concurrently can produce the desired result, i.e.,

"acceptable" surface decontamination with only 0.05 seconds of exposure, at a

"reasonable" ratio of square meter/kW-hr?

1.2 Project Objective

Past research has shown that electrical technologies can be very effective tools to

rapidly decontaminate large surface areas of biological and chemical agents. For the first

year, certain tasks have been addressed to provide the infrastmcture for future work on

electrically driven biological and chemical decontamination. The selection of energy

sources and setting up the experimental apparatus are the first goals that have been

reached. Following the development of a flexible experimental set-up, the group has

determined the chemical and biological simulants to be used, who will prepare them, and

who will evaluate them. After making the above decisions, the first year's experiments

have been performed with the intention of perfecting and evaluating sources, data

Page 14: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

acquisition instrumentation, agent handling, and results evaluation. The final work has

been to address the problem of mail contaminated with anthrax as a specific need

application of the equipment and processes developed over the last year.

The group that is referred to throughout this paper consists of faculty, staff, and

student workers from three different departments at Texas Tech University. Dr. O'Hair

and Dr. Dickens are the faculty members in Electrical Engineering who have directed

their expertise towards supervising the development of the pulsed corona discharge,

adding the environmental enhancements, and all other equipment necessary to collect

data for energy calculation and system monitoring. They have also been responsible for

organizing the efforts of the Chemistry and Microbiology departments to provide and

analyze the test specimens. As an Electrical Engineering graduate student, I have

worked imder the supervision of Dr. O'Hair and Dr. Dickens. In the Microbiology

department in the Health Sciences Center (HSC) at Texas Tech University, Dr. Fralick

and his student assistant, Kathryn Royse, have been responsible for acquiring, preparing,

fransporting, and evaluating the E. coli and Bacillus specimens after exposure. Dr.

Dasgupta and a post-doctoral researcher. Dr. Sathmgnan, prepared and evaluated the

Malathion and Parathion test specimens for the project. Naturally, the faculty and staff in

the Chemistry and Microbiology departments used their backgrounds to help make

critical decisions concerning appropriate test conditions and test specimen sfrains, when

necessary. Collaboration among all three disciplines was necessary to properly interpret

results obtained from different exposure techniques.

The first objective of the project is to explore the effectiveness of the Pulsed

Corona Discharge (PCD) in decontaminating large surfaces. The PCD is created by

pulsing a high negative voltage of about 60,000 Volts through a wire array generating an

approximately uniform discharge between the wire array and a ground plate. The voltage

pulse has a pulse width of about 20 nanoseconds (ns) and the pulse repetition frequency

is 100 Hz. Killing mechanisms generated by the PCD include a very high electric field,

kinetic energy of charge particles (electrons and ions), IR radiation, UV radiation, and the

production of OH' radicals. The specimens of interest are exposed to the PCD by placing

them on the ground plate, below the wire array. An important condition to note is that all

Page 15: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

experiments are run at atmospheric pressure in order to more accurately imitate the

environmental conditions in which the system is to be used.

In addition to investigating the effectiveness of the PCD, the group will also be

evaluating methods to enhance the capability of the PCD to desfroy biological and

chemical agents. The main goal is to explore the possibility that a combination of the

PCD and another killing mechanism, or some environmental enhancement, can resuh in a

more energy efficient method to decontaminate the agents. The combinations of killing

mechanisms that have been explored include the PCD combined wdth thermal energy and

the PCD combined with water vapor. Thermal energy and water vapor are also tested

individually in order to establish a point of reference that helps determine the

confributions of each killing mechanism when combined wdth the PCD. The biological

agents have been exposed to all combinations of exposure mentioned above. The

chemical agents have not been exposed to heat or humidity. The different agents are only

exposed to conditions that the researchers feel will lead to meaningful real world

applications.

The biological and chemical agents chosen as simulants are thought to represent

agents that are already known to be weaponized and tend to be robust in survivability.

The chemical simulants are Parathion and Malathion. These compounds are commonly

used in insecticides, but in increased concentrations can be very dangerous. These

chemical agents are felt to be good simulants for chemical agents that may be

encountered in the field. The biological agents that have been tested include Escherichia

coli (E. coli) strain 802 and Bacillus stearothermophilus. The E. coli has been chosen as

a preliminary test specimen to make sure that each stage of the testing will kill the E. coli

before using the more expensive and much more robust bacillus spores. The bacillus

spores that have been chosen simulate the well-known biological agent. Anthrax.

The experiments throughout the first year of research have been stmctured to

perfect and evaluate the electrically driven energy sources in addition to determining their

effectiveness as killing mechanisms. Multiple experimental apparatus have been

developed and problems concerning data acquisition have been addressed. One of the

biggest challenges conceming data acquisition for electrical performance has been to

Page 16: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

minimize the amount of noise induced on the signal and power lines that is from

elecfromagnetic interference (EMI) generated by the pulsed corona discharge. Methods

and protocols for handling, transporting, storing, and evaluating the chemical and

biological agents have also been developed and fine-timed. Finally, a method for

calculating energy levels of exposure and comparing the killing capability of the different

"energy" sources based on this energy level and the survival of the simulants has been

established. Basing the survivability of a simulant on the amount of energy to which it is

exposed allows one to compare the killing capability of one energy source with the

killing capability of a different energy source.

In all tests, the main parameter that is used to determine electrical efficiency is the

total energy applied to the specimen. Because of equipment hmitations, the time duration

of exposure is on the order of minutes and a certain amount of "rest" time must be

allowed in between shorter exposure intervals in order to allow the equipment to cool.

For this reason, the results one will view will be in terms of total energy applied to the

specimen for a given duration. At this point, it is assumed that increasing the frequency

of pulses using the proper Power Conditioning Unit (PCU) and removing these "resf'

periods will resuh in a more realistic time frame for decontamination while maintaining

the effectiveness of the PCD technology. However, this is an assumption left for later

investigation.

A specific need has developed during the past year prompting the group to

research the effectiveness that the PCD technology has when applied toward the

decontamination of mail. It is of interest to explore the different types of materials that

one may encounter in the mail system and limits that the different materials may impose

on the effectiveness of the pulsed corona discharge. Experiments combining the PCD

with heat and with water vapor have also been performed during the investigation of mail

decontamination. Materials tested include 20 lb. printer paper in varying numbers of

sheets, cushion mailer envelopes with shredded newspaper as the cushioning, cardboard

mailer envelopes, and plastic mailer envelopes. The collection of data under these

various conditions is an important resource for further project development.

Page 17: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER II

EXPERIMENTAL SETUP

2.1 Generation of Killing Mechanisms

2.1.1 Pulsed Corona Discharge

As mentioned in Chapter I, the investigation of individual killing mechanisms

such as the PCD, conceming their potential to decontaminate biological and chemical

agents, stems from the previous research performed using an arc jet. Like an arc jet, a

pulsed corona discharge generates multiple killing mechanisms. The main elements

thought to cause the destmction of biological and chemical agents include the presence of

an extremely high electric field, the kinetic energy of elecfrons and ions bombarding the

surface of the specimens, IR radiation, UV radiation, ozone, and hydroxide (OH)

radicals. The tips of filaments extending toward the ground plate due to the partial

breakdown of air that exists during a corona discharge are one source of this very high

elecfric field. The tips of the breakdown filaments have a great deal of field enhancement

and in turn expose the biological samples to a very high electric field. The field

enhancement of the wdres in the wire array is another source of the very high electric

field. For this reason, particular interest lies in the viability of biological and chemical

specimens when exposed to a pulsed corona discharge (PCD).

Since the 1960s there has been an interest in the effects of a pulsed electric field

on the viability of microorganisms, mainly bacteria, in liquid. Schoenbach explains that

"experimental results obtained over a large range of electrical and microbiological

parameters, point towards an irreversible formation of pores in the cell membrane as

mechanism for lysing" [1]. These studies show that both the magnitude of the applied

electric field and the duration of the pulse affect the viability of the microorganism.

There is some question about how large a role the electric field plays in the tests nm with

the pulsed corona discharge because the pulse duration is so short (nanoseconds)

compared wdth the pulses used by Schoenbach (microseconds range). Schoenbach, while

speaking of the theory of irreversible elecfroporation, states that "only when we approach

the range of very short pulses at high elecfric fields, must the conventional theory be

Page 18: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

modified to include internal effects of electric fields" [1]. ft has been assumed throughout

the tests with the PCD that the effects of the applied electric field are cumulative. The

resuhs presented in later chapters will support this assumption. The repetition rate for

these tests is 100 Hz. ft has been necessary to expose the specimen to one minute of

PCD, and then let the power conditioning unit (PCU) rest for 5 minutes. Another one

minute of exposure is applied after the PCU has cooled down and this process is repeated

until the desired duration of exposure is accumulated.

Research conducted by Mounir Laroussi demonstrates the effectiveness of using a

uniform glow discharge at atinospheric pressure to sterilize biological media. "The

plasma generated by the atmospheric pressure glow discharge ... is a source of elecfrons,

ions, excited atoms and molecules, active free radicals, and radiation (from the infrared to

ultraviolet)" [2]. In a similar way, the PCD is also a source of electrons, ions, excited

atoms and molecules, active free radicals, and radiation. Tests by Laroussi give sfrong

support that the effects of the electrical field and the charged particles in an atmospheric

glow discharge play a significant role in the destmction of the specimens in the

atmospheric glow discharge. The tests performed at Texas Tech with the PCD have not

been compared to an equivalent level of UV radiation. Laroussi comments on the

contributions of UV radiation by saying "... that the UV radiation generated by the

plasma was not the only agent of sterilization [and that charged] particles and active free

radicals appeared to play a significant role" [2]. Based on the work by Laroussi, the

assimiption that the UV radiation is not the dominant killing mechanism in the PCD has

been made. The PCD has been pursued because it is thought to be an energy-efficient,

electrically driven energy source capable of decontaminating chemical and biological

agents and could be easily realized as a mobile unit for field use.

The main considerations in generating an appropriate pulsed corona discharge are

uniformity and reproducibility. The generation of a uniform pulsed corona discharge

enables one to assume that the multiple specimens being tested during each test nm are

exposed to the same intensity electric field. If the PCD is non-uniform, one must

consider the relative locations of the simulants underneath the wire array in addition to

the duration of exposure when comparing the viability of the specimens. Because the

Page 19: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

tests have been mn over an 8 month period, one must be able to reproduce similar

conditions each time a test is mn or unwanted variables are introduced into the

experiment. Limiting the test environment to as few variables as possible is essential to

drawing accurate conclusions about the effectiveness of the PCD in killing the biological

specimens and breaking down the chemical specimens. Various wire spacings have been

used to address the issues of generating a uniform discharge and providing a high degree

of reproducibility.

Two different stmctures have been used for supporting the wire array. The first

stmcture suspends the wire array above the ground plane with no physical contact

between them. The wire spacing is varied by threading the wdre through a copper rod

with holes in it every 0.25 inches and back toward a sheet of plexi-glass with slits cut

every 0.25 inches into the bottom of it. Lifting the ground plane closer to the wire array

using a scissor jack varies the distance between the wire array and the grotmd plane. This

stmcture is limited in its ability to vary the wire spacing to distances less than 0.25 inches

and its ability to curve the edges of the wire array away from the ground plane. The

distance between the wire array and the ground plane is also very difficult to keep

uniform across the entfre exposure area because the scissor jack has a certain degree of

flex in it. This stmcture requires that one thread the wire array with one piece of wire,

which makes tightening the wdres of the array and replacing a broken wire cumbersome

and time consuming. The first stmcture was used for tests from June 15,2001 through

October 17, 2001. A drawing of the first wire array stmcture may be seen in Figure 2.1.

The second and present wire array stmcture addresses the lack of flexibility

experienced with the first stmcture when trying to vary height, wire spacing, wire size,

and curvature of the wdre array (Figure 2.2). The second stmcture employs the same

concept used when stringing guitars. Instead of a single, very long wire, multiple wdres

are used to create the wire array. The individual wires are thread through the center of a

bolt, which can then be used to tighten the wire by turning the bolt and securing it with a

pair of nuts. The wires are thread through a set of fiberglass templates secured at either

end of the array that have a minimum spacing of 0.125 inches, which is half the spacing

allowed by the first stmcture. The fiberglass templates used to space the wires are easily

Page 20: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

and quickly machined enabling one to reduce the spacing further should one desire to do

so. The consistency of the spacing of the wire array is much better because the fiberglass

templates at either end are much more accurate than the spacing slits used with the first

stmcture. The stmcture is secured directly to the ground plane using nylon bolts at each

comer enabling one to adjust the distance between the wire array and ground plane wdth

greater accuracy and consistency. The curvature of the wire array can easily be varied by

inserting curved fiberglass forms underneath the wires at each end of the array and

tightening the wires to conform to the curves of the form.

Figure 2.1: First Wire Array Stmcture

The main weakness with the second stmcture is the tendency for breakdown to

occur at the edges of the stmcture where the edge of the plexi-glass stmcture is in

between the wire array and the ground plane. It is at this location that surface flashover

plays a role in premature breakdown along the edges of the stmcture that supports the

wire array. Even though the stmcture does not extend all the way to the groimd plain, it

does provide a smooth surface between the two conductors along which charged particles

can more easily accumulate, boimce across, and continue down to the groimd plane. This

creates a lower breakdown voltage along the edge of the stmcture than exists between the

wire array and the ground plane in the area intended for exposure. A patch of silicon is

10

Page 21: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

placed on the ground plane beneath these trouble spots to minimize the occurrence of this

breakdown.

Figure 2.2: Second Wire Array Stmcture

An array of wires closely spaced will generate a discharge that is approxunately

uniform assuming that the wire array is sufficiently separated from the ground plane. A

wire array is favorable to a parallel plate geometry because the field enhancement

experienced at the edges of a parallel plate will discourage the generation of a uniform

discharge throughout the volume between the plates. When generating the PCD using the

array, four main variables have been taken into account. The distance between the wdre

array and the ground plane, the spacing between the vsdres of the wire array, the size of

the wire, and the curvature of the wdre array each affect the uniformity of the discharge.

While the wire array approximates a parallel plate geometry, it also takes

advantage of the field enhancement of the wires generating a corona discharge along the

length of each wire. As the array is moved away from the ground plane, the effects of

charge diffusion cause the individual discharges from the wires to overlap. The overlap

of each individual corona discharge causes the electric field experienced at a distance just

above the ground plane to be approximately uniform.

The level of field enhancement is determined by the geometry of the electrode

configuration. For the wire array, a cylinder-plane geometry is assumed. The formula

11

Page 22: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

for the field enhancement factor is presented in Equation 2.1, where d is the distance

between the wire and the ground plane and r is the radius of the wire [3]. The field

enhancement factor is muhiplied by the mean field strength to get the maximum field

strength in a non-uniform field. The maximum field sfrength occurs at the wire. The

mean field strength is based on the voltage per distance for a given geometry. The

average pulse voltage used during the experiments is around -60 kV at a distance of about

1 inch (or 2.54 cm). These conditions yield a mean field strength of about -23.6 kV/cm.

Given that 22 AWG wire has a radius of 0.032 cm and is separated from the ground plane

by 2.54 cm, the field enhancement factor (f) is about 16.4. With a field enhancement

factor of 16.4 and a mean field sfrength of-23.6 kV/cm, the maxunum field sfrength is -

387 kV/cm at the wire. There is not an immediate arc to the ground plane under these

conditions because the maximum field sfrength dies away fairiy quickly in space and

because the voltage pulses are only about 20 ns in duration, leaving little time for a short

circuit path to develop to ground.

r-li

0.9d

r + d f = ~ 7 T - A (2.1)

r J

As the wire array is moved further from the ground plane, the wires in the array

tend to act less like a single wire above a plane and the wire array begins to exhibit more

of the properties consistent with a parallel plate geometry, namely field enhancement at

the edges of the wdre array. According to Equation 2.1, one can also see that as the

distance between the wires of the array and the ground plane increases, the field

enhancement factor increases, but the maximum field strength decreases. By multiplying

the field enhancement factor by the mean field strength, the only influence of the distance

(d) that is left is in the denominator of the equation. This causes the maximum field

strength to be inversely proportional to the distance between the electrodes. As the wire

array is moved closer to the ground plane, the field enhancement due to the individual

wdres of the array becomes more significant, there is less distance for charged particles to

diffuse, and the discharge becomes less uniform and more concentrated under each wire.

12

Page 23: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Both the first and second wire an-ay stmctiires suspend the array between 0.8

inches and 1 inch from the ground plane depending on the spacing between the wires of

the wire array. The grounded mounting ring at the face of the PCU is 1.55 inches from

the aluminum bar that is attached to the high voltage terminal. If the distance between

the wire array and the ground plane exceeds or even closely approaches the distance from

the grounded mounting ring to the aluminum cylinder, a portion of the energy used to

generate the PCD will be lost to corona discharge and arcing between the aluminum

cylinder and the mounting ring. Considering the factors explained above, it is estimated

that the maximum possible distance between the wire array and the ground plane is about

1.4 inches, using the second wire array stmcture.

The spacing between the wires of the wire array and the size of the vsdre must also

be considered when adjusting the wire array to generate a uniform PCD. If the vvdres are

more closely spaced, the charge distribution across the array becomes more uniform

approaching characteristics similar to the parallel plate geometry. Likewdse, spreading

the wires apart decreases the imiformity of charge accumulation across the array leading

to a less uniform discharge very close to the groimd plane, keeping all other conditions

constant. A wdre spacing of 0.25 inches is used during experiments between June 15,

2001 and December 14, 2001. Experiments run after December 14, 2001 are performed

using a wire spacing of 0.125 inches. The wire spacing has been decreased from 0.25

inches to 0.125 inches in order to attempt a more uniform discharge. However, no

significant improvement has been experienced since reducing the wire spacing.

Changing the wdre diameter of the wires is another factor that affects the

imiformity of the PCD very close to the ground plane. One may refer to Equation 2.1 to

see that reducing the wdre diameter (or radius) will increase the field enhancement factor.

Increased field enhancement generates a higher electric field between an individual wire

and the ground plane. In the current application, this is favorable only to a certain extent.

Too much field enhancement will provide for a concentrated discharge between each

wire and the ground plane limiting the ability to generate a uniform pulsed corona

discharge and quickly leading to voltage breakdown conditions (arcing).

13

Page 24: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

It is desirable to create conditions such that the effective voltage at each wire is

just below the breakdown voltage in air at atmospheric pressure. The level of field

enhancement increases with decreasing wire diameter which in turn decreases the

breakdown voltage as seen in Equation 2.2 [3]. In Equation 2.2 Vb is the breakdown

voltage of a gas in a non-uniform electric field, E^ is the breakdown field of the gas, r is

the radius of the wire, and d is the distance between the vvdre and the ground plane.

Based on Equation 2.2, varying the wire diameter (or radius) directly changes the

(r+d) ' Vb= Lll-Em-r-h (2.2)

performance of the wire array with respect to the uniformity of the PCD. Equation 2.2

demonstrates a concept relating the field enhancement of a certain geometry to the

voltage at which the gas between the elecfrodes wdll break down and form a conducting

path. While the concept is the same. Equation 2.2 is meant for the static voltage case and

cannot be direcfiy applied to the case for this project where the voltage is pulsed at 100

Hz. The breakdown voltage in air will be higher for a pulsed voltage as compared to a

static voltage as seen when using Equation 2.3 [3]. This equation is valid for relatively

large gaps with divergent fields where Vb is the breakdown voltage in kV, p is the

pressure in atmospheres, d is the gap separation in cm, and t is the time in ]xs at which the

voltage exceeds 89% of the breakdown voltage, and K_ is equal to 22 and n is equal to

0.6 and both are constants dependent on the gas and the polarity of the voltage pulse.

Plugging in the appropriate values, a rough estimate for the breakdown voltage of air

under the operating conditions used on the project is about 100 kV/cm.

r 5 - l A

Vb = K VP -d -t J (2.3)

Twenty-two AWG wire has been used during all exposures of the specimens. 26

AWG wire has been used when attempting to find a combination of variables that

generate a more unifonn discharge. The effects of smaller diameter are known, however

the ease with which it can be handled limits its desirability when stnnging the wires of

the an-ay. Smaller v\dre is much easier to kink leaving points along the line with even

14

Page 25: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

higher field enhancement leading to certain spots along the wires that consistently

generate voltage breakdown.

Despite the field enhancement of each wire, the wires in the array are spaced

close enough together that they still approximate a parallel plate geometry at a distance

between 0.8 inches and 1 inch from the ground plane leading to a higher electric field at

the edges of the anay. Because of this, if the distance from each wire to the plate is the

same, the discharge will be more concentrated along the outer wires of the anay. To

resolve this issue, the wires toward the outer edges of the anay curve away from the

ground plane. It is desirable to anange the wdres in a manner such that the electidc field

very close to the ground plane is uniform. Electric field is defined as voltage per

distance. Realizing that the elecfric field is stronger at the edges of the anay, one may

increase the distance between the vsdres at the edges anay and the ground plane in order

to counter the effects of field enhancement on the uniformity of the electric field close to

the ground plane.

A considerable source of breakdown that can occur is not between the wire anay

and the ground plane, but between the high voltage connector and the grounding ring at

the face of the pulser. The high voltage connector is a 6/32 machine screw protmding

from the center of a circular region on the front face. There are six grounding bolts

which sunoimd the high voltage connector at a radius of 49 millimeters (mm) spaced

every 45 degrees starting at the 0 degree point as viewed on a polar graph. These six

bolts are connected by an aluminum ring. Because of the location of the aluminum ring,

considerable care has to be taken to prevent breakdown at the pulser face when using the

wire anay. Breakdown at the pulser face is undesirable because it results in wasted

energy that is not applied to the intended specimen.

The geometry between the high voltage post and the grounding ring may be

approximated as coaxial. The inner conductor of this coaxial design is the high voltage

post to which a 0.75 inch (1.905 cm) piece of solid aluminum cylindrical rod is fastened.

The outer conductor diameter is the inner diameter of the aluminum ring that is fastened

to the face of the pulser connecting the grounding studs. The inner diameter of the outer

conductor is 7.7 cm. Choosing appropriate diameters for the inner and outer conductors

15

Page 26: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

can maximize the breakdown voltage between the conductors at the face of the PCU.

The conductor radius ratio that yields a maximum breakdown voltage in a coaxial

geometry is provided by Equation 2.4 [4]. The variable Douter represents the inner

diameter of the outer conductor (the aluminum ring). The variable du er represents the

diameter of the inner conductor (the aluminum cylindrical bar). The ratio between the

outer and inner conductors is equal to 4.042. A figure presented by Smith in his article

shows that a ratio of 4.042 will present a breakdown voltage that is about 95% of

the maximum possible breakdown voltage. The maximum possible breakdown voltage is

attained with a diameter ratio of 2.718. Assuming the diameter of the outer conductor

cannot change, the diameter of the inner conductor would have to equal 2.833cm (1.115

inches) to attain the maximum breakdown voltage. However, given the performance of

the existing inner conductor and the availability of materials, no changes have been made

to the 0.75-inch aluminum rod, which delivers the high voltage pulse to the wire anay.

- 2.718 Ideal Diameter Ratio (2.4) dinner

Additional steps have been taken to insulate the high voltage rod from the

grounding posts at the face of the PCU. A lexan mounting bracket is fastened to the face

of the PCU by screwing it to the grounding studs. The mounting bracket extends inward

toward the high voltage post replacing the volume between the two conductors that is

filled with air wdth lexan, which has a higher relative permittivity, thereby increasing the

breakdown voltage. Silicon is used to seal any gaps that may exist between the lexan

mounting bracket and the face of the PCU. By appropriately sizing the diameter of the

inner conductor and insulating between the two conductors, energy loss due to corona

discharge or voltage breakdown at the face of the PCU has been minimized. Additional

steps could be taken to minimize any corona discharge between the two conductors at the

face of the PCU by filling the volume with a breakdown suppressing gas such as SFe

Considering the adequate performance of the existing layout, these additional steps have

not been taken to allow more time to be spent performing tests on the biological and

chemical simulants.

16

Page 27: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The wire anay has been designed with the intention of delivering all the energy

output by the PCU to the test specimen by means of a PCD. However, there are some

specific locations on the wire anay where it is virtually impossible to escape charge

accumulation resulting in energy loss. One of the main locations that unwanted

breakdown could occur is at the face of the pulser, which has already been addressed.

The other main locations for charge accumulation are at the ends of the copper rod

around which the wires of the anay are wrapped and at the ends of the wdres, which are

wrapped around the nylon tightening bohs. The ends of the cylindrical, 0.5 inch copper

rod have been rounded and sanded as smooth as possible to minimize field enhancement

at the ends. However, one cannot escape the fact that charge wdll accumulate at the ends

of a metalhc rod. The ends of the wdre are also an inescapable source of energy loss in

the system since each wire comes to a sharp point, which also tends to accumulate

charge. There is no arcing from these sources to the ground plate during operation of the

system because these sources of energy loss are located far from the ground plate. The

energy loss is assumed to be minimal compared to the energy delivered to the specimens

since there is no direct arcing or discharge to the ground plate from these locations.

2.1.2 Heat

When exposed to the arc jet plume, the test specimens are exposed to very high

temperatures by way of natural and forced convection of the air. It is of great interest to

know how much effect the heat has on the biological specimens during exposure. Is

heating the specimens a major force in killing the organisms? Does heating the simulants

tend to weaken them thereby making them more susceptible to destmction the killing

mechanisms of the PCD? These are questions that have been addressed in the

experiments performed. Tests have been performed that heat the biological specimens,

that expose the simulants to only the PCD, and that expose the simulants to a

combination of heating and PCD. The tests involving a combination of the two killing

mechanisms are performed as a cumulative process. The organisms are heated to certain

temperature for varying durations, then removed from the heat source and exposed to the

PCD for varying durations.

17

Page 28: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The biological specimens have been heated for increasing durations at different

temperatures. The chemical specimens have not been heated because the slides used to

plate the chemical compounds cannot withstand the temperatiires to which the specimens

would be exposed. In general, there is little interest in the effects of heat on the chemical

compounds. Refening to the biological specimens, there is a large amount of support and

data for the effects of temperature on living organisms. The spores must be exposed to a

higher temperature and longer duration in order to be killed. The E. coli, however, is

more sensitive to heat.

Studies have been conducted by Holwitt, Kiel, et al. conceming the thermal

sensitivity of biological warfare agents. The three biological warfare simulants studied

by this group for their susceptibility to heat were Bacillus anthracis var Sterne, Bacillus

thuringiensis var Kurstald, and Bacillus globigii var niger. "The survival of the spores of

all three organisms was studied at 130°C. This experiment... reaffirmed that BG

[Bacillus globigii] is perhaps the most heat resistant of the three types of spores studied.

Also, BA and BT spores are inactivated within 20 minutes at this temperature" [5].

Results wdll be presented in a later chapter that will closely agree wdth the work done by

Holwitt and his associates conceming the thermal sensitivity of spores. Based on

previous research such as that described above, it is necessary to explore the individual

and combined contributions of heat and the PCB in killing the biological simulants.

The heat source being used is a hot plate powered by 120 VAC. A layer of silicon

heat sink compound is placed on top of the hot plate and a thin sheet of copper that has

slightly larger dimension (length and width) than the top of the hot plate is placed on top

of the heat sink compound. The hot plate surface is not completely flat because of a lip

that extends around the edge. The heat sink compound is used to provide a good

conducting medium in the low regions of the hot plate where the copper plate may not

normally touch. On top of the thin copper plate is the grounding plate on which the test

samples wdll be placed. The mechanisms for heat transfer to the test specimens include

conduction and convection of the air directly above the hot plate. The purpose of the

above setup is to improve the uniformity of the temperature distribution across the

surface of the plate on which the test samples are placed. After taking temperature

18

Page 29: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

readings with an infrared temperature probe at the four comers and the center of the top

copper plate (refened to from now on as just the copper plate), it has been detemiined

that the temperature across the surface of the plate does not vary by more than 1 degree

Celsius CC).

The copper plate has been divided into four quadrants in which each set of test

samples is placed. Figure 2.3 shows the layout and location of the test samples, IR

temperatiire probe, and surface temperattire probe. A piece of masking tape is placed

beneath the IR probe to ensure the most accurate temperatiire reading. Masking tape has

a known and constant emissivity. The emissivity is set in the IR probe confrol module

and then fine-tuned wdth the help of the surface probe. The surface temperattire probe is

a type K thermocouple made by Omega Engineering hic. The emissivity of copper can

be detemiined, but the emissivity changes as the surface of the copper oxidizes. Using

the masking tape minimizes enors in temperature reading that may occur as the surface

of the copper plate changes over time and in response to the increasing temperatures.

Copper Plate

IR Temp. Probe

Masking Tape

•-0

i r

lOmin exposure

Smin. exposure

Surface Probe taped down

5min. exposure

1 min. exposure

(Overhead View of Test Slide Location)

Figure 2.3: Layout for Heat Exposure Tests

Because the test sample lies on a fiberglass coupon, it is important to determine

how well the fiberglass conducts heat from the copper plate. Experiments have been

19

Page 30: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

perforaied to detennine how long it takes the top surface of the fiberglass coupon to heat

to the same temperature as the copper plate. The infrared (IR) temperattire probe is used

to measure the temperatures of the hot plate and the top of the coupons. The maximum

time period of exposure is 10 minutes so this is used as the maximum time the coupon

has to heat up to the same temperature as the copper plate. The coupon is heated to

wdthin 5 degrees Celsius of the hot plate in approximately 1 minute. After 10 minutes,

the temperature of the surface of the fiberglass coupon remains 5 degrees cooler than the

copper plate.

Based on the experiments described above, the low temperature alarm set in the

data acquisition unit is 5 degrees warmer than the desired test temperature and a 1 minute

"heating" time is added to all duration times. This ensures that the top surface of the

coupon, where the specunen is located, is at least the target temperature for the entire

specified duration. Due to the nature of the heating coil within the hot plate, there are 5

degrees of hysteresis wdth the temperature confroller. If the temperature drops below the

set threshold, the controller applies power to the hot plate until the temperature rises

above the threshold. The previously applied energy continues to heat the hot plate by 3-5

degrees Celsius, and then the plate begins to cool. The slow response of the hot plate

wdth respect to the power applied to it is due to the slow speed wdth which the electrical

energy is converted to thermal energy and then conducted from the heating coil, through

the surface of the hot plate, through the silicon heat sink compound, through the 3/8 inch

copper grounding plate, and finally through the fiberglass coupon. Figure 2.4 displays a

graph plotting the variation of the temperature during exposure. The temperature

threshold is set to 80 °C using an Agilent data acquisition unit.

Exposures of different durations can be run at the same time by timing when the

specimens are placed on the hot copper plate and removing them after the desired

duration. The 10-minute samples are laid down first, then the 5,3, and 1-minute duration

samples. There are three specimens per test duration. The time of exposure is started and

stopped as the second specimen (of 3) is placed down on the copper plate and removed

from the copper plate, respectively. It takes about 5 seconds for a set of specimens to be

either placed on the copper plate or removed from the copper plate.

20

Page 31: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Temperature Variation at 80C

15

Time (minutes)

20 25 30

Figure 2.4: Temperature Variation of Hot Plate at 80°C

Voltage and cunent data has been collected for each of the temperature settings.

The voltage recorded is the RMS voltage of the AC input voltage and the conesponding

cunent. The RMS voltage is measured directly by the HP data acquisition unit. The

RMS cunent delivered to the hot plate is measured with a Pearson cunent monitor,

model number 4997. This cunent monitor outputs 0.01 volts per 1 Ampere. A droop of

0.3 % per millisecond (ms) is well within acceptable levels for use with the 60 Hz signal

from a 120 volt outlet. A 50 Q matching resistor is placed across the conductors of the

coax cable connected to the cunent monitor. A gain of 2 is set in the data acquisition

unit on the channel monitoring the hot plate cunent to account for the voltage division

that occurs between the 50 Q coax cable from the cunent monitor and the matching

resistor. Because of the inexact nature of the temperature and its fluctuation, it is not

necessary to log the voltage and cunent data during tests. One may refer to Figure 2.5 to

view the layout of the data acquisition system showdng connections for the hot plate.

2.1.3 Hydroxide (OH) Radicals

A pulsed corona discharge in atmospheric air will generate, among other things, a

certain number of OH" radicals by splitting water molecules apart. The OH" radicals are

very unstable and exist for only a few microseconds. Because of their reactivity, it is

believed that OH" radicals may play a role in the deactivation of both biological and

chemical simulants. By increasing the amount of water vapor in the air between the wdre

21

Page 32: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

anay and the ground plane, it is believed that a significant increase in the number of OIT

radicals directly around the specimens wdll occur. A fairiy complete set of data has been

accumulated on the effects that the increased levels of OH" radicals have on the spores.

Experiments exposing the chemical simulants to the increased levels of OH" radicals were

not performed because the main interests of the group concerned the viability of the

Bacillus spores when exposed to increased levels of OH" radicals.

An ultrasonic humidifier is used to generate the water vapor that is blown over the

wdre anay to increase the probability of OH" radical generation. The Sunbeam humidifier

(Model 696) has a 1.7 gallon capacity and is powered by 120 volts AC. The humidifier is

swdtched on and off through the relay formerly used to switch the hot plate and is

confroUed by the trigger pulse that triggers the function generator which in tum triggers

the PCD. The humidifier is turned on only during the PCD exposure. All of the

specimens to be exposed, regardless of intended exposure duration, are started at the

same time and as the appropriate time of exposure is accumulated, the conesponding

coupons are removed from underneath the wire anay and placed in sterile tubes. The

estimated volume of water vaporized by the humidifier over 10 minutes is 33.382 mL.

The estimated volume of water vapor applied to the volume directiy sunounding the wire

anay, ground plane, and simulants is calculated by measuring the amount of water

depleted from the humidifier's storage tank over a 10-minute period. Because the

ulfrasonic humidifier begins producing water vapor only 2 to 3 seconds after power is

applied, it is assumed that the amount of water vapor directed toward the wire anay

during the 1 minute exposure durations is immediate, consistent, and uniform. The water

vapor is directed to the volume of air between the wire anay and the ground plane using a

plastic snout approximately 3 inches in diameter. The snout increases the concenfration

of water vapor directed toward the desired volume directly sunounding the simulants.

2.2 Data Acquisition

The data acquisition system consists of all equipment needed to trigger the PCU

and record the cunent and voltage waveforais, tidgger voltages, trigger frequency, and

temperature. The cunent and voltage waveforms output from the PCU front face panel

22

Page 33: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

are recorded using a Hewlett Packard Infinium oscilloscope. A computer mnning

Windows 98 and the Agilent 34970A Data Acquisition/Switch Unit are used to record

and monitor the temperatiire as recorded by the infrared temperature sensor, the tidgger

pulse voltage output by the Global Specialties function generator, the tidgger pulse

frequency delivered to the PCU, and the AC voltage and cunent applied to the hot plate.

As a Senior project Mike Hoffman was responsible for the initial set up and trouble

shooting of the data acquisition unit and the opto-isolation circuits. Mike also aided in

general trouble shooting of the data acquisition system until graduating December 2001.

Because of the high levels of electromagnetic interference and noise generated by

the PCD, considerable steps have been taken to filter and minimize noise on the signal

and power lines. Techniques used to minimize noise include opto-isolation, the use of

low pass fihers, the use of fenite toroids to increase in-line inductance, and the use of

screen rooms around both the data acquisition equipment and the PCU. The main

purposes for acquiring the data specified above are to enable the group to calculate the

electrical energy delivered to the test specimens and to monitor the performance of the

system during exposures to ensure consistency and repeatability or make note of the lack

thereof Figure 2.5 shows a schematic of the data acquisition system.

2.2.1 Oscilloscope Settings

The Hewlett Packard Infmium 500MHz, 2 giga-samples per second

(GSa/s) oscilloscope is used to monitor and record the voltage and cunent waveforms

output by the PCU which generate the pulsed corona discharge. The oscilloscope is set to

sample at a rate of 1 GSa/s for the tests using the nanosecond desktop SOS-based pulser.

The voltage waveform measured is a scaled version of the voltage delivered to the load.

The scaling of the voltage is performed by a resistive voltage divider within unit A2 of

the PCU circuitry. The cunent waveform measured is a measurement of the cunent

passing through the solid state opening switches. The cunent is measured with the use of

a cunent viewing resistor having a resistance of about 0.51 Q. The oscilloscope actually

measures the voltage drop across this cunent viewdng resistor. One must appropriately

set the gain for the probes of the respective channels to accurately display the cunent and

23

Page 34: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

voltage delivered to the load. Coax connectors labeled "U" and "Us" located at the front

face of the PCU enable one to connect the measurement points to the oscilloscope using

50 Q coax cable.

Programmable F u n c t i o n G e n e r a t o r

.jvoc „a Trifln., HP P u l s e

Sci reen Room

G e n e r a t o r lOOH; I3.3y>

Dscilloscope

•DptoIsolator ~ircuit

Agilent DAQ Unit

DC Power Supply

<S5>

T .:. Ul

<S22> <22>

<S&

12 0VAC Isolated Power Supply

IDOHz 13.3-Nt

^ 1 ^ [Relay

Fer r i t e toroid around ohich the l ine is crrapped t:a miiDliiize noise on the l i ne .

S c r e e n Box

t ^

<s>

IR C o n t r o l l o d u l e

<2&

PI Network

SOS-Based P u l s e r

C,f2 ., 85uf "

85uF

IprobeL AC Cunent lAagijrimtrft

/iC ^Atoflft MeasvjrwTvent

Power S t r i p

X

120 vac

Hot P l a t e

H u m i d i f i e r

IR P robe

Figure 2.5: Schematic of the Data Acquisition System

The cunent waveform is displayed on channel 1 of the oscilloscope. The cunent

conesponding to the measured voltage can be calculated using Ohm's law shown in the

equation below where V is the voltage, I is the cunent, and R is the resistance.

1 = V R

(2.5)

Refening to the circuit in Figure 2.6, one can see that the value of the cunent viewdng

resistor is 0.51 Q. This resistor is also in parallel wdth an equivalent resistance of 50 Q

because of the 50 Q termination at the oscilloscope; however, because 0.51 Q is much

smaller than 50 Q, this parallel combination has been neglected. Plugging the value for

the cunent viewing resistor into Ohm's law and reananging the fonnula to show volts per

24

Page 35: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

ampere, one sees that the necessary gain is the magnitiide of the cunent viewdng resistor.

Explicitly, the relationship between the voltage measured and the cunent flowdng through

the switch is such that 0.51 volts is equal to 1 ampere. The external gain for Channel 1 is

set to 0.51. A 20-decibel (dB) attenuator is attached at the input to Channel 1 of the

oscilloscope to fiirther reduce the magnitiide of the signal from the PCU so that the

voltage maximum input for the oscilloscope is not exceeded. The additional 20 dB of

attenuation are accounted for in the oscilloscope channel properties by setting the

attenuation of the channel to 20 dB. By choosing the appropriate gain and attenuation

settings, the voltage measured across the cunent viewdng resistor is viewed and recorded

at channel 1 of the oscilloscope as the cunent passing through the solid state opening

switch in unit A2 of the PCU.

L1

(Magnetic Switch section of unit A2)

025... 040

206.25 pF

V S0S1

V S0S2

R16...R31 75.2 kohm

R6...R15 0.51 ohm

RIoad

R32 10 ohm

'Is'(Current Monitor)

R33 10 ohm

'Us'(Voltage Honitor)

Figure 2.6: Location of Cunent and Voltage Monitors in A2 Unit of PCU

The voltage waveform is viewed on Channel 2 of the oscilloscope. The voltage

waveform recorded by the oscilloscope is a scaled version of the voltage pulse delivered

to the load. The amount of scaling is determined using the formula for a resistive voltage

divider. The values of the resistors may be seen in the circuit of Figure 2.6. Resistors

R32 and R33 are also in parallel wdth an equivalent resistance of 50 Q because of the

50 Q. termination at the oscilloscope yielding an equivalent resistance of 4.545 Q for the

low voltage leg of the voltage divider denoted as RLV in the voltage divider equation

below. The equivalent resistance of the series connection of resistors R16 through R31,

25

Page 36: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

denoted as Ri6 31 in the equation below, is 75.2 Ul. (Notice that this is sixteen 4.7 kD

resistors in series). In Equation 2.6, V„,easured is the voltage measured by the oscilloscope

and Vioad is the voltage delivered to the load. Using these values for the resistors of the

voltage divider yields a voltage divider ratio of 6.0 x 10 ^ However, the minimum

external gain that can be set on the oscilloscope is 1 x 10" To overcome this limitation,

the external gain of the oscilloscope is set to 6.0 x 10" , a 20 dB attenuator is connected to

the Channel 2 input, and the attenuation is set to 20 dB in the probe settings menu.

RLV ^measured - Vioad'T T (2.6)

R L V + R l 6.31

2.2.2 Agilent Data Acquisition/Switch Unit

The Agilent 34970A Data Acquisition/Switch Unit is used wdth a computer

running Windows 98 to record the additional data required to monitor the performance of

the system, control the temperature of the hot plate when desired, and record the data

necessary to calculate the energy applied by the heat tests using the hot plate. The data

acquisition unit communicates wdth the computer by way of an RS-232 communication

port. The data acquisition unit comes wdth software that enables communication with the

computer and provides a graphical user interface for the real time display of data. From

the computer's user interface, one can easily export the data recorded during a test run to

a '.CSV' file for later use with a spreadsheet or other appropriate software tool.

Refer to Figure 2.5 for a schematic representation of the explanations provided in

this paragraph. The temperature is measured at Channel 1 using an infrared probe that

outputs a voltage between 0 and 5 volts conesponding to the temperature in degrees

Celsius. The duration of the trigger pulse output by the Global Speciahies function

generator is monitored at Channel 2. The trigger pulse is passed through an opto-isolator

before entering the unit so the voltage recorded is the voltage output by the opto-isolator,

not the voltage output by the Global Specialties function generator. This function

generator is programmed to output a positive voltage of about 10 volts for a duration of 1

minute, then an output of 0 volts for a duration of 5 minutes. The output of the Global

Speciahies function generator is sent to the external tidgger connection of the Hewlett

Packard pulse/function generator; the low voltage terminals of the relay which switches

26

Page 37: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

the 120 VAC power sfrip in the PCU screen box on or off; and the input to channel 2 of

the data acquisition unit via an opto-isolator circuit. The cycle of 1 minute on and 5

minutes off is used to control the duration of exposure of the specimens and is necessary

to prevent over heating of the PCU as mentioned previously in the chapter. Channel 3 of

the data acquisition unit measures the frequency of the trigger pulse delivered to the PCU

by the Hewlett Packard 811 lA pulse/function generator. The output from the Hewlett

Packard function generator is set to a 100 Hz square wave wdth amplitude of 13.3 volts

and an offset of 7.2 volts. The resulting waveform delivered to the PCU is a 100 Hz

square wave with a peak voltage of 10 vohs and a minimum voltage of 0 volts which

meets the requirements by the PCU to have a drive pulse of 10 to 15 volts. The voltage

drop is assumed to be due to loading by the PCU. By operating this function generator in

gated mode, it wdll only apply the set output when it receives the 10 volt trigger pulse

output from the Global specialties function generator. Channels 4 and 5 are reserved for

use when running experiments wdth the hot plate. Channel 4 directly measures the AC

voltage applied to the hot plate. Channel 5 measures the voltage output of a Pearson

cunent monitor that is positioned around the conductor of interest. The Pearson cunent

monitor outputs 1 mV per ampere. The gain for Channel 5 is set to a value of 2 to

account for the 50 Q terminating resistor added to the line which matches the line

impedances, but also creates a voltage divider circuit that cuts the desired voltage in half

The only output from the data acquisition unit is an alarm that is only used during

the experiments with the hot plate. When the temperature of the hot plate falls below the

threshold set using the alarm, the data acquisition unit sends a +5 voh signal to the relay

which switches the AC power strip inside the PCU screen box on and off. The

temperature sensed by the infrared sensor, which is recorded at Channel 1, is the

temperature monitored by the alarm. This allows automation of the temperature control

and increased consistency conceming the temperature variation for a given temperature

during tests.

27

Page 38: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

2.2.3 Solutions to Electromagnetic Interference (EMI)

The pulsed corona discharge radiates elecfromagnetic interference in all

directions. The signal lines and power lines leading to and from the PCU are injected

wdth noise because of this EMI. The Agilent data acquisition unit, the Global Specialties

synthesized function generator, and the HP Infinium oscilloscope are the most affected

by increased levels of noise on the lines. The four main techniques that have been

employed to try to minimize the presence of the noise on the lines include a PI network to

filter the DC supply line to the PCU, opto-isolation to filter the trigger duration and

frequency inputs to the data acquisition unit, fenite toroids used to increase in-line

inductance presenting a higher impedance to high frequency signals, and physical

enclosure of the PCU wdth a copper screen box and enclosure of all instrumentation wdth

a different screen room.

When refening to the data acquisition system schematic shown in Figure 2.5 one

wdll notice a block labeled PI network. The PI network is a low pass filter used to filter

the DC input power to the PCU (Figure 2.7). Because the power being filtered is DC,

the values of the capacitors and inductors merely need to be sufficient in size. No tuning

is necessary as one may find the need to do wdth other applications. The circuit takes

advantage of the impedance characteristics of the circuit components. The impedance of

a capacitor is equal to 1/jcoC and the impedance of an inductor is jcoL; co being the

frequency of the signal [6]. By inspection, one can see that the impedance of a capacitor

is very low for a high frequency signal and decreases with increasing signal frequency

and capacitance. The impedance of an inductor is relatively high and increasing wdth an

increase in the frequency of the signal and the inductance. To the noise on the DC power

fine, the capacitors act like shorts and the inductor acts like a high impedance, while the

DC power passes through the circuit with little hindrance because of its zero frequency

content. The PI network is encased in a metal box and placed very close to the PCU.

28

Page 39: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

T C1

L1 C2 T

Figure 2.7: PI Network

A simple circuit employing the 5601 opto-isolator is used to minimize the effects

of noise on channels 2 and 3 of the data acquisition unit (Figure 2.8). Channels 2 and 3

record the trigger durafion output by the Global Specialties function generator and the

frequency of the trigger pulse delivered to the PCU, respectively. The opto-isolator

physically separates the function generators from tiie data acquisition along with any

noise that may be induced on the lines. Notice that the voltage of the signal measured by

the data acquisition unit is the voltage from the opto-isolator circuit, not the voltage

output by the function/pulse generators. Mike Hoffman designed and built the opto-

isolator used for this project.

From G e n e r a t o r

R1 ^AA vvv 330

1

7

^Q>

Nb(. PS2

t 501

ro

VCC = +9VDC 4 V

< R2 > 331k

To DAQ

Figure 2.8: Opto-isolator Circuit

Ferrite toroids are used at multiple places in the signal and power lines to increase

the in-line inductance which tends to inhibit the passage of high frequency signals such as

the noise induced on the lines from the EMI. By wrapping the signal line around the

ferrite toroid, one is essentially creating an inductor with an inductance based on the

number of turns. As mentioned during the explanation of the PI network, the impedance

of the element increases as both the inductance and frequency of the signal increases.

29

Page 40: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

One may refer to the data acquisition system schematic shown in Figure 2.5 to note the

locations of the ferrite toroids, which are designated as an oval filled wdth hash marks.

The inductance of a closely wound rectangular toroidal coil can be found using Equation

2.7. In the equation, N is the number of turns, 'a' is the inner radius of the toroidal core,

b is the outer radius of the toroidal core, h is the height or thickness of the toroidal core,

and ^ is the permeability of the toroidal core [7]. Assuming that the dimensions of the

toroidal core are constant, notice that the inductance changes wdth the square of the

number of turns (N). Increasing the number of turns around the fenite core increases the

inductance, which increases the impedance seen by the high frequency noise signals.

f -N -h 2.U ru\ L= In -

2n

h

va; (2.7)

The last method used to minimize noise on the lines induced by EMI is the use of

a copper mesh screen room that sunounds all data acquisition equipment and a copper

mesh screen box that sunounds the PCU and all equipment needed to change

environmental conditions during the exposure process. The screen room limits the EMI

that radiates into the volume sunounding the data acquisition equipment. In addition, the

power strips provided inside the screen room are isolated from outside power sources by

way of an isolating transformer. The screen box performs the opposite objective as the

screen room; that being to keep as much of the EMI inside the box as possible. The

screen room and screen box have BNC connector panels that allow one to pass the signal

lines through the box with minimal gaps in the panel. There are also holes in the panel

that permit one to run lines directiy into the screen room/box. These holes are patched

with metal tape after all desired lines are run through them.

2.2.4 Power Supplies

There are two DC power supplies used in this data acquisition system. One of the

power supplies is used to power the infrared temperature sensor control module. The

second power supply is a 600 volt, 1.7 ampere DC power supply made by Xantiex and

used to supply the main power to the PCU. While the ratings of the power supply exceed

the power required by the PCU, the Xantrex power supply cannot recharge the primary

30

Page 41: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

storage capacitors of the PCU fast enough to maintain the desired voltage at a pulse

repetition rate of 100 Hz. This inability to recharge the primary storage capacitors causes

the pulsed corona discharge at the load to cut out sporadically. In order to help the power

supply maintain the necessary charging voltage, three 85 |xF capacitors have been added

in parallel to the power supply. One may view the locations of these capacitors in the

data acquisition system schematic shown in Figure 2.5. The capacitors are an unlytic

type UL30 lOOOV DC capacitor manufactured by Electronic Concepts wdth a tolerance of

plus or minus 10%. These capacitors are initially charged by the power supply and

during the firing of the PCU, act to provide quick bursts of charge as needed whenever

the power supply cannot keep up with the repetition rate.

31

Page 42: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER III

CHEMICAL AND BIOLOGICAL AGENTS

3.1 Biological Agents

3.1.1 Background

The microorganisms that were used in the pulsed power experiments were

selected after consideration of their sensitivity to the separate elements of the

experiments, and the simplicity in analyzing them in the resuhs of the experiment.

Escherichia coli strain 802 was chosen for its tetracycline marker, which was used to

screen against contaminants that may have been inttoduced in tiansport and

experimentation of the microorganism. The tefracycline marker would allow the E. coli

to grow on tetracycline culture medium while preventing growth of any contaminants.

The E. coli used in pulsed power experiments is a general bacterial paradigm, and was

used preliminarily to check for difficulties or variations in the field. E. coli also tested for

the effectiveness of the pulsed power machine in reducing cell number. Experimental

analysis of E. coli was mostly uncomplicated because it is one of the most researched and

documented microorganisms, and therefore planning for preparation and analysis could

not be more uncomplicated.

Bacillus stearothermophilus is a spore forming microorganism, generating a

microorganism that is more thermally and chemically resistant than many other known

Bacillus. This microorganism was chosen because it is documented to be less thermally

and chemically sensitive to spore inactivation than bacillus anthracis and because it

could be purchased from NAMSA containing the desired number of cells per milliliter

for experimentation. Also, the germination temperature of the bacillus was 55 degrees

Celsius, which is higher than virttially any other possible contaminant, thereby almost

eliminating contaminants from analysis.

3.1.2 Preparation

The fiberglass coupons that carried the microorganisms for ttansport and

experimentation were obtained from the elecfrical engineering departinent and sterilized.

32

Page 43: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The sterilization process included first washing the coupons with biological detergent,

and then multiple times wdth DI water. The coupons were then placed inside a sterile

glass petri dish and sterilized in an autoclave for twenty minutes and 18 to 20 psi. The

coupons were then removed and dried in a sterilization oven at 136 °C as a final

safeguard.

E. coli was prepared for experunentation by first isolating a colony to be cultured

then aseptically placing it in two milliliters of sterile, liquid cultiire medium containing

tetracycline for ovemight incubation at °C. This temperature is ideal for maximum cell

growth. After ovemight incubation the broth culture typically contained 10 cells per

milliliter. Twenty microliters of broth culture were then ttansfened aseptically to sterile

fiberglass coupons, and placed in sterile glass petid dishes for ti-ansport. The cell

concentration per coupon is approximately 10 per milliliter.

Bacillus spores were prepared for experimentation by aseptically removing them

from a spore suspension with a sterile syringe and placing them in a sterile container.

Twenty microliters were removed for each individual coupon and these were then placed

in a sterile glass petri dish for fransport.

There were many controls used in transport and experimentation of both types of

microorganisms to assure the most accurate results for analysis. The coupons containing

the microorganisms were placed in a Styrofoam container during transport to attempt to

control temperature equilibrium, and therefore try to reduce cell death. Also, there were

coupons that remained in the microbiology laboratory to determine if transport was a

factor in cell loss for every experiment, and these coupons were used in analysis of the

results of the experiment. Coupons that were fransported to the pulsed power laboratory

but not exposed to any tests were used as a confrol to try to determine any possible cell

loss that might have occuned. These coupons were also used in analysis of coupons

exposed to any temperature or electrical field.

3.1.3 Coupon Analysis

After coupons were exposed and transported back to the microbiology lab they

were analyzed. The exposed coupons were placed in sterile plastic tubes along wdth two

33

Page 44: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

milliliters of buffer salts, and then vortexed to remove cells from the coupon. The

microorganisms were then assayed using a serial dilution technique for each individual

coupon, which was suspended in Phosphate Buffered Saline solution and placed in sterile

test tubes. After serially diluting them, the bacterial cells were plated on nutrient agar

plates, incubated ovemight (12-14 hrs) at 37°C for E.coli 802 and at 55°C for Bacillus

stearothermophilus. The percent survival was determined and compared wdth all the

confrolled samples.

The explanation for the use, source, preparation, and evaluation of the biological

agents was provided by Kathryn Royce and verified by Dr. Joe Fralick in the Texas Tech

University Health Sciences Center Microbiology department. Kathryn Royce prepared

and evaluated all biological agents that were used for this project. Kathryn also helped

handle the test specimens during the exposure process. Dr. Fralick has been the main

source for expertise and advice on any issues conceming the biological specimens.

3.2 Chemical Agents

The chemical agents chosen for use in this project are Malathion and parathion.

The agents have been chosen based on their relatively low vapor pressure, comparatively

low human toxicity, and availability. Dr. Karthikeyan Sathmgnan, a post-doctoral

researcher in the Texas Tech University Chemistry department, under the supervision of

Dr. S. Dasgupta, has prepared the chemical simulants for exposure and has evaluated the

test slides after exposure. Preparation of the slides consists of plating 1-inch square glass

slides wdth 20 nanomoles (nmole) each of Malathion or parathion. Two hundred fifty

nmoles conesponds to 82.5 jug of Malathion and 72.75 j^g of parathion. A normal test

group consisted of around fifteen slides providing three different exposure durations and

one control set wdth three test samples in each set. After exposure, the slides are washed

with 0.5 mL of acetonitrile and the washes are analyzed by liquid chromatography for

their parent compound and metabolites. There have not been any significant problems

with the preparation or evaluation of the chemical compounds. Dr. Dasgupta has been

the main source for expertise and advice on any issues conceming the chemical

specimens.

34

Page 45: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER IV

CHEMICAL AND E. COLI TEST RESULTS

4.1 Method for Calculating Results

Before presenting the specific resuhs for each simulant under different exposure

conditions, the method by which the results have been calculated wdll be explained.

Because different energy sources, such as the PCD and heat, are being used as killing

mechanisms, a common method to compare the various killing schemes has to be

established. The total energy applied to the specimen has been chosen to be this point of

comparison. The energy applied will be presented on two different scales. One scale

describes the total energy, in Joules (J), applied per specimen area. The specimen area is

the area of the slide on which the coupon is placed. The specimen area for the chemical

slides is 6.25 cm^ and the specimen area for the biological slides is 1.21 cml The second

scale describes the equivalent total energy, in kilo-Joules (kJ), that it is estimated would

be necessary to provide the same results on a 1 square meter area of surface. The main

assumption made when calculating the energy is that the electric field of the PCD to

which the simulants are exposed is relatively uniform.

In order to calculate the energy of the PCD, the voltage and cunent applied to the

load are recorded by the oscilloscope and saved as data files. Voltage and cunent

waveforms are saved for the different durations of exposure during a given test. For

example, if a test consists of 1-minute, 3-minute, and 5-minute exposure durations, there

will be three sets of voltage and cunent waveforms saved. The oscilloscope is also set up

to record the mean peak voltage and cunent and the standard deviation of these mean

values. The voltage and cunent waveforms cannot be multiplied together, because they

are not recorded at the same point in the circuit within the PCU. However, the cunent is

assumed to be at its maximum value and fairly constant during the peak of the voltage

pulse. The mean current value may come from one of two different sources. The

prefened source is the mean peak cunent value recorded by the oscilloscope. If this

value is not available, the mean peak cunent value is calculated from the cunent pulses

recorded for a given test. The voltage waveforms for a given test are averaged together.

35

Page 46: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The power delivered to the load is then calculated by multiplying the averaged voltage

waveform by the mean peak cunent value. The energy of a single pulse is then found by

integrating the power pulse over the duration defined by the Full Width at Half Maxunum

(FWHM) of the power pulse. The total energy applied over a certain duration is

obtained by multiplying the energy for one pulse by the repetition frequency of the PCU

and the total number of seconds of exposure. This total amount of energy is estimated to

be the energy applied over the area directly beneath the wire anay. This energy is

appropriately scaled down to obtain the energy applied to a single specimen and

appropriately scaled up to obtain the energy per square meter.

The energy applied to the hot plate in order to heat the plate and maintain the

given temperature is calculated using the electidcal energy applied to the hot plate.

Voltage and cunent data has been collected for each of the temperature levels used for

exposure during separate tests mn for the purpose of estimating the energy applied to the

simulants. The voltage and cunent recorded are multiplied together and then integrated

over the duration of exposure in order to obtain the energy applied to the entire surface of

the hot plate. As with the PCD, the energy applied to the area of the hot plate is

appropriately scaled to give the energy applied to a single specimen and the energy

needed to be equally effective across a square meter surface. The characteristics of the

heating and cooling of the hot plate have been observed to be very consistent, so the

energy calculations are assumed to be sufficient and representative of any test at a given

temperature.

The level of destmction will be presented in terms of either percent survival of the

simulant or the number of cells or spores killed (labeled "Cell Count Reduction" or

"Spore Count Reduction"). The resultant percent survival, or percent growth for

chemical degradation products, is calculated as a percentage of the mass of the control for

the chemical compounds and a percentage of the cell or spore count of the confrol for the

biological agents. Except for being exposed, the control specimens are submitted to the

same conditions as the exposed specimens. This minunizes inconsistencies in results due

to preparation, storing, transporting, or evaluation of the specimens. The other main

factor that may cause variations in results for individual specimens is the location of the

36

Page 47: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

specimen under the wire anay because the PCD generated by the wire anay is not

entirely uniform. The test specimens are more quickly destroyed by the PCD when they

are placed directly beneath a wire of the wire anay. For the biological simulants, the

most consistent tenninology lies with the percent survival because not all biological tests

are conducted wdth the same concentration of simulant. The final results are averages

wdth a given standard deviation and are shovm in graphs and tables that are provided

within the text of this paper or in Appendix A. Layouts for test run have also been

included in the Appendix C.

4.2 Chemical Experimental Results

4.2.1 Exposure of Malathion to the Pulsed Corona Discharge

Malathion has been exposed to the PCD for durations ranging from 2 seconds to 5

minutes. Repetition frequencies of 10 Hz, 100 Hz, and 200 Hz are also tried. Based on

the PCU performance, a pulse repetition frequency of 100 Hz is used for the tests that

yield noticeable results. No noticeable reduction in the amount of Malathion begins until

an exposure duration of 1 minute is reached. The height of the wdre anay above the

ground plane is 2.5 cm and the wdre anay spacing for these tests is 0.25 inches. The first

wdre anay stmcture has been used for these tests. In addition to the reduction of

Malathion, the generation of its degradation product, DDTP (Dimethyldithiophosphate),

can be seen in the chromatograms which show the Malathion before exposure to the PCD

and after 3 minutes of exposure (Figure 4.1). Remember from the previous chapter that

the concentration of Malathion on each slide is 250 nmole, which conesponds to 82.5 ^g

of Malathion. The chromatograms have been provided by the Chemistry department at

TTU.

Graphs showing the full scope of the results obtained from this set of tests may be

seen in the Figures 4.2 and 4.3. The energies along the x-axis of the graphs conespond to

0, 1, 2, 3, and 5 minutes of exposure. Each minute of exposure is separated by 5 minutes

of rest time that allows the PCU to cool. Notice that after five minutes of exposure, about

15% of the Malathion still remains and the DDTP has not begun to diminish. Increasing

the exposure time to durations greater than five minutes would obviously results in

37

Page 48: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

complete destmction of the Malathion. At some point past the five-minute mark, the

generation of the DDTP would be expected to reach a maximum and proceed to be

eliminated by the PCD.

Malathion control Malathion after 3 min expsoure

0.12 1 £ 0.1 -S 0.08 -T- 0,06 -d 0.04 -= 0,02 -

0

Unkmow f

Mala&ion

/-"-

i^ ^ ^

Time in seconds

0.08 -,

•S 0.08 •

g 0.04 •

S 0,02 •

0 J

DDTP

> '

h - T^ T - 00 U l CN O) <D (O CN CN (O •«• ^ Wl CD

Time in seconds

o

Figure 4.1: Destmction Process of Malathion

The data points in Figures 4.2 and 4.3 indicate the average percent survival after

exposure to the given energy levels. The bars through the data points on the graphs

indicate the standard deviation of these averages. The inconsistencies in the percent

survival of the Malathion are atfributed to the non-uniformity of the PCD and preparation

and evaluation enor. The test specimens are placed at different locations undemeath the

wdre anay in order to obtain a complete survey of the killing effectiveness of the wire

anay. During the experiments, it has been found that the Parathion is a more robust

compound so further tests wdth Malathion were not performed in an effort to concenfrate

on the destmction of Parathion.

4.2.2 Exposure of Parathion to the Pulsed Corona Discharge

Parathion has been exposed to the same initial conditions to which the Malathion

has been exposed. These tests are also performed with the same experimental setiip as

the tests perforaied on Malathion. Like Malathion, Parathion also has a degradation

product, Paraoxon, which is of interest and that must be broken down to completely

eliminate the threat of the compound. The initial concenfration of tiie Parathion is 250

38

Page 49: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

nmole conesponding to 72.5 |Lig of mass. The Parathion has been exposed to durations

that exceed 5-minutes in order to determine the energy necessary for a complete kill.

100%

3 75%

e ^ 50%

1 0) (J

I 25% Q.

Malathion Survival (PCD Only)

0% -+- -+•

470 850

Energgf(kjy»n' 2)

1340 2230

Figure 4.2: Malathion Survivability When Exposed to the PCD

DDTP Present (PCD Only)

10%

470 850

Energy (kJ/m*2)

1340 2230

Figure 4.3: Percentage of DDTP Present When Exposed to the PCD

Figures 4.5 and 4.6 show the average percent survival of the Parathion and the

generation of Paraoxon as a fimction of the total applied average energy representing

exposure durations of 0,3, 5,10, and 15 minutes. After 15 minutes of exposure to the

PCD, the Parathion is completely desfroyed and the Paraoxon is reduced to about 18

percent of the initial mass of tiie Parathion Because it has been observed tiiat the

39

Page 50: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Malathion is more easily killed by the PCD, it is believed that the Malathion would also

be completely destroyed after a 15-minute exposure to the PCD.

Paratliion Survival (PCD Only) 100%

s e O

a.

1300 1990

Energy (kJ/ni 2)

3980 5B40

Figure 4.4: Parathion Survivability When Exposed to the PCD

100% Paraoxon Present (PCD Only)

1990

Energy (lcj/m*2)

Figure 4.5: Paraoxon Present When Exposed to the PCD

Chromatograms showing the degradation of the Parathion and the growth of the

Paraoxon may be seen in Figure 4.6. The first chromatogram (on tiie left) shows the

amount of Parathion present before exposure to the PCD. The second chromatogram (on

the right) shows the amount of Parathion and Paraoxon present after five minutes of

exposure to the PCD. The first step in neufralizing the Parathion is breaking the

Parathion down into its degradation product, Paraoxon. The Paraoxon is then broken

down into a harailess compound. This is why one notices that the Paraoxon first grows

in percent concentration, reaches a maximum value, and then begms to degrade toward

zero percent. The chromatograms have been provided by the Chemistty department at

TTU.

40

Page 51: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Parathion Control

0,3

02 5 O-is •

S. 0,1 • 3 0,05

Patathion

Time in seconds

Parathion after 6 min exp.

0.3

ii 0.1

0

PNP

A Paraoxon

y FaratMon

g ' 67 133 199 265 331 397 483 529 595 661 72;j|

-0.2

Time h seconds

Figure 4.6: Parathion Chromatograms

4.3 E. coli Experimental Results

While E. coli is a harmful microorganism that should be taken very seriously, the

main focus of the project was not aimed toward finding methods that effectively kill E.

coli. There are already many methods for killmg E. coh which work very well and are

very convenient. When compared to a Bacillus spore, E. coli is relatively easy to kill.

The purpose for including E. coli in the experiments was partly as a pre-test. If a certain

exposure method cannot kill E. coli, it wdll certainly not be able to kill Bacillus spores.

Because of the expense of the Bacillus spores, it has been the practice of this group to

validate the effectiveness of a given exposure method by first killing E. coli and then

attempting to kill Bacillus spores. In addition, the data taken while running the tests with

E. coli may be a very good source for later research. The results and any considerations

regarding specific test conditions are therefore included in the following sections.

The exposure methods that have been used on the E. coli include exposure to:

only the PCD, only heat, heat and the PCD, and PCD wdth water vapor. One may notice

in some of the graphs that the standard deviation of the percent survival is quite large.

Keep in mind that E. coli is relatively easily killed and tiie task of maintaining a secure,

controlled environment while fransporting these microorganisms between two labs is not

a trivial task. The specimens are prepared and later evaluated in the Micro-Biology Lab

in the Health Sciences Center (HSC). This requires fransporting the specimens across

campus (3 miles each way). The number of cells on the conti-ol samples (Control 1)

taken to the Electrical Engineering building have always been less than the number of

41

Page 52: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

cells on the control samples (Control 2) that are left at the HSC. If E. coli

decontamination experiments should contmue, improvements for preparing and handling

the specimens will need to be made. A new source for obtaining the E. coli specimens

should also be considered.

4.3.1 E. coli Exposed to the Pulsed Corona Discharge

The durations for which E. coli has been exposed to the PCD include 2 seconds, 5

seconds, 10 seconds, 30 seconds, 1 minute, 3 minutes, 5 minutes, and 10 minutes.

Different pulse repetition frequencies have also been tried, which include 10 Hz, 100 Hz,

and 200 Hz. The DC voltage input to the PCU was varied between 450 V and 600 V.

While all of these conditions were tried at one time or another, the optimum operating

conditions based both on the PCU performance and the results were found to be exposure

durations between 3 and 10 minutes, a pulse repetition frequency of 100 Hz, and a DC

voltage input to the PCU of 600 V. The other main variable present during the tests is the

location of the specimen under the wire anay. Better kill rates are obtained when the

specimen is placed directly beneath a wire of the wire anay because the PCD is not

completely uniform. For shorter exposure durations, the percent survival is half as high

directiy beneath the wdre. For the peak exposure duration in these experiments, there is

littie or no difference between results based on the location of the coupon. The first anay

stmcture is being used with a wdre spacing between the wdres of the anay of 0.25 inches

for these tests.

The graphical results presented in this paper are an average of all the results

regardless of the location of the specimen. This is a significant contribution to the

standard deviation of the averages. Even though the PCD is a cold plasma, one is conect

to assume that it does generate some thermal energy. However, the surface temperature

of the specimen coupons never exceeded more tiian about 10 °C above the ambient room

temperature and is assumed to not be a significant killing factor during the PCD

exposure. One may refer to the Heat Only section to see data showing temperatiires

necessary to facilitate a timely kill of the E. coli.

42

Page 53: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The process of exposing the specimen consisted of exposing four slides at a time

with two coupons directly beneath a wire and two coupons centered between two wires.

The four coupons are exposed to a certain duration. After the desired duration of

exposure has been reached four new coupons are placed in the same four locations under

the wire anay and a different duration of exposure is performed. The exact location of

the four slides did change between tests, but there were always two coupons directly

beneath a wdre and two coupons centered between wdres. Placing the coupons at different

locations under the wire anay helped to verify the relative uniformity of the PCD.

Figure 4.7 shows the survivability of the E. coli and one can see that the average

killing rate is relatively linear and decreasing to zero from 3 minutes to 10 minutes of

exposure. The data points in the graph conespond to 0, 3, 5, and 10 minutes of exposure

time, respectively. While the standard deviations for the 3 minute and 5 minute durations

are affected by the factors previously discussed, 10 minutes of exposure to the PCD

results in a virtually complete kill on a consistent and repeatable basis. The actual values

after 10 minutes of exposure reveal a 0.39% survival with a 0.47% standard deviation.

This standard deviation is attributed to the differences in specimen location.

E. Coll Survival (PCD Only) 100%

h H

S e (J

'S •£ w u 09

75%

6U%

25%

0% 1230 1880

Energy <kJrtii*2)

3580

Figure 4.7: E. coli Survivability When Exposed to the PCD

4.3.2 E. coli Exposed to Only Heat

It is believed that the heat generated by the PCD is not a significant factor in the

killing of the biological or chemical agents. However, in order to verify this assumption,

both E. coli and Bacillus spores are heated to increasing temperatures for varying

43

Page 54: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

durations. A secondary and more important motive behind the investigation of this

environmental enhancement is to determine at what temperature the destiiiction of the E.

coli becomes significant. Because a hot plate is used as the thermal energy source, it is

assumed that the air immediately above the hot plate is the same temperature as the hot

plate itself However, even though the specimen is plated on a thin piece of fiberglass,

the fiberglass still has some degree of thermal insulation. Because of the thermal

properties of the fiberglass coupon, the tests are mn so that there is a one minute heating

period allowed at the beginning of each test to give the fiberglass coupon time to heat to

the desired temperature.

The tests are performed by placing the coupons on the hot plate in groups of three

in increments of 30 seconds per group of three. There are normally three or four different

exposure durations run per test. As a desired duration of exposure is reached, three or

four coupons are removed from the hot plate and placed in sterile tubes. The main

benefit of this process is that it saves tune by allowdng one to run different exposure

durations at the same time. Figure 4.8 displays the survivability of E. coli when exposed

to a temperature of 50°C.

E. coli Heat Only Test (COC) 100%

% 75% c o if 50% o

2 25%

0%

+

Exposure Duration (minutes)

Figure 4.8: E. coli Survivability When Exposed to 50°C

10

The data points in Figure 4.8 conespond to exposure durations of 0, 1,3, 5, and

10 minutes. Notice that there is still about a 25% survivability after 10 minutes of

exposure. This tends to confirm the assumption that the heat generated by the PCD,

which has a maximum temperature of about 35''C, is not a significant contributor to the

destiiiction of the E. coli during its exposure to the PCD. Pursuing the heat tests fiirtiier,

it can be seen in Figure 4.9 that significant destmction begins to occur at a temperature of

44

Page 55: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

75°C. At 75°C there is less than a 25% survivability after only one minute of exposure.

The data points in Figure 4.9 conespond to 0 ,1 , 3, 5, and 10 minutes of exposure. After

five minutes of exposure, one can see that the E. coli is almost completely desfroyed.

The results shown below suggest that exposing E. coli to both heat and the PCD could

dramatically decrease the exposure time necessary to produce a complete kill.

E. coli Heat Only Test (75C) 100%

2 75% *i c o if 50%

c H « a.

26%

0%

\

\

\l 1

^ — 4——^ ^ T ~ ~ — * - 1 * 1

3 5 Exposure Duration (minutes)

10

Figure 4.9: E. coli Survivability When Exposed to 75°C

4.3.3 E. coh Exposed to Heat Then the PCD

Continuing with the train of thought begun at the end of the last section, a test

combining the killing mechanisms of the PCD and of increased temperature has been

conducted. The second wire anay stmcture is used for these tests and the spacing

between the wdres of the wire anay is 0.25 inches. The increased temperature is

considered to be an enhanced en-vironmental condition imposed before exposure to the

PCD. The E. coli is heated to 75^0 and exposed to this environment for the specified

duration and then the specimens are exposed to the same duration of the PCD. The heat

and the PCD have been applied consecutively and it is suggested that the effects of tiie

two killing mechanisms are cumulative. The energies given in Figure 4.10 represent the

energy applied by the pulsed corona discharge. The durations for which the spores were

exposed to heat is given in the parenthesis next to the conesponding data point. After five

minutes, there is about a 7% survival of the E. coli. Comparing this wdth the heat only

test, one can see that there is no great improvement over the individual exposures to tiie

PCD or to heat. While these resuhs may seem to be inconsistent wdth the previous heat

results, one must keep in mind that the data set from which Figure 4.10 is derived is

45

Page 56: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

limited, and would need to be expanded for a more accurate representation of the process.

However, since no great improvement has been experienced, the use of heat as an

environmental enhancement to the PCD has been abandoned.

E. coli Exposed to Heat (75C) Then PCD

1250 Energy (kJ/m*2)

Figure 4.10: E. coli Survivability When Exposed to Heat Then the PCD

4.3.4 E. coli Exposed to the PCD With Water Vapor

The second and final environmental enhancement applied with the PCD is

increased humidity. The humidity in the air frnmediately above the specimens is

increased by using a water vapor spray from a small, room-sized humidifier. The major

impact on the survivability of the E. coli from the water vapor in the pulsed corona

discharge is the generation of OH" radicals which come in contact wdth the specimens.

No attempt has been made to measure or vary the density of the OH" radicals. As with

the tests conducted with heat and the PCD, the second wire anay stioicture is used for

these tests with a wdre spacing of 0.125 inches.

The resuhs for this test show a complete kill, destroying an average of 9.67 x 10*

cells after five minutes of exposure time, which conesponds to an average energy of

1177 kJ/m^ Based on the limited amount of data collected for this type of exposure

environment, it appears that increasing the humidity in the air directiy above tiie

specimens is a more effective environmental enhancement technique for increasing the

destmction capability of the PCD than is heat. The main question conceming these

results is whether the E. coli cells were "washed" off tiie coupons by the water vapor

46

Page 57: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

during the exposure process. Tests were performed wdth only the water vapor in an effort

to clarify this question, but the results were inconclusive. Because E. coli is so easily

killed, it is unknown whether the percent survival recorded is a direct resuh of the water

vapor or whether the E. coli were killed by some other environmental factor. Further

investigation of this exposure method has been continued with the more robust Bacillus

spores. The survivability of the spores under these environmental conditions wdll be

presented in the next chapter.

47

Page 58: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER V

SPECIFIC APPLICATION TO THE

ANTHRAX PROBLEM

5.1 Motives and a Re-Evaluation

In its conception, this project was scheduled as a three year exploration of

electrically driven energy sources which are capable of rapid biological and chemical

decontamination of surfaces without destroying the object that is contaminated. Because

of a lack of funding renewal, this project has become both the first and the last year of the

proposed three year program. With tiie use of biological and chemical weapons

becoming an increasing threat to the nation and the world, it has been concluded by the

faculty members of the research group that the last few months of research should be

focused on a specific problem area in order to complete the research efforts wdth the most

immediately practical solution. The specific problem that seems most appropriate for the

research infrastmcture that has been developed by this project is the contamination of

mail by Anthrax.

The background research for a realizable mail decontamination system has been

conducted over the last three years by the faculty and students working on this project

and projects such as the arc jet decontamination system presented m Chapter I.

Experiments evaluating the decontamination of Bacillus spores, which are shnulants for

Anthrax, using a pulsed corona discharge have been performed over the last quarter of

the first year of research in order to obtain specific information about the possibility of

using a PCD to decontaminate the mail. As with the previous focus of the research, it is

desirable to find the most energy efficient exposure method, or combination of methods

that produces the most rapid destmction of the agent.

While the arc jet was very effective in killing Bacillus spores very quickly (on the

order of micro-seconds) and would not even bum paper when passed by the arc jet

plume, the energy efficiency of this system is something left to be desired. The exposure

methods that have been explored for the application of decontaminating mail include:

exposure to only the pulsed corona discharge, exposure to only heat, exposure to heat and

48

Page 59: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

then to the PCD, exposure to the PCD with water vapor, and exposure to the PCD wdth

water vapor while the coupons are covered by 20 lb. printer paper or placed inside

various types of envelopes. The different types of envelopes used to simulate the mail

include padded envelope mailers with shredded newspaper as the padding, a stiff

cardboard mailer, and a plastic mailer. The number of the sheets of paper placed above

and below the coupons is also varied from 1 sheet above and below the coupons to 3

sheets above and below the coupon. The realization of a PCD decontamination system

that could be used by the US Post Office may be the size of a 24 inch television or a

standard microwave and be powered by 220 VAC. The importance of the last few

months of this research project has been to define the energy level as a fimction of the

decontamination of Bacillus spores. Once the energy levels necessary for

decontamination are established, a suitable power conditioning unit could be chosen or

designed wdthin a reasonable amount of time and the application of the research could be

completed.

5.2 Bacillus Experimental Results

5.2.1 Bacillus Exposed to the Pulsed Corona Discharge

The first experiment conducted involves exposing the Bacillus spores to just the

pulsed corona discharge. The experimental setup of these tests is the same as the similar

tests performed on the E. coli and the chemical compounds. The first wdre anay structure

is used with a wire spacing of 0.25 inches and separated from the ground plane by 1 inch.

The coupons are placed mainly beneath the wdres of the wire anay. Tests were

performed using pulse repetition frequencies of 10 Hz, 100 Hz, and 200 Hz, and input

voltages of 450 volts and 600 volts, but due to initial problems with the spores, these tests

did not yield results. The tests that did yield results were performed wdth a pulse repetion

rate of 100 Hz and an input supply voltage of 600 volts.

One can see in the in Figure 5.1 that it took almost 1000 k j W to reduce the

number of spores to just under 50% of the confrol count and an energy of almost 2000

kJ/m^ is necessary to reduce the number of spores to about 13% of the conttol count. The

energies at the data points on the graph conespond to exposure durations of 0, 5, and 10

49

Page 60: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

minutes. Based on previous experiments performed on the E. coli, it is believed that

using environment enhancement techniques such as heat or increased humidity may be

very effective in increasing the rate of kill while decreasing the necessary energy.

100% -r-

o 75%

o 'd 50% o 0)

S 25% Q.

0%

Bacillus Exposed to the PCD

990

Energy (kJ/m*2)

1980

Figure 5.1: Bacillus Survivability When Exposed to the PCD

5.2.2 Bacillus Exposed to Only Heat

Once again, the experimental setup established by the heat tests performed on

E. coli, is also used for the experiments wdth Bacillus spores. The spores have been

exposed to temperatiires of 50, 75, 100,125,130,135, and 150°C for durations of 1,3, 5,

and 10 minutes. As explained in the hot plate section of Chapter H, the temperatiire of

the hot plate is confrolled in such a way that it oscillates above and below the intended

exposure temperature. The results for these tests show that a temperature around 100°C

must be reached before the heat begins to kill any of the Bacillus spores at all. The most

significant results show a complete kill of the spores at around 130°C. This data supports

previous research performed by Holwitt, Kiel, et al. presented eariier in the paper that

sets a 130°C threshold for significant destmction of Bacillus spores.

Figures 5.2 and 5.3 show the destiiiction of tiie Bacillus spores at 100°C and

125°C, respectively, after exposure durations of 0,1, 3, 5, and 10 minutes. Notice tiiat at

100°C, the spore survival is significantly reduced after five minutes, but even after 10

minutes of exposure, there is not a complete kill. Referring now to Figure 5.3 showdng

the desti^ction of spores at 125°C at the same duration increments, one can see that after 50

Page 61: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

five minutes, there is only a trace of the Bacillus spores surviving. Comparing the

gradual, but evident killing of the Bacillus spores occuning at 125°C to the complete kill

of spores at 1 minute, 130''C exposures supports previous research suggesting that 130°C

is a threshold temperatiire for this strain of Bacillus spores. Tables wdth all temperatiire

and duration results can be found in Appendix A. While these results demonstrate a

complete kill of the Bacillus spores, maintaining such a high temperature on a large open

surface for even tiie shortest exposure duration is not practical and is destioictive to tiie

contaminated surface. Because of this destiaictiveness, the concept of using heat as a

killing mechanism is not a desirable solution to the problem at hand.

Bacillus Heat Only Test (100C) 100%

1 3 5 Exposure Duration (minutes)

Figure 5.2: Bacillus Survivability When Exposed to 100°C

10

Bacillus Heat Only Test (125C)

c o o

c a> u

a.

100%

0 1 3 5

Exposure Duration (minutes)

Figure 5.3: Bacillus Survivability When Exposed to 125°C

51

Page 62: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

5.2.3 Bacillus Exposed to Heat Then the PCD

While the sole use of heat has been eluninated as a feasible option for

decontamination, the possibility of using a lower temperature to pre-treat the spores

before exposure to the pulse corona discharge is an idea worth evaluating. The goal in

such a test is to attempt to use about the same amount of energy as that used in other

tests, but apply it using the two methods being discussed. These tests have been

conducted in the same way that the E. coli tests were perfonned. The second wire anay

stiiicture is used for these tests wdth a wire spacing of 0.25 inches and a separation

between the wire anay and the ground plane of 1 inch.

If one compares the energies of exposure for this method with the exposure

methods using only heat or the PCD, one can see that combining the two individual

energy sources provides no improvement over either method. The energies in Figure 5.4

conespond to a heat exposure of 0, 5, and 10 minutes followed by an exposure to the

PCD for 0, 5, and 10 minutes, respectively. The energies given in the graph represent the

energy applied by the pulsed corona discharge. The durations for which the spores were

exposed to heat is given in the parenthesis next to the conesponding data point. Given

that the test usmg only heat is more efficient than this combined method of exposure; h is

safe to say that this is not a more efficient method for decontamination.

Bacillus Exposed to Heat (75C} Then PCD 100%

is 75% s o U -- 50% S 0)

Q.

25%

0%

(0 min!r"'" -.v, ^^^

^^"---.v^ (5 min.)

" ^ " • - f - 1 ^ ^'""^^---^^..^.^(lO min.)

1 1 ^"^"^^^ 1

990 Energy (lcJ/m*2)

1980

Figure 5.4: Bacillus Survivability When Exposed to Heat Then the PCD

5.2.4 Bacillus Exposed to the PCD With Water Vapor

Increasing the percent kill of Bacillus using an environmental enhancement such

as water vapor is a very atfractive possibility. This is because it takes very little energy to

52

Page 63: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

generate water vapor using very common, readily available equipment. A common

household humidifier has been used to fill the volume between the wdre anay and the

ground plane wdth water vapor for these experiments, just as was done with the E. coli

experiments. One may remember that tiie increased water vapor in the air is believed to

cause an increased number of hydroxide radicals to be produced, which is thought to be

the actual mechanism that improves the killing efficiency of the PCD exposure. The

second wdre anay stmcture is used for these tests and the wire spacing has been reduced

from 0.25 inches to 0.125 inches in an effort in increase the uniformity of the discharge.

The water vapor is directed toward the wire anay using a snout. A continuous

flow of water vapor is applied to the volume sunounding the Bacillus coupons during the

entire duration of the PCD exposure. The energies at the data points in Figure 5.5

conespond to exposure durations of 0, 1,3,4, 5, and 10 minutes. One wdll notice that

there is almost a complete kill at an energy of 1258 kJ/m . Comparing this wdth the other

exposure methods, one wdll notice that this is a significant improvement over just

exposing the Bacillus spores to the pulsed corona discharge and is comparable wdth the

heat test performed at 125°C (Figure 5.3). While it is comparable in energy to the heat

test, using the PCD with water vapor is a much more practical solution and is much less

likely to harm the surface of the object bemg decontaminated. One may argue that using

the heat test wdth 130°C requires only 266 kJ/m^ of energy (1 minute) for a complete kill,

but the fact remains that heating a surface to such high temperatures for the necessary

duration will most likely be destmctive to the surface as well as to the agent.

Bacillus PCD and Water Vapor Exposure

260 750 990

Energy (kJ/m*2)

1260 2540

Figure 5.5: Bacillus Survivability When Exposed to the PCD wdth Water Vapor

53

Page 64: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

5.2.5 Bacillus Exposed to the PCD With Water Vapor when Covered

In continuing with the possibility that the pulsed corona discharge when

combined with water vapor could be a feasible method to decontaminate the mail,

experiments have been performed testing the effect that paper, plastic, and cardboard

barriers have on the survivability of the Bacillus spores. It is reasonable to believe that

there is a limitation to the effectiveness of the PCD through thick objects such as

cardboard boxes. The main purpose of these tests is to determine what common

materials and in what thickness of these materials does the PCD become ineffective. The

materials that have been chosen to be used include: up to 3 sheets of copy paper placed

above and below the coupons, a US Post Office padded mailer with shredded newspaper

as the padding, a cardboard document mailer, and a plastic mailer. When possible, the

coupons are slid inside the envelope being tested to most accurately simulate conditions

that would be encountered with mail. The same equipment setup has been used for this

test that was used for the previous PCD with water vapor tests.

Figure 5.6 shows the results after exposure to the PCD wdth the test specimens

sunounded by the US Post Office padded mailing envelope. Because the envelope would

not fit under the wire anay, large strips of the envelope were cut and placed above and

below the coupons plated wdth Bacillus spores. The envelope was cut so as to cover the

entire surface of the ground plane and positioned in a manner that prevented the coupons

from direct access to the water vapor. This was done in order to more closely simulate

the situation of Anthrax inside such an envelope. The energies at the data points in

Figure 5.6 conespond to exposure durations of 0,3,4, and 5 minutes. Comparing these

resuhs wdth those for the PCD with water vapor exposure without any barriers (Figure

5.5), it is clear that the presence of the envelope increases the survivability of the Bacillus

spores. The results for the plastic mailer, provided in Appendix A, tend to suggest that

the more effectively the banier prevents the water vapor from coming in close contact

with the coupons, the greater tiie survivability of the coupons. The plastic mailer is much

thinner than the padded mailer; however, tiie survivability of the coupons in the plastic

mailer is higher tiian tiiat for the padded mailer. This supports the idea that tiie water

vapor is a key factor in the effectiveness of the pulsed corona discharge.

54

Page 65: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The inconsistencies with some of the results are attributed to the probability that

some of the barriers being used are wiping spores off the coupons when they lay on top

of the coupons or when the baniers are removed from on top of the coupons. At this

point, it is not completely clear how much of a factor this wiping action is playing in the

survivability of the spores. However, it is clear that as the number of sheets of paper is

increased or as a thicker envelope is used, the survivability of the spores is increased.

These baniers are hindering the effectiveness of the pulsed corona discharge.

Bacillus Exposed to PCD & Water Vapor & Covered (Cushion Mailer) 100%

940 1250

Energy (kJ/m*2)

1560

Figure 5.6: Bacillus Survivability When Placed Inside the Cushion Mailer

55

Page 66: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER VI

FUTURE RESEARCH

The objectives for the first year of work on this project have been met and great

strides toward the immediate application of this technology to the decontamination of

mail have been made. In addition to accumulating meaningful data, the infrastmcture for

further testing has been established. The lab space is now fully operational and is

ananged in such a manner that adapting to new pieces of equipment and new test setups

would be a sfraightforward task. The necessities such as available power and necessary

venting are now available and convenientiy accessed. While the goals for the first year

have been met, the continuance of the project would necessitate some basic

improvements.

The main improvements lay wdth the power conditioning unit (PCU) and the wdre

anay stmcture. It is desirable to be able to perform the tests in a continuous manner

wdthout the time gaps that are cunently allowed in order for the PCU to cool down. In

addition, it is believed that the duration of exposure could be reduced down to seconds or

micro-seconds by merely increasing the pulse repetition frequency into the hundreds of

kHz range. Shortening the duration to these time periods would demonstrate the

capabilities of the pulsed corona discharge on a more realistic time scale. This is an

assumption that cannot be tested with the cunent PCU.

The main limitation wdth the wire anay is the uniformity of the electric field that

it produces, which is directly related to the uniformity of the pulsed corona discharge.

Generating a more uniform pulsed corona discharge directly translates into more

consistent and repeatable results. Alternative architectures that have been considered

include the use of dielectric barriers between the wdre anay and the ground plane (a

barrier discharge) or completely abandoning the wire anay and moving to a pin anay. In

either case, the main point is to improve the uniformity of the pulsed corona discharge.

Anotiier source of inconsistency throughout the project has been the biological

specimens. As has already been explained, the methods for preparing, fransporting, and

evaluating the biological specimens must be improved. It is believed that a major source

56

Page 67: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

of variation wdth the biological resuhs lies with the transporting of test specimens

between laboratories. No matter how many precautions one takes to ensure safe fransport

of these specimens, the 6 mile round trip voyage of the biological specimens consistentiy

results in the reduction of cells or spores whether the sample had been exposed or not.

This is a situation that must be addressed for further tests.

With respect to further experiments, one of the main sources that has not yet been

applied is the electron beam. This source has been studied by many other research

groups and is actually cunently being used by the US Post Office to decontaminate mail.

However, it is important to perform these tests using the same basic setup and test

specimen that have been used so far in order to establish a point of comparison with the

pulsed corona discharge. Tests evaluating the effectiveness of the elecfron beam as the

sole source and the elecfron beam combined wdth the pulsed corona discharge are of

particular interest. The survivability of the chemical compounds when exposed to the

electron beam source would also be of particular interest. Because of the concentration

on Bacillus spores, there has not been much data accumulated for the chemical

compounds. This database needs to be expanded and experunents wdth an elecfron beam

would be one source by which to expand the chemical database.

Other techniques that would help one to better understand which killing

mechanisms are most dominant within the pulsed corona discharge include tests wdtii UV

light, tests that limit the production of hydroxide radicals, such as using a Nifrogen

blanket, and tests that vary the amount of ozone to which the specimens are exposed. It

would also be of interest to use a spectrometer to determine the specfral elements of tiie

pulsed corona discharge. This would help in deteraiining how much UV or IR light is

actually being generated by the PCD. It would be interesting to plate the specimens on

different materials and evaluate thefr survivability based on the material to which the

specimen is attached. While the type of material is limited by the ability to recover the

specimens after exposure, this could be a very important factor which detennines the

survivability of the specimen.

57

Page 68: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

CHAPTER VII

CONCLUSIONS

After multiple experiments testing the ability of the pulse corona discharge (PCD)

to destroy both biological and chemical agents, it has been verified that the PCD is an

effective destmction technique. As expected, the Bacillus spores were not as easily killed

by the PCD as the E. coli and the Parathion proved to be more robust than the Malathion.

The most promising combination of destmction mechanisms was found when combining

the pulsed corona discharge with water vapor. After exposing the Bacillus to just the

PCD, it was found that 1980 kjW of energy reduced the number of spores to around

13% of the confrol specimen count. When a constant flow of water vapor across the wdre

anay was added as an environmental enhancement, a 1258 kJ/m^ exposure yielded a 3%

survivability. Increasing the amount of water vapor in the air directly surroimding the

specimens proved to be an energy-efficient and non-destmctive additive that increased

the capability of the pulsed corona discharge to kill Bacillus spores and left no residue.

After further tests with the Bacillus spores, it has been verified that barriers such

as cushion mailers, cardboard document mailers, and plastic mailers inhibit the effective­

ness of the PCD. When placed inside a cushioned mailing envelope and exposed to 1560

kJ/m^ of energy from the pulsed corona discharge with the water vapor, the percent

survival increased to about 30%. Using the same exposure method, the survivability of

the spores is further increased when placed inside a plastic mailing envelope resulting in

a 70% survival of the spores exposed to an energy of about 1580 kJ/m .

Based on the results, it is believed that the PCD combined wdth water vapor is an

energy efficient, non-destmctive technique that could be used to successfully destroy

biological agents such as Anthrax that may contaminate the mail. Continued research is

necessary to verify that the duration of exposure could be significantly reduced by

increasing the repetition rate of the pulsed corona discharge. A possible decontaminating

machine utilizing the pulsed corona discharge could be realized as an appliance the size

of a large microwave powered by 220 VAC into which a bag of questionable mail could

be placed for exposure durations of a few minutes or less.

58

Page 69: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

REFERENCES

1. Schoenbach, K. H., R.P. Joshi, R.H. Stark, F.C. Dobbs, and S.J. Beebe, "Bacterial Decontamination of Liquids wdth Pulsed Electric Fields," IEEE Transactions on Dielectrics and Electrical Insulation, Vol. 7 No. 5, pp. 637-645, October 2000.

2. Laroussi, Mounir, "Sterilization of Contaminated Matter with an Atmospheric Pressure Plasma," IEEE Transactions on Plasma Science, Vol. 24, No. 3, pp. 1188-1191, June 1996.

3. Pai, S.T. and Qi Zhang, Introduction to High Power Pulse Technology, River Edge, New Jersey: Worid Scientific, pp. 50-58, 1995.

4. Smtth, Phillip H., "Optimum Coax Diameter," Electronics for Communication Engineers, Ed. John Markus and Vin Zeluff New York: McGraw Hill, pp. 548-551, 1952.

5. Holwitt, Eric, Johnathan L. Kiel, John L. Alls, Pedro J. Morales, and Homer Gifford, "Thermal Sensitivity of biowarfare simulants," Proceedings ofSPIE Reprint, vol. 4036, pp. 31-39, 2000.

th

6. Nilsson, James W. and Susan A. Riedel, Electric Circuits, 5 Edition, New York: Addison-Wesley, p. 372, 1996.

7. Cheng, David K., Field and Wave Electromagnetics, 2°'' Edition, New York: Addison-Wesley, p. 270, 1989.

8. Electrophysics Institute, Ural Branch, Russian Academy of Sciences, Nanosecond Desk-Top SOS-Based Generator Operating Manual, Ekaterinburg, 1997.

59

Page 70: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

APPENDIX A

RESULT TABLES FOR ALL TESTS

60

Page 71: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Chemical Test Results

Table A. 1: Malathion Exposed to the Pulsed Corona Discharge

Malathion Initial Concenfration:

Duration 1 min. 2 min. 3 min. 5 min.

Total Energy

(kJ/m^) 470 850

1340 2230

Std. Dev. (Energy)

0 0

63 104

82.5 mg (250 nmole)

% Survival 83.22% 50.33% 30.43% 14.64%

Std. Dev. (% Survival)

9.14% 6.21% 7.05% 5.73%

% DDTP 2.63% 2.60% 4.90% 7.18%

Std. Dev. (% DDTP)

1.86% 1.84% 0.92% 1.69%

Table A.2: Parathion Exposed to the Pulsed Corona Discharge

Parathion Initial Concentration:

Duration 3 min. 5 min. 10 min. 15 min.

Total Energy

(kJ/m^) 1300 1990 3980 5640

Std. Dev. (Energy)

0 130 270

0

72.75 mg (250 nmole)

% Survival 46.38% 26.91%

8.64% 0.00%

Std. Dev. (% Survival)

5.16% 13.66% 15.37% 0.00%

% DDTP 55.80% 71.57% 60.44% 17.14%

Std. Dev. (% DDTP)

4.51% 13.66% 19.88% 4.67%

61

Page 72: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

E. coli Test Results

Table A.3: E. coli Exposed to the Pulsed Corona Discharge

E. coli Initial Concenfration:

Duration 3 min. 5 min. 10 min.

Total Energy

(kJ/m^) 1230 1880 3580

Std. Dev. (Energy)

8 270 600

10^ cells

% Survival 52.78% 15.78% 0.39%

Std. Dev. (% Survival)

31.29% 14.35% 0.47%

Table A.4: E. coli Exposed to Only Heat

E. coli 1

Temp.

CC) 50

75

100

nitial Concentiation:

Duration

1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min.

Total Energy

(kJ/m^)

70 200 330 660 130 390 640

1400 180 540 900

1810

Std. Dev. (Energy)

0 0 0 0

20 10 2 2 0 0 0 0

10^ cells

%

Survival

81.16% 55.43% 42.12% 23.33% 21.79% 10.15% 2.23% 2.06% 0.33% 0.08% 0.01% 0.00%

Std. Dev. (% Survival)

18.42% 17.49% 29.64%

5.77% 13.90% 5.38% 2.68% 0.82% 0.48% 0.10% 0.01% 0.00%

Cell Count Reduction

1.30E+05 1.60E+05 1.28E+05 7.67E+04 5.12E+06 5.70E+06 5.66E+06 6.23E+06 7.47E+06 7.50E+06 7.51E+06 7.51E+06

Std. Dev. (Cell Count Reduction)

1.27E+05 1.28E+05 7.19E+04 5.77E+03 5.33E+06 6.04E+06 6.00E+06 6.72E+06 8.15E+06 8.19E+06 8.30E+06 8.20E+06

Table A.5: E. coli Exposed to Heat Then the Pulsed Corona Discharge

E. coli Initial Concenfration:

Temp.

CC) 75 75

Duration 5 min. 10 min.

Total Energy

(kJ/m^) 1870 3040

Std. Dev. (Energy)

0 0

4.67 X 10 cells

%

Survival 7.14% 0.25%

Std. Dev. (% Survival)

4.95% 0.03%

Cell Count Reduction

4.34E+06 4.66E+06

Std. Dev. (Cell Count Reduction)

2.33E+05 1.53E+03

62

Page 73: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

E. coli Test Results

Table A.6: E, coli Exposed to the Pulsed Corona Discharge wdth Water Vapor

E. coli Initial Concenfration:

Duration 5 min. 10 min.

Total Energy (kJ/m^)

1180 2360

Std. Dev. (Energy)

0 0

% Survival 0.00% 0.00%

9.67 X 10 cells

Std. Dev. (% Survival)

0.00% 0.00%

Cell Count Reduction

9.67E+06 9.67E+06

Std. Dev. (Cell Count Reduction)

0 0

63

Page 74: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Bacillus Test Results

Table A.7: Bacillus Exposed to the Pulsed Corona Discharge

Bacillus Initial Concenfration: 2.00E+4 spores

Duration: 5 min. 10 min.

Total Energy

(kJlrn) 990

1980

Std. Dev. (Energy)

0 0

% Survival Bacillus

45.45% 12.73%

Std. Dev. (% Survival)

0.00% 11.53%

Spore Count Reduction:

1.20E+04 1.92E+04

Std. Dev. Spore Count Reduction

0.00 2535.74

Table A.8: Bacillus Exposed to Heat Then the Pulsed Corona Discharge

Initial Bacillus Concentration:

Temp.

CC) 75 75

Duration 5 min. 10 min.

Total Energy

(kJ/m^) 2020 4040

Std. Dev. (Energy)

0 0

2.20E+04

%

Survival 37.42%

0.00%

Std. Dev. (% Survival)

10.85% 0.00%

Spore Count Reduction

1.38E+04 2.20E+04

Std. Dev. Spore Cnt. Reduction

2.39E+03 O.OOE+00

Table A. 9: Bacillus Exposed to the Pulsed Corona Discharge wdth Water Vapor

Initial Bacillus Concenfration:

Duration 1 min. 3 min. 4 min. 5 min. 10 min.

Total Energy

(kjW) 260 750 990

1260 2540

Std. Dev. (Energy)

1 15 0

25 20

% Survival 39.00% 12.55% 14.14% 2.09% 0.00%

Average 1.58E+04

Std. Dev. (% Survival)

15.44% 14.71% 14.79% 2.53% 0.00%

Std. Dev. 2.62E+03

Spore Count Reduction

1.06E+04 1.30E+04 l.lOE+04 1.48E+04 1.73E+04

Std. Dev. Spore Cnt. Reduction

2668 2224 1893 2634

0

64

Page 75: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Bacillus Test Resuhs

Table A. 10: Bacillus Exposed to Heat Only

Bacillus Heat Only Tests (hiifial Concentration on order of 5E+4 spores)-

Temp.

CC) 75 C

100 C

125 C

130 C

135 C

150 C

Duration

1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min. 1 min. 3 min. 5 min. 10 min.

Total Energy

(kJ/m^)

140 400 670

1350 150 440 730

1460 250 720

1200 2400

270 800

1330 2660

210 620

1400 2080

270 800

1330 2660

Std. Dev. (Energy)

0 0 0 0 0 5

10 20 10 5

10 15 0 0 0 0

40 120 600 500

10 30 60

110

%

Survival

73.91% 91.30% 86.96%

104.35% 100.00% 68.70% 39.22% 34.87% 23.67%

7.34% 0.50% 0.44% 0.00% 0.00% 0.00% 0.00% 0.00%) 0.00% 0.00% 0.00% 0.00% 0.00% 0.00% 0.00%

Std. Dev. (%

Survival) 45.76% 15.06% 4.35% 3.83% 0.00%

22.24% 26.31%

0.00% 18.26%) 6.89% 0.55% 0.73%) 0.00% 0.00% 0.00%) 0.00% 0.00% 0.00% 0.00% 0.00% 0.00% 0.00% 0.00% 0.00%)

Spore Reduction

2.95E+05 1.20E+05 9.00E+04

-3.00E+04 O.OOE+00 2.26E+04 4.94E+04 5.04E+04 2.83E+04 3.59E+04 4.07E+04 4.06E+04 1.30E+04 1.30E+04 1.30E+04 1.30E+04 1.88E+04 1.88E+04 1.88E+04 1.88E+04 2.39E+04 2.39E+04 2.39E+04 2.39E+04

Std. Dev. Spore Reduction

3.46E+05 O.OOE+00 3.00E+04 2.65E+04 O.OOE+00 2.87E+04 4.01E+04 3.62E+04 2.95E+04 3.93E+04 4.49E+04 4.47E+04 O.OOE+00 O.OOE+00

• O.OOE+00 O.OOE+00 6.40E+03 O.OOE+00 O.OOE+00 O.OOE+00 1.20E+03 O.OOE+00 O.OOE+00 O.OOE+00

65

Page 76: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Bacillus Test Results

Table A. 11: Bacillus Exposed to the PCD with Water Vapor when Covered

Bacillus Efield & Paper & Humidity Tests: Average Bacillus Initial Concentration;

Duration

1 min. 3 min. 3 min. 3 min. 3 min. 3.5 min. 3.5 min.

4 min. 4 min. 4 min. 4 min.

4.5 min. 4.5 min. Smin. 5 min. 5 min. Smin. 5 min.

5.5 min. 5.5 min. 10 min. 10 min. 3 min. 4 min. Smin.

3 min. 4 min. S min. 5 min. 10 min.

Sheets Above

1 1 2 3 3 1 2 1 2 3 3

1 2 1 1 2 3 3

1 2 1

1 1* 1* 1* ] * *

J** 2** J*** } * * *

Sheets Below

1 I 2 2 3 1 2

1 2 2 3

1 2 0 1 2 2 3

1 2 0

1

1* 1* 1* ] * *

1** J** J*** J***

Total Energy

(kJ/m^)

250 800 800 800 800 920 920

1070 1070 1070 1070 1190 1190

13S0 13S0 1350 1350 1350 1450 14S0 2720 2720 940

1250 1560

930 1240 1550 1580 3180

Std. Dev. (Energy)

0 95 95 95 95 0 0

135 13S 135 135

0 0

180 180 180 180 180

0 0

430 430

5 S S

40 50 65

20 40

Average: 1.33E+04

% Survival

11.66% 22.62% 19.27% 15.09% 28.86% 72.56% 39,07%

3.83% 10.71% 11.47% 9.73%

42.31% 33.01% 2.66% 0.50% 4.04% 6.69% 0.00%

66.25% 25.10%

1.03% 0.04%

51.94% 35.69% 30.67% 21.01% 11.92% 18.51% 68.33% 42.15%

Std. Dev. 5.46E+03

Std. Dev. (% Survival)

6.43% 40.94% 18.04% 3.34%

20.06% 0.00% 0.00% 4.03%

14.09% 3.83%

11.05% 8.97%

19.34% 0.58% 1.08% 6.11% 3.38% 0.00%

48.32% 43.48%

1.79% 0.06% 6.63% 6.70% 8.20%

3.23% 2.62% 8.19% 7.92% 5.37%

Spore Count Reduction

1.82E+04 1.23E+04 8.39E+03 1.09E+04 7.89E+03 1.97E+03 4.37E+03 8.17E+03 9.23E+03 1.13E+04 l.OOE+04 4.14E+03 4.80E+03 1.26E+04 1.45E+04 9.80E+03 1.19E+04 l.llE+04 2.42E+03 5.37E+03 1.28E+04 2.06E+04 4.72E+03 6.32E+03 6.81E+03 1.64E+04 1.83E+04 1.70E+04 6.S8E+03 1.20E+04

Std. Dev. (Spore Count Red)

1.32E+03 8.45E+03 2.90E+03 4.27E+02 2.22E+03 O.OOE+00 O.OOE+00 1.43E+03 2.72E+03 4.90E+02 1.23E+03 6.43E+02 1.39E+03 5.58E+01 6.42E+03 2.22E+03 4.33E+02 O.OOE+00 3.46E+03 3.12E+03 2.31E+02 1.29E+01 6.52E+02 6.57E+02 8.05E+02 6.71E+02 S.45E+02 1.70E+03 1.64E+03 1.12E+03

: US Post Office Cushion Mailer (Shredded Newspaper as Cushion) ** : Airborne Express Cardboard Mailer *** : Airbome Express Plastic Mailer

66

Page 77: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

APPENDIX B

SLIDE LAYOUTS FOR

SELECTED TESTS

67

Page 78: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Ground Plane

1

4

7

10

(From Pulser)

2

5

8

11

3

6

9

12

Wire Array

(Overhead View of Test Slide Location)

Figure B. 1: Layout for Bacillus K, L, N, O Tests

Ground Plane

Wire Array

9

6

3

(F rom Pul<

8

5

2

>er)

7

4

1

(Overhead View of Test Slide Location)

Figure B.2: Layout for Bacillus X, Y, Z, AA, BB, CC, DD, FF, KK, LL, NN, 0 0 and Parathion M Tests

68

Page 79: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Ground Plane

Wire Array

1

4

(From Pulser)

2

5

3

6

(Overhead View of Test Slide Location)

Figure B.3: Layout for Bacillus J, M, P, S and E. coli Q Tests

Ground Plane

Wire Array

4

5

6

(F rom Puis >er)

1

2

3

(Overhead View of Test Slide Location)

Figure B.4; Layout for Bacillus V and W Tests

69

Page 80: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Ground Plane

ire Array

13

14

15

16

17

18

(From Pulser)

7

8

9

10

11

12

1

2

3

4

5

6

(Overhead View of Test Slide Location)

Figure B.S: Layout for Bacillus T and U Tests

Ground Plane

Wire Array

1

7

4

(F rom Puis

6

3

1

ser)

5

2

(Overhead View of Test Slide Location)

Figure B.6: Layout for Bacillus EE and GG Tests

70

Page 81: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Ground Plane

\

Wire Array

, 6

4

2

(From Pulser)

5

3

1

(Overliead View of Test Slide Location)

Figure B.7: Layout for Bacillus HH Test

Ground Plane

Wire Array

, 1,4

1.4

(F rom Pul!

2,5

2,5

ser)

3.6

3,6

Malathion H

Parathion K

(Overhead View of Test Slide Location)

Figure B.S: Layout for Parathion K, Malathion H Tests

71

Page 82: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

Ground Plane

Wire Array

,

6,9

3, 12

(F rom Pul;

5,8

2, 11

ser)

4,7

1, 10

(Overhead View of Test Slide Location)

Figure B.9: Layout for Malathion I Test

(Overhead View of Test Slide Location)

Figure B. 10: Layout for Parathion L Test

(Empty)

Ground Plane

Wire Array

12

9

6

3

(F rom Pul!

11

8

5

2

5er)

10

7

4

1

20 min.

15 min.

10 min.

5 min.

72

Page 83: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

APPENDIX C

POWER CONDITIONING UNIT (PCU)

73

Page 84: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

In order to generate the PCD, a power source capable of delivering high voltage,

nanosecond pulses is necessary. A Russian built nanosecond desktop semiconductor

opening switch (SOS) based generator is used to deliver the necessary high vohage

pulses. The nanosecond desktop SOS-based generator is referred to as the power

conditioning unit (PCU) throughout this paper. The PCU is capable of delivering voltage

pulses with amplitude of up to 150 kV and 400 Amps at a pulse repetition frequency in

burst mode of up to 100 Hz. The maximum recommended operating time in burst mode

is one minute, but may also be operated in a continuous pulsing mode if the pulse

repetition frequency is reduced to 10 Hz. The typical pulse width of the delivered

waveform varies between 20 and 30 ns at Full Width Half Maximum (FWHM). The

PCU output pulses are controlled by input drive-pulses with voltage amplitude of 10-15

volts and a pulse length of 5-10 ^s at a 50 Q load. The main supply voltage to the PCU

is 600 Volts DC and the mains power consumption of the PCU should not exceed 1 kVA

at the maximum pulse repetition frequency of 100 Hz. The PCU is capable of delivering

pulses at the stated voltage levels and a pulse repetition frequency of 100 Hz for up to

two minutes, but one risks overheating the pulser under such operating conditions. A

diagram showing how the PCU is connected to the rest of the data acquisition and power

supply systems can be viewed in Figure 2.5.

The PCU consists of three stages of power conditioning units. The first unit is a

thyristor converter, referred to as the A1 unit, which converts the 600 volt DC input into a

microsecond (^s) long pulse using a thyristor switch. A schematic diagram of the Al

unit may be seen in Figure C. 1. The A2 unit is the high voltage nanosecond (ns) pulse

shaper, which compresses the pulse from unit Al to a 1 |us pulse with pulse amplitude of

50-60 kV using a magnetic switch and pulse transformers. The semiconductor opening

switch at the output of unit A2 converts the energy of the microsecond pulse to the

nanosecond range of tune simultaneously increasing the voltage and current by about 3

times. The load is connected to the output of unit A2. Figure C.2 provides a schematic

diagram of the A2 unh. The third unit (A3) of the PCU is the confrol unit and the circuit

diagram for this unit may be seen in Figure C.3. Umt A3 amplifies the input dnve pulses

74

Page 85: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

from the external generator. Transformed pulses of the required amplitude and length are

sent to control the electrode of the thyristor in the Al unit [8].

The PCU operating specifications limit the ability to deliver pulses for continuous

exposure durations of more than 1 minute at a pulse repetition frequency of 100 Hz. The

components within the PCU that limit the output specifications due to overheating are the

charging resistors for the primary storage capacitors in the Al unit and the unknown

limitations of the semiconductor opening switch in the A2 unit. In order to limit power

dissipation by the charging resistors, it would not be difBcuh to replace these ceramic,

high power resistors with resistors of a higher power rating and higher resistance.

Replacing the resistors would lower the current flowing through the resistors thereby

lowering the amount of heating at the resistors assuming a constant input voltage. The

next major component of concern is the semiconductor opening switch which switches

the power in the final stage delivering the nanosecond high voltage pulse at the output.

More detailed specifications considering the switch's power rating and recovery times

would be preferred before operating the pulser for longer continuous durations or at

higher frequencies than 100 Hz. Because of limited information about the specifications

of individual components within the pulser, the replacement of the charging resistors has

not been pursued. Due to the limitations of the PCU, the experiments have been

restricted to applying the pulsed corona discharge exposures in one minute increments

wdth a five minute cooling period in between each exposure.

The output voltage and current through the SOS interrupter are measured via

BNC connectors labeled "Us" and "Is" that are located at the front face of the pulser. The

front face of the pulser is also the face from which the high voltage nanosecond pulse is

delivered to the desired load. The connector Us is intended to measure the output voltage

delivered to the load and the connector labeled Is is intended to measure the current

through the interrupter. The voltage and current waveforms are monitored using a

Hewlett-Packard Infinium Oscilloscope capable of measuring frequencies up to 500 MHz

at a 2 GSa/s resolution. The peaks of the voltage and current waveforms will not be

aligned when monitored with the oscilloscope because the voltage and current are not

measured at the same points in the circuit.

75

Page 86: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

The connector Is measures the current passing through the SOS interrupters by

measuring the voltage across the parallel connection of resistors R6...R15 and the 50 Q

termination at the oscilloscope. The equivalent resistance of the parallel elements is

0.505 Q. This can be rounded to 0.51 Q. The current is calculated using Ohm's law. The

resistance is a constant value of 0.51 Q, neglecting any changes due to heating and the

oscilloscope records the voltage drop across the 0.51 Q resistance. The current may be

viewed on the oscilloscope by appropriately setting the external gain and the attenuation

in the channel properties for channel 1 of the oscilloscope. A more detailed explanation

for the setup of the oscilloscope is provided in the Data Acquisition section of Chapter II.

The current delivered to the load is approximately the same as the current measured by

the current monitor Is. During the reverse pumping of the SOS switches, the current

passing through the switches charges LI, an inductive storage component. The load

resistance is estimated to be much less than the resistance presented by the voltage

divider. Because of the low load resistance, virtually all of the energy stored in the

inductive storage component is transferred to the load. Maintaining a very low load

resistance allows one to assume that the current delivered to the load is approximately the

same as the current passing through the SOS switches a few nanoseconds beforehand as

measured by the current monitor.

The voltage (Us) delivered to the load is measured using a resistive voltage

divider. The series connection of resistors R16 through R31, summing to 75.2 kQ, make

up the high voltage leg of the divider and the parallel combination of R32, R33, and the

50 Q termination at the oscilloscope, equaling 4.545 Q, make up the low voltage leg of

the divider. The voltage across the low voltage leg of the divider is a known fraction of

the voltage delivered to the load. A mathematical explanation is provided in the Data

Acquisition section of Chapter II to explain the setup of the oscilloscope conceming the

proper gain settings for the signals used to measure the voltage and current delivered to

the load. Figure C.2 is a schematic diagram for part of the A2 unit showing the locations

in the circuit where the current and voltage are measured.

76

Page 87: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

fll

+ o-

'POUER"

P.7M :3 . . ,C !2

- ^ ^ J L ^ 2] ^ fT8 H ijl

'5fl"

s <r''^^fc 0 to ! R2

R3,R4

Tl

"=600 U" ^ ' ^ tu

UDl RS

t o i n3

* 0 i

o to 2 fl2

Figure C. 1: Circuit Diagram of the Thyristor Charging Unit Al

ft2 C 1 3 . . . C 2 1

111 d

L ^ u^ 4- L.

n C ! . . . c i ; R3 1

1

2

— I I . '

il i l^o-vv-

"\k

p " T-XI

Ul- -u i -

HO) H(J)

R5...fll5

9i

U)" kD

Jfi lS. . .Rjl

I 4 i 1 ' •

R 3 2

' ^ '

fjBM....

Figure C.2: Circuit Diagram of the High Voltage Nanosecond Pulse Shaper A2

77

Page 88: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

fl3

R I . . . R 3 R9. . .R11 to 1 fll 1 o—r—I 1—, 1 f ^

Ct

t o 2 fll 2 <y-^ I-

Cl C3

T2

3

: : osi

R7 ^1 R8 R12

"TRIG-T i f xl

UD2

x_3

. 1 R13 • RM

CZh'

-N- UD3

3

Figure C.3: Circuit Diagram of the Confrol Unit A3

-o to 3 fll Lo to 4 fll

78

Page 89: LARGE SURFACE BIOLOGICAL/CHEMICAL DECONTAMINATION …

PERMISSION TO COPY

In presentmg this thesis m partial frilfiUment of the requuements for a master's

degree at Texas Tech University or Texas Tech University Health Sciences Center, I

agree that the Library and my major department shall make it freely available for

research purposes. Permission to copy this thesis for scholariy purposes may be

granted by the Dfrector of the Library or my major professor. It is understood that

any copying or publication of this thesis for financial gain shall not be allowed

without my further written permission and friat any user may be liable for copyright

infringement.