The Ocular Response Analyzer (ORA) is a new instrument that measures the corneal biomechanical response (corneal hysteresis, CH). It measures the effect of viscous and elastic qualities with a calibrated air pulse in vivo. The ORA measures Intraocular Pressure (Goldmann). Corneal hysteresis is an indication of viscous damping in the cornea, reflecting the capacity of corneal tissue to absorb and dissipate energy. It is the capacity of tissue to recover its original shape after external force is applied.

TRANSCRIPT

Page 1: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

Corneal HysteresisRole of Corneal Biomechanics

in Refractive Surgery

Luis Rodriguez, MDAnny Villegas, MDClínica de Córnea

Centro Médico Docente La TrinidadCaracas, Venezuela

Page 2: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

Corneal Biomechanics is the science that studies the equilibrium and deformation of tissues submitted to any force. This science includes the following extrinsecal factors: Intraocular Pressure, Cilliar Muscle, Extraocular Muscle, Atmospheric Pressure and Eyelids.

Intrinsecal factors of Corneal Biomechanics include: Central Pachymetry, Viscosity, Elasticity, Hydration and Regional Pachymetry.

Page 3: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

The Ocular Response Analyzer (ORA) is a new instrument that measures the corneal biomechanical response (corneal hysteresis,

CH). It measures the effect of viscous and elastic qualities with a calibrated air pulse in vivo.

The ORA measures Intraocular Pressure (Goldmann).

Corneal hysteresis is an indication of viscous damping in the cornea, reflecting the capacity of corneal tissue to absorb and dissipate energy.

It is the capacity of tissue to recover its original shape after external force is applied.

Page 4: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

The Corneal Resistance Factor, also derived from corneal hysteresis, is an indicator of the overall resistance of the cornea.

Corneal-compensated IOP is a pressure measurement that utilizes the new information provided by the corneal hysteresis measurement to provide an IOP that is less affected by corneal properties.

The objective of this study is to discuss the Ocular Response Analyzer readings of corneal biomechanics in refractive surgery and corneal pathology.

This study began in July 2006 and ended in January 2007.

Page 5: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

Methods

ORA was used to obtain four measurements in each patient, and the mean of these four readings was used in the analysis according to the manufacturer’s instructions.

The study included 244 eyes from the Corneal Department of the Centro Medico Docente La Trinidad divided into the following nine categories:Normal cornea, Radial Keratotomy, PostLasik Myopia, PostLasik Ectasia, Keratoconus, Post Intracorneal Segments (Intacs®), Penetrating Keratoplasty, Fuchs’ Dystrophy and Graft rejection.

Page 6: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

244 Eyes

Results

10 Post-LASIK Ectasia 3,42 – 7,43

20 Normal cornea 7,96 – 11,57

70 Keratoconus 5,74 – 7,71

34 Post-LASIK Myopia 7,24 – 9,82

70 Post-Intracorneal Segments (INTACS®) 7,09 – 9,97

Ranges of Corneal Hysteresis (mmHg)

10 Radial Keratotomy 3,50 – 5,75

20 Penetrating Keratoplasty 7,74 – 11,71

5 Fuch’s Distrophy 6,25 – 7,71

5 Graft Rejection 5,35 – 7,40

Page 7: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

Frequency of Corneal Hysteresis

6.95 6.47

8.93 8.62 8.24

0

1

2

3

4

5

6

7

8

9

10

Nor

mal

Cor

nea

Pos

t - L

AS

IK

Pos

t -IN

TAC

S

Ker

atoc

onus

Pos

t -LA

SIK

Ect

asia

8.247.73 7.71

6.35.42

0123456789

Normal CorneaPost-LASIKPost-INTACSKeratoconusPost-LASIK Ectasia

Frequency of Corneal Resistance Factor (CRF)

Page 8: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

02468

1012141618

Normal Cornea Keratoconus Post-LASIK Post- LASIK Ectasia

Post-INTACS®

IOPg IOP cc

14,86

16,36

14,49

9.02

14,23

11,54 11,22

14,73

11,64

15,05

Frequency of IOPg and IOPcc values

Page 9: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

QuickTime™ and aCinepak decompressor

are needed to see this picture.

Page 10: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

Keratoconus

Lasik

Ectasia Post Lasik

Page 11: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

Discussion

Patients with -6.00D (equal to or less), Hysteresis of more than 11mmHg, and normal topography are good candidates for LASIK.

Patients with -6.00D or less, suspicious topography and Hysteresis between 8mmHg and 10mmHg are good candidates for PRK.

Patients with -6.00D or more, suspicious topography and Hysteresis between 7mmHg and 10mmHg are good candidates for Phakic IOL.

Patients, with Keratoconus and Hysteresis of more than 7mmHg are good candidates for intracorneal segments.

Patients with Keratoconus and Hysteresis of equal to or less than 5mmHg are candidates for corneal transplant.

When Hysteresis is less than 7mmHg PostLasik, patients have Ectasia.Hysteresis is normal in Corneal Transplants, but if the patient has a transplant rejection Hysteresis decreases significantly.

Page 12: Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas,…

• Biomechanical studies must be analyzed individually

• Ectasia is a known risk of LASIK and PRK

• Hysteresis measures corneal biomechanical properties

• Patients who are at risk of developing post-LASIK ectasia should be identified by analyzing Hysteresis

• Corneal Hysteresis can guide the physician in the choice of treatment.

• ORA helps us understand the IOP in patients with corneal pathologies and surgeries.