Sugar Sensitive Phenylboronic Acid-Based Hydrogels: Towards glucose analysis in sweat
1
by
HENOK TESFAY 4th year project presentation
Supervisors: Dr. Aoife Morrin,
Dr Larisa Elena Florea Dr. Colm Delaney
School of Chemical Sciences May 2016
Phenylboronic acid (PBA)
PBA bearing hydrogel
2
Sugar
Photochemical polymerization(Gel forming reagents)
Acrylamide(AAm)
linker
3-(Acrylamido) phenylboronic acid – (PBA)
sugar reactive monomer
2-Hydroxy-2-methpropiophene-(HMPP)
photo initiator
Methylenebisacrylamide (BIS)
Cross-linker
3PolyDimethylsiloxane(PDMS) mould
Uvp CL 1000 ultraviolet crosslinker
Source: http://www. idf.org/about-diabetes [1] 4
Diabetes is becoming a big problem in our society, affecting currently one in every 20 people in the world.[1]
Prevalence of Diabetes • Diabetes caused 5.1 million
deaths in 2013. [1]• Every six seconds a person
dies from diabetes.[1]
Source: http://www. idf.org/about-diabetes [1]
5
Is there a cure for Diabetes?• No cure for diabetes • Can be controlled but not
easy. • Must prick their finger to
get drops of blood for test. • Limitation – don’t know
blood glucose between checks?
• Need advanced sensor
Source: http://www.slideshare.net/funk97 / glucose-monitoring-smart-watch
6
• Non-invasive *Blood free * not causing pain or trauma *real-time glucose sensors. *Ables to monitor during the day and night *use sweat, saliva or tears .
Source: http://ingenuitylab.ca/about
7
Source: Health Tech Insider
Time dependent swelling of Hydrogel in 10mM Fructose solution (5mol% PBA)
8
gravimetric method
Behaviour of 5mol% PBA Hydrogel in Glucose, Sucrose & Fructose Solutions
9
Behaviour of Hydrogel (20mol% PBA) in different concentration of Glucose, Fructose & Sucrose
10
20mol%PBA gel in Glucose
• Control
• In 4mM
• In 6mM
• In 10mM
• In 20 mM
20mol%PBA gel in Fructose
•Control
• In 4mM
• In 6mM
• In 10mM
• In 20mM
11
Linear regression Analysis of Calibration Plots
Sugars R2 (1-5mM) Slope (Swelling ratio/[]
LOD (3σ/M) LOQ(10σ/M)
Fructose 0.9971 0.067 0.30 1.0
Glucose 0.9933 0.026 0.50 1.50
Sucrose 0.8661 0.003 2.20 7.22
12
Glucose in 20mol% & 5mol% PBA Hydrogel
13
Conclusion
• The hydrogel shown a swelling in response to the Glucose and Fructose. But didn’t show a response to the sucrose.
• Challenges: Accuracy, the PBA has more sensitive for fructose than glucose
14
Future work
• Measuring using impedance. • Adding tertiary amine
15
fixed voltage 20 mV frequency range 0.1 – 10 MHz
Solartron (Impedance)
0 5 10 15 20 25 30 35 40 45 500
1
2
3
4
5
C (n
F)
glucose concentration (mM)
1 kHz
Acrylamide20 mol % PBA1 mol % bisAm1mol % HMPP(1:1) water/DMSO
UV light30 minutes
Electrode Functionalization with Hydrogel
Glucose Additions
Electrical Impedance Spectroscopy
Source: Insight centre for Data analytics (DCU)
16