Page 1: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

GC Best Practices & Troubleshooting

Page 2: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 2

Troubleshooting Tips

1. Isolate the problem.

(Blank Runs, Inject Un-retained Compound, Know what it is not)

2. Change only one variable at a time.

3. Compare before/after chromatograms.

(Peak shape, response, retention, baseline rise, background, look for trends, etc.)

Page 3: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 3

4. Make sure it makes sense to do what

you’re doing…

Page 4: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 4

5. Be careful of distractions

Page 5: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 5

6. Sometimes just a fresh set of eyes is all that is

needed.

Page 6: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 6

INJECTOR – contamination, flow settings, flow path issues,

overload, valve settings, faulty consumables

COLUMN – contamination, flow settings, flow path issues,

damage (activity & bleed), breakage

DETECTOR – contamination, flow settings, flow path issues,

electronics

What Can Possibly Go Wrong?

Page 7: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 7

Troubleshooting Starts with Isolating the problem

There are 5 basic areas from where the problem arises

FLOW

INJECTOR

COLUMN

DETECTOR

ELECTRONICS

(Temperature)

But of course it can always be some COMBINATION

Knowing what can & can’t cause the symptom is the key

Logical Troubleshooting

Page 8: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Column

FlowController

Regulators

Air

Hyd

rog

en

Ca

rrie

r G

as

Mol-SieveTraps

Fixed

Injection

Port Detector Electrometer

Recorder/Integrator

Restrictors

Cylinders or Generators

Typical Gas Chromatographic System

Selection of

type and

velocity

influences

efficiency and

retention time

Page 8

Page 9: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

van Deemter Curves

10 20 30 40 50 60

0.25

0.50

0.75

1.00

u (cm/sec)

h

He

N2

H2

Page 9

Page 10: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Introduction to Capillary GC

Agilent Restricted

February 11, 2009Page 10

CARRIER GAS

Type Velocity Range (uopt – OPGV)

Nitrogen 8-16

Helium 20-40

Hydrogen 30-55

Page 11: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Gas Clean Filters

Page 11

Page 12: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Column

FlowController

Regulators

Air

Hyd

rogen

Carr

ier

Gas

Mol-Sieve

Traps

Fixed

Injection

Port Detector Electrometer

Recorder/Integrator

Restrictors

Cylinders or Generators

Typical Gas Chromatographic System

Have to be able to

get things to their

gas state

Key to starting the

Chromatographic

process right.

Page 12

Page 13: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Influence of Injection Efficiency

Short

Concentrated

Long

Diffuse

Solute Bands

Same column, same chromatographic conditions

Page 13

Page 14: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Injectors

Split

Splitless

Page 15: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Septumpurge

Split vent

Split InjectorFlow Path

Carrier gassource

Page 16: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Splitless InjectorPurge Off At Injection

Flow through injector = Column flow only

Carrier gassource Septum

purge

Split vent

Page 17: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Splitless Injector Purge On After Injection

Flow through injector = Column flow + Split Vent Flow

Carrier gassource Septum

purge

Split vent

Page 18: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Split InjectorMajor Variables

Split ratio - determines amount of sample onto column and efficiency of injection (sensitivity vs peak shape)

Liner - influences efficiency of vaporization/discrimination

Temperature - hot enough to vaporize sample without degradation or causing backflash

Injection volume - typically 1-3uL, increasing it does not have as much of an effect as one might think

Page 19: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Split Liners – What’s What?

Straighttube

Straighttube withglass wool

Invertedcup

BaffleFixed glass

wool

Page 20: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Split Liner

C10

C40

C10

C40

Packed with Glass Wool

Without Glass Wool Packing

Peak Area Ratio

n-C40/n-C10 = 0.64

Peak Area Ratio

n-C40/n-C10 = 0.37

Page 21: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Split Liner

C10

C40

C10

C40

Packed with Glass Wool

Without Glass Wool Packing

Peak Area Ratio

n-C40/n-C10 = 0.64

Peak Area Ratio

n-C40/n-C10 = 0.37

Page 22: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Splitless InjectorOverview

Most of the sample is introduced into the column

Used for low concentration samples

Wider peaks are obtained than for split injections

Page 23: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Splitless InjectorMajor Variables

Purge activation time - determines amount of sample onto column and efficiency of injection (sensitivity vs peak shape)

Liner - preventing backflash more critical than vaporization properties (double tapered type recommended)

Injection volume - typically 1uL or less (backflash)

Temperature – long residence times allow for lower temps

Page 24: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Splitless Injector Liners

Straighttube

Bottomrestriction

Dualrestriction

Page 25: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Solvent Vapor Volume Calculator

Page 25

Page 26: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 26

Backflash (carry-over) can give false positives!

Page 27: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Cleaning the Split/Splitless Injector

Ace

toneMeCl2

Injector body

Remove column,

reducing nut, gold seal,

washer and liner

Carrier gas flow off

GC Off

Disconnect split vent line

Replace split vent trap

Page 27

Page 28: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Finding the Split Vent Trap

Follow the split vent line back to the EPC

Page 29: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Finding the Split Vent Trap

Remove cover at Split Vent

Page 29

Page 30: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Replacing the Split Vent Trap

Finger Tight Knurled Nut

G1544-80530

Page 31: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Split Vent Trap Changed (Column Bleed?!?)

Page 31

Before

After

Page 32: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Pulsed Splitless

- sample containment more critical than in split injection

- much sharper peaks than in traditional splitless injection

Pulsed Split

- the most volatile components and solvent effected most

- faster sample transfer not as critical since it’s already fast

Split/Splitless Pulsed Injection

Pressure Pulse contains sample expansion

and transfers analytes to the column faster.

Page 33: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Select Pulsed Splitless Mode in Inlets

Page 34: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Check the Splitless Pressure

Page 35: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Double or Triple the Pressure for ~1 sec less than

the Purge Activation Time

Page 36: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

The BIGGEST Problem in GC is…

There are more things that DON’T go through a GC than DO!

….therefore, don’t inject anything and you’ll never have

problems.

OK, inject, but realize that everything just got dirty…deal with it!

Page 36

Page 37: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Where Does it Get Dirty?

Here

Here

Here

Here

Here

Here

Page 37

Page 38: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

What Are You Doing!?

Page 38

Page 39: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 39

Bonus Peaks or Ghost Peaks

Contamination in INJECTOR or FLOW (carrier gas)

-Contaminated consumables

-Carryover from a backflash or previous sample

-Bad tank of gas or traps have expired

-Septum bleed

*TIP = Run a blank run…it should be blank!

Before After

Page 40: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Bonus Peaks - Ferrule Contamination

min5 10 15 20 25

pA

18

20

22

24

26

28

FID2 A, (E:\FERRULE\RTKVSPGR.D) FID2 A, (E:\FERRULE\AGLTVSPG.D)

Agilent

Vespel/Graphite Ferrule from a bag

Page 40

Page 41: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

More “Off-Brand” O-Ring IssuesControlled Substances Analysis, H2 Carrier

Residue on top of

inlet weldment

Problem Resolution:

Agilent Non-Stick Liner O-Ring

p/n 5188-5365, 10PK

Page 41

Page 42: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 42

Septa Bleed vs Column Bleed (MSD)

Source of peaks from outside of the column Columns: HP 5MS

30mx0.25mmx0.25um

Oven: 80 to 160C at 25 C/min,

160 to 320 C at 3 C/min(4),

320 to 325 C at 20C/min(4)

Injection: split 100:1; 1ul of 100ng/ul

Detector: MSD (HP-5973)

run at max sensitivity- full scan

4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00

Peak height approx. equiv. to 5 ppb of PAH

(actually impurities)

Page 43: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 43

Septa Bleed vs Column Bleed (MSD)

Source of peaks from outside of the column ELIMINATED

Columns: HP 5 MS

30mx0.25mmx0.25um

Oven: 80 to160 C at 25 C/min,

160 to 320 C at 3 C/min(4)

320 to 325 C at 20C/min(4)

Injection: split 100:1; 1ul of 100ng/ul

Detector: MSD (HP-5973)

run at max sensitivity- full scan

6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0 2 0 . 0 0

5 0 0 0 0

1 0 0 0 0 0

1 5 0 0 0 0

2 0 0 0 0 0

2 5 0 0 0 0

3 0 0 0 0 0

3 5 0 0 0 0

4 0 0 0 0 0

4 5 0 0 0 0

5 0 0 0 0 0

T im e - - >

A b u n d a n c e

T I C : P C 2 4 1 N G 1 . DNew Septa Installed

HP Advanced Green Septa

P/N 5183-4594 ( 11 mm)

320 deg

325 deg

Page 44: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 44

Peak Tailing

INJECTOR or COLUMN is Active

-Reversible adsorption of active compounds

(-OH, -NH, -SH)

FLOW problem - dead volume, obstruction, poor installation, or severe column contamination

Miscellaneous – temperature issues for late eluters, overloading of PLOT columns, co-elution, polarity mismatch between phase, solute or solvent, and some compounds always tail

*Tip = Inject a light hydrocarbon, should not tail unless flow path problem.

Before After

Page 45: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

good cut

bad cut

bad cut

A flat, 90° square cut will be optimal for all connections

Flow Path Matters

Page 45

Page 46: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Symptom – Tailing of Active Compounds1. n-Nitrosodimethylamine2. Aniline3. 1,4-Dichlorobenzene-d44. Benzoic Acid5. Naphthalene-d86. Acenaphthene-d107. 2,4-Dinitrophenol8. 4-Nitrophenol9. 2-Me-4,6-dinitrophenol10. 4-Aminobiphenyl11. Pentachlorophenol12. Phenanthrene-d1013. Benzidine14. Chrysene-d1215. 3,3’-Dichlorobenzidine16. Benzo[b]fluoranthene17. Benzo[k]fluoranthene18. Perylene-d12

Sample: 0.5 ng on column loading with ISTD

Column: 20m 0.18mm 0.18µm

Carrier: Helium 37cm/sec, Ramped flow; 0.7ml/min (0.1min) to 1.3ml/min (15ml/min2)

Oven: 35oC (2.5 min) to 80oC (40oC/min), 15oC/min to 200oC, 8oC/min to 275oC (2 min)

Injection: 0.5µl, Splitless, 280oC, purge flow 30ml/min at 0.75 min

MSD: Transfer Line 290oC, Source 300oC, Quad 180oC

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00

0

50000

100000

Time (min)

Abundance

1

2

3 56

7

8

10

11

9

12

13

14

15

16

17

18

4

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00

0

50000

100000

Time (min)

Abundance

1

2

3 56

7

8

10

11

9

12

13

14

15

16

17

18

4

Page 46

Page 47: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Solution – Ultra Inert GC Column1. n-Nitrosodimethylamine2. Aniline3. 1,4-Dichlorobenzene-d44. Benzoic Acid5. Naphthalene-d86. Acenaphthene-d107. 2,4-Dinitrophenol8. 4-Nitrophenol9. 2-Me-4,6-dinitrophenol10. 4-Aminobiphenyl11. Pentachlorophenol12. Phenanthrene-d1013. Benzidine14. Chrysene-d1215. 3,3’-Dichlorobenzidine16. Benzo[b]fluoranthene17. Benzo[k]fluoranthene18. Perylene-d12

Sample: 0.5 ng on column loading with ISTD

Column: 20m 0.18mm 0.18µm

Carrier: Helium 37cm/sec, Ramped flow; 0.7ml/min (0.1min) to 1.3ml/min (15ml/min2)

Oven: 35oC (2.5 min) to 80oC (40oC/min), 15oC/min to 200oC, 8oC/min to 275oC (2 min)

Injection: 0.5µl, Splitless, 280oC, purge flow 30ml/min at 0.75 min

MSD: Transfer Line 290oC, Source 300oC, Quad 180oC

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00

0

50000

100000

Time (min)

Abundance

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00

0

50000

100000

Time (min)

Abundance

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

Page 47

Page 48: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Improved PerformanceImproving the Entire Flowpath

October Agilent

Restricted

48

UltiMetal – TCD,

FPD, NPD/FID Jets

UltiMetal Capillary Flow Technology

Devices, Ultimate Union

UltiMetal Inlet Weldment, Shell

and Transfer Lines

Ultra Inert GC Column

Ultra Inert Gold Seal

Ultra Inert Inlet

Liner

Ne

w

New UltiMetal

FlexiMetal Ferrules

…now from a single supplier

Page 49: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 49

Split Peaks

INJECTOR (poor sample introduction)

-Injecting the sample twice (some how?)

-Mixed sample solvent (polarity difference)

-Sample in syringe needle (manual inject)

INJECTOR (activity)

-Breakdown (not really a split peak, 2 peaks)

-Sample degradation in injector

VOLATILITY

High boilers dropping out on Cold Spots

-Transfer line temps

-Unions or fittings not tracking column temp

Page 50: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Broad Peaks

Page 50

INJECTOR

-Poor Installation

-Change in settings (temps/flows)

-Poor sample focusing

-Large change in sample concentration

FLOW

-Change in gas velocity

-Constant Flow vs Constant Pressure

COLUMN

-Contamination

-Damaged/old stationary phase

-Reverse Solvent Effect

Page 51: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 51

No Peaks

DETECTOR (not on or not operational)

INJECTOR (not working)

-Plugged syringe/plunger not moving

-Wrong injector (or detector)

-Huge leak/no carrier gas flow (older systems)

NOT the COLUMN Unless…

-Broken column

-No column

Page 52: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 52

Symptom – No Peaks

10 20 30 40 50

14

16

18

20

22

24

26

28

30

Page 53: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 53

Solution - Unplugged Syringe

0 10 20 30 40 50

14

16

18

20

22

24

26

28

30

Page 54: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 54

Peak ResponseAll Change in Size

DETECTOR (response problem)

-Settings or flows changed

-Electronics failing

-Split ratio set incorrectly

-Wrong purge activation time

-Septum purge flow too high

-Injector temperature too low*

INJECTOR

-Leaky syringe

*Tip = Ask is it all of them or some of them, if all then injector or detector

Page 55: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 55

Peak ResponseSome Change in Size

INJECTOR or COLUMN is active/contaminated

-Irreversible adsorption of active compounds (-OH, -NH, -SH)

-Decomposition of sample

-Temperature Change – Discrimination

-Evaporation from sample

*Tip = If only some change, then ask which ones? If active compounds

then activity. If tracks volatility then cold spots or inlet discrimination.

Page 56: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Group/Presentation Title

Agilent Restricted

Month ##, 200X

Column

FlowController

Regulators

Air

Hyd

rog

en

Ca

rrie

r G

as

Mol-SieveTraps

Fixed

Injection

Port Detector Electrometer

Recorder/Integrator

Restrictors

Cylinders or Generators

Typical Gas Chromatographic System

Picking the appropriate

stationary phase and optimum

dimensions for the column will give the greatest resolution

in the shortest analysis time.

Page 57: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 57

Peak FrontingShark Fin Shaped or Just Slight

COLUMN (contaminated)

-Overload (More pronounced with large solute and phase polarity differences)

INJECTOR

-Poor efficiency (flow/temp)

-Column installation

-Compound very soluble in injection solvent (need retention gap)

-Mixed sample solvent

OTHER

-Co-elution

-Breakdown

Page 58: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 58

Retention Time Shift

2.00

3.25

4.75

2.75

4.00

5.50

INJECTOR

-Change in injection solvent

-Large change in sample concentration

FLOW

-Leak in the septum

-Change in gas velocity

COLUMN

-Contamination

-Damaged stationary phase

-Loss of stationary phase

-Change in temperature

1.75

3.00

4.50

Page 59: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Constant Pressure vs Constant Flow Retention

C10

C20

C30

C40

C10

C20

C30

C40

Under constant pressure conditions, flow decreases as temperature increases.

(viscosity of a gas increases as temperature increases)

Page 59

Page 60: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Gas Viscosity vs Temperature

J.V. Hinshaw, Column Connections, LCGC Asia Pacific, 12(2), 1100 (2009).

Page 60

Page 61: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 61

Loss of Resolution

Resolution is a function of separation and peak width

Separation

Peak Width

Page 62: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 62

Loss of Resolution - Separation Decrease

COLUMN

-Different column temperature

-Contamination (more phase?)

-Matrix components co-eluting

-Different column phase?

Separation

Peak Width

Page 63: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 63

Loss of Resolution - Peak Broadening

FLOW

-Change in carrier gas velocity

-Make-up gas

COLUMN

-Contamination

-Phase degradation

INJECTOR (efficiency)

-Settings, Liner, Installation, etc.

Peak Width

Separation

Page 64: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 64

Typical Problems of Optimized Methods becoming

Unoptimized…and the Reason Why.

• Peak Tailing – Flow Path or Activity

• Bonus Peaks – In Sample or Back Flash (Carry Over)

• Split Peaks – Injector Problems, Mixed Solvent

• No Peaks – Wasn’t Introduced, Wasn’t Detected

• Response Changes – Activity, Injector Discrimination, Detector Problem

• Peak Fronting – Overload or Solubility Mismatch, Injector Problems

• Shifting Retention – Leaks, Column Aging, Contamination or Damage

• Loss of Resolution – Separation Decreasing, Peak Broadening

• Baseline Disturbances – Column Bleed, Contamination, Electronics

• Noisy or Spiking Baseline – Electronics or Contaminated Detector

• Quantitation Problems – Activity, Injector or Detector Problems

Page 65: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 65

Quantitation Problems

DETECTOR

-Poor stability (electronics) or Baseline disturbances (contamination)

-Outside detector's linear range or wrong settings

Activity (adsorption) in INJECTOR or COLUMN

INJECTOR

-Technique, settings, conditions

-Syringe worn

OTHER

-Co-elution

-Matrix effects

-Sample evaporation – leaky vials

-Sample decomposition

Page 66: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 66

Baseline Disturbances

Sudden Changes, Wandering, or Drifting

COLUMN or DETECTOR

-Not fully conditioned or stabilized (electronics)

-Contamination

FLOW

-Changes in carrier and/or detector gas flows

-Valves switching, leaks

WANDER

DRIFT

Page 67: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 67

Noisy Baseline

FLOW

-Contaminated gas

-Incorrect detector settings

COLUMN

-Bleed if at high temperature

-In detector flame (poor installation)

MILD

SEVERE

DETECTOR

-Air leak - ECD, TCD

-Electronics malfunction

Page 68: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 68

Spiking Baseline

DETECTOR

-Particles entering the detector

-Random: poor connection

-Regular: nearby "cycling" equipment (electronics)

Page 69: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Page 69

Remember

Multiple cause and effect

Do not change too many variables at

once

Complete system = Carrier Gas + Injector +

Column + Detector + Data System

Page 70: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Self-Tightening Column Nuts

Page 70

Video at agilent.com/chem/STnutvideo

• Less wasted time: No retightening

needed after repeated thermal cycles

• Ease of use: Finger-tight,

consistent connections without tools

• Leak Free = Lower column bleed:

Longer column life

Innovative spring-driven piston continuously presses against ferrule

Page 71: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Self-Tightening Column Nuts

Page 71

Page 72: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Changes in Column Dimensions, Gas Type or

Velocity Require Changes in Temp Program Rates

Method Translation Software to the Rescue!

Page 72

Page 73: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Troubleshooting Resources

Online Troubleshooting and Maintenance Videos

http://www.chem.agilent.com/en-US/Technical-Support/Instruments-Systems/Gas-

Chromatography/Pages/troubleshootingvideos.aspx

GC Troubleshooting Guide

http://www.chem.agilent.com/en-US/Products-Services/Instruments-Systems/Gas-

Chromatography/pages/gp6770.aspx

Method Translation Software

http://www.chem.agilent.com/en-US/Technical-Support/Instruments-Systems/Gas-

Chromatography/utilities/Pages/gcmethodtranslation.aspx

Page 73

Page 74: GC Best Practices & Troubleshooting · GC Best Practices & Troubleshooting. Page 2 Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what

Agilent Better GC Connections

www.agilent.com/chem/betterGCconnections

Order the poster… View the video…