First Impressions on the
Agilent 7200 GC / Q-TOF
Frank David
RIC and Ghent University
Mass Spectroscopy
Structure
Elucidation
Selective and
Sensitive Detection
Identification Target Compound
Analysis
Mass Spectroscopy
Structure
Elucidation
Selective and
Sensitive Detection
Identification
(NIST, Wiley)
Target Compound
Analysis
Mass Spectroscopy
Structure
Elucidation Selective and
Sensitive Detection
Identification: accurate mass
“Open Methods”
Target Compound
Analysis by MS/MS
The (GC) MS compromise triangle
Selectivity
(Resolution) Mass Accuracy
(ID) Robustness
Sensitivity
“Scan” versus “SIM” (MRM)
QQQ
(Q)TOF
SQ (scan)
S/N
n (solutes)
SQ (SIM)
Applications & Expectations
• Analysis of pesticide mixture – Sensitivity check on aldrin (organochlorine pest)
– Searchable spectrum at low pg level?
• Analysis of urinary metabolites in kidney disease study – Compare with MSD
• Analysis of Biomarkers in Crude Oil – Resolution, extract accurate mass chromatograms
– MS/MS operation
Analysis by GC-MSD (5975C) - SCAN 10 pg on-column - EIC (262.8)
Aldrin: 15.60 min
1 2 .0 0 1 3 .0 0 1 4 .0 0 1 5 .0 0 1 6 .0 0 1 7 .0 0 1 8 .0 0 1 9 .0 00
1 0 0
2 0 0
3 0 0
4 0 0
5 0 0
6 0 0
7 0 0
8 0 0
9 0 0
1 0 0 0
1 1 0 0
1 2 0 0
1 3 0 0
1 4 0 0
1 5 0 0
1 6 0 0
1 7 0 0
1 8 0 0
1 9 0 0
2 0 0 0
2 1 0 0
2 2 0 0
2 3 0 0
T im e -->
A b u n d a n c e
Io n 2 6 2 .8 0 (2 6 2 .5 0 to 2 6 3 .5 0 ): s q te s t-1 4 .D \d a ta .m s
1 5 .6 0 2
1 6 .1 3 4
1 6 .3 7 5
1 7 .5 1 0
1 7 .9 0 2
Analysis by GC-MSD (5975C) - SCAN MS (background substrated) – NIST search
Scan 1047 (15.602 min): sqtest-14.D\ data.ms (-1045) 1,4:5,8-Dimethanonaphthalene, 1,2,3,4,10,10-hexachloro-1,4,4a,5,8,8a-hexahydro-, (1à,4à,4aá,5à,8à,8aá)-Head to Tail MF=449 RMF=523
50 70 90 110 130 150 170 190 210 230 250 270 290 310 330 350 370 390 410 430 450 470 490
0
50
100
50
100
51
64
66
73
79
81
91
91
96 116
121
143
143
159
160186
191
201
207
216
220
235
237
248
255
263
263
293
298 324
329
341 355
364
390428 450462 480492
Hit quality: match = 449; Rmatch = 523; Prob = 28.9
Analysis by GC-QQQ (7000B) - SCAN 10 pg on-column - EIC (262.8)
Aldrin: 17.93 min
Analysis by GC-QQQ (7000B) - SCAN MS (background substrated) – NIST search
+EI Scan (17.934 min) test04.D Subtract 1,4:5,8-Dimethanonaphthalene, 1,2,3,4,10,10-hexachloro-1,4,4a,5,8,8a-hexahydro-, (1α,4α,4aβ,5α,8α,8aβ)-Head to Tail MF=612 RMF=635
50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 360 370 380 390 400 410 420 430 440 450 460
0
50
100
50
100
50
51
55
66
66
75
79
84
85
91
92
97
101
101
110
111
127
133
147
149 158
160
176
186
191203
205
207
220
227
229
241
257
263
265
276
281 293
293
314
315
327
329
341 356
364
366 405 417 452
Hit quality: match = 612; Rmatch = 635; Prob = 67.9
Analysis by GC-QTOF (7200) 10 pg on-column - EIC (262.8564)
Aldrin: 18.53 min S/N = 147
Analysis by GC-QTOF (7200) 10 pg on-column - Spectrum
Analysis by GC-QTOF (7200) 1 pg on-column - EIC (262.8564)
GC-MS based metabolomics
• Agilent Fiehn GC/MS Metabolomics RTL Library (G1676AA)
• A retention time locked (RTL) GC-MS method on a 7890A GC with 5975C inert XL MSD
• Library containing searchable EI spectra and retention times from appx. 700 common derivatized metabolites
• DB5-MS (30 m x 250 µm x 0.25 µm) • 60°C - 1 min - 10°C/min - 325°C • Splitless, 250°C • CF, 1.1 mL/min • MS in scan (50-600 amu)
Data handling: differential analysis Healthy
Disease
Feature extraction
Alignment, normalization and filtering
(MassProfilerPro)
H1
H2
H3
H4
H5
D1
D2
D3
D4
D5
Data handling: differential analysis
Feature extraction
Alignment, normalization and filtering
(MassProfilerPro)
PC1
PC2
PC3
Statistical analysis (MassProfilerPro)
GC-MSD analysis of Urine sample Fiehn method (oximation – silylation)
8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0 2 0 . 0 0 2 2 . 0 0 2 4 . 0 0 2 6 . 0 0
5 0 0 0 0 0
1 0 0 0 0 0 0
1 5 0 0 0 0 0
2 0 0 0 0 0 0
2 5 0 0 0 0 0
3 0 0 0 0 0 0
3 5 0 0 0 0 0
4 0 0 0 0 0 0
4 5 0 0 0 0 0
5 0 0 0 0 0 0
T i m e - - >
A b u n d a n c e
T I C : 1 1 0 4 2 0 _ C K D 3 - 1 6 . D \ d a t a . m s
CKD3
GC-QTOF: diluted 1/10 versus MSD
Healthy versus CKD3
From differential analysis
First hit: Formula (M): C10 H16 O Si Mass (MFG): 180.097 m/z (Calc): 180.0965 Diff (ppm): 2.77 = p.cresol (TMS der)
Analysis of Biomarkers in Crude Oil
• Biomarkers: more stable against weathering
• Include: (heterocyclic) PAHs, hopanes, steranes (can be > 100 analytes)
• Analysed for oil characterization, oil spillage identification,…
• Classical method: fractionation (SPE, HPLC) + GC-MS (EI) using multiple SIM methods
Agilent 7200 GC/Q-TOF conditions
• 1 µL splitless injection of 1 mg/mL oil sample
• 30 m x 0.25 mm ID x 0.25 µm DB-5MS
• 1.5 mL/min helium, constant flow
• 50°C (1 min) – 10°C/min – 320°C (8 min)
• MS:
– 40 – 500 Da acquisition, 5 Hz
– Source 280°C, quad: 150°C
PAH mix with dibenzothiophene (0.38 ng OC)
NIST 2260 mixture DBT = C12H8S 184.0341
Number 1 hit = DBT
Analysis of DBT – Mass Accuracy
Mass error: - 1.09 ppm 184.0339
April 8, 2012 27
Analysis of Biomarkers in Crude Oil
April 8, 2012 28
Analysis of Biomarkers in Crude Oil
Mass extraction window: +/- 0.5 amu Equivalent to MSD
EIC 184
Interferences
dibenzothiophene
April 8, 2012 29
Analysis of Biomarkers in Crude Oil
Mass extraction window: +/- 5 ppm
EIC 184.0341 dibenzothiophene
April 8, 2012 30
Analysis of DBT – Mass Accuracy (in matrix)
Mass error in matrix: - 1.63 ppm
184.0338
DBT, mono- and dimethyl DBTs
184.0341
198.0498
212.0645
Hopanes
(mainlib) 28-Nor-17 a( H)-hopane
100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 360 370 380 390 400 410
0
50
100
109 123
135
149
163
177
191
205
217
245 260
355 383 398
Analysis of Hopanes in Crude Oil
191 +/- 0.5
191.1794 +/- 10 ppm
Interferences hopanes
hopanes
Confidentiality Label
April 8, 2012 34
Analysis of Steranes in Crude Oil
EIC 217.1951
EIC 400.4064
Confidentiality Label
April 8, 2012 35
Steranes in Crude Oil by GC-Q-TOF - MS/MS mode
MS: EIC 400.4064
MS/MS: 400 > 217.1951
Conclusions
• Mass accuracy and resolution: excellent
• Sensitivity: higher than expected
• Stability: excellent
• Outlook:
– Excellent tool for compound ID (“beyond EI MS library search”)
– “open methods” – “retrospective analysis”