evaluation of 6-year application of the enzymatic colorimetric phenylalanine assay in the setting of...
TRANSCRIPT
Evaluation of 6-year application of the enzymatic colorimetric
phenylalanine assay in the setting of neonatal screening
for phenylketonuria1
Andreas Schulze *, Ertan Mayatepek, Georg F. Hoffmann
Division of Metabolic and Endocrine Diseases, University Children’s Hospital, Im Neuenheimer Feld 150, 69120 Heidelberg, Germany
Received 19 June 2001; accepted 5 September 2001
Abstract
Background: Most reports on phenylketonuria (PKU) screening focused solely on the result of the initial investigation of
the neonatal screening sample. The aim of this study was to evaluate an enzymatic phenylalanine (Phe) determination in the
whole context spanning from the initial investigation over the recall period, up to the confirmation or exclusion of the disease.
Methods: Phe of dried blood spot specimens was analysed colorimetrically in a microtitre-plate assay based on the L-
phenylalanine dehydrogenase reaction coupled with an intermediate electron acceptor system. This assay was evaluated for
analytical variables and for neonatal PKU screening in a total number of 423,773 neonates during a 6-year period. Results:
Method validation with respect to linearity, precision (within-run CVs 3.4–4.2%, between-run CVs 6.2–10.4%), and accuracy
fulfilled all requirements for a screening method. Mean Phe (F SD) of 130,000 healthy neonates was 84 (F 22) mmol/l with a
cut-off point (mean + 3 SD) of 150 mmol/l. From 423,773 neonates, hyperphenylalaninemia was confirmed in 155 cases and
further differentiated into PKU (41 cases, 27%), BH4 deficiency (3, 2%), non-PKU HPA (67, 43%), transient neonatal HPA
(28, 18%), and secondary HPA (16, 10%). The number of false-positives (recall-rate) was 0.23%, and no false-negatives were
noted. Conclusions: Detailed studies over a period of 6 years including more than 400,000 neonates clearly show that the
enzymatic assay is a reliable and sensitive method for neonatal screening of PKU. The proven prevalence of non-PKU HPA in
the German population disclosed by the assay was twice as high as compared to the ‘‘Guthrie test’’ used previously. The
growing use and application of tandem mass spectrometry in neonatal screening will not derogate the usefulness of the
enzymatic assay in PKU screening in the foreseeable future. Careful analysis of our screening results and monitoring of all
pathological samples resulted in an evidence-based flow chart for a rational PKU screening. D 2002 Elsevier Science B.V. All
rights reserved.
Keywords: Phenylketonuria; Phenylalanine determination; Screening
1. Introduction
More than 30 years after the inauguration of the
bacteriological inhibition assay (BIA) for phenylala-
nine (Phe) measurements by Guthrie and Susi [1],
0009-8981/02/$ - see front matter D 2002 Elsevier Science B.V. All rights reserved.
PII: S0009-8981 (01 )00736 -7
* Corresponding author. Fax: +49-6221-563-714.
E-mail address: andreas�[email protected]
(A. Schulze).
www.elsevier.com/locate/clinchim
1 In memoriam Horst Bickel.
Clinica Chimica Acta 317 (2002) 27–37
the starting point for neonatal screening of phenyl-
ketonuria (PKU) and other inherited metabolic dis-
eases, different methods for detecting PKU are used
in the screening laboratories throughout the world.
All those methods overcame the main disadvantage
of the BIA, its visual interpretation, and were devel-
oped in order to improve precision, sensitivity, prac-
ticability, and running time of screening assays.
These different methods include fluorometric [2,3],
HPLC [4], enzymatic colorimetric [5], and more re-
cently tandem mass spectrometric [6] applications for
the determination of Phe in dried blood spot speci-
mens (DBS).
The quality of the screening process mainly de-
pends on the methodology and consequent tracking,
which means the individual follow-up of each ab-
normal result. However, the latter is hard to achieve
by the screening laboratory because of insufficient
legal regularisation for feedback from the responsi-
ble physicians, thus often leading to loss of informa-
tion on the further fate of patients. Therefore, most
reports about PKU screening methods were focused
solely on the result of the initial investigation of the
neonatal screening sample. The aim of this study
therefore was to evaluate an enzymatic colorimetric
method for Phe determination in the whole context
spanning from the initial investigation over the recall
period, up to the confirmation or exclusion of the
disease.
About 10 years ago, Wendel et al. [7] described a
colorimetric method for the determination of plasma
Phe using L-Phe dehydrogenase (L-PheDH; EC 1.4.1.-)
coupled with an intermediate electron acceptor system
(Fig. 1) and adapted this method as a microtitre-plate
assay for DBS [8]. Further improvement of specificity
with respect to cross reactivity towards tyrosine (Tyr)
was achieved by elevation of pH to 10.8 in the L-
PheDH reaction [9]. Upon this basis, Porton Cam-
bridge launched a commercial Phe screening kit
(QuantasekR) applicable for neonatal PKU screening
[10–12].
In 1994, after the investigation of more than 1.5
million neonates with the BIA in the past 25 years, our
laboratory which is responsible for the regional neo-
natal screening program (Baden-Wurttemberg, Ger-
many) changed their routine screening procedure from
the BIA to the enzymatic Phe determination. Our
report focuses on the evaluation of an enzymatic
colorimetric Phe determination based on the neonatal
screening results from over 400,000 investigations
during a 6-year period leading to an evidence-based
flow chart for rational PKU screening.
Fig. 1. Reaction scheme of the enzymatic Phe determination. Coupling of the L-PheDH reaction with an intermediate electron acceptor (IEA)
system for the colorimetric measurement.
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–3728
2. Materials and methods
2.1. Reagents and materials
All reagents were purchased from Sigma or
Merck. The QuantasekR Phe screening assay, the
QuantasekR Phe monitoring assay, 96-well Milli-
pore microfilter-plates (PCBH034), and the Vacuum
Manifold (PCNH001) were purchased from Quan-
tase (Perth, Scotland, UK). Filter paper, S&S 2992,
was used from Schleicher & Schuell (Dassel, Ger-
many).
2.2. Calibration materials
With the beginning of 1998, the European Work-
ing Standard [13] was used as reference material for
Phe calibrators. These blood spot calibrators were
purchased from Quantase. Before that date calibra-
tion materials were prepared by using whole blood
from healthy adult blood donors. Phe concentrations
were measured by means of the enzymatic plasma
method mentioned below and thereafter spiked with
a stock solution of 6 mmol/l to final Phe concentra-
tions of 0.120, 0.240, 0.480 and 1.2 mmol/l, respec-
tively. Thirty microliters of the calibration materials
was spotted onto filter paper, air-dried over night and
stored desiccated at 3–5 �C up to 1 month, at the
latest.
Four 4.25-mm punches of each blood spot calibra-
tor were measured in duplicate every day for calcu-
lation of the calibration curve.
2.3. Enzymatic Phe screening assay
4.25-mm discs of blood spot calibrators, of two
different internal Phe blood spot standards as well
as of the DBS from neonates, were punched out
into 96-well Millipore microfilter-plates, incubated
with 60 ml TCA 3% (w/v) and agitated for 60 min
at RT. Eluates were then transferred to microtitre-
plates using a vacuum manifold. One hundred
microliters of working enzyme (L-PheDH)/coenzyme
(NAD + ) reagent was added with a 12-channel
pipette to each well. After incubation for 30 min
at ambient temperature 100 ml colour reagent (citratebuffered solution of tetrazolium salt, intermediate
electron acceptor, and detergent) was added. Absorp-
tion at 570/690 nm was measured after 2 min on a
Multiscan MCC 349 (Labsystems, Finland) micro-
titre-plate reader.
2.4. Further methods for phenylalanine determina-
tion
In plasma, enzymatic determination of Phe was
performed with the QuantasekR Phe monitoring
assay. In this assay, the NADH production coupled
to a tetrazolium/intermediate electron acceptor detec-
tion system was measured at 570 nm with a Cobas Bio
(Hoffmann-LaRoche, Grenzach-Villen, Switzerland)
in 20 ml plasma after deproteinisation with 0.6 mol/l
perchloric acid.
Determination of plasma Phe by automated amino-
acid analysis was performed using a Biotronic 3000
analyser (Biotronic, Munich, Germany) following
standard procedures [14].
Phe, Tyr, and the Phe/Tyr-ratio in DBS were
measured by means of electrospray tandem mass
spectrometry (ESI-MS/MS) as already described in
detail [6].
2.5. Calculations and statistics
Blood Phe concentration was calculated from the
calibration curve obtained by linear regression analy-
sis (least-square method). For that purpose, the on-line
data were transferred in the Porton Cambridge Data
Management SystemR producing the report as a
Microsoft ExcelR file.
Data are expressed as the meanF SD. Statistical
significance was determined using Student’s t-test.
Differences with P values < 0.01 were considered
significant.
2.6. Dried blood spot specimens
DBS of all neonates (n= 423,773) born between
1994 and 1999 were investigated within the scope of
the regional neonatal screening program (Baden-
Wurttemberg, Germany). The blood regularly
obtained by a heelstick on the fifth day of life (range
1–10, for detailed distribution see Fig. 4) was spotted
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–37 29
onto filter paper, allowed to dry and sent per post mail
to our laboratory.
3. Results and discussion
3.1. Method validation (analytical variables)
3.1.1. Calibration curves, linearity, and limit of detec-
tion
For the study of the calibration curve, DBS of 98,
282, 568, and 1077 mmol/l Phe were measured four-
fold on 20 days. Results are shown in Fig. 2. Analysis
of the daily obtained calibration curve over a period of
4 years revealed a slope of 0.144F 0.053 and inter-
cept of 0.014F 0.003 (meanF SD, respectively).
The detected absorbance for the calibrators was
linear in the range of 98–1077 mmol/l (r2 = 0.974).
Limit of detection was determined as the concen-
tration corresponding to a signal 3 SD above the
mean for a calibrator free of analyte. This limit of
detection for Phe in DBS was 43 mmol/l, highlight-
ing the limitation of the method in the lower Phe
range. However, this is not of any relevance for
neonatal screening, but must be taken into account
for potential use in the monitoring of long-term PKU
treatment.
3.1.2. Precision
We assessed the precision of the whole Phe screen-
ing assay according to NCCLS EP5-T2 (Wayne, PA:
National Committee for Clinical Laboratory Stand-
ards, June 1984). Estimates of within-run and bet-
ween-run SD were obtained by determination of DBS
with four different Phe concentrations. Two repli-
cates per specimen per run and two runs per day for
20 days were performed. The results are presented in
Table 1.
The CV for the determination of the within-run
precision was between 3.4% and 4.2%. The CV for
the determination of the between-run precision was
between 6.2% and 10.4%.
3.1.3. Analytical recovery and comparison-of-meth-
ods study
For the determination of the accuracy of the Phe
screening assay, Phe recovery in the DBS was ana-
lysed by spiking whole blood with Phe amounts
reaching final Phe concentrations of 121, 485, and
1212 mmol/l of pooled blood, respectively. The
pooled blood was spotted onto filter paper and, in
addition, deproteinised plasma was prepared. Subse-
Fig. 2. Calibration curve of the enzymatic Phe screening assay.
Dried blood spot Phe (DBS) calibrators were measured fourfold on
20 consecutive days. Phe concentrations are plotted as meanF SD.
Calibration curve was obtained by linear regression analysis. Slope
and intercept are expressed as meanF SD. SEE, standard error of
estimate (standard deviation about the regression line).
Table 1
Precision of the enzymatic phenylalanine screening assay
DBS phenylalanine Within-run Between-run
concentration (mmol/l)Concentration (meanF SD),
(mmol/l)
CV
(%)
Concentration (meanF SD),
(mmol/l)
CV
(%)
98 95.5F 3.7 3.8 95.5F 10.0 10.4
282 260.3F 9.0 3.4 260.3F 23.7 9.1
568 600.5F 21.7 3.6 600.5F 44.8 7.5
1077 1065.8F 44.8 4.2 1065.8F 66.2 6.2
DBS: dried blood spot.
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–3730
quently, Phe concentrations were measured in DBS
with the Phe screening assay and by ESI-MS/MS
15 times in the same batch, and in plasma by the
QuantasekR Phe monitoring assay on Cobas Bio
and by automated amino-acid analysis 12 times in the
same batch (Table 2).
The enzymatic Phe determination in DBS re-
vealed a recovery of 89–101% and CVs of 6.8–
11.2%, thus fulfilling the requirements for a screen-
ing method.
Phe determination in DBS by ESI-MS/MS (recov-
ery 106–125%, CVs 6.5–16.9%) revealed increased
recoveries and CVs comparable to the enzymatic
method.
The enzymatic Phe measurement in plasma
showed the best accuracy (recovery 92–109%, CVs
1.1–2.9%), whereas automated amino-acid analysis,
an established method for quantitative plasma amino
acid analysis, was less good as expected (recovery
74–120%, CVs 3.1–14.1%). This was especially true
for the low (121 mmol/l) and medium (485 mmol/l)
Phe concentrations which yielded recoveries of about
80–120% and 74–108% as well as CVs of 10.2% and
14.1%, respectively (Table 2).
3.1.4. Reference interval (normal range) of neonates
Data of a total number of 130,000 healthy neo-
nates were analysed for the determination of the Phe
reference interval, distribution analysis, as well as
Phe concentration in relation to age of sampling.
Mean (F SD) Phe concentration was 84 (F 22)
mmol/l with a median of 83 mmol/l. Cut-off point cal-
culated by mean + 3 SD was 150 mmol/l. Frequency
distribution of Phe values followed a nearly Gaussian
distribution (Fig. 3). The curve only slightly flattens
on the right side.
The Phe concentration related to age of the neo-
nates at the time of blood sampling was investigated
in order to estimate the need of changing the cut-off
value in cases which deviate from the time of
sampling regularly performed on the fifth day of life.
That is of special importance for early discharge of
neonates, because of the well-known elevations of
Phe concentrations immediately post-natally [15]. In
only 2% of our investigated population of 130,000
neonates blood sampling was performed before the
fourth day of life, whereas 5% were investigated after
the sixth day of life (Fig. 4). The mean (F SD) Phe
concentrations on the first and second days of life
were 96 (F 27) and 93 (F 24) mmol/l, respectively,
thus, significantly higher than on the fifth day. From
the third to the 10th day of life concentrations did not
differ significantly from the latter (Fig. 4). Higher
Phe levels on the first and second days resulted in
Fig. 3. Frequency distribution of Phe concentration in DBS from
130,000 neonates (circles). The lined curve represents the ideal
Gaussian distribution.
Table 2
Recoveries of phenylalanine spiked in whole blood. Comparison between different methods of measurement in dried blood spots (DBS) and
plasma
Method Specimen N Recovery mean (range) (%)
121 mmol/l
adjusted (%)
CV
(%)
485 mmol/l
adjusted (%)
CV
(%)
1212 mmol/l
adjusted (%)
CV
(%)
Screening assay DBS 15 94 (75–107) 10.4 89 (69–108) 11.2 101 (90–111) 6.8
ESI-MS/MS DBS 15 106 (87–160) 16.9 125 (102–143) 10.5 125 (108–136) 6.5
Enzymatic, Cobas Bio Plasma 12 104 (95–109) 2.9 94 (92–96) 1.1 100 (96–102) 1.7
Amino Acid Analyzer Plasma 12 99 (80–120) 10.2 83 (74–108) 14.1 99 (94–104) 3.1
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–37 31
2.5% and 0.8% of samples over the cut-off (150
mmol/l), respectively, in contrast to 0.2% on the fifth
day of life.
3.2. Diagnostic outcome of neonatal screening for
PKU
DBS of a total number of 423,773 neonates which
were investigated from 1994 to 1999 in the context of
the neonatal screening program of Baden-Wurttem-
berg, Germany, were included in the study. 93% of
samples were taken on the fourth to the sixth day of life
(Fig. 4). Measurement of Phe from the same DBS was
repeated if Phe concentration was found to be over the
cut-off of 150 mmol/l. This was necessary in 0.78% of
all samples. In case that Phe elevation was confirmed,
either a written recall was sent out (0.26% of all
samples) or when Phe levels were > 360 mmol/l the
sender of the DBS was contacted by phone and asked
for transferal of the neonate to a specialised hospital
(60 neonates, 0.01% of all samples) (Table 3).
Result of all positive tested screening samples,
number and prevalence of confirmed diagnoses as
well as rates of false positives are summarised in
Table 3. A more detailed consideration of the different
diagnosis groups is the content of the following
paragraphs. Quantitative data for all different groups
are shown in (Table 4).
3.3. Phenylketonuria
Phenylketonuria is caused by phenylalanine
hydroxylase (PAH; EC 1.14.16.1) deficiency (McKu-
sick 261600). Following the German recommenda-
tions of PKU treatment [16], patients with Phe >600
mmol/l while on normal diet were classified as PKU.
We detected a total number of 41 neonates suffering
from PKU. The resultant prevalence in our population
was estimated to be 1:10,336. In general, confirma-
tion of PKU was made on the 10th day of life. At that
time, the Phe concentration was twofold increased in
comparison to the screening value (Fig. 5). In 39/41
cases, the initial screening Phe levels were >360
mmol/l and in 35/41 >600 mmol/l. From the six
neonates with Phe < 600 mmol/l, Phe restriction was
Table 3
Results of neonatal screening in 423,773 neonates
True positive False positive False negative
Phe 150–360 mmol/l 95 (8.2%) 995 (86.2%) 0 1090 (94.4%)
Phe>360 mmol/l 60 (5.2%) 4 (0.3%) 0 64 (5.5%)
155 (13.4%) 999 (86.5%) 0 1154 (100%)
Confirmed diagnoses in true
positives (n= 155)
Number Prevalence
PKU 41 1:10,336
BH4 Deficiency 3 1:141,258
Non-PKU HPA 67 1:6325
Transient neonatal HPA 28 1:15,135
Secondary HPA 16 1:26,486
HPA: hyperphenylalaninemia.
Fig. 4. Phe concentration in DBS from 130,000 neonates related to
age of blood sampling. The line plot shows the mean (open circles)
and meanF 3 SD (filled circles) of Phe concentration. ** marks
significant ( p< 0.01) difference of the mean of Phe concentration
compared to that obtained on the fifth day of life. The number of
investigated DBS on the respective days is plotted as columns.
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–3732
started in two of them before screening was taken
because of already known older siblings affected with
PKU. In one other case (screening Phe 480 mmol/l),
DBS was taken on the second day of life, and in two
neonates which were primarily classified as non-PKU
hyperphenylalaninemia (HPA) (screening Phe 310
and 260 mmol/l), a follow-up revealed a ‘‘mild’’
PKU with delayed increase of Phe exceeding 600
mmol/l after 6 weeks and 10 months, respectively.
Only in one neonate (Phe 490 mmol/l on fifth day of
life) did we not find a lucid explanation.
3.4. Tetrahydrobiopterin deficiency (BH4 deficiency)
BH4 is the cofactor of PAH. HPA in the inves-
tigated neonates was caused in a total number of
three patients by a defect in BH4 metabolism. Two of
them suffered from 6-pyrovoyl-tetrahydropterin syn-
thase deficiency (McKusick 261640), and one from
dihydropteridine reductase deficiency (McKusick
261630). Their initial screening Phe levels ranged
between 479 and 596 mmol/l. BH4 deficiency was
suspected because of decrease of Phe levels after oral
BH4 administration. Final diagnoses were made by
enzyme measurements. The proportion of BH4 defi-
ciencies of the proven HPAs accounted for 2.7%,
reflecting a prevalence of about 1:140,000 in our
population.
3.5. Non-PKU hyperphenylalaninemia (mild PKU)
Non-PKU HPA (defined as 150 mmol/l < Phe < 600
mmol/l while on normal diet) due to mild PAH
deficiency was confirmed in 67 cases. The resultant
prevalence of 1:6325 was markedly higher as those
obtained by the BIA (1:14,000, former results of our
screening centre). This is of special importance for
females with regard to the risk of maternal PKU,
which can be prevented by achieving Phe levels < 360
mmol/l preconceptually and during pregnancy [17].
Therefore, it is important to identify women at risk.
Screening Phe levels of the non-PKU HPA group
ranged between 158 and 636 mmol/l (median 246
mmol/l) and remained normally unchanged during
the recalls (Fig. 6). If Phe exceeds levels >240
mmol/l in the first recall, we recommend initiation of
further confirmation tests including exclusion of
BH4 deficiency, instead of requesting further recalls
(Fig. 9).
In three cases with positive first and second recalls,
further confirmation analyses revealed Phe levels of
about 120–150 mmol/l. Long-term Phe-monitoring
performed in monthly intervals, however, confirmed
the diagnosis of non-PKU HPA.
Two neonates, both breast fed, suspected of suf-
fering from PKU because of their screening Phe
levels of 600 and 636 mmol/l, respectively, showed
a decline of their Phe levels already at the time of
their hospitalisation. Further work-up also revealed
non-PKU HPA.
Table 4
Phenylalanine concentration in the different HPA groups at neonatal
screening, during the recall period, and at time of confirmation
Diagnosis Number Age of Sampling Phenylalanine
Median
(range) day
Median (range)
mmol/ l
PKU (n= 41)
Screening 41 5 (2–7) 970 (261–2024)
Confirmation 35a 10 (5–18) 1697 (448–3212)
BH4 Deficiency (n = 3)
Screening 3 5 (5–6) 564 (479–596)
Confirmation 3 17 (14–17) 970 (824–1515)
Non PKU-HPA (n= 67)
Screening 67 5 (1–10) 246 (158–636)
1st Recall 61 15 (7–122) 261 (164–588)
2nd Recall 40 29 (15–180) 245 (170–570)
Confirmation 44a 41 (10–330) 255 (121–564)
Transient neonatal-HPA (n= 28)
Screening 28 5 (1–21) 172 (152–521)
1st Recall 28 17 (5–42) 182 (121–309)
2nd Recall 27 36 (12–90) 133 (61–267)
3rd Recall 13 49 (20–270) 109 (36–145)
Secondary-HPA (n= 16)
Screening 16 6 (5–13) 355 (170–1357)
1st Recall 7b 13 (8–18) 218 (79–539)
2nd Recall 4 29 (23–39) 173 (79–261)
3rd Recall 1 65 145
Healthy
neonates
130,000 5 (1–10) 84.0 (F 22.1)c
aDifference of cases and number of confirmations result from
failed quantitative reports in a part of confirmed cases.b In 9 cases no recall was done because of medical reports
explaining secondary Phe elevation.cMean (F SD).
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–37 33
3.6. Transient neonatal HPA
Increased Phe concentration in 28 DBS was caused
by transient HPA of the neonate. In 27/28 cases, the
screening Phe was 152–194 mmol/l (median 172
mmol/l). Phe in the first recall samples (taken on
17th day of life on average) tended to result in higher
levels, but declined in the further recall samples. After
7 weeks, the Phe levels in all infants were found to be
< 150 mmol/l (Fig. 7). In order to omit unnecessary
Fig. 5. Phe concentration of the PKU group in neonatal screening (DBS) and in confirmation test (plasma). Plasma Phe was measured by means
of different methods used in the respective metabolic centres where the confirmation analyses were made.
Fig. 6. Phe concentration of the non-PKU HPA group in neonatal screening, in recalls (DBS), and in confirmation test (plasma). Plasma Phe was
measured by means of different methods used in the respective metabolic centres where the confirmation analyses were made.
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–3734
recalls, we recommend to perform the first and all
succeeding recalls with an interval of 2 weeks if the
initial screening Phe levels are < 200 mmol/l (Fig. 9).
In one neonate, we measured a high Phe concen-
tration (521 mmol/l) which normalised in subsequent
samples (120 as well as 60 mmol/l, at 16th and 41st
Fig. 7. Phe concentration in DBS of the transient neonatal HPA group in neonatal screening and in recalls.
Fig. 8. Phe concentration versus molar Phe/Tyr-ratio in the different HPA groups. Phe concentrations of all neonatal screening DBS in which
the Phe/Tyr-ratio was available are plotted against the latter. The Phe concentration in transient neonatal HPA was below 210 mmol/l in 17 of
17 cases (horizontal line). Note that the Phe of 521 mmol/l from one infant with transient HPA, discussed in the text, is omitted because of the
missing Phe/Tyr-ratio. The Phe/Tyr-ratio in 16/17 DBS from transient HPA and in 7/7 DBS from secondary HPA was found below 2.7
(vertical line).
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–37 35
day of life, respectively). We could exclude a mix-up
of samples, an error of measurement, and intercurrent
illness of the child which could explain such a unusual
finding.
3.7. Secondary HPA
Increased Phe in neonatal screening not caused by
genetic PAH deficiency was detected in 16 babies. In
them, Phe levels were found to be elevated because of
liver impairment in the neonates due to galactosemia
(in 4 of 16 cases), multi-organ failure (5/16), cardio-
vascular disease (2/16), severe asphyxia (cord pH 6.9)
(1/16), chromosomal aberration (trisomy 18 as well as
21) (2/16), and extreme immaturity (25 as well as 26
weeks of gestational age) (2/16). The wide range of
Phe concentrations within these group (170–1357
mmol/l) disallowed a differential diagnosis of HPA
by the Phe levels. As recently shown by our group, the
use of the Phe/Tyr-ratio obtained by means of ESI-
MS/MS, therefore, is a good aid in the differential
diagnosis [6]. In the DBS of seven neonates with
secondary HPA, their molar Phe/Tyr-ratio could be
estimated and was found to be below 2.7 in all cases
(Fig. 8). In comparison, Phe/Tyr-ratio in PKU ranged
between 6.7 and 47.7, except for two milder cases of
PKU mentioned above, where a Phe/Tyr-ratio of 5.3
as well as 3.4 was found.
3.8. False positives and pitfalls
In 995 of 423,773 DBS, the increased Phe con-
centration (150–360 mmol/l) of the neonatal screening
was found to be < 150 mmol/l in the recall sample,
reflecting a false positive rate of about 0.23%. Phe
determination in four DBS revealed concentrations
>360 mmol/l. Two of them were caused by intra-
venous amino-acid administration (Phe 582 and 988
mmol/l; Phe/Tyr 10.2 and 38.3, respectively). In two
filter cards, a high Phe concentration (679 and 1760
mmol/l, respectively) was traceable in a circumscribed
area of the blood spot only. In both of them, measure-
ments of other punches from the same spot and of
other blood spots from the same filter card revealed
decreased or normal Phe concentrations, respectively.
Thus, the high Phe may have been caused by con-
tamination of the filter card and highlights the need
for internal confirmation tests of abnormal results
before initiating recalls or hospitalisations of neo-
nates.
Fig. 9. Evidence-based flow chart for PKU screening.
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–3736
4. Conclusion
Detailed studies of an enzymatic colorimetric Phe
determination in DBS over a period of 6 years
including more than 400,000 neonates clearly show
that this represents a reliable and sensitive method for
neonatal screening of PKU. During the whole study
period, we got no information and were not aware of
any false negative result. The overall recall rate of
0.23% was found to be in an acceptable range. More-
over, the prevalence of non-PKU HPA detected by
this method in our centre was twice as high compared
to the BIA, demonstrating the improved sensitivity of
the former.
Based on careful analyses of the screening results
and the monitoring of all abnormal samples, a new
evidence-based flow chart for neonatal PKU screen-
ing is proposed (Fig. 9). Its application should reduce
recalls without loss of sensitivity or any delay in
diagnosis and early treatment.
The observation of two PKU cases with delayed
Phe increase within the first weeks up to several
months highlights the need of continuous investiga-
tion and careful follow-up of infants with Phe levels in
the non-PKU HPA range. This is also true for those
neonates whose screening Phe levels are distinctly
elevated, because of the possible Phe decrease as we
have seen in two other babies. Furthermore, it must be
a general rule to confirm all abnormal results firstly by
internal re-measurement before initiating recalls or
hospitalisation of neonates in order to avoid unneeded
anxiety of the families.
The growing use and application of tandem mass
spectrometry in neonatal screening, based on the
advantage of this method to obtain information about
many other metabolites in addition to Phe at the same
time and within one analytical step, will not derogate
the usefulness of the enzymatic colorimetric Phe
determination in PKU screening in the foreseeable
future. This is especially true and important for
screening laboratories which are not able to afford
the greater expenses for tandem mass spectrometry.
References
[1] Guthrie R, Susi A. A simple phenylalanine method for detect-
ing phenylketonuria in large populations of newborn infants.
Pediatrics 1963;32:338–43.
[2] Blau K. Determination of phenylalanine in filter paper blood
spots by a simplified automated fluorimetric method without
dialysis. Clin Chim Acta 1983;129:197–200.
[3] Gerasimova NS, Steklova IV, Tuuminen T. Fluorometric meth-
od for phenylalanine microplate assay adapted for phenylke-
tonuria screening. Clin Chem 1989;35:2112–5.
[4] Vollmer DW, Jinks DC, Guthrie R. Isocratic reverse-phase
liquid chromatography assay for amino acid metabolic disor-
ders using eluates of dried blood spots. Anal Biochem 1990;
189:115–21.
[5] Dooley KC. Enzymatic method for phenylketonuria screening
using phenylalanine dehydrogenase. Clin Biochem 1992;25:
271–5.
[6] Schulze A, Kohlmueller D, Mayatepek E. Sensitivity of elec-
trospray-tandem mass spectrometry using the phenylalanine/
tyrosine-ratio for differential diagnosis of hyperphenylalanine-
mia in neonates. Clin Chim Acta 1999;283:15–20.
[7] Wendel U, Hummel W, Langenbeck U. Monitoring of phenyl-
ketonuria: a colorimetric method for the determination of plas-
ma phenylalanine using l-phenylalanine dehydrogenase. Anal
Biochem 1989;180:91–4.
[8] Wendel U, Koppelkamm M, Hummel W, Sander J, Langen-
beck U. A new approach to the newborn screening for hyper-
phenylalaninemias: use of l-phenylalanine dehydrogenase and
microtiter plates. Clin Chim Acta 1990;192:165–70.
[9] Campbell RS, Brearley GM, Varsani H, et al. Development
and validation of a robust specific enzyme mediated assay for
phenylalanine in serum. Clin Chim Acta 1992;210:197–210.
[10] Dhondt JL, Paux E. Evaluation of an enzymatic, colorimetric
method for the neonatal screening of phenylketonuria. Sreen-
ing 1993;2:141–7.
[11] Keffler S, Denmeade R, Green A. Neonatal screening for phe-
nylketonuria: evaluation of an automated enzymatic method.
Ann Clin Biochem 1994;31(Pt 2):134–9.
[12] Elvers LH, Diependaal GAM, Blonk HJ, Loeber JG. Phenyl-
ketonuria screening using the Quantase phenylalanine kit in
combination with a microfilter system and the dye tartrazine.
Sreening 1995;5:209–23.
[13] Dhondt JL, Loeber J, Elvers LH, Paux E. Preparation of the
first European working standard for phenylalanine determina-
tion in dried blood spots. J Med Screening 1998;5:63–6.
[14] Slocum RH, Cummings JG. Amino acid analysis of physio-
logical samples. In: Hommes FA, editor. Techniques in diag-
nostic human biochemical genetics: a laboratory manual. New
York: Wiley-Liss; 1991. p. 87–126.
[15] Chace DH, Sherwin JE, Hillman SL, Lorey F, Cunningham
GC. Use of phenylalanine-to-tyrosine ratio determined by tan-
dem mass spectrometry to improve newborn screening for
phenylketonuria of early discharge specimens collected in
the first 24 hours. Clin Chem 1998;44:2405–9.
[16] Burgard P, Bremer HJ, Buhrdel P, et al. Rationale for the
German recommendations for phenylalanine level control in
phenylketonuria 1997. Eur J Pediatr 1999;158:46–54.
[17] Recommendations on the dietary management of phenylketo-
nuria. Report of Medical Research Council Working Party on
Phenylketonuria. Arch Dis Child 1993;68:426–7.
A. Schulze et al. / Clinica Chimica Acta 317 (2002) 27–37 37