Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories Slide 1 Slide 1 Transparent Accountable Scientifically Defensible Slide 1 National Water Quality Monitoring Council Webinar Series January 25, 2017 Beverly H. van Buuren, Principal Investigator Quality Assurance Services Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories Moss Landing, California Email: [email protected] Phone: 206-297-1378 Introduction to Reporting Limits © All slides and content copyright protected. Must receive permission prior to use or reproduction.

Upload: others

Post on 11-Mar-2020

13 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 1 Slide 1 Transparent Accountable Scientifically Defensible Slide 1

National Water Quality Monitoring Council Webinar Series

January 25, 2017

Beverly H. van Buuren, Principal Investigator Quality Assurance Services

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Moss Landing, California

Email: [email protected]

Phone: 206-297-1378

Introduction to Reporting Limits

© All slides and content copyright protected. Must receive permission prior to use or reproduction.

Page 2: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 2 Slide 2 Transparent Accountable Scientifically Defensible Slide 2

Marine Pollution Studies Laboratory

Moss Landing Marine Laboratories

The Marine Pollution Studies Laboratory (MPSL) is a collaborative research

consortium of scientists at Moss Landing Marine Laboratories (MLML).

Services

• Quality Assurance

• Data Management

• Field Sampling

Scope

• Monitoring

• Assessment

• Compliance

• Research

• Spill Response

• Emerging Contaminants

• Wastewater Treatment

• Litigation Support

Page 3: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 3 Slide 3 Transparent Accountable Scientifically Defensible Slide 3

Quality Assurance Services

Since 1998, partnering with the academic, government, and private sectors to

build tools and processes that enhance the transparency, accountability, and

scientific defensibility of environmental data collection, analysis, and reporting

California Department of Fish and Wildlife

CALFED Science Program

California State Water Resources Control Board

National Oceanic and Atmospheric Administration

United States Geological Survey

United States Environmental Protection Agency

Pesticide Industry

Pharmaceutical Industry

Mining Industry

Timber Industry

Build Large-Scale QA Programs

Provide Project QA Services

Page 4: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 4 Slide 4 Transparent Accountable Scientifically Defensible Slide 4

• Introduction

• Definitions

• Examples

• Intro to Determining Program/Project Reporting Limits

• Reporting Limits in Databases and Reports

• Working with a Laboratory and Reporting Limits

• Documents for Communicating Reporting Limits

• Conclusion

Agenda

Page 5: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 5 Slide 5 Transparent Accountable Scientifically Defensible Slide 5

Desired Outcome

• A general understanding of common detection

and quantitation terms

• An understanding that there are differences

between detection and quantitation limits

• An introduction to determining RLs

• An appreciation for linking RLs to data use (e.g.,

decision making)

Page 6: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 6 Slide 6 Transparent Accountable Scientifically Defensible Slide 6

Text and References

• There are several slides with a significant

amount of text and definitions.

• There are also several slides that show tools

and list web site addresses.

• We will not be going over these verbatim; they

are included so that you may use the slides later

as a reference.

Page 7: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 7

Introduction

Page 8: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 8 Slide 8 Transparent Accountable Scientifically Defensible Slide 8

Why Reporting Limits Matter

Page 9: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 9 Slide 9 Transparent Accountable Scientifically Defensible Slide 9

Data Use - Examples

Ability to make recommendations and/or decisions

related to…

Improved Water

Supply

Critical Species

and Habitat

Long-term

Water Resources

Why Reporting Limits Matter

Page 10: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 10 Slide 10 Transparent Accountable Scientifically Defensible Slide 10

• A general understanding of common detection

and quantitation terms

• An understanding that there are differences

between detection and quantitation limits

• An introduction to determining RLs

• An appreciation for linking RLs to data use (e.g.,

decision making)

Desired Outcome

Page 11: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 11 Slide 11 Transparent Accountable Scientifically Defensible Slide 11

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Action Limit

(or water quality standard)

Practical Quantitation Limit

Page 12: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 12 Slide 12 Transparent Accountable Scientifically Defensible Slide 12

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit =

PQL

Action

Limit

Concentration

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

CTR Water

0.05 ug/L

Non-Detect Detected but not Quantified Quantified Quantified w/ Stat. Rigor

RL = PQL

0.0015 ug/L

Page 13: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 13

Definitions

Page 14: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 14 Slide 14 Transparent Accountable Scientifically Defensible Slide 14

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Action Limit

(or water quality standard)

Practical Quantitation Limit

Page 15: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 15 Slide 15 Transparent Accountable Scientifically Defensible Slide 15

Method Detection Limit

Sample Prep + Analyses + Lab = MDL

Method

Page 16: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 16 Slide 16 Transparent Accountable Scientifically Defensible Slide 16

Method Detection Limit

• 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision 2

o Google: e-CFR title 40 part 136, go to App. B

• Definition: The method detection limit (MDL) is defined as the minimum measured concentration of a substance that can be reported with 99% confidence that the measured concentration is distinguishable from method blank results. (EPA 821-R-16-006 December 2016)

• The MDL for an analytical procedure may vary as a function of sample type. The procedure requires a complete, specific, and well-defined analytical method. It is essential that all sample processing steps of the analytical method be included in the determination of the method detection limit.

Page 17: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 17 Slide 17 Transparent Accountable Scientifically Defensible Slide 17

https://www.epa.gov/sites/production/files/2016-12/documents/mdl-

procedure_rev2_12-13-2016.pdf

Page 18: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 18 Slide 18 Transparent Accountable Scientifically Defensible Slide 18

Method Detection Limit

• Final rule update signed on December 15, 2016

• The new process takes background contamination into consideration in

the determination of detection limits. This will reduce false positives due

to blank bias.

• MDLs will be representative of lab performance over time, compared to

capturing MDL data on a single day.

• Allows the lab to combine data from more than one instrument to

calculate a lab-wide MDL, rather than individual instrument-specific

MDLs.

• Fact sheet available at: www.epa.gov/cwa-methods/methods-

update-rule-support-documents

• FAQ available at: www.epa.gov/cwa-methods/method-detection-

limit-frequent-questions#questions-mur

Page 19: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 19 Slide 19 Transparent Accountable Scientifically Defensible Slide 19

Method Detection Limit Process Summary

• A lab determines its MDLs based on a minimum of seven spiked samples and

seven method blank samples that go through all steps of the method.

• The spiking concentrations used to determine an MDL are between 1 and 10

times the estimated MDL and should be re-evaluated annually.

• The samples used for the MDL must be prepared in at least three separate

batches and analyzed on three separate days. Existing data may be used for

MDL calculation as long as it is collected on different days.

• Calculate the spiked sample MDL (MDLs) by using standard deviation of the

results and the appropriate student’s t-value and the blank sample MDL (MDLb)

by using the mean results and the appropriate Student t-value.

• Select the greater value between MDLs and MDLb as the initial MDL.

• During any quarter in which samples are being analyzed, prepare and analyze

a minimum of two spiked samples on each instrument, in separate batches.

Routine method blanks can be used to calculate MDLb

• At least once every thirteen months, re-calculate MDLs and MDLb

Page 20: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 20 Slide 20 Transparent Accountable Scientifically Defensible Slide 20

Method Detection Limit

Sample Prep + Analyses + Lab = MDL

The higher value of seven spike replicates or seven blank replicates

MDL = lowest level signal produced

A signal is detected

Method

Page 21: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 21 Slide 21 Transparent Accountable Scientifically Defensible Slide 21

Concentr

ation

Method Detection Limit

Minimum Level

Page 22: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 22 Slide 22 Transparent Accountable Scientifically Defensible Slide 22

Minimum Level

Method + MDL + Factor = ML

ML = lowest point on calibration curve

EPA/

Std

Method

or

SOP

Page 23: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 23 Slide 23 Transparent Accountable Scientifically Defensible Slide 23

Page 24: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 24 Slide 24 Transparent Accountable Scientifically Defensible Slide 24

Minimum Level

A minimum level at which the analytical system shall give

recognizable mass spectra (background corrected) and acceptable

calibration points.

Page 25: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 25 Slide 25 Transparent Accountable Scientifically Defensible Slide 25

Example: EPA Method 1631 – Mercury in Water

“The method detection limit for Hg has

been determined to be 0.2 ng/L when no

interferences are present. The minimum

level of quantitation (ML) has been

established as 0.5 ng/L. An MDL as low as

0.05 ng/L can be achieved for low Hg

samples by using a larger sample volume,

a lower BrCl level (0.2%), and extra

caution in sample handling.”

Page 26: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 26 Slide 26 Transparent Accountable Scientifically Defensible Slide 26

Minimum Level

• US EPA Method 1631 Revision E, 2002, Page 32, Office of Water

• The lowest level at which the entire analytical system must give a

recognizable signal and acceptable calibration point for the

analyte. It is equivalent to the concentration of the lowest calibration

standard, assuming that all method-specified sample weights,

volumes, and cleanup procedures have been employed.

• The ML is calculated by multiplying the MDL by 3.18 and rounding

the result to the number nearest to (1, 2, or 5) x 10n, where n is an

integer.

• Minimum levels are used in some US EPA methods.

Page 27: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 27 Slide 27 Transparent Accountable Scientifically Defensible Slide 27

Minimum Level

Method + MDL + Factor = ML

ML = lowest point on calibration curve

A signal is quantified

EPA/

Std

Method

or

SOP

Page 28: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 28 Slide 28 Transparent Accountable Scientifically Defensible Slide 28

Method Detection Limit

Sample Prep + Analyses + Lab = MDL

The higher value of 7 spike replicates or

7 blank replicates

MDL = lowest level signal produced

A signal is detected

EPA/

Std

Method

or

SOP

Method + MDL + Factor = ML

ML = lowest point on calibration curve

A signal is quantified

Minimum Level

Page 29: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 29 Slide 29 Transparent Accountable Scientifically Defensible Slide 29

Concentr

ation

Method Detection Limit

Minimum Level

Practical Quantitation Limit

Page 30: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 30 Slide 30 Transparent Accountable Scientifically Defensible Slide 30

Practical Quantitation Limit

Instrument + Analyst + Factor = PQL ** or **

PQL = 3 times lowest point on calibration curve

Page 31: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 31 Slide 31 Transparent Accountable Scientifically Defensible Slide 31

Page 32: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 32 Slide 32 Transparent Accountable Scientifically Defensible Slide 32

Practical Quantitation Limit

• Subjective definitions ?

• A quantity set at two to ten times above the method detection

limit (MDL). By raising the MDL by a factor of two to ten, serving

as a “safety factor,” commercial labs hope to quantify the

environmental sample concentrations with a degree of certainty.

• The degree of the factor (2-10) is decided by the analytical lab

depending upon the skill and experience of the analyst, the

quality of the instrument, and the nature of the sample objectives.

Page 33: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 33 Slide 33 Transparent Accountable Scientifically Defensible Slide 33

Practical Quantitation Limit

Unnamed Western State Water Quality Control Division

US EPA Region III Fact Sheet 2006

Common Practice – 3 times the lowest level standard

Page 34: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 34 Slide 34 Transparent Accountable Scientifically Defensible Slide 34

Practical Quantitation Limit

Instrument + Analyst + Factor = PQL

A signal is quantified

** or **

PQL = 3 times lowest point on calibration curve

A signal is quantified with statistical rigor

Page 35: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 35 Slide 35 Transparent Accountable Scientifically Defensible Slide 35

Method Detection Limit Minimum Level

Practical Quantitation Limit

Page 36: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 36 Slide 36 Transparent Accountable Scientifically Defensible Slide 36

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Page 37: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 37 Slide 37 Transparent Accountable Scientifically Defensible Slide 37

Reporting Limit

Program + Data Use = RL

(Action Limit)

Program

Documents

Page 38: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 38 Slide 38 Transparent Accountable Scientifically Defensible Slide 38

The minimum value below which data are

documented as non-detects.

Question: Does this mean the RL = MDL

Question: What does “are documented” mean

Page 39: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 39 Slide 39 Transparent Accountable Scientifically Defensible Slide 39

https://water.usgs.gov/owq/OFR_99-193/level.html

Page 40: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 40 Slide 40 Transparent Accountable Scientifically Defensible Slide 40

http://pubs.acs.org/doi/pdf/10.1021/ac60259a007

Richard M. Lindstrom, NIST – See Slide 41

Page 41: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 41 Slide 41 Transparent Accountable Scientifically Defensible Slide 41

http://nvlpubs.nist.gov/nistpubs/sp958-lide/164-166.pdf

Page 42: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 42 Slide 42 Transparent Accountable Scientifically Defensible Slide 42

Page 43: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 43 Slide 43 Transparent Accountable Scientifically Defensible Slide 43

Different Definitions - Examples

• An instrument-dependent quantity based on the lowest point on the calibration curve. ((Unnamed

North Eastern State) Department of Environmental Protection)

• A limit imposed upon the reporting lab. The RL is usually demanded by the client or regulatory

guidelines, and is basically associated with method detection limits (MDLs) or practical quantitation

limits (PQLs). (Unnamed Western State Regional Water Quality Control Board )

• Reporting Limit (RL)—is the lowest concentration at which an analyte can be detected in a sample and

its concentration can be reported with a reasonable degree of accuracy and precision. A criterion of ±

20% accuracy and 20% RSD for replicate determinations is often used to define “reasonable”. The

acceptable ranges depend somewhat on the analytical methodology used. For samples that do not

pose a particular matrix problem, the RL is typically about three to five times higher than the MDL.

Similar to the MDL, the RL is a laboratory-specific number, which may change with time. When a

sample has to be diluted before analysis, either because of matrix problems or to get the instrument

response within the linear dynamic range, the RL is raised by a factor corresponding to the dilution

factor. (Unnamed Federal Sanitation and Radiation Laboratory)

Page 44: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 44 Slide 44 Transparent Accountable Scientifically Defensible Slide 44

Term is Undefined

Reporting Limit

as a term

pizza

Project or program

must have a written definition

Page 45: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 45 Slide 45 Transparent Accountable Scientifically Defensible Slide 45

Reporting Limit

Program + Data Use = RL

(Action Limit)

Program

Documents

Page 46: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 46 Slide 46 Transparent Accountable Scientifically Defensible Slide 46

Data Use - Examples

Ability to make recommendations and/or decisions

related to…

Improved Water

Supply

Critical Species

and Habitat

Long-term

Water Resources

Why Reporting Limits Matter

Page 47: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 47 Slide 47 Transparent Accountable Scientifically Defensible Slide 47

Method Detection Limit

Minimum Level

Practical Quantitation Limit

Reporting Limit

Dependent on METHOD

Dependent on USE

Analytical Method

Data Use in Decisions

Page 48: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 48 Slide 48 Transparent Accountable Scientifically Defensible Slide 48

Reporting Limit

Program + Data Use = RL

(Action Limit)

A signal is quantified and is more robust if it incorporates

statistical rigor

Program

Documents

Page 49: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 49 Slide 49 Transparent Accountable Scientifically Defensible Slide 49

Method Detection Limit Minimum Level

Practical Quantitation Limit Reporting Limit

Page 50: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 50 Slide 50 Transparent Accountable Scientifically Defensible Slide 50

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Page 51: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 51 Slide 51 Transparent Accountable Scientifically Defensible Slide 51

• MDL = Method Detection Limit

Detected

• ML = Minimum Level

Quantified

• PQL = Practical Quantitation Limit

Quantified

If 3x the lowest point on calibration curve

Quantified with Statistical Rigor

• RL = Reporting Limit

Quantified

May be defined as quantified with statistical rigor

LOD = Limit of Detection

LOQ = Limit of Quantification

Detected

Quantified

Page 52: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 52 Slide 52 Transparent Accountable Scientifically Defensible Slide 52

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Action Limit

(or water quality standard)

Practical Quantitation Limit

Page 53: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 53 Slide 53 Transparent Accountable Scientifically Defensible Slide 53

Reporting Limit

Program + Data Use = RL

(Action Limit)

Program

Documents

Page 54: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 54 Slide 54 Transparent Accountable Scientifically Defensible Slide 54

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Action Limit

(or water quality standard)

Practical Quantitation Limit

Protective of Standard

Page 55: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 55 Slide 55 Transparent Accountable Scientifically Defensible Slide 55

0 Method

Detection

Limit

Minimum

Level

Reporting

Limit

Action

Limit

Protection of Standard

Concentration

Page 56: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 56

Examples

Page 57: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 57 Slide 57 Transparent Accountable Scientifically Defensible Slide 57

Example: Unnamed Midwestern State

Environmental Protection Agency

Parameter Units Typical MDL Practical

Quantitation Limit

Maximum

Contaminant

Level

Carbon Tetrachloride μg/L 0.02 (EPA 180.1) 0.5 25

Chlorobenzene μg/L 0.03 (EPA 200.8) 2.0 14

Chloroform μg/L 0.2 (EPA 200.8) 2.5 117

Same as “Action Limit”

From Document’s Text:

“PQL variation may be due to such issues as ground water matrix interference,

analytical method, laboratory, laboratory personnel or a change in

analytical instruments. Such variability is not unexpected and reflects the

nature of PQLs.”

Page 58: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 58 Slide 58 Transparent Accountable Scientifically Defensible Slide 58

Example: Unnamed Western State

Department of Ecology

Parameter Units Typical MDL Laboratory

Quantitation Level Benchmark

Malathion μg/L 0.02 (EPA 8141A) 0.5 25

Mirex μg/L 0.03 (EPA 8081A) 2.0 14

Simazine μg/L 0.2 (EPA 200.8) 2.5 117

Same as “Minimum Level” Same as “Action Limit”

Concentration

Page 59: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 59 Slide 59 Transparent Accountable Scientifically Defensible Slide 59

Example: Unnamed Southwestern State

Commission on Environmental Quality • Surface Water Quality Standards are written by the ### under the authority of the

Clean Water Act and the ### Water Code. The standards are effective for Clean

Water Act purposes when approved by the EPA.

• Limit of Quantification (LOQ) Criteria are located in Appendix A

Example: Unnamed North Eastern State

Department of Environmental Protection • Reporting Limit (RL) is defined as the concentration of the lowest standard in the

calibration curve for organics and the lowest concentration standard used in the

calibration of the method and for inorganics, derived from the concentration of that

analyte in the lowest level check standard (which could be the lowest calibration

standard in a multi-point calibration curve).

Page 60: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 60

Determining

Reporting Limits

Page 61: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 61 Slide 61 Transparent Accountable Scientifically Defensible Slide 61

How do I determine program/project RLs?

• Program/project RLs should be based on the data use.

o This may include, but is not limited to, water quality standards, assessment

thresholds, TMDLs, regulatory contexts, and the use of results with other testing

(such as toxicity testing).

• Labs are great resources for information!

• Other programs, states, projects, etc are great resources for

information

• It is important to understand that you cannot set RLs below the

lowest level of state-of-the-art analytical capabilities.

• Consider cost/benefit

o Lower RL may mean fewer resources for field samples

Page 62: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 62 Slide 62 Transparent Accountable Scientifically Defensible Slide 62

Data Use – Examples

Ability to make recommendations and/or decisions

related to…

Improved Water

Supply

Critical Species

and Habitat

Long-term

Water Resources

Why Reporting Limits Matter

Page 63: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 63 Slide 63 Transparent Accountable Scientifically Defensible Slide 63

How do I determine program/project RLs?

• Program/project RLs should be based on the data use.

o This may include, but is not limited to, water quality standards, assessment

thresholds, TMDLs, regulatory contexts, and the use of results with other testing

(such as toxicity testing).

• Labs are great resources for information!

• Other programs, states, projects, etc. are great resources for

information

• It is important to understand that you cannot set RLs below the

lowest level of state-of-the-art analytical capabilities.

• Consider cost/benefit

o Lower RL may mean fewer resources for field samples

Page 64: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 64 Slide 64 Transparent Accountable Scientifically Defensible Slide 64

Page 65: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 65 Slide 65 Transparent Accountable Scientifically Defensible Slide 65

Example: California Toxics Rule

• On May 18, 2000, EPA promulgated numeric water quality criteria

for priority toxic pollutants and other provisions for water quality

standards to be applied to waters in the state of California. EPA

promulgated this rule - also known as the California Toxics Rule

(CTR) - based on the Administrator's determination that numeric

criteria are necessary in California to protect human health and the

environment. (EPA)

• CTR values can be used as guidance for setting action limits and

RLs that are protective of human health and aquatic life.

• Specific CTR standards can be found in 40 CFR 131.38

www.gpo.gov/fdsys/pkg/CFR-2012-title40-vol23/pdf/CFR-2012-

title40-vol23-sec131-38.pdf

Page 66: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 66 Slide 66 Transparent Accountable Scientifically Defensible Slide 66

Select Parameter

Determine Action Limit

Check Methods

Establish RL

Check Lab MDL & RL

Process for Determining Reporting Limits

Page 67: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 67 Slide 67 Transparent Accountable Scientifically Defensible Slide 67

Select Parameter

Determine Action Limit

Check Methods

Establish RL

Check Lab MDL & RL

Mercury, total water

CTR water 0.05 ug/L

Page 68: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 68 Slide 68 Transparent Accountable Scientifically Defensible Slide 68

0 Method

Detection

Limit

Minimum

Level

Reporting

Limit

Action

Limit

Protection of Standard

Concentration

Establish RL

10x < CTR = 0.005 ug/L

RL

0.005 ug/L

CTR Water

0.05 ug/L

Page 69: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 69 Slide 69 Transparent Accountable Scientifically Defensible Slide 69

Select Parameter

Determine Action Limit

Check Methods

Establish RL

Check Lab MDL & RL

Use National Environmental Methods

Index to find methods: www.nemi.gov

Mercury, total water

CTR water 0.05 ug/L

10x < CTR = 0.005 ug/L

Page 70: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 70 Slide 70 Transparent Accountable Scientifically Defensible Slide 70

Page 71: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 71 Slide 71 Transparent Accountable Scientifically Defensible Slide 71

Method # Source Detection

Level

Detection

Type

200.8 EPA 0.2 ug/L MDL

200.7 EPA 7 ug/L MDL

245.1 EPA 0.2 ug/L RNGE

I-1462 USGS 0.5 ug/L RNGE

I-7462 USGS 0.5 ug/L RL

D6502 ASTM 1 ug/L ML

1631 EPA 0.0002 ug/L MDL

245.7 EPA 0.0018 ug/L MDL

RL = 0.005 ug/L

Page 72: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 72 Slide 72 Transparent Accountable Scientifically Defensible Slide 72

Select Parameter

Determine Action Limit

Check Methods

Establish RL

Check Lab MDL & RL

1631 Minimum Level = 0.0005 ug/L Cost $51-200

245.7 Minimum Level = 0.005 ug/L Cost $51-200

Page 73: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 73 Slide 73 Transparent Accountable Scientifically Defensible Slide 73

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit

Action

Limit

Protection of Standard

Practical Quantitation Limit

3x lowest level point on calibration curve

0.0015 ug/L

Concentration

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

RL

0.005 ug/L

CTR Water

0.05 ug/L

Page 74: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 74 Slide 74 Transparent Accountable Scientifically Defensible Slide 74

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit

Action

Limit

Protection of Standard

Practical Quantitation Limit

3x lowest level point on calibration curve

0.0015 ug/L

Concentration

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

RL

0.005 ug/L

CTR Water

0.05 ug/L

Quantified

Page 75: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 75 Slide 75 Transparent Accountable Scientifically Defensible Slide 75

Select Parameter

Determine Action Limit

Check Methods

Establish RL

Check Lab MDL & RL

Back and forth

process

Page 76: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 76 Slide 76 Transparent Accountable Scientifically Defensible Slide 76

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit

Action

Limit

Protection of Standard

Practical Quantitation Limit

3x lowest level point on calibration curve

0.0015 ug/L

Concentration

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

RL

0.005 ug/L

CTR Water

0.05 ug/L

Quantified

Page 77: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 77 Slide 77 Transparent Accountable Scientifically Defensible Slide 77

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit =

PQL

Action

Limit

Protection of Standard

Concentration

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

RL = PQL

0.0015 ug/L

CTR Water

0.05 ug/L

Page 78: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 78

Reporting Limits in

Databases and Reports

Page 79: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 79 Slide 79 Transparent Accountable Scientifically Defensible Slide 79

Reporting

When reporting data, other terms worth

understanding include:

• Not Detected (ND): The sample result is less than the MDL. The

analyte being tested cannot be detected by the equipment or method.

• Detected Not Quantifiable (DNQ): The sample result is between the

MDL and the ML. These results may be reported as the measured value

(not negative) with a flag that is carried all the way through data storage,

handling, and reporting.

Page 80: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 80 Slide 80 Transparent Accountable Scientifically Defensible Slide 80

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit =

PQL

Action

Limit

Protection of Standard

Concentration

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

CTR Water

0.05 ug/L

Non-Detect Detected but not Quantified Quantified Quantified w/ Stat. Rigor

RL = PQL

0.0015 ug/L

Page 81: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 81 Slide 81 Transparent Accountable Scientifically Defensible Slide 81

Page 82: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 82 Slide 82 Transparent Accountable Scientifically Defensible Slide 82

Page 83: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 83 Slide 83 Transparent Accountable Scientifically Defensible Slide 83

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit =

PQL

Action

Limit

Protection of Standard

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

CTR Water

0.05 ug/L

Non-Detect Detected but not Quantified Quantified Quantified w/ Stat. Rigor

RL = PQL

0.0015 ug/L

National Functional Guidelines Flags – U, J, UJ

Other Laboratory Flags – ND, DNQ

Page 84: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 84

Working with a Laboratory

Page 85: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 85 Slide 85 Transparent Accountable Scientifically Defensible Slide 85

Laboratory Reporting Limits

• Supply information to lab

o Table with matrix/analyte combinations with MDL, ML, and RL

o Written definitions for each term

• Generally, labs will establish RLs

o based on the lowest point in the calibration curve, or

o as 3x the lowest point in the calibration curve, or

o at 2-5x the MDL.

• Ask labs how

o Ask how dilutions are handled

o Ask how results are reported (esp. between 0-MDL, MDL-ML)

Page 86: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 86 Slide 86 Transparent Accountable Scientifically Defensible Slide 86

0

Method

Detection

Limit

Minimum

Level

Reporting

Limit =

PQL

Action

Limit

Protection of Standard

EPA 1631

0.0002 ug/L

EPA 1631

0.0005 ug/L

CTR Water

0.05 ug/L

Non-Detect Detected but not Quantified Quantified Quantified w/ Stat. Rigor

RL = PQL

0.0015 ug/L

National Functional Guidelines Flags – U, J, UJ

Other Laboratory Flags – ND, DNQ

Page 87: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 87 Slide 87 Transparent Accountable Scientifically Defensible Slide 87

Laboratory Reporting Limits

• Supply information to lab

o Table with matrix/analyte combinations with MDL, ML, and RL

o Written definitions for each term

• Generally, labs will establish RLs

o based on the lowest point in the calibration curve, or

o as 3x the lowest point in the calibration curve, or

o at 2-5x the MDL.

• Ask labs how

o Ask how dilutions are handled

o Ask how results are reported (esp. between 0-MDL, MDL-ML)

Reporting Limit

as a term

Page 88: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 88

Documents for Communicating

Reporting Limits

Page 89: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 89 Slide 89 Transparent Accountable Scientifically Defensible Slide 89

Documents

• Request for Proposals (RFPs)

• QA Project Plans

• QA Manuals or other program documents

• Data Qualification/Flagging Manuals (or procedures)

• Permits

• Contracts

Tip – Have QA staff review

Tip – Have QA staff review

Measurement Quality Objectives

Holding Times

Methods

MDLs, MLs, RLs

How to handle NDs and DNQs

Reporting Formats

Timelines

Subcontracting Work If it’s in the contract, people pay attention

Page 90: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Transparent Accountable Scientifically Defensible

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 90

Conclusion

Page 91: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 91 Slide 91 Transparent Accountable Scientifically Defensible Slide 91

Why Reporting Limits Matter

Page 92: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 92 Slide 92 Transparent Accountable Scientifically Defensible Slide 92

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Action Limit

(or water quality standard)

Practical Quantitation Limit

Page 93: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 93 Slide 93 Transparent Accountable Scientifically Defensible Slide 93

Scientifically

Defensible Accountable Transparent

Defensible Comparable

Usable

QUALITY ASSURANCE COMPONENTS

Measurable Metrics

Page 94: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 94 Slide 94 Transparent Accountable Scientifically Defensible Slide 94

Concentr

ation

Method Detection Limit

Minimum Level

Reporting Limit

Action Limit

(or water quality standard)

Practical Quantitation Limit

Page 95: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 95 Slide 95 Transparent Accountable Scientifically Defensible Slide 95

Why Reporting Limits Matter

Page 96: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 96 Slide 96 Transparent Accountable Scientifically Defensible Slide 96

0 Method

Detection

Limit

Minimum

Level

Reporting

Limit

Action

Limit

Protection of Standard

Concentration

Why Reporting Limits Matter

Reporting limits must be protective of our water quality standards

Page 97: Introduction to Reporting Limits - ACWIMethod Detection Limit • 40 CFR Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision

Marine Pollution Studies Laboratory at the Moss Landing Marine Laboratories

Slide 97 Slide 97 Transparent Accountable Scientifically Defensible Slide 97

Marine Pollution Studies Laboratory

Moss Landing Marine Laboratories

Beverly H. van Buuren, Principal Investigator

[email protected]

206-297-1378

QUESTIONS?