electron impact mass spectrometry of alkanes in supersonic ...94)00095-h... · electron impact mass...
TRANSCRIPT
Electron Impact Mass Spectrometry of Alkanesin Supersonic Molecular Beams
Shai Dagan and Aviv AmiravSchool of Chemistry, Sackler Faculty of Exact Sciences, Tel Aviv University, Tel Aviv, Israel
The electron impact mass spectrometry of straight chain alkanes CSH1S-C411Hs2' squalane,methylstearate, 'l-chlorohexadecane, I-bromohexadecane, and dioctylphthalate was studiedby sampling them with supersonic molecular beams. A fly-through Brink-type electronimpact ion source was used, utilizing a vacuum background ion filtration technique based ondifferences between the kinetic energy of the supersonic beam species and that of thermalmolecules. The 70-eV electron impact mass spectra of all the alkanes were characterized by apronounced or dominant molecular weight peak together with all the fragment ions normally exhibited by the standard thermal 70-eV EI mass spectra. In contrast, the NIST libraryof most of these molecules did not show any molecular weight peak. By eliminating theintramolecular thermal vibrational energy we gained control over the degree of molecularion fragmentation by the electron energy. At an electron energy of 18 eV the molecular iondissociation was further reduced considerably, with only a small absolute reduction in thepeak height by less than a factor of 2. The effect of vibrational cooling increased with themolecular size and number of atoms. Pronounced differences were observed between themass spectra of the straight chain triacontane and its branched isomer squalane. Similar massspectra of octacosane (C 2sH ss) achieved with 70-eV EI in a supersonic molecular beam wereobtained with a magnetic sector mass spectrometer by using an electron energy of 14 eV andan ion source temperature of 150°C. However, this ion source temperature precluded the gaschromatography-mass spectrometry (GC-MS) of octacosane. The GC-MS of alkanes wasstudied with an ion trap gas chromatograph-mass spectrometer at an ion source temperatureof 230°C. Thermal peak tailing was observed for C 211H 42 and heavier alkanes, whereas forC 2sH s8 and heavier alkanes the severe peak tailing made quantitative GC-MS impractical. Incontrast, no peak tailing existed even with C 411H s2 for GC-MS in supersonic molecularbeams. The minimum detected amount of eicosane (C 211H 42) was shown to be 60 fg. This wasdemonstrated by using single ion monitoring with the quadrupole mass analyzer tuned tothe molecular weight peak of 282 u. The coupling of electron impact mass spectrometry insupersonic molecular beams with hyperthermal surface ionization and a fast GC-MS inlet isbriefly discussed. (J Am Soc Mass Spcctront 1995, 6, 120-13])
Electron impact (ED is the major ionization methodused in mass spectrometry and gas chromatography-mass spectrometry (GC-MS). EI is a rela
tively Simple ionization method, yet it provides highsensitivity and a uniform semiquantitative response toall molecules and atoms. In addition, the rich fragmentation pattern observed, combined with the very richlibraries of 70-eV EI mass spectra, result in substantialmolecular identification and structural informationcapability.
Despite its strength, EI suffers from several limitations:
1. In many important cases the molecular weight peakis either unobserved or it is too small to be reliable.
Address reprint requests to Professor Aviv Arnirav, School of Chemistry. Sackler Faculty of Exact Sciences, Tel Aviv University, RamalAviv 69978, Tel Aviv, Israel.
© 1995 American Society for Mass Spectrometry1044-0305/95/$9.50SSDl1044-0305(94){)0095-H
2. The ion source must be heated to reduce the response time, and the ion source metal surface caninitiate thermal or catalytic molecular and molecular ion decomposition.
3. Relatively nonvolatile molecules undergo lengthycycles of thermal adsorption-desorption in the ionsource that result in gas chromatography (GC) peaktailing. Eventually, this peak tailing dictates theupper limit of molecular volatility in GC-MS.
We note that these three limitations are interrelated.The actual ion source temperature is set to a compromised value of - 150°C for direct sample insertionand 230-250 °C for GC-MS. Although increasing theion source temperature can decrease its response timeand ability to handle relatively nonvolatile molecules,it also substantially reduces the observed relativeheight of the molecular weight peak and promotescatalytic and thermal molecular decomposition.
Received June 1, 1994Revised September 7, 1994
Accepted September 7,1994
J Am Soc M.1SS Spectrorn IYY5. b. 120-131 ElMS OF ALKANES IN SUPERSONIC BEAMS 121
This compromise can be eliminated by the use ofsupersonic molecular beams [1-3] for molecular sampling in a slightly modified EI ion source. Supersonicmolecular beams (SMB) [4-8] are characterized byunidirectional motion with controlled hyperthermalmolecular kinetic energy (0.1-30 cv). intramolecularvibrational-rotational supercooling, mass focusing asin an ideal high load jet separator, a very high flowrate in the range of 100-500 ml.yrnin, and a convenient method for introducing samples at atmosphericpressure.
Supersonic molecular beams are used in mass spectrometry predominantly in conjunction with laserinduced multiphoton ionization followed by time-offlight mass analysis [9-14]. In these experiments thevibrational supercooling is used mostly to improve thewavelength selectivity in resonance-enhanced two ormultiphoton ionization schemes.
In a supersonic molecular beam, the vibrationalIy supercooled molecules undergo a single passagethrough the "fly-through" EI ion source. Thus, the lowmolecular vibrational temperature is independent ofthe ion source temperature, the molecular weightpeak height is substantially increased, and molecularthermal decomposition and adsorption-desorption onthe ion-source walls is avoided.
Amirav and Danon [2] have demonstrated the considerable effect of vibrational supercooling encountered in 5MB on the electron impact mass spectrometry (EI-MS) of bromopentane isomers. This effect wasalso studied in cholesterol [15] and various isomers ofoctane [3]. In these studies electrostatic backgroundion filtration [2, 3, 16, 17] was used. This method isbased on the observation that in a seeded supersonicmolecular beam, when hydrogen or helium is used asthe carrier gas, heavy organic molecules acquire hyperthermal kinetic energy. The ions formed in the ion cagepossess a kinetic energy that is the sum of their electrically given ion energy and their directional hyperthermal kinetic energy. Because the thermal molecules ofthe vacuum background have only small « O.l-eV)random kinetic energy, the use of a simple smallretarding field along the ion trajectory results in effective background ion filtration. This effect was used formass spectrometry of atoms and small molecules toimprove the beam signal-to-vacuum background ratio[16,17]. We have found that this procedure could havea major impact on organic mass spectrometry for thefollowing reasons: (1) Background ion filtration alsoeliminates the self-background of sample moleculesscattered from the ion source and vacuum chamberwalls. This self-background elimination uncovers thegenuine mass spectra of the vibrationally coldmolecules. (2) Eliminating the self-background alsoresults in a very fast ion source « 1 ms) response timeregardless of the molecular volatility because any surface-scattered molecule loses its kinetic energy and isfiltered out after ionization. (3) Because only moleculesin the 5MB are analyzed, the electron emission fila-
ment can be close to the ion cage and errussion currents over 10 mA can be routinely used. Therefore,increased ionization emission current and reducedbackground can compensate for the loss of sensitivityassociated with the fast passage and lack of multiplescattering of the molecule in the EI ion source.
Although the combination of EI and 5MB alreadyhas been demonstrated, no detailed study of its manifestation in organic mass spectrometry is available.This paper presents our results from the systematicstudy of the homologous alkane series CRH1/l-C40H /l2and their derivatives. These molecules have the samecharacter and their size and vibrational heat capacitysystematically increase. This group of molecules waschosen because it represents molecules that are important to petrochemical mass spectrometry, whose massspectra usually do not exhibit a molecular weight peakat the 1% relative height scale. In addition, chemicalioniza tion is less effective with alkanes both in termsof extensive fragmentation and also because of reduced sensitivity that results from their relatively smallproton affinity.
Experimental
In Figure 1, the supersonic molecular beam-massspectrometry (SMB-MS) apparatus is shown (upper),together with an expanded view of the inlet, whichalso serves as a fast gas chromatograph (lower). Thecarrier gas (H 2 or He) is introduced via needle valveVI at a flow rate of about 120-150% above the flowrequired by the column. The excess gas flows to theatmosphere and serves as a "septum flush". The samples are introduced as in a direct insertion device fromthe lower portion of the vaporizing oven in a conventional small test tube. No air lock and bypass pumpingis required, and the upper entrance remains open tothe atmosphere. The temperature gradient in the inletallows easy control of the sample vaporization ratewith mass spectrometry feedback. The upper openingserves for fast GC injection [18] with conventionalsyringes such as the SGE 1.0-JLL syringe (SGE, Australia). Manual injection is fast and simple because noseptum has to be penetrated. Only below a certainpenetration depth of the syringe into the injector doesthe carrier gas vaporize the injected amount and sweepit into the column (above this level the sample is sweptoutside as in septum flush devices). The injector istemperature controlled up to 500 °C, and the injectorliner is made either of Vespel'" or glass. All seals aremade of graphite.
The injector is mounted on a 3/4-in. stainless steeltube that contains a 50-cm-long, 0.53-mm-i.d. megabore column [DB-lor DB-5MS, 1.5-JL film thickness(J & W Scientific, Folsom, CA)]. The GC column isseparately heated and its temperature is controlled. Ata relatively high temperature it merely serves as atransfer line for the vaporized molecules. The outputof the column is mixed with make up gas (helium or
122 DAGAN AND AMIRAV J Am Soc Mass Speclrom 1995. 6, 120-131
Figure 1. The sample inlet and mass spectrometer apparatus.The carrier gas flow He or H 2 , is controlled by needle valves 0)and (2). The sample is injected or introduced into the vaporizingoven (3) and flows through the temperature-controlled capillarycolumn (4). The gas mixture expands through a supersonic nozzle (5) and is double differentially pumped. The sample moleculesare ionized in the EI ion source (6), deflected by 90° by the iondeflector (7) and mass analyzed by the quadrupole mass spectrometer (8). The molecular beam also can be square wavemodulated by the beam chopper (9). The data are processed andpresented on a PC (10). The lower section shows an expandedview of the fast gas chromatograph and sample inlet.
diffusion pump) and enters the electron impact ionsource (6) in the high vacuum mass spectrometrychamber (63-mm diffusion pump). The nozzle can bealigned (XYZ) with respect to the skimmer from outside the vacuum chamber to maximize the beam intensity as measured by the increase of pressure in themass spectrometry chamber.
The supersonic molecular beam is ionized by thefly-through Brink-type [20] EI ion source (6), which isopen both at the entrance and exit of the 5MB. Theionization cage is 25 mm long and 5 mm in diameter.The filament is made of a O.2-mm-diameter thoriatedtungsten wire purchased from Vacuumetrics. The ionsare extracted by the ion optics attached to the EI ionsource and guided by 90° deflection (7) into aquadrupole mass analyzer (8) [VG-SXP 600 (VG Analytical Ltd., Manchester, UK)]. This quadrupole massanalyzer does not have a prefilter element and exhibitsa significant reduction in ion transmission with mass.We measured an exponential transmission reductionfactor of about 60 between masses 69 and 502 underconventional unit mass resolution conditions. The original commercial EI ion source was removed from themass analyzer and its entrance contains an ion opticslens system.
Between the quadrupole mass analyzer and thechanneltron ion detector there is a circular plate with a3-mm hole denoted "exit lens". This plate is externallyvoltage biased to filter out ions produced from thethermal molecules of the vacuum background, and ittransfers only ions produced in the hyperthermal 5MB,which are distinguished by their higher kinetic energy[2, 3, 16, 17]. As mentioned, in addition to eliminatingthe vacuum background, the background-free EI ionsource facilitates ultrafast response time and exposesthe mass spectrum of vibrationally cold molecules.
The 5MB also can be square wave modulated by abeam chopper (9), which can be inserted into the beamtrajectory or removed. This beam modulation allowsthe use of lock-in amplification for background subtraction when conventional EI is used.
The 900 ion deflector can also be moved upward onan XYZR vacuum manipulator and be replaced by asuitable surface for hyperthermal surface ionization(HSI) [3, 21, 22] of the hyperthermal beam species.
Compatibility with standard GC was investigatedby using a conventional gas chromatograph [CarloErba (Milan, Italy) Vega 6000] coupled to the fast gaschromatograph via a homemade, separately heated,transfer line. The GC column (Ifi-m megabore J& WDB-I; 1.5-p. film thickness) ended inside the injectornear the fast GC column, which served as a transferline at that time by heating it to above the conventional GC column temperature. We used the ShraderSystem TM as the GC-MS software and for most of thedata presentation.
Some additional important dimensions are thenozzle-skimmer distance, which is 5-8 mm, theskimmer-surface (or ion deflector) distance of - 60
sc AND SYRINGEINLET
M5 in 5MB
AT".OSPHERIC PRESSURE SAMPLING
VACUUM
FAST GC
HEATERNOZZLE \
hydrogen) provided by needle valve Vz. This needlevalve controls the pressure behind the nozzle and thusaffects the flow rate through the column. When Vz isclosed the linear velocity of the carrier gas is about2000 cmys, whereas when it is open the column flowrate can be regulated to any value, including a reverseprotective flow. The supersonic beam nozzle is anall-ceramic nozzle with either a 100-p.m pinhole in awatch jewel [19] or a I-mm-long channel that has adiameter of 105 p.m drilled in alumina (Kulicke andSoffa-Israel), Both the gas mixing zone and the nozzleare separately heated. The carrier gas mixture and thesample molecules expand through the nozzle (5) (inthe upper scheme), into the first vacuum chamber,which is pumped by a 4-in. diffusion pump [VarianVHS 4 (Varian Associates, Lexington, MA) backed by a550-L/min rotary pump]. The emerging supersonicfree jet is skimmed and differentially pumped (63-mm
J Am Soc Mass Speclrom 1995. 6, 120-131 ElMS OF ALKANES IN SUPERSONIC BEAMS 123
Figure 2. EI mass spectra in 5MB of octacosane and a comparison with the NIST library thermal 70-eV mass spectrum (trace c).Trace a was obtained in 5MB with 18-eV electron energy. Trace bwas obtained with 70-eV electron energy.
Octocosone- C28H58394
A .IBeV-SMB
B.70eV-SMB 394
57M
71
4j r k1/1111
M'2,.
43C. 70eV-L1BRARY57
71
C
a molecular weight of 394 u. Octacosane and two of itsisomers appeared to be the best three candidates. Onthe other hand, we failed to locate octacosane throughthe mass spectrum obtained from an ion trap GC-MS[Finnigan ITO model 8100 (Finnigan-MAT, San Jose,CA)] with the standard library search algorithm. Thesesearch results demonstrate that when the molecularweight peak is a prominent mass spectrometry peak,the chances to obtain high confidence level librarysearch results are increased. This observation emergesfrom the existence of all the usual fragments with onlysmall changes in their relative peak heights. This resultis rationalized by noting that the molecular weight ioncan only dissociate, whereas all other peak heights canbe increased by the dissociation of a higher mass peakand can be decreased by their own fragmentation.Therefore they are in a quasi-steady state. The molecular weight peak in octacosane is increased by morethan 3 orders of magnitude compared with thermal70-eV EI mass spectrum, whereas the peak heightsfrom other fragments changed only slightly.
Because in 5MB the thermal intramolecular vibrational energy is removed, the electron energy is theonly parameter that governs the degree of fragmentation. The effect of the electron energy on the degree offragmentation of the molecular weight ion is wellknown. However, the effect of reduced electron energyto achieve a noticeable increase in the relative molecular weight peak height also is associated with a consid-
rnm, the beam diameter of - 3.5 mm in the EI ionsource, and the 900 ion deflector of 15 mm from thequadrupole mass spectrometer entrance lens.
Vibrational Cooling Effect on the ElectronImpact-Mass Spectrometry of Octacosane
Vibrational temperature effects on EI-induced ion fragmentation are well known [23-28] including in longchain alkanes [29]. In many books on mass spectrometry , at least one example of the pronounced effect ofthe ion source temperature on the degree of ion fragmentation appears [30, 31]. The subject of low electronenergy, low temperature EI-MS is reviewed by Maccolland references therein [32].
Octacosane [CH3(CHz)Z6CH3] is a relatively longstraight chain alkane molecule that has 86 atoms. It is asolid at room temperature and has a boiling point of705 K. Octacosane, like all the other long chain alkanes,is known to exhibit a mass spectrum that consists oflow mass fragment ions without a molecular weightpeak. The high temperature limit of the vibrationalheat capacity of octacosane is 11.4 eV at an ion sourcetemperature of 250 °C, whereas it is estimated, basedon model calculations [33], that the actual heat capacity is about 3.5 eV. This intramolecular vibrationalenergy, together with the electron imparted vibrationalenergy, is sufficient to break the majority of all molecular ions. In a supersonic molecular beam, however, weanticipated that this thermal contribution would bepractically eliminated. In Figure 2 the electron impactmass spectra of octacosane are shown. In trace a, the18-eV electron impact mass spectrum is plotted. Intrace b the 70-eV mass spectrum is shown. The lowertrace (c) was taken from the NIST mass spectra libraryand it represents the standard 70-eV EI mass spectrumof thermal molecules at unspecified ion source temperature (probably in the range 150-250 0C). In contrast totrace b, trace c does not show any molecular weightpeak nor does it show any other high mass fragmentpeaks. In fact, the molecular weight peak does notexist in the library list of ions even at the level of 0.1%.On the other hand, the 70-eV EI mass spectrum in asupersonic molecular beam is characterized by a dominant molecular weight peak that is the most prominentfeature in the mass spectrum. The expanded massscale near the molecular weight shows that all themolecular weight peaks exist, and their relative heightsreflect 13C natural abundance of 1.08 ± 0.02%, which isin good agreement with the literature value of 1.10 ±0.03%. Thus, one can obtain the elemental informationdirectly from the relative peak height ratios and determine the empirical formula. We note that the mass 394u reflects the genuine molecular weight without anychemical ionization-related process of proton transfer.
We used a library search to identify the mass spectrum of trace b by eliminating its molecular weightpeaks and by limiting the search to molecules that had
20 120 220 320
MASS(omu)
420 520
124 DAGAN AND AM IRAV J Am Soc Mass Spectrum 1995,6,120--131
562I
450
338
L:'H"226 -----I
220 320 420 520
MASS (o rnu)
oscillator. In Figure 3 it is observed clearly that eachion has a different normalized electron energy dependence. We also observe that the smaller the ion fragment is, the weaker is its dependence on the electronenergy. This observation suggests that fragmentationoccurs via a chain of fragmentation events, each to aclose, but lower mass. It is believed that the measurements outlined in Figure 3 contain structural and energetic information, but this subject is beyond the scopeof this article.
85
70eV (I-SMB
Effects of the Length of the Alkane Chainon its Mass Spectrum in SupersonicMolecular Beams
114
Because the vibrational heat capacity depends approximately linearly on the number of atoms in the molecule,one can expect an increase of the vibrational coolingeffect with the length of the alkane chain.
We have studied the mass spectra of alkanes CHH1Hthrough C 4oH H2, and in Figure 4 the 70-eV EI massspectra in 5MB of the indicated straight chain alkanesare portrayed. All these mass spectra are characterized
Figure 4. 70-eV electron impact ionization mass spectra in supersonic molecular beams of the indicated straight chain alkanesCMH1M-C40HR2'
80
oElo-
oZ
(J)
uco-0c~
.o«-0(J)
N
erable reduction in the EI ionization cross section. Wehave measured an EI ionization cross section reductionfactor of 8 at 18 eV, in comparison with 70-eV electronenergy. In trace a we show that at 18 eV, the EI massspectrum in 5MB is dominated almost totally by themolecular weight peaks, whereas all other fragmentpeak heights are less than 5% of the normalized scale.More important is the observation that this considerable simplification of the mass spectrum was achievedwhere the absolute height of the molecular weightpeak was reduced only by a factor of 1.5. This observation results from the demonstrated degree by whichthe molecular ion dissociation is reduced, which compensates for the reduced ionization cross section.
We believe that the ability to achieve EI mass spectra that consist almost entirely of the molecular weightpeak, without losing sensitivity, is of considerable importance. If any molecule in a complex mixture exhibits a single or very few mass spectral peaks, thiscould considerably extend the ability of GC-M5 tohandle its analysis in complex mixtures, because boththe mass spectrometer and the gas chromatographcould serve as orthogonal separation tools with simplified deconvolution of overlapping GC peaks.
In addition, because the electron energy alone determines the degree of fragmentation of the various ions,this dependence can be explored and monitored. InFigure 3 we plot the normalized abundance of theindicated fragment and molecular weight ions versusthe electron energy. Each ion peak height was normalized to its maximum value in the range of 10-80 eV.The electron energy was scanned by programming theelectron energy power supply through a slow ramping
Figure 3. The dependence of the normalized octacosane fragment ion peak heights on the electron energy.
J Am Soc Mass Spectrom 1995. 6. 120-131 ElMS OF ALKANES IN SUPERSONIC BEAMS 125
by a dominant or strong molecular weight peak. Infact, the molecular weight peak is always the strongestpeak in the mass spectra. However, these mass spectraare distorted due to the substantial reduction in transmission at high masses of our quadrupole mass analyzer. In addition, a very long progression of peaksevery 14 u (CH z) is observed, which extends to fragments heavier than those in the library mass spectra ofthese compounds. Note that the library mass spectraalso do not show a molecular weight peak for most ofthese alkanes.
It seems, therefore, that electron impact mass spectrometry in supersonic molecular beams exhibits theinformation content usually achieved by electron impact together with chemical ionization (Cl) mass spectra in one scan. However, CI provides a degree ofselectivity against alkanes and in favor of compoundsthat have high proton affinity. Moreover, the molecularweight peaks in EI provide elemental and isotopicinformation through the relative M+, [M + 1]+, and[M + 2]+ peak height rat ios.
In Figure 5 we show the EI mass spectra in 5MB ofthe same alkanes as in Figure 4, obtained at 18-eVelectron energy. It is clear that as the length of the
18eV EI-SMB
85 114
A .C8 HI8
,iI170
B . CI2H26
IIII226
C . CI6H34
338
D. C24 H50
450
E .C32H 6 6
562
F. C40H82
20 120 220 320 420 520MASS (ornu )
Figure 5. 18-eV electron impact ionization mass spectra in supersonic molecular beams of the indicated straight chain alkanesCSHlS -C40Hs2'
alkane chain increases, the 18-eV EI mass spectrum isdominated further by the parent undissociated molecular weight peak. For certain noncritical applications,one can say that above an alkane chain length of C16'
the 18-eV EI mass spectra are essentially characterizedby a single molecular weight peak and only smallindication of other fragments . This claim is furthersupported by consideration of the strong transmissionreduction of our quadrupole mass analyzer with mass.
As mentioned before, this simplification of the massspectra at reduced electron energies also involves areduction in the ionization cross section by a factor of 8at 18 eV in comparison with 70 eV (these data wereobtained by total ion current measurements normalized to the emission current in a DuPont magneticsector mass spectrometer). However, this reduction inthe ionization cross section is compensated by thesubstantially reduced degree of fragmentation encountered in EI mass spectra in 5MB. In Figure 6, thereduction factor of the molecular weight signal at 18eV in comparison with that of 70-eV EI mass spectra isplotted versus the chain length of the straight alkane.We note that the electron energy value is accurate towithin ± 1.5 V due to a voltage gradient on the electron emission filament. From Figure 6 we observe that(1) the reduction of M +, and thus the reduction indetection sensitivity, is much smaller than the reducedionization cross section by a factor of 8; (2) for alkaneswith over 12 carbon atoms, this reduced sensitivity isless than a factor of 2; (3) the factor by which thesensitivity falls is reduced with the length of the alkanechain and approaches unity (no reduction) for longchain alkane molecules. We conclude therefore, that in5MB the simplification of the mass spectra that resultsfrom lowering the electron energy is achieved with asmaller penalty in sensitivity as compared with thermal electron impact mass spectrometry.
Comparison of Figures 4 and 5 indicates that although the effect of reduced electron energy on themass spectra is relatively small in octane, it graduallyincreases with the length of the alkane chain. In Figure7, we plot the ratio of the molecular weight peak to the57-u prominent fragment peak achieved at 18-eV electron energy, divided by the same ratio achieved at anelectron energy of 70 eV. This normalized ratio is ameasure of the relative degree of control over thedegree of fragmentation achieved by the electron energy. Note that this measure is normalized and unaffected by inconsistencies in the response of thequadrupole mass analyzer with the ion mass. Figure 7demonstrates a purely exponential increase of the degree of control (C) over the fragmentation achievedwith increased chain length of the alkane. Althoughthe mass spectrum of octane is hardly affected by theelectron energy in the ranges 18-70 eV, where C ::::; 1.2,the degree of control is increased to more than 50 forthe C 36 and C 40 alkanes. The small deviation of C 4oH82
from the straight line is attributed to incomplete cooling in the 5MB.
126 DAGAN AND AMlRAV
2.8
2.6 ..2.4
;;- 2.2u00......~
2::E-.;;- 1.8 ..u
0 ..t:,
1.6::E ....1.4 ..1.2 .. ..
1+----~-~-.______,_-......____,,.____r-.___.-__14 8 12 16 20 24 28 32 36 40 44
Number of Carbon Atoms
Figure 6. The ratio of the height of the molecular weight peaktaken with 70- and 18-eV electron energy, plotted against thenumber of carbon atoms of straight chain alkanes C~H 1~-C.lIH6Z'
The observation shown in Figure 7 can be rationalized by using the following assumptions: (1) The vibrational energy given by the electron to the moleculeis independent of the length of the alkane chain both at18- and 70-eV electron energies (localized ionization);(2) the amount of intramolecular vibrational energy isnegligible and can be ignored in comparison with thevibrational energy imparted by the electrons. By usingthese reasonable assumptions one finds that the amountof vibrational energy is molecular size independent,
100..------------,--,
........;>.,0t:,ell~..~.......::E....... 10
;>.,00-......ell
'"..~::E
1+--.-L..,----r-..-----.----r-____,-.---,-_.---14 8 12 16 20 24 28 32 36 40 44
Number of Carbon Atoms
Figure 7. The ratio of the molecular weight peak divided by the57-u fragment peak height achieved at 18- and 70-eV electronenergies, plotted against the number of carbon atoms of straightchain alkanes CgHlS-C.oHgz.
J Am Soc Mass Spectrom 1995, 6, 120-131
but the amount of vibrational energy per atom isreduced linearly with the number of atoms (chainlength).
Accordingly, as the molecular size increases, itsdissociation to small fragments such as the 57-u fragment is less likely, especially in view of the consecutive fragmentation processes implied from the resultsshown in Figure 3. If we further assume an exponentialdependence of the rate of ion fragmentation on theexcess vibrational energy and number of atoms, thedegree of fragmentation will be reduced exponentiallywith reduced intramolecular vibrational energy content per atom at lower electron energy. In addition, theratio of these reduction ratios also will behave in anexponential fashion as is demonstrated in Figure 7.Although the foregoing conjectured explanation maynot be based on theoretically solid arguments, ourexperimental results clearly demonstrate increasedcontrol over the degree of molecular ion fragmentationwith electron energy for larger molecules. We notethat the extrapolation of our measurements to largeralkanes should result in the appearance of the molecular ion exclusively even at 70-eV electron energy. Infact, if we correct our results for the reduced sensitivity of our quadrupole mass analyzer with increasedmass, the molecular weight of C 4oH H2 at 562 u will behigher than all other fragments by a factor close to 20in the 70-eV electron energy mass spectrum.
Up to now we have addressed the question of thevibrational supercooling effect on the EI mass spectraof straight chain alkane molecules. In Figure 8 the70-eV EI mass spectrum in 5MB of the straight chaintriacontane molecule (C 30H 62 ) is compared with thatof its branched isomer squalane. Pronounced isomermass spectral differences are observed and new highmass major peaks appear through the enhanced fragmentation at the branching points that form secondaryions. A closer examination of the NI5T library massspectra of these molecules reveals that these isomereffects exist even in the thermal mass spectra, but theyare considerably amplified in the EI-M5 of the vibrationally cold molecules in the 5MB. The mass spectrumof squalane contains relatively abundant ions at 182,238, 266, and 337 u as well as neighbors to thesemasses. These ions are not straightforward cleavagesof the carbon chain and are hard to explain. Again allthese ions appear in the NI5T library, whereas in 5MBtheir relative abundance is largely amplified. In Figure9 the 70-eV EI mass spectra in 5MB of the indicatedalkane derivatives are shown. Figure 9 demonstratesthat the feature of a considerably enhanced molecularweight peak is universal to all molecules. In fact, thegeneral trend we observed in these and many othermolecules [2, 3, 15, 20] is that the smaller the molecular weight peak is in thermal 70-eV EI-M5, the larger isthe enhancement achieved by vibrational cooling in5MB. Methylstearate exhibits a 21% molecular weightpeak in the library (NI5T) thermal 70-eV EI massspectra. In the 5MB its relative peak height is increased
J Am Soc Mass Spectrom 1995. 6. 120-1 31 ElMS OF ALKANES IN SUPERSON IC BEAMS 127
4H
70 eV EI- SM B-MS of (30 H62 70 eV EI-SM8 298
"'.
A. TriacontoneA .Methylsteorote
7487
14 3
I I
8 .1,Chlorohexodecone260
91
262
135
C. I , 8romohexodecane238
B. Squalane
T14 9
422
D.Dioctylphtholote279
167
71
Mass (amu)
Figure 8. 70-eV electron energy EI mass spectra of the straightcha in triacontane (a) and its branched isomer squalane (b) .
by about 1 order of magnitude. 1-Chlorohexadecaneand 1-bromohexadecane do not show any molecularweight peaks in their thermal 70-eV EI mass spectraand thus an enhancement of at least 3 orders of magnitude in peak height is obse rved. Dioctylphthalate in5MB exhibits only a 4% molecular weight peak, whichis the lowest we have measured to date. We havestud ied the 70-eV EI mass spectrum of dioctylphthalate in a DuPont magnetic sector instrument and foundthat at an ion source temperature of 250 °C, the relati veheight of its M + peak is - 2 X 10- 6 of the 149-u peak,and that the [M + 1] peak is higher than expected ,probably due to the ionization of dimers or due toresidual chemical ion ization. Thus, the small M+ peakof dioctylphthalate in EI-SMB represents an increase ofover 4 orders of magnitude in the relative height of themolecular weight peak.
Figure 9. 70-eV electro n ene rgy El mass spectra in 5MB of theindicated alkane der ivati ves .
Tradeoff of Response Time, Sensitivity,and Relative Molecular WeightPeak Height
The effect of the vibrational temperature on the degreeof molecular ion fragmentation is a very well knownand established phenomenon [23-33]. Reduction of theion source temperature indeed improved the molecular ion pea k appearance , but involved a tradeoff oflowered ion source response time and reduced cleanliness. The relative height of the molecular weight peakalso can be increased through the use of a better massanal yzer. A magnetic sector, in contrast to a quadrupolemass analyzer, does not exhibit a marked reduction insensitivity with increased mass and enhances the apparent M + peak height. Another way to increase therelati ve height of the molecular weight peak is toreduce the electron energy with the tradeoff of detection sensitivity . In Figures 10 and 11 we demonstratethese three methods and discuss the tradeoffs involved. In Figure 10 mass spectra of octacosane(C2sH ss) measured by a magnetic sector mass spectrometer are shown. They were taken at the indicated
170 220 270 320 370
MASS (ornu )
12070
5,7" d J
20
182
11:3
127 266
155 237
t j It I I T IDO 180 230 280 330 380 430
j
~7
30 80
128 DAGAN AND AM IRAV
C28 H S8 Thermal EI-MS
J Am Soc Mass Spectrom 1995, 6, 120-131
ALKANES GC - MS
A .GC-ITD-MS
230°C
Magnetic Sector
A. T1s=230°C
Ee=18eV
o 5 10Cl6
15 20TIME(m in}
25 30 35
394
B. TIS =ISO°C
Ee= 18eV
o
B GC-SMB-MS
2 3TIME (m in)
4 5
c.Tls=ISO°C
Ee=14eV
o
C FAST GC-SMB-MS
8 10
L.U.,LL.,........y...u,..I...L~+-'-.J+LJ..,JLJ..--.
40 120 200 280 360 440MASS(amu)
Figure 10. Electron impact ionization-magnetic sector massspectrometry of octacosane achieved with a thermal ion source atthe ind icated ion source temperatures and electron energies.
electron energies and ion source temperatures. In contrast to the library 70-eV EI mass spectra, the magneticsector instrument does provide a 6% normalizedmolecular weight peak at 18-eV electron energy and anion source temperature of 230 °C as shown in trace a.This increased M + peak is the result of the flat massresponse of the magnetic sector mass analyzer as wellas the effect of reducing the electron energy to 18 eV.However, the tradeoffs are both in price and complexity through the use of a magnetic sector instrumentinstead of a quadrupole one and reduction of theionization cross section (sensitivity) at the reducedelectron energy by a factor of 8. We note that reducedelectron energy cannot be achieved with an ion trapmass spectrometer unless an external ion source isused. In trace b, the effect of the ion source temperature is demonstrated by its reduced temperature from230 (trace a) to 150 °C. The molecular weight peak isnow strongly increased to about 90% of the normal-
Figure 11. GC-MS and fast GC-MS of a mixture containingC8H 18-C40H S2 alkanes achieved us ing the following instru ment s: (a) Commercial Finni gan model 8100 ion trap gas chr omatograph-mass spectrometer with a 30-m narrow bore capillary column. (b) Superson ic molecular beam mass spectrometrycombined with a gas chromatograph with a 4-m narrow borecolumn <injection at 100 °C and ramping at 49 °Cjmin to 300 °c).(e) Fast GC-MS in a supersonic molecular beam, achieved withthe fast gas chromatography interface shown in Figure 1.
ized scale. This is ach ieved together with the additional penalty that at 150 °C octacosane cannot bequantified with GC-M5 due to a severe loss of ionsource response time and associated peak tailing. Intrace c the magnetic sector mass spectrum of octacosane was measured at an ion source temperature of150 °C and 14-eV electron energy. At this electronenergy the ionization cross section is lower than at 70eV by a factor of 30, but now the apparent massspectrum seems similar to that of octacosane achievedwith our quadrupole mass analyzer with 70-eV EI-M5in 5MB, as shown in Figure 2. Therefore, from Figure10 it can be concluded that the use of supersonicmolecular beams for 70-eV EI-M5 can provide massspectra like those achieved by magnetic sector instruments at low electron energy and reduced ion sourcetemperature, without the involved penalties of reduced sensitivity, much longer ion source response
J Am Soc Mass Spcclrom 1995,6,120-131 ElMS OF ALKANES IN SUPERSONIC BEAMS 129
time, and increased complexity of the mass analyzer.Among these three features we believe that the ionsource response time is the most significant because itconcerns the mere ability to perform GC-M5 experiments.
In Figure 11 we compare a total ion chromatogramobtained by GC-MS (Finnigan ITD 8100) with a conventional ion source and a chromatogram taken withsupersonic molecular beam mass spectrometry. A mixture of C
IIH 2 11 + 2 (11 = 8,12,16,20,24,28,32,36,40)was injected into the gas chromatograph. It is observedclearly that the ion trap response is characterized byextensive chromatographic peak tailing for the C 24H 50
and heavier alkanes. A closer examination revealedthat even the eicosane (C 2oH 42 ) exhibited noticeableion source-related peak tailing. As shown in the expanded scale ( X 16), the extensive peak tailing rendersGC-MS measurement of C 32H 66 , C 36H 74 , and C 4oH R2
molecules impractical. In addition, the peak tailingfound with C 2oH 42 , C 24H 50 , and C 2RH 4R results in areduction of detection sensitivity and hampers quantitative measurement by considerably poorer reproducibility, as the exact amount of peak tailing maychange due to long term thermal instabilities of the iontrap. Trace b shows the EI-SMB GC-MS of the samealkane mixture by using a Carlo Erba Vega 6000 gaschromatograph with a 4-m narrow bore column (250-jLi.d. J& W DB1; 0.25-jL film thickness). No peak tailingis observed although the chromatographic peak heightsabove hexadecane are reduced due to chromatographicpeak broadening and due to the reduced response ofthe quadrupole mass spectrometer at higher masses. InFigure 11c we also demonstrate the ultrafast GC-~S[18] of a similar mixture of C 24-C 40 alkanes. In thiscase, with the help of flow programming after theelution of octacosane, the GC-MS trace ended after 8 sand even the C 4oH R2 alkane molecule was characterized by a l-s tail-free peak width. Thus, we concludethat, as expected from the use of background ionfiltration, no tailing exists and the full temperature andvolatility limit of the gas chromatograph can be exploited without mass spectrometric-related limitations.In addition, we note that we have examined the expanded mass spectra of hexadecane and eicosane obtained with the ion trap gas chromatograph-massspectrometer and failed to find even a sing~e ion. attheir molecular weight masses at the conventional IOnsource temperature used (230 °C).
Detection Sensitivity with Electron ImpactMass Spectrometry in SupersonicMolecular Beams
One of the most important parameters of any massspectrometry ionization method is its associated detection limits. However, detection limits are usuallyquoted for a given molecule and experimental conditions that do not reveal the other interrelated parame-
ters. Electron impact in supersonic molecular beams isdifferent from conventional EI in several importantaspects that affect the detection limits:
1. The supersonic molecular beam is inherently associated with jet separation, which reduces the molecular flux by a factor in the range of 2-20. The heavierthe molecule is, the smaller is the reduction factor.On the other hand, the very high flow (240 ml.yrnin)jet separator allows fast sampling of very largesplitless injections and thus actually improves theachievable minimum detected concentration limit.
2. The molecules pass the ion source in fast motiononly once instead of an estimated number of 10-20cycles of intraion source adsorption-desorptionrounds of thermal molecules which are 10 timesslower. This means that the molecular ionizationprobability is smaller in the 5MB by a factor of100-200 times.
3. Because pyrolysis on the electron emitter filamentand temperature gradients inside the ion source arenot a problem, an open Brink-type ion source [20]. isused with a filament close to the ion cage and WIthan electron emission current of up to 20 mAo Theconsiderably increased electron emission current andthe high efficiency of the open Brink-type ion source,which allows several electron trajectories through it,compensate for most of the sensitivity losses due. tothe single fast molecular flight through the IOnsource.
4. The ion source ion optics and quadrupole massanalyzer transmission obviously affect the detectionsensitivity. Typically, we detect less than 2% of theions that exit the ion source at the channeltron iondetector (mostly quadrupole losses). The quadrupoletransmission is further reduced exponentially athigher masses.
5. The vacuum background ion filtration process reduces the noise. This feature does not provide betterdetection limits, but rather allows peak performancenear the detection limits to be achieved everydayregardless of the "history" of the apparatus. Novacuum system bake-out is needed to approach theminimum detection limit.
6. The supersonic molecular beam allows coupling ofa fast gas chromatograph with narrower peak widthsand thus higher peak flux, which contributes toimproved detection limits. .
7. The considerably enhanced molecular weight peakimproves the single ion monitoring of chosenmolecules through their molecular weight peak by afactor that can be over 103
.
Figure 12 demonstrates the detection sensitivityperformance of our apparatus through the fast GC-MSof eicosane (C 2oH 42 ) . The upper trace (a) demonstratesthe single ion monitoring of eicosane through itsmolecular weight peak at 282 U. A solution of 100-ppb
130 DAGAN AND AMlRAV J Am Soc Mass Speclrom 1995.6,120-131
Discussion and Overview
We have shown that the EI-M5 of alkanes in supersonic molecular beams is characterized by a substantial increase in the relative height of the molecularweight peak. The electron energy provides good control over the degree of molecular ion fragmentationwith only a small sacrifice in the absolute signal of themolecular weight ion. In addition, the EI-M5 in 5MBprovides tail-free GC-M5 chromatograms without massspectrometry ion source related restrictions on themolecular volatility.
As is well known and established, the enhancedmolecular weight peak is of considerable importancefor the following reasons:
1. This is a very important piece of information fordetecting unknown molecules.
2. It helps to improve library searches.3. This is the best mass spectrum peak to resolve
overlapping gas chromatography peaks.4. M + is the best peak for the single ion monitoring of
target compounds.5. The M+, [M + 1]+, and [M + 2]+ peak ratios pro
vide elemental information.
ions per time constant of 0.3 s. The noise originatespredominantly from a mass-independent backgroundion signal, which depends on the total pressure. Webelieve this background signal originates from the insufficient ion filtration dynamic range of our quadrupole mass analyzer.
The lower trace (b) shows the full scan total ionchromatogram obtained by five consecutive injectionsof 1 ng of eicosane (0.1 JLL, splitless, 10 ppm), eachevery 12 s. The extrapolated MDA in this full scanmode is 30-40 pg. The lower trace (c) represents thesame chromatogram as in trace b, but with a computerreconstructed single ion chromatogram at the molecular weight peak. In this case the detection limit isimproved to below 10 pg . We note again that evenwith 100 ng of eicosane we could not find even a singleion at the eicosane molecular weight in the conventional GC-M5 shown in Figure 11. Thus, in this modethe 5MB-M5 is much more sensitive than an ion trapgas chromatograph-mass spectrometer. It should bementioned also that when one searches for low detection limits, mass spectrometry in supersonic molecularbeams also facilitates the addition of hyperthermalsurface ionization [3, 21, 22] which is a very sensitiveionization method with a demonstrated MDA of below1 fg [18]. However, hyperthermal surface ionization isrelatively inefficient with alkanes and results in extensive molecular ion fragmentation. It is, however, anoptimal ion source for the detection of aromaticmolecules and amines that supplement the capabilityof the EI ion source and that is present in the sameapparatus.
48
x to~
A 10 Picogrom
SIM 2820mu
12o
I , ,6 i
0 2 4 8 10TI ME (sec)
B. 1 Ng FU LL SCAN TIC
~\ 1
c. 1 Ng FULL SCAN RSIM 2820mu
24 36TIME (sec)
Figure 12. Sensitivity evaluation of EI-M5 in 5MB. (a) O.l -J.tLsolution of lOO-ppbeicosane 00 pg ) was injected into the fast gaschromograph with the mass spectrometer tuned to the molecularweight mass of 282 u. (b) Five repetitive O.l-J.tL injections, eachafter 12 s, of a lO-ppm eicosane solution into the fast gas chromatograph-mass spectrometer. A full scan GC mass spectrum isshown with the mass spectrometer scanning. (c) As in b, butcomputer reconstructed single ion monitoring is shown at themolecular weight of 282 u.
eicosane in cyclohexane (0.1 JLL) was injected (splitless) into the O.5-m megabore column of the fast gaschromatograph inlet [18]. The GC-MS measurementwas completed in 3 s and the observed chromatogramof 10 pg of eicosane was practically noise-free. Fromthe expanded ( X 10) noise trace we estimate the signalto-two rms noise ratio to be 180, and the linearlyextrapolated minimum detected amount (MDA) to be60 fg (this value is 250 fg if a three peak-to-peak noisevalue is considered). We believe this result is betterthan that which can be achieved with any other massspectrometer for heavy alkanes by single ion monitoring at their molecular weight peak. We also havemeasured the signal by ion counting and found theoverall signal to be about one molecular weight ion atthe channeltron ion detector per femtogram of eicosane.The background noise count rate was 2000 ions persecond and the rms noise fluctuations were about 25
J Am Soc Mass Spectrum 1995, 6, 120-131 ElMS OF ALKANES IN SUPERSONIC BEAMS 131
6. The existence of a noticeable molecular weight peakcan provide isotope abundance geochemical information as well as enable deuterium exchange andisotope dilution analysis.
Obviously, all these aspects have long been attributedto EI-M5 for those molecules that exhibit a pronouncedmolecular weight peak. The contribution of supersonicmolecular beam sampling is merely in the extension ofthe range of molecules that exhibit molecular weightpeaks. In this article we have demonstrated this foralkanes.
This work on EI-M5 of alkanes in 5MB is part ofour continuous research on mass spectrometry insupersonic molecular beams. Our goal is to create anew mass spectrometric approach that will combinefast GC-M5, extended temperature limits, increasedcapability for handling thermolabile molecules, enhanced molecular weight peak height with the abovementioned features, possible selective ionization andincreased sensitivity with hyperthermal surface ionization, possible coupling of gas chromatographs withany column and gas load, and the ability to use verylarge amounts of splitless injections.
References
1. Arnirav, A.; Danon, A. u.s. Patent 5055677,1991; Israel Patent90970, 1989.
2. Arnirav, A.; Danon, A. lilt. [. Mass Spec/rolll. 1011 Processes1990, 79, 107.
3. Arnirav, A. Org, Mass Spec/rolll. 1991, 26, 1.4. Milne, T. A.; Greene, F. T. Adl'. High Telllp. Chem. 1969, 2,
107.5. Pauly, H.; Toennies, j. P. Methods or Experimental Phveics;
Bederson, B.; Fite, W., Eds.; Academic: New York, 1968; Vol.7A, P 227.
6. Anderson, j. B.; Andres, R. P.; Fenn. j. B. Adz'. Chcm, Phus,1965, 10, 275.
7. Kolodney, E.; Amirav, A. Chem. Phys. 1983, 82, 269; Danon,A.; Amirav, A. Rev. Sci. lnsirum. 1987, 58, 1724.
8. Scoles, G., Ed. Atomic and Moleclliar Beam Methods; OxfordUniversity Press: New York, 1988; Vol. 1.
9. Boesl, V.; Walter, K.; Schlag, E. W. lilt. [, Mass Spec/rom. 1011
Processes 1986, 71, 309; Croterneyer, J.; Boesl, V.; WaIter, K.;Schlag, W. [, Am. Chem. Soc. 1986, 108, 4233; Org. MassSpec/rom. 1986, 21,596.
10. Hager, J. W.; Wallace, S. C Anal. Chem. 1988, 60, 5.11. Lubman, D. M.; Jordan, R. M. Rev. Sci. IIIS/ntlll. 1986, 56,
1299; Ternbreull, R; Lubman, D. M. Anal. Chelll. 1986, 58,1299; Ternbreull, R.; Sin, C H.; Pang, H. M.; Lubrnan, D. M.Optics News 1986, 12, 16.
12. Leutuyler, S.; Even, U. Chelll. Phys. Lelf. 1981, 81, 578.13. Meijer, G.; de Vries, M. S.; Hunziker, H. E.; Wendt, H. R
Appl. Phys. 1990, B51, 395; '. Chem. Phys. 1990, 92, 7625.14. Lubman, D. M., Ed. Laser and Mass Spcctrometrv; Oxford
University Press: New York, 1990; Lubman, D. M. MassSpccirom. Rel'. 1988, 7, 535.
15. Arnirav, A. ,. PIIYs. Chcm. 1990, 94, 5200.16. Ballenthin, J. 0.; Nier, A. O. Rc-v. Sci. Ills/ntm. 1981, 52, 1016.17. Nier, A. 0.; Potter, W. E.; Kayser, D. C; Fenslak, R G.
Geophys. Res. Lett. 1974, 1,197.18. Dagan, S.; Arnirav, A. lilt. [, Mass Spec/rom. 1011 Processes,
1994, 133, 187.19. Darton, A.; Amirav, A. Rel'. Sci. 1115/1'1111I. 1987, 58, 1724.20. Brink, G. O. Rev. Sci. lnsirum, 1966, 37, 857 and 1626.21. Danon, A.; Arnirav, A. '. Phys. Chem. 1989, 9.3, 5549.22. Danon, A.; Amirav, A. lilt. [, Mass Spectrom. 1011 Processes
1990, 96, 1390.23. Danon, A.; Amirav, A.; Silberstein, J.; Salman, Y.; Levine,
R. D. '. Phys. Chcm. 1989, 93, 49.24. Lifshitz, C; Tieman, 1. O. '. Chelll. Phys. 1973, 59, 6143;
Lifshitz, C; Weiss, M. Chem. PIIYs. Lett. 1972, 15, 266.25. Chupka, W. A. '. Chem. Phys. 1971, 54, 1936.26. Cenuit, W.; Nibbering, N. M. M. lilt. [. Mass Spectrolll. 1011
Processes 1986, 73, 61.27. Amorebieta, V. T.; Colussi, A. J. '. Phys. Chem.1982, 86,3056;
1986, 90, 1225; Choll. Phys. Lett. 1982, 89, 193.28. Milne, T. A.; Beachey, J. E.; Greene, F. T. ,. Chem. Phys. 1972,
56,3007.29. Remberg, G.; Remberg, E.; Spiteller, M.; Spiteller, G. Org.
Mass Spectrom. 1968, 1, 87.30. Budzikiewics, H. In Biochemical Applications of Mass Spectrom
etry; Waller, G. R, Ed.; Wiley-Interscience: New York, 1972; p251.
31. Davis, R.; Frearson, M. Mass Spectrometry; Wiley: New York,1987.
32. Maccoll, A. Org. Mass Spec/rom. 1986, 21, 601.33. Turecek, F. Org. Mass Spec/rom. 1991, 26, 1074.