histamine test for food safety - hyserve assay.pdf · 2011-09-05 · histamine test for food...
TRANSCRIPT
1/HyServe
Histamine TestHistamine Test for Food Safetyfor Food Safety
Easy, Rapid and Specific!
2/HyServe
Histamine Histamine causes Allergycauses Allergy--like food poisoninglike food poisoning
tuna,sardine, Mahi-Mahi, herring, etc.
Bacteria
Allergy-like food poisoning
Poor handlingTemperature abused
HistamineHistidine (amino acid)
3/HyServe
Regulation ofRegulation of HistamineHistamine in Fish and Fisheries productsin Fish and Fisheries products
US (FDA/HACCP) 500ppm based on toxicity 50ppm defect action level
EU (EC 91/493/EEC) 9samples for each batch ≦200ppm, AV≦100ppm,
JAPAN (MHLW) 100ppm for control
*MHLW=the Ministry of Health and Welfare
4/HyServe
Performance comparison with other methodPerformance comparison with other method
Histamine Test Commercial test kit HPLC method AOAC method
Analytical Methodology Enzymatic method EIA method HPLC method Fluorometric method
Detection range 20 - 300 ppm 1 - 100 ppm 0.1 ppm 50 ppm
Total Test Time 1 h 1 - 1.5 h >6 h >6 h
Reproducibility Good Variation may occur. Excellent Variation is large.
Sensitivity Quite Accurate Accurate Very Accurate Accurate
Standard 1 point is required. 5 points are required. Required 4 points are required.
Complex procedure
5/HyServe
Performance comparison with other methodPerformance comparison with other method
Analytical Methodology Enzymatic method EIA method HPLC method Fluorometric method
Detection range 20 - 300 ppm 1 - 100 ppm 0.1 ppm 50 ppm
Total Test Time 1 h 1 - 1.5 h >6 h >6 h
Sensitivity Quite Accurate Accurate Very Accurate Accurate
Standard 1 point is required. 5 points are required. Required 4 points are required.
Simple Somewhat complicated Complicated Very complicated
Many stepsPreparation of analytical
apparatusMany steps, Reagent :
ready to use
Procedure
Histamine Test
Reproducibility Good Variation may occur. Excellent Variation is large.
6/HyServe
Comparison with other methodsComparison with other methods
10% trichloroacetic acidextraction
75% methanol extractionBoiling or water extraction
Histamine test HPLC method AOAC methodSample Sample Sample
Sample preparationSample preparation(Extraction) Neutralization(Extraction)
Add 0.4N acetic acid
Ion-exchange treatment Ion-exchange treatment
Enzymatic reaction
Fluorescence reaction
Fluorescence reactionAssayAssay
Assay using HPLC Assay using fluorometerAssay using spectrophotometer
7/HyServe
Histamine TestHistamine Test
Specific for Histamine
Rapid results
Easy procedure
8/HyServe
Principle of Principle of Histamine TestHistamine Test(colorimetric enzyme assay)
(reduced form)
Histamine dehydrogenase (HDH)*
Histamine
4-Imidazolylaldehyde 1-methoxy PMS
1-methoxy PMS
WST- 8
WST- 8 formazan(at 460nm)
* patent pending
9/HyServe
Composition of Composition of Histamine TestHistamine Test
60 assays / kit
Histamine standard solution 30ml×1
Colorimetric reagent(desiccated) ×6
Bufffer24ml×2
Enzyme reagent(desiccated)×6
10/HyServe
Assay procedureAssay procedure of of Histamine Test (1)Histamine Test (1)
2. Sample 1 g, and0.1M EDTA/Na (pH8) buffer 24 ml.
4. Filtrate.3. Boil for 20 min and then, cool the tube on ice (until it becomes <20ºC). *
1. Homogenize a sample.
* You may skip this step for canned fish and fish meal.
11/HyServe
Assay procedureAssay procedure of of Histamine Test (2)Histamine Test (2)
8. Measure the absorbance at 460 nm.
7. Incubate at 25-37ºC for 15 min.
6. Add reagents and sample to test tube and mix well.
5. Add distilled water to vial of reagents.
12/HyServe
Linearity of Linearity of Histamine TestHistamine Test
y = 0.0378x + 0.0127r = 1
0.00
1.00
1.20
0 5 10 15 20 25 30 35
(2.5)
(1.9)
Histamine concentration (Histamine concentration (mMmM/assay)/assay)
0.20
0.40
0.60
0.80
(2.3)
(1.7)
(CV:%)Abs
A
bs 46
046
0
13/HyServe
Recovery of histamine added to canned tunaRecovery of histamine added to canned tuna
Canned tuna in oil Canned tuna in soup
Added Recovered ( ppm )histamine(ppm )
10 11.3 10.8
20 21.6 21.2
50 51.0 49.5
75 72.0 73.5
Recovered ( )histamine(ppm )
10 11.3 10.8
20 21.6 21.2
50 51.0 49.5
75 72.0 73.5
5 5.0 5.05 5.0 5.0
Recovery Recovery 98 – 108 %RecoveryRecovery 96 – 103 %
Good recovery!
14/HyServe
Comparison of Comparison of Histamine TestHistamine Testand other methods resultsand other methods results
5662,545H PLC m ethod
5302,646AO AC m ethod※1
5742,728V e ※2
4632,764H istam arine※2
5372,886H istam ine Test
C anned tuna (ppm )
R aw tuna (ppm )
5662,545H PLC m ethod
5302,646AO AC m ethod※1
5742,728V eratox ※2
4632,764H istam arine※2
5372,886H istam ine Test
C anned tuna (ppm )
R aw tuna (ppm )
Storaged at 20℃ for 2 daysS toraged at 20℃ for 2 days
※※11 fluorometric fluorometric analysis established as AOAC official method analysis established as AOAC official method ※※22 commercial EIA test kitcommercial EIA test kit
15/HyServe
Performance comparison with other methodPerformance comparison with other method
Histamine Test Commercial test kit HPLC method AOAC method
Analytical Methodology Enzymatic method EIA method HPLC method Fluorometric method
Detection range 20 - 300 ppm 1 - 100 ppm 0.1 ppm 50 ppm
Reproducibility Good Variation may occur. Excellent Variation is large.
Sensitivity Quite Accurate Accurate Very Accurate Accurate
Standard 1 point is required. 5 points are required. Required 4 points are required.
Simple Somewhat complicated Complicated Very complicated
(Many steps)(Preparation of analytical
apparatus)(Many steps, Reagent :
ready to use)
Procedure
Histamine Test
Total Test Time 1 h 1~ 1.5 h >6 h >6 h
16/HyServe
Histamine TestHistamine Test for Food Safetyfor Food Safety
Easy, Rapid and Specific!