GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i) Complete Tagger. (ii) Vacuum Chamber Tests.

Upload: jerome-morrison

Post on 17-Jan-2016

219 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

GlueX Two Magnet Tagger

G. Yang University of Glasgow

Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i) Complete Tagger. (ii) Vacuum Chamber Tests.

Page 2: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Software used : Opera 3D, version 8.010.

What we have calculated: • 2 D, 3 D magnetic field distributions.• The Effective field boundary along the magnet edge.• Electron trajectories calculated by using Opera 3D,

post processor.• Optical properties using a grouping of 9 electrons

trajectories.

Part 1 3 D Tosca analysis

Page 3: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

2 D magnetic field histogram calculated by TOSCA.

magnetic field along a line perpendicular to the magnet output edge.

Page 4: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

• Electron trajectories have been calculated using Opera 3 D post processor.

• By using the calculated electron trajectories, beam spot size, and focal plane position have been determined.

Page 5: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Calculated electron trajectories

Beam trajectories calculated from TOSCA in the mid plane for 1 GeV and 9 GeV. Those trajectories having the same direction focus on position 1, and those trajectories having the same starting position focus on position 2. ( electrons travelling in the direction shown by the top arrow ).

Electron trajectory bundles according to their directions at the object position.

(1 GeV) (9 GeV)

Page 6: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Other interesting results

The intersections of the beam trajectories with the plane through the focusing point and perpendicular to the beam ((a):1 GeV, (b): 9 GeV), the dotted lines show the beam spot profile.

(We have used a rectangular object shape. So, ideally we expect to obtain a rectangular image at the focal plane.)

(a) (b)

Page 7: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

• Preliminary drawings of the two identical magnet tagger system are shown. The magnet dimension used here are the same as used in the Tosca analysis.

• The drawings include the two magnets, a vacuum chamber.

• Assembly procedure are proposed.

Part 2 Preliminary drawings

Page 8: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

The two identical magnets tagger

Magnet 1

Magnet 2

Vacuum chamber

The vacuum force is around 70 tons, so the vacuum chamber needs external support

Page 9: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

The electron entrance angle :5.9 degrees

Main beam exit angle: 6.6 degrees

Main beam bending angle 13.4 degrees

The angle between the photon exit beam and the focal plane: 9.94 degrees

1

1

1

Page 10: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

• Length: 3.11 m.

• Width: 1.09 m.

• Height:1.41m.

• Weight: ~38 Tonns for one magnet.

• Conductor area: 84 cm2.

• Current: 154 A.

• Magnetic field:

1.5 T.

• Pole gap: 2 cm

1

Vertical section through one of the dipole magnets showing pole profile and coil geometry

Page 11: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Vacuum chamber

1

1

Page 12: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Vertical section showing how vacuum O-ring compression is determined

*

O-ring sealVacuum chamber

Page 13: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Vertical section showing how coils are supported against magnetic forces (0.5 ton)

Page 14: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Vertical section showing how weight of coils are supported (each coil ~0.4 ton)

Page 15: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

The brackets and spacers used in the tagger system

1 2

3 4

Page 16: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Part 3. (i) Proposed Assembly procedure for complete tagger

Page 17: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 18: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 19: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 20: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 21: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 22: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 23: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 24: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 25: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 26: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 27: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 28: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 29: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 30: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 31: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Part 3. (ii) Proposed Assembly procedure for vacuum chamber testing

Page 32: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 33: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 34: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 35: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 36: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 37: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 38: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)
Page 39: GlueX Two Magnet Tagger G. Yang University of Glasgow Part 1, 3 D Tosca analysis. Part 2, Preliminary Drawings. Part 3, Proposed Assembly Procedures (i)

Conclusions

1. Optical properties meet design specification

2. Sufficient details is provided so that a realistic cost estimate of the tagger magnets and vacuum chamber can be obtained from potential manufactures.