  • Joja Surgicals Pvt. [email protected]

  • This presentation is issued in scientific interest to promote better marking techniques for Toric IOL implantation, LRIs and refractive surgeries

    The tool and software described in this presentation are patented / copyrighted

  • This is not an advertisement

  • Why is Marking Important?For every degree of rotation 3.3% of the effect is lostWith 30 degree rotation all effect of the Toric lens is lostBeyond that you start adding to the astigmatism

  • Why is Marking Important?Higher the Toric correction greater the chances of errorse.g. For a 5 degree misalignmentIn a 6 D Toric error is 1 DIn a 1.5 D Toric error is 0.25 D

  • So lets get the marking spot on !Two issues in markingMaking sure our 0 180 meridian is true horizontalMaking sure that the marks bisect the cornea through the optical centre

  • Current MarkersHand held markersWith spirit level to ensure horizontality

  • Current MarkersDo not ensure that the marks pass through the centre of the cornea

  • Are you bisecting the cornea?

  • Current MarkersYou also need a second intra-op mark if the placement axis is not 0 180

  • Current MarkersCome in only one sizeBut the corneas are not of the same size or shape!

  • IssuesStill does not solve the problem of bisecting the cornea through the optical centre

  • A New ApproachA slit lamp based systemUses gravity to define horizontalUses Purkinje images to ensure centring on the cornea

  • 3 Simple StepsSet the slit lamp beam to the desired placement axisAlign the beam to pass through the corneal centre (described later)Mark the cornea

  • Step 1 Setting the AngleThe slit beam can be set to the axis of placementNo need for reference markingClamp device to chin rest of slit lamp

  • Setting the Angle

  • Step 2Unclamp the calibrator devicePlace the patients head normally on the slit lamp

  • Centring on the CorneaPatient looks at the light with the other eye closedMove the slit beam across the cornea until you get a bright Purkinje image

  • Centring on the CorneaImages bordered in red show the correct positioning of the slit beam

  • Centring on the Cornea

  • How do we know it is accurate?

  • Planiware 2.0This is a IOL planimetry software designed for measuringCentration of IOL, rhexis, pupilMeasurement of rhexis / pupil areas and diametersAngular measurements of Toric IOLs

    Planiware can be downloaded from www.suryodaya.in/planiware

  • Verifying your markings

  • *