fundamentals of cardiac devices
TRANSCRIPT
-
8/13/2019 Fundamentals of Cardiac Devices
1/86
1
Welcome
Chris Stamper
Technical Field Engineer
-
8/13/2019 Fundamentals of Cardiac Devices
2/86
2
The Fundamentals of Cardiac DevicesModule 2
-
8/13/2019 Fundamentals of Cardiac Devices
3/86
3
Objectives
After completing this introduction to Cardiac Devices, you ill!e a!le to"
# $dentify the various cardiac device systems
#$dentify the !asic indications for implantation for eachdevice
# %ecogni&e each device type on a chest '(ray
-
8/13/2019 Fundamentals of Cardiac Devices
4/86
)
Cardiac Devices
# Designed to"* %estore or maintain a rhythm and rate sufficient to meet
meta!olic needs
* +rovide diagnostic information a!out
# Device operation
# The patient
-
8/13/2019 Fundamentals of Cardiac Devices
5/86
Cardiac Devices
# +acema-ers or $mplanta!le +ulse .enerators /$+.0* +rovide a rate to support meta!olic needs
* +rovide various diagnostics
* Single and dual cham!er
*A!out 1( years longevity
* Some of the neer pacema-ers include therapies hich
pace(terminate AT4AF
-
8/13/2019 Fundamentals of Cardiac Devices
6/86
5
Cardiac Devices
# $mplanta!le Cardioverter Defi!rillators /$CDs0* %estore sinus rhythm in the presence of tachycardia
# Defi!rillate
# Cardiovert
# Anti(Tachy +ace /AT+0
* +rovide a rate to support meta!olic needs
# $ncludes single or dual cham!er pacing
* +rovide various diagnostics
*A!out 5(6 years longevity
* There are also devices designed to terminate AF via
cardioversion4AT+ in addition to standard $CD therapies
-
8/13/2019 Fundamentals of Cardiac Devices
7/86
7
Cardiac Devices
# Cardiac %esynchroni&ation Therapy /C%T0* $+. ith C%T, or an $+. 8 $CD ith C%T
* %estore ventricular synchrony
# 9ses a specially designed lead placed usually on the posterior(lateral
all of the :; via the Coronary Sinus circulation
# :; epicardial lead placement is an option
# +rovides %; and :; synchronous pacing
* May restore rhythms in presence of lethal tachycardia
# C%T pacing 8$CD /
-
8/13/2019 Fundamentals of Cardiac Devices
8/86
1
Cardiac Devices
# $mplanta!le :oop%ecorders /$:%0
* +rovides rate(!ased monitoring
# Fast rates
# Slo rates
* +rovides E.M during patient
triggered events
-
8/13/2019 Fundamentals of Cardiac Devices
9/86
6
IndicationsPacemakers
# The A
-
8/13/2019 Fundamentals of Cardiac Devices
10/86
IndicationsDefibrillators
# +rimary vs Secondary +revention for SCA* +rimary
# +atients ho have e'perienced a previous SCA or ventricular arrhythmia
# Studies such as A;$D, C$DS2, CAS
-
8/13/2019 Fundamentals of Cardiac Devices
11/86
IndicationsDefibrillators (cont.)
Bell defined !y
-
8/13/2019 Fundamentals of Cardiac Devices
12/862
IndicationsDefibrillators (cont.)
# $CD Class $ %ecommendation
# +atients at least ) days post(M$
# :;EF 3 * )
# =
-
8/13/2019 Fundamentals of Cardiac Devices
13/863
IndicationsCardiac Resynchronization hera!y
# =
-
8/13/2019 Fundamentals of Cardiac Devices
14/86)
Indications"oo! Recorders
# Transient, infreLuent !ut recurrent syncope
-
8/13/2019 Fundamentals of Cardiac Devices
15/86
Reco#nizin# $ystemsPacemakers
Dual Cham!er +acema-er
%; :ead at the Ape'
%A :ead in Appendage
-
8/13/2019 Fundamentals of Cardiac Devices
16/865
Reco#nizin# $ystemsICD or %i#h Po&er
%ight Atrial :ead
Appro'imate position outlined
%ight ;entricular :ead
Bith %; and S;C coils
-
8/13/2019 Fundamentals of Cardiac Devices
17/867
Reco#nizin# $ystemsCR "o&'Po&er
%ight Atrial :ead
:eft ;entricular :ead
+laced on the surface of the
:; via the Coronary Sinus
%ight ;entricular :ead
-
8/13/2019 Fundamentals of Cardiac Devices
18/861
Reco#nizin# $ystemsCR %i#h'Po&er
%ight Atrial :ead
:eft ;entricular :ead
+laced on the surface of the
:; via the Coronary Sinus
%ight ;entricular :ead
=ote the 2 high voltage defi!
coils
Surface EC. leads
-
8/13/2019 Fundamentals of Cardiac Devices
19/866
Pacin# hera!y
#Senses underlying heart rate* Delivers lo energy electrical
pulses hen rate falls !elo
programmed limit
# :o energy pulses capture and
depolari&e the heart musclecausing it to contract
* Dual Cham!er $+.s provide
A(; synchrony
-
8/13/2019 Fundamentals of Cardiac Devices
20/862
# Atrium and ;entricle
* Sensing
* +acing
# ;entricle
* Antitachycardia pacing /AT+0
ICD Pacin# hera!y
Atrial :ead
;entricular :ead
-
8/13/2019 Fundamentals of Cardiac Devices
21/862
%i#h'olta#e hera!y
Cardioversion and
Defi!rillation are delivered in
!iphasic avesFor e'ample" First, from the S;C
coil and the Can to the %; coil,
and then reverse
The device must detect * charge
* confirm * and deliver the shoc-
Fast, accurate detection, and fast
charge times are critical
-
8/13/2019 Fundamentals of Cardiac Devices
22/8622
ICD hera!y
Brady and
Anti-Tachypacing (ATP)
# The $CD lead is
designed to carry !oth
high voltage and
pacing therapies
* >rady
* AT+
* Cardioversion
* Defi!rillation
-
8/13/2019 Fundamentals of Cardiac Devices
23/86
-
8/13/2019 Fundamentals of Cardiac Devices
24/862)
Im!lant vol*tion
+acema-ersesterday
# First $mplants in early 65s
# Single Cham!er, non
programma!le
# A!out 2 year longevity
# A!out 2 cc
# A!dominal implants, sternotomy
for epicardial leads
-
8/13/2019 Fundamentals of Cardiac Devices
25/862
Im!lant vol*tion
+acema-ersToday
# +ectoral implants
# 5(7 F transvenous lead
placement
* Hutpatient4overnight stay
# J 3 cc
# 1( years longevity
# Dual cham!er, multi(
programma!le
# Advanced diagnostics and
trending information
-
8/13/2019 Fundamentals of Cardiac Devices
26/8625
Im!lant vol*tion
$CD +ast" N(2 years ago $CDs
# %eLuired maGor surgery
* A!dominal implants
* Median sternotomy to suturedefi! patches on heart
* :ength of hospital stay I ee-
# =onprogramma!le
#
-
8/13/2019 Fundamentals of Cardiac Devices
27/8627
Im!lant vol*tion
$CDs Today
# Similar to a pacema-er implant
# Transvenous, single incision
* +ectoral implant
* Hvernight stay
# :ocal anesthesia, conscious
sedation
# +rogramma!le therapy options
# Single, dual and triple cham!er
/C%T0
# 9p to 6 years longevity
# A!out 3 cc
# I N, implants4year
-
8/13/2019 Fundamentals of Cardiac Devices
28/8621
$tat*s Check
# .iven the folloing patient
* 57 H male patient, had an anterior M$ 2 years ago
* EF 2
* loc-er0
* C4H
# Severe shortness of !reath at rest, fatigue, una!le to perform AD:, vertigo, rare
syncope
-
8/13/2019 Fundamentals of Cardiac Devices
29/86
26
$tat*s Check
# +atientPs EC."
=S%
%ate 71
:>>
Bhat device
system /if
any0 is the
patient li-ely
to getKClic- for Anser
-
8/13/2019 Fundamentals of Cardiac Devices
30/86
3
$tat*s Check
# Could this !e the previous
patientPs chest '(rayK
* $s it C%TK
* $s it
-
8/13/2019 Fundamentals of Cardiac Devices
31/86
3
$tat*s Check
# Consider the folloing patient* Elderly female patient
* Medical history includes
# E'(smo-er * Luit years ago
# Mild e'ertional angina
* Cardiac cath shos mild disease right coronary artery
# 3 adult children
*Medications# =itroglycerine +%= # Aspirin 1 mg 4day
* C4H
# Fatigue, una!le to perform normal activities
-
8/13/2019 Fundamentals of Cardiac Devices
32/86
32
$tat*s Check
# +atientPs stress test indicates"
* Stopped after ) minutes for fatigue
* EC. immediately after stress test" Clic- for Anser
:i-ely a pacema-er, as the patient has chronotropic
incompetence * her heart rate does not increase ith e'ercise
Bhat device
system ould you
recommendK
=S%, %ate 7
-
8/13/2019 Fundamentals of Cardiac Devices
33/86
3333
+asic Conce!ts,lectricity andPacemakers
-od*le
-
8/13/2019 Fundamentals of Cardiac Devices
34/86
3)3)
Objectives
9pon completion you ill !e a!le to"# Descri!e the relationship !eteen voltage, current, and
resistance
# Descri!e the clinical significance of alterations in voltage,
current, and resistance
-
8/13/2019 Fundamentals of Cardiac Devices
35/86
33
Characteristics of an electrical circ*it/Incl*din# a !acemaker circ*it
# ;oltage# Current
# $mpedance
-
8/13/2019 Fundamentals of Cardiac Devices
36/86
35
olta#e
# ;oltage is the force, or ?push,@ that causes electrons tomove through a circuit
# $n a pacing system, voltage is"
* Measured in volts /;0
* %epresented !y the letter ?;@
* +rovided !y the pacema-er !attery
* Hften referred to as amplitude or pulse amplitude
-
8/13/2019 Fundamentals of Cardiac Devices
37/86
37
C*rrent
# The flo of electrons in a completed circuit# $n a pacing system, current is"
* Measured in milliamps /mA0
* %epresented !y the letter ?$@
* Determined !y the amount of electrons that move through a circuit
-
8/13/2019 Fundamentals of Cardiac Devices
38/86
31
Im!edance
# The opposition to current flo# $n a pacing system, impedance is"
* Measured in ohms /)
* %epresented !y the letter ?%@
* The measurement of the sum of all resistance to the flo of current
-
8/13/2019 Fundamentals of Cardiac Devices
39/86
36
olta#e0 C*rrent0 and Im!edance are Interde!endent
# The interrelationship of the three components is analogousto the flo of ater through a hose
* ;oltage represents the force ith hich
* Current /ater0 is delivered through
*A hose, here each component represents the total impedance"
# The no&&le, representing the electrode
# The tu!ing, representing the lead ire
lt C t d I d
-
8/13/2019 Fundamentals of Cardiac Devices
40/86
))
olta#e0 C*rrent0 and Im!edanceReca!
# ;oltage" The force moving the current /;0* $n pacema-ers it is a function of the !attery chemistry
# Current" The actual continuing volume of flo of electricity /$0
* This flo of electrons causes the myocardial cells to depolari&e /to
?!eat@0
# $mpedance" The sum of all resistance to current flo /% or
or sometimes )
* $mpedance is a function of the characteristics of the conductor /ire0,the electrode /tip0, and the myocardium
lt d C t 1l
-
8/13/2019 Fundamentals of Cardiac Devices
41/86
)
olta#e and C*rrent 1lo&lectrical 2nalo#ies
Spigot /voltage0 turned up, lots of
ater flos /high current drain0
Spigot /voltage0 turned lo, little flo
/lo current drain0
Bater pressure in system
is analogous to voltage *
providing the force to
move the current
R i t d C t 1l
-
8/13/2019 Fundamentals of Cardiac Devices
42/86
)2
Resistance and C*rrent 1lo&lectrical 2nalo#ies
# =ormal resistance* friction caused !y the hose and no&&le
More ater discharges, !ut is all of it going to
the no&&leK
#
-
8/13/2019 Fundamentals of Cardiac Devices
43/86
)3
Ohm3s "a&
# Descri!es the relationship
!eteen voltage, current,
and resistance
#; Q $ %
# $ Q ; 4 %
# % Q ; 4 $
II
R
I
R
I
R
I Q
Q
Q
-
8/13/2019 Fundamentals of Cardiac Devices
44/86
))
Ohm3s la& tells *s/
$f the impedance remains constant, and the voltagedecreases, the current decreases
2 $f the voltage is constant, and the impedance decreases,
the current increases
So BhatK
-
8/13/2019 Fundamentals of Cardiac Devices
45/86
))
$tat*s Check
# Start ith"
* ;oltage Q ;
* $mpedance Q
* Current Q mA
# Solve for Current /$0"
* $ Q ;4%
* $ Q ; U Q Amps
* Current is mA
# %educe the voltage to 2 ;
* ;oltage Q ;
* $mpedance Q
* Current Q K
# $s the current increased4
decreased or unchangedK
* $ Q ;4%
* ; Q 2 ; U Q
Amps or mA
# he c*rrent is red*ced
Bhat happens to current if the voltage is reduced !ut the
impedance is unchangedK
-
8/13/2019 Fundamentals of Cardiac Devices
46/86
)5)5
$tat*s Check
# Start ith"
* ;oltage Q ;
* $mpedance Q
* Current Q mA
# Solve for Current /$0"
* $ Q ;4%
* $ Q ; U Q Amps
* Current is mA
# %educe impedance to 2
* ;oltage Q ;
* $mpedance Q 2
* Current Q K
# $s the current increased4
decreased or unchangedK
* $ Q ;4%
* ; Q 2 ; U 2 Q
2 Amps or 2 mA
# he c*rrent is increased
Bhat happens to current if the impedance is reduced
!ut the voltage is unchangedK
-
8/13/2019 Fundamentals of Cardiac Devices
47/86
)7)7
Other terms
#Cathode" A negativelycharged electrode
* For e'ample, the electrode on
the tip of a pacing lead
# Anode" A positivelycharged electrode
* E'amples"
# The ?ring@ electrode on a !ipolar lead
# The $+. case on a unipolar system
* More on this later /see"
+acema-er >asics0
2node
Cathode
+attery +asics
-
8/13/2019 Fundamentals of Cardiac Devices
48/86
)1)1
+attery +asics$o &here does the c*rrent come from4
#A !attery produces electricity as a result of a chemicalreaction $n its simplest form, a !attery consists of"
*A negative electrode /anode0
*An electrolyte, /hich conducts ions0
*A separator, /also an ion conductor0 and
*A positive electrode /cathode0
=egative terminal
Anode
Separator
Cathode
+ositive terminal
-
8/13/2019 Fundamentals of Cardiac Devices
49/86
)6)6
2!!lyin# lectrical Conce!ts toPacemakers
-od*le 5
-
8/13/2019 Fundamentals of Cardiac Devices
50/86
Objectives
# 9pon completion you ill !e a!le to"* %ecogni&e a high impedance condition
* %ecogni&e a lo impedance condition
* %ecogni&e capture threshold
* Determine hich sensitivity value is more
/or less0 sensitive
-
8/13/2019 Fundamentals of Cardiac Devices
51/86
lectrical Information
# Bhy is this electrical information relevantK
# A pacema-er is implanted to"
* +rovide a heart rate to meet meta!olic needs
# $n order to pace the heart, it must capturethe myocardium
# $n order to pace the heart, it must -no hen to pace, ie, it must !e a!le to
sense
# A pacema-er reLuires an intact electrical circuit
Ohm3s "a&
-
8/13/2019 Fundamentals of Cardiac Devices
52/86
2
Ohm s "a&Relevance to Pacemaker Patients
#
-
8/13/2019 Fundamentals of Cardiac Devices
53/86
33
e ect o ead e o a ce o
-yocardial Ca!t*re
Bhat ould you e'pect to happen if a lead as partially
fracturedK
# $mpedance /or %esistance0 ould rise
# Current ould decrease and !attery energy conserved
( !ut (
Could you guarantee that enough current /$0 can flo
through this fractured lead so that each time the
pacema-er fired the myocardium ould !eatK
%i#h Im!edance Conditions
-
8/13/2019 Fundamentals of Cardiac Devices
54/86
)
Lead wire fracture
Increased resistance
%i#h Im!edance Conditions2 1ract*red Cond*ctor
# A fractured ire can cause
$mpedance values to rise
* Current flo from the !attery
may !e too lo to !e effective
# $mpedance values maye'ceed 3,
Hther reason for high
impedance" :ead not seated
properly in pacema-er header
-
8/13/2019 Fundamentals of Cardiac Devices
55/86
"ead Im!edance al*es Chan#e as a Res*lt of/
# Bire fractures
# $nsulation !rea-s
Typically, normal impedance reading values range from 3to ,
* Some leads are high impedance !y design These leads ill
normally sho impedance reading values greater than , ohms
# Medtronic
-
8/13/2019 Fundamentals of Cardiac Devices
56/86
5
"o& Im!edance Conditions
# $nsulation !rea-s e'pose the
lead ire to the folloing* >ody fluids, hich have a lo
resistance, or
* Another lead ire /in a !ipolar
lead0
# $nsulation !rea- that e'poses a
conductor causes the folloing
* $mpedance values to fall
* Current to drain through the
insulation !rea- into the !ody, or
into the other ire
* +otential for loss of capture
* More rapid !attery depletion
Current ill follo the path of
:EAST resistance
-
8/13/2019 Fundamentals of Cardiac Devices
57/86
7
Ca!t*re hreshold
# The minimum electrical stimulus needed to consistently capture the
heart outside of the heartPs on refractory period
;entricular pacema-er 5 ppm
Capture Non-Capture
-
8/13/2019 Fundamentals of Cardiac Devices
58/86
1
ffect of "ead Desi#n on Ca!t*re
# :ead maturation
* Fi!rotic ?capsule@ develops around the electrode folloing lead
implantation
* May gradually raise threshold
* 9sually no measura!le effect on impedance
-
8/13/2019 Fundamentals of Cardiac Devices
59/86
6
$teroid l*tin# "eads
# Steroid eluting leads
reduce the inflammatory
process
* E'hi!it little to no acute
stimulation threshold pea-ing
* :eads maintain lo chronic
thresholds
Silicone ru!!er plugcontaining steroid
Tines for
sta!lefi'ation
+orous, platini&ed tip
for steroid elution
-
8/13/2019 Fundamentals of Cardiac Devices
60/86
5
ffect of $teroid on $tim*lation hresholds
+ulse Bidth Q msec
$mplant Time /Bee-s0
Te'tured Metal Electrode
Smooth Metal Electrode
Steroid(Eluting Electrode
2
;olts
References/ +acing %eference .uide, >a--en Education Center, 66, 9C665)7aE= Cardiac +acing,2nd Edition, Edited !y Wenneth A Ellen!ogen 665
-
8/13/2019 Fundamentals of Cardiac Devices
61/86
5
-yocardial Ca!t*re
# Capture is a function of"
*Amplitudethe strength of the impulse e'pressed in volts
# The amplitude of the impulse must !e large enough to cause depolari&ation /ie, to ?capture@
the heart0
# The amplitude of the impulse must !e sufficient to provide an appropriate pacing safety
margin
* +ulse idththe duration of the current flo e'pressed in
milliseconds
# The pulse idth must !e long enough for depolari&ation to disperse to the surrounding tissue
Com!arison
-
8/13/2019 Fundamentals of Cardiac Devices
62/86
52
Com!arison6.7 olt 2m!lit*de at Different P*lse Widths
Amplitude
;
2 ms ms ms
-
8/13/2019 Fundamentals of Cardiac Devices
63/86
53
he $tren#th'D*ration C*rve
# The strength(duration
curve illustrates the
relationship of
amplitude and pulse
idth
*Any com!ination of
pulse idth and voltage,
on or a!ove the curve,
ill result in capture
+ulse Bidth
.67
8.7
8.6
9.7
.96
;olts
7.6 8.7 8.6
Capture
=o Capture
%heo!ase
Chrona'ie
-
8/13/2019 Fundamentals of Cardiac Devices
64/86
5)
D*ration
P*lse Width (ms)
Clinical :tility of the $tren#th'D*ration C*rve
# >y accurately determining
capture threshold, e canassure adeLuate safety margins
!ecause"
* Thresholds may differ in acute or
chronic pacing systems
* Thresholds fluctuate slightly daily
* Thresholds can change due to
meta!olic conditions or
medications
7.6 8.7 8.6
.67
8.7
8.6
9.7
.96$tim
*lationhresh
old(olts)
+rogrammed Hutput
-
8/13/2019 Fundamentals of Cardiac Devices
65/86
5
Pro#rammin# O*t!*ts
# +rimary goal" Ensure patient safety and appropriate device
performance
# Secondary goal" E'tend the service life of the !attery
* Typically program amplitude to J 2 ;, !ut alays maintain
adeLuate safety margins# A common output value might !e 2 ; at ) ms
*Amplitude values greater than the cell capacity of the pacema-er
!attery /usually a!out 21 ;0 reLuire a voltage multiplier, resulting in
mar-edly decreased !attery longevity
-
8/13/2019 Fundamentals of Cardiac Devices
66/86
5555
Pacemaker $ensin#
# %efers to the a!ility of the pacema-er to ?see@ signals
* E'pressed in millivolts /m;0
# The millivolts /m;0 refers to the si&e of the signal the
pacema-er is a!le to ?see@
m; signal
2 m; signal
$ensitivity
-
8/13/2019 Fundamentals of Cardiac Devices
67/86
57
yhe al*e Pro#rammed into the IP;
ime
6.7 m
9.6 m
8.96 m
$ensitivity
-
8/13/2019 Fundamentals of Cardiac Devices
68/86
51
Time
m;
2 m;
2 m;
m; sensitivity
At this value the pacema-er ill not
see the 3 m; signal
yhe al*e Pro#rammed into the IP;
$ensitivity
-
8/13/2019 Fundamentals of Cardiac Devices
69/86
56
yhe al*e Pro#rammed into the IP;
At this value, the pacema-er can see !oth the 3 m; and the
3 m; signal So, is ?more sensitive@ !etter, !ecause the
pacema-er sees smaller signalsK
2 m; Sensitivity
Time
m;
2 m;
2 m
>ut hat
a!out thisK
-
8/13/2019 Fundamentals of Cardiac Devices
70/86
7
$ensin# 2m!lifiers
-
8/13/2019 Fundamentals of Cardiac Devices
71/86
7
$ensin# 2cc*racy
# Affected !y"
* +acema-er circuit /lead0 integrity
# $nsulation !rea-
# Bire fracture
* The characteristics of the electrode
* Electrode placement ithin the heart
* The sensing amplifiers of the pacema-er
* :ead polarity /unipolar vs !ipolar0
* The electrophysiological properties of the myocardium
* EM$ * Electromagnetic $nterference
"ead Cond*ctor Coil Inte#rity
-
8/13/2019 Fundamentals of Cardiac Devices
72/86
72
2ffect on $ensin#
# 9ndersensing occurs hen the cardiac signal is una!le to
get !ac- to the pacema-er
* $ntrinsic signals cannot cross the ire fracture
# Hversensing occurs hen the severed ends of the ire
intermittently ma-e contact* Creates signals interpreted !y the pacema-er as +( or %(aves
"ead Ins*lation Inte#rity
-
8/13/2019 Fundamentals of Cardiac Devices
73/86
73
2ffect on $ensin#
# 9ndersensing occurs hen inner and outer conductor coils
are in continuous contact
* Signals from intrinsic !eats are reduced at the sense amplifier, and
amplitude no longer meets the programmed sensing value
# Hversensing occurs hen inner and outer conductor coilsma-e intermittent contact
* Signals are incorrectly interpreted as +( or %(aves
: i l P k
-
8/13/2019 Fundamentals of Cardiac Devices
74/86
7)
:ni!olar Pacemaker
#Bhere is the sensing circuitK
_
Cathode
Anode
:ead tip to canThis can produce a large potential
difference /signal0 !ecause the cathode
and anode are far apart
Clic- for Anser
+i l P k
-
8/13/2019 Fundamentals of Cardiac Devices
75/86
7
+i!olar Pacemaker
# Bhere is the sensing
circuitK
Anode and
Cathode
Clic- for Anser
:ead tip to ring on the lead
This usually produces a smaller potential
difference due to the short inter(electrode
distance#>ut, electrical signals from outside the
heart /such as myopotentials0 are less
li-ely to !e sensed
-
8/13/2019 Fundamentals of Cardiac Devices
76/86
t d ; di t
-
8/13/2019 Fundamentals of Cardiac Devices
77/86
7777
ectors and ;radients
Sense
The ave of depolari&ation produced !y
normal conduction creates a gradient
across the cathode and anode This
changing polarity creates the signal
Hnce this signal e'ceeds the
programmed sensitivity * it is
sensed !y the device
Clic- for More
2 m;
Ch i th t
-
8/13/2019 Fundamentals of Cardiac Devices
78/86
7171
Chan#in# the ector
Sense
2 m;
A +;C occurs, hich is conducted
a!normally Since the vector relative
to the lead has changed, hat effect
might this have on sensingK
Clic- for More
$n this case, the ave of
depolari&ation stri-es the anode and
cathode almost simultaneously This
ill create a smaller gradient and
thus, a smaller signal
P*ttin# It 2ll o#ether
-
8/13/2019 Fundamentals of Cardiac Devices
79/86
7676
P*ttin# It 2ll o#ether
# Appropriate output programming can improve device longevity
* >ut, do not compromise patient safetyX# :ead design can improve device longevity via
* Steroid eluting leads
# Can help -eep chronic pacing thresholds lo !y reducing inflammation and scarring
*atteries, circuit !oards, capacitors, etc, specificto needs, can lead to
improved efficiencies and loered static current drain
*
-
8/13/2019 Fundamentals of Cardiac Devices
80/86
11
P*ttin# It 2ll o#ether
# +acema-er :ongevity is"
*A function of programmed parameters /rate, output, time pacing0
*A function of useful !attery capacity
*A function of
# Static current drain
# Circuit efficiency
# Hutput $mpedance
# The loer the programmed sensitivity the MH%E sensitive
the device
* :ead integrity also affects sensing
$tat*s Check
-
8/13/2019 Fundamentals of Cardiac Devices
81/86
11
$tat*s Check
# Determine the threshold amplitude
Capture threshold Q loest value ith consist capture
This is at 2 ;
Clic- for Anser
2 ; ;7 ; ;
$tat*s Check
-
8/13/2019 Fundamentals of Cardiac Devices
82/86
1212
$tat*s Check
H%
Clic- for Anser
+acema-er A is a!le to ?see@ signals as small as m; Thus, it is
more sensitive
+rogrammed
Sensitivity m;
+acema-er
A
+rogrammed
Sensitivity 2 m;
+acema-er
>
# Bhich of these pacema-ers is more sensitiveK
$tat*s Check
-
8/13/2019 Fundamentals of Cardiac Devices
83/86
1313
$tat*s Check
# A pacema-er lead must fle' and move as the heart !eats
Hn average, ho many times does a heart !eat in yearK
Clic- for Anser
3 M$::$H= times $t is not a simple tas- to design a
lead that is small, relia!le, and lasts a lifetime
$tat*s Check
-
8/13/2019 Fundamentals of Cardiac Devices
84/86
1)1)
$tat*s Check
Clic- for Anser
:ead Fracture"
#
-
8/13/2019 Fundamentals of Cardiac Devices
85/86
11
What &o*ld yo* e=!ect4
# Bhich value is out of rangeK
# Bhat could have caused thisK
+acema-er $nterrogation %eport
Mode" DDD%
:oer" %ate 5 ppm
9T%" 3 ppm
9S%" 3 ppm
Atrial :ead $mpedance" )7 Hhms
;entricular :ead $mpedance" 6 Hhms
$nsulation failure
Clic- for Anser
-
8/13/2019 Fundamentals of Cardiac Devices
86/86
hank >o*