1 2 3 ECHELON FLEX™ ENDOPATH® Stapler In-service Best Practices: Troubleshooting the Device Will Not Open If the surgeon needs to manually return the knife and/or release the anvil, the following steps may be followed: 1. Ensure the firing trigger is fully open. 3. Fire one stroke fully plastic to plastic with the firing trigger to return the knife to the home position. Note: This position can now be confirmed using the stroke count indicator (at “0”). 4. Ensure the firing trigger is fully open. 5. Press the anvil release button and open the jaws of the device. 2. Move the red knife reverse switch fully down until the knife direction arrow is pointing in the reverse direction (e.g., toward the user). Note: The red knife reverse switch will not go down unless Step 1 has been completed.

Upload: others

Post on 24-Aug-2020

9 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: In-service Best Practices: Troubleshooting the Device Will ... · In-service Best Practices: Troubleshooting the Device Will Not Open If the surgeon needs to manually return the knife

1

2

3

ECHELON FLEX™ ENDOPATH® StaplerIn-service Best Practices: Troubleshooting the Device Will Not Open

If the surgeon needs to manually return the knife and/or release the anvil, the following steps may be followed:

1. Ensure the firing trigger is fully open.

3. Fire one stroke fully plastic to plastic with the firing trigger to return the knife to the home position.

Note: This position can now be confirmed using the stroke count indicator (at “0”).

4. Ensure the firing trigger is fully open.

5. Press the anvil release button and open the jaws of the device.

2. Move the red knife reverse switch fully down until the knife direction arrow is pointing in the reverse direction (e.g., toward the user).

Note: The red knife reverse switch will not go down unless Step 1 has been completed.

Page 2: In-service Best Practices: Troubleshooting the Device Will ... · In-service Best Practices: Troubleshooting the Device Will Not Open If the surgeon needs to manually return the knife

For complete product details, see Instructions For Use.

The third-party trademarks used herein are trademarks of their respective owners.

©2013 Ethicon Endo-Surgery, Inc. All Rights Reserved. 002908-130917

ECHELON FLEX™ ENDOPATH® StaplerIn-service Best Practices: Cartridge Loading

Properly Loaded Cartridge

Result of Improperly Loading the Cartridge

The sled is flush to the back of the cartridge. Proper loading technique prevents accidental sled movement.

“U” shape formed by the inadvertently advanced sled.

Movement may cause staples to deploy and the safety lock-out to engage.

Proper Loading Steps

Improper Loading Examples

Inadvertently Pushing The Sled Forward