e.m.g.. emg is a continuosly changing voltage in the time domain emg - electromyography

20
Electromyography E.M.G.

Upload: allan-tipple

Post on 16-Dec-2015

234 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electromyography

E.M.G.

Page 2: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

E.M.G.

EMG is a continuosly changing voltage in the time domain

EMG - Electromyography

Page 3: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

E.M.G. Characteristics

EMG - Electromyography

The amplitude of the EMG signal is stochastic (random)

The amplitude can range from 0 to 6 mV (peak-to-peak)

The energy of the signal is is in the 0 to 500 Hz frequency range, with the dominant energy being in the 50-150

Hz range.

Page 4: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Frequency AnalysisFourier transformation quantifies the amount of each frequency in the signal

Frequency domain

EMG - Electromyography

Page 5: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Time

Vo

lta

ge Time

Vo

lta

ge

10 20 30 40 50 60 70 80 90 100

Frequency

Am

plit

ud

e

Resultant signal produced by summing the 10 signals to the left

10 individual sinusoids each with different amplitudes and frequencies

Frequency spectrum. Plot of amplitude of signal at each frequency

Page 6: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Detecting and Recording the EMG signal

Two main issues of concern that influence the fidelity of the signal:Signal to noise ratio.

the ratio of the energy in the EMG signal to the energy in the noise signal.

Noise is defined as electrical signals that are not part of the wanted EMG signal.

Distortion of the signal relative contribution of any frequency component in

the EMG signal should not be altered.

EMG - Electromyography

Page 7: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electrode Stability

When an electrode is placed on the skin a chemical reaction takes place which requires some time to stabilize,

The chemical reaction should remain stable and should not change significantly if the electrical characteristics of the skin change from sweating or

humidity changes.

EMG - Electromyography

Page 8: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electrical Noise

Inherent noise in the electronics components in the detection and recording equipmentAll electronics equipment generates

electrical noise.

EMG - Electromyography

Page 9: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electrical Noise

Ambient noise - originates from sources of electromagnetic radiation.

The dominant concern for the ambient noise arises from the 60 Hz (or 50 Hz) radiation from power sources. The ambient noise signal amplitude may be 3 times the magnitude of the EMG signal.

EMG - Electromyography

Page 10: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electrical Noise

Motion artifacts - two main sources of motion artifact:

Interface between the detection surface of the electrode and the skin

Movement of the cable connecting the electrode to the amplifier.

Most of their energy in the frequency range from 0 to 20 Hz

EMG - Electromyography

Page 11: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electrical Noise

Inherent instability of the signal The frequency components between 0 and

20 Hz are particularly unstable Because of the unstable nature of these

components of the signal, it is advisable to consider them as unwanted noise and remove them from the signal.

EMG - Electromyography

Page 12: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Electrode Placement

Place 1 - 2cm apart –A distance of 1 – 2cm is adequate to obtain a representative sample of the EMG signal from a muscle.

Midline placement – Electrodes should be placed away from the edges of muscles where crosstalk from signal of adjacent muscles is more likely to occur. Electrode orientation - The axis connecting the two electrodes should be parallel with longitudinal axis of the.

Place away from musculotendonous junction – Near the musculotendonous junction there are less fibers and they are thinner, thus the amplitude of the EMG will be less. The signal also becomes more susceptible to crosstalk from adjacent muscles.

Place away from the motor point – Action potentials travel distally and proximally along the muscle fibers from the motor point. Because the positive and negative phases of the action potentials (detected by the differential configuration) will add and subtract with minor phase differences there will be erroneous higher frequency components in the resulting EMG signal.

Page 13: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Surface Electrode Placement

Page 14: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

FILTERING

Kin 304 Spring 2005 EMG - Electromyography

Analog filters are electronic circuits which are designed to transform an input to a desired output by eliminating signals of specified undesired frequency bands

Page 15: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

FILTERING

High Pass

Low Pass V

Band Pass

Band Elimination

Frequency

When filters are used, some of the desired signal may be lost

Preprocessing of signal is desirable and may minimize computing demands

Page 16: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Filtering

EMG signal will be contaminated by some noise.

The signal to noise ratio can be increased by filtering between 20 - 500 Hz. (Band Pass)

EMG - Electromyography

Page 17: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Amplification

EMG - Electromyography

Single-ended AmplificationSkinfold caliper output from potentiometer

Differential Amplification EMG - preamplifier

Page 18: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Differential Amplification

Page 19: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Differential Amplification

The output voltage (Vo) results from the difference between two inputs (V1 and V2). Then Vo = K(V2-V1). T

This type of recording minimizes noise

This assumes that the two signals are subtracted perfectly, which unfortunately they are not. How well your differential amplifier subtracts the two signals is quantified by the Common Mode Rejection Ratio (CMRR). Perfection would be a CMRR of infinity.

EMG - Electromyography

Page 20: E.M.G.. EMG is a continuosly changing voltage in the time domain EMG - Electromyography

Ground Electrode

The ground or reference electrode provides the common reference for the differential amplifier.

Placed on an electrically neutral tissue a long way from the detecting electrodes.

Large electrode with electrically conductive gel to ensure a very good electrical contact with the skin.

Reduces power line interference noise.EMG - Electromyography