Irradiator Technology H-315 Session 07 Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical Timing Circuits Radiation Detection Mechanical or Physical

Upload: others

Post on 20-Mar-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Irradiator Technology H-315 Session 07

Irradiator Technology H-315 Session 07

1

Session 07Safety Systems

Session 07Safety Systems

Timing CircuitsRadiation Detection

Mechanical or Physical

Timing CircuitsRadiation Detection

Mechanical or Physical

Page 2: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Safety FeaturesSafety Features

• Fault: (in Red) will stop irradiator immediately

• Caution (in yellow) will not stop

• Fault: (in Red) will stop irradiator immediately

• Caution (in yellow) will not stop

2

( y ) pirradiator immediately but will not restart

( y ) pirradiator immediately but will not restart

Page 3: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Timing circuits

Source rack timerSource pass timer

overall and cylinders timers

Overdose timer Console & Touchtime or Nortrack

3

Console & Touchtime or Nortrack

Product movement timer

Barrier doorsStart-up timer & Conveyor start movingCylinder timer

Ozone delay timer

Uninterruptible power supplyResearch Loop Door timer

Page 4: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Radiation detection

Ludlum Area Monitor L396Or DL-2 monitors (deionizer)

Ludlum Area Monitor L396

4

Ludlum Area Monitor L396or DL-2 monitors (product exit)

L118 or L385 (coming soon)at access door

Page 5: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Mechanical devices

High temperature sensorsSmoke detector Exhaust fan (flow switch)EarthquakeLow air

5

Stop

TurntablesIntruder detectionFloat switchSource valveAir valve (chain or barrier at door)Pressure switch at source hoist

Computer interlock

Page 6: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Mechanical devices

Product inlet + outlet doorsElevators, hydraulic, electric pneumaticCylinders (pushing or pulling)

6

Carrier collision device

Warning:Warning:Be aware of always on devicesBe aware of always on devicesBefore fixing any jam all energyBefore fixing any jam all energymust be released on any mechanical devicesmust be released on any mechanical devices

Page 7: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Source rack10CFR36.31

Timer in PLC

Source hoist switchIrradiation chamber switch

What:

Where:

7

The source rack takes too long to reach the bottom or up position

To prevent damage to rack or equipment

When:

Why:

How: Change time on PLC to 10 sec.

Page 8: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Source pass

Timer - Overall timer,Cylinder timer

Any motion or group of motions in the

Within PLC

What:

Where:

When:

8

Any motion or group of motions in theIrradiation chamber exceeds theselected time

To avoid overdose and To prevent damage to the equipment

When:

Why:

How: Change timer on PLC to a short value

Page 9: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Overdose timer10CFR36.31

Timer

Mounted in the console

Th t Ti h t i iti t d

What:

Where:

Wh

9

The master Timer has not initiateda new cycle

To avoid overexposure of product

When:

Why:

How: Set to a value less than Master Timer

Page 10: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault:Product Movement Timer

10CFR36.31

Within PLC

No movement of product inside the

TimerWhat:

Where:

Wh

10

No movement of product inside the irradiation chamber for a period exceeding the allotted time.

To keep product from catching fire due to static irradiation.

When:

Why:

How: Change timer in PLC to a short value.

Page 11: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault:Barrier doors10CFR36.23a

Mechanical switches, PLC timer

Mounted at the product portals

O f th d t d h

What:

Where:

When:

11

One of the product doors has exceeded the short time limit in which it is allotted to reach its intended position (open or close).

To prevent entry of personnelTo prevent collision with a carrier

When:

Why:

How: Remove or bend switch

Warning: Always on device on some machine

Page 12: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Deionizer monitor10CFR36.33e

Ludlum Area Monitor or DL-2

Mounted on deionizer system

What:

Where:

When: Detects radiation levels above background

12

Prevent contamination or exposure of personnel

When:

Why:

How: Simulate radiation levels with check source

Detects radiation levels above background.The personnel entry door locksThe water circulation systems stops

Page 13: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault:Product Outlet Monitor10CFR36.29a

Ludlum Area monitor or DL-2

Mounted at product exit

What:

Where:

When: Detection of background radiation

13

Prevent spread of contaminationor exposure to personnel

When:

Why:

How: Simulate radiation levels with check source

Detection of background radiationLocks personnel door

Page 14: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: L-118 Monitor10CFR36.23a,c

L-118 or Ludlum 385 (now current)

Mounted at personnel entry doorProbe inside source passInterlocked with door (off when source up)

What:

Where:

14

Interlocked with door (off when source up)

Prevent exposure to personnel orspread of contamination

When:

Why:

How: Simulate radiation levels with check source + saturation testusing bypass button on the console.Personnel entry door locks. Alarm on

Detection of background radiationlocks personnel door,

Page 15: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault:High temperature10CFR36.27a

Heat sensors

Mounted on the ceiling between thesource rack and the exhaust fan

What:

Where:

15

To protect the source rack againstexcessive heat

When:

Why:

How: Use portable heat source to slowlyheat the sensors

Whenever one of the sensors is trippedat 160 F. This will also shut down the exhaust fan

Page 16: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Smoke detector10CFR36.27a

Smoke detector

Mounted in the penthouse or equipment room. (Intake inside the irradiation chamber)

What:

Where:

16

irradiation chamber)

The smoke detector is tripped.

To protect the source rack against possible excessive heat.

When:

Why:

How: Use smoke generator to feed the intake pipe with smoke.

Page 17: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Flow switch

Mounted in the ventilation stack, just before the filters.

What:

Where:

Fault: Exhaust fan

17

Whenever there is a loss of air flow.

To prevent ozone from getting into the storage area.

When:

Why:

How: Manually shut down the ventilation.

Page 18: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Earthquake10CFR36.39j

Seismic detector

Mounted on the concrete shield in the equipment room.

What:

Where:

18

Whenever the detector is triggered by any vibration.

To protect source rack and equipment.

When:

Why:

How: Use a rubber hammer and "gently" hit the detector supporting plate.

Page 19: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Carrier collision

Plates with mounted microswitches.

Located at both ends of the source rack

What:

Where:

19

A carrier or product hit the plate.

To protect the source rack from collision that could create a source jam situation

When:

Why:

How: Manually push the two plates to activate the microswitches.

Page 20: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Low air

Pressure switch

Mounted in the source hoist interlock valve box located in the personnel corridor

What:

Where:

20

corridor

Whenever there is low air pressure to the source hoist.

Cylinders cannot operate at normal speed if the air pressure is too low

When:

Why:

How: Shut off the air at the compressor

Page 21: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Stop10CFR36.23h-36.31c

Emergency push button on consoleMicroswitches at the plugEmergency stop cable (2)Tampering with or opening the personnel entry door

What:

21

personnel entry door

See above

Stop operation due to emergencyTo prevent exposure of personnel

When:

Why:

How: Push button on consoleBend switch at the inner plugPull emergency stop cables

Page 22: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Computer Interlock

Nortrack

What:

Where:

If Nortrack computer senses anysafety system trip or malfunction

22

Fault or Nortrack failure

Backup to console, Stop operation if Nortrack failsStop operation when off carrier timerreaches zero

When:

Why:

How: Any other fault will also give the“Computer interlock fault

Page 23: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: End of Batch

In the PLC

What:

Where:

Not a real fault. Stops the machineafter a batch load, run or unload

23

To stop machine, see above

When:

Why:

How: This is not a fault to be tested

See above

Page 24: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Turntable

Switch

Mounted on a turntable

Wh t t bl t t ti

What:

Where:

Wh

24

Whenever a turntable stops rotating

To prevent non-uniform irradiation of product

When:

Why:

How: Manually, jam the turntable so that the switch will not be actuated within the allotted time.

Page 25: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: High water

Float switch

Mounted at the pool

Wh th t l l i b

What:

Where:

Wh

25

Whenever the water level is above a preset level

To avoid overflow of the pool

When:

Why:

How: Manually, raise the float until the caution light comes on.

Page 26: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: Low water10CFR36.33d – 36.41

Float switch

Mounted at the pool

Whenever the water level is below a

What:

Where:

When:

26

Whenever the water level is below a preset level. The personnel entry door will be locked

To avoid loss of shielding water and to protect personnel from exposure to radiation

When:

Why:

How: Manually lower the float until the caution light comes on

Page 27: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: Interim not clear10CFR36.23

Keyswitch, interlock

Mounted in the interim area.

The interim access door has been

What:

Where:

When:

27

The interim access door has been opened and the keyswitch not actuated.

To prevent raising the source rack with someone still being in the interim area.

When:

Why:

How: Open the interim door and close without resetting the keyswitch.

Page 28: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: Entry delay

Timer to allow purging of ozone

Within PLC

Whenever the source rack has just

What:

Where:

When:

28

Whenever the source rack has just reached the safe position.

The personnel entry door is locked.

To protect personnel from ozone poisoning.

When:

Why:

How: Try to open the personnel entry door while the "entry delay" light is on.

Page 29: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: PLC Fault

In the PLC

The PLC failed an internal self

What:

Where:

The PLC

29

The PLC failed an internal self check or has sensed a problem in the hardware relays

Irradiator cannot operate without the PLC

When:

Why:

How: No need to test, self checking

Page 30: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: PLC Run

In the PLC

What:

Where:

The PLC

30

The PLC Light is on when the PLCfunctions properly

Visual indication the PLC functionsproperly (not on all machines)

When:

Why:

How: No need to test, light stays on

Page 31: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Source valve

Two pressure switches betweenThe two source hoist valves

In the penthouse

What:

Where:

31

If one of the two valves sticks inthe on position

To prevent damage to the rack

When:

Why:

How: Not tested, self checking

Page 32: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Building Power10CFR36.37

UPS (Uninterruptible power supply)

In the equipment room

What:

Where:

32

Whenever power to the UPS is removedfor more then 5 seconds

To supply power to the control consoleIf power goes off

When:

Why:

How: Turn power off for more then 5 seconds

N.B. If the power is loss for too long & the PLC battery is dead, you will loose the PLC program, (need to reload)

Page 33: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Backup Access Control10CFR36.23b

Photosensor arrays

Product entrance & exit and personnel door

What:

Where:

33

Sensor activated. Causes STOP FAULT alarms, lights

To prevent unauthorized entry with source not down.

When:

Why:

How: Activate sensors using a pole while source up or moving up (access door)

N.B. Pre 1992 Nordion Irradiators needed retrofit

Page 34: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: PLC Battery

The PLC board

On the PLC board

C ti

What:

Where:

When:

34

Board data hold

ContinuousWhen:

Why:

How: Self checking on some systems

Need to change the PLC battery every 3 years with Allen Bradley PLC, others??

Page 35: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Smart Time Fault

PC and PLC

Synchronization between PLC & PC

S b t PC & PLC

What:

Where:

When:

35

Sync. timer check between thetwo computers (each has independent timers)

Sync. error between PC & PLCWhen:

Why:

How: Self checking

Page 36: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Fault: Smart Time Communication

PC server and the PLC

Between PC & PLC

C ti

What:

Where:

When:

36

Must communicate for proper operation

ContinuousWhen:

Why:

How: Self checking

Page 37: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: I/O Fuse

I/O electronic fuses

In the PLC

If h t d f ti

What:

Where:

When:

37

Prevent damage to the PLC

If short or defective component input/output

When:

Why:

How: Self checking

Page 38: Irradiator Technology H-315 Session 07 Safety Systems ...Irradiator Technology H-315 Session 07 1 Session 07 Safety Systems Timing Circuits Radiation Detection Mechanical or Physical

Caution: Storage Stop

Three emergency stop buttons

In the warehouse storage area

Wh f th b b tt

What:

Where:

When:

38

Machine would eventually run out of product

When any of the above buttonsare pressed

When:

Why:

How: Physically pressing any button