Page 1: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

Spectrophotometer

CM-2300d

Instruction Manual

Page 2: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

Safety SymbolsThe following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument.

Denotes a sentence regarding a safety warning or note.Read the sentence carefully to ensure safe and correct use.

Denotes a prohibited operation.The operation must never been performed.

Denotes an instruction.The instruction must be strictly adhered to.

Denotes a prohibited operation.Never disassemble the instrument.

Denotes an instruction.Disconnect the AC power cord from the AC outlet.

Notes on this Manual• Copying or reproduction of all or any part of the contents of this manual without KONICA MINOLTA

SENSING’s permission is strictly prohibited.• The contents of this manual are subject to change without prior notice.• Every effort has been made in the preparation of this manual to ensure the accuracy of its contents.

However, should you have any questions or find any errors, please contact a KONICA MINOLTA SENSING -authorized service facility.

• KONICA MINOLTA SENSING will not accept any responsibility for consequences arising from the use of the instrument.

Page 3: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

1

Safety PrecautionsTo ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises.

WARNING (Failure to adhere to the following points may result in death or serious injury.)

Do not use the instrument in places where flammable or combustible gases (gasoline etc.) are present. Doing so may cause a fire.

Do not disassemble or modify the in-strument or the AC adapter. Doing so may cause a fire or electric shock.The instrument should not be operated if it is damaged or AC adapter is damaged, or if smoke or odd smells occur. Doing so may result in a fire. In such situations, turn the power OFF immediately, dis-connect the AC adapter from the AC outlet (or remove the batteries if they are used) and contact the nearest KONICA MINOLTA SENSING-authorized serv-ice facility.

Always use the AC adapter supplied as a standard accessory or the optional AC adapter, and connect it to an AC outletof the rated voltage and frequency. Use of AC adapters other than those speci-fied by KONICA MINOLTA SENS-ING, may result in damage to the unit, fire or electric shock.If the instrument is not to be used for a long time, disconnect the AC adapter from the AC outlet. Accumulated dirt or water on the prongs of the AC adapter’s plug may cause a fire and should be re-moved.

Do not disassemble or modify the in-strument. Doing so may cause a fire or electric shock.Do not insert or disconnect the AC adapter with wet hands. Doing so may cause electric shock.

Take special care not to allow liquid or metal objects to enter the instrument. Doing so may cause a fire or electric shock. Should liquid or metal objects enter the instrument, turn the power OFF immediately, disconnect the AC adapter from the AC outlet (or remove the batteries if they are used), and con-tact the nearest KONICA MINOLTA SENSING-authorized service facility.

Do not dispose of batteries in fire, short their terminals, apply heat to them, or disassemble them. Also, do not recharge them (if they are not chargeable). Doing so may cause explosion or heat genera-tion, resulting in fire or injury.

CAUTION (Falling to adhere to the following points may result in injury or damage to the instrument or other property.)

Do not perform measurement which the measurement aperture directed towards your face. Doing so may damage them.

Do not place the instrument on an unsta-ble or sloping surface. Doing so may re-sult in its dropping or overturning, causing injury. Take care not to drop the instrument when carrying it.

Do not use batteries other than those specified by KONICA MINOLTA SENSING.When installing batteries in the instru-ment, make sure that they are correctly oriented according to the (+) and (–) marks. Failure to adhere to these in-structions may cause batteries to ex-plode or leakage of electrolyte, resulting in fire, injury or air pollution.

When using the AC adapter, make sure that the AC outlet is located near the in-strument and that the AC adapter can be connected to and disconnected from the AC outlet easily.

Page 4: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

2

Notes on Use<Operating Environment>• The CM-2300d should be used at ambient temperatures between 5°C and 40°C, and at a relative hu-

midity of 80% or less (35°C), without condensation. Do not use it in areas of rapid temperature change.

• Do not leave the CM-2300d in direct sunlight or near sources of heat, such as stoves etc. The internal temperature of the instrument may become much higher than the ambient temperature in such cases.

• Do not use the CM-2300d in areas where dust, cigarette smoke or chemical gases are present. Doing so may cause deterioration in performance or breakdown.

• Do not use the CM-2300d near equipment which produces a strong magnetic field (such as speakers etc.).

• The CM-2300d belongs to installation category II products (equipment which is powered by an AC adapter connected to a commercially available power).

• The CM-2300d belongs to pollution level 2 products (equipment which may cause temporary electrical hazards due to contamination or condensation or products which are used in such an environment).

• Do not use the CM-2300d at altitudes of higher than 2000 m.• This CM-2300d and the AC adaptor supplied as the standard accessory have been designed exclusively

for indoor use. Do not use this instrument outdoors. If this instrument is used outdoors, it may be dam-aged by rain or other environmental conditions.

<Measurement>• When using the instrument upside-down, make sure no dirt or dust get into the aperture.• When using the instrument for long periods of time, the displayed value may change depending on

changes in the environment. Therefore, in order to achieve accurate measurements, we recommend that white calibration be done regularly using the White Calibration Plate.

<White Calibration Plate>• The calibration data for the White Calibration Plate was measured at 23°C. To achieve the highest ac-

curacy when measuring absolute values (colorimetric values), calibration and measurement should be performed at 23°C.

• Do not allow the White Calibration Plate to get scratched or stained.• If you are not going to use the White Calibration Plate, attach the cap to the White Calibration Plate to

prevent entry of ambient light.

<Target Mask>• Do not touch the Target Mask’s inner surface by hand, scratch it or make it dirty.

<Measuring Base>• When removing the “Measuring Base”, make sure that the screws used to attach the base to the instru-

ment are stored properly and are not mislaid. If the screws do become mislaid, use M3 cross-headed screws that are 4 to 5 mm long as replacements. (For details, see page 14.)

• Do not tighten the screws too tightly when attaching the “Measuring Base”. This could damage the “Measuring Base” or the instrument itself.

• Remove the “Measuring Base” before using the optional Zero Calibration Box or Dust Cover Set.

<Power Source>• Make sure that the power switch is set to OFF (“O”) when the CM-2300d is not in use.• Always use the AC adaptor (AC-A17) supplied as a standard accessory and connect it to an AC outlet

of the rated voltage and frequency. Use the AC power supply voltage of the rated supply voltage (with-in ± 10%).

Page 5: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

3

<System>• Do not subject the CM-2300d to strong impact or vibration. Doing so may cause deterioration in per-

formance or breakdown.• Since the specimen measuring port and integrating sphere are extremely precise optical components,

great care should be taken to prevent them getting dirty or exposing them to impact. If you are not go-ing to use the CM-2300d, put it on the White Calibration Plate (CM-A145).

• The CM-2300d may cause interference if used near a television, radio, etc.• Since the CM-2300d uses a microcomputer, the LCD may go blank if it is exposed to strong static elec-

tricity. In this case, turn the power OFF, then turn it ON again. If black smudges appear on the LCD, wait until they disappear naturally.

• When turning the power OFF and then ON again, wait several seconds after turning the power OFF.

<Backup Battery>• Measured data and various settings are stored in the memory backed up by batteries. The backup bat-

teries are automatically charged during operation of this instrument, and can retain the contents of the memory for 4.5 months if they have been fully charged. At the time of purchase, the backup battery may not be fully charged. To charge the backup battery, set the power switch to ON. Charging of the backup battery is performed continuously while the instrument is switched on, even while the instru-ment is being used. Full charging is completed in 25 hours, and there is no danger of overcharging.

• It is recommended to keep a backup of your important data in another recording medium using optional color data software (sold separately).

Note• The backup batteries’ model number is VL2020 (3 V).• Do not try to replace the backup batteries by yourself. Contact a KONICA MINOLTA SENSING-authorised

service facility.

Notes on Storage• The CM-2300d should be stored at temperatures of between 0 and 45°C and maximum relative humid-

ity of 80%. Do not store it in areas subject to high temperatures, high humidity, or rapid changes of temperature, or where condensation may occur. For added safety, it is recommended that it be stored with a drying agent (such as silica gel) at near room temperature.

• Do not leave the CM-2300d inside a car or in the trunk of a car. Under direct sunlight in summer, the increase in temperature can be extreme and may result in malfunction.

• Do not store the CM-2300d in areas where dust, cigarette smoke or chemical gases are present. Doing so may cause deterioration in performance or breakdown.

• Accumulation of dust inside the integrating sphere will hinder accurate measurement. Block the meas-uring port to prevent entry of the dust.

• The White Calibration Plate may become discolored if left exposed to light. Therefore, make sure that the lid is closed to prevent entry of ambient light when it is not in use.

• The Target Masks may discolor if they are left exposed to light. When they are not in use, keep them in a safe place to prevent exposure to light and to protect them from scratches and dust.

• Be sure to keep all packing materials (cardboard box, cushioning material, plastic bags, etc.). They can be used to protect the instrument during transportation to service facility for maintenance (re-calibra-tion etc.).

• If you are not going to use the CM-2300d for more than two weeks, the batteries must be removed. If the batteries are left in the instrument, leakage may occur resulting in damage to the instrument.

Page 6: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

4

Notes on Cleaning• If the CM-2300d becomes dirty, wipe it with a soft, clean dry cloth. Never use solvents such as thinner

and benzene.• If the White Calibration Plate becomes dirty, wipe it gently with a soft, clean dry cloth. If dirt is diffi-

cult to remove, contact the nearest service facility listed on the attached sheet.• If the inner surface of the Target Masks or the inside of the integrating sphere get dirty, contact a

KONICA MINOLTA SENSING-authorized service facility.• Should the CM-2300d break down, do not try to disassemble and repair it by yourself. Contact a

KONICA MINOLTA SENSING-authorized service facility.

Page 7: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

5

ContentsSafety Precautions .............................................................................................................................. 1

Notes on Use .................................................................................................................................. 2Notes on Storage ............................................................................................................................ 3Notes on Cleaning .......................................................................................................................... 4Conventions ................................................................................................................................... 7

Chapter 1 Before Using the Instrument

Accessories........................................................................................................................................... 10Standard Accessories ..................................................................................................................... 10Optional Accessories ..................................................................................................................... 11

Names and Functions of Parts .......................................................................................................... 12Preparation ........................................................................................................................................ 14

Attaching/Removing the “Measuring Base” ................................................................................. 14Cleaning Each Part ......................................................................................................................... 15Inserting the Batteries .................................................................................................................... 17Connecting the AC Adapter ........................................................................................................... 18Turning Power ON ......................................................................................................................... 19Turning Power OFF ....................................................................................................................... 19

System Configuration ........................................................................................................................ 20Items You Must Know ...................................................................................................................... 21

Language Mode ............................................................................................................................. 21Screen Display ............................................................................................................................... 21Battery Alarm ................................................................................................................................ 21Data Saving .................................................................................................................................... 21

Chapter 2 Preparation for Measurement

Flow of Measurement .................................................................................................................... 24Turning Power On for the First Time ............................................................................................. 25

Setting the Display Language ........................................................................................................ 25Setting the Date and Time ............................................................................................................. 26Setting the Mode ............................................................................................................................ 27Setting the Auto Save .................................................................................................................... 28Setting the Display Type ................................................................................................................ 29Setting the LCD Contrast ............................................................................................................... 30

Setting Measurement Conditions ..................................................................................................... 31Setting the Observer ....................................................................................................................... 32Setting the Illuminant ..................................................................................................................... 33Setting the Color Space .................................................................................................................. 34Setting the Number of Measurements for Auto Averaging ........................................................... 35Setting the Specular Component Mode ......................................................................................... 36

Zero Calibration ................................................................................................................................ 37White Calibration .............................................................................................................................. 39Setting a Color Difference Target Data ........................................................................................... 41

Measuring a Color Difference Target Data ................................................................................... 41Selecting a Color Difference Target Data ...................................................................................... 42Deleting a Color Difference Target Data ....................................................................................... 43

Setting Color Difference Tolerances ................................................................................................ 44

Page 8: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

Chapter 3 Measurement

6

Contents

Measurement ...................................................................................................................................... 48Selecting Screens ........................................................................................................................... 49

Displaying the Measurement Results ............................................................................................... 50Sample data screen (color difference value or absolute value) ..................................................... 50Sample data screen (spectral reflectance) ...................................................................................... 51Target data screen (absolute value) ............................................................................................... 51Target data screen (spectral reflectance) ....................................................................................... 52Abbreviations on LCD Display ..................................................................................................... 53Sample Data No. and Target Data No. .......................................................................................... 54Wavelength of Spectral Reflectance Data ..................................................................................... 54Saving Temporary Data ................................................................................................................. 55Switching the Display Types ......................................................................................................... 56Switching between SCI/SCE ......................................................................................................... 57Deleting Data ................................................................................................................................. 58

Chapter 4 Other Functions

Connecting to an External Device .................................................................................................... 62Connecting a Personal Computer .................................................................................................. 62Outputting to a Printer ................................................................................................................... 64Printing Displayed Data ................................................................................................................. 67

Chapter 5 Troubleshooting

Error Messages .................................................................................................................................. 70Troubleshooting ................................................................................................................................. 72

Chapter 6 Appendix

Principles of Measurement ............................................................................................................... 74Illuminating/Viewing System ........................................................................................................ 74Simultaneous SCI/SCE Measurement ........................................................................................... 75

Specifications ...................................................................................................................................... 76Dimensions ......................................................................................................................................... 78

Page 9: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

7

Contents

ConventionsThis manual describes how to set up the CM-2300d which the firmware version is 3.00 or higher and use it to take measurements.

• Page layoutSymbols used in this manual are explained below.*Note that the page shown in the illustration is for explanatory purposes only, and is not an actual page

from this manual.

36

Zero Calibration

Zero Calibration

When you use this instrument for the first time or when you have initialized it, zero calibration must beperformed.

Note

• If the instrument is not used for long periods of time, the zero calibration data stored in the instrument may belost. If the data is lost, zero calibration must be performed again.

Memo

• The effects of stray light inside the measuring part (i.e. light generated due to the flare characteristics of the opticalsystem) will be compensated automatically with the zero calibration data.

[Setting Procedure]

Start the procedure from the <CALIBRATION> screen.

1 When the instrument is turned ON, the <CALIBRATION> screen will appear automatically.

2 Turn B to select “ZERO”.

3 Direct the specimen measuring port into the air.

• Never direct the specimen measuring port toward a light source.• Keep the specimen measuring port more than 1 m away from any reflective items (hands, desks, walls etc.).

4 Make sure that is displayed, then press C (MEAS. button).

Screen when zero calibration is selected

Zero calibration will be performed.

When zero calibration is complete,

the <CALIBRATION> screen will

reappear.

<Settings>• SCI: Measurement area: ø8 mm, SCI (Specular Component Included)• SCE: Measurement area: ø8 mm, SCE (Specular Component Excluded)

Keeping B held down to right or left will cause the value change continuously.

S K I P

W H I T E

Z E R O

T I O NI B R AC A L< >

1 m or more

E S .AMP R E S S

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

MemoGives useful Infor-mation and addi-tional explanations.

NoteGives the points that you should know to perform operations correct-ly. Make sure that you read the notes.

ProcedureShows the operat-ing procedure.

C......... Indicates the <MEAS.> button.

B ... Indicates request to turn the naviga-tion wheel to the right or left.

A ... Indicates request to press the naviga-tion wheel.

Start screenShows the screen from which opera-tion must be started.

ScreenShows the contents of the screen in effect when the given oper-ation is carried out.

TIP (screen)Gives explanation about the screen and operations that can be carried out from this screen.

Screen noteGives the points to note on the opera-tions to be performed from this screen.

SettingsGives the range and explanation of the values to be set in this screen.

For the version of the instrument firmwareThe version of the instrument firmware can be confirmed on the screen which is displayed first after switching power on.

Page 10: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

8

Contents

Page 11: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

9

Chapter 1

Before Using the Instrument

Page 12: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

10

AccessoriesStandard and optional accessories are available with the instrument.

Standard AccessoriesMake sure that all the following items are present.

White Calibration Plate CM-A145Used to perform white calibration.A data disk containing white calibration data is supplied with this accessory.

Memo• This accessory can be used as a table on which to store the CM-2300d.

Target Mask CM-A146 (for ø 8 mm)Used to switch the illumination area (specimen measuring port area) according to the specimen.

Memo• CM-A146 (for ø 8 mm) is already attached to the CM-2300d when it is

supplied.

AC Adapter AC-A17Used to supply power from an AC outlet to the instrument.Input: Voltage 100 to 240 Vac (50-60 Hz)Output: Voltage 5 Vdc Current 2.8 A (Max.)

RS-232C Cable IF-A16(for IBM PC/AT, 9-pin, 2 m)Used to connect the instrument to a personal computer (PC).

AA-size battery (×4)

Page 13: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

11

Accessories

Optional Accessories

Zero Calibration Box (CM-A32)Used to perform zero calibration.

Hard Case (CM-A148)Can be used for storing the CM-2300d, the instruction manual and standard accessories, such as the White Calibration Plate and AC adapter.

NoteThe Hard Case is designed purely for storing the above items and must not be used for transportation purposes.

Dust Cover Set (CM-A149)Used when measuring powder or wet surfaces.It can also be used when woven fabric needs to be laid flat and measured.

MemoThe Dust Cover (CM-A152) can be used as a vinyl cover for the re-placement.

Color Data Software “SpectraMagic NX” (CM-S100w)This software allows you to operate the instrument from your PC, and to process data and manage files.

Printer Cable (CR-A75)*metric screw typeUsed to transfer data to a printer. Connect a printer to the exter-

nal output terminal on the instrument with this cable. The D-sub connector (9-pin) of the cable must be connected to the printer.

Page 14: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

12

Names and Functions of Parts

0

6

9

87

43

2

1

5

A

1 Viewfinder OpenUsed to check the position of the specimen. By sliding the lever you can check whether the specimen is set correctly.

2 Viewfinder leverUsed to open/close the viewfinder. By sliding the lever in the direction of the arrow, the white LED will light up and illuminate the specimen, so the specimen can be seen through the viewfinder to check that it is set correctly.

MemoThe specimen cannot be measured if the viewfinder is open and the white LED is lit.

3 Navigation wheelTo select To enter

MemoWhen selecting an item or setting a value, holding down the navigation wheel will switch the item or value from one to another continuously.

Use this navigation wheel to select an item or set the select-ed item.To select an item, turn it to the right or left until the desired item is shown. To set the selected item, press it.

4 LCD displayDisplays the setting items and measured data.

5 Specimen measuring portA port provided to measure the specimen.

Page 15: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

13

Names and Functions of Parts

6 Measuring button (MEAS. button)Press this button to perform calibration or measurement.

MemoWhen setting measurement conditions or tolerances, this button can be used as UNDO button to return to the previous item.

7 External output terminalTo transfer data to an external device, connect the RS-232C cable (IF-A16) or printer cable (CR-A75) to this terminal.

8 Battery coverA cover for the battery chamber. Four AA-size batteries must be set in the battery chamber in the cor-rect polarity direction.

9 AC adapter terminalWhen using the AC adapter (AC-A17), connect the adapter’s plug to this terminal.

0 POWER switchUsed to turn ON/OFF power. Setting this switch to “ ” turns the power OFF, and setting it to “|” turns the power ON.

A Measuring BaseUse this base to attach the specimen securely to the instrument when analyzing small specimens.

White Calibration Plate CM-A145

1

2

1 CapA cap provided to protect the White Calibration Plate.

NoteIf you are not going to use the White Calibration Plate, attach the cap to the White Calibration Plate to prevent exposure to ambient light and protect it from scratches and dust.

2 White Calibration PlateUsed to perform white calibration of the CM-2300d.If you are not going to use it, attach the cap to prevent exposure to ambient light and protect it from scratches and dust.

Page 16: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

14

Preparation

Attaching/Removing the “Measuring Base”A “Measuring Base” has been included with the Spectrophotometer CM-2300d.This allows small specimens to be attached securely to the instrument when they are being measured, and this enables more accurate measurements to be made.

Measuring Base

“Measuring Base” attachment hole

The “Measuring Base” is shown in the illustration to the right, and is attached to the base of the CM-2300d with two screws.

Note• Make sure that the “Measuring Base” is securely attached to the

instrument before calibrating the instrument or taking measure-ments.

• You must remove the “Measuring Base” before using the optional Zero Calibration Box or Dust Cover Set. Specifications for “Measuring Base” attachment screws

Thickness of screw head (head height): less than 3 mm

Length of screw body (body length): 4-5 mm

Screw diameter: φ 3 mm

Diameter of screw head (head diameter): φ 5-6.5 mm

Terms in parenthesis () are JIS defined terms for parts of screws

• When removing the “Measuring Base”, make sure that the screws used to attach the base to the instrument are stored properly and are not mislaid. If the screws do become mislaid, replace them with screws that meet the specifications shown to the right.

[Screw name]Cross-Recessed Pan Head Machine Screw M3, Nominal Length 4-5 mm

Attaching/Removing the “Measuring Base” to/from the Instrument

Removing the “Measuring Base”

1. Place the CM-2300d so that its base is facing up and it is stable.

2. Using a crosshead screwdriver, turn the two attach-ment screws counter-clockwise and remove them. Use a crosshead screwdriver that is a suitable size for the screws.

Attaching the “Measuring Base”

1. Place the CM-2300d so that its base is facing up and it is stable.

2. Place the “Measuring Base” on the base of the CM-2300d as shown in the illustration. Place the “Measuring Base” so that it is aligned with the attach-ment holes in the base of the CM-2300d. When placing the “Meas-uring Base” on the base of the CM-2300d, make sure that nothing is trapped between the instrument and the base.

3. Using a crosshead screwdriver, turn the two at-tachment screws clockwise and tighten them se-curely. Do not tighten the screws too tightly.

Page 17: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

15

Preparation

Cleaning Each PartThis section explains how to clean the White Calibration Plate and inside of the integrating sphere.

White Calibration Plate

Gently wipe off dirt with a soft dry cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe. Then remove the liquid with a cloth dampened with water, and leave it to dry.

NoteTake care not to scratch the White Calibration Plate.

Target Mask

Use a blower to remove dirt and dust from the Target Masks.

NoteDo not touch the inner surface of the Target Masks with your fingers or wipe them with a cloth. If the Target Masks are so dirty that dirt cannot be removed with a blower, contact the nearest KONICA MINOLTA SENSING-authorized service facility.

Inside the Integrating Sphere

MemoFor attaching/removing a Target Mask, refer to “Attaching/Removing a Target Mask to/from the Instrument” on page 16.

1. To prevent dust and dirt entering the optics sec-tion from the integrating sphere, slide the view-finder lever to open the viewfinder.

2. Remove the Target Mask.

3. Use a blower to remove dirt and dust from the in-tegrating sphere.

NoteDo not touch the white-coated inner surface of the integrating sphere, wipe it with a cloth or put any object inside it. If the surface is so dirty that dirt cannot be removed with a blower, contact the nearest KONICA MINOLTA SENSING-authorized service facility.

Page 18: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

16

Preparation

Attaching/Removing a Target Mask to/from the Instrument

MemoTo facilitate attaching/removing a Target Mask, turn the instrument over so that the specimen measuring port is face up.

Note• When attaching/removing a Target Mask, take care not to allow dirt and dust to enter the integrating sphere

though the measuring port.• Do not exert excessive force on the latch of the Target Mask. Doing so may damage the latch, disabling use of

the Target Mask.

Target Mask

Latch

Positioning marker

Specimen measuring port

Attaching a Target Mask

1. Place the Target Mask on the specimen measuring port so that the marker on the latch is aligned with the “ ” marker on the port.

2. Hold the outer edge of the mask, and turn it in the direction of the arrow (clockwise) until the marker on the latch is aligned with the “o” marker on the port.

Removing the Target Mask

1. Hold the outer edge of the mask, and turn it in the opposite direction to the arrow (counter-clock-wise) until the marker on the latch is aligned with the “ ” marker.

2. Hold the outer edge of the mask and remove it.

Page 19: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

17

Preparation

Inserting the BatteriesTo supply power to the instrument, the AC adapter (AC-A17) or four AA-size batteries (Alkaline or Ni-MH battery is recommended for better service life) must be used. Use either the AC adapter or batteries, according to which suits your application.

Note• If you are not going to use the instrument for more than two weeks, make sure that the batteries are removed.

If the batteries are left in the instrument for long periods of time, battery electrolyte may leak and damage the instrument.

• Do not use batteries of different types or mix new batteries with old ones. Doing so may result in battery explo-sion or reduction of battery life.

• Do not touch or short-circuit the terminals inside the battery chamber. Doing so may result in breakdown of the instrument.

[Operating Procedure]

1 Make sure that power is OFF (i.e. the POWER switch is set to “ ”).

2 Slide the battery cover on the rear of the instru-ment to open it.

3 Place four AA-size batteries in the battery chamber. Make sure that the batteries are placed in the correct direction.

4 Align the marker on the battery cover with that on the instrument, and slide the battery cover to close it.

Page 20: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

18

Preparation

Connecting the AC Adapter

MemoUse of the AC adapter (AC-A17) rather than batteries is recommended, since more power will be required when the external output terminal is used to output data to an external device or print it.

Note• To supply AC power to the instrument, always use the AC adapter (AC-A17) supplied with the instrument.(Rat-

ed: 5 V, 2.8 A)• Before connecting or removing the AC adapter, make sure that power is turned OFF.

[Operating Procedure]

1 Make sure that power is OFF (i.e. the power switch is set to “ ”).

2 Connect the AC adapter’s connector plug to the AC adapter terminal on the rear of the in-strument.

3

AC outlet

Insert the AC adapter’s power plug to an AC outlet (100-240 Vac, 50-60 Hz).

Page 21: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

19

Preparation

Turning Power ON

NoteWhen turning the power ON for the first time, the display language and measurement mode must be set.For details, refer to page 25.

[Operating Procedure]

1 Set the POWER switch to “|”.The power will be turned ON.

Turning Power OFF

[Operating Procedure]

1 Set the POWER switch to “ ”.The power will be turned OFF.

Auto Power Save Function

Power save mode will be activated if the MEAS. button and navigation wheel are not operated for more than three minutes. During power save mode, the flash circuit will not be charged. To cancel power save mode, press the MEAS. button to start measurement.

Note• If the MEAS. button is pressed to cancel power save mode, start of measurement will be delayed by a few sec-

onds, so keep the instrument still until the lamp flashes and measurement is taken.• The auto power save function is not available in remote mode (see page 63).

Page 22: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

20

System Configuration

System Configuration

Standard accessories

Optional accessories

AA-size battery

( 4)

AC Adapter

AC-A17

(Dust Cover CM-A152)

Target Mask 8mm

CM-A146

RS-232C Cable

IF-A16

Printer Cable

CR-A75

Printer (commercially available)

Hard Case CM-A148

Dust Cover Set CM-A149

Zero Calibration Box CM-A32

Personal computer

(commercially available)

White Calibration Plate CM-A145

* For details to output data to a printer,

refer to page 62.

* with Measuring

Base

Color Data Software

• SpectraMagic NX

CM-S100w

Page 23: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

21

Items You Must Know

Language ModeContents on the LCD screen can be displayed in English or Chinese.In this manual, explanation of operating methods and display is given for English mode.

Screen DisplayThe basic screen structure is shown below.

Screen title and data No. are displayed.

Conditions are displayed.

Measurement results are displayed.

Battery AlarmWhen using CM-2300d with batteries, battery alarm indication appears when their power level lowers.

[Half battery indication]If this indication (Half battery indication) appears, new batteries should be prepared for replacement in near future. Even with this indication displayed, measurements are still possible.

[Empty battery indication]If this indication (Empty battery indication) appears, measurement or calibration is no longer possi-ble. Replace batteries with new ones.

Data SavingWhen the auto-save function is set to ON, data created with this instrument are saved automatically. When the auto-save function is OFF, data can be saved by selecting “MENU” - “SAVE”.

Page 24: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

22

Items You Must Know

Page 25: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

23

Chapter 2

Preparation for Measurement

Page 26: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

24

Flow of Measurement

Turning Power On for the First Time (see page 25)

Setting Measurement Conditions(see page 30)

Zero Calibration (see page 36)

White Calibration (see page 38)

Turning Power ON (see page 19)

Measuring a Color Difference Target Data (see page 40)

Selecting a Color Difference Target Data (see page 41)

Measurement (see page 46)

Complete Measurement

When Checking Color Difference When Not Checking Color Difference

For the first time For the second and subsequent times

Page 27: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

25

Turning Power On for the First TimeWhen turning the power ON for the first time, the following six items must be set.

(1) Display language <LANGUAGE>(2) Date and time <TIME>(3) Mode <MODE>(4) Auto-save function <AUTO SAVE>(5) Display type <DISPLAY TYPE>(6) LCD contrast <CONTRAST>

These items are saved even when the power is turned OFF. To set these items, perform the procedure given below:

[Setting Procedure]

1 While pressing A, turn the power ON.

The <SETTING> screen will ap-pear. Use this screen to set the nec-essary items.

(Back side)

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

Setting the Display LanguageTo set the display language, perform the procedure given below:

[Setting Procedure]Start the procedure from the <SETTING> screen.

1 Turn B to select “LANGUAGE”, then press A.

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

Page 28: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

26

Turning Power On for the First Time

2 Turn B to select the desired language, then press A. The setting is determined and the <SETTING> screen will reappear.

E N G L I S H

C H I N E S E

S E T T I N G

EGAUGNA

M

L

EMIT

O ED

<Settings>• ENGLISH• CHINESE

Setting the Date and Time

[Setting Procedure]Start the procedure from the <SETTING> screen.

1 Turn B to select “TIME”, then press A. The cursor will move to the “year” field.

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

2 Turn B to select the item to change from “day/month/year, hour: minute: sec-ond”. Select the item, then press A.E will appear to allow you to change the value.

0 9 / 0 8 / 2 0 0 5

1 3 : 4 5 : 2 9

EGAUGNA

M

L

EMIT

O ED

S E T T I N G

• The cursor moves in the order of “day → month → year → hour → minute → second → → ”.

3 Turn B to change the value, then press A. The selected value is set.

0 9 8 / 2 0 0 5

1 3 : 4 5 : 2 9

S E T T I N G

EGAUGNA

M

L

EMIT

O ED

• Keeping B held down to right or left will cause the value change continuously.

<Settings>• Day: 01 to 28, 29, 30 and 31 (varies with the selected month)• Month: 01 to 12• Year: 00 to 37 (lower two digits)• Hour: 00 to 23• Minute: 00 to 59• Second: 00 to 59

Page 29: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

27

Turning Power On for the First Time

4 Turn B to select “ ”, then press A. The setting is determined and the <SETTING> screen will reappear.

Selecting “ ” and pressing A will cancel the setting values and the <SETTING> screen will re-appear.

Setting the Mode

[Setting Procedure]Start the procedure from the <SETTING> screen.

1 Turn B to select “MODE”, then press A.

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

2 Turn B to select the desired mode, then press A. The setting is determined and the <SETTING> screen will reappear.

R E M O

R I N T

LAMRON

T E

P

S E T T I N G

EGAUGNA

M

L

EMIT

O ED

<Settings>• NORMAL: Use this mode when a serial printer or a PC is not used.• PRINT: Use this mode when the instrument is connected to a serial printer. Data is automatically output

to the serial printer every time measurement is performed. • REMOTE: Use this mode when the instrument is connected with a PC. You can use the PC to set up the

instrument, and load data to the PC.

Page 30: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

28

Turning Power On for the First Time

Setting the Auto Save

[Setting Procedure]Start the procedure from the <SETTING> screen.

1 Turn B to select “AUTO SAVE”, then press A.

A V EA U T O S

S E T T I N G

O N

C O N T R A S T

D I S P L A Y T Y P E D I F F

2 Turn B to select “ON” or “OFF”, then press A. The setting is determined and the <SETTING> screen will reappear.

O F F

NOA V EA U T O S

S E T T I N G

D I S P L A Y T Y

C O N T R A S T

<Settings>• ON: All sample data is saved automatically as the measurement is performed.• OFF: Sample data is not saved automatically, but is treated as temporary (TEMP) data.

One TEMP data each for target data and sample data can be used. To save TEMP data, you need to select “MENU” - “SAVE”.

Temporary data screen

T GT G T

M

T E M PT E M P

E N U

NoteIf the next measurement is performed before the current temporary data are saved, the temporary data are automatically overwritten. If the Auto Save setting is changed from OFF to ON, the current temporary data (TEMP data) are automatically deleted. When this happens, the sample data linked to the temporary data of target data turns into absolute value data. The link will not be recovered even when the Auto Save setting is set to OFF again.

Page 31: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

29

Turning Power On for the First Time

Setting the Display Type

[Setting Procedure]Start the procedure from the <SETTING> screen.

1 Turn B to select “DISPLAY TYPE”, then press A.

A V EA U T O S

S E T T I N G

O N

D I S P L A Y T Y P E D I F F

C O N T R A S T

2 Turn B to select the desired display type, then press A. The setting is determined and the <SETTING> screen will reappear. F F

F

S & D I

F

A B

D I

A V EA U T O S

S E T T I N G

D I S P L A Y T Y

C O N T R A S T

<Settings>• ABS&DIFF: Absolute value and color difference are displayed.• DIFFERENCE: Only color difference is displayed. (As for target data, absolute value data are displayed

regardless of this setting. As for the sample data which have been measured as absolute data, only absolute value data are displayed because they have no color difference.)

ABS&DIFF screen

DIFFERENCE screen

T GT G T

Page 32: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

30

Turning Power On for the First Time

Setting the LCD Contrast

[Setting Procedure]Start the procedure from <SETTING> screen.

1 Turn B to select “CONTRAST”, then press A.

A D J U S T

D I S P L A Y T Y- +

S E T T I N G

C O N T R A S T

F I N I S H

2 Turn B to select “ADJUST”, then press A.D will appear to allow you to change the contrast.

D I S P L A Y T Y

S E T T I N G

C O N T R A S T

A D J U S T

- +

F I N I S H

3 Turn B to adjust the LCD contrast, then press A.

4 Turn B to select “●”, then press A.The contrast you set will be confirmed, and the <SETTING> screen will reappear.

Selecting “ ” and pressing A will cancel the setting contrast and the <SETTING> screen will re-appear.

• Turning B to the right will move the contrast adjuster bar to the right, increasing the contrast. Turning B to the left will move the contrast adjuster bar to the left, decreasing the contrast.

• Keeping B held down to left or right will cause the contrast adjuster bar to move continuously.

Note• On the LCD screen of this instrument, the display looks darker in a high temperature environment, and

looks paler in a low temperature environment. When this occurs, adjust the display level so that the dis-play shows its contents properly.

• The display contrast will reset to the default setting when you enter the <SETTING> screen.

Page 33: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

31

Setting Measurement Conditions

Setting Measurement ConditionsThe instrument will make measurements according to the selected measurement condition, which must be set before measuring. The selected measurement conditions are applied to all data stored in the instrument. The following five condition items can be set:

(1) <OBSERVER>(2) <ILLUMINANT 1>, <ILLUMINANT 2>(3) <COLOR SPACE>, <EQUATION>(4) <AUTO AVG> (No. of measurements for auto averaging)(5) <GLOSS> (Specular component)

NoteThe measurement conditions displayed on the target data screen and sample data screen (observer/illuminant, specular component, etc.) are determined by these settings. Consequently, changes in the measurement condi-tions will automatically affect the conditions for the old data (data measured in different environment). Exercise due caution before changing the measurement conditions.

[Setting Procedure]Make sure that power to the instrument is turned OFF.

1 While pressing B to left, turn the power ON. The <ENVIRONMENT> screen will appear. (Back side)

2 Turn B to select “CONDITION”, then press A. The <CONDITION> screen will ap-pear. Use this screen to set the meas-urement conditions.

IVNE MNOR TNE

OT

C

AREL

I O N

ECN

N D I TO

F I N I S H

N T 2

ULLI ANIM 1TN

NOITIDNOC

REVRESBO

MULI L I N A

D 6 5

1 0

ENON

MemoThe current measurement conditions are displayed on the right of the items.

Page 34: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

32

Setting Measurement Conditions

Setting the Observer

[Setting Procedure]Start the procedure from the <CONDITION> screen.

1 Turn B to select “OBSERVER”, then press A.

N T 2

ULLI ANIM 1TN

NOITIDNOC

REVRESBO

MULI L I N A

D 6 5

1 0

ENON

2 Turn B to select the desired observer, then press A. The setting is determined and the <CONDITION> screen will reap-pear.

2

1 0

MI L L U

I L L U M

O B S E R

C O N D I

<Settings>• 10°: 10° observer (CIE1964)• 2°: 2° observer (CIE1931)

Page 35: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

33

Setting Measurement Conditions

Setting the Illuminant

[Setting Procedure]Start the procedure from the <CONDITION> screen.

1 Turn B to select “ILLUMINANT 1” or “ILLUMINANT 2”, then press A.

I L L U M I N A N T 1

I L L U M I N A N T 2

NOITIDNOC

REVRESBO

D 6 5

1 0

ENON

2 Turn B to select the desired illuminant, then press A.

The setting is determined and the <CONDITION> screen will reap-pear.

D 6 5 F 2 F 1 0

D 5

C

A

0 F

F

F

6 F 1 1

7 F 1 2

8I L L U M

D IC O N

I L

O B S E R

L U M

When “ILLUMINANT 1” is selected: Illuminant 1 setting screen

N O N E

D 6 5 F 2 F 1 0

D 5

C

A

0 F

F

F

6 F 1 1

7 F 1 2

8I L L U M

D IC O N

I L

O B S E R

L U M

When “ILLUMINANT 2” is selected: Illuminant 2 setting screen

<Settings>• D65: Standard Illuminant D65; Daylight, Color temperature: 6504K • D50: Illuminant D50; Daylight, Color temperature: 5003K • C: Illuminant C; Daylight (Ultraviolet part is small in comparison with the daylight), Color temperature:

6774K • A: Standard Illuminant A; Incandescent lamp, Color temperature: 2856K • F2: Cool white (fluorescent lamp) • F6: Cool white (fluorescent lamp) • F7: Color rendering A; Daylight white (fluorescent lamp)• F8: Color rendering AAA; Natural white (fluorescent lamp)• F10: 3-band type natural white (fluorescent lamp)• F11: 3-band type cool white (fluorescent lamp)• F12: 3-band type warm white (fluorescent lamp)• NONE: No setting (Not selectable for “ILLUMINANT 1”)

• To display MI (metamerism index), ILLUMINANT 2 must be set.

Page 36: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

34

Setting Measurement Conditions

Setting the Color Space

[Setting Procedure]Start the procedure from the <CONDITION> screen.

1 Turn B to select “COLOR SPACE” or “EQUATION”, then press A.

NOITIDNOC

C O L O R S P A C E

1

L a b

E Q U A T I O N

A U T O A V G .

2 Turn B to select the desired color space, then press A. When “COLOR SPACE” is selected: Color space setting screen

L a b

L CC h

Y x y

X Y ZE Q U A T

D IC O N

A U T O

C O L O R

The setting is determined and the <CONDITION> screen will reap-pear.

When “EQUATION” is selected: Equation setting screen

C M C ( 1 : 1 )

C M C ( 2 : 1 )

C I E 9 4

C I E 0 0

W I A S T M E 3 1 3

E Q U A T

D IC O N

A U T O

C O L O R

<Settings for COLOR SPACE>• L*a*b* : L*a*b* color space• L*C*h : L*C*h color space• Yxy : Yxy color space• XYZ : XYZ color space

<Settings for EQUATION>• E* : Color difference by L*a*b* color space• CMC(1:1) : Color difference by “CMC1:1” color- difference formula• CMC(2:1) : Color difference by “CMC2:1” color- difference formula• CIE94 : Color difference by “CIE94” color- difference formula• CIE00 : Color difference by “CIE DE2000” color- difference formula• WI ASTM E313 : Whiteness index (ASTM E313-73)• YI ASTM D1925 : Yellowness index (ASTM D1925)• If E*, CMC(1:1), CMC(2:1), CIE94 or CIE00 is selected as the color space and ILLUMINANT 1 or

ILLUMINANT 2 as the illuminant, it will be possible to caluculate MI (metamerism index) and display the result.

Page 37: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

35

Setting Measurement Conditions

When “CIE00” is selectedWhen “CIE00” is selected as EQUATION, its parameters can be set.

1. Turn B to select the value of the desired parameter, then press A.E will appear to allow you to change the value. C M C (

C I E 9

W I A

Y I A

C I E 0

l = 1 . 0 0

c = 1 . 0 0

h = 1 . 0 0

C O N D I

C O L O R

E Q U A T

A U T O

• Holding B down to the left or right will cause the value to change continuously.

2. Turn B to change the value, then press A.

3. Turn B to select “ ”, then press A.The setting is determined and the previous pop-up screen will reappear.

Selecting “ ” and pressing A will cancel the setting values and the previous pop-up screen will reappear.

Setting the Number of Measurements for Auto AveragingTo use the average of multiple measurements as sample data, set the number of measurements for auto averaging.Measurements are performed for the specified number during a single measurement.

[Setting Procedure]Start the procedure from the <CONDITION> screen.

1 Turn B to select “AUTO AVG”, then press A.

NOITIDNOC

E

L a bC EAPSROLOC

E Q U A T I O N

1A U T O A V G .

2 Turn B to select the desired number of measurements for auto averaging, then press A.The setting is determined and the <CONDITION> screen will reap-pear.

1 2 3 4 5

6 7 8E Q U A T

D IC O N

A U T O

C O L O R

<Settings>• Specify the number of measurements to be performed (1 to 8).

Page 38: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

36

Setting Measurement Conditions

Setting the Specular Component Mode

[Setting Procedure]Start the procedure from the <CONDITION> screen.

1 Turn B to select “GLOSS”, then press A.

NOITIDNOC

A U T O A V G . 1

G L O S S S C I

P R E V I O U S

2 Turn B to select the desired specular component mode, then press A.The setting is determined and the <CONDITION> screen will reap-pear. S C I

S C E

G L O S S

D IC O N

P R E V I

A U T O

<Settings>• SCI: Measurement area: ø 8 mm, SCI (Specular Component Included)• SCE: Measurement area: ø 8 mm, SCE (Specular Component Excluded)

MemoThe specular component data specified on this screen are displayed on the sample data screen as a measure-ment result. The measurement is performed for both SCI and SCE regardless of the setting on this screen.

Page 39: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

37

Zero Calibration

Zero CalibrationWhen you use this instrument for the first time or when you have initialized it, zero calibration must be performed.Once zero calibration is completed, the zero calibration data will be kept even if the power is turned OFF. Thus, it is not necessary to perform zero calibration each time the power is turned ON.

If neither zero calibration nor white cali-bration is required, you can immediately go to the sample data screen by selecting “SKIP”.

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

N o . 1 2 3 4 5 6 7

Note• Once you skip zero calibration and go to the sample data screen, you cannot perform zero calibration unless

you turn OFF the power and turn it ON again. • If the instrument is not used for long periods of time, the zero calibration data stored in the instrument may be

lost. If the data is lost, zero calibration must be performed again.

Memo• The effects of stray light inside the measuring part (i.e. light generated due to the flare characteristics of the optical

system) will be compensated automatically with the zero calibration data. • The amount of stray light may change because of dust or dirt accumulated on the optical system, temperature, re-

peated operation, or the vibration and shock exerted on the instrument. In such cases, periodic zero calibration is recommended.

[Setting Procedure]Start the procedure from the <CALIBRATION> screen.

1 When the instrument is turned ON, the <CALIBRATION> screen will appear auto-matically.

2 Turn B to select “ZERO”.

UUB T T O NS .E AMP R E S S

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

N o . 1 2 3 4 5 6 7

Page 40: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

38

3 Direct the specimen measuring port into the air.

1 m or more

• Never direct the specimen measuring port toward a light source.

• Keep the specimen measuring port more than 1 m away from any reflective items (hands, desks, walls etc.).

4 Make sure that is displayed, then press C (MEAS. button). Screen when zero calibration is selected

UUB T T O NS .E AMP R E S S

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

N o . 1 2 3 4 5 6 7

Zero calibration will be performed. When zero calibration is complete, the <CALIBRATION> screen will reappear.

Screen during zero calibration

331 2C A L I B R A T I N G

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

N o . 1 2 3 4 5 6 7

Memo• The use of the optional Zero Calibration Box ensures more reliable zero calibration.

To use the Zero Calibration Box, remove the Measuring Base attached to the instrument. • If does not appear due to the auto power save function, it may take some time to start calibration after C is pressed.

Note• When zero calibration is complete, perform white calibration. Measurement cannot be performed if

white calibration has not been completed. • Even if white calibration was performed after power ON, if zero calibration was performed after that,

white calibration must be performed again.

Page 41: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

39

White Calibration

White CalibrationWhite calibration must be performed prior to start of measurement after the power is turned ON.

Memo• Own calibration data has been registered to the White Calibration Plate supplied with the instrument.• Although white calibration data is retained in internal memory even after power is turned off, it is still necessary

to repeat white calibration each time you switch the power back on.• The reading may fluctuate slightly due to changes in the ambient temperature or due to heat generation caused by

repeated operation of the instrument. In this case, make sure to perform white calibration regularly.• White calibration is not required if you just want to see the measured data without performing measurement. In

such a case, select “SKIP”, and you can immediately go to the sample data screen.

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

N o . 1 2 3 4 5 6 7

Note• Once you skip white calibration and go to the sample data screen, you cannot perform white calibration unless

you turn OFF the power and turn it ON again. After turning OFF the power, wait several seconds before turn-ing it ON again.

• White calibration must be performed at the same temperature as the one at which measurement will be per-formed.

• Before starting white calibration, make sure that the instrument is sufficiently acclimatized to the ambient tem-perature.

[Setting Procedure]Start the procedure from the <CALIBRATION> screen.

1 When the instrument is turned ON, the <CALIBRATION> screen will appear auto-matically.

2 Turn B to select “WHITE”.

N o . 1 2 3 4 5 6 7

P R E S S

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

UB T T O NS .E AM

Page 42: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

40

3 Place the instrument on the correct White Calibration Plate (i.e. the calibration plate whose plate No. is displayed in the screen).

4 Make sure that is displayed, then press C (MEAS. button). Screen when white calibration is selected

N o . 1 2 3 4 5 6 7

P R E S S

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

UB T T O NS .E AM

White calibration will be performed. When white calibration is complete, the <SAMPLE> screen will reap-pear.

Screen during white calibration

331 2C A L I B R A T I N G

S K I P

W H I T E

Z E R O

T I O NI B R AC A L

N o . 1 2 3 4 5 6 7

NoteThe instrument's memory contains the calibration data for the White Calibration Plate supplied with the instrument. Thus, the White Calibration Plate bearing the same serial number as that displayed in the <CALIBRATION> screen must be used for white calibration.

Memo• Although it is not necessary to perform white calibration each time the power is turned ON, to obtain more

accurate data, it is recommended to perform white calibration immediately before measurement. • If does not appear due to the auto power save function, it may take some time to start calibration after C is pressed.

Page 43: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

41

Setting a Color Difference Target Data

Setting a Color Difference Target DataTo measure the color difference between two specimens, the color of one of the specimens must be set as target data. Sample data are linked to this target data and saved. Consequently, you need to select one target color as the target data to be linked before starting color difference measurement. This target datum is called the active target. The current active target can be confirmed on the target data screen (See page 51.). Once measurement is performed, you cannot change the target data to which sample data are linked into another target data. (The active target can be changed for each measurement.) Up to 1,700 target data can be set for color difference measurement. To set target data, follow the procedure given below.

Memo• One set of tolerances data can be set for pass/fail judgment with this instrument. (The initial tolerance value is 1.00

for all cases. For the tolerance setting procedure, refer to page 44.)

Note• You cannot change target data to sample data. • When the active target data are set to temporary data (TEMP data), the color difference changes every time the

temporary data are overwritten.• Before setting target data, you need to select measurement environment and perform white calibration. • For accurate measurement, make sure to keep measurement atomosphere such as ambient temperature constant.

Measuring a Color Difference Target DataAfter the calibration is complete, the <SAMPLE> screen (sample data screen) appears automatically. When the <TARGET> screen (target data screen) already appears, start from step 2.

MemoAs long as the target data screen has been displayed, the data displayed on the screen does not affect the procedure. The target data cannot be overwritten. It is always registered as new data.

[Setting Procedure]Start the procedure from the <TARGET> screen.

1 Turn B to select “TGT”, then press A.

T GT G T

2 Place the specimen measuring port to the specimen.

Page 44: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

42

Setting a Color Difference Target Data

3 Make sure that is displayed, then press C (MEAS. button). The lamp will flash, the measure-ment will be taken, and the result will appear on the LCD.

Memo• Even temporary data (TEMP data) can be active target data. • If does not appear due to the auto power save function, it may take some time to start measurement after C is pressed.

4 To set the next target data, repeat steps 2 and 3.

Memo• Measurement will no longer be possible once the total number of target data and sample data reaches

1,700. In this case, delete some target data or sample data. • Measurement may not be possible even when the total number of target data and sample data is less than

1,700. For details, refer to “Deleting Data” on page 58.

Selecting a Color Difference Target DataTo measure the color difference between two specimens, the target data to be used for measurement must be selected from those set in the procedure “Setting a Color Difference Target Data”. To select target data, follow the procedure given below.

[Setting Procedure]Start the procedure from the <TARGET> screen.

1 Turn B to select a target data number, then press A. D will appear to allow you to change the number.

2 Turn B to select the desired target data number, then press A. The setting is determined and the <TARGET> screen will reappear. Rotating the navigation wheel changes the display and allows you to select the desired data number by browsing the list of data.

NoteTo make pass/fail judgment during measurement, it is necessary to set tolerances for the target data to be used. For details, refer to “Setting Color Difference Tolerances” on page 44.

Page 45: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

43

Setting a Color Difference Target Data

Deleting a Color Difference Target DataFor the instructions on how to delete target data, refer to “Deleting Data” on page 58. You can do it by deleting the target data one by one, or by deleting all the target data set in the instrument at once.

Note• When target data is deleted, the tolerance setting will not be deleted. • You will not be able to view the color difference for the sample data that used the deleted target data.

Page 46: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

44

Setting Color Difference Tolerances

Setting Color Difference TolerancesThis instrument allows you to set tolerances to the color difference target data for the sample data, in or-der to make pass/fail judgment.Only one set of tolerances can be set with this instrument. The tolerances can be set within the range of 0.01 to 19.99 with increments of 0.01.

NoteNote that changing the tolerance setting will automatically affect the results of all judgment. For example, the sample data which has been judged as passed is now judged as failed, and vice versa.

[Setting Procedure]Make sure that power to the instrument is turned OFF.

1 While pressing B to left, turn the power ON. The <ENVIRONMENT> screen will appear. (Back side)

IVNE MNOR TNE

OT

C

AREL

I O N

ECN

N D I TO

F I N I S H

2 Turn B to select “TOLERANCE”, then press A.

IVNE MNOR TNE

OT

C

AREL

I O N

ECN

N D I T

F I N I S H

O

3 Turn B to select “SCI” or “SCE”, then press A.

IVNE NOR

OT

C

AREL

I

N

N D I TO

F I N I S H

S C I

S C E

Page 47: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

45

Setting Color Difference Tolerances

4 Turn B to select the color space and other item ( L*, E*, etc.) used as judgment criteria, then press A. E will appear allowing you to change the value. S C I

L * = 1 9 . 9 9��

a * =�

b * =

E * = + 1 . 0 0

1 . 5 7

0 . 0 1

±±±

IVNE NOR

OT

C

AREL

I

N

N D I TO

F I N I S H

5 Turn B to change the value, then press A. The selected value is set.

��

S C I

L * =�

a * =�

b * =

E * =

±+

±±

1 9 . 9 9

1 9 . 9 9

1 . 5 7

0 . 0 1

IVNE NOR

OT

C

AREL

I

N

N D I TO

F I N I S H

• Keeping B held down to right or left will cause the value change continuously.

6 Turn B to select “ ”, then press A. The setting is determined and the previous pop-up screen will reappear.

Selecting “ ” and pressing A will cancel the setting values and the previous pop-up screen will reappear.

Page 48: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

46

Setting Color Difference Tolerances

Page 49: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

47

Chapter 3

Measurement

Page 50: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

48

MeasurementAfter the calibration is complete, the <SAMPLE> screen (sample data screen) appears automatically. To perform sample data measurement, follow the procedure given below.

Note• Prior to start of measurement, be sure to perform white calibration. For details, refer to “White Calibration”

on page 39. • To measure color difference, it is necessary to set target data first and then select active target data. (The sam-

pled data are linked to the target data specified as active target data.) • The target data linked to the sample data cannot be changed. • Sample data cannot be used as target data. • For accurate measurement, make sure to keep measurement conditions such as ambient temperature constant.

[Setting Procedure]Start the procedure from the <SAMPLE> screen.

1 Make sure that the <SAMPLE> screen is displayed.

• As long as the sample data screen has been displayed, the data displayed on the screen do not affect the procedure.

• The sample data are not overwritten. They are always registered as new data.

2 Place the specimen measuring port to the specimen.

• If necessary, slide the viewfinder lever to check the position of the specimen.

3 Make sure that is displayed, then press C (MEAS. button). The lamp will flash, measurement will be taken, then the result will ap-pear on the LCD.

Page 51: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

49

Measurement

Memo• When active target data have been set, the target data No. (T****) of the target data linked to the sample

data is displayed on the sample data screen. When no target data are linked, “ABS” will be displayed.• If does not appear due to the auto power save function, it may take some time to start measurement after C is pressed.

4 To set the next target data, repeat steps 2 and 3.

Memo• Measurement will no longer be possible once the total number of target data and sample data reaches 1,700. In

this case, delete some target data or sample data. • Measurement may not be possible even when the total number of target data and sample data is less than 1,700.

For details, refer to “Deleting Data” on page 58.

Selecting ScreensThe sample data and measurement result can be displayed on 4 types of screens: Sample data screen (color difference value or absolute value), target data screen (absolute value), sample data screen (spec-tral reflectance), and target data screen (spectral reflectance). These screens can be switched in the pro-cedure given below:

• On the <SAMPLE> screen (sample data screen), turn B to select “TGT”, then press A to change to the <TARGET> screen (target data screen).

MemoThe most recently displayed datum is displayed. If TEMP was displayed most recently, TEMP is displayed. If No. 1 was displayed, No. 1 is displayed.

• On the <TARGET> screen (target data screen), turn B to select “SPL”, then press A to change to the <SAMPLE> screen (sample data screen).

MemoThe displayed target data screen always shows the active target data. (It is not the target data linked to the sample data.)

Sample data screen (color difference value or absolute value)

Target data screen (absolute value) Target data screen (spectral reflectance)

Sample data screen (spectral reflectance)

• On either of the <SAMPLE> screen (sample data screen) or <TARGET> screen (target data screen), when you turn B to select “MENU” and press A, then turn B to select “DISPLAY” and press A, the color difference/absolute value display and the spectral reflectance display are switched. For details, “Switching the Display Types” on page 56.

Page 52: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

50

Displaying the Measurement ResultsThis section describes the typical measurement result screens.

Sample data screen (color difference value or absolute value)After measurement is performed, the LCD displays the measurement results according to the preset measurement conditions.

NoteThe absolute value data and color difference value are displayed. (The display depends on the display type. When target data are set, a color difference value is displayed. When target data are not set, an absolute value is displayed.)

1 Indicates that this is a sample data screen.2 Result of pass/fail judgment3 Ready to measure4 Sample data5 Battery mark (Indicates the remaining battery power. When appears, replace the

batteries with new ones.) 6 Linked target data No.7 Specular component of the displayed data8 Relative gloss (luster) value9 Information of the currently selected observer/illuminant0 Date/time when the displayed data was measuredA Sample data No.

For details, refer to “Sample Data No. and Target Data No.” on page 54. B Changes to the target data screen.

For details, refer to “Selecting Screens” on page 49. C Menu

Provides the following operations: • Setting the Illuminant (page 33)• Printing displayed data (page 67)• Saving temporary data (page 55)• Switching the display types (page 56)• Switching between SCI/SCE (page 57)• Deleting data one by one (page 58)• Deleting all data at once (page 59)

MemoAs for A, B and C, turn B to select a desired option, then press A to determine the option or to go to the corresponding screen.

Page 53: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

51

Displaying the Measurement Results

Sample data screen (spectral reflectance)

1 to C are same as the case of “Sample data screen (color difference value or absolute val-ue)”. D Wavelength of the spectral reflectance dataE Spectral reflectance of the selected wavelength

MemoAs for A, B, C and D, turn B to select a desired option, then press A to determine the option or to go to the corresponding screen.

Target data screen (absolute value)After target data measurement, the LCD shows the measurement results.

NoteOnly absolute value data are displayed.

1 Indicates that this is a target data screen.2 Ready to measure3 Target data4 Battery mark (Indicates the remaining battery power. When appears, replace the

batteries with new ones.) 5 Active target data No.6 Specular component of the displayed data7 Relative gloss (luster) value8 Information of the currently selected observer/illuminant9 Date/time when the displayed data was measured

Page 54: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

52

Displaying the Measurement Results

0 Target data No. For details, refer to “Sample Data No. and Target Data No.” on page 54.

A Changes to the sample data screen. For details, refer to “Selecting Screens” on page 49.

B Menu Provides the following operations: • Setting the Illuminant (page 33)• Printing displayed data (page 67)• Saving temporary data (page 55)• Switching the display types (page 56)• Switching between SCI/SCE (page 57)• Deleting data one by one (page 58)• Deleting all data at once (page 59)

MemoAs for 0, A and B, turn B to select a desired option, then press A to determine the option or to go to the corresponding screen.

Target data screen (spectral reflectance)

1 to B are same as the case of “Target data screen (absolute value)”.C Wavelength of the spectral reflectance dataD Spectral reflectance of the selected wavelength

MemoAs for 0, A, B and C, turn B to select a desired option, then press A to determine it or to go to the cor-responding screen.

Page 55: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

53

Displaying the Measurement Results

Abbreviations on LCD DisplayTo facilitate the user’s understanding of various types of information displayed on the instrument’s LCD, the following abbreviations are commonly used.

Screen/Item Notation Meaning

<TARGET> screen<SAMPLE> screen

Data number

TEMP Temporary data

G*** 8-degree gloss value

<SAMPLE> screen NONE No data stored

<TARGET> screen ABS Absolute data

<TARGET> screen<SAMPLE> screen T Target data

<TARGET> screen ACT T Linked target data

<SAMPLE> screen Short cut to target data screen TGT Target data

<TARGET> screen Short cut to sample data screen SPL Sample data

<SAMPLE> screen (Color difference & Absolute value & Pass/fail screen)

Color space, colorimetric data

CMC1 CMC1:1

CMC2 CMC2:1

E94 E for CIE94

E00 E for CIE00

MI(**) Metamerism index (** is the light source of the data which is not displayed.)

Page 56: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

54

Displaying the Measurement Results

Sample Data No. and Target Data No.Sample data numbers and target data numbers can be changed in the procedure given below:

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select a sample data number or a target data number, then press A.D will appear to allow you to change the number.

2 Turn B to select the desired sample data number or a target data number, then press A.

Wavelength of Spectral Reflectance DataOn the sample data screen (spectral reflectance) and target data screen (spectral reflectance), the spectral reflectance and wavelength can be changed. Perform the procedure given below:

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select a wavelength, then press A.D will appear to allow you to change it.

2 Turn B to select the desired wavelength, then press A.

Page 57: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

55

Displaying the Measurement Results

Saving Temporary DataYou can save the data displayed on the target data screen or sample data screen. This saving option is applicable only for temporary data. When the auto save (page 28) is set to ON, no temporary data exists. Consequently, this saving option is invalid and cannot be performed. Also when the data has already been saved, this menu option is grayed out and invalid.

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select “MENU”, then press A.

2 Turn B to select “SAVE”, then press A. “Do you want to save this data?” is displayed.

3 Turn B to select “YES”, then press A.

The status of the data changes from temporary saved to saved.

When the data are temporarily saved

When the data are saved

Page 58: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

56

Displaying the Measurement Results

Switching the Display TypesOn either of the sample data screen or target data screen, the displays of color difference value/absolute value and spectral reflectance can be switched.

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select “MENU”, then press A.

2 Turn B to select “DISPLAY”, then press A.

The displays of color difference val-ue/absolute value and spectral re-flectance are switched.

Color difference value/absolute value display

Spectral reflectance display

Page 59: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

57

Displaying the Measurement Results

Switching between SCI/SCEOn either of the sample data screen or target data screen, you can switch between “SCI” and “SCE” to display the data on the screen.

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select “MENU”, then press A.

2 Turn B to select “SCI” or “SCE”, then press A.

The display changes between “SCI” and “SCE”.

SCI display

SCE display

Page 60: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

58

Displaying the Measurement Results

Deleting DataThere are two methods of deleting sample data or target data. One is to delete the data items displayed on the LCD one by one. The other is to delete all the data items at once.

Deleting Data One by One

On either of the sample data screen or target data screen, data currently displayed can be deleted.

MemoAfter sample data or target data was deleted, the sample data screen or target data screen displays data which has the number smaller than that of the deleted data by one. For example, when data No. 50 is deleted, the sample data screen or target data screen displays data No. 49. If data No. 49 does not exist, data No. 48 is displayed instead.

Note• Once data are deleted, they cannot be recovered. • When either sample data or target data are deleted, the data Nos. do not change.

For example, when three data, No. 1, No. 2 and No. 3, exist and data No. 2 is deleted, data No. 1 remains as No. 1, and data No. 3 remains as No. 3. No. 2 has no data and regarded as an unassigned number. During the next measurement, the sample data are assigned to No. 2. (The data are not assigned to No. 4. Data are always registered with the smallest available data No.)

• You can store only sample data to an unassigned number resulting from the deletion of sample data. Similarly, you can store only target data to an unassigned number resulting from the deletion of target data. Consequent-ly, when 1,700 data, the maximum, are saved, and No. 2 becomes unassigned because target data No. 2 was deleted, only target data measurement is possible and sample data measurement is impossible.

• When target data are deleted, the sample data linked to the target data will become absolute value data. (The link to the target data is deleted.) As a result, its color difference is no longer displayed. The target data linked to sample data will be changed only when the target data are deleted. In such a case, sample will automatically become absolute value data. In other cases, target data linked to sample data cannot be changed.

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select “MENU”, then press A.

2 Turn B to select “DELETE”, then press A. “Are you sure to delete this data?” is displayed.

Page 61: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

59

Displaying the Measurement Results

3 Turn B to select “YES”, then press A.

The data are deleted and preceding data are displayed.

Deleting All Data at Once

You can delete all the data at once on either the sample data screen or target data screen.

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1 Turn B to select “MENU”, then press A.

2 Turn B to select “ALL DELETE”, then press A. “Are you sure to delete All data?” is displayed.

3 Turn B to select “YES”, then press A.

All the data are deleted and a screen without data will appear.

Page 62: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

60

Displaying the Measurement Results

Page 63: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

61

Chapter 4

Other Functions

Page 64: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

62

Connecting to an External DeviceConnecting a PC or printer to the external output terminal on the instrument allows the transfer of data between the instrument and PC or printing of the measurement results.

Connecting a Personal ComputerConnecting a PC (PC-AT compatible) to the instrument with a RS-232C cable will allow uploading the data stored in the instrument’s memory to the PC or downloading data from the PC to the instrument. When connecting this instrument to a PC, use of the optional Color Data Software “SpectraMagic NX CM-S100w” is recommended.

MemoWhen the instrument is connected to a PC, make sure that the instrument is set in remote mode. Remote mode allows the following operations: 1. Outputting sample data and target data to the PC 2. Downloading target data from the PC For details, refer to “Switching to Remote Mode” on page 63.

Note• TEMP data (of both target and sample data) will not be output to a PC.

Consequently, when sample data linked to TEMP data of target data is output to a PC, the sample data turns into absolute value data. To use a PC to check the color difference between the absolute-value sample data and the TEMP data of target data, follow the procedure below:1)Save the TEMP data of target data on the CM-2300d (For details, refer to “Saving Temporary Data” on

page 55.), and output the saved target data to the PC.2)Use the software to link the target data output to the PC and the sample data (absolute value data) output

earlier. Then check the color difference.• The CM-2300d saves both SCI and SCE data for simultaneous SCI/SCE measurement.

When using SpectraMagic NX (CM-S100w) or other PC software to download target data to the CM-2300d, you can download the SCI or SCE data together or separately. When downloading only one type of data, dum-my data is automatically created for the data that is not downloaded. The dummy reflectance data is set to 0.01% over the entire wavelength range. When dummy data is used, the chromaticity, color difference and relative gloss values are also calculated and displayed using the dummy data.

Note• When connecting, make sure that the connectors are correctly oriented and fastened securely with screws. • Before connecting, make sure that the power to both the instrument and PC is turned OFF. • When disconnecting the cable, be sure to grip the plug or connector and pull it. Do not pull on or forcibly bend the cable.• Do not touch the connector terminals with your hand. Keep the terminals clean and do not apply excessive force to them.• Make sure that the cable has sufficient length. Tensioning the cable may cause connection failure or wire breakage.• When using a cable other than the one supplied as the standard accessory (I F-A16), make sure that the cable

conforms to “RS-232C Pin Assignment and Cable Wiring Diagram” on page 63. If it does not conform, a mal-function may occur or incorrect data may be transferred between the instrument and PC.

Communication ParametersItem Setting

Baud rate 9600 bpsCharacter bits 8 bitsParity NoneStop bits 1 bit

Page 65: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

63

Connecting to an External Device

RS-232C Pin Assignment and Cable Wiring Diagram

23578

Pin No.RXDTXDGNDRTSCTS

Signal

Instrument

5

9 7 68

4 3 2 1

Personal Computer(For D-sub, 9-pin)

RXDTXDGNDRTSCTS

Signal23578

Pin No.

1 8

Switching to Remote Mode

To transfer data between the instrument and the PC, the instrument must be switched to “REMOTE MODE”. To switch to remote mode, follow the procedure given below.

[Setting Procedure]Make sure that power to the instrument is turned OFF.

1. Connect the instrument to the PC with the RS-232C cable.

2. While pressing A, turn the power ON. The <SETTING> screen will appear.

(Back side)

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

Page 66: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

64

Connecting to an External Device

3. Turn B to select “MODE”, then press A.

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

4. Turn B to select “REMOTE”, then press A. The setting is determined and the remote screen will appear.

R S 2 3 2 C

E N D : < E N T E R > K E Y

R E M O T E

Outputting to a PrinterConnecting the instrument to a printer with a printer cable allows printing the sample data or the color difference target data stored in the instrument’s memory. Two printing methods are available. One prints automatically each time measurement is taken (this is called “Auto Print”; both sample and target data.), and the other prints the currently displayed sample data and color difference target data.

MemoOn the <SAMPLE> screen, the spectral reflectance data of the displayed sample data will be printed. On the <TAR-GET> screen, the spectral reflectance data of the displayed color difference target data will be printed. If “ABS. & DIFF.” is selected, the color difference and absolute data will be printed.

Note• When connecting, make sure that the connectors are correctly oriented and fastened securely with screws. • Before connecting, make sure that the power to both the instrument and printer is turned OFF. • When disconnecting the cable, be sure to grip the plug or connector and pull it. Do not pull on or forcibly bend

the cable. • Do not touch the connector terminals with your hand. Keep the terminals clean and do not apply excessive force

to them. • Make sure that the cable has sufficient length. Tensioning the cable may cause contact failure or wire break-

age. • The instrument and printer must be connected using a printer cable that is wired as shown in “Printer Cable

Pin Assignment and Wiring Diagram”.

Communication ParametersItem Setting

Baud rate 9600 bpsCharacter bits 8 bitsParity NoneStop bits 1 bitFlow control RTS/CTS

Page 67: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

65

Connecting to an External Device

Applicable Printer: DPU-H245AS-A03 (Seiko Instruments Inc.) * Serial connection: Metric screw type

In addition to the above printer, printers that meet the following specifications can be used.Item Specifications

Number of print lines 27 or more lines, Data input: RS-232CData control BUSYBaud rate 9600Character bits 8 bitsParity NoneStop bits 1 bitBasic function CR (0DH) carriage return

Printer Cable Pin Assignment and Wiring Diagram• For D-sub, 9-pin connector

5

9 7 68

4 3 2 1

1 8

358

TXDGNDCTS

Pin No. Signal

Printer(For D-sub, 9-pin)

Instrument

DATAGNDBUSY

358

Signal Pin No.

• For D-sub, 25-pin connector

1 8

358

TXDGNDCTS

Pin No. Signal

Printer(For D-sub, 25-pin)

Instrument

DATABUSYGND

257

Signal Pin No.

Page 68: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

66

Connecting to an External Device

Auto Print

If the mode has been set to “PRINT” beforehand, the sample data can be printed as each measurement is made.

NoteThe instrument and printer must be connected using a printer cable that is wired as shown in “Printer Cable Pin As-signment and Wiring Diagram” on page 65.

MemoWhen the printer’s connector is a D-sub 9-pin type, an optional printer cable CR-A75 can be used. * Serial connection of printer cable CR-A75: Metric screw type

[Setting Procedure]Make sure that power to the instrument is turned OFF.

1. Connect the instrument to the printer with a printer cable.

2. While pressing A, turn the power ON. The <SETTING> screen will appear.

(Back side)

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

3. Turn B to select “MODE”, then press A.

S E T T I N G

EGA GNEUGNA

M

L

EMIT

O

HSIL

51:3180/90

N O R M A LED

4. Turn B to select “PRINT”, then press A.The setting is determined and the <SET-TING> screen will reappear. R I N TP

R E M O T E

S E T T I N G

EGAUGNA

M

L

EMIT

O ED

After the setting is complete, sample data will be printed automatically each time measure-ment is taken.

Page 69: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

67

Connecting to an External Device

Printing Displayed DataThe data displayed on either of the sample data screen or target data screen is printed in a specified format. Printing is performed even when the instrument is not connected to the dedicated printer. As a result, it is assumed that printing has been complete even if printing was not performed in reality.

MemoTo print sample data, the instrument and printer must be connected using a printer cable beforehand. When the printer's connector is a D-sub 9-pin type, an optional printer cable CR-A75 can be used.

Note• The instrument and printer must be connected using a printer cable that is wired as shown in “Printer Cable Pin

Assignment and Wiring Diagram” on page 65. • Before connecting, make sure that the power to both the instrument and printer is turned OFF.

[Setting Procedure]Start the procedure from the <SAMPLE> or <TARGET> screen.

1. Turn B to select “MENU”, then press A.

2. Turn B to select “PRINT”, then press A. “Do you want to print this data?” is dis-played.

3. Turn B to select “YES”, then press A.

The printer outputs the data according to a specified format.

Page 70: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

68

Connecting to an External Device

• Print ExamplePrint Example 1

Type of screen: “Sample data screen (color difference value and absolute val-ue)”

Display mode: “ABS&DIFF”

SAMPLE No.0001 T0001

25-Oct-2005 13:09:45

FAIL GLOSS=0

SCI 2/D65

dE*= 13.27

L*= 84.99 dL*= 5.87

a*= 0.16 da*= 0.82

b*= -4.61 db*= -11.87

Print Example 2

Type of screen: “Sample data screen (spectral reflectance)”

SAMPLE No.0001 T0001 25-Oct-2005 13:09:45 FAIL GLOSS=0 SCI 360nm 30.18 550nm 64.89 370nm 33.98 560nm 66.31 380nm 37.85 570nm 66.18 390nm 43.06 580nm 64.65 400nm 52.86 590nm 65.14 410nm 64.38 600nm 63.80 420nm 70.96 610nm 62.89 430nm 72.43 620nm 64.06 440nm 72.97 630nm 65.17 450nm 72.99 640nm 65.72 460nm 72.07 650nm 64.08 470nm 70.38 660nm 64.11 480nm 69.56 670nm 62.98 490nm 69.33 680nm 61.19 500nm 69.09 690nm 61.32 510nm 70.00 700nm 62.49 520nm 67.01 710nm 60.90 530nm 65.42 720nm 58.40 540nm 65.90 730nm 58.62 740nm 57.59

Page 71: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

69

Chapter 5

Troubleshooting

Page 72: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

70

Error MessagesThe following messages may appear while you are using the instrument. If such messages appear, take the necessary actions shown in the table below. If the trouble does not go away in spite of taking action, contact a KONICA MINOLTA SENSING-authorized service facility.

MemoMessages that may be displayed on the LCD are given in the table below. For communication error check codes, refer to the separate document.

MessageCorrective Action Symptom/Possible Cause Corrective ActionOK02LOW ILLUMINATION

Amount of light from the xenon lamp has dropped to 65% of its initial level.• Deterioration of xenon lamp• Dirt on integrating sphere

Clean the integrating sphere as explained in “Cleaning Each Part” on page 15. If the problem still remains, contact a KONICA MINOLTA SENSING-author-ized service facility.

OK10MAKE WHITE CALIBRATION !

The reliability of the white cali-bration data has dropped because a certain period of time has passed since the last white cali-bration.

Performing white calibration is recommended.

ER01MEMORY FULL

The instrument can store up to a total of 1,700 measurement and color difference target data.

Delete some data for securing nec-essary memory space. This instru-ment can retain up to 1,700 data sets (counting both measurement results and color difference target data).

ER02LOW BATTERY

(When the instrument is powered by batteries)• Battery voltage drop

Replace the batteries with new ones.

ER05ERROR IN FLASHING

Xenon lamp does not flash prop-erly.• Xenon lamp has blown.• Faulty flash circuit• Faulty sensor

Contact a KONICA MINOLTA SENSING-authorized service fa-cility.

ER07ZERO CALIBRATION INCOMPLETED

Zero calibration has not been performed.

Perform zero calibration.

ER11FAILED IN CALIBRATION

Zero calibration or white calibra-tion has not been performed cor-rectly.

For zero calibration, the speci-men measuring port must be di-rected into the air. For white calibration, the White Calibra-tion Plate must be used.

The Target Mask is not correct. Attach the correct Target Mask.ER13ERROR IN A/D

Failed during A/D conversion.• Faulty A/D converter• Break down of circuits relating

to the A/D converter

Turn the power OFF, and then turn it ON again. If this message keeps appearing, contact a KONICA MINOLTA SENSING -authorized service facility.

Page 73: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

MessageCorrective Action Symptom/Possible Cause Corrective ActionER17INCORRECT CLOCK OPERATION

Clock IC is not working correct-ly.• Clock IF’s power voltage drop

or breakdown of circuits relat-ing to the clock IC.

Contact a KONICA MINOLTA SENSING-authorized service fa-cility.

ER24WHITE CALIBRATION INCOMPLETED

White calibration has not been performed.

White calibration must be per-formed each time the instrument is turned ON. White calibration must also be performed when the measurement area or specular component setting is changed.

ER27ERROR IN CHARGING

Charging for flashing of Xenon lamp can’t be completed.• Breakdown of charging circuit• Battery voltage drop

Contact a KONICA MINOLTA SENSING-authorized service fa-cility.

ER30NO DATA IN MEMORY

Data has been lost since the memory’s backup battery is ex-hausted.

Turn the power ON to charge the memory’s backup battery. The backup batteries can be fully charged in 25 hours when the power of this instrument is turned ON. After the batteries are fully charged, data backup can be stored for approximate 4.5 months.

71

Error Messages

Page 74: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

72

TroubleshootingIf an abnormality has occurred with the instrument, take necessary actions as given in the table below. If the instrument still does not work properly, turn the power OFF, and then turn it ON again. If the symp-tom remains, contact a KONICA MINOLTA SENSING-authorized service facility.

Symptom Check Point ActionLCD is blank. Are batteries inserted? Is the AC adapter

connected?Insert the batteries. Connect the AC adapter.

Is the LCD contrast appropriate? Ajust the LCD contrast. (page 30)Isn’t displayed? Or isn’t the batter-ies extremely low?

Replace the batteries with new ones.

“ ” does not appear. Are batteries inserted? Is the AC adapter connected?

Insert the batteries. Connect the AC adapter.

Isn’t displayed? Or aren’t the bat-teries extremely low?

Replace the batteries with new ones.

MEAS. button is not effective.

Is measurement still in progress? Wait until measurement is complete, and then press the button.

Is a measurement screen displayed? MEAS. button must be pressed while a measurement screen is displayed.

Abnormal measure-ment result

Is the instrument directed perpendicular to the specimen?

Make sure that the instrument is directed perpendicular to the specimen, to prevent leakage of light.

Is the White Calibration Plate correct? Use the correct White Calibration Plate to perform white calibration. (page 39)Was white calibration performed correct-

ly?Was zero calibration performed correct-ly?

Direct the specimen measuring port into the air or use the optional Zero Calibra-tion Box and perform zero calibration.

Is the target data dummy data (all wave-lengths 0.01%)?When using SpectraMagic NX (CM-S100w) or other PC software to download target data to the CM-2300d and down-loading either SCI or SCE data, dummy data is automatically created for the data that is not downloaded. (page 62)

Select the appropriate specular compo-nent mode “SCI” or “SCE” to match the actual target data. (page 57)To perform measurements for both SCI and SCE, download both SCI and SCE data at the same time when setting target data from PC software.

Measurement results fluctuate.

Is instrument kept stationary during measurement?

Do not allow the instrument to move dur-ing measurement.

Not possible to input data to the computer. No commands from computer are accepted.Commands cannot be accepted correctly.

Is the RS-232C cable connected correct-ly?

Connect the instrument’s external output terminal to the computer’s RS-232C con-nector with the RS-232C cable supplied with the instrument.

Is the RS-232C cable the one supplied with the instrument used?Do communications settings on the in-strument match those on the computer?

Set the same communications settings as those on the instrument to the computer.

Not possible to print. Is MODE set to PRINT? Set MODE to PRINT. (page 27)Measurement data or settings are not held in memory, and dis-appear immediately.

Is AUTO SAVE set to OFF? When AUTO SAVE is set to OFF, new sample data and their measurement set-tings are saved as temporary data. To save these data and settings, save the temporary data. (page 28, 55)

Page 75: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

73

Chapter 6

Appendix

Page 76: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

74

Principles of Measurement

Illuminating/Viewing SystemThe flow of measurement is shown below.

This instrument utilizes the D/8 geometry conforming to CIE No.15, ISO 7724/1, ASTM E1164, DIN 5033 Tei17, and JIS Z8722-1982 (diffused illumination/8º viewing system) standards, and offers simul-taneous SCI (specular component included) and SCE (specular component excluded) measurements.

1 IlluminationLight from xenon lamps diffuses in the in-tegrating sphere and il-luminates the specimen uniformly.

2 Receivinga: Light reflected by the

specimen is received.b:Light diffused in the

integrating sphere is received.

3 SensingLight is transmitted to the specimen-measur-ing and illumination-monitoring optical sys-tem, where the light in the wavelength range of 360 to 740 nm is di-vided into 10 nm-pitch components and signal proportional to the light intensity of each com-ponent are output to the analog processing cir-cuit.

1 Light from xenon lamps diffuses on the inner surface of the integrating sphere and illuminates the specimen uniformly.

2 a. The light reflected by the specimen surface at an angle of 8 degrees to the normal of the surface is received by the specimen-measuring optical system.

b.The diffused light in the integrating sphere is received by the illumination-monitoring optical sys-tem and guided to the sensor.

3 The light reflected by the specimen surface and the diffused light are divided into each wavelength component by the specimen-measuring optical system and illumination-monitoring optical sensor, respectively, and then signal proportional to the light intensity of each component are output to the analog processing circuit.

By using the outputs from the specimen-measuring optical system and the illumination-monitoring sen-sor for calculation, compensation for slight fluctuation in spectral characteristic and the intensity of the illumination light is performed. (Double-beam system)

Page 77: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

75

Principles of Measurement

Simultaneous SCI/SCE MeasurementThe CM-2300d employs simultaneous SCI/SCE measurement method.With the conventional models, SCI (Specular Component Included) and SCE (Specular Component Ex-cluded) need to be switched mechanically by opening/closing the optical trap provided inside the inte-grating sphere.This means mechanical switching has to be made each time the mode is switched from SCI to SCE or vice versa, and it is not possible to start measurement in the new mode until switching is complete.With the CM-2300d, both SCI and SCE data can be obtained simultaneously by calculating the data ob-tained by two illuminantions.

Simultaneous SCI/SCE Measurement MethodTo illuminating/viewing system

specimen

Two light sources 1 and 2 are located as shown in the figure, and light source 1 flashes first.

Light source 1 is the diffused illumination type, and SCI measured data can be obtained as it flashes.

Then, light source 2 flashes next.

Light source 2 is provided for numeric control of the specular component. The data (i.e. amount of light reflected specularly from the surface of the specimen) obtained when this light source flashes and that obtained when light source 1flashed are used to calculate SCE data.

By carrying out the above measurement and calculation during a single measurement, both SCI and SCE data can be obtained simultaneously without mechanical switching that is required in the case of the con-ventional models.

Page 78: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

76

Specifications

Specifications

Model CM-2300dIlluminating/viewing system d/8 (diffuse illumination, 8-degree viewing angle), equipped with

simultaneous measurement (not switched mechanically) of SCI (specular component included)/SCE (specular component exclud-ed). (Conforms to CIE No.15, ISO 7724/1, ASTM E1164, DIN 5033 Teil7, JIS Z8722 Condition c, standard.)

Size of integrating sphere ø 52 mmDetector Silicon photodiode array (dual 40 elements)Spectral separation device Diffraction gratingWavelength range 360 nm to 740 nmWavelength pitch 10 nmHalf bandwidth Approx. 10 nmReflectance range 0 to 175%, Display resolution: 0.01%Light sources 2 pulsed xenon lampsMeasurement time Approx. 1.5 secondsMinimum measurement interval 3 seconds (at 23°C)Battery performance Approx. 1000 times at 10-second intervals (when alkaline batteries

used) (at 23°C)Measurement/illumination area ø 8 mm/ø 11 mmRepeatability Spectral Reflectance: Standard deviation within 0.2% (360 to 380

nm: Standard deviation within 0.4%)Chromaticity Value: Standard deviation E*ab within 0.08 (When a White Calibration Plate is measured 30 times at 10-sec-ond intervals after white calibration)

Inter instrument agreement E*ab within 0.4 (SCI) (Based on the average of 12 BCRA Series II color tiles, compared to values measured with master body at 23°C)

Language modes English/ChineseMeasurement mode Single/averaging (auto mode: 1 to 8 times/manual mode)Interface Conforms to RS-232CObserver 2/10 degreesObservation illuminant A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (simultaneous evalu-

ation is possible using two illuminant)

Page 79: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

Model CM-2300dDisplay Spectral value/graph, colorimetric value, color difference value,

PASS/FAIL result, relative gloss valueColor space/colorimetric data L*a*b*, L*C*h, Yxy, XYZ, MI, E*ab, CMC (1:1), CMC (2:1),

CIE94, CIE DE2000, WI (ASTM E313), YI (ASTM D1925)Storable data sets 1,700 (each SCI/SCE pair constitutes a single data set)Tolerance judgment Tolerance for color difference (1 set of tolerances can be set.)Power AA-size battery (×4), AC adapterSize 69 (W) × 96 (H) × 193 (D) mmWeight approx. 670 g (including “Measuring Base”/excluding batteries) Operating temperature/ humidity range

5°C to 40°C, relative humidity 80% or less (at 35°C) with no con-densation

Storage temperature/humidity range

0°C to 45°C, relative humidity 80% or less (at 35°C) with no con-densation

Standard accessories White Calibration Plate: CM-A145Target Mask ø 8 mm: CM-A146RS-232C cable (D-sub 9-pin, for IBM, PC/AT): IF-A16AC adapter: AC-A17AA (R6) alkaline battery × 4

Optional accessories Zero Calibration Box: CM-A32Hard Case: CM-A148Dust Cover Set: CM-A149Dust Cover: CM-A152Color Data Software “SpectraMagic NX”: CM-S100wPrinter cable (D-sub 9-pin, for serial connection): CR-A75

77

Page 80: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

78

Dimensions

Dimensions

193

96 9669

(mm)

Page 81: Instruction Manual · 2017-08-29 · Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incor-rect use of the instrument

© 2005 KONICA MINOLTA SENSING, INC. AGXbGA(2) Printed in Japan9222-1856-25