effects of error in radius of curvature on the corneal power measurement before and after laser...
TRANSCRIPT
TECHNICAL NOTE
Effects of error in radius of curvature on the corneal powermeasurement before and after laser refractive surgery formyopiaYongji Liu1, Yan Wang2, Zhaoqi Wang1 and Tong Zuo2
1Institute of Modern Optics, Key Laboratory of Optical Information Science and Technology, Ministry of Education, Nankai University, Tianjin, and2Refractive Surgery Center, Tianjin Eye Hospital, Tianjin, China
Citation information: Liu Y, Wang Y, Wang Z & Zuo T. Effects of error in radius of curvature on the corneal power measurement before and
after laser refractive surgery for myopia. Ophthalmic Physiol Opt 2012, 32, 355–361. doi: 10.1111/j.1475-1313.2012.00921.x
Keywords: corneal power, intraocular lens,
measurement error, refractive surgery
Correspondence: Yongji Liu
E-mail address: [email protected]
Received: 21 December 2011; Accepted: 14
May 2012
Abstract
Purpose: To investigate the sources of error in corneal power measurement
before and after corneal refractive surgery for myopia.
Methods: The study comprised 28 eyes of six males and eight females with a
mean age of 26 (range 18–39 years). The radius of curvature of anterior and
posterior corneal surface, Q-Values of anterior and posterior corneal surface
and corneal central thickness were measured by rotating Scheimpflug imaging
(Pentacam). The true net power Fg, back vertex power Fv, and keratometric
power SimK, were calculated respectively at the apex and at a paracentral area
on the 3 mm ring.
Results: For virgin eyes, the overestimation (0.53 ± 0.11 D) of the corneal
power by using a keratometric index of 1.3375 was balanced by the underesti-
mation ()0.21 ± 0.09 D) of the corneal power by the error in the radius of
curvature, resulting in a relatively small corneal power error with a mean value
of 0.33 ± 0.11 D. With the Q-value changing from )0.09 to )0.41, the percent-
age balanced by the error in radius of curvature increased from 16% to 73%.
However, for eyes after laser refractive surgery, the radius of curvature error
lead to an overestimation (0.54 ± 0.16 D) of the corneal power and the kerato-
metric index of 1.3375 again overestimated (1.59 ± 0.26 D) the corneal power,
resulting in a large measurement error with a mean value of 2.12 ± 0.40 D.
With the Q-value changing from 0.35 to 1.89, the percentage added by the
error in radius of curvature increased from 14% to 32%.
Conclusions: For virgin eyes, the overestimation of the corneal power by using a
keratometric index of 1.3375 is balanced by the underestimation of the corneal
power by the error in the radius of curvature, resulting in a relatively small cor-
neal power error. However, for eyes after laser refractive surgery, the flatter
anterior corneal surface means that the use of a keratometric index of 1.3375
significantly overestimates the corneal power and the radius of curvature error
now adds to this overestimation and results in a large measurement error.
Introduction
Due to the difficulty of measuring the posterior corneal
radius of curvature, the corneal power is conventionally
determined by the anterior corneal radius with the for-
mula (n)1)/r, where n is taken as 1.3375 for most kera-
tometers and corneal topography systems and r is the
corneal anterior radius. There are two prior assumptions
behind this formula: The first assumption is a constant
posterior/anterior curvature ratio, resulting in the anterior
and posterior corneal surfaces being considered as one
surface with a fictitious single refractive index of 1.3375.
The second assumption is that the corneal anterior sur-
face is considered to be spherical with a zero Q-value
Ophthalmic & Physiological Optics ISSN 0275-5408
Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists 355
(which is a factor that describes the shape of the surface)
over the paracentral zone of 2 mm or 3 mm, resulting in
a uniform corneal power in this zone.
It is well known that after laser refractive surgery the
posterior/anterior curvature ratio is not a constant and
the anterior corneal surface is changed from prolate to
oblate.1 However the validity of the assumptions made
using keratometers did not draw much attention until
hyperopic refractions2,3 were observed in patients who
had phacoemulsification with intraocular lens (IOL)
implantation after previously having laser refractive sur-
gery for myopia. After this observation was made, the
first assumption has been been extensively studied4 and
the general conclusion is that the index of 1.3375 tends to
overestimate the corneal power, which helps to explain
the hyperopic surprise after IOL exchange surgery. Since
the role of the second assumption on the corneal mea-
surement for eyes with previous laser refractive surgery
has received little attention, the focus of this research is
try to address this issue.
Another issue in question is how accurate both
assumptions are for virgin eyes. Studies5,6 have shown
that the standard keratometric index of 1.3375 tends to
overestimate the true corneal power for virgin eyes. In
addition, studies7 have reported that the normal anterior
corneal surface is prolate with a negative Q-value, so that
the second assumption is also questionable. However,
both assumptions work surprisingly well for virgin eyes in
terms of IOL power calculation which has been generally
explained8,9 by the formulae for calculating IOL power
having built-in compensations for the bias introduced by
the conventional keratometric index of 1.3375.
Methods
Subjects
Both eyes of 14 patients were randomly selected from the
Refractive surgery Center of the Tianjin Eye Hospital of
the Tanjin Medical University. The study followed the
tenets of the Declaration of Helsinki. Informed consent
was obtained from all subjects after the nature and possible
consequences of the study had been explained. Those who
had corneal or retinal disease or had previous ocular sur-
gery were excluded. Subjects who had worn contact lens at
any time 2 weeks before the examination were excluded as
well. All subjects had a full ophthalmic examination.
Corneal topography was performed with the rotating
Scheimppflug imaging system (Pentacam; http://www.
pentacam.com) preoperatively and about 1 month after
laser refractive surgery. Values recorded from the topog-
raphy included mean radius of corneal anterior surface
(R1), mean radius of posterior surface (R2) and Q-values
for anterior and posterior corneal surface. According to
the Pentacam instruction manual, the mean radius of cur-
vature represents the average of the radius of curvature in
the central 3.0 mm zone. A diameter of 8.0 mm (corre-
sponding to 30�) was chosen to obtain the corneal shape
factor Q-value because in most eyes the ablation zone is
about 8.0 mm. The laser in situ keratomileusis (LASIK)
was performed under topical anesthesia, with the Moria
M2 microkeratome (http://www.moria-surgical.com) being
used to create a superior hinged 110-lm flap measuring
9.0 mm in diameter. The flap was superiorly reflected,
and the stromal bed was ablated using the VISX STAR S4
excimer laser system (http://www.visx.com).
Mathematical method to calculate measurement errors in
corneal power
We suppose the anterior corneal surface is a rotationally
symmetric conicoid surface and for simplicity, the corneal
surface is represented in the x–z plane by the following
formula:
y2 þ ð1þ QÞz2 � 2zR ¼ 0 ð1Þ
where the origin is chosen at the anterior corneal surface
apex, y is the vertical meridian, z is the axis of revolution,
R is the apical radius and Q is the aspheric parameter
that specifies the type of conicoid. Clinical data10 showed
that the cornea anterior surface could be represented by
an ellipse, corresponding to )1 < Q < 0, representing a
prolate, and Q > 0, representing an oblate.
With the measured anterior and posterior corneal radii
and the cornea central thickness, a cornea model was
constructed. In the present paper, we used a corneal
model to study the errors in measuring eyes with and
without laser refractive surgery using keratometry. Since
the keratometry gives the axial radius11 of points on the
3.0 mm ring, then the axial radius of the cornea model
was selected and was calculated using the formula12:
Ra ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiR� Qy2
pð2Þ
In addition, the back vertex power of the cornea at the
posterior surface was calculated by:
Fv ¼Fg
1� dnc� Fa
ð3Þ
where the back vertex power of the cornea, the true net
power obtained from Gaussian formula, the actual power
of anterior corneal surface, the corneal central thickness
and the index of cornea (1.376) are represented by Fv, Fg,
Fa, d, nc respectively.
Using the cornea model, the measurement errors
caused by using an index of 1.3375 (corresponding to the
first assumption) and by error in radius of curvature
Effects of error in radius for the corneal power measurement Y Liu et al.
356 Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists
(corresponding to the second assumption) are calculated
by the following steps:
1 At the apex, the true net power Fg, back vertex power
Fv by Equation (4), and keratometric power obtained by
(1.3375)1)/R, SimKapex, were calculated respectively.
2 At the paracentral area on the 3.0 mm ring, kerato-
metric power obtained by (1.3375)1)/Ra SimKmanual was
calculated, where SimKmanaul corresponds to the reading
from the keratometery.
3 DFapex¼SimKapex� Fv;DSimK¼ SimKmanual � SimKapex;
DF ¼SimKmanual�Fvwere calculated respectively. DFapex =
SimKapex ) F reflects the error solely caused by using an
index of 1.3375. DSimK = SimKmanual ) SimKapex reflects
the error solely caused by the error in the radius of cur-
vature. DF = SimKmanual ) Fv reflects the corneal error
caused by using both an index of 1.3375 and the error in
corneal radius measurement, because 1.3375 and the
radius of curvature on the 3 mm ring are used to get
SimKmanual.
The above calculations were conducted in Matlab (Ver-
sion 7.0, http://www.mathworks.com) and the statistics
analysis was conducted with SPSS (Version 11.5; http://
www.ibm.com/SPSS_statistics).
Results
Subjects
The study comprised of 28 eyes from six males and eight
females with a mean age of 26 (range 18–39 years). The
mean manifest refraction spherical equivalent before sur-
gery was )7.12 ± 1.72 D (range )4.00 to )10.50 D). After
surgery, the mean manifest refraction spherical equivalent
was )0.26 ± 0.71 D (range )1.69 to 1.44 D).
The mean radius of the anterior cornea surface was sig-
nificantly changed from 7.75 to 9.02 mm after laser
refractive surgery, with the radius of the posterior cornea
surface almost unchanged. The average Q-value of the
anterior corneal surface was prolate with a negative
Q-value of )0.25 (range )0.09 to )0.41). After surgery,
the average anterior corneal surface was oblate with a
positive Q-value of 1.05 (range 0.35–1.89), whereas the
posterior cornea surface remained prolate.
The effects of error in the radius for virgin eyes
In Figure 1, for virgin eyes, the error due to using a cor-
neal refractive index of 1.3375 is shown by the solid tri-
angles, the error in the measured radius of curvature is
shown by the solid squares and the resultant error is
shown by the open circles. The corneal power measure-
ment error caused by the error in the measured radius
of curvature increased negatively with the Q-value. With
the Q-value changing from )0.10 to )0.40, the error
changed from )0.08 to )0.35 D with a mean value of
)0.21 ± 0.09 D.
As shown in Figure 1, the error due to using a corneal
refractive index of 1.3375 overestimates the true corneal
power. The average overestimation is 0.53 ± 0.11 D, rang-
ing from 0.33 to 0.74 D. However, the total corneal
power measurement error caused by using an index of
1.3375 and the error in radius of curvature, was relatively
small with a mean value of 0.33 ± 0.11 D, ranging from
0.12 to 0.53 D.
The effects of the error in the radius for eyes with laser
refractive surgery
For the eyes after laser refractive surgery, the corneal
power measurement error due to using an index of
1.3375, the error in radius of curvature and the resultant
total corneal power measurement error are shown in
Figure 2. In contrast with the results for virgin eyes, the
error in the radius of curvature lead to an overestimation
of the corneal power for eyes after laser refractive surgery.
The overestimation increased as the Q-value increased.
With the Q-value ranging from 0.35 to 1.89, the overesti-
mation increased from 0.2 to 0.78 D with a mean value
of 0.54 ± 0.16 D. The corneal power measurement error
caused solely by using an index of 1.3375 shows an over-
estimation of the corneal power with a mean value of
1.59 ± 0.26 D (range 1.04 to 2.03 D). However, the mag-
nitude of the overestimation increased statistically signifi-
cantly (t27 = 32.41 p < 0.001) in the laser refractive
Figure 1. For eyes before refractive surgery, with prolate corneas,
the error due to using a corneal refractive index of 1.3375 is shown
by the solid triangles and has an average of +0.53 D. The error in the
measured radius of curvature is shown by the solid squares. It
increases negatively with Q-value, helps neutralize the error caused by
using the 1.3375 refractive index, and has an average of )0.21 D.
The resultant error is shown by the open circles and has an average
of +0.33 D.
Y Liu et al. Effects of error in radius for the corneal power measurement
Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists 357
surgery eyes compared to that for virgin eyes, which had
a mean value of 0.53 ± 0.11 D (range 0.33 to 0.74 D).
Discussion
Because the physical meaning of keratometric power
obtained from the keratometric index of 1.3375 is of vital
importance to the discussion, it is first explained. The
keratometric index was proposed13 based on the Gull-
strand eye model, in which the radius of the anterior cor-
neal surface, the radius of the posterior corneal surface
and the thickness of the central cornea are 7.7, 6.8 and
0.5 mm respectively.
From Gaussian Optics, the refractive power of the cor-
nea, which is considered as a thick lens, is calculated by:
Fg ¼ Fa þ Fp �d
nc� Fa � Fp ð4Þ
where Fg is the corneal power, Fa is the power of the
anterior surface, Fp is the power of the posterior surface,
d is the thickness of the central cornea, nc is the index of
the cornea. With the data of Gullstrand eye model, we
obtained:
Fa ¼nc � 1
Ra¼ ð1:376� 1Þ � 1000
7:7¼ 48:83D ð4:1Þ
Fp¼ na � nc
Rp¼ ð1:336� 1:376Þ � 1000
6:8¼ �5:88D ð4:2Þ
d
nc� Fa � Fp ¼
0:5
1:376� 1000� 48:83� ð�5:88Þ ¼ �0:1D
ð4:3Þ
Fg¼ 48:83þ ð�5:88Þ � ð�0:1Þ ¼ 43:05D ð4:4Þ
The cornea back vertex power at the cornea posterior
surface is:
Fv ¼Fg
1� dnc� nc�1
Ra
¼ 43:05
1� 0:51:376� 1:376�1
7:7
¼ 43:83D ð5Þ
If we apply a refractive index to the front surface that
would produce the same result as the thick lens calcula-
tion, the index can be calculated from Fg by:
ng ¼ Fg � Ra þ 1 ¼ 43:05� 7:7� 10�3 þ 1 ¼ 1:3315
ð6Þ
If we use the cornea back vertex power at the posterior
corneal surface Fv to get the index, we obtained:
nv ¼ Fv � Ra þ 1 ¼ 43:83� 7:7� 10�3 þ 1 ¼ 1:3375
ð7Þ
From the above calculations, it is quite clear that if
1.3315 is chosen as the constant index to calculate the
corneal power with the radius of anterior corneal sur-
face by (n)1)/R, the resulting corneal power can be
regarded as the power from Gaussian Optics, which is
usually called the true net corneal power. If 1.3375 is
used to calculate corneal power, the result reflects the
back vertex power at the posterior corneal surface. Con-
sequently, the keratometric corneal power obtained from
most clinical instruments, including keratometers or
topography systems can be considered as the cornea
back vertex power at the posterior corneal surface.
Therefore, the corneal power used to calculate IOL
power for virgin eyes is also the back vertex power of
the cornea, rather than the true net corneal power (This
can explain why true net power cannot be entered into
IOL power formulae even for virgin eyes6). Drawing
form this conclusion, in terms of the IOL power calcu-
lation, the error in the back vertex power measurement
before and after laser refractive surgery is our major
focus of the discussion.
Figure 2. For eyes after refractive surgery, with oblate corneas, the
error due to using a corneal refractive index of 1.3375 is shown by
the solid triangles and has an average of +1.59 D. The error in the
measured radius of curvature is shown by the solid squares. It
increases positively with Q-value, adds to the error caused by using
the 1.3375 refractive index, and has an average of +0.54 D. The
resultant error is shown by the open circles and has an average of
+2.12 D.
Effects of error in radius for the corneal power measurement Y Liu et al.
358 Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists
The effects of error in the radius for virgin eyes
Because the corneal power is actually measured on a ring
with 3 mm diameter and the corneal anterior surface is
not spherical, there is a measurement error, which stems
from the error in radius of curvature. The effects of the
error in radius of curvature on the corneal power mea-
surement were studied in this paper. As expected for a
prolate corneal anterior surface, the error in the radius
of curvature underestimates the apical corneal power for
virgin eyes.
It is well known5,6 that the index of 1.3375 overesti-
mates the true net power. The back vertex power, which
is the main focus of our study, is higher than the true net
power. Therefore, our first question is whether an index
of 1.3375 continues to overestimate the back vertex
power? Our results have shown that the apical keratometric
power is higher than the apical back vertex power, which
confirms that the keratometric index of 1.3375 continues
to show an overestimation of the corneal back vertex
power.
An important finding of the present study is that the
overestimation by using the index of 1.3375 is balanced
by the adius of curvature error, resulting in a relatively
small corneal power measurement error for virgin eyes.
To get a more detailed look, we calculated the percentage
of the error due to using an index of 1.3375 that is bal-
anced by the error in measuring the radius of curvature
and the results are shown in Figure 3. When the Q-value
is around )0.4, almost 70% of the error introduced by
using an index of 1.3375 was balanced by the negative
error in the radius of curvature. Around a Q-value of
)0.25, which is the mean Q-value for normal eyes7,
about 30–40% of the error caused by using the index of
1.3375 is balanced. The balance leads to a relatively small
corneal power measurement error with a mean value of
0.33 ± 0.11 D.
An interesting fact is that though the two assumptions
are questionable, both work surprisingly well for virgin
eyes in terms of the IOL power calculation. The phenom-
enon is explained by a built-in compensation8,9 in the
formulae for calculation of the IOL power. If only the
error due to using the keratometric index of 1.3375 is
taken into account, an error of 0.53 D in keratometry
caused by using 1.3375, translates into about 0.78 D
refractive error14 after an IOL implantation. As it is
reported that the mean absolute refractive error after
modern optimized phacoemulsification is approximately
0.40 to 0.50 D,15,16 which is lower than 0.78 D, it is rea-
sonable to use the built-in compensation to explain the
discrepancy. However if the error in the radius of curva-
ture is also taken into account, the total error in keratom-
etry declines to 0.33 D, corresponding to a refractive
error of 0.40 D after an IOL implantation, which is
within the range of clinical results of 0.40 to 0.50 D.
Two reasons may lead to the discrepancy between our
results and the results in previous studies. First, in the
previous studies, only the effect of keratometric index of
1.3375 on the corneal power calculation was considered.
However, in our study, besides the corneal power mea-
surement error caused by 1.3375, the corneal power mea-
surement error caused by error in the radius of curvature
is also considered causing the neutralization between the
error in keratometric index and the error in the radius of
curvature to be observed. Secondly, it is the back vertex
power, rather than the vertex power at the corneal ante-
rior surface that was used in our study.
The effects of the error in the radius of eyes with laser
refractive surgery
After laser refractive surgery, a distinct change is that the
corneal power measurement error caused by the error in
the radius of curvature is positive, which is the result of
the oblated anterior surface. In addition, the absolute
value of measurement error caused by error in radius of
curvature increases in comparison with that for virgin
eyes (0.54 D vs )0.21 D).
After laser refractive surgery, the corneal power mea-
surement error caused by using an index of 1.3375
remains positive. However, the magnitude is significantly
increased from a mean value of 0.53 D for virgin eyes to
a mean value of 1.59 D for eyes with laser refractive sur-
gery, because the ratio between the anterior and the pos-
terior radius of curvature has increased. The calculated
percentage of the corneal power error caused by the error
in the radius of curvature as a function of the Q-value in
eyes after laser refractive surgery is shown in Figure 4. ForFigure 3. For virgin eyes, the relationship between the contribution
of the error in the radius of curvature and Q-value.
Y Liu et al. Effects of error in radius for the corneal power measurement
Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists 359
eyes with laser refractive surgery, the error caused by the
index of 1.3375 plays a dominant role in the corneal
power measurement error. However, the additional effect
of the radius of curvature error should not be neglected
especially for eyes with large Q-values.
There are few studies on this issue. A recent study by
Savini et al.17 found no statistically significant difference
between the corneal curvature measurement in both the
paracentral area and over the entrance pupil for eyes with
excimer laser surgery. It concluded that the radius error
had limited relevance in the context of corneal power
measurements after myopic excimer laser surgery, at least
when a large optical zone (6.5 mm) is used. The discrep-
ancy between our results and Savini’s results can be
explained as follows: In our study, the difference between
the corneal power measurement at the apex and at a
position on the 3 mm ring is used, which is more suitable
to reflect the effects of the error in the radius in compari-
son to the mean central value used in Savini’s study. In
another study, Roberts18 has shown that the error in the
radius of curvature plays an important role in the corneal
power measurement even for virgin eyes. In her study, it
is the instantaneous radius that is used to evaluate the
effects of the error in the radius of curvature. In our
study, it is the axial radius that is used to evaluate the
effects of the error in the radius of curvature on corneal
power measurement, as the reading from the keratometry
is the axial corneal power.10
One limitation in our study is that both eyes from sub-
jects were used in the analyses. It is known that two eyes
of one subject are typically similar so that measurements
such as corneal power are correlated. Therefore, in terms
of statistical analysis, more accurate results will be
obtained if data from one eye of a subject are used.
Another limitation of our study is the relatively small
sample size. However, the main focus of the present study
is to investigate the sources of error in corneal power
measurement before and after corneal refractive surgery,
therefore the tendency of the contributions of different
errors seems to be the same with a larger data base. Cer-
tainly, a larger data base is needed for a better quantifica-
tion of the contributions of different errors.
In summary, for virgin eyes, the overestimation of the
corneal power by using a keratometric index of 1.3375 is
reasonably balanced by the underestimation of the corneal
power by the error in the radius of curvature, resulting in
a relatively small corneal power error. However, for eyes
after laser refractive surgery, the flatter anterior corneal
surface means that the use of a keratometric index of
1.3375 further overestimates the corneal power and the
radius of curvature error now adds to this overestimation
and results in a large measurement error.
Acknowledgements
This research is supported by National Science Founda-
tion of China (NO. 60978068, NO. 11104149), by the
Science and Technology Project of Tianjin city (No.
10ZCKFGX18800) and by the Fundamental Research
Funds for the Central Universities.
References
1. Tuan K & Chernyak D. Corneal asphericity and visual
function after wavefront-guided LASIK.pdf. Optom Vis Sci
2006; 83: 605–610.
2. Seitz B, Langenbucher A, Nguyen NX, Kus MM & Kuchle
M. Underestimation of intraocular lens power for cataract
surgery after myopic photorefractive keratectomy. Oph-
thalmology 1999; 106: 693–702.
3. Gimbel HV, Sun R, Furlong MT, van Westenbrugge JA &
Kassab J. Accuracy and predictability of intraocular lens
power calculation after photorefractive keratectomy. J
Cataract Refr Surg 2000; 26: 1147–1151.
4. Shammas HJ, Shammas MC, Garabet A, Kim JH, Sham-
mas A & Labree L. Correcting the corneal power measure-
ments for intraocular lens power calculations after myopic
laser in situ keratomileusis. Am J Ophthal 2003; 136: 426–
432.
5. Olsen T. Calculation of intraocular lens power: a review.
Acta Ophthalmol Scand 2007; 85: 472–485.
6. Savini G, Barboni P, Carbonelli M & Hoffer KJ. Agree-
ment between pentacam and videokeratography in corneal
power assessment. J Refract Surg 2009; 25: 534–538.
7. Gonzalez-Meijome JM, Villa-Collar C, Montes-Mico R &
Gomes A. Asphericity of the anterior human cornea with
different corneal diameters. J Refract Surg 2007; 33: 465–
473.
Figure 4. For the eyes after laser surgery, the relationship between
the contribution of the error in the radius of curvature and Q-value.
Effects of error in radius for the corneal power measurement Y Liu et al.
360 Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists
8. Tang ML, Li Y, Avila M & Huang D. Measuring total cor-
neal power before and after laser in situ keratomileusis
with high-speed optical coherence tomography. J Cataract
Refract Surg 2006; 32: 1843–1850.
9. Ho J, Tsai C, Tsai R, Kuo L, Tsai I & Liou S. Validity of
the keratometric index: evaluation by the Pentacam rotat-
ing Scheimpflug camera. J Cataract Refract Surg 2008; 34:
137–145.
10. Anera R, Jime¢nez J, Barco LG, Bermu¢dez J & Hita E.
Changes in corneal asphericity after laser in situ keratomi-
leusis. J Cataract Refract Surg 2003; 29: 762–768.
11. Mandel RB. Corneal power correction factor for photo-
refractive keratectomy. J Refract Corneal Surg 1994; 10:
125–128.
12. Salmon THD. Comparison of elevation, curvature, and
power descriptors for corneal topographic mapping.
Optom Vis Sci 1995; 72: 800–808.
13. Cheng ACK, Rao SSK, Lau S, Wong A & Lam DSC. Com-
parison of techniques for corneal power assessment after
myopic LASIK without the use of preoperative data. J
Refract Surg 2008; 24: 539–543.
14. Liu YJ, Wang ZQ & Mu GG. Effects of measurement
errors on refractive outcomes for pseudophakic eye based
on eye model. Optik 2010; 121: 1347–1354.
15. Norrby S, Lydahl E, Koranyi G et al. Clinical application
of the lens haptic plane concept with transformed axial
lengths. J Cataract Refract Surg 2005; 31: 1338–1344.
16. Feiz V, Mannis MJ, Garcia-Ferrer F et al. Intraocular lens
power calculation after laser in situ keratomileusis for
myopia and hyperopia – a standardized approach. Cornea
2001; 20: 792–797.
17. Savini G, Carbonelli M, Barboni P & Hoffer KJ. Clinical
relevance of radius of curvature error in corneal power
measurements after excimer laser surgery. J Cataract
Refract Surg 2010; 36: 82–86.
18. Roberts C. Characterization of the inherent error in a spheri-
cally-biased corneal topography system in mapping a radially
aspheric surface. J Cataract Refract Surg 1994; 10: 103–111.
Y Liu et al. Effects of error in radius for the corneal power measurement
Ophthalmic & Physiological Optics 32 (2012) 355–361 ª 2012 The College of Optometrists 361