Optimization of Protection in Computed Tomography (CT)-What can radiographers

do?

IAEA Regional Training Course on Radiation Protection of patients for Radiographers, Accra, Ghana, 11-15 July 2011

IAEA

Introduction

18: Optimization of Protection in CT Scanner 2

IAEA

Multislice CT scanner

IAEA

The CT image is mathematically reconstructed from the measured data.

IAEA

Outline

• Introduction

• Factors affecting image quality

• Factors affecting dose

• Optimization

18: Optimization of Protection in CT Scanner 5

IAEA 18: Optimization of Protection in CT Scanner 6

• CT scanning offers immense medical benefits (improved low contrast resolution)

• Patient doses can be high in CT

• CT trauma: 59 mGy (normal chest 0.33 mGy x ~ 150)

• Image quality in CT is often higher than necessary for diagnostic confidence

• Need for patient dose management

• Radiographer plays a big role in this endeavor

…………………Introduction

IAEA

…….Introduction

18: Optimization of Protection in CT Scanner 7

• CT scanner is a complex equipment

• User must understand relationship between scan protocols and patient dose and image quality

• Such understanding is not intuitive/training & experience needed

IAEA

•Scanning protocols can not simply be transferred between different manufacturers

IAEA

Factors affecting image quality

18: Optimization of Protection in CT Scanner 9

• Spatial resolution• Contrast discrimination• Spatial uniformity• Noise• Pixel size• Slice thickness• mAs• Tube voltage• Reconstruction algorithm• Sampling frequency• Pitch (in multislice)• Patient size

IAEA

Spatial resolution

18: Optimization of Protection in CT Scanner 10

The ability to resolve neighbouring structures•Size of detectors• Number of detectors• X-ray focus• Distance between source and detector• Sampling frequency

IAEA

Spatial Resolution

IAEA

Contrast Discrimination

• The ability to detect differences between neighbouring structures of similar density (CT number)

IAEA

Spatial Uniformity

• The faithful representation of shapes and contrast throughout the image

Identical

IAEA

Spatial Uniformity

• The faithful representation of shapes and contrast throughout the image

Non-Uniform Contrast

IAEA

Spatial Uniformity

• The faithful representation of shapes and contrast throughout the image

Non-Uniform Shape

IAEA

Noise

What is noise? Anything that contributes to the degradation of CT images

IAEA

Noise

• Decreases the quality of the image

• Makes diagnosis harder by• Masking information

• Not presenting correct information

IAEA

Other Factors - Image Quality & Patient Dose

• Pixel Size

• Slice Thickness

• mAs

• Algorithm

• Sampling Frequency

• Artefacts

IAEA

Slice Thickness

• Thin slice -> smaller signal

• Smaller signal -> more noise

• More noise -> poor image

• Solution increase dose!!

IAEA

mAs

• Current directly proportional to intensity of X-rays

• Low current, low intensity

• Low intensity, low signal

• Low quality image produced

• Solution increase dose!!

IAEA

mAs

100 mAs 50 mAs

IAEA

Paper of Interest

• http://www.oucom.ohiou.edu/ou-microct/Downloads/Tradeoffs_in_CT_Image_Quality_and_Dose_9794-13379.pdf

• Good paper on CT image quality & dose tradeoffs

IAEA 18: Optimization of Protection in CT Scanner 23

Monitoring of dose

• An issue to be considered

• Factors should be

adjusted to produce

necessary image

quality without resulting

to unnecessary dose

IAEA 18: Optimization of Protection in CT Scanner 24

Monitoring of dose

• CTDI: special quantity to express radiation Dose in CT• Useful since it is not easy to measureactual dose to internalorgans• Not a measure of patient dose

IAEA 18: Optimization of Protection in CT Scanner 25

Dose Length Product

• DLP is a practical

way for expressing

total radiation dose

deposited in body

A measure of patient

dose/risk

DLP=CTDIvol x scan

length

IAEA

Factors affecting dose

18: Optimization of Protection in CT Scanner 26

• Tube current (mA)• Scan (rotation) time (s)• Tube voltage• Beam (slice) width (mm)•Helical pitch• Number of slices/tube rotations• Number of slices/tube rotations

IAEA

Effect of scan parameters on CTDIvol

18: Optimization of Protection in CT Scanner 27

• mAs and scan (rotation) time

-At a given pitch, CTDIvol increases linearly with mA and time e.g. 2 x mAs= 2x CTDIvol - Given 100 mA 1s, one can double mAs by 200 mA x 1s or 100 mA x 2s)

IAEA

Effect of scan parameters on CTDIvol

18: Optimization of Protection in CT Scanner 28

• CTDIvol increases with kV (approx. kV squared) - Variation of dose with kV if other parameters are constant kV relative CTDIvol 80 0.4 120 1.0 140 1.4• Beam width: CTDI increases if beam width wider than nominal imaged width (otherwise remains constant)

IAEA

Effect of scan parameters on CTDIvol

18: Optimization of Protection in CT Scanner 29

For some image noise, dose can increase with decreasing kV

kV CTDIvol per unit noise

head body

80 1.6 2.6

110 1.2 1.2

130 1.0 1.0

e.g. Siemens Emotion 6

IAEA

Effect of scan parameters on CTDIvol

18: Optimization of Protection in CT Scanner 30

Slice width : indirect effect of imaged slice width on dose•In theory: halve image slice width means double mAs • In practice: have increased contrast so compromise on increase in mAs

Scan length: Number of slices (tube rotations)• CTDIvol is approx. independent of scan length• DLP is directly related to scan length

IAEA

Pitch

• Pitch = table travel per rotationnominal slice width

Single Slice Multi Slice

IAEA

Pitch

Single Slice CT

Table travel = 5 mm per rotationSlice width = 5 mmPitch = 5/5 =1

Pitch = table travel per rotation nominal slice

width

IAEA

Pitch

Multi Slice CT

Pitchx = table travel per rotation xray beam width

IAEA

Pitch

Multi Slice CT

20 mm

5 mm

15 mm

Pitchx = 20 /15 = 1.33

Pitchd = 20 / 5 = 4

IAEA

Pitch

Pitch = 1, no gaps in helical path

Pitch > 1, gaps

Pitch < 1, overlap

IAEA

Helical considerations

18: Optimization of Protection in CT Scanner 36

• For equivalent scan parameters: helical dose is approx.equal to sequential (axial dose)

• For constant mAs, CTDIvol is inversely proportional to pitch

•On single slice scanners, increasing pitch is used for dose reduction)

IAEA

Multi-slice considerations

18: Optimization of Protection in CT Scanner 37

•Dose differences between single and multi-slice - Use of pitch for dose reduction - Extent of additional rotations in helical scanning -’over-beaming’-penumbra lies outside active detectors

Multi-slice pitch: 200 mAs (200 mA, 1s rotation)• Effective mAs or mAs per slice= True mAs/ pitch• On MSCT, mAs is often adjusted to keep effective mAs constant• Therefore CTDIvol will remain constant with pitch

IAEA

Multislice: scan protocol

18: Optimization of Protection in CT Scanner 38

• Greater flexibility of MSCT allows user to

- increase scan length

- scan more phases in multiphase studies

- increase mAs to keep noise down in thin slices

IAEA

Effect of patient size

18: Optimization of Protection in CT Scanner 39

• Theory of x-ray CT: HVL in tissue is approx. 4 cm• To maintain constant noise, double mAs for extra 4 cm of tissue

• Adjusting mAs for patient size: - small patients: require lower noise -large patients : higher noise accepted

• Some manufacturers recommend doubling mAs for - approx. 10 cm in abdomen scans -approx. 13 cm in lung scans• Automatic tube current control (mA modulation) (20-50% dose reductions without compromising image quality) -manufacturer’s weight/age based pre-programmed protocol - mAs adjusted for lateral patient dimensions

IAEA

CH McCallough: Dose optimization in CT: Implementation and clinical acceptance of size based charts (RSNA, 2002)

18: Optimization of Protection in CT Scanner 40

Patient width* (cm) Relative mAs

> 21-26 0.4

> 26-31 0.5

> 31-36 0.7

> 36-41 1.0

> 41-46 1.4

> 24-51 2.0

* Lateral width based on A-P scout at level of liver

IAEA Radiation Protection in Paediatric Radiology L06. Radiation protection in computed tomography 41

Example of successful story. IAEA TECDOC-1621(2009): Dose reduction in CT

while maintaining Diagnostic Confidence: A feasibility /Demonstration study

• Six hospitals (Canada, Greece, India,Poland,Thailand,UK) developed a relationship between image noise and patient weight.

• Used the relationship to adapt the noise in the image to a pre-selected value.

• Dose reductions: 25-62% (abdomen CT ) + 12-79% (Chest CT)

Paediatric chest CT

IAEA 18: Optimization of protection in CT scanner 42

Quality Control Tests In CT

Quality control test description on:

• CT accuracy, uniformity, linearity and noise,

• Low and high contrast resolution

• Alignment, Couch travel accuracy

• Gantry tilt measurement

• Dosimetry

IAEA

Summary• Scanning parameters should be based on study

indication, patient size and body region being scanned

• Manufacturer protocols should be the starting point. Any adjustments must be done in consultation of radiologist.

• Image quality in CT is often higher than necessary for diagnostic confidence

• Implementation of QC programme is important for patient dose management

• Training of physicians and CT staff can help in management of protocols and patient dose

18: Optimization of Protection in CT Scanner 43

IAEA